T U R B I N E G A S M E T E R



Similar documents
Fluxi 2000/TZ Turbine Gas Meter

TURBINE GAS METER SM-RI-X TURBINE GAS METER EM-SMRI-X-17 NATIONAL ACCREDITATION OF CERTIFICATION BODIES

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

How To Use A Corona Mdi 110

Turbine Flow Sensor Series VTR

Thermal flow sensors TA10-ZG8c and TA10-ZG9c for exact and stable, long-term measuring of lower flow velocities (Laminar Flow)

The number of revolutions is proportional to the passed volume. The rotation is transferred to a mechanical index, which indicates this volume.

Technical Specification. Generating Set with Waukesha engine burning natural gas

How To Use A Flowmeter

Air Eliminators and Combination Air Eliminators Strainers

3/2-, 5/2- and 5/3-way Solenoid Valves for process pneumatics

Control- and feedback head for integrated mounting on Robolux valves Type 2036

Bulk water meters. All that counts. EnergyMetering

PRESSURE REGULATOR SERIES R 4100 R 2100 R 5100 VIRTUAL CATALOGUE FAIL TO OPEN REGULATOR BALANCED VALVE HIGH FLOW COEFFICIENT

COMBIMASS. Technical Data COMBIMASS eco-bio +

Flow switches Series AD/VH

1805 Series Relief Valves

HOW TO SIZE A PRESSURE REGULATOR

Micro Motion 3098 Gas Specific Gravity Meter

Thermal Mass Flow Meter (Series MST)

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection.

2-Port Self-acting Temperature Control Valve Selection for Heating and Cooling Applications

1. A belt pulley is 3 ft. in diameter and rotates at 250 rpm. The belt which is 5 ins. wide makes an angle of contact of 190 over the pulley.

Fundamental Principles of Rotary Displacement Meters

Floating Switch FTS 20 Füllstandgrenzschalter Level Limit Switch Détecteur de niveau

FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters

Pressure Switch for gas and air

EVAC ONLINEVAC TECHNICAL DATA EVAC ONLINEVAC 16, 24 V DC. Required maintenance space 700 mm. Required maintenance space 200 mm 360 Ø13

Instruction Manual. T3111 T3111L T3111P T3111Ex

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Seated valves (PN 16) VF 2 2-way valve, flange VF 3 3-way valve, flange

IFC 070 Technical Datasheet

SHARKY 775 COMPACT ENERGY METER ULTRASONIC

VAD Variable Area Desuperheaters

Custody Transfer Measurement. with the V-Cone Flowmeter

Certified according to DIN EN ISO 9001 Technical Datasheet KVO Series Vortex Flow Meter

Valve Proving System VPS 504 Series S06 (120 VAC) Series S05 (24 VDC)

Kompressoren. Adsorption Dryer AD

Fig B36 Stainless Steel 'T' Type or Basket Type Strainers

289 Series Spring-Loaded Relief Valves

SHARKY 775 COMPACT ENERGY METER ULTRASONIC

Pressure transmitter For general industrial applications Model A-10

The Unique Accelabar Flow Meter

For Multi-Parameter Meters see mvx

THE HUMIDITY/MOISTURE HANDBOOK

CORONA E / FLYPPER ELECTRONIC METER MULTI-JET

VAD. Variable Area Desuperheaters

FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases

MEMBRANE VACUUM MINI PUMPS

Ferrostat Speed Sensor Series DSF Explosion Proof Versions EEx

OPTIMASS 1400 Technical Datasheet

Electromagnetic Flow Meter

VSO MAX HP High Flow Proportional Valve

User s Manual Laminar Flow Elements

Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors

PRESSURE REDUCING VALVES

WaterMAG. Electromagnetic Flowmeter for Water Applications. Model MCB10A OVERVIEW APPLICATIONS FEATURES. No. SS2-MCB

SERIES ASM NEOPRENE/EPMD FLANGED SINGLE SPHERE CONNECTOR CONNECTORS. Pressures to 225 PSIG (15.51 barg) Temperatures to 230ºF (110ºC)

Turbine Flowmeters for Gas Applications

CO MPa (abs) 20 C

Linear modules Lifting units Rotary modules Grip modules Inductive proximity switches Plug connectors

System 6000 Electropneumatic Converter for Direct Current Signals Type 6116 i/p Converter

Energy and Flow Measurement for Hydronic Systems

FUNDAMENTALS OF GAS TURBINE METERS John Gorham Sensus Metering Systems 805 Liberty Blvd. DuBois, PA 15801

