MRI QA Technologist s Tests



Similar documents
MRI QA Technologist s Tests

Breast MRI Quality Control

SITE IMAGING MANUAL ACRIN 6698

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

Phantom Test Guidance

INTRODUCTION. A. Purpose

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

Update on ACR Digital Mammography QC Manual

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Quality Control and Maintenance Programs

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

32-Channel Head Coil Imaging at 3T

Current Industry Neuroimaging Experience in Clinical Trials Jerome Barakos, M.D.

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

5 Factors Affecting the Signal-to-Noise Ratio

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Quality Control of Full Field Digital Mammography Units

Musculoskeletal MRI Technical Considerations

How To Improve Your Ct Image Quality

Prepublication Requirements

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

MQSA Quality Control Manual for Monochrome Displays for Mammography

GE Medical Systems Training in Partnership. Module 12: Spin Echo

MGH Adult Diffusion Data Scanning Protocols

Standardized MRI Protocol for Brain Tumor Clinical Trials. Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Digital Mammography Update: Design and Characteristics of Current Systems

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Quantitative Imaging In Clinical Trials Using PET/CT: Update

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Window Glass Design 5 According to ASTM E 1300

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

MRI SEQUENCES. 1 Gradient Echo Sequence

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Accreditation Is Coming

Kap 8 Image quality, signal, contrast and noise

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

MEDIMAGE A Multimedia Database Management System for Alzheimer s Disease Patients

Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants

Imaging Informatics Ann Scherzinger, PhD, CIIP C-TRIC Lecture Series, May 18, 2011

Physics testing of image detectors

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines

Performance testing for Precision 500D Classical R/F System

Automatic Wheel Profile Measuring Device for Tramway Vehicles

Laser Calibration Check

Tube Control Measurement, Sorting Modular System for Glass Tube

Specifying Laser Scanning Services. A Quantapoint White Paper

Advanced MRI methods in diagnostics of spinal cord pathology

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute

Elfring Fonts, Inc. PCL MICR Fonts

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Super-Resolution Reconstruction in MRI: Better Images Faster?

Microsoft Excel 2010 Charts and Graphs

Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities

Fundamentals of Cone-Beam CT Imaging

Digital Breast Tomosynthesis QC Requirements

Assessing Radiation Dose: How to Do It Right

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

DICOM Correction Item

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Purchasing a cardiac CT scanner: What the radiologist needs to know

Drawing a histogram using Excel

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Sample Table. Columns. Column 1 Column 2 Column 3 Row 1 Cell 1 Cell 2 Cell 3 Row 2 Cell 4 Cell 5 Cell 6 Row 3 Cell 7 Cell 8 Cell 9.

Vibration measurements on future supports of mirrors M3 and M4 (ver. 1.0)

1. General Description

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

Atomic Force Microscope

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

The multitude of symptoms following a whiplash injury has given rise to much discussion because of the lack of objective radiological findings.

All V7 registers support barcode printing, except the Sharp 410/420 1A ROM and that limitation is based upon the register.

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota

NI USB-5681 RF Power Meter Specifications

Transcription:

AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD, DABR Associate Professor Department of Physics, Louisiana State University Department of Radiology, The Ohio State University Outline Background QC importance Technologist s role QC test procedures Table OK? CF and TX Gain Phantom distance HR holes LCD spokes Hard copy image QC Additional QC tests Questions 1

Background Dr. Moriel S. NessAiver s experience 174 yearly performance tests (98 magnets over 3.5 years) 18 (10.3%) without deficiencies 19 (21.3%) with minor deficiencies, not affecting image quality 137 (78.7%) with deficiencies, directly affecting image quality 144 phased array coils (19.2%) with significant problems 22 systems (12.6%) with homogeneity problems 10-20% scanners: excessive RF noise, excessive ghosting, poor gradient calibration, poor hard copy (film), and soft copy performance 1 vendor s TSE with slice thickness 18-23% thicker than specified 1 vendor s TSE with slice thickness 20-25% thinner than specified Daily/Weekly QC Program 2

A QC Team Guide initial QC test Assure correct execution of QC procedures Review QC notebook annually Answer image quality questions Help identify/correct image quality problems Inform image quality problem in clinical reading Decide to continue/postpone pending corrective action Participate in QC program establishment Monitor QC results regularly QC Testing Frequency Technologist s QC testing Weekly: minimum recommended frequency Daily: strongly recommended by ACR More frequently: if problems detected often, equipment unstable, repair, upgrade 3

