AN IN-SITU VERIFICATION TECHNOLOGY FOR CORIOLIS FLOWMETERS



Similar documents
Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010

ELECTRON SPIN RESONANCE Last Revised: July 2007

Industry faces unrelenting pressure to increase production

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Energy and Flow Measurement for Hydronic Systems

Active Vibration Isolation of an Unbalanced Machine Spindle

Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.

Micro Motion ELITE CMFS Series Meter

EDUMECH Mechatronic Instructional Systems. Ball on Beam System

INTERFERENCE OF SOUND WAVES

Understanding Poles and Zeros

Dimensional analysis is a method for reducing the number and complexity of experimental variables that affect a given physical phenomena.

Quick Start Guide , Rev DA March Rosemount 8714D (Calibration Standard) Magnetic Flowtube Simulator

Custody Transfer Measurement. with the V-Cone Flowmeter

CHAPTER 3 MODAL ANALYSIS OF A PRINTED CIRCUIT BOARD

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Application Note. So You Need to Measure Some Inductors?

Oilfield Services Measurement Solutions

Controller Design in Frequency Domain

Using meters to measure steam flow

Rock Bolt Condition Monitoring Using Ultrasonic Guided Waves

Transducer Mounting and Test Setup Configurations. Rick Bono The Modal Shop

Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))

Transmission Line and Back Loaded Horn Physics

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Waveforms and the Speed of Sound

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification

Mounting instructions. Acceleration Transducer B12. B 26.B12.10 en

99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, cm

Similar benefits are also derived through modal testing of other space structures.

Title : Analog Circuit for Sound Localization Applications

Experiment # 3: Pipe Flow

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

ISO Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.

Entrained Gas Diagnostic with Intelligent Differential Pressure Transmitter

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

Mapping Assignments for. Micro Motion

The Sonometer The Resonant String and Timbre Change after plucking

THE USE OF A CORIOLIS FLOW METER FOR MEASURING MOLASSES PRODUCTION IN A SUGAR MILL. ABSTRACT

Dynamics of Offshore Wind Turbines

Safakcan Tuncdemir 1, William M. Bradley *2. 1. Introduction

An Electromagnetic Micro Power Generator Based on Mechanical Frequency Up-Conversion

User checks and maintenance of laboratory balances

Electrical Resonance

The use of Operating Deflection Shapes (ODS) to model the vibration of sanders and polishers HSL/2006/104. Project Leader: Author(s): Science Group:

Micro Motion Coriolis Interface Module User Manual (for ROC800-Series and FloBoss 107 Controllers)

Loudspeaker Parameters. D. G. Meyer School of Electrical & Computer Engineering

APPLICATION NOTE AP050830

Magnetic Field of a Circular Coil Lab 12

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

Everything you need for protection scheme testing

Dynamic Process Modeling. Process Dynamics and Control

Part IV. Conclusions

Correcting a Coriolis Meter for Two Phase Oil & Gas Flow

Chemical Measurement Solutions

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

Shaft. Application of full spectrum to rotating machinery diagnostics. Centerlines. Paul Goldman, Ph.D. and Agnes Muszynska, Ph.D.

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Sound absorption and acoustic surface impedance

Σ _. Feedback Amplifiers: One and Two Pole cases. Negative Feedback:

Various Technics of Liquids and Solids Level Measurements. (Part 3)

! n. Problems and Solutions Section 1.5 (1.66 through 1.74)

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Series 6000 Torque measured metal bellow coupling

Adaptation of General Purpose CFD Code for Fusion MHD Applications*

ENS 07 Paris, France, 3-4 December 2007

AC CIRCUITS - CAPACITORS AND INDUCTORS

The Calculation of G rms

Design Validation and Improvement Study of HVAC Plumbing Line Assembly under Random Loading Condition

Equipment: Power Supply, DAI, Transformer (8341), Variable resistance (8311), Variable inductance (8321), Variable capacitance (8331)

Candidate Number. General Certificate of Education Advanced Level Examination June 2014

HITACHI INVERTER SJ/L100/300 SERIES PID CONTROL USERS GUIDE

Determining Measurement Uncertainty for Dimensional Measurements

Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope

The Unique Accelabar Flow Meter

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.

ngenieros Ingenieros Asociados De Control S.L.

