Achieving Peak Performance From Your Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite

Similar documents
Tuning & Mass Calibration

Micromass LCT User s Guide

Appendix 5 Overview of requirements in English

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

Electrospray Ion Trap Mass Spectrometry. Introduction

PREVENTIVE MAINTENANCE OF FIBER OPTIC CABLES AND OPTICS

EQuan MAX Getting Connected Field Service Guide

TOF FUNDAMENTALS TUTORIAL

Background Information

18-270mm F/ Di II VC PZD for Canon, Nikon (Model B008) mm F/ Di II PZD for Sony (Model B008)

AMD Analysis & Technology AG

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

Wi-Fi handheld microscope for use with the Apple iphone, ipod touch, or ipad.*

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

The Determination of an Equilibrium Constant

CHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS

Registry Tuner. Software Manual

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

74 Chapter CO 2 Cleaning

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

13C NMR Spectroscopy

Electric Kettle. Model: SK-1717 INSTRUCTIONS MANUAL

Analyst 1.6 Software. Getting Started Guide

Lab 9: The Acousto-Optic Effect

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17)

COMPACT LABEL APPLIER

BATTERY CHARGING GUIDELINES FOR 6-VOLT DEEP CYCLE BATTERIES

WSD130 SOLDERING STATION

VCR Ion Beam Sputter Coater

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

STANDARD CLEANING AND CALIBRATION PROCEDURE FOR TGA-50(H) AND TGA-51(H)

The front end of the receiver performs the frequency translation, channel selection and amplification of the signal.

MAINTENANCE & TROUBLESHOOTING

The Parts of the System

Physics 221 Experiment 5: Magnetic Fields

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MAINTENANCE & TROUBLESHOOTING

Single Channel Loop Detector

How To Use Gc-Ms

Expectations for GC-MS Lab

Pro Logic Version Diagnostics Manual

Pesticide Analysis by Mass Spectrometry

Coating Technology: Evaporation Vs Sputtering

SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16)

Operator Quick Guide EC SENSOR

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Analysis of Liquid Samples on the Agilent GC-MS

Thermistor Thermometer MODEL NO

Greenline Industrial Monitors IMD-C Series. User Manual

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE

MassLynx 4.1 Getting Started Guide

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

ETL listed for installations within 5 ft. (1.5M) of outer edge of water Rev E Page 1 of 10

23 The Thermal Conductivity Detector

Thermo Top - Troubleshooting Tree

Troubleshooting with ABS Service Data Kit

Manual Analysis Software AFD 1201

Frequently Asked Questions: Cisco Jabber 9.x for Android

TROUBLESHOOTING FLOW CHART

Set Top Box User Guide. Global Reach with a Local Touch

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions

Features: Mounting the Optic

PR 50 Frymaster Universal Holding Cabinet (UHC)

DASH KIT INSTALLATION

VISM CQB Scope Series

IMPORTANT SAFETY INSTRUCTIONS WARNING READ AND SAVE THESE OPERATING AND SAFETY INSTRUCTIONS BEFORE USING THIS HEATER.

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

E/M Experiment: Electrons in a Magnetic Field.

Experiment 5. Lasers and laser mode structure

EZ DUPE DVD/CD Duplicator

The photoionization detector (PID) utilizes ultraviolet

Welcome to the presentation Quartz Watch Knowledge for Professionals

ExStik TM ph Waterproof Meters

4-Channel Thermometer / Datalogger

Digital Photo Picture Frame. Operation Manual

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Tube Control Measurement, Sorting Modular System for Glass Tube

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

PART 1 - INTRODUCTION...

Care and use guide. Contents. a handy solution made easy. page

Thermal Conductivity Detector

Reasons for reissue are in Section 6, REVISION SUMMARY.

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Setting up a Quantitative Analysis MS ChemStation

ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST

DAS202Tools v1.0.0 for DAS202 Operating Manual

Periodic Verification System Model 775PVS

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Operating Manual Color Changing Fountains

Easy Machining Center Setup

Care and use guide. Contents. a handy solution made easy. page

MGL Avionics CAN bus interface for Trig Avionics TT21/TT22 transponders

Time-of-Flight Mass Spectrometry

SCION tc DASH APPLIQUE Preparation

Transcription:

Achieving Peak Performance From Your Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite Ronald Pong, Ph.D.. Julie Horner, Ph.D., and Martin Zeller, Ph.D. Product Support Engineering and Ion Trap Marketing

What We Aim to Achieve This guide will help you: Achieve peak performance with your Velos Pro, Orbitrap Velos Pro, or Orbitrap Elite. Learn more about how to operate and care for your mass spectrometer. Communicate more effectively with us if there is a problem that cannot be solved using this troubleshooting guide. This guide is: 1. A tool to help you tune, calibrate, diagnose problems, and clean the ion optics of your mass spectrometer. 2. Designed to be used with both the Achieving Peak Performance Quick Reference Card and the LTQ Series Hardware Manual. 3. The content of this document will be reviewed by your Field Service Engineer upon completion of installation of your instrument, and again during the first preventative maintenance visit. 2

What You Need To Do First 1. Pump Down Before Using Your Mass Spectrometer Your mass spectrometer performs best when pumped down for a minimum of 15 hours. Using the mass spectrometer before a pump down time of at least 15 hours may result in incorrect transfer lenses calibration and may age the electron multipliers more rapidly After 2 hours of pumping you may view the spectrum as a quick check to see if the system is functioning correctly. Be sure to calibrate the multipole RF and main RF frequency before viewing the spectrum for the first time. 3

What You Need To Do First 2. Install LTQ 2.7 SP1 or highest instrument control version Using outdated versions of instrument control software may result in reduced features and/or improper calibrations. Always verify your instrument control software is up to date. You may determine the latest version for your instrument from your FSE or sales representative. Call your sales representative or field service engineer to obtain the latest version of your instrument s control software. 4

LTQ 2.7 SP1 and Sweep Gas NOTE: After installing LTQ 2.7 SP1, you will notice that the sweep gas is ON and set to 5 units by default when the system is in standby. This is a safety feature to minimize the particulate matter that enters the mass spectrometer as this may result in a charge nucleation site on critical ion optics. Thermo Fisher Scientific recommends you leave your instrument in these default conditions. To disable this safety feature (not recommended): 1. Close all Xcalibur and Foundation occurrences. 2. Open Instrument Configuration. 3. Select the Velos Pro based device. 4. Select Ion Source. 5. Uncheck the box titled Enable sweep gas when in standby. 5

