A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials
|
|
|
- Darrell Armstrong
- 10 years ago
- Views:
Transcription
1 Illicit Trafficking, Sub/cross-national threats, Poster presentation A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Ronald M. Keyser, Timothy R. Twomey, Daniel L. Upp ORTEC 801 South Illinois Avenue Oak Ridge, TN, USA [email protected], [email protected], [email protected] Abstract: The three steps in the interdiction process for illicit trafficking of radioactive and nuclear materials are detection, location and identification. Many handheld radioisotope identifiers (RIDs) have been introduced with the claim to the latter two tasks for gamma-emitting sources. Recent emphasis has been on identification in order to reduce the false positive rate. Previous work compared the search mode of an HPGe-based and a NaI-based RID for neutron and gamma-ray sources. In the present work, the performance of the same two instruments in the identification of radionuclides is compared. The requirements for the identification are given in ANSI N42.34 and related standards. The tests in the standard are for single nuclides and certain mixtures of nuclides for both shielded and unshielded sources. The two systems have been compared under the standard test conditions for a variety of the sources and configurations specified. Results comparing the performance of both systems, with the standard and with each other will be given. The results show the HPGe system provides the correct identification more rapidly and on less spectral data than the NaI system and that it easily exceeds the ANSI N42.34 requirement. Keywords: radioisotope; identifiers; germanium detectors; HPGe; illicit trafficking 1. Introduction The Detective EX is a handheld radioisotope identifier (RID) based on a high purity germanium detector (HPGe) and moderated 3 He tubes for the gamma ray and neutron detection respectively. The size of the HPGe and neutron detectors was based on the efficiency requirements of ANSI N42.34 for the detection of differing amounts of material and on the ability to correctly identify the various nuclides in mixtures. The mixtures specified in the standard are those which could be used to hide prohibited material by masking it with other, innocent, radioactive materials. HPGe is arguably the only material available today which has both the necessary high resolution and adequate detection efficiency in a single detector. The neutron detectors were not used for these identifications. Because of the relatively small size of the HPGe detector (about 12% relative efficiency), there is a widely held misconception that accurate nuclide identification will require a long counting time in order to obtain the necessary spectrum counts. However, these results show that the improved resolution of the HPGe more than compensates for the fewer overall counts than in the scintillator spectrum. In the expected mode of operation of the RIDs, the location of the source is determined in the search mode and then the identification of the nuclides is determined in the identify mode. Prior work [1,2,3,4]
2 showed that high resolution gamma-ray detectors have the same or greater sensitivity than the most common size of low resolution detectors and that the HPGe detector systems can correctly perform the identification, even for mixtures, in less time than the requirements in the standard. This work compares the time for correct analysis for the HPGe, high resolution RID with a scintillator based RID. 2. Experimental Configuration The two RIDs were described earlier [1] and are shown in Fig. 1. The tests were conducted using the two instruments sequentially. The sources were placed in front of the instrument being tested at 10 cm from the front face of the housing (not necessarily the detector) and aligned with the active center of the detector as defined by the manufacturer. The sources were positioned with low mass and low atomic number materials to reduce scatter. The DU samples were positioned so that the maximum surface area was facing the detector. When the point sources were used with the DU, the point sources were positioned beside the DU to eliminate any shielding by the DU. The average dose, as recorded by the instruments, was about 50 :Sv/h above background, consistent with the ANSI N42.34 [5] requirements. The Detective EX continuously collects and analyzes the spectrum. Once a second, the current analysis result is updated on the display. The running live and real