Soil Dynamics Prof. Deepankar Choudhury Department of Civil Engineering Indian Institute of Technology, Bombay

Tamura Closed Loop Hall Effect Current Sensors

Experiment 3 Pipe Friction

Abaqus Technology Brief. Automobile Roof Crush Analysis with Abaqus

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

Address for Correspondence

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

Troubleshooting accelerometer installations

CORRECTION OF DYNAMIC WHEEL FORCES MEASURED ON ROAD SIMULATORS

Manufacturing Equipment Modeling

Using the Spectrophotometer

Accurate Loss-of-Signal Detection in 10Gbps Optical Receivers using the MAX3991

Upon completion of this lab, the student will be able to:

Chapter 3 Non-parametric Models for Magneto-Rheological Dampers

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

Chapter 11 SERVO VALVES. Fluid Power Circuits and Controls, John S.Cundiff, 2001

Transcription:

Timothy J. Cunningham Micro Motion, Inc 77 Winchester Circle Boulder, CO 81 () 5-859 () 5-8596 (fax) tim.cunningham@emerson.com Abstract: Structural Integrity Meter Verification is a robust new Coriolis verification technology which uses the onboard electronics to very accurately measure the stiffness of the flowtubes. The flow tube stiffness is directly related to the flow calibration factor and is uninfluenced by process conditions. Meter Verification compares the measured stiffness to the factory baseline stiffness to confirm that the flow calibration factor is unchanged from the factory value. Structural Integrity Meter Verification also performs additional electronics and software checks to ensure accurate measurement. This new technology allows users to save money and reduce downtime by verifying Coriolis meters in situ. Sometimes Coriolis meters are used with corrosive fluids that can etch away the tube or with erosive fluids that can cause localized thinning of the tubes. In these applications the precision and accuracy of the Coriolis flowmeter outweighs the replacement costs. A Coriolis meter has been deliberately corroded while tracking the flow calibration factor and the stiffness. Finite element analysis was used to analyze the relationship between stiffness and flow calibration factor for erosive applications. Meter Verification results for these cases are compared to the experimental and analytical data. 1. INTRODUCTION AN IN-SITU VERIFICATION TECHNOLOGY FOR CORIOLIS FLOWMETERS Flowmeters are commonly validated by comparing the indicated flow measurement to a reference flow measurement. Flowmeters are also commonly verified by tracking a secondary variable that is highly correlated to the flow measurement. For example, orifice plates can be measured to verify accuracy. Other verification techniques include spindown tests for turbine meters and speed of sound and transducer gain checks for ultrasonic meters. Coriolis meters have historically used secondary variables to verify performance, e.g. drive gain. Unfortunately drive gain is only loosely correlated with the flow measurement. A method of verification using a known density fluid has been used successfully, but this approach is prone to user error. In response to customer demand for an easy to use meter verification methodology for Coriolis flowmeters, Micro Motion has developed Structural Integrity Meter Verification that uses the onboard electronics to verify the integrity of the flow tube as well as the electromechanical components, the transmitter electronics, and the transmitter software.. CORIOLIS FLOWMETER BACKGROUND Structural Integrity Meter Verification uses the stiffness of the flow tubes as the secondary variable to An In Situ Verification Technology 1/5 verify the correctness of the Flow Calibration Factor (FCF). The FCF is the proportionality constant that relates the time delay,δt, to the mass flow rate, m. m = FCF δt (1) Equation (1) can be derived from first principles, for example starting with the Housner differential equation describing a fluid-conveying beam [1, ]. These derivations result in a term corresponding to the FCF as shown in Equation (). EI m = C t L δ () where C is a dimensionless geometric constant related to the boundary conditions and beam properties. The EI term, corresponding to the FCF, has L units of force/length, the units of stiffness. Going through these derivations in detail to show this relationship between stiffness and flow calibration factor is beyond the scope of this paper. However a much simpler dimensional analysis of Equation (1) shows that the FCF has units of stiffness. Rearranging equation (1) FCF = m () δt shows that the units of the FCF are mass flow rate/time delay. This is shown dimensionally as