Information in this Guide: Section I: Ion Trap Achieving peak performance on the Velos Pro ion trap as a stand alone instrument and as the front end for the Orbitrap Velos Pro and Orbitrap Elite hybrid instruments. Section II: Orbitrap Achieving peak performance on the orbitrap portion of the Orbitrap Velos Pro and Orbitrap Elite hybrid instruments. Note: Most of this information is also pertinent to the LTQ Velos and LTQ Orbitrap Velos mass spectrometers. 6

Section I: Ion Trap Velos Pro Ion Optics and Ion Trap

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 6. Communicating with Thermo 8

1. Achieving and Maintaining a Stable Spray Prior to performing any experiment, tuning, calibration, or diagnostic procedure, establishing a stable and consistent spray is essential. This is true for all source types, ESI, HESI, and NSI. Failure to maintain a stable spray will compromise the data quality or result in a poor tune, failed calibration or diagnostic result. The goal is a spray with less than 15% RSD over at least 100 scans. 9

1. Achieving and Maintaining a Stable Spray A stable spray is the first step in a successful calibration! Run the API stability evaluation: Signal stability < 15 % RSD indicates a stable spray Parameters Affecting Spray Stability 1. Spray voltage 2. Sheath gas, auxillary gas 3. Syringe pump speed 4. Sprayer position 5. HESI II heater temperature 10

1. Achieving and Maintaining a Stable Spray Set the syringe pump flow rate to 3 ml/min. Adjust the following parameters while running the API Stability Diagnostic Probe position: A or B Sheath: 1-8 Aux: 0-2 Sweep gas: 0 Spray voltage: 3.0-4.5kV Probe temperature (HESI): 30-50 C Minimize the %RSD shown in the graph; this must be at most 15%. 11

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 12

2. Calibrating the Velos Pro Key Starting Requirements: 1. Make sure that your instrument has been continuously under vacuum for at least 15 hours since the last vent. Failure to do so may cause faulty calibrations. 2. Install LTQ 2.7 SP1 or highest version available. Many of the key diagnostics in this document are new to LTQ 2.7 SP1 and are not available in earlier versions. 3. Ensure the spray is stable as described in 1. Achieving and Maintaining a Stable Spray. 13

2. Calibrating the Velos Pro Calibrating the Velos Pro, Orbitrap Velos Pro, or Orbitrap Elite enables peak performance and ensures mass accuracy. Calibration is always done using either the HESI or the ESI probe. Never use NSI to calibrate the instrument. Use the appropriate Calibration Mixture (Calmix). Never use expired or contaminated Calmix. These may compromise calibration results. Calibration Check is a way to periodically determine if the instrument calibration is within specification. If a Calibration Check Fails, run the actual Calibration procedure. If an actual Calibration repeatedly fails, contact your local Field Service Representative for further assistance. Two Calibrations that are critical to maintaining expected sensitivity are: a.) Electron Multiplier Gain (both positive & negative) b.) Transfer Lenses (both positive & negative) 14

2.a) Electron Multiplier Gain Calibration The gain will change more rapidly for new electron multipliers (either for new instrument installations or replacements in older systems) versus older ones. Consequently, the voltage necessary to maintain the correct gain will change rapidly until a voltage plateau is reached. This effect is amplified in fast scanning instruments (e.g. LTQ (Orbitrap) Velos, (Orbitrap) Velos Pro and Orbitrap Elite) due to the higher ion currents. The graph on the next slide shows an example of the rate of voltage change for two new electron multipliers in an LTQ Velos instrument. Typically a plateau is reached at operating voltages of approximately 900 to 1100 (ITT multipliers used in non-pro instruments) or 1700 to1800 (SGE multipliers used in Velos Pro, Orbitrap Velos Pro and Orbitrap Elite instruments). Once the plateau is reached, the gain of the multipliers will change more slowly, and therefore less frequent calibration is required. The time to reach the plateau will vary with instrument conditions and use electron multipliers in heavily used instruments will plateau faster than instruments that are used only occasionally. 15

2.a) Electron Multiplier Gain Calibration When the multiplier voltage is between 750 and 1100 V (ITT Multipliers) or 1200 and 1800 V (SGE Multipliers), run the Multiplier Calibration Check every 3 to 5 days. If the calibration check fails, the actual semiautomatic electron multiplier gain calibration should be run. Once the rate of change in voltage decreases, the multiplier check and multiplier calibration may be carried out less frequently, for example, every 2 to 4 weeks. Both negative and positive electron multiplier checks and calibrations should be run, even if you normally run only positive mode. All other calibrations are also recommended to be performed monthly. The time required to reach the plateau will vary with use. 16

2.b) Transfer Lenses Calibration Run the Transfer Lenses Calibration Check once per month, or if trap charging/contamination is suspected. Typically, all transfers should have efficiencies greater than 90% after calibration. Persistently failing calibration (<70%) may indicate charging/contamination in the trap. Call your local Field Service Representative if you suspect trap charging. Low Pressure Trap Transfer Lens High Pressure Trap 17

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 6. Communicating with Thermo 18

3. Tuning the Velos Pro for Optimal Performance Key Starting Requirements: 1. Ensure your instrument has been continuously under vacuum for at least 15 hours since the last vent. 2. Install LTQ 2.7 SP1 or highest version available. Many of the key diagnostics in this document are new to LTQ 2.7 SP1 and are not available in earlier versions. 3. Ensure the spray is stable as described in 1. Achieving and Maintaining a Stable Spray. 4. Verify all calibrations are up to date as described in 2. Calibrating the Velos Pro. 19

3. Tuning the Velos Pro for Optimal Performance During the normal use of an instrument, sample/solvent residue will accumulate on the components of the ion optics. Over time, the accumulated residue can alter the tuning characteristics of the ion optics components. Accumulated residue can result in sub-optimum tuning parameters. This can be observed as a reduction in sensitivity. This reduction in sensitivity can happen more rapidly if non-optimized tune values are used, in addition to poor sample clean up and instrument usage. In order to obtain high levels of sensitivity over a longer period of time, the ion optics must have an appropriate voltage gradient to resist changes to tuning conditions due to usage. 20

