BC-8000 Field Calibration Procedure 55 AMP

Similar documents
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

Transmitter Interface Program

BC-5000 OPERATIONS MANUAL BATTERY CAPACITY TESTER COFKO LLC.

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION

CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13

Essential Electrical Concepts

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

How to use the OMEGALOG software with the OM-SQ2010/SQ2020/SQ2040 Data Loggers.

1. SAFETY INFORMATION

Real Time Clock USB Evaluation Board V3.0

User Manual for CH-PFC76810

OPERATING INSTRUCTIONS

User's Guide. Integrating Sound Level Datalogger. Model Introduction

Troubleshooting and Diagnostics

AutoRanging Digital MultiMeter

Sierra S-Series Smart Interface Software User Manual

Solar Home System. User Manual. AEH-SHS01-10W2L Solar Home System 2 Lamps

Solar Controller / Battery Charger User s Manual

User's Guide DylosLogger Software Version 1.6

DC Electronic Loads 8500 series

Hand-held thermometer Model CTH7000

Troubleshooting Guide, Freedom and Fleet Power Inverter/Chargers

12 Volt 30 Amp Digital Solar Charge Controller Installation & Operation Manual

KCU Communication. PC Software Installation and User s Guide For KCU-30 / KCU-40 / KCU-50 modules. System requirement: Windows XP or above

Lab 3 - DC Circuits and Ohm s Law

12 Volt 30 Amp Digital Solar Charge Controller

User Guide. 4 wire Earth Resistance Tester Model GRT300

Lab E1: Introduction to Circuits

Renewable Energy Monitor User Manual And Software Reference Guide. (979)

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

Ground Resistance Clamp On Tester

13 Watt PORTABLE SOLAR POWER KIT

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor

Instructions: Retain these instructions for future reference SmartChargePro35 RSCPR35-12v, 2 / 8 / 16 / 35A

INSTALLATION GUIDE. Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930

NEWTECH PRO KVA ONLINE UNINTERRUPTIBLE POWER SUPPLY USER MANUAL

COLOR TFT LCD MONITOR. User Manual

SPY-BATT Battery Tutor Device Installation Manual Rev /04/2016

M 140i. Multifunction Calibrator. Operation Manual

SOLAR POWER - THE NEW WAY TO RV!

510 / 511 Cardiorespiratory Monitor Preventative Maintenance Test Procedure

7 FUNCTION DIGITAL MULTIMETER

EMBEDDED ACCESS CONTROL Hardware Installation Guide

EDD-600 FLOW TOTALIZER RATE INDICATOR OPERATION MANUAL. P.O. Box Odessa, Texas USA.

Duct Humidity Transmitter

User's Guide. True RMS Industrial Multimeter

DCX300 - DCX400 - DCX600

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

ANALOG AND DIGITAL METERS ANALOG VS. DIGITAL METERS VOLTMETERS ANALOG AND DIGITAL

SYSTEM 4C. C R H Electronics Design

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

INSTALLATION AND OPERATION MANUAL

PC Software for VAV compact and modular controllers

EVAL-UFDC-1/UFDC-1M-16

EasyNote TJ Series. Disassembly Manual

User Guide. Model Insulation Tester / Megohmmeter

LG Air Conditioning Multi F(DX) Fault Codes Sheet. Multi Split Units

Back-UPS Pro 1300/1500 Installation and Operation

What is a multimeter?

162 CB CABLE TRACER. Filter Probe & Tone Generator INSTRUCTION MANUAL

IPX AUTOMATIC IP NETWORK LOSS BACKUP A/B SWITCH INSTRUCTION BOOK IB

Installation and Operation Back-UPS 1250, 1300, 1500

Digital Photo Bank / Portable HDD Pan Ocean E350 User Manual

INSTRUCTION MANUAL CIRRO MK 3/TYPE 2 MIST GENERATOR

MGL Avionics CAN bus interface for Trig Avionics TT21/TT22 transponders

ExStik TM ph Waterproof Meters

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

BlueSolar Pro Remote Panel For BlueSolar PWM-Pro charge controllers 12/24V 5, 10, 20, 30A Article number SCC

OCSM Series. High Resolution Digital Crane Scale. User Guide

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

PHD User Manual. Table of Contents

Field Calibration Software

MAVO-MONITOR / MAVO-SPOT Instrument Set for Contact or Distant Measurements of Luminances

PDA Software Manual For the Windows CE and Mobile Operating Systems

BMW CAR-PC MONITOR MODEL SPECIFICATION BMW3/5 SERIES. Model Name : BMW-100P Paper Version : Ver 1.0

Automatic taper of charge rate for superior battery life through good equalization of cells and low water use rate.

