The Effect of Temperature on Conductivity Measurement



Similar documents
The Effects of Temperature on ph Measurement

The Relationship between ph and Deionized Water

Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19

High Purity Water Resistivity/ Conductivity Measurement

This value, called the ionic product of water, Kw, is related to the equilibrium constant of water

Conductivity Theory and Practice

ph: Measurement and Uses

Inferred ph in Steam Plant Water Chemistry Monitoring

Review of Chemical Equilibrium 7.51 September free [A] (µm)

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

the importance of ph in water purification the importance of ph in water purification Water Purification Systems UK

Chem101: General Chemistry Lecture 9 Acids and Bases

Q.1 Classify the following according to Lewis theory and Brønsted-Lowry theory.

Refractive Index Measurement Principle

Q.1 Classify the following according to Lewis theory and Brønsted-Lowry theory.

Chemistry Ch 15 (Solutions) Study Guide Introduction

Understanding electrical conductivity

Chemistry 51 Chapter 8 TYPES OF SOLUTIONS. A solution is a homogeneous mixture of two substances: a solute and a solvent.

Acid Dissociation Constants and the Titration of a Weak Acid

Electrical Conductivity of Aqueous Solutions

Lab 7 Soil ph and Salinity OBJECTIVE INTRODUCTION Soil ph active

Validation and Calibration. Definitions and Terminology

THEORY AND APPLICATION OF CONDUCTIVITY

Gravimetric determination of pipette errors

Molecular Models in Biology

4.2 Bias, Standards and Standardization

Experiment 8 - Double Displacement Reactions

Sketch the model representation of the first step in the dissociation of water. H 2. O (l) H + (aq) + OH- (aq) + H 2. OH - (aq) + H 3 O+ (aq)

One problem often faced in qualitative analysis is to test for one ion in a

Lab #11: Determination of a Chemical Equilibrium Constant

Work hard. Be nice. Name: Period: Date: UNIT 1: Introduction to Matter Lesson 4: A Fine Line Between Compounds and Mixtures

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

CHM1 Review for Exam 12

Conductivity Sensor Calibrations to Meet Water Industry Requirements

DEGASSED CATION CONDUCTIVITY MEASUREMENT

Carolina s Solution Preparation Manual

(Uncertainty) 2. How uncertain is your uncertainty budget?

Exp 13 Volumetric Analysis: Acid-Base titration

HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY

stoichiometry = the numerical relationships between chemical amounts in a reaction.

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

Milliequivalents, Millimoles,, and Milliosmoles

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

Care and use guide. Contents. a handy solution made easy. page

Chemical Reactions in Water Ron Robertson

Measuring the Conductivity of Pure Water Using USP<645>

Features, Limitations & Potential Errors

EXPERIMENT 20: Determination of ph of Common Substances

Formulae, stoichiometry and the mole concept

THE CONDUCTIVITY OF LOW CONCENTRATIONS OF CO 2 DISSOLVED IN ULTRAPURE WATER FROM C

Tutorial 4 SOLUTION STOICHIOMETRY. Solution stoichiometry calculations involve chemical reactions taking place in solution.

5s Solubility & Conductivity

Acids, Bases, and ph

The Determination of an Equilibrium Constant

Liquid Conductivity: Measuring Conductivity in Saline Water Solutions (Teacher s Guide)

Understanding hydrogen ion concentration (ph)

Design of Experiments for Analytical Method Development and Validation

Corrosion of Copper in Water

Molarity of Ions in Solution

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

EXPERIMENT 2 THE HYDROLYSIS OF t-butyl CHLORIDE. PURPOSE: To verify a proposed mechanism for the hydrolysis of t-butyl Chloride.

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Review for Solving ph Problems:

Relative Humidity Calibration Kit

Ion Exchange Design Hand calculation. Brian Windsor (Purolite International Ltd)

6 Reactions in Aqueous Solutions

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

IB Chemistry. DP Chemistry Review

Eutech 2700 Series. Bench Meters. ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature

Chapter 8: Chemical Equations and Reactions

Acid Base Titrations

SUPPLEMENTARY MATERIAL

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Brønsted-Lowry Acids and Bases

Titrations. Acid-Base Indicators and Titration Curves. Shapes of Titration Curves. A titration curve is a graphical history of a titration

ph Measurement and Control

Auto-ionization of Water

Acids and Bases: A Brief Review

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

Unit 2: Quantities in Chemistry

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED

How Sensors Work. How Oxygen, Electrochemical Toxic, and Metal Oxide Semiconductor Sensors Work *

SAMPLE PROBLEM 8.1. Solutions of Electrolytes and Nonelectrolytes SOLUTION STUDY CHECK

DEIONIZATION IN A "NUT SHELL"

Optimize Your Process Operations by Improving Process Monitoring

Determining the Identity of an Unknown Weak Acid

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

Institut Technologique FCBA. Ecotoxicology Chemistry Laboratory. Allée de Boutaut - BP Bordeaux Cedex

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

A Comprehensive Look at Conductivity Measurement in Steam and Power Generation Waters

Practical description of how to measure conductivity. Laboratory environment. Theory and Practice of Conductivity learn Applications easily

