Vista 120 Patient Monitor



Similar documents
D A cost effective monitor that meets your clinical needs VISTA 120

Vista 120 Central Monitoring System

Moving beyond telemetry INFINITY M300

Polaris 5xx/7xx Medical Lights and Video

D Move life-critical patient data on your existing network INFINITY ONENET

OMNI TOUCH SCREEN PATIENT MONITOR

Oxylog 3000 plus Emergency & Transport Ventilation

Sekusept Cleaner Liquid Cleaning Agent Cleaning, Desinfecting and Drying

Dräger X-dock Release Notes Software-Update and Manager Software 1.7.0

Agila Ceiling Supply Unit Medical Supply Units

Infinity M540 monitor

Reference Case: Continuous mobile patient monitoring

Movita Ceiling Supply Unit Medical Supply Units

MT Monitoring and IT Solutions for supporting patient safety and care INFINITY IN-HOSPITAL TRANSPORT SOLUTIONS

Dräger PSS 3000 Self Contained Breathing Apparatus

INNOC. Innocare-T. Innomed. Innomed s intensive and anesthesiological monitor system can be used both as a transport or bed-side care unit

Dräger Globe-Trotter GT5400 Neonatal Transport

Your personal anesthesia assistant DRÄGER PRIMUS

Dräger Alcotest 9510 Evidential Alcohol Measuring Device

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one.

The bridge to more efficient care PONTA SUPPLY BEAMS

Monitoring and It solutions for supporting patient safety and care. InfInIty

D Next-generation LED phototherapy NanoBlu 500

Dräger Evita Infinity V500 ventilator ICU Ventilation and Respiratory Monitoring

Infinity Acute Care System Monitoring Solution

EC-12R EC-12RM EC-12R/S EC-12SResting and Stress ECG Family. Technical specification

DC Electronic Loads 8500 series

PIN CONFIGURATION FEATURES ORDERING INFORMATION ABSOLUTE MAXIMUM RATINGS. D, F, N Packages

Op Amp Circuit Collection

1701/22 FieldMonitor Management Interface Module

Dräger Babylog VN500 Neonatal Open Care

Analog signal converters CC-E I/I Current / current isolators

MPC 4. Machinery Protection Card Type MPC 4 FEATURES. Continuous on-line Machinery Protection Card

NI USB-5681 RF Power Meter Specifications

Introducing the first 12-lead for the BLS team.

Dräger Communication Systems

Voltech DC1000. Precision DC Bias Current Source

Rosemount 333 HART Tri-Loop

MEDICAL SYSTEMS. Monnal T60. Touch and Breathe.

All Products of MEDIANA. All because of you,mediana will be better...

U1602A Handheld Oscilloscopes, 20 MHz

Carina Pocket Guide Software version 3.n. The Carina Pocket Guide is not a replacement or substitute for the Instructions for Use.

How To Use A Welch Allyn 1500 Patient Monitor

MEDIANA Corporate Profile. Contents

HEARTSTART XL DEFIBRILLATOR/MONITOR Model: M4735A PRODUCT DESCRIPTION

iva Cable & Antenna Analyzer

M-D,T,Q-series BroadLighter. Benchtop Broadband Light Source with Extended Bandwidth. Technical Product Specification

The Anesthesia Ventilator

TL084 TL084A - TL084B

Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

iva Cable & Antenna Analyzer

Type: EASY719 DC RC Article No.: Ordering information Relay outputs Quantity 6 Power supply V DC 24 V DC. Description

PSM 900 and PSM 1000 Personal Monitor Systems TWO SYSTEMS. INFINITE POSSIBILITIES.

Digital Pressure Measuring Instrument MDR480

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

AIS 3000 APC AIS kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Process modules Digital input PMI for 24 V DC inputs for 120 V AC inputs

Cardiac Conduction System (1) ECG (Electrocardiogram) Cardiac Conduction System (2) The ECG (1) The ECG (1) The ECG (1) Achmad Rizal BioSPIN

Welch Allyn 1500 Patient Monitor Train the Trainer

Aliseo Anesthesia system

Arbitrary/Function Generators AFG3000C Series Datasheet

Rigel 333 Multi Parameter Patient Simulator Version 2.0

Differential Charge Amplifier

Multi-Range Programmable DC Power Supplies 9115 Series

SOLAR300N Rel /01/10

RRV934. Multi-controller. Synco living

SpO2 PULSE OXYMETER MD-300

PHOENIX CONTACT Features

CONTROLS DATA MANAGEMENT PROCESS AUTOMATION EUROCUBE. General purpose single phase thyristors and solid state relays Product data.