INSTALLATION & SERVICE MANUAL. Display Panel

ELECTRONIC THERMOSTAT AND THERMOMETER With SPEED CONTROL

Automatic Transfer Switch FT-10 Network Control Communications Module (CCM-T) Kit

Auto-ranging Digital Multimeter INSTRUCTION MANUAL

ST Series POWER SUPPLIES USER INSTRUCTIONS

Model HPX10 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

BATTERY BOX INSTALLATION WARNING WARNING CAUTION ATTENTION AVERTISSEMENT ADVERTENCIA PRECAUCIÓN SOLBBOX12V CARTON INVENTORY COMPLETED OVERVIEW

Weather Direct Displays show Lost Forecast (blank boxes in the picture icons)

e-4 AWT07MLED 7 Q TFT LCD MONITOR (LED Backlighted) USER MANUAL

Contents. Document information

Application Note - Connecting an Electricity Meter to SolarEdge Devices (Europe and APAC)

Integrating Sound Level Meter and Datalogger

OUTPUT CABLE CONNECTIONS:

430 Power/Electronics Replacement

Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual

Current Monitoring Kit

Leaf River Outdoor Products Vibrashine, Inc. P.O. Box 557 Taylorsville, MS (601) FAX (601)

Six-servo Robot Arm. DAGU Hi-Tech Electronic Co., LTD Six-servo Robot Arm

Digital and Analog I/O

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

SYSTEM 45. C R H Electronics Design

Transcription:

BC-8000 Field Calibration Procedure 55 AMP CF1_FIELDCAL_BC8000 Revised 9/8/2012 Charger section can be used on a BC-9000 Version 1.00 INTRODUCTION: The BC-8000 Battery Capacity Analyzer and Charger will perform capacity testing of 12 and 24 volt lead-acid batteries and charge the batteries. Capacity testing amperage setting resolution is 0.1A for tests in the range of 0.5A to 15A and resolution of 1A for tests from 15A to 55A. The micro processor controlled unit uses quartz crystal timing and will accurately measure the capacity of a fully charged battery with a rating between 0.5 and 55 amp-hours. Amperage selection resolution is 0.1A for a load of 0.5 to 14.9 amps and 1A resolution for 15 to 55 amps. The BC-8000 saves the date, time, test parameters, and test results. It is equipped with a USB interface. With the BC Report software the user can retrieve data from the BC-8000 to create a printable report that includes test settings, test results, time and date of test, and a battery discharge voltage graph. The BC-8000 is line powered and easy to operate. The recommended calibration interval is 1 year. Battery charging amperage resolution is 0.01 amp from 0.050 to 2.50 amps and 0.1 amp from 2.6 to 25 amps. The battery charger voltage range is 3.0V to 36.2V. When operating at 3.0V the maximum amperage is 2 amps. The AC power input is universal 100V to 240V. The charger can charge in 1 or 2 steps, maximum run time limit on each step, and operate in constant voltage or constant current mode. SCOPE: This document provides instructions on how to perform the annual calibration for the BC-8000. The calibration is broken down into the following categories: Verify timing accuracy Calibrate Cutoff at 10.00 Volts Calibrate Cutoff at 20.00 Volts Calibrate 0.5A, 2.0A, 8.5A, 15A, and 55A calibration points Calibrate the Seiko clock chip (This is optional. The clock is calibrated at the factory and requires adjustment if the time keeping accuracy is unacceptable. The date and time are set to the current PC time by the BC-8000 report utility every time test results are transferred from the BC-8000 to the report utility) Calibrate the charger voltages and currents. Page 1