A Review of Statistical Outlier Methods

Technology: Heating, Ventilation and Air Condition (HVAC) Zone Controls

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

CHEMICAL DETERMINATION OF EVERYDAY HOUSEHOLD CHEMICALS

(1) e.g. H hydrogen that has lost 1 electron c. anion - negatively charged atoms that gain electrons (1) e.g. HCO 3 bicarbonate anion

Transcription:

The Effect of on Measurement Authors: John J. Barron & Colin Ashton Technical Services Department, Reagecon Diagnostics Ltd, Shannon Free Zone, County Clare, Ireland Abstract The conductivity of all solutions changes as the solutions temperature changes. It is a fundamental, practical requirement in the field of conductivity measurement that test measurements taken at different temperatures can be compared. This comparison can be facilitated by the use of, which provides an estimate of the samples conductivity at a common reference This paper describes the principle of and provides details of the different types of that are available to the analyst so that the most appropriate type can be selected for the analyst s measurement application. 1 Introduction An increase in a solution s temperature will cause a decrease in its viscosity and an increase in the mobility of the ions in solution. An increase in temperature may also cause an increase in the number of ions in solution due to dissociation of molecules. As the conductivity of a solution is dependent on these factors then an increase in the solution s temperature will lead to an increase in its conductivity. The Coefficient of Variation is the rate at which a solution s conductivity increases with an increase of temperature and is expressed as the percentage increase in conductivity for a temperature change of 1 C. The Coefficient of Variation will be different for each solvent / solute mixture - examples are outlined in Table 1. As can be seen from Table 1, temperature has a significant effect on the measured conductivity. For ultrapure water, a variation in temperature of 0.1 C will cause a change in conductivity of 0.55%. This example shows that accurate temperature measurement and control is of paramount importance for accurate conductivity measurement. 2 In many applications, including field or process measurement, temperature control may prove impractical; but comparison of results is still required. This is achieved by using temperature compensation to calculate the samples conductivity values at a common reference To allow the analyst to compare conductivity measurements, results should be referenced to a specified Most conductivity measurements are referenced to 25 C, but some guidelines reference results to other temperatures examples are given in Table 2. Solution Coefficient of Variation %/ C at 25 C Ultrapure Water 5.5 NaOH 5% 2.01 NaOH 30% 4.50 HCl 5% 1.58 HCl 30% 1.52 KCl 5% 2.01 KCl 20% 1.68 Fresh water ~ 2.0 Table 1: Coefficient of Variation for Common Solutions (1,2) Page 1 of 5

Standard Procedure Reference United States Pharmacopoeia (3) 25 C ISO 7888-1985 (4) 25 C European Pharmacopoeia (5) 25 C ASTM D 1125 95 (6) 25 C U.K. Water Quality Supply Regulations (7) 20 C Table 2: Reference Cited in Commonly used Procedures Reported at Reference (25 C) Measured 10 C 25 C Graph 1: compensation requires an algorithm to be applied to the measured conductivity reading and measured temperature to produce a compensated reading at a reference Graph 1 shows how a reading taken at 10 C is used to generate a temperature compensated conductivity value at 25 C by applying the algorithm represented by the curve. 2.1 Linear For linear temperature compensation, it is assumed that the temperature coefficient of variation has the same value for all measurement temperatures. This assumption is not true; but for many measurements this does not result in a substantial contribution to the combined uncertainty of measurement of the reported result. The temperature coefficient of variation at 25 C, α θ,25, is calculated as shown in Equation 1, with its value being expressed as a percentage change in conductivity value per 1 C change in α θ,25 = κ θ - κ 25 x 100 κ 25 (θ 25) Equation 1 Where θ and 25 C are temperatures at which conductivities κ θ and κ 25 respectively are measured. When the measurement is not carried out at 25 C, α θ,25 can be used to calculate the solution s compensated conductivity at 25 C, as shown by Equation 2. Page 2 of 5

κ 25 = κ θ 1 + (α θ,25 /100) (θ - 25) Equation 2 warns that the data is not applicable to potassium chloride calibration solutions a different temperature compensation algorithm must be applied during calibration. 2.2 Non-Linear As previously mentioned the relationship between temperature and conductivity does not follow a linear relationship. When the assumption of a linear relationship is not adequate, a polynomial may be applied to improve the accuracy of the calculated result. Non-linear temperature compensation data for a variety of solutions is widely available, e.g. ISO 7888-1985 (4) provides data for natural waters having κ 25 of 60 to 1000 μs/cm over the temperature range 0 to 35 C. The standard Graph 2 demonstrates how linear and non-linear temperature compensation are used to provide temperature compensated readings at 25 C from a measured value at 10 C. The graph demonstrates that the two forms of temperature compensation will result in different reported conductivity values at the reference temperature of 25 C. It should not be assumed that non-linear temperature compensation will lead to lower measurement errors than linear compensation - this will only be the case if the non-linear temperature compensation is a good fit for the sample s conductivity variation with Linear to 25 C Non-Linear to 25 C Measured Value at 10 C 10 C 25 C Graph 2: Representation of Linear & Non-Linear Whatever form of compensation is used, the reading will not be as accurate as taking a reading of the sample s conductivity at the reference Page 3 of 5