The information in this document is subject to change without notice.

Limited Ericsson Internal

3500/40M Proximitor* Monitor

1.44 kw Programmable DC Power Supplies XLN Series

Wahl C50 Calibrator FEATURES

Application Note TMA Series

5000V Graphical Megohmmeter Model 5070

Datex-Ohmeda Critical Care Monitor

Fluke 1650B Series Multifunction Installation Testers

Disturbance Recoder SPCR 8C27. Product Guide

A compact, lightweight, portable optical spectrum analyzer for DWDM system installation and maintenance.

Energy Management Energy Meter Type EM23 DIN

Hand-held thermometer Model CTH7000

RISH EM 3490 DS Dual Source Energy Meter RISH EM 3490 DS. Application : Product Features:

High-ohmic/high-voltage resistors

Portal Software GSM-PRO Run LED displays module activity Com LED displays activity on the GSM network GSM-PRO

+ A B V B NO NC V COM PV PV

SINEAX U553 Transducer for AC Voltage

Agilent 4338B Milliohm Meter

The Electrocardiogram (ECG)

FM Radio Transmitter & Receiver Modules

Concept. Monitoring Unit. Patient Box. Defibrillator/Pacer Unit.

Building a Simulink model for real-time analysis V Copyright g.tec medical engineering GmbH

CardioCard System. Product Specification and Sample Reports. PC-Based ECG CardioCard. PC-Based Stress. PC-Based Holter. CardioCard.

Company Profile & product presentation.

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

D Gas supply secure at a glance DRÄGER ALARM MANAGEMENT SYSTEM

Product Information S N O. Portable VIP protection CCTV & Alarm System 2

MIT510/2, MIT520/2 and MIT1020/2

APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out

Transcription:

Vista 120 Patient Monitor The Vista 120 offers essential monitoring capabilities at an exceptional value and meets the needs of a variety of care environments for adult, paediatric and neonatal patients. The 380 mm (15 inch) colour touch screen provides a clear view of vital patient data, and the quick access menu, hard keys and rotary knob enable easy navigation for the clinician. D-32041-2011 D-3531-2011 FEATURES Versatile: provides a core set of essential parameters, plus optional advanced measurements Scalable: offers a choice of two models to meet diverse clinical needs Visible: large 380 mm (15 inch) colour touch screen provides a clear view of patient data Easy to use: provides quick access menu, hard keys, and a rotary knob Built-in recorder: saves time by providing documentation when needed OR support: includes gas analyser capability TECHNICAL DATA D-3527-2011 Vista 120 SUPPORTED PARAMETERS ECG Lead mode 3-lead wire: I, II, III 5-lead wire: I, II, III, avr, avl, avf, V Waveform 3-lead wire: 1-channel waveform 5-lead wire: 2-channel waveform, max. seven waveforms Lead naming style AHA, IEC Display sensitivity 1.25 mm/mv ( 0.125), 2.5 mm/mv ( 0.25), 5 mm/mv ( 0.5), 10 mm/mv ( 1), 20 mm/mv ( 2), AUTO gain Sweep 6.25, 12.5, 25, 50 mm/s Bandwidth (-3 db) Diagnosis: 0.05 to 150 Hz Monitor: 0.5 to 40 Hz Surgery: 1 to 20 Hz CMRR Diagnostic: > 95 db (the Notch filter is off) (Common Mode Rejection Ratio) Monitor: > 105 db (the Notch filter is on) Surgery: > 105 db (the Notch filter is on) Notch 50 Hz/60 Hz (Notch filter can be selected manually) Differential input impendance > 5 MΩ Input signal range ±10 mvpp Electrode offset potential tolerance ±500 mv