Series 7500 Torque Sensor

Typical Aluminum GFM Mass Flow Meter

Chapter 3.5: Fans and Blowers

T1-40 T1-40BF T1-40V T1-38 T1-38BF T1-38V

SWIMMING POOL HEAT PUMP Owners Manual

SMARTY 160 POWER page 4-5 SMARTY 180 / 220 XL page 4-6 SMARTY TX 250 page 4-7. SMARTY TX 160 Alu page 4-8 SMARTY TX 220 Alu page 4-9

M/MR Modulating Valves TABLE OF CONTENTS DESCRIPTION SPECIFICATIONS. For Atmospheric, Infrared, and Direct Fired Burners

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER

SIGMA PUMPY HRANICE CENTRIFUGAL MULTISTAGE HORIZONTAL PUMPS

D.C. Motors. Products and specifications subject to change without notice.

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

Selecting and Sizing Ball Screw Drives

Size : Ends : Min Temperature : Max Temperature : Materials : Cast iron body

Digital inductive conductivity transmitter

STEAM TURBINE 1 CONTENT. Chapter Description Page. V. Steam Process in Steam Turbine 6. VI. Exhaust Steam Conditions, Extraction and Admission 7

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

1 DESCRIPTION OF THE APPLIANCE

Pilot Operated Pressure Reducing Valve PRV 47 (Steel) PRV47I (St.Steel)

G234X-55 Conventional Gas Boiler

CRN A-F-G-E-HQQE 3x400D 50 HZ

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater

MAKING MODERN LIVING POSSIBLE. Thermostatic valve AVTA. Technical brochure

HFS and HFD Series. High Flow Single and Duplex Filters Max l/min. 10 bar. Heavy Duty Reliability. Applications: Contact Information:

VdS en. VdS Guidelines for water extinguishing systems. Non-return valves. Requirements and test methods. VdS en : (01)

red-y smart series product information Thermal Mass Flow Meters and Controllers for Gases

DS 600. A contact free flux gate based current measurement sensor 600A rms

Thermostatic valve Type AVTA

Series 6000 Torque measured metal bellow coupling

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.

Series 240 Type and Type Pneumatic Control Valves Type 3244 Three-way Valve

Micro Motion Model D and DL Coriolis Flow and Density Meters

Proportional pressure regulators VPPE

Q&A Session for Advanced Ball Screws 102: Troubleshooting for Design Engineers

Duplex Filters 40 FLDK 0008(C)-0120(C)

Operating instructions Mechatronic flow sensor. SBY33x /06 03/2011

Transcription:

TURBINE GAS METER

TURBINE GAS METER CGT 1 2 3 4 5 6 7 Design and function page 2 General technical data page 3 Measurement outputs page 4 Dimensions and weights page 5 Performance page 7 Pressure loss page 8 Installation and operation recommendations page 8 CGT series turbine gas meters are flow meters designed to measure quantity of gases. The meters are mainly used for gas flow ranges from 6.5 up to 60 m 3 /h. The CGT series gas meters are applied in measurement systems where high accuracy is required: transportation of natural gas primary and secondary measurements control metering of the natural gas and non aggressive technical gases in industry flow measurement for technical purposes The meters may be applied for measurements of aggressive gases. Any application of that kind should be individually proved and agreed between the client and COMMON S.A. 1

1 DESIGN AND FUNCTION The turbine gas meter measures the quantity of gas basing on the flow principle. The gas flows through an integrated flow conditioner, which distributes the flow proportionally in the annular slot and guides it to the turbine wheel. The wheel is driven by the gas flow, and the angular velocity of the rotation is proportional to the gas flow rate. The energy consumption, perceived as pressure loss, is reduced to absolute feasible minimum due to the application of the flow conditioner, highest precision ball barings, most accurate tolerances of all measuring parts and their proper alignement. The rotary motion of the turbine wheel is transferred mechanically by gear wheels, and the incorporated gas tight and hermetic magnetic coupling, to the index unit, which is mounted on the top of the body, and shows the operating volume on the totalizer. The basic components of the COMMON CGT series turbine gas meter are as follows: pressure resistant meter body inlet flow conditioner measuring cartridge with the turbine wheel magnetic coupling as the transferring element between measuring cartridge and the index index lubricating system (some meters are provided with self lubricating bearings) Index head Magnetic coupling Meter body Flow conditioner 2 Measurement cartridge