Mass (4) For example, FCF is commonly expressed in units of (gm/sec)/μsec. In a consistent system of units, mass can be represented by force/(acceleration of gravity), taking advantage of Newton s Second Law. Plugging this into equation (4) Mass Force * / Length Force = = (5) Length shows very simply that the flow calibration factor has units of stiffness (Force/Length). The equivalence of FCF and stiffness shows why stiffness is the secondary variable that is highly correlated to the FCF. The problem now becomes one of how to determine the stiffness of the flow tubes..1 Meter Verification Theory Coriolis Structural Integrity Meter Verification uses techniques from Experimental Modal Analysis and Structural Dynamics theory to very accurately measure the stiffness of the flow tubes using the embedded electronics and onboard pickoff and drive coil and magnets. Figure 1 shows a typical Coriolis mass flow meter. The drive coil and magnet at the top center inbetween the tubes is used to drive the Coriolis flowmeter at resonance. A feedback control system in the flowmeter electronics applies a sinusoidal current to the drive coil to maintain resonance at a specific amplitude. The two pickoff coils and magnets produce a voltage in response to the resonance motion. The pickoffs are used as the feedback signal to control amplitude. The transmitter s digital signal processing uses the pickoff responses to estimate the frequency of vibration, used in the density measurement, as well as the time delay between the two pickoff sinusoids, δt, needed for the mass flow measurement. Further details discussing the operation of a Coriolis flowmeter are given in Reference []. Meter Verification runs on top of the standard Coriolis signal processing and drive control. A series of tones are added to the drive signal. These tones excite off-resonance responses in the two pickoffs. The embedded flowmeter electronics measures these tonal inputs and responses to produce a frequency response function (FRF). Figure 1. Typical Coriolis Flowmeter A structural dynamics FRF can be modeled as a second order system with the parameters of stiffness (K), mass (M), and damping (C). Applying electromagnetic theory to the problem, the FRF can be defined by pickoff voltage/input current. X ( ω) jω FRF = H ( ω) = = (6) F( ω) Mω + jcω + K The Meter Verification results are based on fitting the measured FRF to the second order model to independently estimate K, M, and C. Figure shows this graphically. The lower frequency portion of the FRF is dominated by the stiffness. The higher frequency portion is dominated by the mass. These mass and stiffness lines, as they are called, are shown in the plot, and are actually reciprocals of the mass and stiffness. FRF Magnitude 1 1 1 1 1 1 Frequency Response Function Freq=sqrt(K/M) Nominal K Frequency (Hz) Peak ~1/ζ Nominal M 1 Figure. Nominal Frequency Response Function An In Situ Verification Technology /5