3. Tuning the Velos Pro for Optimal Performance MP00-MP0 Gradient The most critical components for tuning the ion optics is the difference between the MP0 offset and the MP00 offset (the Gradient ). The difference between these offsets should be maximized and not less than -5.5 volts. An MP0/MP00 difference of -7 volts is the default setting. This difference provides a high enough voltage gradient so that the ions will have sufficient energy to overcome any local charging due to normal rates of ion optic contamination. MP00 MP0 MP1 Front Lens The next most critical component is the Front Lens. The Front Lens voltage depends upon the MP0 offset. For the default MP0 offset range (-8 to -12) the Front Lens voltage will optimize at a value slightly more positive than the MP0 voltage. A more positive front lens potential decelerates the ions allowing more efficient trapping of the ions as they enter the high pressure trap. -7 Volts Ions Fall Down Hill! 21

3. Tuning the Velos Pro for Optimal Performance Front Lens Tuning Always start with the default tune file for the source you re using. The ion optics voltages should match the values in the table below. Autotune is generally not needed. Use semi-automatic tune to set the Front Lens voltage for the MP0 voltage setting. Optics Element Voltage MP00-2 L0-3 MP0-9 L1-15 MP1-20 Front Lens -8.5 Multipole RF 800Vp-p 22

TIC TIC 3. Tuning the Velos Pro for Optimal Performance Front Lens Tuning 4000000 3500000 Front Lens Tuning Curve for m/z 195 MP0 Offset: -9V 2000000 1800000 Front Lens Tuning Curve for m/z 1522 MP0 Offset: -9V 1600000 3000000 1400000 2500000 1200000 2000000 1000000 1500000 800000 1000000 600000 500000 0-30 -28-26 -24-22 -20-18 -16-14 -12-10 -8-6 -4-2 0 Lens Potential (V) 400000 200000 0-30 -28-26 -24-22 -20-18 -16-14 -12-10 -8-6 -4-2 0 Lens Potential (V) 23

3. Tuning the Velos Pro for Optimal Performance Multipole rf Voltage In order to effectively focus and transport the ions, especially the higher m/z ions and at the higher MP0 offsets, a greater Multipole RF Amplitude becomes necessary. Increasing the Multipole RF Amplitude from 600 Vp-p to 800 Vp-p will aid in focusing the ions and increase the robustness of the instrument. Default tuning parameters: 24

3. Tuning the Velos Pro for Optimal Performance Always use the released version of the instrument control software, currently LTQ 2.7 SP1 Earlier versions of instrument control software, for example LTQ 2.6.0 SP3, do not enforce the minimum MP00/MP0 difference. If your instrument uses these earlier versions of instrument control software, upgrade your instrument control software to LTQ 2.7 SP1. 25

3. Tuning the Velos Pro for Optimal Performance Tune parameters are linked to Ion Source Type Tune the instrument with the ion source you will use for your experiment When a new ion source type is installed, the software automatically loads the last saved tune file for that source type. For example, when NSI is installed after tuning and calibrating the Velos Pro with HESI, the last saved NSI tune file is loaded. The title in the Instrument Control window (e.g., Tune Plus) updates only when a tune file is saved, so you may be unaware of the change. The tune parameters established using the HESI will NOT be transferred to the NSI tune file, rather the tune parameters will be overwritten by the parameters in the NSI tune file. Tune parameters determined using a different ion probe type must be entered manually and saved after the desired probe has been installed. 26

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite: LTQ 2.7 SP1 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 6. Communicating with Thermo 27

4. Troubleshooting and Diagnostics When following the standard operating procedure described in the previous slides, your Velos Pro system should provide you with routine, high performance results. However, you may occasionally be required to troubleshoot your system in order to maintain optimum performance. To facilitate the troubleshooting process, a number of system diagnostics have been created in LTQ 2.7 SP1. These new diagnostics are described in the remainder of this document. 28

4. Troubleshooting and Diagnostics: Contents a) System Diagnostics b) Loss of Sensitivity/Signal Diagnostics: a series of 10 new diagnostics and procedures to help determine the cause of reduced sensitivity: 1. Injection Time Check 2. API Stability Evaluation 3. TIC Oscillation Check 4. Multiplier Gain Calibration Check 5. Transfer Lenses Calibration Check 6. Multipole Gradient Evaluation 7. Source Optics Flight Time Evaluation 8. Multipole M0 Flight Time Evaluation 9. Ion Optics Charging Evaluation 10. Ejection and Multiplier Gain Ratio Evaluation c) Diagnostics Resolution See the LTQ Series Hardware Manual for troubleshooting other problems with your instrument. 29

4. Troubleshooting and Diagnostics Key Starting Requirements: 1. Make sure that your instrument has been continuously under vacuum for at least 15 hours since the last vent. 2. Install LTQ 2.7 SP1 or highest version available. Many of the key diagnostics in this document are new to LTQ 2.7 SP1 and are not available in earlier versions. 3. Ensure the spray is stable as described in 1. Achieving and Maintaining a Stable Spray. 4. Verify all calibrations are up to date as described in 2. Calibrating the Velos Pro. 5. Ensure the tune parameters are set as described in 3: Tuning the Velos Pro for Optimal Performance. 30

4.a) System Diagnostics Click on Diagnostics in the main menu of the LTQ Tune page. 31

4.a) System Electronics Diagnostics 1. Click on Tests, select All 2. Check the All tests checkbox. 3. Click on Start to begin diagnostics. 2 3 1 Test results appear here. These system diagnostics report a Pass/Fail result. 32

4.a) System Diagnostics 1. Click on Tests, select Scan device 2. Check the All Readbacks checkbox. 3. Click on Start to begin diagnostics. 2 3 1 33

4.a) System Diagnostics If any of the diagnostics tests fail, please record the data so that a service engineer can better help get your problem resolved. Specifically, please: 1. Note and report any failures. 2. Copy and paste all text in the Testing box into a Word document. 3. Take screen captures of or right click and copy any System Diagnostic Graphs. Paste these results into a Word document as well. Any System failures must be addressed by qualified repair personnel. Contact your local Field Service Representative to report any failures. So that your FSE can arrive prepared, please email the Word document containing the System Diagnostic Results to your local Field Service Engineer prior to setting the appointment. System Log files are stored in C:\Thermo\Instruments\LTQ\system\logs. Include the 3 most recent files in the email to your FSE. 34

