Low kv Multispectral Imaging in a Field Emission SEM

Similar documents
7. advanced SEM. Latest generation of SEM SEM

Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts

View of ΣIGMA TM (Ref. 1)

Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope

The Basics of Scanning Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging

Name: Due: September 21 st Physics 7230 Laboratory 3: High Resolution SEM Imaging

Scanning Electron Microscopy Primer

Preface Light Microscopy X-ray Diffraction Methods

CSCI 4974 / 6974 Hardware Reverse Engineering. Lecture 8: Microscopy and Imaging

Physics 441/2: Transmission Electron Microscope

Lenses and Apertures of A TEM

THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER. Insert Nickname Here. Operating Instructions

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts

Microscopy and Nanoindentation. Combining Orientation Imaging. to investigate localized. deformation behaviour. Felix Reinauer

Fundamentals of Scanning Electron Microscopy

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

Electron Microscopy SEM and TEM

ORIENTATION CHARACTERISTICS OF THE MICROSTRUCTURE OF MATERIALS

The Focused Ion Beam Scanning Electron Microscope: A tool for sample preparation, two and three dimensional imaging. Jacob R.

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Coating Thickness and Composition Analysis by Micro-EDXRF

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM

Scanning Electron Microscopy: an overview on application and perspective

Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

h e l p s y o u C O N T R O L

LIFE SCIENCE I TECHNICAL BULLETIN ISSUE N 11 /JULY 2008

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications

Looking through the fish-eye the Electron Ronchigram. Duncan T.L. Alexander CIME seminar May 24, 2012

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

X-ray thin-film measurement techniques

LASER ENGRAVING REFLECTIVE METALS TO CREATE SCANNER READABLE BARCODES Paper P516

Rodenstock Photo Optics

Introduction to microstructure

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Lectures about XRF (X-Ray Fluorescence)

NATIONAL NETWORK OF ELECTRON MICROSCOPY RNME. NETWORK MANAGEMENT MODEL a ARTICULATION AND GENERAL OPERATION. (English translation draft)

Plastic Film Texture Measurement With 3D Profilometry

Coating Technology: Evaporation Vs Sputtering

Mass production, R&D Failure analysis. Fault site pin-pointing (EM, OBIRCH, FIB, etc. ) Bottleneck Physical science analysis (SEM, TEM, Auger, etc.

COMPARISON OF TEXTURE IN COPPER AND ALUMINUM THIN FILMS DETERMINED BY XRD AND EBSD *

Usage of AFM, SEM and TEM for the research of carbon nanotubes

Electron Microprobe Analysis X-ray spectrometry:

Energy Dispersive Spectroscopy on the SEM: A Primer

Demonstration of sub-4 nm nanoimprint lithography using a template fabricated by helium ion beam lithography

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

Using the PDF for material identification using elemental data. from XRF and SEM EDS.

Scanning Electron Microscopy tools for material characterization

Scanning He + Ion Beam Microscopy and Metrology. David C Joy University of Tennessee, and Oak Ridge National Laboratory

Phase Characterization of TiO 2 Powder by XRD and TEM

Keywords: Planar waveguides, sol-gel technology, transmission electron microscopy

INTRODUCTION TO THE XL30-FEG SEM. 1.1 The mouse The monitor. Figure THE USER INTERFACE

CREOL, College of Optics & Photonics, University of Central Florida

Near-field scanning optical microscopy (SNOM)

Zeiss 780 Training Notes

DOE Solar Energy Technologies Program Peer Review. Denver, Colorado April 17-19, 2007

EDXRF of Used Automotive Catalytic Converters

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Realization of a UV fisheye hyperspectral camera

Lapping and Polishing Basics

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

Introduction to Energy Dispersive X-ray Spectrometry (EDS)

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

WOOD WEAR TESTING USING TRIBOMETER

DualBeam Solutions for Electrical Nanoprobing

SEMTech Solutions. Leaders in Refurbished SEMs. SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM

Introduction to X-Ray Powder Diffraction Data Analysis

Scanning Electron Microscopy Services for Pharmaceutical Manufacturers

CHARACTERIZATION OF POLYMERS BY TMA. W.J. Sichina, National Marketing Manager

ICC Recommendations for Color Measurement

Blue Laser Technology Applied to the Microtrac Unified Scatter Technique for Full- Range Particle Size Measurement.

