Update on ACR Digital Mammography QC Manual



Similar documents
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Quality Control of Full Field Digital Mammography Units

Update on MQSA and ACR Mammography Accreditation

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Digital Breast Tomosynthesis QC Requirements

American College of Radiology Mammography Accreditation Program

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

INTRODUCTION. A. Purpose

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Quality Control and Maintenance Programs

Performance testing for Precision 500D Classical R/F System

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Aspire FCRm/CRm Fujifilm CR for Mammography

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Digital Mammography Update: Design and Characteristics of Current Systems

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

MQSA and ACR Digital Breast Tomosynthesis Mammography Accreditation

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

MQSA Quality Control Manual for Monochrome Displays for Mammography

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Physics testing of image detectors

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Breast MRI Quality Control

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Digital Mammography Quality Control

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Understanding Digital Modalities: System Integration and Use

MQSA, FDA Regulations, and Inspections

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

MRI QA Technologist s Tests

Supports screening, diagnostic, and multimodality workflows

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Show that when a circle is inscribed inside a square the diameter of the circle is the same length as the side of the square.

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

Protocol Management and Review Strategies The MD Anderson Experience

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Measuring Size and Distance in Digital Mammography

The Exposure Index and Its Standardization

How To Compare Mammography To A 3D System

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)

Diagnostic Exposure Tracking in the Medical Record

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Tallahassee Community College PERIMETER

MRI QA Technologist s Tests

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

MIN-R 2000 Plus Film

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Chapter 15. Creating the Digital Image. Objectives. Key Terms

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Assessing Radiation Dose: How to Do It Right

2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Quantitative Imaging In Clinical Trials Using PET/CT: Update

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Checklist for submission of application form for Registration

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

The Mammography Quality Standards Act Final Regulations Document #4; Final Guidance for Industry and FDA

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

MediCal QAWeb. Secured online service for Quality Assurance of medical displays

Proton tracking for medical imaging and dosimetry

Elements of required physical infrastructures: space, schielding, and patient flow..

Butt Fusion Welding of HDPE Pipes Work Procedure

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

Worldwide Quality Assurance networks for radiotherapy dosimetry

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

DR Retrofit Kit 1417 WL Configuration Integrated PACS

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Perimeter is the length of the boundary of a two dimensional figure.

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

Perimeter, Area, and Volume

The Mammography Quality Standards Act Final Regulations Document #1

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

WHERE IN THE WORLD JILL LIPOTI?

Transcription:

Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual Explain what we feel is important Thank you for being here today 1

Digital Mammography in the US (as of 3/1/15) 13,493 units at 8720 facilities Over 96% of all units in US are FFDM *4 accrediting bodies *Over 30 models and mfrs approved by FDA 2

Quality Control: What It and Why Is It Important? Primary Purpose Reduce exposure to patients and personnel Ensure adequate and consistent patient image quality Detect and correct for potential problems, before they impact patient image quality and care What it s not: Not a detailed technical evaluation of a unit Not a detailed measure of a limits of a unit Not the optimization of a unit QC Mandates MQSA Mammography Quality Standards Act ACR American College of Radiology 3

FDA s Current DM QC Requirements Follow latest version of mfr s QC manual procedures for unit tested Lorad (Hologic) allows facility to follow any of their manuals Meet mfr s performance standards Failures must be fixed before use on patients Most mfrs applied for alternative standards to allow 30 days for some QC tests ACR s Current DM QC Requirements Same as FDA s Which are the same as the manufacturer s ACR suggests using manufacturer s data forms 4

Screen-Film QC Single detector technology Single processing technology First QC manual in 1990 Manufacturer s DM QC Requirements Tests vary for each manufacturer and model Some tests same but names different Some tests not required by some manufacturers Frequencies vary for each manufacturer and model Procedures vary for each manufacturer and model Pass/fail criteria vary for each manufacturer and model 5

Film Printers FDA recommends only using printers cleared by FDA s Office of Device Evaluation for FFDM (but may legally use others) Facility must have access to a laser printer (either on-site or someplace else) Printer must exist and be tested by MP before the facility performs mammography Monitors and Workstations FDA regs require facilities to comply with a QA program substantially the same as that of FFDM mfr (i.e., GE, Lorad, etc.) Impractical; even impossible since tests are softwarebased FDA says If the monitor/workstation has been cleared by FDA s ODE for FFDM, the monitor s QC manual is substantially the same and OK to follow If monitor was not cleared by FDA ODE for FFDM facilities must follow one by FFDM mfr FDA ODE clears monitors/workstations Over 500 total;? have been cleared for FFDM 6