4.b) Loss of Sensitivity/Signal Diagnostics Keys to Success in Diagnosing Sensitivity Issues: Run the following list of checks (next slide) in the polarity that you noted the loss of sensitivity. Use a full syringe of Calmix. Do not switch the polarity until the final diagnostic check has been run. All calibration checks, tests and diagnostics are automated unless otherwise noted. Details of each are given in subsequent slides. First run calibration checks or tests 1 5 (see next slide) and remedy any failed checks or tests. Then, if the problem persists, continue with diagnostics 6 10 (see next slide) in order without stopping or skipping steps. Note: Diagnostics 6-10 should be run in order, since item 9 can discharge a system and temporarily mask charging symptoms! Run all diagnostics (6-10) before attempting to fix any issues, compiling data in a word document throughout. 35

Run Checks & Diagnostics In This Order Item LTQ Instrument Control Version Check 1 All Injection Time Check (manual) 2 All API Stability Check 3 All TIC Oscillation Check (manual) 4 All Multiplier Gain Calibration Check 5 All Transfer Lenses Calibration Check LTQ Instrument Item Control Version Diagnostic 6 LTQ 2.7 SP1 and higher Multipole Gradient Evaluation 7 LTQ 2.7 SP1 and higher Source Optics Flight Time Evaluation 8 LTQ 2.7 SP1 and higher Multipole M0 Flight Time Evaluation 9 LTQ 2.7 SP1 and higher Ion Optics Charging Evaluation (always run after diagnostic procedures 6-8 are run) 10 LTQ 2.7 SP1 and higher Ejection and Multiplier Gain Ratio 36

1. Injection Time Check Injection Time 37

1. Injection Time Check Injection Time (also called Ion Injection Time, Ion Time) is the time in milliseconds that ions accumulate in the mass analyzer (the ion trap) prior to being scanned out. For this check use: Full MS Target of 3E4 Full Scan MS from 150-2000. For the Velos Pro, the Injection Time is typically less than 0.200ms when infusing Velos Pro calibration mixture ( Calmix ). If the Injection Time > 0.200 ms, do some quick checks of: Spray stability Source operating conditions (voltage, sheath gas, needle position) Multiplier gain calibration Fresh calmix for the Velos/Velos Pro from Pierce Chemical Co. Clean or change the ion transfer tube. Caution: Lower the ion transfer tube temperature to <200ºC and allow the API stack to cool down prior to removing or inserting the ion transfer tube. Inserting or removing the ion transfer tube at a higher temperature may permanently damage it. 38

2. API Stability Check 3. Click Start 1. Highlight System Evaluation 2. Check API stability evaluation checkbox 39

2. API Stability Check This was covered in the Spray Stability Section. The lower the %RSD the better. 15% or less is a good target. 40

3. TIC Oscillation Check 1. Click on the Manual Tune tab. 2. Choose the TIC radio button. 3. Click the Start button. 1 2 3 41

3. TIC Oscillation Check Use the Manual Tune feature in Tune Plus to plot the Total Ion Current while infusing Calmix. 1. Click on the Tuning Fork 42

3. TIC Oscillation Check TIC oscillations (scan to scan changes between an average high level and an average low level, not just an unstable graph) may indicate that the transfer lenses are not calibrated or that the trap may be contaminated. TIC Oscillations between a distinct high level and a low level 43

3. TIC Oscillation Check If the TIC Oscillates Run the Multiplier Gain Check (next section) Run the Transfer Lens Calibration Check Run the Multiplier Gain Calibration, if necessary Run Transfer Lens Calibration If the problem persists, the ion trap may be charging. Take screen captures or make a raw file of your observations to share with your local field service representative. All ion trap work must be done by a qualified Field Service Representative 44

4. Multiplier Gain Check To run the Multiplier Gain (and any other) Calibration Check, click on the calibration button: Click on the Calibration Button 45

4. Multiplier Gain Check 1. Click on the Check tab. 2. Check the Positive (or Negative if the failure was noted in this mode) Ion Electron Multiplier Check checkbox. 3. Click the Start button. 2 1 3 46

4. Multiplier Gain Check If the Multiplier Gain Check Fails Run the Multiplier Gain Calibration. The Multiplier Gain Check, and if needed the Multiplier Gain Calibration, should be run on a weekly basis regardless of loss of signal. See the section on Calibration above. If the Multiplier Gain Calibration fails, make sure the spray is stable. Repeated failures with a stable spray suggests a Field Service call is in order. Be sure to take screen shots of all results, including all of the calibration text, and email these data to your local Field Service Representative. 47

4. Multiplier Gain Check To run the Multiplier Gain Calibration, click on the Calibration Button. 1. Click on the Semi-Automatic tab. 2. Check the Positive (or Negative) Ion Electron Multiplier Gain checkbox. 3. Click the Start button. 2 1 3 48

5. Transfer Lenses Calibration Check 1. Click on the Check tab. 2. Check the Positive (or Negative if the failure was noted in this mode) Transfer Lenses Calibration Check checkbox. 3. Click the Start button. 2 1 3 If the Transfer Lenses Check Fails Run the Transfer Lenses Calibration. If the Transfer Lenses Calibration fails, make sure the spray is stable. Repeated failures with a stable spray suggests trap charging. A Field Service call is in order. Be sure to take screen shots of all results, including all of the calibration text, and email these data to your local Field Service Representative. 49

Run Diagnostics 6-10 In This Order Item LTQ Instrument Control Version Diagnostic 6 LTQ 2.7 SP1 and higher Multipole Gradient Evaluation 7 LTQ 2.7 SP1 and higher Source Optics Flight Time Evaluation 8 LTQ 2.7 SP1 and higher Multipole M0 Flight Time Evaluation 9 LTQ 2.7 SP1 and higher Ion Optics Charging Evaluation 10 LTQ 2.7 SP1 and higher Ejection and Multiplier Gain Ratio Note: Please run these diagnostics one at a time, in the order above, and without stopping, until all are completed. Do not select all diagnostics at once!! This will result in tests being run out of order. Results of each test will overwrite the previous one; please copy and paste the contents of the results window before running the next test. 50

6. Multipole Gradient Diagnostic What Information Does It Provide? Provides information on the Sensitivity of the system as a function of the Multipole Voltage Gradient. Am I giving up sensitivity for long term robustness? The shape of the curve can indicate the contamination level of the Optics. How Does It Work? Sets all optics to Standard Default Settings. Turns AGC off to save time. Changes MP0 Offset, and therefore MP Gradient, to various values. At each value, optimize the front lens voltage for various masses. Curve fit each data set to more confidently determine optimum value. Plot Signal versus Multipole Gradient value for various m/z. Plot Normalized Signal versus Multipole Gradient value for various m/z. Contaminated optics will show loss of signal at lower gradients. 51