time are also displayed. For each source, the identification was started with the source in place and Figure 1. The Detective and scintillator-based RIDs. the screen was continuously observed so that when the correct result was displayed, the real time (representing the data collection time) could be recorded. The identification was repeated and the results shown are the average of the multiple analyses. The scintillator-based RID operates by collecting a spectrum for a fixed time, storing the spectrum, analyzing the spectrum and displaying the result. For this test, the collection time was set to the minimum time and the manual time extension mode was used to extend the time of the spectrum collection until a result was observed. The recorded times are only for the spectrum collection and do not include the extra time for processing, typically about 25 seconds Both instruments were operated at the same confidence level on the peak areas for the peak to be recognized as present in the spectrum. This level was set such that the uncertainty of the net peak area was less than 5 sigma for most peaks. This level ensures a low level of false positives. For either instrument, if the correct analysis was not observed in 180 seconds, the process was stopped and the result recorded as a failure. The maximum counting time is not given in ANSI N42.34, but is listed in the IAEA draft TECDOC [6] as 60 seconds for unshielded sources. Some of the sources were point sources and some were extended sources. The list of sources is shown in Table 1. Source Nominal activity Form Correct report Cs uci Point source Cs 137 Co 60 1 uci Point source Co 60 Ba uci Point source Ba 133 Depleted uranium 98 g Bar (5 x 5 x 30) Depleted uranium Depleted uranium 18 g Thin plate Depleted uranium
3 Mixed uci each Point source Am241, Cs137, Co60, Cd109 Thorium 200 g natural thorium Bulk powder, about Th 232, Th 228, or Ra mm x 15 mm diameter Natural Uranium Unknown Uranium in glass Uranium Natural Uranium 5000 bq Compressed powder, 5 mm x 40 mm diameter Uranium 2.1. Example spectra As soon as the correct analysis was observed in the Detective, the spectrum was manually saved. An example spectrum for identifying 60 Co is shown in Fig. 2. In this spectrum, note the relatively low number of counts needed to make the identification. The DU and 133 Ba combination is shown in Fig. 3. In this spectrum, note the closely spaced gamma-ray peaks in the spectrum, which are easily separated by the HPGe detector. Figure 2. Short time spectrum of 60 Co from HPGe detector. Figure 3. HPGe Spectrum of depleted uranium and 133 Ba. The corresponding spectra for the scintillator system are shown in Figs. 4 and 5. Figure 4. Scintillator spectrum of 60 Co.
4 The scintillator spectrum does not show the separation of the gammaray lines, making the identification difficult. 3. Results 3.1 Single nuclides Figure 5. Scintillator spectrum of depleted uranium and 133 Ba. For the single nuclide cases, both RIDs were able to correctly identify the nuclide, however the Detective determined the result in less time than the scintillator-based system. 3.2 DU and DU with other nuclides The HPGe RID was able to identify DU as depleted uranium both alone and with other nuclides in less than 1 minute, and in some cases less than a few seconds. The time to identify increases as the number of gammaray lines in the spectrum increases, but remains at or near the 60 second requirement except for the DU and mixed nuclide standard. The DU plus mixed 1888 standard is identified in about 100 seconds as DU plus the 4 nuclides shown in Table 1. The scintillator RID reported unknown in all cases where DU was the sample. For the mixtures with the 60 Co 133 Ba and natural thorium, the Figure 6. Comparison of time to identify single nuclides alone and with DU by HPGe and Scintillator RIDs. addin nuclide was reported with no DU reported. For the mixture of 137 Cs and DU, the scintillator RID only reported 137 Cs. These results are shown in Fig Other nuclides and combinations Several other combinations of nuclides were tested. The results are shown in Fig. 7. The HPGe RID is able to identify the material in less than 130 seconds in all cases and generally less than 40 seconds. The background test was done in a room with concrete floors, walls and ceiling and the expected result was 40 K. The background dose is about 5 :Sv/h. Figure 7. Comparison of time to identify multiple nuclide mixtures, uranium, and natural material by HPGe and Scintillator RIDs.