TEST EQUPMENT REQUIRED: 1. CF1 test device (this is available from Concorde). 2. 12 Volt 55 Amp or greater fully charged battery used for 0.5A, 2.0A, 8.5A, 15A, and 55A 3. 24 Volt 1 Amp or greater fully charged battery or a 24 Volt 1 Amp regulated power supply used to calibrate voltage and verify timing accuracy. 4. Calibrated Fluke 87 DMM or equivalent. (for 450 Hz measurement) 5. Calibrated 5½ digit DMM Fluke 8840A or equivalent. 6. Calibrated Frequency counter with 1 second period measurement and 7 digit or more readout. 7. PC computer with BC Report utility software installed. 8. Anti-static wrist strap. 9. 24 Volt 25 Amp or greater battery to calibrate charger. PROCEDURE: INITIAL SETUP: 1. Photo copy the Calibration Work Sheet on page 8. 2. Attach anti static wrist strap to chrome fan guard and your wrist. 3. Remove the 8 lid screws. Remove the lid. 4. Connect CF1 test device ring terminals to the 24V power source. RED lead to the positive terminal of the power source and the BLACK lead to the negative terminal of the 24V power source. 5. Verify unit is OFF. Connect the BC-8000 to the VOLTAGE OUTPUT CONNECTOR on the CF1 test device. It is the connector that has two #16 AWG smaller wires. 6. Locate the display circuit board located on the inside of the front panel and REMOVE the plastic jumper shunt across pins 1 & 3. See Figure 1 for the location of the jumper shunt. VERIFY TIMING ACCURACY MEASUREMENT: 1. Set Fluke 87 DMM to measure frequency. 2. Use shielded test probe wires or twist non shielded probe wires together to maximize measurement accuracy. 3. Connect DMM minus - to TP1 on the display circuit board. See figure 1 for the location of TP1 and TP2. 4. Connect DMM plus + to TP2 on the display circuit board. 5. Turn ON the BC-8000 power switch. 6. After the sign on message the unit will display Code:. The UP and DOWN buttons will increase or decrease the digit's number value. The NEXT button will advance to the next digit. Using the UP, DOWN, and NEXT buttons, enter the code of 1234. Press NEXT. 7. The unit will display CAL VOLTS. Press UP 2 times. The unit will display CHK 450Hz. Press NEXT. Page 2

8. Verify the Fluke 87 DMM frequency is 449.8 Hz to 450.2 Hz (450 Hz ±0.2 Hz). 9. Write the Frequency measurement reading into the calibration work sheet. (450 Hz) If the frequency is not 449.8 Hz to 450.2 Hz, the unit must be sent in for repair. The crystal oscillator should provide an accurate time base for the entire life of the BC-8000. If the unit fails this test, it is an indication that the crystal time base may have been damaged. 10.Turn OFF the BC-8000 power. ADJUST 10.00 and 20.00 VOLT CUTOFF: 1. Set DMM to a voltage range of 0-200 Volts. 2. Connect DMM minus - to VOLT minus - test point on the CF1 test device. 3. Connect DMM plus + to VOLT plus + test point on the CF1 test device. 4. Turn ON the BC-8000 power switch. 5. After the sign on message the unit will display Code:. The UP and DOWN buttons will increase or decrease the digit's number value. The NEXT button will advance to the next digit. Using the UP, DOWN, and NEXT buttons, enter the code of 1234. Press NEXT. 6. On the CF1 test device, select the 10 Volt range and turn the voltage adjustment knob to obtain a voltage of 10 volts ±0.010 volts on your DMM. 7. The unit will display: CAL VOLTS. Press NEXT. The unit will display CAL Volts? 10, press NEXT. The unit will display the voltage on the right and the word ADJUST. 8. Use the UP and DOWN buttons to make the display read 10.00. Press NEXT to save the calibration data. Wait for the word SAVED to appear then disappear. 9. Write the DMM Voltage reading into the calibration work sheet. (10 volt cutoff) 10.On the CF1, select the 20 Volt range and turn the voltage adjustment knob to obtain a voltage of 20 volts ±0.010 volts on your DMM. 11.The unit will display: CAL VOLTS. Press NEXT. Press UP. The unit will display CAL Volts? 20, press NEXT. The unit will display the voltage on the right and the word ADJUST. 12.Use the UP and DOWN buttons to make the display read 20.00. Press NEXT to save the calibration data. Wait for the word SAVED to appear then disappear. 13.Write the DMM Voltage reading into the calibration work sheet. (20 volt cutoff) 14.Turn OFF the BC-8000 power. 15.Unplug the BC-8000 from the CF1 VOLTAGE OUTPUT CONNECTOR. 16.Disconnect the CF1 test device ring terminals from the 24 volt power Page 3