ACR DM QC Manual VS. All of the above may vary with QC manual revisions of same manufacturer/model ACR DM QC Manual Project Subcommittee on Quality Assurance Chair Eric Berns, PhD University of Colorado Subcommittee all volunteers (with day jobs) Clinical representatives MITA representatives (equipment and phantom manufacturers) 7

ACR DM QC Manual Project Subcommittee Charge: Design ACR Accreditation Phantom for FFDM Write QC Manual for ACR FFDM Mammography Accreditation Program Subcommittee Goals: Standardize all QC tests for all digital manufacturers Standardize test frequencies Standardize performance criteria ACR DM QC Manual Project QC tests: Will apply to all manufacturers Tests come from a variety of sources (MQSA, SFM, ACRIN DMIST, Manufacturer s QC programs, MITA, subcommittee clinical experience, etc.) Clinically relevant User friendly Eliminate non-productive testing Just because you can test something, doesn t mean you should! 8

ACR Digital QC Manual - Structure Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section Appendices **Clinical Image Quality Section (w/patient Positioning and Compression and Clinical Image Quality Evaluation) will be revised at a later date and posted on ACR website ACR Digital QC Manual - Tools Will be downloadable to computer or tablet Future revisions fully described (in new Revisions section) Sections and references fully linked Optional procedures and forms available Management forms available to aid organization Procedures liberally illustrated Forms in PDF and Excel formats Excel formats contain formula and functionality Excel forms downloadable from ACR website 9

ACR DM QC Manual Project Subcommittee Goals: Standardize all QC tests for all digital Mfrs Standardize test frequencies Standardize performance criteria What Will Be New? 10

The ACR DM Phantom Prototype Phantom Prototype Design Principles 11

Phantom Prototype Design Principles Phantom Prototype Design Principles Based on existing ACR Accreditation Phantom Similar imaging and scoring to current phantom Build on experience of QC techs and physicists at ~8,700 US facilities who already know how to use and score the existing phantom (~25,000+ techs) 12

Phantom Prototype Design Principles Can be used on both SFM & FFDM Total attenuation matched to current SFM phantom Similar thickness Similar total dose Permits testing of the MQSA 3.0 mgy dose limit (single CC view) Proposed Scoring Changes Eliminate subtraction for artifacts Add Fail for artifacts New pass/fail criteria from 4,3,3 To: 2,3,2 **But, objects are the same (effective) size as SFM Phantom 13

ACR Phantom Prototype Depth of CNR Cavity = 0.1 + 0.005 cm Air Gap = 0.027 cm Nominal Tolerances (Insert Well & CNR Cavity) Wax insert well depth : + 0.005 cm (+ 2 mils). Wax insert well width and length : + 0.04 / -0.00 cm. CNR cavity depth : + 0.005 cm (+ 2 mils). CNR diameter : + 0.05 cm. Cover =Nominal 0.3 cm Total Thickness = 4.10 + 0.03 cm Wax = 0.70 cm + 0.02 cm Compensator = 0.023 cm Total Insert Depth = 0.75 cm Total Thickness Under Insert = 3.05 cm Test object distance from base of wax = 0.35 + 0.10 cm 31.0 + 0.1 cm 2.0 + 0.05 cm Centered Left to Right CNR Cavity (0.1 + 0.005 cm Deep) 9.5 cm 19.0 + 0.1 cm 1.0 + 0.05 cm Milled out wax insert area 13.0 cm (+0.04, -0.00 cm) ID Tag Milled out wax insert area 7.0 cm (+0.04, -0.00 cm) 2.5 + 0.05 cm 1.0 + 0.05 cm 1.0 + 0.05 cm Screws Wax Insert Specifications with Virtual Placement Grid ID Tag Specs 12.98 cm (+ 0.00, - 0.04 cm) 1.5 cm Virtual Box: height = 0.5 cm, length = 1.8 cm Location of Virtual Box ID Tag 0.49 cm 2.0 cm 6.98 cm (+ 0.00, - 0.04 cm) 2.0 cm 2.0 cm 0.49 cm 0.49 cm 2.0 cm each 0.49 cm Notes: Test objects to be centered on their respective placement grid locations. 0.49 cm perimeter around test object placement grid. 0.635 cm (1/4 inch) radius on corners of wax insert. Fiber Placement specs Fiber specifications Fiber Fiber Fiber Length = 1.0 cm + 0.1 cm Fiber Diameter = See Table 45 o + 5 o 45 o + 5 o Speck Placement & Specs 1. Specks to be placed at points on star and middle of star 2. Speck Size (spherical) = See Table 3. Center speck placement to be within + 0.1 cm of center of virtual grid 4. Distance from center speck to center of speck on perimeter = 0.5 cm + 0.1 cm Mass Placement & Specs 1. Mass pre-cut sphere diameter = 5/8 inch 2. Mass placement to be within + 0.1 cm of center of virtual grid 14