5 The scintillator RID reported unknown for the small DU sample and for the natural uranium sample. For the mixed gamma, only 241 Am and 60 Co were reported with 109 Cd and 137 Cs missing. For the background case, no nuclides were reported. For the thorium ore, any of the possible components (e.g., 232 Th, 228 Th, or 226 Ra) were accepted as the correct result, since the different requirements can dictate which result is displayed. 4. Conclusions The results show that the HPGe-based RID can correctly analyze single nuclide samples and mixtures of materials in less time than a RID based on scintillation detectors. More importantly, the depleted uranium sample was correctly identified by the HPGe and not identified as DU by the scintillator, both in mixtures and even when DU was the only material present. This is possible because of the greatly improved resolution of HPGe compared to scintillators, which greatly improves the signal-to-noise ratio. Further work will concentrate on more complex mixtures. 4. References 1 Keyser, R. M., Twomey, T. R., and Upp, D. L., An Improved Handheld Radioisotope Identifier (RID) for both Locating and Identifying Radioactive Materials, HPS Midyear Meeting, January Keyser, R. M., Twomey, T. R., and Upp, D. L., Performance of Light-Weight, Battery-Operated, High Purity Germanium Detectors for Field Use, INMM Annual Meeting, July Twomey, T. R. and Keyser, R. M., Hand-Held Radio Isotope Identifiers for Detection and Identification of Illicit Nuclear Materials Trafficking: Pushing the Performance Envelope, World Customs Organization Conference, Sept Keyser, R. M. and Twomey, T. R., Performance of a Portable HPGe-Based Nuclide Identifier on Multiple Nuclide Mixtures, INMM Annual Meeting, July American National Standards Institute, Performance Criteria for Hand-held Instruments for the Detection and Identification of Radionuclides, ANSI N42.34, January, International Atomic Energy Agency, Technical / Functional Specifications for Border Radiation Monitoring Equipment IAEA-TECDOC-xxxx, Rev 20E final draft April 2005
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831
ORTEC. High Resolution Detection Systems for Interdiction of Nuclear Material Trafficking. Abstract. Introduction
ORTEC High Resolution Detection Systems for Abstract The interdiction of nuclear materials during illegal transportation requires both detection of radiation and identification of radionuclides in order
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,
ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software
ORTEC DET-SW-UPG Latest Software Features Three Search Modes: Gamma/Neutron total count rate. SNM search mode. Sliding average "monitor" mode. (NEW) User choice of identification schemes: Classify mode
Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software
ORTEC AN55 APPLICATION NOTE Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software 1. Introduction Quality Assurance (QA) measurements are made by analysis laboratories
Overview of Nuclear Detection Needs for Homeland Security. Abstract
Overview of Nuclear Detection Needs for Homeland Security Timothy E. Valentine * Oak Ridge National Laboratory, P.O. Box 2008, Oak Ridge, Tennessee, 37831, USA Abstract The need for advanced and improved
Sepye SeptembMay 2015- Volume 3, Issue 1 In This Issue Thank You for Your Feedback NEW! Detective-Remote V3 NEW! Mobile MCB Server NEW! Detective-Pro NEW! Programmer's Toolkit NEW! M-1-T-2 Tripod NEW!
Micro-Detective. Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier
Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier Accurate nuclide identification based on high resolution gamma spectroscopy with confirmatory neutron detection. Definitive detection of illicit
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA
ABSTRACT Comparison of Activity Determination of Radium 226 in FUSRAP Soil using Various Energy Lines - 12299 Brian Tucker*, Jough Donakowski**, David Hays*** *Shaw Environmental & Infrastructure, Stoughton,
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
ORTEC Radioactivity and its Measurements in Foodstuffs Where Does Radioactivity in Food Come From? Natural Radioactivity: Industrial Radioactivity:
ORTEC Food Monitoring Note Radioactivity and its Measurements in Foodstuffs Where Does Radioactivity in Food Come From? Radioactivity is, and always has been, present in all foodstuffs to some extent.
