HOW SENSITIVE IS THE GAMMA-RAD?

Similar documents
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Introduction to Geiger Counters

GAMMA-RAY SPECTRA REFERENCES

Radiation Detection and Measurement

Atomic and Nuclear Physics Laboratory (Physics 4780)

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Measurement of Germanium Detector Efficiency

Tutorial 4.6 Gamma Spectrum Analysis

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE

EDXRF of Used Automotive Catalytic Converters

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Reflection and Refraction

Introduction to the Monte Carlo method

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, Bologna, Italy

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Gamma and X-Ray Detection

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction

Main properties of atoms and nucleus

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Uses of Derivative Spectroscopy

Jitter Measurements in Serial Data Signals

Radiation Strip Thickness Measurement Systems

Chapter NP-5. Nuclear Physics. Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 NUCLEAR REACTIONS 2.0 NEUTRON INTERACTIONS

ELECTRON SPIN RESONANCE Last Revised: July 2007

FLUID COMPOSITION ANALYSIS BY MULTIPLE GAMMA-RAY BEAM AND MODALITY MEASUREMENTS

The accurate calibration of all detectors is crucial for the subsequent data

X Ray Flourescence (XRF)

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Radioactivity III: Measurement of Half Life.

13C NMR Spectroscopy

Using the Spectrophotometer

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Electron-Muon Ranger (EMR)

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

Calorimetry in particle physics experiments

A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

Appendix A. An Overview of Monte Carlo N-Particle Software

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

Fiber Optics: Fiber Basics

WHITE PAPER. Source Modeling for Illumination Design. Zemax A Radiant Zemax Company

Determination of source parameters from seismic spectra

CHAPTER 7 SCINTILLATION COUNTING

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Exploitation of geometric occlusion and covariance spectroscopy in a gamma sensor array

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Blue Laser Technology Applied to the Microtrac Unified Scatter Technique for Full- Range Particle Size Measurement.

Least-Squares Intersection of Lines

Lecture 14. Point Spread Function (PSF)

Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

2.4 The deflection of beta radiation in a magnetic field. Task. How does β-radiation behave in a magnetic field?

Introduction to Fourier Transform Infrared Spectrometry

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE

Treasure Hunt. Lecture 2 How does Light Interact with the Environment? EMR Principles and Properties. EMR and Remote Sensing

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues

GRID AND PRISM SPECTROMETERS

Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector

with functions, expressions and equations which follow in units 3 and 4.

Experiment 5. Lasers and laser mode structure

4.4 WAVE CHARACTERISTICS 4.5 WAVE PROPERTIES HW/Study Packet

Testing of Kaonetics Devices at BNL

ACCELERATORS AND MEDICAL PHYSICS 2

Magnetic Fields and Their Effects

Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient

Mathematics Pre-Test Sample Questions A. { 11, 7} B. { 7,0,7} C. { 7, 7} D. { 11, 11}

Lab #11: Determination of a Chemical Equilibrium Constant

Electron Muon Ranger (EMR) Software Development

A Short Guide to Significant Figures

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

Measuring Line Edge Roughness: Fluctuations in Uncertainty

[2] At the time of purchase of a Strontium-90 source, the activity is Bq.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Efficiency calibration of the Ge(Li) detector of the BIPM for SIR-type ampoules

(Refer Slide Time 1:13 min)

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function.

Module 13 : Measurements on Fiber Optic Systems

Uncertainty evaluations in EMC measurements

ENERGY LOSS OF ALPHA PARTICLES IN GASES

The KaleidaGraph Guide to Curve Fitting

Understanding Poles and Zeros

Math 0980 Chapter Objectives. Chapter 1: Introduction to Algebra: The Integers.

Characterizing Digital Cameras with the Photon Transfer Curve

Principle of Thermal Imaging

LA-UR- Title: Author(s): Submitted to: Approved for public release; distribution is unlimited.

