PS3000 Weighing Indicator USER GUIDE



Similar documents
OCSM Series. High Resolution Digital Crane Scale. User Guide

Digital Weight Indicator Setup / Operation Manual

TI-500E Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual

How To Use A Ci 2001A/B (I2) With A Power Supply (I3)

CASTON. Crane Scale. OWNER S MANUAL (BT Ver.) 9007-CT3-BT

Back-UPS Pro 1300/1500 Installation and Operation

Transmitter Interface Program

Digital indicator 3011

Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS

LEN s.r.l. Via S. Andrea di Rovereto 33 c.s CHIAVARI (GE) Tel Fax mailto: len@len.it url: http//

BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER. Installation, Operation, and Maintenance Instructions. Manual No.

Model 5511 Filler Controller User s Manual Version 1.1 October 2011

User's Guide. Integrating Sound Level Datalogger. Model Introduction

Operation of the 500 computer. Menu. Meaning of the different knobs. Type K. Esc

Digital Indicator. Revision 1.4 September 24, Transcell Technology, Inc. Contents subject to change without notice.

OCS-S High Resolution Digital Crane Scale

User's Guide Integrating Sound Level Datalogger Model

ELECTRONIC COUNTING SCALE. Operation Manual. l JCS- A SERIES

Installation and Operation Back-UPS 1250, 1300, 1500

HCS-3300/3302/3304 USB Remote Programmable Laboratory Grade Switching Mode Power Supply

DC Electronic Loads 8500 series

BlueSolar Pro Remote Panel For BlueSolar PWM-Pro charge controllers 12/24V 5, 10, 20, 30A Article number SCC

M O D U L E - 7 E Model CS-Caliprompter Operator s Manual

Digital I/O: OUTPUT: Basic, Count, Count+, Smart+

BW-500 BED WEIGHER USER MANUAL

Back-UPS 650 VA 120 V with AVR (BX650CI-LM)

GSM Alarm System User Manual

Ron 3025 Crane Scale with 1"/25mm Display

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

GENERATOR START CONTROL MODULE - MINI (2 Wire to 3 Wire)

HIGH PRECISION BALANCE MWP OWNER S MANUAL. MWP ver.1.0

PAC1 Door Access Controller

4-Channel Thermometer / Datalogger

BMW CAR-PC MONITOR MODEL SPECIFICATION BMW3/5 SERIES. Model Name : BMW-100P Paper Version : Ver 1.0

ACS-3S/NEW SU ELECTRONIC SINGLE/DUAL DISPLAY WEIGHING/COUNTING SCALE USER MANUAL

Features, Benefits, and Operation

High precision counting scale. Operation Manual. l RCZ-A Series

Smarthome SELECT Bluetooth Wireless Stereo Audio Receiver and Amplifier INTRODUCTION

INSTRUCTION MANUAL OVERVIEW Remote Meter: MT-5 Remote meter (Model MT-5) is available to connect with solar controller Tracer MPPT series.

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure

XP Series Balances Operation Manual

Operating instructions Platform/floor scales

Electrical Safety Tester

Controller Automation, Model II+

User Manual for CH-PFC76810

Operation Manual. Smart-UPS C. Uninterruptible Power Supply. Tower 1000/1500 VA. 120/230 Vac. su0813a

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

SATEX SA Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg DA Zeewolde HOLLAND

INSTRUCTION MANUAL WEIGHING INDICATOR TI-500 E

Parts List. Navigation

Troubleshooting and Diagnostics

SNMP Web Management. User s Manual For SNMP Web Card/Box

is then retained absolutely without interruption.

Lighting Controls ! WARNING RISK OF ELECTRIC SHOCK. Installation Instructions DESCRIPTION

NC-12 Modbus Application

T Sentry 4 Multi-Point Digital Alarm Instruction Manual

Supplementary Certificate of Approval NMI S669

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

I STAR COMPUTER CO., LTD

G-100/200 Operation & Installation

INDUSTRIAL INSTRUMENTATION

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

IO-DIM6 6-Channel, Digital to Analogue Converters

Instruction Manual FL

The Foundation User Manual LED Battery Slim Par Fixture (RGBA, DMX 7Ch/4CH)

1. SAFETY INFORMATION

Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers

INTELLIGENT CONTROL MODULE SA-2000-II HARDWARE MANUAL. Access Technologies International, Inc.