Antoine Henri Becquerel was born in Paris on December 15, 1852
Discovery Antoine Henri Becquerel was born in Paris on December 15, 1852 Summit Environmental Technologies, Inc. Analytical Laboratories 3310 Win Street Cuyahoga Falls, Ohio 44223 Fax: 1-330-253-4489 Call
Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
Radiological mapping and characterization at the Barsebäck nuclear power plant
Radiological mapping and characterization at the Barsebäck nuclear power plant Leif Spanier, Scandpower AB Lars Håkansson, Barsebäck Kraft AB Radiological Characterisation for Decommissioning Workshop
Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N
Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N Berthold Technologies has more than 60 years experience in manufacturing radiation protection measuring
RADIATION DETECTION & ANALYSIS SYSTEMS FOR WATER
RADIATION DETECTION & ANALYSIS SYSTEMS FOR WATER On Line Remote Sensing Systems On-line remote monitoring of gamma radiation in seawater using spectroscopy and gross monitoring detectors with remote command
identifinder R-Series Handheld Radiation Detectors
identifinder R-Series Handheld Radiation Detectors Portable Gamma Spectroscopy Systems Entry Control Checkpoints Vehicle Screening Checkpoints Package/Baggage Inspection Mailroom Safeguards Critical Infrastructure
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
Radiological Protection Principles concerning the Natural Radioactivity of Building Materials
Radiological Protection Principles concerning the Natural Radioactivity of Building Materials European Commission Radiation protection 112 Radiological Protection Principles concerning the Natural Radioactivity
Micro-Detective-HX. Enhanced Capability, Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier
Enhanced Capability, Ultra-Light, High-Fidelity Hand-Held Radioisotope Identifier Wireless Remote Monitoring Remotely control and monitor from a central location. Portable Light weight, long battery life,
Performance of a portable, Digital-Signal-Processing MCA with Safeguards Germanium Detectors
Performance of a portable, Digital-Signal-Processing MCA with Safeguards Germanium Detectors Bingham, R. D., Keyser, R. M., Twomey, T. R. ORTEC, PerkinElmer Instruments, Inc 801 South Illinois Avenue Oak
GammaVision 8. Gamma Spectroscopy Software. Compatible, Efficient, and Defendable Results for Gamma Spectroscopy Applications.
GammaVision 8 Gamma Spectroscopy Software Compatible, Efficient, and Defendable Results for Gamma Spectroscopy Applications. GammaVision is an All-Inclusive gamma spectroscopy application for High and
EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction
ACTA PHYSICA DEBRECINA XLVI, 25 (2012) EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES L. Csedreki 1 1 Institute of Nuclear Research of the Hungarian Academy of Sciences,
arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014
Monte Carlo Simulation of Indoor External Exposure due to Gamma-emitting Radionuclides in Building Materials arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014 DENG Jun 1) CAO Lei SU Xu 2) Key Laboratory
Networks and Connections in the Counting Room
Networks and Connections in the Counting Room Ronald M. Keyser and Timothy R. Twomey EG&G ORTEC Oak Ridge, TN 37830 Abstract The use of networks in the laboratory to connect PCs, printers and MCAs has
ORTEC. Micro-trans-SPEC. Ultra-Light, Battery-Powered, Portable, LN 2 -Free, HPGe Gamma Spectrometer
ORTEC Micro-trans-SPEC Ultra-Light, Battery-Powered, Portable, LN 2 -Free, HPGe Gamma Spectrometer A Lightweight and Compact Solution for Heavyweight Spectroscopic Problems Micro-trans-SPEC Overview Amazingly
An Innovative, portable MCA based on Digital Signal Processing. R.D. Bingham, R.M. Keyser, T.R. Twomey
An Innovative, portable MCA based on Digital Signal Processing R.D. Bingham, R.M. Keyser, T.R. Twomey ORTEC, PerkinElmer Instruments, Inc Oak Ridge TN 37831-0895 USA Abstract Digital Signal Processing
Waste Management 04 Conference, February 29 - March 4, 2004, Tucson, AZ Copyright WM Symposia, Inc. All Rights Reserved. Reprinted with permission.