Record Notebook Examples Data form for weekly QC Template from ACR 2004 MRI QC Manual Examples Record Notebook Examples QC visual checklist Template from ACR MRI QC Manual Examples 4

Table OK? Table OK? A test about the accuracy of the laser localizer If the table positioning system functions properly and the center of the sagittal image of the phantom is within +/-2mm of the central grid structure on the phantom--- ACR Manual 2004 version Frequently ignored by technologists Most technologists misunderstand that it is a check of whether the table move smoothly Laser Light Table OK? on Siemens Scanner Procedure Open the sagittal localizer image User PixelLens to check the top edge of the vertical bars The location should be within ± 2mm < H(ead)2 or F(oot)2 5

Table OK? on GE Scanner Procedure Open the sagittal localizer image User Crosshair to check the top edge of the vertical bars The location should be within ± 2mm < S2 or I2 Table OK? Offline Measurement Philips EBW Open the sagittal localizer image User Crosshair to check the top edge of the vertical bars 6

Table OK? Offline Measurement Philips EBW 5.8 mm (> 2mm) MRI service is required to adjust the laser Central Frequency Drift Rate Definition Center frequency run down over time Drift reasons Windings are not perfect superconductors Eddy current interactions Allowed drift level 1.5 ppm per week (ACR MRI QC guide, 2004) Corrective action Ask medical physicist or MRI scientist to re-check Ask service to monitor the units to ensure 1. B 0 field within RF transmit/receive frequency range 2. Drift rate does not increase 7

Center Frequency on GE scanner GE Signa Horizon 1.5T Place the phantom Click Auto Prescan before run the localizer sequence Record TG (transmit gain) and AX (central frequency) Center Frequency on GE scanner GE Signa Horizon 1.5T Text page Series page Record central frequency and transmit gain 8

Center Frequency on Siemens Scanner Procedure Before run a sequence, e.g. ACR T1 axial Options Adjustments Record central frequency and transmit gain Distance Measurement Set display window width and level/center To have the edges at the half-maximum value of the signal intensity Skipped by some technologists First set window width to 0 or 1 Adjust level/center about ½ white and ½ black, record level value window width = recorded level value window level/center = ½ recorded level value W = 606 C = 303 W = 1 C = 256 W = 256 C = 128 W = 768 C = 384 W = 1 C = 500 W = 500 C = 250 9

Distance Measurement: Sagittal Sagittal Image Measure phantom length < 148 ± 2 mm Avoid going through the black bar Sagittal Loc Length 145.3 mm 146.4 mm 147.0 mm Distance Measurement: Problem 10

Distance Measurement: Solution Receiver bandwidth (rbw) Low rbw for high SNR May cause measurement failure Due to field inhomogeneity Bandwidth Increase rbw may improve accuracy 15.6 khz 31.3 khz 62.5 khz Distance Measurement: Axial Axial slice #5 Diameters of the phantom One vertical; One horizontal < 190 ± 2 mm Examples: 11

Distance Measurement: Problem Vertical distance 194 mm Out of spec Technologist chose Slice #1, instead of #5 Axial Curved border Sagittal Out of spec Coronal Distance Measurement: Solution Turn on geometric distortion correction 194 mm 192 mm 2D Geometric Correction Curved border Sagittal Straight border Sagittal 12

Spatial Resolution Measurements Procedure Magnify image x2-4 Set a small window width Increase window level to identify hole arrays Record smallest hole size An example (1.0) Rows 1 through 4 of the UL array are resolved (1.1) no columns or column 4 of the LR array could be resolved at 1.0 (Out of spec) Zero filled interpolation (ZIP) Original image matrix: 256x256 Resolution: ZIP FT Unfiltered Image 256x256 k-space data 256x256 Courtesy of Dr. Moriel S. NessAiver 13

Zero filled interpolation (ZIP) ZIP512: zero fill up to 512x512 Better image resolution Cost 1. Slower reconstruction time 2. Longer to transfer 3. Takes up more disk space Resolution: ZIP FT Unfiltered Image 512x512 Courtesy of Dr. Moriel S. NessAiver k-space data 512x512 Resolution: ZIP No ZIP UL: 1.0 LR: 1.1 Not passing ACR criteria with ZIP512 UL: 0.9 LR: 0.9 Passing ACR criteria 14

