PACEMAKERS AND ICDS INDICATIONS, FUNCTION AND TROUBLESHOOTING

Similar documents
«Δυσλειτουργία βηματοδότη. Πως μπορούμε να την εκτιμήσουμε στο ιατρείο.» Koσσυβάκης Χάρης Καρδιολογικό Τμήμα Γ.Ν.Α. «Γ. ΓΕΝΝΗΜΑΤΑΣ

Basics of Pacing. Ruth Hickling, RN-BSN Tasha Conley, RN-BSN

Michael Hartman, CRNA, DNP, MSN, BSN, BA

Tachyarrhythmias (fast heart rhythms)

Diagnosis Code Crosswalk : ICD-9-CM to ICD-10-CM Cardiac Rhythm and Heart Failure Diagnoses

UCSF Anesthesia Resident Pearls: Cardiac Implantable Electronic Devices: Pacemakers and ICDs Table of Contents

PRO-CPR Guidelines: PALS Algorithm Overview. (Non-AHA supplementary precourse material)

MRI Case Study MRI Cervical Spine

Management of Pacing Wires After Cardiac Surgery

Introduction. Planned surgical procedures

Introduction to Electrophysiology. Wm. W. Barrington, MD, FACC University of Pittsburgh Medical Center

DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Medicare & Medicaid Services

Atrial Fibrillation and Cardiac Device Therapy RAKESH LATCHAMSETTY, MD DIVISION OF ELECTROPHYSIOLOGY UNIVERSITY OF MICHIGAN HOSPITAL ANN ARBOR, MI

NEONATAL & PEDIATRIC ECG BASICS RHYTHM INTERPRETATION

QRS Complexes. Fast & Easy ECGs A Self-Paced Learning Program

Pacemakers 12/04. Pacemakers. 1. What is a pacemaker?

HTEC 91. Topic for Today: Atrial Rhythms. NSR with PAC. Nonconducted PAC. Nonconducted PAC. Premature Atrial Contractions (PACs)

INTRODUCTORY GUIDE TO IDENTIFYING ECG IRREGULARITIES

Catheter Ablation. A Guided Approach for Treating Atrial Arrhythmias

Automatic External Defibrillators

Pacers use a 5-letter code: first 3 letters most important

MRI SureScan Pacing System

Worldwide Experience with S-ICD: Early results from the EFFORTLESS Registry

Potential Causes of Sudden Cardiac Arrest in Children

What Can I Do about Atrial Fibrillation (AF)?

What Are Arrhythmias?

Electrophysiology study (EPS)

Medtronic Cardiac Rhythm and Heart Failure ICD-10 Coding for Physicians

By the end of this continuing education module the clinician will be able to:

Normal Sinus Rhythm. Sinus Bradycardia. Sinus Tachycardia. Rhythm ECG Characteristics Example (NSR) & consistent. & consistent.

Pacemaker Counter and Histogram Operation and Interpretation. Percentage of total count. Figure 2. Detailed ALTRUA Paced and Sensed Event Counters

CCAD Training Manual. Cardiac Rhythm Management (CRM)

Purpose: To outline the care of patients with permanent or temporary pacemakers.

Team Leader. Ensures high-quality CPR at all times Assigns team member roles Ensures that team members perform well. Bradycardia Management

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

The science of medicine. The compassion to heal.

Atrial & Junctional Dysrhythmias

Treatments to Restore Normal Rhythm

Emergency Insertion of Temporary Trans-venous Pacing Catheter in ICU

Advanced Cardiovascular Life Support Case Scenarios

the basics Perfect Heart Institue, Piyavate Hospital

How To Understand What You Know

Electrophysiology Daymar College. Lisa H. Young, RN, BSN, MAE 2011

Activity 4.2.3: EKG. Introduction. Equipment. Procedure

Introduction. What is syncope?

Basic principles of pacing

Signal-averaged electrocardiography late potentials

Department of Emergency and Disaster Medicine Medical University of LODZ

Electrophysiology Heart Study - EPS -

The Patient s Guide to the Electrophysiologic Study (EPS) and Catheter Ablation

Syncope. enough to include disorders such as epileptic seizures and concussion. January 14-15, 2011 SCA Conference 1

Adult Cardiac Surgery ICD9 to ICD10 Crosswalks

Cardiac Rhythm Management Devices (Part II)

ACLS RHYTHM TEST. 2. A 74-year-old woman with chest pain. Blood pressure 192/90 and rates her pain 9/10.

Atrioventricular (AV) node ablation

Christopher M. Wright, MD, MBA Pioneer Cardiovascular Consultants Tempe, Arizona

PACEMAKERS AND ANTITACHYCARDIA DEVICES WILLIAM BATSFORD, M.D.