The resonant frequency is determined by the square root of ratio of the mass and stiffness. The height of the resonant peak is determined by the non-dimensional damping coefficient ζ, which is related to the damping, C, by Equation (7). ζ = C (7) KM The embedded core processor performs the signal processing necessary to generate the FRF; curve fits the FRF to generate estimates for K, M, and C; and handles all of the bookkeeping to keep track of the results generated by Meter Verification.. RESULTS Meter Verification distills all of its results down to two simple numbers that it presents to the customer. Meter Verification starts with the factory baseline verifications during the standard meter calibration process, which Micro Motion performs on both air and water as part of its comprehensive diagnostic program. However since the process fluid does not change the stiffness of the meter, the factory baseline stiffnesses on air and water are statistically the same. Each meter verification measurement is normalized by the average of these stiffnesses and converted into a stiffness uncertainty, which is the percentage change in the measured stiffness from the factory baselines. stiffness = uncertainty stiffnessmeasured ( ) ( stiffnessfactory, air + stiffnessfactory, water ) 1 % Normalizing the stiffness uncertainty in this way makes it easy to track any changes in the flowmeter by using a format that is convenient to view. (This stiffness uncertainty should not be confused with the term measurement uncertainty as it is used in metrological terms.).1 Meter Verification Stability An In Situ Verification Technology /5 (8) The signal processing used in Meter Verification has been designed to enhance the stability of the measurement. Each stiffness uncertainty measurement is the average of the stiffness estimates from many FRFs. In turn, each FRF that is fit is averaged from many individual FRF measurements. This averaging results in a very stable stiffness uncertainty estimate. In line with standard measurement techniques, the variation in the meter verification uncertainty is several times better than the base flow accuracy. Figure shows a typical meter verification uncertainty plot with a standard deviation of less than.1% under laboratory conditions. Note that stiffness uncertainty is calculated for each of the two pickoffs, further increasing the confidence in the measurement. Uncertainity % 5 4 1 - - -4 Structural Integrity Normalized Stiffness u spec l spec -5 5 4 45 5 55 6 65 7 Meter Verification Counter Figure. Meter Verification Stability K LPOoutlet K RPOinlet Meter Verification uncertainty variation is of course subject to field effects. The specification limits for stiffness uncertainty are set such that under the full range of field effects there is a σ probability against giving a false alarm. Meter Verification, specification limits and field effects are discussed more fully in References [4] and [5].. Meter Verification Corrosion Detection Because Coriolis meters operate with no moving parts, the calibration is expected to be unchanged over the life of the meter. Most customers using Meter Verification will expect to see stiffness uncertainty results around %, indicating that the meter is unchanged from the factory baseline. However, because of their value proposition, Coriolis meters are sometimes used with incompatible, corrosive, fluids where the lifetime of the meter may be on the order of 1 to years. For these applications, Meter Verification can be used to detect changes due to corrosion in the tubes. Finite element analysis (FEA) is used extensively in the design of Coriolis flowmeters [6]. FEA can also model the effects of corrosion in flowmeters. The details of this FE work are beyond the scope of this paper, but these analyses illustrate some of the technology behind Meter Verification. Figure 4 is an extension of Figure, where the nominal FRF is shown in blue. The FE model was modified to simulate corrosion and used to generate

a second FRF, shown in green. Note that with the corroded tubes the frequency has decreased. A key feature of the meter verification technology is that the test tones used to generate the FRF are dynamically determined, based on the instantaneous drive frequency. These test tones are shown as blue x s for the nominal case and as green o s for the corroded case. Most importantly, the independent estimations of the mass, stiffness, and damping are unaffected by the actual resonant frequency. The change in stiffness instead is detected by the shift in the estimated value for K. Figure 4 shows that the corroded tubes have an FRF with a higher stiffness line, corresponding to less stiffness, than the nominal case. The curve fit correctly estimates the change in stiffness independently from the change in resonant frequency. Frequency Response Function FRF Magnitude 1 1 1 1 1 1 Freq=sqrt(K/M) Nominal K Frequency (Hz) Peak ~1/ζ Less Stiff K Nominal M 1 Figure 4. Change in FRF with Stiffness Reduction To experimentally verify the ability of Meter Verification to detect corrosion, a flowmeter was corroded using a very strong, incompatible, acid. At each corrosion step the flowmeter was recalibrated to get the new FCF. Figure 5 shows the change in FCF versus the change in stiffness for each corrosion step. Note that even with careful laboratory technique, the corrosion is not uniform as indicated by the different change in stiffness for the left and right pickoffs (LPO, RPO). However note that the FCF change has an approximately 1 to 1 relationship with the averaged change in stiffness, as expected due to their essential equality. These results show that corrosion is very detectable with Meter Verification. Another application of Coriolis flowmeters where wear is expected is with erosive slurries, the subject of next section.. Meter Verification Erosion Detection It is difficult to do a controlled experiment for erosive wear of a Coriolis meter. However finite element analysis can be used to investigate the detectability of erosion with Meter Verification. Figure 6 shows a beam finite element model used to analyze erosion. The red elements on the inlet tube bend are where most of the erosion is expected. The wall thickness of these elements was progressively decreased with the stiffness uncertainty and FCF calculated at each iteration. The FE results are shown in Figure 7. Note how the LPO and RPO stiffnesses change it opposite directions, a hallmark of nonuniform changes to the flow tubes. These stiffness uncertainties have diverged from the baseline with a total spread of about 1% for an FCF decrease of ~1.% Percent Change in FCF Percent Change in FCF vs. Percent Change in Stiffness LPO -.5 RPO.5 - -.5 - -.5-4 -5-4 - - 1 Percent Change in Stiffness Figure 5. FCF vs Stiffness with Corrosion Figure 6. Beam FE Model for Erosion Study An In Situ Verification Technology 4/5