6. Multipole Gradient Diagnostic 1. Click on System Evaluation 2. Click on Multipole gradient evaluation 3. Click Start 52

Normalized Area Normalized Area 6. Multipole Gradient Diagnostic Lightly Used Heavily Used Normalized Signal Intensity vs q0-q00 Gradient Normalized Signal Intensity vs q0-q00 Gradient 1.35 1.30 1.25 1.20 1.15 1.10 1.05 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 m/z 195.1 m/z 524.3 m/z 1022.0 m/z 1822.0-13 -12-11 -10-9 -8-7 -6-5 -4-3 -2-1 0 q0-q00 Gradient (V) 1.35 1.30 1.25 1.20 1.15 1.10 1.05 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 m/z 195.1 m/z 524.3 m/z 1022.0 m/z 1822.0-13 -12-11 -10-9 -8-7 -6-5 -4-3 -2-1 0 q0-q00 Gradient (V) At higher MP0-MP00 gradient signal is better maintained At low MP0-MP00 gradient, signal is lost 53

7. Source Optics Flight Time Diagnostic What Information Does It Provide? Helps determine the contamination/charging level of the ion source optics, which includes S-lens, Exit Lens, MP00, L0, and MP0 for LTQ Velos, or MP00,L0, and MP0 for Velos Pro. How Does It Work? Measures the transit time of ions of various m/z through the source optics. Plots the signal intensity for various masses as a function of the time delay after the S-lens RF (for LTQ Velos) or Exit lens (for Velos Pro) is gated to allow ions to travel to the Gate electrode. Charging, and other issues, can cause the flight times to be long, especially at low voltage gradients (low ion energies). (LTQ Velos) Flight Time 54

55 7. Source Optics Flight Time Diagnostic

Normalized Peak Height Normalized Peak Height 7. Source Optics Flight Time Diagnostic Example of a Lightly Used System Example of a Heavily Used System Evaluation of q00/l0/q0 Flight Time - Normalized Evaluation of q00/l0/q0 Flight Time - Normalized peak height 195.09 peak height 524.26 peak height 1221.99 peak height 1521.97 peak height 1821.95 peak height 195.09 peak height 524.26 peak height 1221.99 peak height 1521.97 peak height 1821.95 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 Flight Time (ms) 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 Flight Time (ms) 56

8. Multipole MP0 Flight Time Diagnostic What Information Does It Provide? Helps determine the contamination/charging level of the MP0 device. How Does It Work? Measures the transit time of various mass ions through the MP0 device. Plots the signal intensity for various masses as a function of the time delay after the L0 lens is gated to allow ions to travel through MP0 to the Gate electrode. Charging, and other issues, can cause the flight times to be long, especially at low voltage gradients (low ion energies). Flight Time 57

58 8. Multipole MP0 Flight Time Diagnostic

Normalized Peak Height Normalized Peak Height 8. Multipole MP0 Flight Time Diagnostic Example of a Lightly Used System Example of a Heavily Used System Evaluation of q0 Flight Time - Normalized Evaluation of q0 Flight Time - Normalized peak height 195.09 peak height 524.26 peak height 1221.99 peak height 1521.97 peak height 1821.95 peak height 195.09 peak height 524.26 peak height 1221.99 peak height 1521.97 peak height 1821.95 1.50 1.45 1.40 1.35 1.30 1.25 1.20 1.15 1.10 1.05 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 Delay Time (ms) 1.50 1.45 1.40 1.35 1.30 1.25 1.20 1.15 1.10 1.05 1.00 0.95 0.90 0.85 0.80 0.75 0.70 0.65 0.60 0.55 0.50 0.45 0.40 0.35 0.30 0.25 0.20 0.15 0.10 0.05 0.00 EXAMPLE OF FAILED MP0 FLIGHT TIME DIAGNOSTIC 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 Delay Time (ms) 59

9. Ion Optics Charging Diagnostic What Information Does It Provide? Helps to Identify which Ion Optic Components are Charging Helps Identify which parts may need to be cleaned How Does It Work? Sequentially tests each optical element Plots positive Ion TIC for 30 sec Switches to Negative Polarity and transmits negative ion beam (100 ms Injection times) for 80 seconds up to the ion optical element being tested Switches back to Positive Ion Polarity and Continues Plotting positive Ion TIC for 30 sec If the element is charged, a spike in the intensity will be observed after exposure to negative ions Plot Post/Pre Negative Ion Exposure Intensity Ratios for various mass window. Discharging process is usually very m/z dependant Measures Pos Ion/Neg Ion Flux Ratio, since this can directly effect the results Common reasons for low Neg Ion Flux Never Calibrated Gain and Transfer Lens Never Tuned for Negative Ions Sheath Gas not appropriate for Neg Ions Scans are in Turbo Scan Rate to save time Note that once this diagnostic is run, it can discharge the system. Consequently, running it a second time may not show the same behavior. 60

9. Ion Optics Charging Diagnostic: Example MP0 (+) ions Scan & save 30s DC Offsets (-) (-) (-) (-) (-) (-) (-) ions Reflect & neutralize Pause 80s DC Offsets (+) (+) (-) (-) (-) (-) (+) ions Scan & Save 30s DC Offsets (-) (-) (-) (-) (-) (-) 61

9. Ion Optics Charging Diagnostic Positively charged contaminant During 80 sec pause Reflected (-) beam neutralizes (+) charged contaminant; polarity switching may also dislodge charged lint or particulate contaminants 62

9. Ion Optics Charging Diagnostic Diagnostic Result Graph View Graph 1 Graph 2 63

Signal 9. Ion Optics Charging Diagnostic: Graph 1 Example of a Lightly Used System Ratio of Positive/Negative Ion Flux : 22.80 750000000 Up to Exit Lens Up to MP00 Up to Lens0 Up to MP0 Up to LENS1 Up to MP1 Up to FRONTLENS Up to Center Lens Discharge Time Up to SPLITLENS 700000000 650000000 600000000 550000000 500000000 450000000 400000000 350000000 300000000 250000000 200000000 Ideally, the Positive/Negative Ion Flux ratio should be low, e.g. 0.5 25.0. A higher ratio may result from the tune file or multiplier gain calibration for the opposite polarity being out of date. Be sure to update tune file(s) and multiplier gain calibration for the opposite polarity that you normally run. For example, if you normally run in positive mode, switch to negative mode at least once every three months. Update the negative ion mode tune file by saving it after tuning. 150000000 100000000 50000000 0 0 1 2 3 4 5 6 7 8 9 10 Time(min) 64