Transcription:

HOW SENSITIVE IS THE GAMMA-RAD? This is one of the most commonly asked questions. Sensitivity is very important, arguably the most important performance parameter in many applications of gamma-ray spectroscopy. However, there is a great deal of confusion surrounding sensitivity, for two reasons: (1) When a user asks about sensitivity, he or she might mean the intrinsic efficiency, or the intrinsic photopeak efficiency, or the absolute efficiency, or the minimum detectable activity, and so on. There are many different quantities related to sensitivity. (2) Many of these quantities depend not only on the instrument itself, but also on the measurement techniques, on the analysis software used, on background radiation levels, and so on. The various definitions of efficiency relate the fraction of pulses which are detected to those emitted or incident on the detector. Efficiency may be defined as intrinsic vs. absolute, total vs. photopeak, etc. These quantities are used to estimate the count rate from a detector in a particular measurement. Efficiency depends on the scintillation material, its geometry, the distance to the source, and the energy of γ-rays. The minimum detectable activity answers the question: what is the minimum amount of radioactive material which can be detected with confidence? It is related to the minimum number of counts required to distinguish a source from background. It depends not only on the efficiency of the detector but also on the background radiation level, the signal processing algorithms, measurement time, etc.. This note will briefly define the various parameters used to quantify sensitivity and will show typical values for Amptek s Gamma-Rad. The information in this note is a summary of information described in more detail in various references. For a general discussion of sensitivity, we recommend the text by Knoll 1 and the on by Gilmore and Hemingway 2. Most of this discussion is not unique to the Gamma-Rad, but would be applicable to any detector using a scintillator and PMT with similar properties. 1 EFFICIENCY 1.1 INTRINSIC TOTAL EFFICIENCY The intrinsic total efficiency is defined as the ratio of the total number of events which are detected to the total number of γ-ray photons incident on the detector. It is determined by the total attenuation coefficient. The plot below shows the total interaction probability, the probability of any interaction occurring, for γ-rays passing through 7.6, 10.0, and 15.2 cm of NaI(Tl). At 662 kev, for example, about 87% of the incident γ-rays will interact when passing through 7.6 cm of NaI(Tl). 100% Total Interaction Probability 75% 50% 25% 152 mm 100 mm 76 mm 0% 100 1,000 10,000 Incident Photon The values in this plot are a good estimate but there are additional effects. First, the scintillator is surrounded by 1.5 mm Al, which attenuates photons below ~30 kev. Second there are geometric effects: if γ-rays are incident along the side of the cylindrical scintillator, then the path length varies with incident location. For sources at a large distance, this is a small correction: for a 7.6 cm diameter cylinder, at 662, Amptek Inc. AN-GRD-001 Page 1 of 9

the efficiency correction is ~1%. But if the distance to the source is less than 20 times the detector dimension, then the correction may be quite important 3. Third, it is possible for the signal processing electronics to fail to detect an interacting γ-ray. If the amplitude of the pulse is very small, below the noise threshold of the electronics, it will not be detected, but in the Gamma-Rad the noise threshold is very low so this is usually negligible. Since radioactive decays occur at random times, the electronics may be processing one pulse when a second interaction occurs. These dead time losses are usually negligible at low counting rates but at sufficiently high count rates are important. 1.2 PHOTOFRACTION The plots below show a spectrum measured from a 137 Cs source and a background spectrum. The plot on the left is a linear plot of the data. Many of the counts occur in a Gaussian peak at 662 kev, which is the energy at which the 137 Cs emits γ-rays. This is termed the photopeak. But there are many counts from the 137 Cs at lower energies, arising from scattering events. In these, the incident γ-ray deposits a portion of its energy in an interaction which produces a secondary particle, which may then exit the detector. The plot on the right is a logarithmic plot of the same data. The signal to background ratio is clearly improved if one uses only the photopeak or full energy counts. Photopeak counts are produced by photoelectric interactions or if the incident γ-ray is scattered and the secondary particle is stopped in the detector. 1.6E+04 1.0E+05 1.2E+04 137 Cs 1.0E+04 137 Cs 1.0E+03 Background Counts 8.0E+03 Counts 1.0E+02 4.0E+03 Background 1.0E+01 0.0E+00 0 100 200 300 400 500 600 700 800 900 1000 1.0E+00 0 500 1000 1500 2000 2500 The photofraction is defined as the ratio of the number of events which deposit their full energy in the detector, forming the photopeak, to the total number of events which are detected. There is no easy way to calculate it. It is easy to compute the probability of photoelectric interactions: at 662 kev, about 20% of the incident γ-rays undergo a photoelectric interaction in 7.6 cm of NaI(Tl). But in a 7.6x7.6 cm NaI(Tl) detector, the photopeak efficiency is closer to 50%, so most of the photopeak events are not due to photoelectric interactions. The photofraction can be computed using Monte Carlo simulation methods or measured. The figure below shows results which were measured for a 3 x3 cylindrical NaI(Tl) scintillator 4. The photofraction increases for larger detectors, as more secondaries are captured. It depends on the geometry of the detector and also on the distance to the source. 1.0 0.8 Peak to Total Ratio 0.6 0.4 0.2 0.0 0 500 1,000 1,500 2,000 2,500 3,000 Amptek Inc. AN-GRD-001 Page 2 of 9

