OLYMPUS STUDENT MICROSCOPES INSTRUCTIONS ST. HS&HSC MODELS OLYMPUS



Similar documents
Care and Use of the Compound Microscope

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.

Name Class Date Laboratory Investigation 4B Chapter 4: Cell Structure

OPERATING INSTRUCTIONS XJL 101/101A REFLECTED LIGHT METALLURGICAL MICROSCOPE

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012

EXPERIMENT #1: MICROSCOPY

FirstView 3 Reflector Telescope Owner s Manual

THE COMPOUND MICROSCOPE

National Optical & Scientific Instrument Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Chapter 17: Light and Image Formation

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY

Compound microscope (Hund)

Microscope Lab Introduction to the Microscope Lab Activity

PROFESSIONAL REFRACTOR MODEL

A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31

ISO9001 ISO Environmentally-friendly product All optical components in CX41 microscope use lead-free eco-glass.

Operator s Maintenance Manual STEINER BINOCULARS. Military ruggedness and precision 07/03

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

PlaneWave CDK Telescope Instructions CDK12.5, 17, 20 and 24

C) D) As object AB is moved from its present position toward the left, the size of the image produced A) decreases B) increases C) remains the same

VISM Evolution Scope Series

Leica DM300 User Manual

9/16 Optics 1 /11 GEOMETRIC OPTICS

Measuring. User Manual

euromex microscopen bv The Netherlands page1/16

TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions

Be careful not to scratch or hit front edge of the side viewing micro prisms onto hard objects!

How To Use An Asbestos Microscope

MACRO PHOTO LENS. MP-E 65mm f/ ENG. Instruction

30 mm Diameter Main Tube 1-4x24 Models

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

TEC APO140. This telescope is closer to optical perfection more than any instrument I have ever used before. R.Renzi, Italy.

CALIBRATION FOR LAL20X & LAL24X

Lecture 17. Image formation Ray tracing Calculation. Lenses Convex Concave. Mirrors Convex Concave. Optical instruments

3-D Workshop AT A GLANCE:

ST-80 MICROSCOPE With Electronic Eyepiece #6810. User Guide

Installation and use of Millett Tactical TRS-1 and TRS-2 scopes TRS-1 TRS-2. Before starting, make sure firearm is UNLOADED!!

Light and its effects

VISM CQB Scope Series

waves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object

EnduraLAM 150 User Manual

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

The Krasnogorsk-3. Operating Instructions. Sold by NCS Products, New York

INSTRUCTION. Industrial Sewing Machines. No First published : June 1997 Second edition : March 2001

V-500-PRE. 500mm F/8.0 Telephoto Lens INSTRUCTION MANUAL

RAY OPTICS II 7.1 INTRODUCTION

Science In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light

OLYMPUS STUDENT MICROSCOPES IINSTRUCTION MANUAL I CBA& CBB MODELS

Electric Meat Grinder

The Vortex Spitfire 1x Prism Scope

OLYMPUS SYSTEM MICROSCOPE. Model BHT OLYMPUS

M A R C O. CP-670 AUTOMATIC CHART PROJECTOR Instruction Manual

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between

R/T TACTICAL MONOCULAR

Forensic Science: The Basics. Microscopy

Lenses and Telescopes

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

MICROSCOPY MARKETING GUIDE

Protocol for Microscope Calibration

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

Features: Mounting the Optic

LIGHT SECTION 6-REFRACTION-BENDING LIGHT From Hands on Science by Linda Poore, 2003.

Pet hair clipper. Model Diagrams within this manual may not be drawn proportionally.

Exercise 2. The Compound Light Microscope

Instruction Manual Genesis 200 EQ

CrossFireII. 1-Inch Diameter Main Tube RIFLESCOPE MANUAL RIFLESCOPE RIFLESCOPE

RIFLESCOPE RIFLESCOPE MANUAL

Physics 41, Winter 1998 Lab 1 - The Current Balance. Theory

NIGHT VISION MONOCULAR PHANTOM

Triple Threat 3-in-1 Game Table 3 IN 1 GAME TABLE

Handheld USB Digital Endoscope/Microscope

Lesson 26: Reflection & Mirror Diagrams

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

BIO 10 Lab 1 Introduction Pre Lab Test

Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University

Brief description and possible applications of the BAADER HYPERION EYEPIECES

Auto Feed Screwdriver

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

7.2. Focusing devices: Unit 7.2. context. Lenses and curved mirrors. Lenses. The language of optics

MP-4V Heavy Duty Riveter / 39048

SERIE SZR. Stereomicroscopi da laboratorio. 13

A Fishy Tale. Observing the Circulatory System of a Goldfish with a Compound Light Microscope

Microscopes 2014 Microscopes PB Microscopes 1

18-270mm F/ Di II VC PZD for Canon, Nikon (Model B008) mm F/ Di II PZD for Sony (Model B008)