Pre/Post Signal Ratio 9. Ion Optics Charging Diagnostic: Graph 2 Example of a Lightly Used System Average Ratio Post/Pre TIC: 1.03 Up to Exit Lens Up to MP00 Up to Lens0 Up to MP0 Up to LENS1 Up to SPLITLENS Up to MP1 Up to FRONTLENS Up to Center Lens Test Boundary 2.5 2.4 2.3 2.2 2.1 2.0 1.9 1.8 1.7 1.6 1.5 1.4 1.3 1.2 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0-0.1 0-2000 Exit Lens 0-2000 MP00 0-2000 Lens 0 0-2000 MP0 0-2000 Lens 1 0-2000 Split Lens 0-2000 MP1 0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000 m/z window (* Lens Element) 0-2000 Gate Lens 0-2000 Centre Lens Failure Threshold (2X) Axis represents Cumulative Mass Range 0-2000 for each Optic 65

Signal Pre/Post Signal Ratio 9. Ion Optics Charging Diagnostic: Exit Lens Charging Example of Heavily Used System with Exit Lens Charging Ratio of Positive/Negative Ion Flux : 20.30 Up to Exit Lens Up to MP00 Up to Lens0 Up to MP0 Up to LENS1 Up to SPLITLENS Up to MP1 Up to FRONTLENS Up to Center Lens Discharge Time Average Ratio Post/Pre TIC: 1.11 Up to Exit Lens Up to MP00 Up to Lens0 Up to MP0 Up to LENS1 Up to SPLITLENS Up to MP1 Up to FRONTLENS Up to Center Lens Test Boundary 190000000 180000000 170000000 160000000 150000000 Instability in Signal 2.3 2.2 2.1 2.0 1.9 1.8 140000000 130000000 120000000 110000000 100000000 90000000 80000000 70000000 60000000 50000000 1.7 1.6 1.5 1.4 1.3 1.2 1.1 1.0 0.9 0.8 0.7 0.6 Above Threshold Indicates Exit Lens Charging 40000000 30000000 20000000 10000000 0 Indicates Exit Lens Charging 0.5 0.4 0.3 0.2 0.1 0.0-0.5 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 Time(min) 0 2000 4000 6000 8000 10000 12000 14000 16000 18000 m/z window (* Lens Element) 66

9. Ion Optics Charging Diagnostic: MP00 Charging Example of Heavily Used System with MP00 Charging Indicates MP00 Charging Instability in Signal Above Threshold Indicates MP00 Charging 67

9. Ion Optics Charging Diagnostic: MP0 Charging Example of Heavily Used System with MP0 Charging Indicates MP0 Charging Instability in Signal Above Threshold Indicates MP0 Charging 68

10. Ejection and Multiplier Gain Ratio Evaluation What Information Does It Provide? Helps determine if Ion trap ejection slots are distorted/contaminated AND if multiplier gains are not matched How Does It Work? Measures signal intensity for various masses on detectors at each side of the trap independently Calculate the ratio of signal and plot as a function of m/z Mismatches in signal are often due to traps with contaminated slots causing asymmetric ejection even at optimum resonance ejection phase. Both Positive and Negative Ions can be tested, depending on current polarity Note: To eliminate gain mismatch effects, this diagnostic should be run immediately following a multiplier gain calibration. Note: This test has been eliminated from the multiplier gain calibration check since, predominantly, this test fails due to trap contamination issues, and multipliers were getting changed erroneously. 69

10. Ejection and Multiplier Gain Ratio Evaluation Ejection and EM Gain Ratio Diagnostic Output Positive Mode 19:02:50: Evaluate Trap Ejection Signal Ratio and Multiplier Gain Match... 19:02:50: Checking Ejection Signal Match 19:02:50: Signal Match Range: 0.60-1.67 19:03:03: Ejection Signal Ratio at m/z 74.06 = 0.909024 19:03:12: Ejection Signal Ratio at m/z 195.09 = 0.952128 19:03:22: Ejection Signal Ratio at m/z 262.65 = 0.997340 19:03:32: Ejection Signal Ratio at m/z 393.28 = 0.910877 19:03:41: Ejection Signal Ratio at m/z 524.26 = 0.879310 19:03:50: Ejection Signal Ratio at m/z 922.01 = 0.856755 19:03:59: Ejection Signal Ratio at m/z 1022.00 = 0.851542 19:04:08: Ejection Signal Ratio at m/z 1221.99 = 0.877723 19:04:17: Ejection Signal Ratio at m/z 1521.97 = 0.892436 19:04:27: Ejection Signal Ratio at m/z 1821.95 = 0.897192 19:04:27: Ejection Signal Match OK at all masses Negative Mode 15:26:07: Evaluate Trap Ejection Signal Ratio and Multiplier Gain Match... 15:26:08: Checking Ejection Signal Match 15:26:08: Signal Match Range: 0.60-1.67 15:26:18: Ejection Signal Ratio at m/z 1080.00 = 1.065100 15:26:27: Ejection Signal Ratio at m/z 1180.00 = 1.216347 15:26:36: Ejection Signal Ratio at m/z 1280.00 = 1.248680 15:26:45: Ejection Signal Ratio at m/z 1380.00 = 1.397896 15:26:54: Ejection Signal Ratio at m/z 1480.00 = 1.482011 15:27:03: Ejection Signal Ratio at m/z 1580.00 = 1.526921 15:27:13: Ejection Signal Ratio at m/z 1680.00 = 1.556537 15:27:22: Ejection Signal Match (1.69) was not OK. Trying again... 15:27:31: Ejection Signal Ratio at m/z 1780.00 = 1.476162 15:27:41: Ejection Signal Match (1.73) was not OK. Trying again... 15:27:50: Ejection Signal Ratio at m/z 1880.00 = 1.610880 15:28:01: Ejection Signal Match (1.76) was not OK. Trying again... 15:28:10: Ejection Signal Match (1.79) was not OK. Trying again... 15:28:20: Ejection Signal Ratio at m/z 1980.00 = 1.548889 15:28:21: Ejection Signal Match OK at all masses Note: Both positive and negative spectra can be used for this diagnostic. 70