source. CALIBRATE AMP SETUP: 1. Connect the CF1 test device ring terminals to a fully charged 12 Volt 55 Amp battery. RED lead to the positive terminal of the battery and the black lead to the negative terminal of the battery. 2. Set DMM to a range of 0-200 mv. 3. Connect 5 ½ digit DMM minus - to the AMP minus - test point on the CF1 test device. Use caution as the DMM meter leads will be connected to a high amperage battery. 4. Connect 5 ½ digit DMM plus + to the AMP plus + test point on the CF1 test device. 5. Verify the BC-8000 is OFF. 6. Connect the BC-8000 to the HIGH AMPERAGE CONNECTOR on the CF1 test device. It is the connector wired with two #6 heavy duty wires. 7. The scale factor is 1 A/mV. Example: when the 5 ½ digit DMM reads 10.000 mv, it is indicating 10.000 Amps. 0.500 mv is indicating 0.500 Amps. CALIBRATE 0.5A, 2.0A, 8.5A, 15A, and 55A 1. Keep hands, arms, test leads, and clothing away from the FAN BLADE. The FAN BLADE will rotate rapidly during the amp 2. Turn ON the BC-8000 power switch. 3. After the sign on message the unit will display Code:. The UP and DOWN buttons will increase or decrease the digit's number value. The NEXT button will advance to the next digit. Using the UP, DOWN, and NEXT buttons, enter the code of 1234. Press NEXT. 4. The unit will display: CAL VOLTS. Press the UP button until the LCD screen displays CAL AMP. Press NEXT, the unit will display Volts? 12. Press NEXT, the unit will display Amps: 0.5. 5. Press NEXT to start the process. The FAN will start. 6. Press the UP or DOWN button to obtain a DMM reading of 0.497 to 0.503 mv (0.500 mv ±0.003 mv). Allow a few seconds for the reading to settle. Repeat this step to get as close as possible to 0.500 mv. 0.500 mv is 0.5 Amps actual current flow out of the battery. 7. Write the mv reading into the calibration work sheet. ( 0.5 Amp calibration) 8. Press NEXT to save the calibration data into memory. Wait for the word SAVED to appear then disappear. 9. Press the UP button to select Amps: 2.0. 10.Press NEXT to start the process. The FAN will start. 11.Press the UP or DOWN button to obtain a DMM reading of 1.996 to 2.004 mv (2.000 mv ±0.004 mv). Allow a few seconds for the Page 4

reading to settle. Repeat this step to get as close as possible to 2.000 mv. 2.000 mv is 2 Amps actual current flow out of the battery. 12.Write the mv reading into the calibration work sheet. (2 Amp calibration) 13.Press NEXT to save the calibration data into memory. Wait for the word SAVED to appear then disappear. 14.Press the UP button to select Amps: 8.5. 15.Press NEXT to start the process. The FAN will start. 16.Wait 30 seconds. 17.Press the UP or DOWN button to obtain a DMM reading of 8.483 to 8.517 mv (8.500 mv ±0.017 mv). Allow a few seconds for the reading to settle. Repeat this step to get as close as possible to 8.500mV. 8.500 mv is 8.5 Amps actual current flow out of the battery. 18.Write the mv reading into the calibration work sheet. (8.5 Amp calibration) 19.Press NEXT to save the calibration data into memory. Wait for the word SAVED to appear then disappear. 20.Press the UP button to select Amps: 15.0. 21.Press NEXT to start the process. The FAN will start. 22.Wait 30 seconds. 23.Press the UP or DOWN button to obtain a DMM reading of 14.970 to 15.030 mv (15.000 mv ±0.030 mv). Allow a few seconds for the reading to settle. Repeat this step to get as close as possible to 15.000mV. 15.000 mv is 15 Amps actual current flow out of the battery. 24.Write the mv reading into the calibration work sheet. (15 Amp calibration) 25.Press NEXT to save the calibration data into memory. Wait for the word SAVED to appear then disappear. 26.Press UP to select Amps: 55.0. 27.Press NEXT to start the process. The FAN will start. 28.Wait 30 seconds. This will allow the current sensor to reach operating temperature. 29.Press UP or DOWN button to obtain a DMM reading of 54.890 to 55.110 mv (55.00 mv ±0.110 mv). Allow a few seconds for the reading to settle. Repeat this step to get as close as possible to 55.00 mv. 30.Write the mv reading into the calibration work sheet. (55 Amp calibration) 31.Press NEXT to save the calibration data into memory. Wait for the word SAVED to appear then disappear. 32.Turn OFF the BC-8000 power. Page 5