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

Handbook on the Ultrasonic Examination. Austenitic Welds

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. IV Femtosecond Measurements Combined With Near-Field Optical Microscopy - Artyom A.

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

Institute s brochure. Microstructure Analysis, Metallography and Mechanical Testing of Materials. Institute of Materials Research

3D EDX MICROANALYSIS IN A FIB/SEM:

Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture

Displays. Cathode Ray Tube. Semiconductor Elements. Basic applications. Oscilloscope TV Old monitors. 2009, Associate Professor PhD. T.

5. Scanning Near-Field Optical Microscopy 5.1. Resolution of conventional optical microscopy

Silicon Drift Detector Product Brochure Update 2013

The promise of ultrasonic phased arrays and the role of modeling in specifying systems

Quality Control on Aerospace Components Using Handheld X-ray Fluorescence (XRF)

Nanotribology of Hard Thin Film Coatings: A Case Study Using the G200 Nanoindenter

Advanced Research Raman System Raman Spectroscopy Systems

Fiber Optics: Fiber Basics

Raman spectroscopy Lecture

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

Measuring the Point Spread Function of a Fluorescence Microscope

X-ray diffraction techniques for thin films

Transcription:

Low kv Multispectral Imaging in a Field Emission SEM N. Erdman, N. Kikuchi, A. Laudate and V. Robertson JEOL USA Inc., Peabody, MA 01960 enable unmatched imaging and analytical performance, including the Aperture Angle Control Lens (ACL), the new patented LABe detector (low-angle backscatter detector) and the Gentle Beam mode (application of a specimen bias) that allows operation down to 100 volts (see Figure 2). Successful imaging and characterization of nanomaterials, composites and biological specimens require unsurpassed performance in the low and ultra-low kv range in FEG-SEM for both imaging and analytical modes. The state-of-the-art FEG-SEM should show superb resolution with various detectors at a wide range of accelerating voltages and have the ability to perform fast and reliable microanalysis at low kv using high probe currents, without loss of spatial resolution and with an ease of operation. The new JSM-7600F Schottky FEG-SEM from JEOL combines two proven technologies an electron column with semi-in lens detectors and an in-the-lens Schottky field emission gun - to deliver ultrahigh resolution combined with wide range of probe currents for all applications (from 1pA to more than 200nA). It delivers unsurpassed imaging performance at ultra low kv, with 1.5 nm resolution at 1 kv and 5 nm at 0.1 kv. Figure 1. Au on C specimen imaged with JSM-7600F at 100V accelerating voltage. Figure 2. General schematic of the detector system in the JSM-7600F. In particular, we focus on new analytical performance (EDS, WDS and CL) as well as backscatter imaging data acquired at low and ultra low kv. Vector Filtered BSE Imaging Traditionally BSE imaging required high or moderately high kv due to smaller electron yield in BSE, detector sensitivity and bandwidth at lower probe current, low energy electron sensitivity and a single (typically solid state) detector. Even with all of these limitations BSE detectors have been very useful at low to medium magnifications, especially for highlighting compositional or topographic differences in the samples [1]. Recent developments by JEOL in detector and SEM technology have allowed significant progress towards BSE imaging at low kvs with high resolution. JSM-7600F incorporates a new and unique LABe backscatter detector that is based on this patented JEOL technology. This paper highlights some of the unique features of the new JEOL JSM-7600F SEM that