Operational Overview and Controls Guide

Electronic Pressure Switch EDS 3000

User Manual. Humidity-Temperature Chart Recorder. Model RH520

M 140i. Multifunction Calibrator. Operation Manual

VT300 Weighbridge Indicator User s Guide

CONTIENTS. 1. SPECIFICATIONS (mm) CAUTIONS INSTALLATION INSTRUCTION

CelluLine CGW-TS GSM Cellular Gateway. Installation and Programming Manual

Virtual Measurements & Control. Flip/Flop Filler

AquaTrend Interface. Model AquaTrend Model AquaTrend/SOM Model51400 PortableAquaTrend. Instrument Manual. Catalog Number

PDA Software Manual For the Windows CE and Mobile Operating Systems

USER S MANUAL. MaxPower UPS. Uninterruptible Power System 28-2MAXPO0018

FIREDEX Conventional Fire Panels

IO-RM3 3-Stage Relay Module

omega.com ΩOMEGA RS232 Multi-Drop FMA1600-MDB Multi-Drop Box

PYRODIGITAL CONSULTANTS PD Firing Module Continuity Reader

Product Information Sheet UH28CS-12kV Hi Pot Tester

SOUND LEVEL METER 94405

MSI3460. User Guide. Measurement Systems International. Challenger 3 Crane Scales

DC-8706K Auto Dial Alarm System

Model SR463iR-H, SR463iR-3H Series In-Floor Platform Scale Operating and Service Manual Part No. MAN463iR-H_ Page 1 of 18

INSTALLATION MANUAL XM3 Reader

Model 349KLX Remote Display Digital Scale User Instructions

DYNA-LINK TENSION LOAD SENSORS

User Guide. Model Insulation Tester / Megohmmeter

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

PPM Users Manual Signature Software

Software Manual RS232 Laser Merge Module. Document # SU Rev A

PRT3 Printer Module: ASCII Protocol Programming Instructions

DAB+ / FM Tuner Model: TU-201

10 Thomas, Irvine, CA USA Tel: (949) Fax: (949) Toll Free: (800) 23 FUTEK

Physical Specifications (Custom design available.)

Application Note - Connecting an Electricity Meter to SolarEdge Devices (Europe and APAC)

Transcription:

PS3000 Weighing Indicator USER GUIDE Edition: 08-06-2012

Safety Instruction WARNING Calibration, inspection, and maintenance of the indicator must be performed by a qualified technician. WARNING The indicator can be damaged by static electricity and should be considered sensitive equipment. Remove power during electrical connections. Touching internal components by hand is prohibited.

INDEX Section 1. Instruction... 3 1.1 Battery Instruction (Optional)... 3 Section 2. Installation and Calibration... 4 2.1 Power Supply... 4 2.2 Connecting Load Cell and Indicator... 4 Section 3. Operation... 5 3.3 Zero operation... 7 3.4 Tare operation... 7 3.5 Accumulation operation... 7 3.6 Print and Print Formatting... 8 Section 4. Calibration and Parameters... 9 4.1 Enter Calibration... 9 4.2 Calibration Procedure... 10 4.3 Parameter Settings... 11 Section 5. Error Codes... 15

Section 1. Instruction This weighing indicator is designed for use with bench or floor scales. Its basic weighing functions include: Zero, Tare, Gross/Net, Hold, Print, kg/lb conversion and Total (accumulation). 1.1 Battery Instruction (Optional) 1. When first using the internal battery, charge the battery fully, to prevent low voltage resulting from the self-leakage of battery. 2. A flashing Red light means low battery. 3. The Red light turns on during charging. 4. The light turns Green when fully charged. 5. If the battery is not used for long periods of time, take it out to avoid the leakage. 6. For best results, fully discharge the battery every month (using the indicator until automatically powers off. 3

Section 2. Installation and Calibration Your scale is pre-calibrated at the factory and only requires the load cell cable to be connected from the base to the indicator. The following information is provided for a qualified service technician only. 2.1 Power Supply The indicator is powered by plugging the adapter directly into the DC Power connector located at the bottom of the indicator. Power Input: 100-240V 50/60Hz 1.6A Power Output: 9.0V 1.2A Polarity: Center Positive Approvals: ULc and ULus 2.2 Connecting Load Cell and Indicator The indicator can be used with either 4- or 6-wire 350Ω load cells. Quick Disconnect Pinout: 4

Section 3. Operation Front Panel LED Indicators Functions when Activated kg Weighing in Kilograms Tare A Tare weight is present lb Weighing in Pounds Weight is stable Hold Indicates a held weight 0 Zero weight confirmation Gross Gross weight is displayed Total Go to accumulation mode Net Net weight is displayed Battery AC is present or indicates optional battery charge level. 5

Function Press to print current weight. Zero the displayed weight. Press to tare any non-negative displayed weight. The indicator automatically switches to NET mode. When in Net mode, to clear the tare weight in memory. The indicator automatically switches to Gross mode. When in Net mode, to display the gross weight. The displayed weight will automatically switch back to Net mode after three seconds. Toggles between lb and kg units. Holds the most current weight value on the display until ed again. (see C11 for settings) 1. Accumulation 2. Works together with Print to perform the accumulation function and check the accumulation result Turns the indicator On and Off. Hold for 2 seconds to power Off. 6