1.3 INTRINSIC PHOTOPEAK EFFICIENCY The intrinsic photopeak efficiency is the ratio of the number of full energy events to the total number of γ-ray photons incident on the detector. It is the product of the intrinsic total efficiency and the photofraction. For 137 Cs and a 7.6x7.6 cm NaI(Tl), it is (50%)(87%)=43%. 1.4 ABSOLUTE TOTAL EFFICIENCY The absolute total efficiency is defined as the ratio of the total number of events which are detected to the total number of γ-ray photons emitted by the radioactive source. It is the product of intrinsic total efficiency and a geometric factor G F, which yields the fraction of the emitted γ-rays which are incident on the detector. Consider the geometry sketched below. A source is located a distance R from a detector with area A facing the source. The γ-rays are emitted isotropically, i.e. with equal probability in all directions, with solid angle 4π steradians. Seen from the source, the detector subtends a solid angle Ω, defined by the area of the detector, which determines the geometric factor. At a large distance, the fractional solid angle is just the area A divided by the area of a sphere with radius R, and this is the fraction of γ-rays incident on the detector. For a 7.6 cm dia detector at 60 cm, the geometric factor is 0.001, as calculated below. G F A = 4π R 2 Area A π ( 3.8 cm) 2 4 π ( 60 cm) 2 = 1.003 10 3 Ω R The simple formulas only holds if the separation between the source and detector is large. If we consider the limit of small separation, the detector subtends half the solid angle, and this does not change even if the source is moved slightly away from the detector. For intermediate values, analytical approximations have been computed for a variety of different geometries 5. Note that the absolute efficiency is defined in terms of the number of γ-rays emitted by the source, rather than by the number of radioactive decays. Not every decay is accompanied by the emission of a γ- ray. For example, in 137 Cs the 662 kev γ-ray is emitted in 85% of the decays. Note also that the absolute efficiency assumes there is no attenuation between the source and the detector and no scattering from adjacent materials. If the source is shielded, or is embedded in a matrix which attenuates the γ-rays, this will affect the count rate. Even more subtle is the effect of surrounding materials. Consider the geometry below, where the source is located in a collimator which does not restrict the solid angle impinging on the detector. Some photons from the source will be emitted away from the detector, but then will backscatter from the collimator and thus add to the signal. Computing or measuring absolute count rates is quite challenging. 1.5 ABSOLUTE PHOTOPEAK EFFICIENCY The absolute photopeak efficiency is defined as the ratio of full energy events to the total number of γ- ray photons emitted by the radioactive source. It is the product of absolute total efficiency and photofraction. 1.6 RELATIVE EFFICIENCY The relative efficiency is defined as the ratio of the efficiency of a detector to that of a 3 x3 NaI(Tl) scintillator at 1332 kev, in the same geometry and under the same conditions. This parameter can be confusing, since it is expressed in the same units as the other efficiencies. The relative efficiency may certainly be >100%, which adds further confusion. Amptek Inc. AN-GRD-001 Page 3 of 9