Calibrate Charger (Optional): 1. Attach anti static wrist strap to chrome fan guard and your wrist. You will need a discharged 24V battery that can accept a 25A charge, a CF1 current shunt, and a DMM. All Voltage OUT and DISP menu items require the output of BC-8000 connected to a DMM and can never be connected to a battery as the unit being calibrated will output various voltages. All Amperage OUT and DISP menu items require a battery and a DMM connected to the CF1 current shunt. Note: the current flow is into the battery instead of out of the battery which causes the polarity of the current flow to be reverse of doing a capacity test. 2. Connect the DMM to the grey connector. After the sign on message the unit will display Code:. The UP and DOWN buttons will increase or decrease the digit's number value. The NEXT button will advance to the next digit. Using the UP, DOWN, and NEXT buttons, enter the code of 1234. Press NEXT. The unit will display CAL CHARGER. Press NEXT. Press UP. The unit will display OUT 3.0V. Press NEXT. Press the UP or DOWN button so your DMM reads as close as possible to 3.0V. Note: 2.9V is closer to 3.0V than 3.2V. If the difference is equal choose the higher voltage. Press NEXT to save the 3. Press UP to select DISP 3.0V. Press NEXT to start. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 3.0V on the LCD display. Press NEXT to save the 4. Press UP to select OUT 7.9V. Press NEXT to start. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 7.9V on the DMM. Press NEXT to save the 5. Press UP to select DISP 7.9V. Press NEXT to start. Wait for the possible to 7.9V on the LCD display. Press NEXT to save the 6. Press UP to select OUT 8.0V. Press NEXT to start. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 8.0V on the DMM. Press NEXT to save the 7. Press UP to select DISP 8.0V. Press NEXT to start. Wait for the possible to 8.0V on the LCD display. Press NEXT to save the 8. Press UP to select OUT 16.9V. Press NEXT to start. Wait for the possible to 16.9V on the DMM. Press NEXT to save the 9. Press UP to select DISP 16.9V. Press NEXT to start. Wait for the possible to 16.9V on the LCD display. Press NEXT to save the Page 6

10.Press UP to select OUT 17.0V. Press NEXT to start. Wait for the possible to 17.0V on the DMM. Press NEXT to save the 11.Press UP to select DISP 17.0V. Press NEXT to start. Wait for the possible to 17.0V on the LCD display. Press NEXT to save the 12.Press UP to select OUT 36.0V. Press NEXT to start. Wait for the possible to 36.0V on the DMM. Press NEXT to save the 13.Press UP to select DISP 36.0V. Press NEXT to start. Wait for the possible to 36.0V on the LCD display. Press NEXT to save the 14.The next set of calibration steps require measuring the current flow into a battery. Connect you DMM in series between the battery negative and charger negative. Use the 200mA connections on you DMM to get the maximum resolution. Connect a alligator clip wire between the battery plus and charger plus. Set your DMM to the 200mA scale. Press UP to select OUT 0.05A. Press NEXT. Wait for the possible to 50mA on the DMM. Press NEXT to save the 15.Press UP to select DISP 0.05A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 0.05A on the LCD display. Press NEXT to save the 16.Reconnect you DMM to use the 10A measurement. Press UP to select OUT 2.5A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 2.5A on the DMM. Press NEXT to save the 17.Press UP to select DISP 2.5A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 2.5A on the LCD display. Press NEXT to save the 18.Press UP to select OUT 2.6A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 2.6A on the DMM. Press NEXT to save the 19.Press UP to select DISP 2.6A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 2.6A on the LCD display. Press NEXT to save the 20.The next steps require the CF1 calibration shunt. Connect the battery to the CF1 calibration shunt. Connect the DMM to the CF1 and select the 200mV scale. 1mV will be 1A. Press UP to select OUT 25.0A. Press NEXT. Wait for the relays to sequence. Press UP or DOWN button to get as close as possible to 25.0A. Press NEXT to save the 21.Press UP to select DISP 25.0A. Press NEXT. Wait for the relays to Page 7