RADIOLOGICAL FALSE POSITIVES IN ENVIRONMENTAL SOIL AND GROUNDWATER DATA FROM COMMERCIAL LABORATORIES Walter Kubilius Westinghouse Savannah River Company Thomas Coffey, Paul Mark EXR, Inc. Allen Volesky
For convenience, we may consider an atom in two parts: the nucleus and the electrons.
Atomic structure A. Introduction: In 1808, an English scientist called John Dalton proposed an atomic theory based on experimental findings. (1) Elements are made of extremely small particles called atoms.
Experiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado
1 Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Introduction Some radioactive isotopes formed billions of years ago have half- lives so long that they are
RN-222 RELEASE TO THE ENVIRONMENT: COMPARISON BETWEEN DIFFERENT GRANITE SOURCES
EG0600055 Proceedings of the Environmental Physics Conference, 24-28 Feb. ~, -...,., - OJr - RN-222 RELEASE TO THE ENVIRONMENT: COMPARISON BETWEEN DIFFERENT GRANITE SOURCES A. Mamoon and Salah M. Kamal
Laboratory Technologies, Inc. Multi-Wiper. Time Saving Multiple Detector Wipe Counter
Laboratory Technologies, Inc. Multi-Wiper Time Saving Multiple Detector Wipe Counter Features Wipe test sets In-vitro tests Patient library Isotope library AutoSpect efficiency correction Advanced 4096-channel
ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy
Equipment Needed from ORTEC GEM10P4/CFG-PV4/DWR-30 Coaxial Detector System (Includes detector, cryostat, dewar, preamplifier, and 12-ft. cable pack); typical specifications: 10% relative efficiency, 1.75
How To Monitor For Radioactive Material In International Mail
IAEA Nuclear Security Series No. 3 Technical Guidance Reference Manual Monitoring for Radioactive Material in International Mail Transported by Public Postal Operators Jointly sponsored by the UPU WCO
Environmental Radiation Monitoring in Taiwan
Environmental Radiation Monitoring in Taiwan Introduction: Ching-Chung Huang Radiation Monitoring Center, Atomic Energy Council 823, Cherng-Ching Road, Kaohsiung, Taiwan cchuang @ trmc.aec.gov.tw Chemical,
CONCENTRATION OF U-238, U-235, RA-226, TH-232 AND K- 40 FOR SOME GRANITE SAMPLES IN EASTERN DESERT OF EGYPT.
CONCENTRATION OF U-238, U-235, RA-226, TH-232 AND K- 40 FOR SOME GRANITE SAMPLES IN EASTERN DESERT OF EGYPT. S. Harb*, A. H. El-Kamel**, A. I. Abd El-Mageed**, A. Abbady*, and Wafaa Rashed** * Physics
Activities at temporary job site utilizing U.S. NRC Radioactive Material License #06-30556-01 Amendment 03 0303531(,
CABRERA SERVICES RADIOLOGICAL. ENGINEERING. REMEDIATION May 17,2011 Regional Administrator U.S. Nuclear Regulatory Commission Region I 475 Allendale Road King ofprussia, P A 19406 ATTN: Director, Division
FORMERLY UTILIZED SITES REMEDIAL ACTION PROGRAM
-_-~---- _.l~- _^--._.---..^- FORMERLY UTILIZED SITES REMEDIAL ACTION PROGRAM ELIMINATION REPORT FOk FORMER QUALITY HARDWARE AND MACHINE COMPANY; 5823/5849 NORTH RAVENSWOOD AVENUE; CHICAGO, ILLINOIS JULY
Application Note. ipix A Gamma imager to support various applications
Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose
PROSPECT: Precision Reactor Oscillation and Spectrum experiment
PROSPECT: Precision Reactor Oscillation and Spectrum experiment DAVID MARTINEZ CAICEDO on behalf of PROSPECT collaboration ILLINOIS INSTITUTE OF TECHNOLOGY NUFACT 2015 AUGUST 14th 2015 1 Outline Motivations:
Nuclear data for science & technology: Centres for development
Nuclear data for science & technology: Centres for development An update on the IAEA's nuclear data centre, its services, and the contributions to the global network from developing by Hans Lemmel f\ typical
Eckert & Ziegler Reference & Calibration Sources
coated Eckert & Ziegler Reference & Calibration Sources Product Information uncoated pantone 193u Table of Contents Eckert & Ziegler Reference & Calibration Sources Product Information General Information
Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System
Application Note Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System Based on the German KTA 1503.1 and respective international regulations,
Measurement of Germanium Detector Efficiency
Measurement of Germanium Detector Efficiency Marcus H. Wiggs 2009 Notre Dame Physics REU Advisor: Dr. Philippe Collon Mentors: Matthew Bowers, Daniel Robertson, Chris Schmitt ABSTRACT: A possible discrepancy
Radioactivity III: Measurement of Half Life.