COVERAGE GUIDANCE: ABLATION FOR ATRIAL FIBRILLATION

Karen Bain James Cook University Hospital Middlesbrough

VIVA QUAD XT CRT-D DTBA1QQ

Electrodes placed on the body s surface can detect electrical activity, APPLIED ANATOMY AND PHYSIOLOGY. Circulatory system

Sign up to receive ATOTW weekly - wfsahq@anaesthesiologists.org

Assessment, diagnosis and specialist referral of adults (>16 years) with an episode of transient loss of consciousness (TLoC) or a blackout.

Bradycardia CHAPTER 12 CODE SCENARIO

Atrial Fibrillation: The heart of the matter

Implantable Cardioverter Defibrillator Therapy

Advanced EKG Interpretation

BIPOLAR LIMB LEADS UNIPOLAR LIMB LEADS PRECORDIAL (UNIPOLAR) LEADS VIEW OF EACH LEAD INDICATIVE ECG CHANGES

PACEMAKER PATIENT INFORMATION

8 Peri-arrest arrhythmias

Patient Information Sheet Electrophysiological study

NCDR ICD Registry V2.1 Data Collection Form Generator & Leads

GUIDELINE 11.9 MANAGING ACUTE DYSRHYTHMIAS. (To be read in conjunction with Guideline 11.7 Post-Resuscitation Therapy in Adult Advanced Life Support)

Exchange solutes and water with cells of the body

CARDIAC ELECTROPHYSIOLOGY, ARRHYTHMIAS AND PACING. Medical Knowledge. Goals and Objectives PF EF MF LF Aspirational

ATRIAL FIBRILLATION (RATE VS RHYTHM CONTROL)

NAME OF THE HOSPITAL: 1. Coronary Balloon Angioplasty: M7F1.1/ Angioplasty with Stent(PTCA with Stent): M7F1.3

Equine Cardiovascular Disease

Michigan Adult Cardiac Protocols CARDIAC ARREST GENERAL. Date: May 31, 2012 Page 1 of 5

BASIC STANDARDS FOR RESIDENCY TRAINING IN CARDIOLOGY

2014: nowadays the one shot technologies and the injectable monitor allow a wide and complete AF patient management. Why shouldn t we use them?

Understanding the Electrocardiogram. David C. Kasarda M.D. FAAEM St. Luke s Hospital, Bethlehem

Section Four: Pulmonary Artery Waveform Interpretation

ACLS Chapter 3 Rhythm Review Instructor Lesson Plan to Accompany ACLS Study Guide 3e

Novartis Gilenya FDO Program Clinical Protocol and Highlights from Prescribing Information (PI)

Managing Patients With RIATA Leads

Introduction to Electrocardiography. The Genesis and Conduction of Cardiac Rhythm

12-Lead EKG Interpretation. Judith M. Haluka BS, RCIS, EMT-P

The abbreviation EKG, for electrocardiogram,

Atrial Fibrillation The Basics

50 years of CRM Device Therapy Past, Present and Future. Richard Sutton Professor of Cardiology Imperial College, London, UK

BASIC CARDIAC ARRHYTHMIAS Revised 10/2001

Management of Atrial Fibrillation in Heart Failure

Specific Basic Standards for Osteopathic Fellowship Training in Cardiology

Palpitations & AF. Richard Grocott Mason Consultant Cardiologist THH NHS Foundation Trust & Royal Brompton & Harefield NHS Foundation Trust

ACLS PRE-TEST ANNOTATED ANSWER KEY

What to Know About. Atrial Fibrillation

ACLS Study Guide BLS Overview CAB

Evaluation copy. Analyzing the Heart with EKG. Computer

NASPE Training Requirements for Cardiac Implantable Electronic Devices: Selection, Implantation, and Follow-Up

Transcription:

PACEMAKERS AND ICDS INDICATIONS, FUNCTION AND TROUBLESHOOTING Tanja Sloan R.N., PA-C St. Vincent Healthcare Cardiac and Vascular Nursing Conference 2014

Disclosures I have no disclosures

Historical Perspective Electrical cardiac pacing for the management of brady-arrhythmias was first described in 1952 Permanent transvenous pacing devices were first implanted in the early 1960 s First ICD was implanted in 1980

Normal conduction system of the heart

Indications For Permanent Pacing Third degree AV block associated with: Symptomatic bradycardia Symptomatic bradycardia secondary to drugs required for dysrhythmia management Asystole > 3 seconds or escape rate < 40 After catheter ablation of the AV node Post-op AV block not expected to resolve Neuromuscular disease with AV block

Indications For Permanent Pacing Symptomatic bradycardia from second degree AV block Bifascicular or trifascicular block with intermittent third degree or type II second degree block Sinus node dysfunction with symptomatic bradycardia Recurrent syncope caused by carotid sinus stimulation