2 SAMPLE COUNT RATE CALCULATION AN-GRD-001 Rev A The worksheet below shows an example count rate calculation for a 1 μci 137 Cs source located 60 cm in front of a cylindrical NaI(Tl) scintillator, 7.6 cm in diameter by 7.6 cm long. The calculation was carried out in MathCad but can easily be carried out using other tools. The activity of the source is S. In this example, we assume 1 μci, or 37 kbq, which corresponds to 3.7x10 4 disintegrations per second. S := 10 6 Ci S = 3.7 10 4 Bq S 3.7 10 4 1 = s The fraction of disintegration which emit a γ-ray is B R, the branching ratio. For 137 Cs, B R =85%, so there are 3.1x10 4 γ-rays per second emitted by the source, isotropically.. B R := 0.85 SB R = 3.145 10 4 1 s The geometric factor GF is the fraction of emitted γ-rays which are incident on the detector. Here we assume a 7.6x7.6 cm NaI(Tl), where the source faces the circular end. Only one out of a thousand of the emitted γ-rays is incident on the detector. π ( 3.8 cm) 2 G F := 4 π ( 60 cm) 2 G F = 1.003 10 3 The rate of γ-rays incident on the detector is R inc. In this example, it is 31 sec -1. R inc := SB R G F R inc = 31.537 1 s The total intrinsic efficiency was shown to be 87%, and the photofraction about 50%. From these we can compute the total rate of interactions R tot, and the photopeak rate, R peak. ε intrinsic := 0.87 ε photofraction := 0.5 R tot := R peak := R inc ε intrinsic R tot ε photofraction For the example here we compute a total rate of 27 sec -1 and the photopeak rate of 14 sec -1. R tot = 27.438 1 s R peak = 13.719 1 s For some applications, knowing the count rate might be sufficient. However, in many real-world applications, the question is: can we distinguish these count rates from background? In our laboratory at Amptek, we typically find background count rates of 300 sec -1. The background count rate in the 662 kev peak is roughly 7 sec -1. All of these counts are subject to statistical fluctuations. Can we confidently detect the presence of this source? With what statistical confidence? How long must we record data to obtain the necessary confidence? The next section will address these questions. Amptek Inc. AN-GRD-001 Page 4 of 9

3 DETECTION LIMITS The plot below shows spectra measured from an 80 μci (2.9 MBq) 137 Cs source at distances from 30 cm to 10 m, using a large volume scintillator (10x10x40 cm 3 NaI(Tl)). Each spectrum was measured for 60 seconds. Also shown is the background spectrum, measured for 10 minutes and averaged. The dose rate from the background and from the source at each distance is also shown. Counts (60 s) 1.0E+05 1.0E+04 1.0E+03 1.0E+02 30 cm 60 cm 120 cm 240 cm 480 cm 960 cm Background 22 μs/hr 5.5 μsv/hr 1.4 μsv/hr 300 nsv/hr 80 nsv/hr 20 ns/hr 500 nsv/hr 1.0E+01 1.0E+00 0 500 1,000 1,500 2,000 2,500 3,000 The background radiation level was 500 nsv/hr, with total count rate of almost 2,970 sec -1. At 10 m, the 137 Cs source added only 20 nsv/hr and 70 sec -1. The signal was much lower than the background. But in the spectrum, one can clearly see the additional counts in the 662 kev photopeak. Looking only at the photopeak, the background rate is 165 sec -1. Even in the photopeak the signal is less than background, yet we can clearly distinguish the signal. So the question is: how well can we distinguish a source from background and random statistical fluctuations? How weak a source can be detected? 400 Counts - Backgroud (60 s) 300 200 100 Background Spectrum N B N S 0 0 200 400 600 800 1,000 1,200 1,400 The figure above illustrates a typical measurement. We define a region of interest (ROI) within the spectrum, marked by the vertical lines, 620 to 670 kev here. N B is the number of counts within the ROI due to background, the total area below the red curve. N S is the number of counts within the ROI due to the signal, the total area between the red and blue curves, sometimes called net counts. N T is the total number of counts within the ROI, sometimes called gross counts: NS = NT NB If there were no fluctuations in the counts or other source of uncertainty, then if N S were greater than zero, we would conclude that a source was present, and if N S were zero, we would conclude no source is present. Amptek Inc. AN-GRD-001 Page 5 of 9