02 VISTA 120 PATIENT MONITOR VISTA 120 PATIENT MONITOR 03 Auxiliary current Active electrode: < 100 na (Leads off detection) Reference electrode: < 900 na Recovery time after defibrillation <5 s Leakage current of patient <10 μa (normal condition) Scale signal 1 mvpp, accuracy is ±5 System noise <30 μvpp ESU protection Cut mode: 300 W Coagulation mode: 100 W Recovery time: 10 s Meets the requirements of ANSI/AAMI EC13-2002: Sect. 4.1.2.1 (a) ESU noise suppression Tested according to the test method in ANSI/AAMI EC13-2002: Sect. 5.2.9.14, it accords with the standard. Minimum Input Slew Rate (Lead II) >2.5 V/s of sinus and SV rhythm Tachycardia Normal Bradycardia of ventricular rhythm Ventricular rhythm Ventricular bradycardia ADU: 120 to 300 bpm PED/NEO: 160 to 350 bpm ADU: 41 to 119 bpm PED/NEO: 61 to 159 bpm ADU: 15 to 40 bpm PED/NEO: 15 to 60 bpm The interval of 5 consecutive ventricular complexes is less than 600 ms The interval of 5 consecutive ventricular complexes ranges from 600 ms to 1,000 ms The interval of 5 consecutive ventricular complexes is higher than 1,000 ms Pace pulse Pulse indicator Pulse rejection Heart rate Sensibility PVC ST value HR averaging method 1 2 Pulses are marked if the requirements of ANSI/AAMI EC13:2002, Sect. 4.1.4.1 are met: Amplitude: ±2 to ±700 mv Width: 0.1 to 2 ms Ascending time: 10 to 100 μs For heart rate meter, pulse is rejected if the requirements of ANSI/AAMI EC13:2002, Sect. 4.1.4.1 are met: Amplitude: ±2 to ±700 mv Width: 0.1 to 2 ms Rise time: 10 to 100 μs ADU: 15 to 300 bpm PED/NEO: 15 to 350 bpm ±1% or ±1 bpm, whichever is greater 1 bpm 300 μvpp ADU: 0 to 300 PVCs/min PED/NEO: 0 to 350 PVCs/min 1 PVCs/min or 2% of measurement, whichever is greater 1 PVCs/min -2.0 to 2.0 mv The max. of ±0.02 mv or 10% (-0.8 to 0.8 mv) 0.01 mv Heart rate is computed by excluding the minimum and maximum values from the 12 most recent RR intervals and averaging the residual 10 RR intervals. If each of three consecutive RR intervals is greater than 1,200 ms, then the four most recent RR intervals are averaged to compute the HR. Startup time for tachycardia Gain 1.0: 10 s 1 mv 206 bpm Gain 0.5: 10 s Gain 2.0: 10 s Gain 1.0: 10 s 2 mv 195 bpm Gain 0.5: 10 s Gain 2.0: 10 s Response time of heart rate meter HR range: 80 to 120 bpm to change in HR : 7 to 8 s, average is 7.5 s HR range: 80 to 40 bpm : 7 to 8 s, average is 7.5 s Tall T-wave rejection Exceeds ANSI/AAMI EC13-2002 Sect. 3.1.2.1 (c) minimum recommended 1.2 mv T-Wave amplitude of heart rate meter and According with ANSI/AAMI EC13-2002 Sect.4.1.2.1 (e) response to irregular rhythm The HR value after 20 s: Ventricular bigeminy: 80 ±1 bpm Slow alternating ventricular bigeminy: 60 ±1 bpm Rapid alternating ventricular bigeminy: 120 ±1 bpm Bidirectional systoles: 91 ±1 bpm Respiration Impedance between RA-LL, RA-LA Baseline impedance range 200 W to 2,500 W (with ECG cables of 1 KW resistance) Measuring sensitivity 0.3 Ω (baseline impedance 200 to 4,500 Ω) Noise <0.12 Ω (3, 5-lead monitoring) Max. dynamic range 500 Ω resistance, 3 Ω variable resistance, no clipping Waveform bandwidth 0.2 to 2.5 Hz (-3 db) RR measuring and alarm range: Adult: 0 to 120 rpm Neo/Ped: 0 to 150 rpm 1 rpm ±2 rpm Gain selection 0.25, 0.5, 1, 2, 3, 4, 5