Wax Insert Test Object Specifications Test Object Fiber Diameter Speck Diameter (Glass Spheres) Mass Thickness mm mm mm 1 0.89 + 0.05 0.33 + 0.0100 1.00 + 0.05 2 0.75 + 0.03 0.28 + 0.0083 0.75 + 0.05 3 0.61 + 0.03 0.23 + 0.0069 0.50 + 0.05 4 0.54 + 0.03 0.20 + 0.0059 0.38 + 0.04 5 0.40 + 0.03 0.17 + 0.0084 0.25 + 0.03 6 0.30 + 0.03 0.14 + 0.0070 0.20 + 0.02 Summary of Test Object Visual Equivalency Test Object Fibers (mm) Specks (mm) Masses (mm) ACR 156 FFDM ACR 156 FFDM ACR 156 FFDM 1.56 1.12 0.54 2.00 0.89 0.89 0.40 1.00 1.00 0.75 0.75 0.32 0.33 0.75 0.75 0.61 0.28 0.50 0.50 0.54 0.54 0.24 0.23 0.38 0.40 0.40 0.20 0.25 0.25 0.30 0.16 0.17 0.20 0.14 15

Pass/Fail Criteria Test Object Fibers (mm) Specks (mm) Masses (mm) ACR 156 FFDM ACR 156 FFDM ACR 156 FFDM 1.56 1.12 0.54 2.00 Fail Pass 0.89 0.89 0.40 1.00 1.00 0.75 0.75 0.32 0.33 0.75 0.75 4 2 3 3 2 0.61 0.28 0.50 0.50 0.54 0.54 0.24 3 0.23 0.38 0.40 0.40 0.20 0.25 0.25 0.30 0.16 0.17 0.20 0.14 The ACR DM Phantom Prototype 16

Image of Entire Phantom Serial Number *Note: Gray dot in lower left corner of wax insert is an artifact due to a bubble in wax insert. Wax Insert Serial Number 17

Expanded View of Wax Insert Pass Criteria: Equivalent to SFM Phantom: 2 Fibers, 3 Specks, 2 Masses 4 Fibers, 3 Specks, 3 Masses Pass Serial Number 18

Effects of Thickness Equalization Serial Number New FFDM phantom equalizes attenuation inside and outside wax insert. This permits evaluation of artifacts over entire phantom area with same WW and WL used to score test objects. Benefits of Prototype Phantom Design Provides view of entire detector artifact evaluation W/L optimized for test objects optimizes for artifact evaluation Finer gradations and smaller sizes of test objects AGD measurement & limit same as SFM meets MQSA Provides single image/exposure for evaluation(s) Minimal training Provides basis for monitor and printer QC ACR Physics Reviewers Can see scores and artifacts on single submitted film (or image) Do not need different WW/WL settings 19

Technologist Section 20

21

22

23

24

25

26

Also: Tally sheet Daily Counting Sheet 27

Technologist Management Forms 28

ACR Technique and Procedure Summaries (continued) Facility MAP ID-Unit# (00000-00) AW & RW Monitor QC Room ID Unit Mfr & Model ACR Technique and Procedure Summaries (continued) Procedure: Display the AAPM TG 18-QC or SMPTE test pattern. (Another pattern that allows relevant measurements may also be used.) Check the monitor screens to verify they are clean. Facility Room ID Verify that 0%-5% and 95%-100% boxes are visible (red circles). Verify that line-pair images are sharp and distinguishable (red rectangles). MAP ID-Unit# (00000-00) Unit Mfr & Model Score the ACR DM Phantom & evaluate for artifacts (RW only). If applicable, perform monitor calibration tests per manufacturer's recommendations. Film Printer QC Monitor manufacturer instructions for finding & viewing TG 18-QC test pattern: Procedure: Required equipment - ACR DM Phantom image and densitometer Print image acquired in "ACR DM Phantom Image Quality" test. Print from workstation/computer/pacs used for printing most clinical films. Do not adjust window and level settings prior to printing. Evaluate the phantom for artifacts & score. Note: If possible, print phantom image from same x-ray unit each time for this test. Note: Use film size most commonly used for mammography. Note: For each printer, only print a single image from a single x-ray unit or workstation. SMPTE Test Pattern TG 18-QC Test Pattern Film Printer Setup Date: Film Size (8 x 10 or 10 x 12): ID of x-ray unit or workstation for printed phantom: Dmax Note: Dmax to be measured on perimeter of film. If not available, print a clinical image and measure in the non-breast area. Background OD Cavity OD Dmax SN-1234 29