Indications For Permanent Pacing Post myocardial infarction with: Persistent second degree AV block with bilateral bundle branch block or third degree AV block Transient second or third degree AV block and bundle branch block Symptomatic, persistent second or third degree AV block

Miscellaneous Pacemaker Indications Hypertrophic obstructive cardiomyopathy Dilated cardiomyopathy Cardiac transplantation Termination and prevention of tachyarrhythmias Congenital heart disease in children and adolescents

Indications for Implantable Cardiac Defibrillator Placement Class I 1.Cardiac Arrest Due to VT or VF Not due to transient or reversible cause 2.Spontaneous sustained VT With structural heart disease 3.Syncope of undetermined origin with: Inducible sustained VT or VF that has clinical relevance and/or hemodynamic significance Class II 1.Sustained VT Normal or near-normal LVEF 2.Unexplained Syncope with: Significant LV dysfunction and NICM

Pacing types Temporary Permanent BiVentricular

Pacemaker Components Pulse Generator Electronic Circuitry Lead system ICD + High power

Pulse Generator Lithium-iodine cell is the current standard battery Advantages: Long life 6 to 10 years Output voltage decreases gradually with time making sudden battery failure unlikely

Lead Systems Endocardial leads which are inserted using a subclavian vein approach Actively fixed to the endocardium using screws or tines Unipolar or bipolar leads

Pacemaker Functions Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing rate responsive pacing Provide diagnostic information stored by the pacemaker

ICD Therapies Brady pacing Antitachycardia pacing Cardioversion (1 to 40J) Defibrillation (up to 40J)

ICD Therapy ICD Therapy Low Power (Pacing Therapies) Anti-tachycardia Pacing (ATP) Bradyarrhythmia Pacing High Power (Shock Therapies) Cardioversion Defibrillation

Pacing Function: Asynchronous/Fixed Rate Does not synchronize with intrinsic HR Used safely in pts with no intrinsic ventricular activity If pt has ventricular activity, it may compete with pt s own conduction system VT may result (R-on-T phenomenon)

Pacing Function: Synchronous/Demand Contains two circuits * One forms impulses * One acts as a sensor When activated by an R wave, sensing circuit either triggers or inhibits the pacing circuit Called Triggered or Inhibited pacers Most frequently used pacer Eliminates competition Energy sparing

Pacing Function Atrial pacing: Stimulation of right atrium produces spike on EKG preceding P wave

Pacing Function Ventricle pacing : Stimulation of RT or LT ventricle produce a spic on ECG preceding QRS complex

Pacing Function AV pacing: Direct stimulation of right atrium and either ventricles to mimic normal heart conduction

Sensing Function Sensing: Ability of the pacemaker to see intrinsic cardiac activity when it occurs. Two types: Demand: pacing stimulation delivered only if the heart rate falls below the preset limit. Fixed: no ability to sense. constantly delivers the preset stimulus at preset rate. Triggered: delivers stimuli in response to (sensing) cardiac event.

Capture Function Capture: Ability of the pacemaker to generate a response from the heart (contraction) after electrical stimulation.

Capture Function Electrical capture: Indicated by P or QRS following and corresponding to a pacemaker spike. Mechanical capture: Palpable pulse corresponding to the electrical event.

Troubleshooting Define the problem Identify the cause of the problem Correct the problem Verify the solution

Pacemaker complications Pacemaker complications are divided into: Implant related complications Lead related complications Pacemaker malfunction

Implanted related complications 1. Procedure related/acute Subclavian puncture Traumatic pneumothorax Hemothorax, air embolism AV fistula Thoracic duct injury Brachial plexus injury

Implanted related 2. Hematoma formation at pulse generator 3. Thrombus formation/embolism/av thrombus Asymptomatic due to collaterals 4. Lead perforation/cardiac tamponade rising threshold, diaphragmatic stimulation, friction rub, hypotension 5. SVT/VT during implantation

Pacemaker Infections Pacemaker insertion is a surgical procedure: 1% risk for bacteremia 2% risk for wound or pocket infection Usually occur soon after pacer insertion Presence of a foreign body complicates management Most common S. Aureus/Epidermis Lead vegetations Local inflammation, fever, +BC

Lead related complications 1. Lead dislodgement A>V 2. Lead fracture Usually occurs adjacent to PM near venous site If fracture of bipolar lead occurs, may be possible to reprogram to unipolor configuration 3. Loss of integrity of insulation Manufacturing defects/wear & tear Crush injury

Lead related complications 4. Exit block Abnormality at myocardial-electrode interface 5. Diaphragmatic or pectoral muscle pacing Dislodgement, insulation defects, common with unipolar leads