04 VISTA 120 PATIENT MONITOR VISTA 120 PATIENT MONITOR 05 NIBP Mode Measuring interval in auto mode Continuous Measuring type Alarm type Measuring and alarm range Adult mode Paediatric mode Neonatal mode Cuff pressure measuring range Pressure resolution Maximum standard deviation Maximum measuring period Adult/Paediatric Neonate Typical measuring period Overpressure protection Adult Paediatric Neonatal PR Oscillometric Manual, Auto, Continuous 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, and 480 min 5 min, interval is 5 s Systolic Pressure, Diastolic Pressure, Mean Pressure SYS, DIA, MAP SYS: 40 to 270 mmhg DIA: 10 to 215 mmhg MAP: 20 to 235 mmhg SYS: 40 to 230 mmhg DIA: 10 to 180 mmhg MAP: 20 to 195 mmhg SYS: 40 to 135 mmhg DIA: 10 to 100 mmhg MAP: 20 to 110 mmhg 0 to 300 mmhg 8 mmhg 120 s 90 s 30 to 45 s (depend on HR/motion disturbance) 297 ±3 mmhg 245 ±3 mmhg 147 ±3 mmhg 40 to 240 bpm ±3 bpm or 3.5%, whichever is larger Nellcor Module Measuring 1% to 100% Alarm 20% to 100% 1% Data update period 1s (70% to 100% SpO2) : DS-100A, OXI-A/N(Adult) ±3% OXI-A/N(Neonate) ±4% D-YS (Infant to Adult) ±3% D-YS (Neonate) ±4% D-YS with D-YSE Ear Clip ±3.5% MAX-FAST ±2% Pulse Rate Measuring 20 bpm to 300 bpm 1 bpm 3 bpm (20 bpm to 250 bpm) Sensor Wave length approximately 660 and 900nm Emitted light energy <15 mw NOTE Information about the wave length range can be especially useful to clinicians (for instance, when photodynamic therapy is performed). Temperature Channels 2 Measuring and alarm range 0 to 50 C (32 to 122 F) Alarm range 0 to 50 C (32 to 122 F) Sensor type YSI 10 kω ±0.1 C (0.2 F) (without sensor) ±0.1 C (0.2 F) Refresh time Every 1 s to 2 s IBP Dynamical pressure measuring range -50 to 300 mmhg ±2% or ±, whichever is greater SpO2 0 to 100% Alarm range 0 to 100% 1% Adult (including Paediatric) Neonate ±2% (70 to 100% SpO2) Undefined (0 to 69% SpO2) ±3% (70 to 100% SpO2) Undefined (0 to 69% SpO2) PI Measuring 0 10. It displays 0 for invalid PI value 1 Pulse rate Pulse rate measuring range Alarm range Data update period 25 to 300 bpm 30 to 300 bpm ±2 bpm 1 s Pressure sensor Sensitivity Impedance Frequency response Zero Measuring and alarm range Art PA CVP/RAP/LAP/ICP P1/P2 CO2 Unit FiCO2 AwRR 5 (μv/v/mmhg) 300 Ω to 3,000 Ω DC to 12.5 Hz or DC to 40 Hz : ±200 mmhg : ± 0 to 300 mmhg -6 to 120 mmhg -10 to 40 mmhg -50 to 300 mmhg Infra-red absorption technique mmhg, %, kpa 0 to 150 mmhg 3 to 50 mmhg 0 to 150 rpm (Mainstream)