Figure 3. LABe images of an uncoated ceramic eutectic. (a) 0.2 kv BSE image showing enhanced surface detail, (b) 5 kv BSE image showing more traditional compositional and crystallographic contrast. The combination of the geometry of the LABe detector, specimen position (Working Distance) with respect to the detector as well as accelerating voltage and application of specimen bias effectively allow Vector Filtered BSE imaging (Figure 3). The LABe detector can serve as a conventional BSE detector under normal operating conditions i.e. moderate kv and analytical WD. In this configuration it is a high angle BSE detector yielding Z contrast and channeling contrast. The LABe detector can also collect only the low angle BSEs which due to the angular dependence convey high spatial resolution. Low angle BSE s escape only from the top few nm although they are generated throughout the entire interaction volume of the specimen. By lowering the accelerating voltage down to few kvs or even few hundred volts the beam/specimen interaction volume is also significantly reduced, yielding true surface information as well as crystallographic contrast which is similar in appearance and information, but fundamentally different in its generation from channeling contrast. Figure 4 shows an image of nano-grained Au imaged at 0.5 kv. The image shows twinning (orientation contrast) in Au crystal with very high spatial resolution compared to a regular BSE image. Figure 4. Nano-Au grain imaged at 0.5 kv with the LABe detector. The sample was prepared using JEOL Cross Section polisher. Applying variable specimen bias via the Gentle Beam (GB) function in combination with the LABe detector offers a remarkably new perspective for imaging nano-materials. An example of the combination of the GB and the LABe for low kv BSE imaging is shown in Figure 5. The 1 kv image clearly shows enhanced surface information (including surface contamination) when compared to the 2 kv image. The images taken with no specimen bias show typical BSE compositional and crystallographic information. An increase in specimen bias (from 0 volts to 2 kv) somewhat suppresses the more traditional compositional contrast and shows more enhanced surface detail as well as additional topographic information. Therefore, choosing a combination of specimen bias and kv will dictate the type of information obtained with the LABe detector.

insignificant. The best analytical data comes from the smallest alpha angle. Therefore, when the ACL is optimized for image resolution, the resulting high current image (large alpha angle) has a somewhat hazy background but shows great resolution. However, when the ACL is optimized for analytical work the ultimate resolution is slightly decreased, yet the analytical signal is no longer affected by the beam tailing, resulting in smaller analytical signal delocalization (Figure 6). Figure 5. An uncoated ceramic eutectic imaged with variations of kv and specimen bias using the LABe detector. Optimization of the Alpha Angle for Analysis One of the unique features of the JSM-7600F is the patented Aperture Angle Control Lens (ACL) that automatically optimizes for both high resolution imaging at the lower probe currents and high spatial resolution X-ray analysis at high probe current with a seamless transition between the two, essential for rapid analysis and optimized image quality. The ability to increase the probe current for fast microanalysis, while still maintaining a small spot size and small volume of excitation for high resolution, has been the holy grail of microanalysis in SEM. This is particularly true for low kv microanalysis. The ACL works by taking into account effects of all aberrations (such as spherical aberration and diffraction limitations) on spot size and optimizing the alpha angle accordingly in an automatic fashion. When the SEM is optimized for the smallest spot size (largest alpha angle) there is some beam tailing that produces X-rays from areas not in the spot. For low beam current applications this is Figure 6. ACL optimization at high beam currents for (a) image resolution, and (b) analytical resolution. Au on C images at 15 kv, 20kX and 198 na. The automatic optimization of the ACL maintains high resolution imaging at a wide range of probe currents, from few picoamps to hundreds of nanoamps, as illustrated in Figure 7. The figure shows images of Au on C taken using 5 kv with various probe currents at the EDS working distance. The resolution at 5 kv is 2.4 nm at 50 pa, 4.5 nm at 5nA, and 12 nm at 100 na. This beam resolution allows very fast acquisition (using high beam current) of EDS or WDS data at low kv with high spatial resolution.

Certain specimens, especially those containing low Z materials, require low kv EDS to correctly identify elemental distribution. Additional advantage of high beam current/small spot size is the ability to collect EDS maps very fast (in a few minutes) even at low accelerating voltages. Therefore, beam sensitive specimens remain damage-free during the analysis. Figure 7. Au on C resolution images at the EDS WD with various probe currents ranging from 50 pa to 100 na. The resolution at 5 kv is 2.4 nm at 50 pa, 4.5 nm at 5nA, and 12 nm at 100 na. Implementation of the ACL function allows collection of very fast EDS maps at high spatial resolution on bulk samples using low kvs. One such example is shown in Figure 8. The map shows an EDS map of Sphene (yellow), Apatite (blue) in a magnesium aluminosilicate matrix (red) with sub-100 nm resolution. The map was collected at 8 kv for 20 min using beam current of 20 na. Figure 8. EDS map of Sphene CaTi(Si0 4 )(O,OH,F) (yellow), Apatite (blue) in a magnesium aluminosilicate matrix (red) with sub-100nm resolution. The map was collected at 8 kv for 20 min using beam current of 20 na. Figure 9. SE image and corresponding EDS maps of graphene layers on a Ni substrate. The maps were collected at 2 kv for 2 min using a beam current of 10 na. Figure 9 shows EDS maps of graphene (C) layers on a Ni substrate. Because the graphene layers are only few Angstroms thick it is necessary to minimize the interaction volume as much as possible to eliminate X-Rays from the bulk when conducting the EDS analysis in order to correctly identify the distribution of graphene on the Ni substrate. Additional consideration was given to minimize the analysis time in order to determine the graphene distribution unambiguously without building up of any possible Carbon contamination on the surface which would obscure the analysis. Thus, the EDS maps were collected for 2 min at 2 kv and a beam current of 10 na. The high beam current used for this analysis ensured adequate data collection for compositional interpretation even with a very short acquisition time. The ability to deliver high beam current also impacts WDS and CL (cathodoluminescence)