30

31

Medical Physicist Section 32

2. ACR DM Phantom Image Quality Facility Name Imaging Facility Room ID, Mfr & Model ACR DM Phantom Mfr and S/N Phantom Mfr 1234 MAP ID-Unit# (00000-00) SID (cm) 70 Survey Date Room 1, Unit Mfr, Unit Model 12345-xx December 1, 2015 Equipment: ACR DM Phantom (required) Phantom Setup: AEC mode: Follow procedure in the Technologist's ACR Technique & Procedure Summaries: Use clinical technique for typical screening exam of 4.2 cm 50/50 breast Largest IR & paddle, 5 dan or 12 lbs, Score on AW Phantom Setup Adjust W/L to optimize test objects, zoom & pan entire image For Config 2 & 3 using kvp & mas closest to phantom techniques Paddle size (IR size): Paddle type (reg or flex): View or selected image: Compression force: AEC cell position (if avail): Phantom patient name: Target/filter (if app): kvp (if app): Phantom patient ID: Image sent to which PACS? Density setting (if app): Mag factor (mag mode only): Contact Mode Clinical - Manual - Manual - ACR DM Target/Filter Target/Filter Phantom Config 2 Config 3 Clinical - ACR DM Phantom Mag Mode Manual - Manual - Target/Filter Target/Filter Config 2 Config 3 Target/filter Resulting Techniques (if available) Image receptor size kvp mas largest Unit-indicated AGD (mgy) ACR DM Phantom Evaluation Artifacts P/F Fiber score Speck group score Mass score Phantom P/F DC offset (if applicable) Mean cavity signal SNR & CNR Raw Image Mean background signal Std dev of background Calculated SNR Calculated CNR #DIV/0! #DIV/0! SNR 40.0 (P/F) CNR 2.0 (P/F) Distance Measurement Parallel to A-C axis (mm) Meas = 70.0 ±14.0 mm (P/F) Analysis Fibers Full Point 8 mm long Overall Pass/Fail Initiated (or updated) technologist's ACR Technique and Procedure Summaries form Half Point 5 & <8 mm Specks Masses 4-6 specks ¾ border 2-3 specks ½ & < ¾ border Distance Required: ACR DM Phantom image must be free of clinically significant artifacts. Fiber score must be 2.0;; speck group score must be 3.0;; mass score must be 2.0. Action Limits SNR must be 40.0;; CNR must be 2.0;; measured wax insert distance must be 70.0 ± 14.0 mm. Timeframe: Failures of required items must be corrected before clinical use. 33

CNR Mean = 542.3 Mean = 498.5 St. Dev = 7.8 Serial Number 34

35

D = Kgcs D = Mean Glandular Dose K = Entrance surface air kerma g = glandularity of 50% c = corrects for difference in composition (age dependent) s = X-ray spectrum correction (Target/Filter) Note: g and c depend on thickness, glandularity, and HVL. Primary Ref: D.R. Dance, et al. Additional for the Estimation of Mean Glandualar Breast Dose Using the UK Mammography Dosimetry Protocol. Physics in Medicine and Biology 45, 3225-3240, 2000. 36

37

38

39

40

41

42

43

44

45

ACR DM QC Manual Status In September 2014, draft sent to manufacturers, select technologists and medical physicists for review We hope manufacturers will adopt this manual Extensive comments received in October Reviewed and incorporated as appropriate Updated draft sent to FDA in February 2015 Comments and instructions for implementation received from FDA late April 2015 ACR DM QC Manual Status (continued) ACR will respond to comments ACR must apply for FDA alternative standard under current regulations Alternative standard will allow facilities to use this instead of the manufacturer s manuals Tomosynthesis QC to be added as Appendix after FDA approval 46

Top Strengths & Take Home Points QC Manual is comprehensive Phantom is of major importance Most failures are artifacts Includes most legacy tests Includes most current manufacturer tests Accommodates mfr specific test (where app) Accommodates growth of QC program Manual can realistically be implemented End of Presentation Questions? 47