PHY 192 Half Life 1 Radioactivity III: Measurement of Half Life. Introduction This experiment will once again use the apparatus of the first experiment, this time to measure radiation intensity as a function
Radiation Protection s Benefits in the Production of Iodine-131
Radiation Protection s Benefits in the Production of Iodine-131 Daniel M. Cestau *, Ariel N. Novello, Cecilia Bravo, Pablo Cristini, Marcelo Bronca, Eduardo Carranza, Ricardo Bavaro, Julián Cestau and
............... [2] At the time of purchase of a Strontium-90 source, the activity is 3.7 10 6 Bq.
1 Strontium-90 decays with the emission of a β-particle to form Yttrium-90. The reaction is represented by the equation 90 38 The decay constant is 0.025 year 1. 90 39 0 1 Sr Y + e + 0.55 MeV. (a) Suggest,
Measurement of RF Emissions from a Final Coat Electronics Corrosion Module
Engineering Test Report No. 37802-02 Rev. A Measurement of RF Emissions from a Final Coat Electronics Corrosion Module For : Canadian Auto Preservation, Inc. 390 Bradwick Drive Concord, Ontario CANADA
SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES
SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES Salgado M.1; Benítez J.C.1; Castillo R.A.1; Berdellans A.1; Hernández J.M1.; Pirez C.J.2; Soto P.G.2 1 Centre for Radiation Protection
Global Value 7. Productivity Suite for GammaVision. Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation.
Productivity Suite for GammaVision Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation. Global Value was designed to transform the standard GammaVision spectroscopy
P O L Y T E C R.M. S.R.L.
INFORMACIÓN FUENTES EXCEPTUADAS PARA ENSEÑANZA RSS-2 G-M Resolving Time Set This set is used for determining the resolving time of GM Counters. It consists of three half discs, two of which contain 5 microcuries
WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA
Upgrade and Certification of the Los Alamos National Laboratory SHENC 2011-12270 ABSTRACT S.B. Stanfield 1, M. Villani 1, P. T. Barton 1, C. Gerlock 1, D. Nakazawa 1, R. C. Baumann 1, R. Mowry 1, J.P Harvill
IDM-200 -V. Interchangeable HPGe Detector Module (IDM) for Systems Integrators
IDM-200 -V Interchangeable HPGe Detector Module (IDM) for Systems Integrators Free from liquid nitrogen, free from maintenance, free from integration complexities. Convenient-to-mount, all-in-one integrated
Perspectives of a method for measuring soil-gas radon by an opened counting vial
Perspectives of a method for measuring soil-gas radon by an opened counting vial Hugo López del Río Juan Carlos Quiroga Cifuentes Unidad Académica de Estudios Nucleares Universidad Autónoma de Zacatecas
SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
Natural radioactivity in waste from former non-uranium mining areas in Norway
Natural radioactivity in waste from former non-uranium mining areas in Norway Bjørn Lind 1, Torbjørn Gäfvert 1 and Anne Liv Rudjord 1. 1 Norwegian radiation Protection Authority, P.O. Box 55, N-1332 Østerås,
CHAPTER 7 SCINTILLATION COUNTING
CHAPTER 7 SCINTILLATION COUNTING Radiation of various types is widely utilized for non-destructive inspection and testing such as in medical diagnosis, industrial inspection, material analysis and other
SECURITY ARRANGEMENTS OF RADIATION SOURCES
SECURITY ARRANGEMENTS OF RADIATION SOURCES 1 General 3 2 The party running the radiation practice shall be responsible for security arrangements 3 3 Places of use and storages of radiation sources shall
