Using Automatic Exposure Control in Digital Radiography



Similar documents
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Physics testing of image detectors

Page: 1 of 6 Page: 1 of 6

INTRODUCTION. A. Purpose

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Digital Mammography Update: Design and Characteristics of Current Systems

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

Quality Control of Full Field Digital Mammography Units

Quality Control and Maintenance Programs

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

The Exposure Index and Its Standardization

Computed Radiography: Acceptance Testing and Quality Control

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Digital Radiography Image Quality: Image Acquisition

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Performance testing for Precision 500D Classical R/F System

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Update on ACR Digital Mammography QC Manual

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Radiation Safety Characteristics of the NOMAD Portable X-ray System

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

X-ray Imaging Systems

Performance evaluation of computed radiography systems

Digital Breast Tomosynthesis QC Requirements

Recommendations for a technical quality control program for diagnostic X-ray equipment

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)

Aspire FCRm/CRm Fujifilm CR for Mammography

Best Practices in Digital Radiography

Radiation safety in dental radiography

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Characterizing Digital Cameras with the Photon Transfer Curve

Modern Fluoroscopy Imaging Systems

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Understanding Digital Modalities: System Integration and Use

Fundamentals of Cone-Beam CT Imaging

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Diagnostic X-ray facilities as per quality control performances in Tanzania

Overview of Digital Detector Technology

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Prepublication Requirements

Digital radiography has become the norm in U.S.

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Atomic and Nuclear Physics Laboratory (Physics 4780)

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

X-ray Imaging Systems

Equipment. Lin muyan. Presented by: CocticaUwititonze Cell:+250 (0)

ELECTRON SPIN RESONANCE Last Revised: July 2007

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Production of X-rays and Interactions of X-rays with Matter

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Digital radiography conquers the veterinary world

COMPUTED RADIOGRAPHY (CR) AND DIGITAL RADIOGRAPHY (DR) STATE X-RAY INSPECTION PROTOCOL

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

EDXRF of Used Automotive Catalytic Converters

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

MATRIX TECHNICAL NOTES

Breast MRI Quality Control

Improving digital image quality for larger patient sizes without compromises

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

Transcription:

Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The purpose of AEC is to deliver consistent, reproducible exposures across a wide range of anatomical thicknesses, tube potentials, and users Detectors used in AEC systems include fluorescent screens with PMTs/photodiodes, ionization chambers, and possibly solid state detectors AEC System Diagram Generator Backup Reference Voltage -3-2 -1 0 +1 +2 +3 AEC Controller Terminate? kvp Amplifier Comparator Loosely based on Bushberg, Seibert, Leidholdt, and Boone, The Essential Physics of Medical Imaging. Fundamental AEC performance characteristics Initial acceptance testing Sensor selector/location Density correction Screen sensitivity adjustment Sensitivity vs. speed setting Reproducibility AEC balance/field sensitivity matching AEC sensitivity Patient thickness tracking kvp tracking Beam quality correction curve Backup timer Cell mapping Minimal exposure duration Field size tracking* Ongoing QC testing AEC sensitivity Density correction kvp tracking Patient thickness tracking Reproducibility AEC balance Backup timer Minimum exposure duration Field size tracking* Remember this Test your system at the SID for which the grid is focused and your system calibrated AEC detectors themselves have an inherent energy dependence Slight variations do exist between cells and should be evaluated upon acceptance testing True vignette: Siemens Sireskop system with reciprocating grid. AEC balance consistently failed when tested at an SID of 100 cm. Moving the tube to an SID of 115 cm resulted in passing test. Cause: Grid reciprocation delay was set to result in appropriately exposed images at an SID of 115 cm, where Siemens calibrates. We were able to adjust the delay to balance the AEC cells within 10% and provide acceptable image quality at 100 cm SID. 1

Inter-cell differences exist Density control Unit 1 100.00 95.00 90.00 85.00 80.00 L+R 75.00 C 70.00 50 70 90 110 130 kvp Unit 2 74.0 72.0 70.0 68.0 66.0 64.0 62.0 60.0 50 70 90 110 130 kvp Right Left Center Adjusts mas upward or downward in increments of 25-30% per step Some newer DR systems do not incorporate this feature Published guidelines/recommendations Tested for proper operation and similar step size (NCRP 99) 0.15 to 0.30 OD per step (AAPM 74) 4-step controller should adjust by 20-25% 25% per step (AAPM 14) Fundamental differences from S/F: Absence of feature on some systems Curious that some systems still include this feature Air kerma vs. sensitivity selection Many DR systems feature selectable sensitivity settings for imaging protocols Refers to AEC, not digital detector These settings adjust the sensitivity of the AEC cells resulting in higher or lower image receptor exposures Akin to using different speed screen/film combinations for different body regions e.g. 200 speed for chest and 400 speed for general While no regulations or guidelines exist, it is prudent to verify that exposure varies logically and predictably with selection Inversely proportional Air kerma vs. sensitivity Fundamental differences from S/F: Similar to screen selector setting Philips DR system Phil Rauch Resolution of EI is not sufficient to identify changes in sensitivity setting 35 120.00 30 mas 100.00 vs. Sensitivity setting 25 80.00 120.00 mas 20 15 60.00 100.00 PMMA Aluminum 10 40.00 80.00 20.00 5 0 0.00 0 200 400 600 800 1000 1200 1400 1600 1800 Sensitivity setting 60.00 40.00 mas S = mas S 2 1 1 2 20.00 0.00 0 200 400 600 800 1000 1200 1400 1600 Sensitivity 2