But there will always be statistical fluctuations in the counts, due to the random nature of radioactive decays, and there are likely to be additional sources of uncertainty and fluctuation in a real instrument. We therefore must require that N S be greater than zero by its uncertainty times factor k, which determines the confidence: N We have and so ( ) k σ S N T σ = σ + σ 2 2 2 NS NT NB σ = N = N + N 2 NT T S B ( N N ) N k ( N N ) σ = + + σ + + σ 2 2 2 2 2 N S B N S S B N S B B This implies that N S must exceed a value which depends on the background rate and on the uncertainty in the estimate of the background rate. There is some lower limit on the total counts, or for a fixed measurement time, on the count rate. For a given absolute efficiency, this implies a minimum activity. The σ 2 NB term arises because we do not directly measure the background in the ROI, we estimate it from something else. This term represents the uncertainty in the background estimate and so depends on the algorithm used to estimated background. In the plot above, we used a prior measurement of the background to estimate N B, but this is not always possible. A common alternative is to estimate it from the spectrum itself, for example using the thin gray line drawn across the ROI. The measured counts in channels just outside the ROI are then used to estimate N B. These two algorithms will yield different values for σ 2 NB. In this example, we used the counts in an ROI around the photopeak to estimate N T and N S. One could also use all the counts in the spectrum or various other algorithms. Sometimes one knows the isotopes of interest before the measurement, and so has known ROIs, while in other cases one is trying to find unknown peaks and so must determine the ROIs from the measured spectrum. The spectral analysis algorithms are critical in determining the MDA. 3.1 CRITICAL AND DETECTION LIMITS One needs to compare the net counts, N S, to a threshold to determine if a source is present or not, typically with 95% confidence. The critical limit L C is defined so that, if N S <L C, we may conclude that no source is present with a false-positive rate no larger than 5%. The detection limit L D is defined so that, if N S >L D, we may conclude that a source is present with a false negative rate no larger than 5%. It can be shown 6 that LC = 2.326 σ NB L = 4.653 σ + 2.706 D N B It may seem surprising that the two limits are different, but if we had a source which gave exactly L C, then half the time the counts would be below L C and half the time below, so we only have 50% confidence. To give 95% confidence, L D must be greater than L C. 3.2 MINIMUM DETECTABLE ACTIVITY The minimum detectable activity is defined as the minimum amount of radioactive material necessary to yield L D. We can rewrite this using the terms from the efficiency: S MDA is the minimum detectable activity, T is the measurement time, B R is the γ-ray yield per disintegration, G F is the geometric factor, ε intrinsic is the intrinsic efficiency, and ε fraction is the fraction of interactions which are summed in the algorithm (this may be the photopeak, all of the counts, or something else). This gives Amptek Inc. AN-GRD-001 Page 6 of 9

L D = TiR meas ( MDA R F )( int rinsic fraction ) = T S B G ε ε Rearranging this, and noting that (G F ε intrinsic ε fraction ) is the absolute photopeak efficiency, yields S L 4.653 σ + 2.706 D MDA = = R F rinsic fraction R F rinsic fraction ( εint ε ) ( εint ε ) B G T B G T N B The σ NB is a strong function of the intensity and spectrum of the background radiation. In many cases, it may be a function of the measurement time. The absolute efficiency is a strong function of distance to the source and of energy. So the MDA can only be properly defined under limited conditions. One cannot state that the MDA is 1 μci. Instead, one must define the energy of the source, the distance to the source, and the background radiation level to properly compare the MDA of two systems. Amptek Inc. AN-GRD-001 Page 7 of 9