analyses where large beam currents are needed for sufficient signal-to-noise collection. Figure 10. Comparison of WDS maps of Miake Island Ore geological thin section collected using 15 kv and 5 kv (only 2 representative maps shown, O-Kα and Fe-Lα). Scale bar = 2.5 microns. highly charging specimens need to be examined it is crucial to be able to work at low kv while still delivering a large probe current into a small spot. These limitations are a thing of the past with the new JSM-7600F design. Using JSM-7600F we can collect the CL signal from specimens using accelerating voltages as low as 0.5 kv (the maximum beam current at 0.5 kv is ~11 na, which is more than sufficient for CL). Figure 11 shows CL images of an uncoated diamond specimen collected using 2 kv and 0.5 kv. While CL data at 2 kv clearly has better resolution, it is quite remarkable that the CL signal can be collected at 0.5 kv in cases where the region of interest may be a thin surface film or layer. In WDS, the detector has a small solid angle for detection, therefore high accelerating voltage and the associated high beam currents are usually employed to detect enough of the signal to efficiently obtain good elemental data and maps. Lower accelerating voltages are desirable when thinner or more surface-oriented features are to be investigated. Less signal is generated, normally requiring longer mapping times unless the beam current is increased, resulting in poorer spatial resolution in the maps. In the JSM-7600F, the ability of the ACL to keep the spot size small when the beam current is high results in efficient WDS mapping with high spatial resolution even at lower kvs. The comparison of maps in Figure 9 illustrates the results that are possible with ACL corrected, high beam current operation for WDS applications. Cathodoluminescence detection in the SEM has been historically associated with high kv, high beam current microanalysis, because of the weak CL signal, and the spot size limitations at low voltages. At high kvs, the CL signal which comes from the same volume of excitation as X-Rays has poor spatial resolution and has often been quite poor due to insufficient probe current in the SEM. When high spatial resolution CL is needed or when Figure 11. SEM and corresponding CL images from diamond collected using a Gatan Chroma CL. Conclusions JEOL s JSM-7600F delivers the ultimate in a Schottky FEG-SEM. In addition to providing the

state of the art in ultra low kv imaging that is expected in today s high end FEG-SEMs, JEOL s patented ACL technology enables the user to automatically move from high resolution imaging at low probe currents to fast high spatial resolution microanalysis. JSM-7600F can perform very fast microanalysis (EDS, WDS or CL) at low kv and high spatial resolution. The LABe detector offers unique insight on materials through Vector Filtered BSE imaging. This detector also delivers unsurpassed compositional and/or topographic contrast with high spatial resolution at low kv that is required in many nanomaterials applications. The JEOL JSM-7600F is a highly versatile Schottky FEG-SEM capable of delivering the highest probe currents in the industry for fast microanalysis with superior spatial resolution without sacrificing the ability to yield ultrahigh resolution imaging whether at high or ultra low kvs. Acknowledgements 1. See for example: J. Goldstein et al, Scanning Electron Microscopy and X-ray Microanalysis, Third Edition, 2003, New York: Plenum 2. All EDS data were collected using a ThermoFisher System 7 with 30 mm 2 SDD. 3. The authors would like to thank SIFCO for providing the sample in Figures 3 and 5. 3. The authors would like to thank Prof. J. Kong and Dr. A. Reina from MIT, Cambridge, MA for providing the sample in Figure 9. 4. The authors would like to thank Dr. D.C. Bell from Harvard University for providing the sample in Figure 11.