AEC Balance Common AEC balance schemes Field sensitivity matching To achieve consistent exposures, AEC cells should be balanced Manufacturers of x-ray x systems use several different schemes for AEC balancing Thus, a one-size size-fits-all test will not be valid if you work with a variety of systems Fundamental difference from S/F: Some manufacturers have changed schemes 1 1 1 1.15 1 1.15 AEC Balance Test AEC balance during acceptance testing to set a baseline standard for balance Ask your service engineer about the calibration/balancing procedure Many manufacturers today use AEC systems that are not serviceable e for balance they can only be replaced In this case you are stuck with manufacturer s s balance scheme Other manufacturers still have tunable AEC cells Pots in generator Software interface Pots in detector housing Published guidelines/recommendations ± 5% across all combinations (AAPM 14) AEC Balance Cells must also be matched when used in combinations AEC systems may terminate the exposure when the most sensitive cell reaches the required current level, or may average the signal between the cells being used Same criteria apply Unit 1 Unit 2 Unit 3 C 6.32 46.6 4.5 L 7.23 69.9 4.32 R 7.16 70.4 4.44 L+R 7.16 69.9 4.39 L+C 6.73 55.7 4.39 R+C 6.7 55.9 4.44 See also X-ray generator and automatic exposure control device acceptance testing by Raymond P. Rossi, M.S., in Specification, Acceptance Testing and Quality Control of Diagnostic X-ray Imaging Equipment, Proceedings of the 1991 AAPM Summer School All 6.83 59.8 4.39 In the case of a system malfunction or technical error, the exposure must be terminated after a certain period of time or delivered mas 600 mas 51 kvp 2,000 mas < 51 kvp 21CFR1020 Many digital radiography systems have preset time limits based on kvp and ma settings, and some will terminate if no signal is detected Backup timer Reproducibility K a delivered by the AEC system should be reproducible Published guidelines/recommendations COV < 0.05 (AAPM 14) ± 5% of average (NCRP 99) I ll leave it as an exercise to prove that these guidelines do not say the same thing 3

Minimum exposure duration Thin or less dense body parts can lead to very short exposure times for AEC-controlled radiography E.g. chest imaging, small patients AEC systems should be capable of delivering appropriate exposures at these exposure times Published guidelines/recommendations T min < 1/60 sec or 5 mas, whichever is greater (21CFR1020) Minimum exposure duration - DR However, for DR systems, some of which require less exposure than film/screen systems, this limit is insufficient Typical PA chest x-ray x at 125 kvp and 320 ma: 3-43 4 ms Most measurement equipment is incapable of measuring these extremely short times kvp divider Spinning top Oscilloscopes Acquire a series of images with a phantom that yields clinically relevant exposure times (3-10 msec) Examine exposure time in DICOM header Measure in a uniform portion of the image COV < 0.05 Also, it is very important to find the minimum response time (MRT) of your AEC/generator and ensure that your AEC exposure times do not fall l below this time AEC sensitivity Typically referenced to the center cell Manufacturer has calibrated the AEC system to the air kerma they believe their detector needs You may need to calibrate the AEC system to the air kerma you and your radiologists know the detector needs AEC Sensitivity AEC system must be calibrated to deliver the necessary but sufficient K a to the image receptor Relatively simple task with screen/film imaging achieve O.D. in linear portion of H+D curve (~ 1.4) Digital imaging is not contrast-limited, but noise limited How can you set up your AEC system to deliver the necessary amount of noise in an image? Fundamental difference from S/F: Wide range of K a will yield usable images, must decide on acceptable noise level in images Rong XJ, Shaw CC, Liu X, Lemacks MR, Thompson SK, Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems A contrast-detail phantom study, Med. Phys. 28(11):2328-35, 2001. Use of scored images of contrast-detail phantom (CDRAD) under clinical image processing conditions to attempt to determine detector exposure required to achieve similar detectability for small low-contrast targets Huda W, Slone RM, Belden CJ, Williams JL, Cumming WA, Palmer CK, Mottle on computed radiographs of the chest in pediatric patients, Med. Phys. 199:242-252, 1996. Use of a rating system based on degrees of mottle to determine what sensitivity number of a Fuji AC series would be needed to match the mottle in a 600-speed s/f system, then use the actual sensitivity numbers measured to compare CR to s/f. Conclusion: 0.43 mr required to achieve similar LCD for 0.5 mm object compared to 2.09 mr for the same screen-film image (21%) Conclusion: To achieve comparable mottle to 600-speed s/f system a twofold increase in exposure is needed as compared to mean sensitivity number. An exposure consistent with a 200-speed s/f system would be needed to achieve negligible mottle. 4