2.3 Manual Manual involves taking a series of conductivity measurements on a sample at various temperatures. A linear or polynomial best fit is then applied to the data to produce an equation for the temperature compensation of results. The temperature compensated conductivity result is manually calculated using the derived equation and the measured conductivity and 2.4 Automatic (ATC) ATC utilises a microprocessor within the instrument to determine a temperature compensated conductivity result. The ATC options available will depend upon the specification of the conductivity instrument: Basic conductivity instruments may only offer linear compensation with the temperature coefficient fixed at 2%/ C. This value is chosen because it is a reasonable approximation of the temperature coefficient of commonly encountered dilute salts. However, many solutions have a markedly different temperature coefficient and measurements of these solutions with basic instruments will result in significant errors in the reported conductivity value. Intermediate conductivity instruments may offer linear compensation with a user-definable temperature coefficient. These instruments may only allow the user to vary the temperature coefficient in relatively large increments, e.g. 0.5%/ C increments. Advanced instruments may offer both linear and non-linear temperature compensation functions as well as an uncompensated mode. These instruments will allow the analyst to input a temperature coefficient (normally 0.00 to 10.00 %/ C) and a reference temperature (normally 20 or 25 C). It is important to note that some procedures, such as USP 26 <645> (3) require that no temperature compensation can be applied to the measured value. The non-linear temperature compensation will carry out temperature compensation using preprogrammed data in the instrument. Non-linear ATC data is available for a variety of solutions including Natural Waters according to ISO 7888 (4) Ultrapure water with traces of ammonia Ultrapure water with traces of HCl Ultrapure water with traces of NaCl Ultrapure water with traces of NaOH In addition to temperature compensation for sample measurements, conductivity instruments may also be pre-programmed to allow temperature compensation to be utilised during calibration. In this instance, the appropriate compensation will be applied during the calibration process for common conductivity calibration standards such as potassium chloride based 147, 1413 & 12,880 μs/cm at 25 C. Many conductivity meters also allow calibration with a user-defined conductivity standard. Standards selected for this purpose should be provided with full characterization of their conductivity s variation with temperature so that this can be accounted for during the calibration process. 3 Conclusion Measurement temperature has a significant influence on conductivity readings; but appropriate temperature compensation is a powerful tool to allow meaningful comparison of readings taken at different temperatures. The analyst must ensure that the type of temperature compensation utilized is appropriate for both the type of sample being analyzed and the required test accuracy. This is an essential factor for determining the suitability of a conductivity instrument for measurement applications. A conductivity measurement taken with the sample at the reference temperature will always be more accurate than a temperature compensated reading taken away from the reference temperature this point is essential for critical applications requiring high accuracy of measurements. Further information on the implications of temperature effects on conductivity measurements is given in further Reagecon practical conductivity measurement publications (8,9). Page 4 of 5

4 References 1. T.S.Light, Dependance and Measurement of Resistivity of Pure Water, Analytical Chemistry, 1984, 56, 1138 2. Instruction Manual for CyberScan CON 500 Meter, Eutech Instruments 3. United States Pharmacopoeia 26-NF 21, Monograph <645> 4. Water Quality Determination of Electrical, ISO 7888:1985 5. European Pharmacopoeia, Edition 5.3, Section 2.2.38: 6. Standard Test Methods for Electrical and Resistivity of Water, American Society for Testing and Materials, D1125-95 7. The Water Supply (Water Quality) Regulations 2000 Water, England and Wales, Statutory Instrument 2000 No. 3184, The Stationery Office Ltd. 8. J.J. Barron & C. Ashton, A Practical Guide to Accurate Measurements * 9. J.J. Barron & C. Ashton, Measurement in the Pharmaceutical Industry * * These papers form part of a comprehensive series of papers that the authors have written covering all of the practical requirements for accurate conductivity measurement. These papers and the authors book, A Practical Guide to Accurate Measurement are available via Reagecon s website - www.reagecon.com. Biographical Notes: John J Barron is Managing and Technical Director of Reagecon Diagnostics Limited. The company, which was founded in 1986, is the largest producer worldwide of Standards and is also a major producer of other chemical standards. Mr. Barron is an expert in several areas of analytical chemistry, including electro-chemical analysis, good laboratory practice (GLP) and chemical metrology. He has written and lectured extensively and is credited with several technical discoveries including stable low level conductivity standards. Colin Ashton has worked in the Reagecon group since 1994 and is currently Head of the Chemical Metrology Department. A graduate of the University of Southampton, he has developed particular expertise in the development, stabilisation, manufacture and validation of cation, anion and electro-chemical standards. He has particular scientific interest in all aspects of on line chemical analysis and has lectured and published on several areas of this field. Acknowledgements The authors wish to extend their gratitude to Ms V. Byrne for her assistance in compiling this paper and Mr L. Geary and Ms R. Cooney for proofing this paper. The authors wish to thank all of their colleagues who have provided technical assistance in compiling Reagecon s series of conductivity measurement papers. John J Barron and Colin Ashton Page 5 of 5