Pacemaker Failure A. Early: electrode displacement/breakage B. Failure > 6 months Premature battery depletion Faulty pulse generator

Pacemaker Malfunction Generally Be Assigned To Five Categories: Undersensing Oversensing Noncapture No output Pseudomalfunction

Undersensing may be caused by: Inappropriately programmed sensitivity Lead dislodgment Lead failure Insulation break; conductor fracture Lead maturation Change in the native signal

Undersensing Pacer fails to detect an intrinsic rhythm Paces unnecessarily Patient may feel extra beats If an unneeded pacer spike falls in the latter portion of T wave, dangerous tachyarrhythmias or V fib may occur TX: Increase sensitivity of pacer

Undersensing An intrinsic depolarization that is present, yet not seen or sensed by the pacemaker

Oversensing may be caused by: Lead failure Poor connection at connector block Exposure to interference

Oversensing Pacer interprets noncardiac electrical signals as originating in the heart Detects extraneous signals such as those produced by electrical equipment or the activity of skeletal muscles (tensing, flexing of chest muscles, EMI) Inhibits itself from pacing as it would a true heart beat

Oversensing On ECG: pauses longer than the normal pacing interval are present Often, electrical artifact is seen Deprived of pacing, the patient suffers CO, feels dizzy/light-headed Most often due to sensitivity being programmed too high TX: Reduce sensitivity

Oversensing The sensing of an inappropriate signal Can be physiologic or nonphysiologic

Noncapture Pacer s electrical stimulus (pacing) fails to depolarize (capture) the heart There is no failure to pace Pacing is simply unsuccessful at stimulating a contraction ECG shows pacer spikes but no cardiac response CO occurs TX: threshold/output strength or duration

Noncapture No evidence of depolarization after pacing artifact

Noncapture

Noncapture can be caused by: Lead dislodgment Low output Lead maturation Poor connection at connector block Lead failure

Less common causes of noncapture Twiddler s syndrome Electrolyte abnormalities Myocardial infarction Drug therapy Battery depletion

Output failure Output failure occurs when a paced stimulus is not generated in a situation where expected. Results in decreased or absent pacemaker function. Very rare, usually a malfunctioning device.

Output failure

Pseudomalfunction Pseudomalfunctions are defined as: Unusual Unexpected Eccentric ECG findings that appear to result from pacemaker malfunction but that represent normal pacemaker function

Pseudomalfunctions May Be Classified Under The Following Categories: Rate AV interval/refractory periods Mode

Rate Changes May Occur Due To Normal Device Operation: Magnet operation Timing variations A-A versus V-V timing Upper rate behavior Pseudo-Wenckebach; 2:1 block Electrical reset Battery depletion PMT intervention Rate response

Magnet operation Magnet application causes asynchronous pacing at a designated magnet rate

Pacemaker Mediated Tachycardia (PMT) A rapid paced rhythm that can occur with atrial tracking pacemakers

PMT Retrograde conduction Tracking fast atrial rates (physiologic or nonphysiologic)

Retrograde Conduction May Be Caused By: Loss of A-V synchrony due to: Loss of sensing/capture Myopotential sensing Premature ventricular contraction (PVC) Magnet application

Pacemaker troubleshooting

Common problems with ICDs Common: Recalled device issues of varying severity SVTs causing inappropriate shocks Less common: Oversensing (electromagnetic interference, myopotentials) causing inappropriate shocks Lead failure causing inappropriate shocks, pacing problems or sensing problems

ICD with repeated shocks Place magnet over ICD Place external defibrillator pads and telemetry Initiate ACLS or critical care management as appropriate OK to do CPR or external defibrillation before magnet is placed

Case Study 1 67 year old male presents to the emergency room 12 hours after insertion of a pacemaker complaining of left sided chest pain and shortness of breath PR96, RR 33, BP 125/85, Oxygen saturation 88% RA CXR as shown

Pneumothorax Occurs during cannulation of the subclavian vien Incidence -?? Cardiologist dependent Treatment: Asymptomatic or small observation Symptomatic or large tube thoracostomy

Case Study 2 72 year old male presents to the emergency room after a fall, tripped over a bath mat, no LOC Shortened and rotated left leg Past history pacemaker, hypertension Nurse does an routine pre-op CXR and EKG

Septal Perforation Usually identified at the time of pacer insertion but leads can displace after insertion Can occur with transvenous pacer insertion Keys diagnosis are a RBBB pattern on EKG and a pacer lead displaced to the apex of the heart on CXR

Septal Perforation Management: Pacer wire has to be removed but not emergently Small VSD which heals spontaneously

Case Study 83 year old female underwent PPM implantation several hours ago Patient is doing well, VS are stable