sequence. Press UP or DOWN button to get as close as possible to 25.0A on the LCD display. Press NEXT to save the Charger menu item and associated BC-8000 output connector hook up. These are the charger menu items and what you should do at each menu item. EXIT OUT 3.0V connect to volt meter (do not connect to a battery) DISP 3.0V connect to volt meter (do not connect to a battery) OUT 7.9V connect to volt meter (do not connect to a battery) DISP 7.9V connect to volt meter (do not connect to a battery) OUT 8.0V connect to volt meter (do not connect to a battery) DISP 8.0V connect to volt meter (do not connect to a battery) OUT 16.9V connect to volt meter (do not connect to a battery) DISP 16.9V connect to volt meter (do not connect to a battery) OUT 17.0V connect to volt meter (do not connect to a battery) DISP 17.0V connect to volt meter (do not connect to a battery) OUT 36.0V connect to volt meter (do not connect to a battery) DISP 36.0V connect to volt meter (do not connect to a battery) OUT 0.05A connect to a 24V battery DISP 0.05A connect to a 24V battery OUT 2.5A connect to a 24V battery DISP 2.5A connect to a 24V battery OUT 2.6A connect to a 24V battery DISP 2.6A connect to a 24V battery OUT 25.0A connect to a 24V battery (requires a discharged battery) DISP 25.0A connect to a 24V battery (requires a discharged battery) Page 8

CALIBRATE SEIKO CLOCK CHIP (OPTIONAL): 1. Turn OFF the BC-8000. 2. Connect the USB cable to the PC computer. Start the BC-8000 report utility. Press the Set Date and Time button. The software will search for the BC-8000 and establish a connection. Enter invalid report into the battery serial number rectangular box. Click your mouse on the rectangular box. The program will expand. 3. Set frequency counter to measure a 1 second period with 1 micro second resolution. Connect the frequency counter TP3 and TP1 gnd. 4. In the BC-8000 report utility Measure Period rectangular box, enter 1.0 and press the Set Clock Calibration button. This will turn off clock correction and allow you to measure the unadjusted 1 second clock at TP3. Press the Get Clock Calibration button and verify it returned a 00 in the blue box. 5. Read the period on your frequency counter. The acceptable adjustment range is 0.999805 to 1.000192 seconds. Enter the value into the Measured Period rectangular box. The entry requires precision down to 1 micro second. Press the Set Clock Calibration button. Wait 1 second. Press the Get Clock Calibration button and if you entered a correction value other than 1.000000, it should return something other than 00. The only time it will return 00 is if you entered a period of 1.000000 seconds. (Values very close to 1.000000 may return with a correction value of 00). 6. Your clock is now calibrated. Every 20 seconds, the Seiko clock chip will correct its crystal time base by adding or subtracting a few cycles. If you entered a correction value that is between 0.999805 to 1.000192 seconds and is not 1.000000, you will see the period change every 20 seconds. A correction value of 1.000000 causes no correction as it is the perfect value that requires no correction. 7. Remove USB cable. Page 9

FINALIZATION: 1. Attach anti static wrist strap to chrome fan guard and your wrist. 2. Locate the display circuit board located on the inside of the front panel and REINSTALL the plastic jumper shunt across pins 1 & 3. See Figure 1 for the location of the jumper shunt. 3. Turn ON BC-8000 power. After sign on message, if the user menu item is showing the unit is ready for use. Turn OFF BC-8000 power. 4. Disconnect the BC-8000 from the CF1 calibration device. 5. Disconnect the CF1 calibration device from the battery. 6. If this wasn t done already, write the calibration due dates of all equipment used in work sheet. Attach a photo copy of the calibration certificates of all of the equipment used in this procedure to the work sheet. 7. Write the BC-8000 serial number in work sheet. The serial number is located on the inside rear panel. 8. Sign and date the calibration work sheet. 9. Install the lid and attach it with the 8 screws. Figure 1 (BC-8000 will not have the IC chip U2 shown on the left) Page 10

CALIBRATION WORK SHEET RECORDS - BC-8000 S/N Item name Measured value 450 Hz 449.8-450.2 Hz 10.00 Volt cutoff 9.90-10.10 Volts 20.00 Volt cutoff 19.80-20.20 Volts 0.5 Amp calibration 0.497 to 0.503 mv 2.0 Amp calibration 1.996 to 2.004 mv 8.5 Amp calibration 8.483 to 8.517 mv Specification Pass Fail 15 Amp calibration 14.970 to 15.030 mv 55 Amp calibration 54.890 to 55.110 mv Digital Multi Meter used to measure 450 Hz calibration expiration date: Digital Multi Meter used to calibrate Voltage and Amperage calibration expiration date: If Seiko Clock chip was calibrated: Frequency counter calibration expiration date: Attach a photo copy of the calibration certificates of all of the equipment used to in this procedure to this work sheet. BC-8000 Calibration performed as specified by: Date: Page 11