TIC Signal Ratio 10. Ejection and Multiplier Gain Ratio Evaluation Top Graph Comparing Signals at Each Detector for m/z 1821.95 Bottom Graph Ejection Signal Ratio vs M/Z 13000000 12000000 11000000 10000000 9000000 8000000 7000000 6000000 5000000 4000000 3000000 2000000 1000000 0-1000000 -2000000-3000000 -4000000-5000000 -6000000 Detector 1 Detector 2 2.0 1.9 1.8 1.7 1.6 1.5 1.4 1.3 1.2 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 2 4 6 8 10 12 14 16 18 20 22 24 26 Scan 200 400 600 800 1000 1200 1400 1600 1800 2000 M/Z 71

4. c) Diagnostic Failures Resolution Based on any failures observed, use the following chart to determine possible causes and resolutions. Table continued on next slide 72

4. c) Diagnostic Failures Resolution Based on any failures observed, use the following chart to determine possible causes and resolutions. 73

4. c) Diagnostic Failures Resolution Based upon the data provided by the diagnostic checks described above, there are several possible steps which may be necessary to restore sensitivity, including: Restoring the default ion optics tune voltages. Calibrating the electron multipliers and/or Transfer Lenses. Cleaning the ion optics. Contacting your local Field Service Representative for assistance. If it is determined that the Ion Optics requires cleaning, the next section briefly describes Ion Optics Cleaning. 74

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite: Introducing LTQ 2.7 SP1 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 6. Communicating with Thermo 75

5. Cleaning the Ion Optics Before removing the API stack or the top cover, be sure that the environment around the instrument is as clean and dust free as possible. After the top cover is removed, it is imperative that the open manifold be covered with a clean sheet of aluminum foil. This will prevent dust and particles from falling out of the air and getting into the manifold causing further charging problems. 76

5. Cleaning the Ion Optics If setting the tune parameters or recalibrating the multipliers do not alleviate the loss of sensitivity, you may need to clean the ion optics. This includes all components on the inner and outer API cages and the top cover. Do not touch the trap itself. If the diagnostic procedures above suggest the problem is in the ion trap itself, contact your local Field Service Representative for further assistance. Do not attempt to service the ion trap yourself! S-Lens MP00 MP0 MP1 Clean only up to MP1, stop at the Trap Front Lens (please do not touch this element). Call Field Service for trap problems. 77

5. Cleaning the Ion Optics: Cleaning Kit The following cleaning tools and optical devices are required to clean the ion optics components according to this procedure. These are available only as a kit, P/N 97455-62051. Please consult appropriate supplier for replenishing individual items. The individual items are not available from Thermo Fisher Scientific. Individual items can be purchased from the suppliers listed below. Item Thermo PN Supplier & Cat. No. Chamois Swabs 00725-01-00028 MCM Electronics 20-1145 Foam Tipped Swabs 00203-99-00031 Fisher 14-960-3L Micromesh Polishing Swabs 00725-01-00027 Micromesh Abrasive Sheets 00725-01-00030 Micro-Surface Finishing Products, Inc SWB-01 Micro-Surface Finishing Products, Inc SHVP-01 2.3X Magna-Visor 00725-01-00029 Adafruit Industries ID: 293 Jeweler s Loupe 00725-01-00032 Illuminated Dual MJ362242L 78

5. Cleaning the Ion Optics: Removing the Optics See the Videos entitled, Top Cover disassembly, Component Cleaning and reassembly, and Accessing the S-lens, Exit Lens, and MP00 for Cleaning, available at http://planetorbitrap.com Be sure not to introduce any scratches or surface abrasions while handling the optics components. Even small scratches can affect performance if they are near enough to the ion flight path. Avoid using tools, such as pliers, that may scratch these components. 79

5. Cleaning the Ion Optics: S-Lens Carefully inspect the S-Lens with magnification for any lint, particulates, and/or sample buildup or coatings. After use, there may be some discoloration on some of the lens elements; this is normal and different from sample build up or coatings. This discoloration, sometimes called ion burn, has not shown to have any effect on instrument performance. If sonication is available, sonicate the S-Lens in either 50:50 methanol:water or a 1% solution of Liquinox in water for 10 15 minutes. If sonication is not available, use a chamois tip swab to clean the S-Lens with a 1% solution of Liquinox in water. Use a 6000 grit Micromesh swab to clean the areas in the lens stack that are not accessible to the chamois tip swab. Rinse the S-Lens thoroughly with water. Rinse the S-Lens with methanol. Dry the S-Lens with a stream of nitrogen gas until it is dry. The source of the nitrogen gas must be oil free. Inspect the S-Lens with magnification for any lint or particulates that may have been introduced by this cleaning procedure. Pay special attention to the orifices and confirm that no lint or particulate is present in the bore of the orifices. Use tweezers or similar tools to remove the lint or particulates. 80

5. Cleaning the Ion Optics: Exit Lens Carefully inspect the Exit Lens with magnification for any lint, particulates, and/or sample buildup or coatings. After use, there may be some discoloration the lens surface; this is normal and different from sample build up or coatings. This discoloration, sometimes called ion burn, has not shown any effect on instrument performance. If sonication is available, sonicate the Exit Lens in either 50:50 methanol:water or a 1% solution of Liquinox in water for 10 15 minutes. If sonication is not available, use a soft toothbrush to clean the Exit Lens with a 1% solution of Liquinox in water. Use rolled 6000 grit Micromesh to clean the bore the Exit Lens orifice. Rinse the Exit Lens thoroughly with water. Rinse the Exit Lens with methanol. Dry the Exit Lens with a stream of nitrogen gas until it is dry. The source of the nitrogen gas must be oil free. Inspect the Exit Lens with magnification for any lint or particulates that may have been introduced by this cleaning procedure. Pay special attention to the orifice and confirm that no lint or particulate is present in the bore of the orifice. Use tweezers or similar tools to remove the lint or particulates. 81