2 GENERAL TECHNICAL DATA DESIGN & TECHNICAL DATA table 1 mm DN inch G Maximum flow Q min [m 3 /h] Minimum flow Q min [m 3 /h] LF pulse rate U a [m 3 /pulse] 2 40 65 65 0 6 0,1 0 0 8 80 3 0 2 2 400 13 1 0 2 13 0 4 2 400 400 6 32 1 1 6 400 6 00 6 00 1 32 80 1 6 00 1 0 8 00 1 1 80 130 00 1 80 2 1 4000 130 0 12 1 4000 4000 60 130 0 3 pressure rating: nominal diameter: PN to PN1, ANSI1 to ANSI other rates on request DN up to DN standard range, other on request meter bodies: cast iron or carbon steel details in table 4 flow rangeability: upstream pipe: 6.5 to 6 0 m 3 /h other on request 1: minimum at atmospheric pressure Some smaller size meters have reduced ranges. minimum 2 x DN; meters meet the requirements of the OIML R32 89 Annex A temperature range: gas temperature C to +60 C ambient temperature 25 C to +70 C allowed medias: see table 2 operating position: horizontal or vertical 3

measurement accuracy: EU requirements and better guaranteed at least: 0.2 Qmax Qmax < ± 1% Qmin 0.2 Qmax < ± 2% fig.2: Measurement error typical curve at low pressure (average 1 bara) green curve at high pressure (over 5 bara) blue curve Error % +2 +1 0 1 2 Q min,ρ Q min Q t Q max Q Gas Chemical symbol (formula) Density ρ [kg/m 3 ] Density related to air Gas meter execution Argon Ar 1,66 1,38 Butane C 4 H 2,53 2, Carbon dioxide CO 2 1,84 1,53 Carbon monoxide CO 1, 0,97 Ethane C 2 H 6 1,27 1,06 Ethylene C 2 H 4 1,17 0,98 Helium He 0,17 0,14 Methane CH 4 0,67 0,55 Natural gas Nitrogen Propane Acetylene Hydrogen Air N 2 C 3 H 8 C 2 H 2 H 2 ~0,75 1, 1,87 1,09 0,084 1, ~0,63 0,97 1,56 0,91 0,07 1,00 special IIC special IIC table 2: Physical properties of most popular gases that may be measured with the CGT turbine gas meters density at 1,325 kpa and at C 3 MEASUREMENT OUTPUTS 4 PRESSURE AND TEMPERATURE OUTPUTS The operating pressure (reference pressure) can be taken from the pressure taps, marked pr, located on both sides of the meter body. The meters DN0 and larger can be optionally equipped with two temperature taps for the measurement of the gas temperature. PULSE SENSORS The mechanical index unit indicates the actual volume of the measured gas at operating temperature and operating pressure. It can be rotated axially by 3 in order to facilitate the readings and enable easier connection of pulse sensor plugs. The index unit is provided with one low frequency LFK reed contact pulse transmitter, as a standard. On request the index may be equipped with: LFI inductive pulse sensor (NAMUR) HF inductive pulse sensor (NAMUR)

MEASUREMENT OUTPUTS HF Ui = V Dc Ii = 25 ma Pi = mw Li = μh Ci = 30 nf LFI Ui = 15,5 V DC Ii = 52 ma Pi = 9 mw Li 40 μh Ci = 28 nf LFK, AFK Ui = 15,5 V DC Ii = 52 ma Pi = 9 mw Li 0 Ci 0 fig. 2. Location of measurement outputs (top view) table 3: Permissible supply parameters of intrinsically safe circuits. The CGT turbine gas meters may be provided with up to pulse sensors for DN0 DN and up to 8 pulse sensors for DN DN80 LFK low frequency reed contact pulse sensor LFK1, LFK2 LFI low frequency inductive pulse sensor LFI1, LFI2 HF inductive pulse sensor in the index unit HF1, HF2 HF inductive pulse sensor over the turbine wheel HF3, HF4 HF inductive pulse sensor over the reference wheel HF5, HF6 AFK antifraud reed contact AFK The turbine wheel, as a standard, is made of aluminium. This allows to provide each CGT turbine gas meters with HF inductive pulse sensors. There are no extra costs due to the replacement of the turbine wheel. fig. 3 Pulse sensor pin numbering in sockets 1 and 2 installed in the index Pin No 1 4 2 5 3 6 Socket 1 pulse sensors LFK 1 (standard) LFI 1 HF 1 or AFK Socket 2 pulse sensors LFK 2 LFI 2 HF 2 The sockets match the TUCHEL plug No C091 31H006 0 2 fig. 4 Pulse sensor pin numbering in socket of the HF pulse transmitter installed in the meter body Pin No 3 4 HF over turbine wheel HF 3 HF 4 HF over reference wheel HF 5 HF 6 The sockets match the TUCHEL plug No C091 31D004 0 2 5