BUTTERFLY SHOWER DOOR INSTALLATION INSTRUCTIONS

Lunette 2 Series. Curved Fixed Frame Projection Screen. User s Guide

Use of the Microscope and Cytology

74 Chapter CO 2 Cleaning

How to make a Galileian Telescope

1. You stand two feet away from a plane mirror. How far is it from you to your image? a. 2.0 ft c. 4.0 ft b. 3.0 ft d. 5.0 ft

Making a reflector telescope

Alignement of a ring cavity laser

MICROSCOPE & OPTICAL CATALOG

Handheld USB Digital Endoscope/Microscope

Transcription:

OLYMPUS STUDENT MICROSCOPES INSTRUCTIONS MODELS ST. HS&HSC OLYMPUS

OLYMPUS MICROSCOPES MODELS ST, HS, HSB, HSC This booklet describes the common features and functions of Models ST, HS. HSB and HSC of biological microscopes for educational use. We always hope to offer the best possible products and for this reason each product we place on the market is accompanied with an inspection certificate. ST HS HSC CONTENTS I. How to Unpack I II. Name ofparts 2 III. Optical Path 3 IV. Construction 4 V. How to Use the Microscope 5 VI. Accessories 7 VII. Important Points in Storing the Microscopes 8

I. How to Unpack 1. Remove the bottom screw from the carrying case with the supplied screwdriver. Keep this screw in the receptacle inside the case. 2. Shift the grayish-tan knob along the groove in the earring handle into the square opening tounlock and open the case. 3. Bring out the microscope from the case and remove the packing and covers. 4. Set up the microscope with the equipment such as objectives. eyepieces. mirror,etc. 5. The removed screw and packing may be rc used for transportation. Tighten the screw when transporting. If the packing is missing, tighten the screw and pack carefully in shipping carlon or the carrying case. Bot tom screw _---- Objective Eyepiece Mirror--~ Space for Spare Eyepiece Immersion oil (HSC only) Piece of wood Receplacle for bottom screw Key -1-

II Name of Parts The principal parts of the microscope are given below with Model HS as an example. (Model HS) Eyepiece -----------:: t-----------eyepiece sleeve Body lube stopper, Coarse adjustment knob----- --------Ilody lube Fine adjustment knob---,..---- Revolving nosepiece <--- Objeclive Arm-- Slage cup--- Slage a------condenser -------- Condenser diaphragm -'-"-=?"r---- Mirror Base -2-

III Optical Path (Model ST) Image on the retina Pupil of the observer's eye ;;.!:>... " F: 0 8 ~, " ~. c :!! ~, c!:> /. -e 4' Body tube---;o.'.' ----':-----Arm Stage Virtural Mirror ------e~~::: -3-

IV Construction 1. Body tube Body tubes can be classified as follows by form and function. Fixed tube Monocular tube Adjustable tube Vertical tube Binocular tube (Trinocular) Fixed tube Adjustable tube The fixed body tube has a constant length, that is, the unchangeable distance between the eyepiece and the objective. In Olympus biological microscopes the tube length is fixed at 160 millimeters. WHh a mechanical tube length of 160mm. if the microscope is focused with a lox objective and Ihen the nosepiece is turned to set a 40X objective in place, it is also focused to such an extent that accurate focusing can be done only by operating the fmc adjustment knob. When the objectives are used on a body lube with a mechanical length other than 160mm, it will be impossible to obtain Ihe rated magnifications, the resolving powers and the par-fol;alty. 2. Revolving nosepiece It is customary for the objectives to be mounted on the nosepiece in 3 manner in which higher powers can be obtained when the nosepiece is turned clockwise with the right hand while looking through the eyepiece. 3. Stage Square fixed stage. Mechanical stage can be filled on HS, HSB and HSC. 4. Focusing mechanism The coarse adjustment mechanism consists of a rack and a pinion by which the body tube is racked up and down. Models HS, HSB and HSC have double-lever fine adjustment mechanism. S. Body tube stopper This stopper is used for preventing the body tube from sliding down farther than a pre-set point. If the stopper is fastened tightly when the body tube has been adjusted at a desired level, the tube can be racked up and down only above the pre-set level but never come down below it. 'This arrangement prevents a damaging collision between the objective and the ~pecimen on the stage, Stopper -4-

V How to Use the Microscope 1. IUumination What is most important in a microscopic work is illumination, because the light source and condenser greatly influence the resolving power of the microscope lenses. For microscope illumination, natural daylight or artificial light source is used. Normally it is sufficient to use natural light coming through a window. However, in some cases such as photomicrography, a specially designed microscope illuminator is used. When daylight is used Avoid the direct rays of the sun. It is most desirable that the microscope is placed by a window facing north. Usually the plane mirror is used but the concave side is used when annoying renections of the window frame or trees come into the field of view. When an illuminator is used A low-voltage bulb is used for microscope illumination. Olympus puts on the market a specially designed microscope illuminator LSD (u!:ing a 6V, SA transformer, optical accessory). When this illumination apparatus is used, the beam of light must be adjusted by adjusting the position of the bulb and the plane side of the mirror should be used. The accessory daylight mter (cobalt color) shohld be inserted into the filter holder. Centering When a microscope illuminator is used, adjust it so that the axis of the illuminating system falls in the center 01 the mirror. Next, rack the body tube up and down SO that it may be roughly in a position in which the microscope is used for observing the specimen. Remove the eyepiece and look through the eyepiece sleeve, and you will see the back lens of the objective. Adjust the mirror SO that the back lens is evenly illuminated. Similar procedures must be adopted when daylight is used. Slide glass and cover glass Ow microscopes are designed to use the slide glass and cover glass of the following thicknesses. Slide glass: 0.8-1.0mm Cover glass: 0.16-0.18mm Use slide and cover glasses that satisfy these requirements. -5-