Liu X and Shaw, CC, a-si:h/csi(tl) flat-panel versus computed radiography for chest imaging applications: image quality metrics measurement, Med. Phys. 31(1):98-110, 2004. Compare indirect digital radiography and cassette-based digital radiography systems in terms of fundamental image quality metrics. Give radiologists a visual idea of what the impact of dose to the image receptor is on image quality CDRAD phantom exposed under PMMA Conclusion: Indirect digital radiography detector has significantly higher DQE and lower NPS than cassette-based digital radiography detector. The magnitude of the differences is illustrated in the graphs from the manuscript. Can also be done with other phantoms ACR R/F phantom, anthropomorphic phantom, or even add noise to a patient image for comparison Caveats Image processing has a large impact on noise (low-contrast detectability) and high-contrast resolution, and thus AEC sensitivity should be configured with this in mind. Also, pixel size has an impact on noise in images this is especially important for digital receptors where pixel size is variable, such as PSP systems I have a confession to make I have been using a very simplistic method of calibrating our AEC systems for sensitivity (cassette- based) Eight inch PMMA phantom imaged using 80 kvp Processed with Sensitivity S number used as indicator of proper calibration Properly calibrated reader DR compare with acceptance testing data It ain t that easy Methods for calibrating sensitivity Noise-based method K a -based methods Use a CR cassette with a cutout for a detector Solid-state detector behind grid Pre-detector K a and primary transmission through grid * Use an exposure indicator (EI) A word on exposure indicators Sometimes the relationship between EI and detector exposure is well understood or intuitive Cassette-based digital radiography Sometimes the relationship between EI and detector exposure is not obvious/proprietary Exposure indicator must be verified Goldman AAPM Summer School Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51:5475-5485, 2006. 5

Noise-based AEC calibration For each kvp, an S number can be associated with the standard deviation of pixel values that produce the desired noise characteristics STDEV is proportional to the square root of the x-ray x fluence absorbed in the IP (after correcting for structured noise) Recommend use of an -based threshold for choosing noise level Example: threshold of 5 Signal difference of 50 can be seen with σ = 10 Still some determination to be done Setting up AEC in Semi-automatic EDR mode yields clinically valid results in Automatic EDR mode Christodoulou, EG, Goodsitt, MG, Chan, H, and Hepburn, T, Phototimer setup for CR imaging, Med. Phys. 27:2652-2658, 2000. K a -based methods CR cassette with cutout S = 200/E, E = exposure (mr( mr) ) for Fuji CR Cutout machined into IP cassette to accept 15 cm pancake ionization chamber Introduces realistic scatter from cassette, overcomes cassette sensing Doyle P, Gentle D, Martin CJ, Optimising automatic exopsure control in computed radiography and the impact on patient dose, Rad. Prot. Dosim. 114:236-39, 2005. K a -based methods Solid-state dosimeter with lead backing should be used to eliminate the effect of backscatter Not possible for all DR systems http://www.rpi.edu/dept/radsafe/public_html/radi ATION%20DETECTORS_files/MOSFET.jpg http://www.unfors.com/products.php?catid=168 Pre-detector K a Simple measurement to make Scatter avoidable Necessitates measurement of primary transmission of anti-scatter grid Some inaccuracy introduced due to beam hardening by grid which is not accounted for in this measurement Doyle P, and Martin CJ, Calibrating automatic exopsure control devices for digital radiography, Phys. Med. Biol. 51:5475-85, 2006. Measuring grid characteristics 180 cm focused grid measurement Anti-scatter grids are labeled with information such as: Line rate Grid ratio Interspacer material Focus distance They are not, however, labeled with other information Contrast improvement factor Primary attenuation Fetterly/Schueler poster SU-GG GG-I-153 Chamber 1 Grid Patient-equivalent phantom Detector assembly Chamber 2 6