It is interesting to note that the RPO stiffness has increased even though the RPO is on the inlet side of the tubes, where the thinning has happened. The LPO stiffness has decreased even though no change occurred on that side of the flow tubes. Non--intuitive results like these occur frequently in structural dynamics problems. A thought experiment gives insight into the increase in stiffness of the RPO with a decrease in tube stiffness. Imagine that the inlet bend was completely removed, the limiting case of stiffness reduction. In that case the RPO would not move at all with a force applied at the driver, essentially becoming infinitely stiff. Intuition must be used with care when interpreting Meter Verification results. Percent Change in FCF -. -.4 -.6 -.8 FCF change vs Stiffness change, inlet tubes eroded LPO. RPO -.4 -...4.6 Percent change in stiffness Figure 7. Fcf vs Stiffness with Erosion 4. DISCUSSION Meter Verification measures stiffness to ensure the integrity of the sensing element, the flow tubes. Additionally the electronics associated with the flow measurement need to be verified. Structural Integrity Meter Verification confirms the integrity of the flowmeter electronics by verifying the stiffness with the same transducers, analog electronics, digital electronics, and software used for the flow measurement. Any change in the electronics will cause the stiffness uncertainty to go out of specification. Therefore good stiffness uncertainty confirms both the sensing element and the electronics. Verification is unlike flowmeter validation methodologies such as proving, in which the unit under test's flow output is compared to a primary flow output. Verifications require several additional checks to confirm overall flowmeter performance. These checks include confirming the software configuration, the flowmeter' s zero, and the proper functioning of the analog outputs. A complete verification might include checking the analog output functionality with the built-in diagnostic/trim functions. Structural Integrity Meter Verification includes a built-in check of the software configuration, comparing it to the previously verified values. Additionally Meter Verification checks the current zero against the factory zero and the last-verified zero. Meter Verification also provides a graphical output of the results and the ability to print a report of the current verification. All of these features combine to completely check the performance of the entire flowmeter. 5. CONCLUSION Structural Integrity Meter Verification is a robust technology for in-situ verification of Coriolis flowmeters. It can be used with confidence as a costeffective, robust, means of verifying Coriolis flowmeter performance. For applications where Coriolis flowmeter degradation is expected, for example with corrosive fluids, Meter Verification is a good tool for detecting flowmeter wear. REFERENCES [1] Effect of detector masses on calibration of Coriolis flowmeters, Lange U., Levien A., Pankratz T., Raszillier H., Flow Measurement Instrumentation, Volume 5 Number 4, 1994. [] A Finite Element for the Vibration Analysis of a Fluid-conveying Timoshenko Beam, Stack C., Garnett R., Pawlas G., 4th SDM conference proceedings, 199, AIAA [] Coriolis Technology Creates Superior Meters, Stack C., Micro Motion White Paper WP-51, www.micromotion.com,. [4] Using Structural Integrity Meter Verification, Cunningham T., Stack C., Connor C., Micro Motion White Paper WP-948, www.micromotion.com, 7. [5] Using Structural Integrity Meter Verification to Track Corrosion in Coriolis Flowmeters, Cunningham T., Micro Motion White Paper WP- 1196, www.micromotion.com, 9. [7] Using IMAT and MATLAB for Coriolis Flowmeter Design, Cunningham T., Hensley D., IMAC XIX Proceedings, February 1, Paper #74 An In Situ Verification Technology 5/5