4 DIMENSIONS AND WEIGHTS Overall dimensions and weights of CGT turbine gas meters are shown in Table 4 fig.5 Dimensions of the CGT turbine gas meter table 4 DN L mm inch mm mm mm mm mm 2 1 3 42 58 60 80 3 240 6 80 95 0 4 2 1 124 1 6 4 247 180 155 0 8 270 212 240 A B 125 2 7 298 270 330 12 900 323 3 C E, F* Flange Weight ANSI PN cost iron steel kg kg * size E is valid for meters PN, PN and ANSI 1, size F is valid for meters PN, PN, PN1, ANSI and ANSI 1 226 146 222 157 233 185 261 2 282 232 308 258 345 1 1 1 1 1 1 1 1 1 1 1 1 1 1 8,5 9,5 19,5 25 26,5 48 47 11 11,5 14 14 25 24,5 27,5 28 32 33,5 34,4 43 40 56 62,5 62,5 80,5 84 6 80 86 1 128 153 142 147 190 6 271 215 235 290 360 6

5 PERFORMANCE DIMENSIONS & PERFORMANCE The meter measures the actual quantity of gas flowing at operating conditions (pressure and temperature). This volume is displayed on the index as actual volume in m 3. The most important factor of the meter size selection (nominal diameter) is the expected minimum and maximum gas flow at the operating conditions. According to standards the turndown ratio is determined at atmospheric conditions (p = 1.01325 bar). The measurement range increases with the increase of operating pressure. Q min value decreases, and it may be calculated from the following formula: Q min m = Q min ρ a / ρ m [m 3 /h] The operating density ρ m may to be determined as follows: ρ m (p + 1) ρ [kg/m 3 ] DEFINITION: Q min m = minimum flow at operating conditions [m 3 /h] Q min = minimum flow of the meter at atmospheric conditions [m 3 /h] p = gauge pressure [bar] ρ = standard density of the gas [kg/m 3 ] see table 2 (for natural gas: 0.75 kg/m 3 ) ρ a = standard density of air [kg/m 3 ] (1.2 kg/m 3 ) In order to convert parameters from base to operating conditions, and vice versa, the following formula may be applied: p s (t V m = V s k V s Z m + 273) T s 273 (p + 1) p m T m DEFINITION: V m V s k Z Z s p s = volume at operating conditions [m 3 /h] = volume at base conditions [m 3 /h] = relative compressibility factor k = Z / Zn = real gas factor at operating conditions = real gas factor at base conditions = base pressure (1.01325 bar) p m p T s T m t m = operating pressure (abs.) at the turbine meter [bar] = operating gauge pressure at the turbine meter [bar] = base gas temperature [K] (273.15K) = operating gas temperature [K] = operating gas temperature [ C] 7

6 PRESSURE LOSS The inevitable pressure loss during the gas flow through the meter is determined at the atmospheric conditions. To determine pressure losses at other, higher pressures, the following formula applies: ρ p+p p 1 = ( ) ( s ) p ρ a p s DEFINITION: p 1 = pressure loss at p p = pressure loss from the diagram, fi g. 6 p = gauge pressure [bar] ρ = standard density of gas [kg/m 3 ] ρ a = standard density of air 1,2 kg/m 3 p s = base pressure (1.01325 bar) fig. 6 Diagram of pressure loss related to ρ = 1,2 kg/m 3 7 INSTALLATION AND OPERATION RECOMMENDATIONS 8 Meters should be shipped in their original package to the place of installation. Meters have to be handled with care and protected against falls, direct infl uence of rain, snow or high humidity. The measured gas should be clean, dry and free from solid impurities. It is recommended that the upstream pipe installation is to be equipped with a filter ( micron). Prior to putting into operation in new installations, it is recommended to install a temporary cone sieve. Prior to installation the upstream and downstream pipe fl anges should be aligned properly. Flange gaskets are to be installed so as not to disturb the gas flow.

The gas fl ow should be in accordance with the arrow placed on the meter body. The piping should be executed so as to avoid unnecessary stresses acting on the meter. When used outdoors the meter should be protected against direct weather infl uence. When starting the gas flow through the installation, the valves should be opened slowly to ensure a gradual increase of pressure. ATTENTION! If valves are opened abruptly or pressure increase takes place in a short time it may happen that the measurement cartridge or the turbine wheel get damaged. ALWAYS REMEMBER TO START UP THE METERS IN A PROPER WAY! COMMON S.A. has continuing program of product research and development. Technical specifications and construction may change due to improvements. This publication serves as general information only, and all specifications are subject to confirmation by COMMON S.A.