Measuring the Bucky factor Rear chamber behind grid Entire setup Typical exposure readings Measuring for clinical grid A level grid is imperative Measurement w/o grid 100 cm focused grid measurement Table Chamber 1 Support for grid Chamber centered in field Patient-equivalent phantom Grid Foam support Chamber 2 Wood support Table Detector assembly Centering and leveling of grid is imperative Complete setup EI-based methods Once the EI has been verified to be accurate across the range of kvps and patient thicknesses used and seen clinically, it should be used to perform these tests EI verification involves some of the same skills and measurements discussed here More from Jeff Shepard coming up You re not done yet 7

The Problem http://www.instablogsimages.com/images/2007/11/23/digital-flatpanel-x-ray-detector_28.jpg Many of our AEC systems have been calibrated for use with screen-film systems The energy response of gadolinium oxysulfide (Gd 2 O 2 S) screens is substantially different from that of image receptors used in digital radiography Thus, to properly expose digital radiographs, we must recalculate the kvp correction curve for our AEC systems to respond correctly considering the image receptor characteristics Absorption cross section (cm 2 /g) 1.00E+04 1.00E+03 1.00E+02 1.00E+01 CsI Gd2O2S BaFBrI The Solution Beam-quality dependent calibration curves for AEC systems used for digital radiography 1.00E+00 1.00E-01 0 20 40 60 80 100 120 140 160 Photon energy (kev) Doyle and Martin have calculated theoretical kvp correction curves for both CR and IDR detectors Also note that the addition of small amounts of Cu filtration does not significantly affect the calibration Pixel value 800 700 600 500 400 300 200 100 0 kvp Tracking Unit #1 120 100 80 60 40 20 0 Noise/ Pixel value Noise 60 70 80 90 100 110 120 130 kvp Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51:5475-5485, 2006. 8

Thickness tracking kvp Tracking - Unit #2 Patient thickness tracking 225 220 80 70 100.00 99.00 98.00 Unit 1 Unit 2 76.0 75.0 74.0 Pixel value 215 210 205 200 60 50 40 30 Noise/ Pixel value Noise Unit 1 97.00 96.00 95.00 94.00 93.00 92.00 91.00 73.0 72.0 71.0 70.0 69.0 68.0 67.0 Unit 2 195 190 185 50 70 90 110 130 kvp 20 10 0 90.00 1 2 3 4 5 6 7 8 9 PMMA (in.) Patient thickness tracking not only adjusts exposure time but also involves beam quality dependencies of detector absorption efficiencies and AEC sensitivity as well as scatter absorption by the AEC system and the contrast improvement factor of the grid. 66.0 Thickness tracking The overall variation I have seen when kvp tracking and thickness tracking are combined is 25-30% for one vendor (60 kvp) and 15% for another vendor When considering the combined impact of kvp and thickness variation, there is likely to be less variation for digital radiography systems () than in S/F (O.D.) due to the fact that the variation in Bucky factor over a range of kvp s no longer matters. Methods to attempt to improve on patient thickness tracking might include Further tightening kvp correction curve Calibrating with and using thin metal filters What about phantoms? Simple, uniform phantoms are desired for AEC testing Also, consulting physicists probably don t want to lug around 65 lbs. of PMMA and engineers certainly won t either But, what about realistic patient scatter Phantoms for AEC calibration Moral of the story: PMMA, water, and aluminum, in the appropriate amounts, deliver similar transmitted spectra to the AEC system and digital receptor So did 0.2 mm of Cu, but fluence was too high to achieve reasonable exposure times. 2 mm of Cu significantly altered the transmitted spectrum Introduction of scatter is a different situation Scatter and AEC calibration Scatter does influence absolute calibration of AEC sensitivity, however, but does not have a large impact on the kvp correction curve Many manufacturers do adjust the sensitivity of the AEC system based on the presence/absence of the anti-scatter grid Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51:5475-5485, 2006. 9

Future of Automatic Exposure Control in Radiography Fluoroscopy/angiography has already altered the way ADRC is performed Monitoring signal value in a region of pixels Detector is read many times per second These methods with TFT arrays do not work in radiography Detector readout is destructive Fischer stereo unit CCD array (Tony Siebert) CMOS (Tony Siebert) Pixel architecture, with multiple transistors, allows for sampling without destroying the contained information US Patent 5937027 Thevenin B, Glasser F, Martin J-L, J Method and device for the taking of digital images with control and optimization of the exposure time of the object to X or g radiation. Sobol W CCD array with two classes of pixels Integrator/comparator Image corrected later Other resources Goldman LW, Yester MV, Specifications, Performance Evaluations, and Quality Assurance of Radiographic and Fluoroscopic Systems in the Digital Era, 2004 Summer School Proceedings Sobol WT, Advances in and specifications for radiographic X-ray systems, pg. 1-681 Goldman LW, Speed values, AEC performance evaluation, and quality control with digital receptors, pg. 271-297 297 10