4 SAMPLE MDA CALCULATION AN-GRD-001 Rev A Consider an example, using the efficiencies and counting rates computed above, if the background count rate in the photopeak is 16 sec -1. Further assume that we characterize the background by a 600 second measurement and then average the result. This leads to a measurement of 9600+98 background counts in the photopeak, for a background rate of 16+0.16 sec -1. In a 30 second measurement, the uncertainty in the background correction will be 4.9 cts. R back := 16 1 s N backgnd T back σ backgnd T back ( ) R back T back := N backgnd ( 600 sec) = 9.6 10 3 ( ) N backgnd ( T back ) σ backgnd ( 600 sec) 600 sec σ back T meas, T back := σ backgnd ( 600 sec) = 97.98 = 0.163 1 s ( ) σ backgnd ( T back ) T meas := σ T back ( 30 sec, 600 sec) = 4.899 back Now assume that we measure a 1 μci 137 Cs source at a distance of 60 cm from a 3.8x3.8 cm NaI(Tl) scintillator. From the efficiency calculations above, this yields a photopeak rate of 13.8 sec -1. If we integrate for 30 seconds, then we obtain 412 cts from the source, 480 background counts, and 892 counts gross in the photopeak. N peak ( T) := T R peak N peak ( 30 sec) = 411.563 N back ( T) := T R back N back ( 30 sec) = 480 N gross ( T) := N back ( T) + N peak ( T) N gross ( 30 sec) = 891.563 If there were no background, then the uncertainty in the peak counts would be 20. With the background, and with the uncertainty in the background removal, the uncertainty in the net counts is 30. If we use a shorter background measurement, e.g. 30 sec, this increases. σ peak ( T) := N peak ( T) σ peak ( 30 sec) = 20.287 ( ) := N peak ( T meas ) N back ( T meas ) σ net T meas, T back ( ( )) 2 + + σ back T meas, T back σ net ( 30 sec, 600 sec) = 30.258 σ net ( 30 sec, 30 sec) = 37.035 We can compute the MDA for the geometry above, using the absolute photopeak efficiency computed earlier. For a 30 second measurement at 60 cm from 3.8x3.8 cm NaI(Tl), in a background of 16 sec -1 which has been measured for 600 sec, the MDA is 62 nci. MDA( T meas, T back ) MDA( 30 sec, 600 sec) = 6.196 10 8 Ci ( ) ( ) 4.653 σ back T meas, T back + 2.706 := T meas B R G F ε intrinsic ε photofraction Amptek Inc. AN-GRD-001 Page 8 of 9

REFERENCES 1 G. Knoll, Radiation detection and measurement, 3 rd edition, John Wiley & Sons, Inc., 2000. The topics of efficiency and MDA are covered on pp 116-119 and 94-96. 2 G. Gilmore, J.D. Hemingway, Practical gamma-ray spectrometry, John Wiley & Sons, 1995 The topics of efficiency and MDA are covered on 119-125 and 137-143. 3 A.A. Mowlavi, R.I. Najafabadi, R.K. Faygh, Calculation of intrinsic efficiency of NaI(Tl) detector using MCNP code, Int. Journ. Pure Appl. Phys., Vol 1, No 2 (2005), pp 129-136. 4 From Figure 1 of Efficiency calculations for selected scintillators, a technical note from Bicron, Inc. 5 N. Tsoulfanidas, Measurement and Detection of Radiation, Hemisphere publishing, 1983, pp 248-258. 6 Derivations are found in Knoll and in Gilmore and Hemingway, pages listed above. Amptek Inc. AN-GRD-001 Page 9 of 9