Comparison of natural radioactivity removal methods for drinking water supplies: A review
Comparison of natural radioactivity removal methods for drinking water supplies: A review E. Esmeray, M. E. Aydin Selcuk University Environmental Engineering Department, Konya Turkey e-mail: [email protected]
International Symposium on Nuclear Security
International Symposium on Nuclear Security The Nuclear Security actions taken in Mexico to get an effective protection system of nuclear and radioactive facilities RUBÉN RAMÍREZ GUERRERO [email protected]
Alpha/Beta Counting Manual Systems
Alpha/Beta Counting Manual Systems Ultra High Performance Alpha/Beta Counting Systems for Maximum Laboratory Productivity ORTEC offers a full suite of MANUAL alpha / beta counters incorporating either
MATERIALS LICENSE. 1. Honeywell International, Inc. 3. License Number: SUB-526, Amendment 6-1
NRC FORM 374 U.S. NUCLEAR REGULATORY COMMISSION Page 1 of 5 Pursuant to the Atomic Energy Act of 1954, as amended, the Energy Reorganization Act of 1974 (Public Law 93-438), and Title 10, Code of Federal
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
PTP. Professional Training Programs. The source for professional radiological sciences training. 2015-2016 Courses
Professional Training Programs The source for professional radiological sciences training 2015-2016 Courses Applied Health Physics September 14 October 16, 2015 February 29 April 1, 2016 September 12 October
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Lecture 2 Macroscopic Interactions. 22.106 Neutron Interactions and Applications Spring 2010
Lecture 2 Macroscopic Interactions 22.106 Neutron Interactions and Applications Spring 2010 Objectives Macroscopic Interactions Atom Density Mean Free Path Moderation in Bulk Matter Neutron Shielding Effective
COMPENDIUM OF EPA-APPROVED ANALYTICAL METHODS FOR MEASURING RADIONUCLIDES IN DRINKING WATER
COMPENDIUM OF EPA-APPROVED ANALYTICAL METHODS FOR MEASURING RADIONUCLIDES IN DRINKING WATER June 1998 Prepared by the Office of Environmental Policy and Assistance Air, Water and Radiation Division (EH-412)
Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:
Gamma Rays OBJECT: To understand the various interactions of gamma rays with matter. To calibrate a gamma ray scintillation spectrometer, using gamma rays of known energy, and use it to measure the energy
Quality Assurance Testing of AtomLab Dose Calibrators
(Manufacturer s* Instructions) Introduction The following procedures are the manufacturer s instructions for qualifying AtomLab 100, 100+, 200, 300, 400, and 500 dose calibrators in accordance with NRC
NUCLEAR BOREHOLE LOGGING TECHNIQUES DEVELOPED BY CSIRO-EXPLORATION AND MINING FOR IN SITU EVALUATION OF COAL AND MINERAL DEPOSITS
AU9817334 NUCLEAR BOREHOLE LOGGING TECHNIQUES DEVELOPED BY CSIRO-EXPLORATION AND MINING FOR IN SITU EVALUATION OF COAL AND MINERAL DEPOSITS M. BORSARU and J. CHARBUCINSKI CSIRO, Division of Exploration
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL
ABSTRACT ENRICHMENT AND URANIUM MASS FROM NMIS FOR HEU METAL J. K. Mattingly, T. E. Valentine, J. T. Mihalczo, L. G. Chiang, R. B. Perez Oa Ridge National Laboratory Oa Ridge, Tennessee 37831-6004 Recently,
State of the art & future developments in the detection of nuclear/radioactive materials
NUCLEONICA training course, Karlsruhe 4-6/11/2009 1 State of the art & future developments in the detection of nuclear/radioactive materials Paolo Peerani European Commission Joint Research Centre, Institute