3.3 Zero operation 1. Initial zero When the indicator powers up it will automatically zero out 10% of the scale capacity. If this is not desired, change parameter C21 to the desired setting. 2. Manual Zero You can zero up to 100% of the scale capacity by ing ZERO. 3.4 Tare operation Press TARE key, the displayed weight is tared. The indicator then displays a zero Net weight and activates the NET and TARE status lights. In Net mode, TARE key to clear the tare weight. The indicator will then enter the Gross mode. 3.5 Accumulation operation Place the item on the scale and wait until stable, Press TOTAL to enter accumulation mode, the TOTAL indicator will turn on and the display will show the count n001 to confirm the item was accumulated. Remove the item from the platform and allow the scale to return to zero. Place the second item on the platform. When the weight is stable TOTAL, n002 is displayed. Repeat until complete. You can accumulate a maximum of 999 items. Display the accumulated weight --- Press and hold the PRINT key; then the TOTAL key, display n***, to display the counter value and then the total weight. There are 8 digits in total. It will show the first 4 digits then the last 4 digits. For example, the first 4 digits is 0012, and the last 4 digits is 34.56. This means the actual weight is 1234.56 EXIT the accumulation function --- When the indicator shows the last 4 digits, Press and hold TOTAL, the indicator will show clr n. To exit without clearing the accumulator: Press PRINT. To exit and clear the accumulator: Press ZERO and clr y will be displayed. Then Press PRINT. 7

3.6 Print and Print Formatting When the weight is stable, ing PRINT will print the displayed weight. Note: When printing in tare mode, if the weight is negative, it will not print. Set C30 for the desired date format. Tare mode Gross mode Date: XX.XX.XX Date: XX.XX.XX Time: XX:XX:XX Time: XX:XX:XX NET XX.X kg GROSS XXX.X kg TARE XX.X kg GROSS XXX.X kg 8

Section 4. Calibration and Parameters The following information is provided for a qualified service technicians only. Calibration of the scale requires the use of certified weights. Any alteration of the settings and/or calibration may require a service call by a certified technician that is not covered under the warranty. 4.1 Enter Calibration 1. Press and hold PRINT, then HOLD to enter calibration. 2. If this procedure does not enter calibration, remove the rear cover and locate the switch on the lower left of the circuit board labeled MARK. This switch must be in the ON position to enter calibration. Once turned on, repeat step 1 above. NOTE: If the indicator has an NTEP seal and a lead and wire seal, an authorized technician must perform this work so the legal for trade status with local weights and measures authorities is not affected. Button PRINT ZERO TARE HOLD LB/KG GROSS COUNT Calibration Function ENTER UP DOWN BACK LEFT RIGHT Description of Function Accept parameter value and scroll to next parameter. Also (where applicable) moves to the next value within a single parameter. Change parameter value Change parameter value Scroll to previous parameter Use to select parameter digit (C##, C#_, C_#). Scroll digit values using Up and Down arrows. Press Enter to advance to that parameter. Exit calibration and save changes ON/OFF POWER 9

4.2 Calibration Procedure Step Display Parameter Notes 1 PRINT + HOLD [C01 ] Calibration mode entry 2 3 4 5 6 7 [C01 ] [C01 #] [C02 ] [C02 #] [C03 ] [C03 #] [C04 ] [#####.#] [C05 ] [C05 #] [CAL 10] [00000.0] [C06 ] [C06 #] [SPAN ] [02500.0] [CAL 10] [CAL End] Choose calibration unit 1=kg, 2=lb Set number of digits after decimal point 0/1/2/3/4 Set scale division (graduation) 1/2/5/10/20/50 Set scale capacity or to change flashing digit or to select digit Zero calibration 0=no zero calibration, 1= zero calibration calibration zero Note: Zero calibration is recommended Span calibration 0= No span calibration; 1= Span calibration Note: Span calibration is recommended Place weights on the scale platform. Input the weight being used to calibrate with. or to change flashing digit or to select digit Calibration complete. Press TOTAL to save and exit. Press to continue to other parameters. 8 or [C07 ] [07 0] [07 1] Default parameters setting 0=Do not default parameters 1=Default all parameters 10