5. Cleaning the Ion Optics: Inter Multipole Lenses Remove the lenses from their mounts on the outer cage and top cover. Carefully inspect all of the ion optic lenses, including Lens L0, Lens L1 and the Gate Lens, with magnification for any lint, particulates, and/or sample buildup or coatings. The farther along the ion path, the less likely you will find sample buildup or coatings. After use, there may be some discoloration the lens surface; this is normal and different from sample build up or coatings. This discoloration, sometimes called ion burn, has not shown any effect on instrument performance. If sonication is available, sonicate the lenses in either 50:50 methanol:water or a 1% solution of Liquinox in water for 10 15 minutes. If sonication is not available, use a soft toothbrush to clean the lenses with a 1% solution of Liquinox in water. Use rolled 6000 grit Micromesh to clean the bore only of the lens orifices. Rinse the lenses thoroughly with water. Rinse the L0 and L1 lenses with methanol. Do not use methanol on the Gate Lens assembly. Dry each lens with a stream of nitrogen gas until it is dry. The source of the nitrogen gas must be oil free. Inspect the lenses with magnification for any lint or particulates that may have been introduced by this cleaning procedure. Pay special attention to the orifice and confirm that no lint or particulate is present in the bore of the orifice. Use tweezers or similar tools to remove the lint or particulates. 82

5. Cleaning the Ion Optics: Quadrupole Ion Guides Remove the quadrupole ion guides from their mounts in the outer cage and top cover. Carefully inspect all of the quadrupoles with magnification for any lint, particulates, and sample buildup or coatings. When inspecting Q0 pay special attention to the incline surface of the lower rod and the decline surface on the upper rod. These two locations are where the sample build up and coatings form the most. See the figure on the next slide for the noted locations. Sonicate the quadrupoles for 10 minutes in a 1% solution of Liquinox in water. Scrub the quadrupoles with a chamois tipped swab soaked in a 1% solution of Liquinox in water. Rinse the quadrupoles thoroughly with water. Sonicate the quadrupoles for 10 minutes in methanol. Scrub the quadrupoles with a chamois tipped swab soaked in methanol. Rinse the quadrupoles thoroughly with methanol. Dry the quadrupoles with a stream of nitrogen gas until it is dry. The source of the nitrogen gas must be oil free. Inspect the quadrupoles with magnification for any lint or particulates that may have been introduced by this cleaning procedure. Confirm that no lint or particulate is present on the inside surface or on the edges of the quadrupole rods. Use tweezers or similar tools to remove the lint or particulates. 83

5. Cleaning the Ion Optics: Quadrupole Ion Guides Likely areas of contamination in MP0 84

Section I Contents 1. Achieving and Maintaining a Stable Spray 2. Calibrating the Velos Pro 3. Tuning the Velos Pro for Optimal Performance 4. Diagnostics and Troubleshooting on the Velos Pro, Orbitrap Velos Pro, and Orbitrap Elite: Introducing LTQ 2.7 SP1 5. Cleaning the Ion Optics on the Velos Pro or the Velos Pro part of the Orbitrap Velos Pro and Orbitrap Elite 6. Communicating with Thermo 85

6. Communicating with Thermo Data describing the problem will help us find you a solution faster. As you run diagnostic tests, please be sure to: Describe the problem fully in a word document. Add screen shots of results and problems into the Word document. Copy all calibration and tune text to the Word document. Collect raw files that demonstrate the problem. Email these items to your FSE along with the following: Model of the instrument (e.g., Orbitrap Velos Pro, Velos, etc) Serial number of the instrument (For Orbitraps, please supply the serial number of both the Orbitrap and the front end instrument (Velos Pro, etc.) When the instrument was installed (how old is the instrument?) When the problem was first observed Length of time the problem has existed 86

Section II: Orbitrap Standard Orbitrap

1. Calibrating Orbitrap Velos Pro and Orbitrap Elite Important calibrations for Orbitrap Velos Pro and Orbitrap Elite: Orbitrap Velos Pro: pagc scaling Mass calibration Orbitrap Elite: pagc scaling Mass calibration Advanced Signal Processing Once per week Once per week 88

predictive AGC scale factor 1.a) Predictive AGC (pagc) Scaling Calibration 120 115 110 105 100 95 90 85 80 75 Rule of thumb: lg(target)=6 should have a pagc scale factor of ~60-100 70 65 60 55 50 45 40 35 30 25 20 15 10 5 4.0 4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 lg(target) 89

1.a) Predictive AGC (pagc) Scaling Calibration With the pagc scaling calibration we plot the pagc scaling factor vs. lg(target). The calibration starts from a target of 1e4 and finishes with a target of 5e6, scaling the abundance(it) / abundance(ft). why do we need this calibration? Eliminate time consumed by prescan execution. Injection time for dependent scans is predicted based on abundance of the precursor in the master scan, this can be an IT or a FT master scan. For the FT master scan the FT abundance needs to be scaled. This scaling is very important when running methods in which the injection times for ITMS2 scans are predicted from the FT master scans. scaling value at 6 lg(target=1e6) should be around 60-100. 90

1.b) Tips to Check Mass Accuracy Chemical background ions at the beginning of a run, for example m/z 214: [C 10 H 15 NO 2 S]H + = 214.08963 plasticizer n-butyl benzenesulfonamide m/z 391: [C 24 H 38 O 4 ]H + = 391.28429 plasticizer Bis(2-ethylhexyl) phthalate m/z 445: [C 12 H 36 O 6 Si 6 ]H + = 445.12002 polysiloxane Proteome Discoverer mass accuracy plot function Use lockmass option (0 %) No lockmass injected, last found mass used for correction if lockmass is suppressed Peptide ID with external calibration 91

1.c) Lock Mass Abundance 0% What happens when the lock mass abundance is set to 0%: If the lock mass is not found in one scan, the system applies the correction from a previous scan where the LM has been found. For example, at the beginning of a nanolc run background ions such as m/z 445: [C 12 H 36 O 6 Si 6 ]H + = 445.12002 polysiloxane are present and can be used as lock mass. When high abundant peptides are eluting, the chance to lose your lock mass in the FTMS spectrum is high. In such a case, when the lock mass isn t found, the system will use the correction from a previous scan where the lock mass was present and will keep this correction until the lock mass is present again in the FTMS spectrum. HCD scans and a LM abundance set to 0%, the LM correction is done based on the LM found in the full Orbi scan. 92

1.c) Lock Mass Abundance 0% WHY? Possible problems that do occur when using lock mass abundance >0%. A LM > 0% can result in a reduced scan rate due to the accumulation of the Lockmass ion. Lower Lockmass abundance than expected (from your selected abundance e.g.5%) may result from different ion source and/or LC characteristics. A Lockmass with lower abundance will have lower signal/noise, leading to a wider spread of the mass accuracy (Lockmass has more jitter). Furthermore, if the Lockmass is lower, there may be occurrences where the Lockmass is not found leading to mass accuracy deviations. Recommendation for best mass accuracy: Use lockmass option 0 % 93