Low Contrast Detectability Procedures Spoke: a row of small disks Rows radiating from the center of a circle like spokes in a wheel Slice #8 to #12: contrast 1.4%, 2.5%, 3.6% and 5.1% Slice #8 Slice #9 Slice #10 Slice #11 LCD Spokes An example 1. Always 10 spokes? 2. Because the technologist used slice #11 3. Should use the specified slice 15

Artifacts Examples 1. Zipper artifacts 1. Most are related to hardware or software problems 2. (RF leakage) perpendicular to frequency direction Phase Frequency Miscellaneous Problems Problems 1. Central frequency 4 digits? (should be 9 at 3T) Units: Hz or MHz? 2. Phantom distance: 3 digits? (should be 4 if possible) 3. Air bubbles in ACR phantom? May not affect QC test 1 2 16

Hard Copy Image QC Ensure artifact-free laser films with consistent gray levels 1. Frequency: Initiation and weekly routine QC 2. Equipment: Densitometer and Laser Film QC Chart 3. SMPTE test pattern 1. Display on the filming console 2. Examine 0/5% and 95/100% patches 3. Film the SMPTE pattern 4. Measure optical density of the 0, 10%, 40%, and 90% 5. Record results on Laser Film QC chart 4. Control Limits Additional Tests: ADNI ADNI (Alzheimer's Disease Neuroimaging Initiative) Magphan phantom Quantitative Imaging Phantom Precise, Comprehensive Measurement of 3d Spatial Distortion Quality Control 165 spheres made from polycarbonate plastic 17

Diffusion Phantom Ice-cold phantom Compare ADC values of a known, temperature-controlled fluid Multiple vendors/sites/field-strength MR systems To confirm proper MR system gradient b-value calibration Acquire a three-axis DWI scan at b = 0, 500, 800, 1000, 2000 Perfusion Phantom Dynamic contrast-enhanced MRI (DCE-MRI) Phantom Tests Provide means for qualitative and quantitative evaluation, including T1 map error estimation, temporal resolution and image quality Evaluate compliance with image acquisition protocols ACRIN CQIE Body Phantom 18

20% 20% 20% 20% Question 1: Allowed weekly CF drift from ACR Manual? 20% 1. 0.1 ppm 2. 1.0 ppm 3. 1.5 ppm 4. 2.0 ppm 5. 2.5 ppm 10 Question 2: Transaxial geometric accuracy measurements should be on slice#? 20% 1. 1 20% 2. 3 20% 3. 5 20% 4. 11 20% 5. Any slice 10 19

20% 20% 20% 20% Question 3: Table OK is for checking? 20% 1. Express table system 2. Table positioning and laser 3. Patient monitors 4. Horizontal bed movement 5. Vertical bed movement 10 References 1. Sobol WT, NessAiver MS, Orton CG, Point/counterpoint. The physics components of the ACR MRI Accreditation Program are overly tedious and beyond what is needed to ensure good patient care.med Phys 35(8)3419 2. ACR 2004 Magnetic Resonance Imaging Qaulity Control Manual 3. Ihalainen TM, Lönnroth NT, Peltonen JI, Uusi-Simola JK, Timonen MH, Kuusela LJ, Savolainen SE, Sipilä OE. MRI quality assurance using the ACR phantom in a multi-unit imaging center. Acta Oncol. 2011 Aug;50(6):966-72 4. Chen CC, Wan YL, Wai YY, Liu HL. Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results. J Digit Imaging. 2004 Dec;17(4):279-84 5. NessAiver, Range of Results from over 534 ACR-mandated Annual MRI Performance Evaluations on over 204 Magnets from 8 Vendors Spanning a 10-year Period, MRI QA RSNA 2011 6. AAPM Report No. 110, Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities, 2010 7. AAPM Report No. 34, Acceptance Testing of Magnetic Resonance Imaging Systems (Reprinted from Medical Physics, Vol. 19, Issue 1) 20

Acknowledgements Department of Radiology Rick Layman Brian Raterman Xiangyu Yang Georgeta Mihai Jun Zhang Department of Physics Wayne Newhauser Kenneth Matthews Kenneth Hogstrom Randell Deen Steven Heymsfield Baylor University Medical Center Tianliang Gu West Physics Consulting Jiali Wang Wright State Research Institute Jason Parker 21