A Review of Emerging Gamma Detector Technologies for Airborne. Radiation Monitoring
A Review of Emerging Gamma Detector Technologies for Airborne Steven Bell ANSRI Dublin 2015 12-14 January Radiation Monitoring Airborne radiation Particulates: sub-μm to sub-mm contaminated with fission
IAEA Support for the Establishment of Nuclear Security Education
IAEA Support for the Establishment of Nuclear Security Education Andrea Braunegger-Guelich, Vladimir Rukhlo Office of Nuclear Security Department of Nuclear Safety and Security International Atomic Energy
Selection, use and maintenance of portable monitoring instruments
HSE information sheet Selection, use and maintenance of portable monitoring instruments Ionising Radiation Protection Series No 7 Introduction The Ionising Radiations Regulations 1999 (IRR99) 1 require
THYROID UPTAKE SYSTEMS LIST OF CONTENTS:
LIST OF CONTENTS: Captus 3000 Portable Thyroid Uptake System.....2 Captus 3000 Thyroid Uptake And Well System...5 Captus 3000 Reports.....7 Captus 3000 Upgrade Package...9 Captus 3000 Custom Protocol...10
FUNDAMENTALS OF A STATE S NUCLEAR SECURITY REGIME: OBJECTIVE AND ESSENTIAL ELEMENTS
IAEA NUCLEAR SECURITY SERIES NO. FUNDAMENTALS OF A STATE S NUCLEAR SECURITY REGIME: OBJECTIVE AND ESSENTIAL ELEMENTS Revision 17.04 Page 1 of 20 FOREWORD [TO BE PROVIDED BY THE SECRETARIAT AT A LATER TIME]
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Basic Nuclear Concepts
Section 7: In this section, we present a basic description of atomic nuclei, the stored energy contained within them, their occurrence and stability Basic Nuclear Concepts EARLY DISCOVERIES [see also Section
Medical Physics and Radioactivity
Medical Physics and Radioactivity Radioactivity Unstable nucleus Electromagnetic wave particle Atoms which emit electromagnetic radiation or a particle by the spontaneous transformation of their nucleus
Objectives 404 CHAPTER 9 RADIATION
Objectives Explain the difference between isotopes of the same element. Describe the force that holds nucleons together. Explain the relationship between mass and energy according to Einstein s theory
Detecting nuclear and radiological materials
Detecting nuclear and radiological materials Summary On 10-11 December 2007 the Royal Society held a two day workshop to explore innovative approaches for detecting the illicit trafficking of nuclear and
Radiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen [email protected] Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
States Environmental Monitoring at Nuclear Power Plants
States Environmental Monitoring at Nuclear Power Plants Alice Hamilton Rogers, P.E. Chair, Conference of Radiation Control Program Directors (Texas Department of State Health Services) Conference of Radiation
NRC s Program for Remediating Polluted Sites J.T. Greeves, D.A. Orlando, J.T. Buckley, G.N. Gnugnoli, R.L. Johnson US Nuclear Regulatory Commission
NRC s Program for Remediating Polluted Sites J.T. Greeves, D.A. Orlando, J.T. Buckley, G.N. Gnugnoli, R.L. Johnson US Nuclear Regulatory Commission Background The U.S. Nuclear Regulatory Commission (NRC)
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
CRIIRAD report N. 10-07. Analyses of atmospheric radon 222 / canisters exposed by Greenpeace in Niger (Arlit/Akokan sector)
CRIIRAD Commission de Recherche et d Information Indépendantes sur la Radioactivité Site : www.criirad.org Tel : + 33 (0)4 75 41 82 50 Fax : + 33 (0)4 75 81 26 48 E-mail : [email protected] Valence,