4.3 Parameter Settings Function Setting Item Parameter Notes (default setting in bold) C08 Warning tone warning tone 0= warning tone On 1= warning tone Off Automatic power off C09 Automatic Shut Off Power saving setting Hold function C10 Sleep Mode setting C11 Hold mode Peak hold: shows the maximum data; application used mainly for materials testing, such as tension and pulling force. Data hold: shows current weight value; main application is for animal weighing. 0= disable Sleep Mode 0= Auto Shut Off disabled 10= Shut Off automatically if no change within 10 min 30= Shut Off automatically if no change within 30 min 60= Shut Off automatically if no change within 60 min 0=disable Hold function 1= Peak hold 2= Data Hold 3= enter Sleep Mode if no change within 3min 5= enter Sleep Mode if no change within 5 min 3= Automatic Hold Function Kg/lb conversion Upper/lower limit alarm C12 kg/lb Button C13 Upper limit alarm value C14 Lower limit alarm value 0= disable kg/lb button 1= kg/lb conversion is active You can set it within the max. capacity limit 11

A/D Raw Display C15 Displays the A/D raw count reading Date and time Communication setting C16 Date C17 Time C18 Serial interface data output method C19 Baud rate In C16 you can set the date, from left to right: year/month/day In C17 you can set the time from left to right: hour/min./sec. 0= Disable serial interface data output 1= Continuous sending for optional remote display 2= Printer setting 3= PC command request method 4= PC continuous format 5= PC/optional remote display continuous format 0= 1200 / 1=2400 / 2=4800 / 3=9600 Zero range C20 Manual zero range (Zero button ) C21 Initial zero range (zero range during indicator power up sequence) 0= Disable manual zero setting 1= ±1% max capacity 2= ±2% max capacity 4= ±4% max capacity 10= ±10% max capacity 20= ±20% max capacity 100= ±100% max capacity 0= no initial zero setting 1= ±1% max capacity 2= ±2% max capacity 5= ±5% max capacity 10= ±10% max capacity Zero tracking C22 Automatic zero tracking range 0= close zero tracking 0.5=±0.5d 1.0= ±1.0d 2.0= ±2.0d 3.0= ±3.0d 12

4.0= ±4.0d 5.0= ±5.0d d = scale division (graduation) as set in parameter C03 Zero tracking timer C23 0= disable zero tracking time 1= 1 second 2= 2 seconds 3= 3 seconds Overload setting Negative display C24 Determines when the scale displays overload error C25 Negative display range 00= disable overload range Valid settings from 01d to 99d Default 05d d = scale division (graduation) as set in parameter C03 0= -20d 10=10% max. capacity 20=20% max. capacity 50=50% max. capacity 100=100% max. capacity Standstill time Motion Filter C26 Standstill time C27 Standstill range C28 When the weight is unstable, you can change the filter setting to make weight display more stable. 0= quick / 1= medium / 2= slow 1=1d 2=2d 5=5d 10=10d d = scale division (graduation) as set in parameter C03 0= disable Motion Aperture 1= 1 Motion Aperture setting 2= 2 Motion Aperture setting 3= 3 Motion Aperture setting 4= 4 Motion Aperture setting 5= 5 Motion Aperture setting 6= 6 Motion Aperture setting 13

C29 Noise filter Note:Set Motion Aperture carefully. For more stable weights use a lower setting. For less stable objects (unbalanced or moving animals) use a higher setting. 0= disable noise filter 1= 1 digital filter strength 2= 2 digital filter strength 3= 3 digital filter strength Date Format C30 0= yy.mm.dd 1= mm.dd.yy 2= dd.mm.yy 3= yy.mm.dd Analog output range C31 0= 0~20mA output 1= 4~20mA output Manually set analog output for device calibration C32 out-4= sets the output current to 4mA Valid settings from 4-20 Relay output setting C33 0= Disable relay output 1= Open relay output function 1 2= Open relay output function 2 3= Reserved (no function) RS485 address C34 C34= 0~99 Add. Code Wireless C35 communication Gravity of calibration C36 location Gravity of destination C37 C35=0~99 signal C36=9.7000~9.9999 (Do not change) C37=9.7000~9.9999 (Do not change) Version No. Reserved menu C38 C39 14

Section 5. Error Codes ERROR REASON SOLUTION UUUUUU 1. overload 2. improper installation of load cell 3. load cell is damaged 1. reduce weight 2. check the load cell connection 3. inspect load cells nnnnnnn 1. improper calibration 2. improper installation of load cell 3. load cell is damaged 1. check that the scale is level 2. check the load cell connection 3. inspect load cells ERR1 ERR2 ERR3 ERR4 ERR5 no input weight or the weight is overload During calibration, the weight is below than Min. required weight During calibration, the input signal is negative During calibration, the signal is unstable input the correct weight The calibration weight Minimum is 10% of scale capacity as set in C04 1. check the connection is correct 2. check load cell for damage 1. Check if platform is level 2. After the platform is stable, start calibration. Microprocessor or A/D failure ERR6 Exceed zero range Remove the load 15