2. Focusing Afte~ completion of the centering of the illuminating system, place the specimen on the stage and focus the microscope on the specimen. If you move the body tube up and down while looking through the eyepiece. you may jam the objective down on to the specimen, and you could quite easily crack the specimen and the front lens as well. First, use a low power objective, and operate the coarse adjustment knob to lower the body tube until the tip of the objectivecomesclose to thespecimen. All this must be done while lookingatthe microscope from the side. Next, look into the eyepiece and raise the body tube to roughly focus the microscope on the specimen, andfasten thestopperat this point. ext, turn the nosepiece to chilnge to an objective with a desired magnification. Usc the fine adjustment knob to get the specimen sharply in focus_ Move the specimen by hand or operate the mechanical stage to bring the target area of the specimen into the center of the field of view. 3. Change of magnification The magnifying power of the microscope is changed either by changing the eyepiece or by changing the objective. I) Change of the eyepiece Eyepieces with different magnifications arc inserted into the eyepiece sleeve to change the magnifying power of the microscope. In this case, the microscope remains roughly in focus but further focusing is required to obtain a sharply focused image. 2) Change of the Objective When the objective is changed to change the magnification of the microscope, you first focus the microscope with a low-power Objective and then turn the nosepiece to change to a higher-power objective. In Model HS you need only operate the fine adjustment knob to obtain an accurate focus. 4. Resolving Power The resolving power of a microscope depends solely on that of the objective and has nothing to do with the eyepiece. The only function of the eyepiece is to magnify the image resolved by the objective. This means that no matter how much the final magnification may be increased by using a high-power eyepiece, you will never see what is not shown in the image that is formed by the objective. For example, you get the same magnific'dtion of 800X by using a 40X objective and a 20X eyepiece, and also by using a IOOX objective and a 8X eyepiecc. However. the latter combination has a resolving power twice higher and show more details than the former combination. -6-

VI Accessories Microscope Illuminator MODEL LSD A radiatof cover is provided. Provided with a condenser and diffused light, a parallel light and convergent light can be obtained. Can be used for photographing and polarizing microscopes 3S well as all other types of microscope observation. Simple Illuminator MODEL LSK This illuminator is used in place of the mirror of the mjcroscope. Easily detachable ~d directly connectable to a household outlet of electric power supply. 15W bulb is used and there is no need for ;1 t.ransformer. A blue filter is provided. When you place an order for this illuminator, please specify that it is to be used for Model S T or Model lis. ~... - I II. I -- LSK Mechanical Stages Attachable to. the plain stage of a microscope FM- II Graduated mechanical stage Model FM-U Range of operation: 50x25mm Vernier scale: read to 0.1 mm KM Ungraduated mechanical stage Model KM Range of operation: 7Sx2Smm -7-

'IIl Important Points in Storing the Microscopes Dampness is the worst enemy to the microscope. The "sou," of any microscope is the optics. especially the objectives and eyepieces. It is desirable that these lenses are stored in a dessicator in which a drying agent (silicagel or calcium chloride) is placed. The next worst enemy to th~ delicate optical instrument is dust. Dust is apt to gather upon the surfaces of lenses, the condenser and mirror under the stage. Use a very soft brush (with a fine point, not starched) to get rid of dust. When the eyepiece is removed, never fail to put the dust cover on the top of the eyepiece sleeve. If the eyepiece sleeve is leftopen. dust will fall through the tube and gather upon the back lens of the objective and it will greatly impair the performance of the lens. All the moving parts (such as the coarse adjustment mechanism, the condenser, the mechanical stage) are properly lubricated with a special grease so that they may move smoothly. Avoid using watch-oil or machine oil on these parts. A microscope. including the objectives. is such a delicate precision instrument that you never try to disassemble any part of it. If you find something wrong with your microscope, better leave it to the Olympus repair personnel. When you clean the lenses, you must be very careful not to scratch them. When dust enters inside objectives, etc. try to blow it off using a rubber blower or brush it away with a clean feather or the like. As we are continually improving and developing our products, the equipment supplied may not liigrce in all derails with the descriptions and/or illustrations shown in this in lructions. -8-

OLYMPUS OPTICAL CO.. LTD. @~~-9B 43 2 HATAGAYA 2 CHOME, SHIBUYA KU, TOKYO, JAPAN Printed in Japan 73115100