06 VISTA 120 PATIENT MONITOR VISTA 120 PATIENT MONITOR 07 FiCO2 AwRR accuracy AwRR accuracy Apnea alarm delay Calculation method Stability Short term drift Long term drift 1 rpm ±2 mmhg, 0 to 40 mmhg ±5% of reading, 41 to 70 mmhg ±8% of reading, 71 to 100 mmhg ±10% of reading, 101 to 150 mmhg ±1 rpm 10, 15, 20, 25, 30, 35, 40 s, default value is 20 s BTPS (Body Temperature Pressure Saturated) Drift over 4 hours <0.8 mmhg specification will be maintained over 120 hours Recorder Record width Paper speed Trace Recording types 48 mm (1.9 inch) 25, 50 mm/s Up to 3 waveforms 8 second real-time recording Auto 8 second recording Parameter alarm recording Trend recording Drug calculation and titration table recording Review recording Display specifications Display screen 380 mm (15 inch) colour TFT 1024 768 Maximum number of waveforms 11 Indicator LEDs 1 power, 2 alarm, 1 charge O2 compensation 0 100% 1% Default 16% Intended patient Adult Measurement method Thermodilution Technique Physical specification Size H W D Weight Electrical specification Power supply Pmax FUSE 316 408 157 mm (12.4 16.1 6.2 inch) 7.0 kg (15.4 lbs) 100 V - 240 V~, 50 Hz/60 Hz 110 VA T 3.15 AH, 250 V TB TI TB, TI 0.1 L/min ~ 20L/min 23 C~43 C -1 C~27 C 0.1 L/min ±0.1 C Classification Anti-electroshock type EMC type Anti-electroshock degree Ingress protection Disinfection/sterilisation method Working system Class I equipment and internal powered equipment Class A CF: ECG (RESP), TEMP, IBP, CO BF: SpO2, NIBP, CO2, AG IPX1 Refer to Instructions for Use: Care and cleaning Continuous running equipment TB TI Trend review Short Long Review ±5% or 0.2 L/min, whichever is greater 0.1 C(without sensor) 0.1 C(without sensor) 1 hr, 1 s. resolution 120 hrs, 1 min. resolution 1200 sets NIBP measurement data Battery specification Type Rechargeable lithium-ion Quantity 1 Capacity 4200 mah 5000 mah Battery Life 4200 mah 180 min (At 25 C, standard configuration, recording set to off, brightness set to 1) 5000 mah 240 min (At 25 C, standard configuration, recording set to off, brightness set to 1) Battery charge time 4200 mah 320 min (Monitor is on or in standby mode) 5000 mah 360 min (Monitor is on or in standby mode) NOTE: Regarding the AG specifications, refer to the Supplement Scio Four modules.

08 VISTA 120 PATIENT MONITOR ENVIRONMENTAL REQUIREMENTS The monitor may not meet the performance specifications given here if stored or used outside the specified temperature and humidity ranges. When the monitor and related products have differing environmental specifications, the effective range for the combined products is that range which is common to the specifications for all products. Temperature range Relative humidity Atmospheric pressure 0 to 40 C (41 to 104 F) -20 to 55 C (-4 to 131 F) 15 to 95 % (non-condensing) 15 to 95 % (non-condensing) 860 to 1060 hpa 700 to 1060 hpa Standards EN 60601-1:2006/AC:2010; EN 60601-1: 1990+A1:1993+ A2:1995; EN 60601-1-2:2007 + AC 2010; EN 60601-6: 2010; EN 60601-1-8: 2007 AC 2010; EN ISO 10993-1: 2009; EN ISO 14971: 2012; EN ISO 17664: 2004; EN 62304: 2006; EN 980: 2008; EN 1041: 2008; EN 60601-1-27:2011; EN 80601-2-30: 2009; EN 60601-2-34:2001; EN 60601-2-49:2001; EN ISO 80601-2-61:2011; EN ISO 80601-2-55: 2011 COR 2012; EN ISO 90601-2-56: 2012; EN ISO 81060-1:2012 The Vista 120 monitors comply with the Medical Device Directive (MDD) 93/42/EEC. VISTA 120 MS31997 MS31998 3/5 lead ECG X X Nellcor SpO2 X X Non-invasive Blood Pressure X X Respiration X X Dual Temperature X X Built-in Recorder X X Networking X X Gas Measurement Module Compatibility X X 2 Invasive Blood Pressures X Cardiac Output X X Market availability Vista 120 monitors are available in selected markets only. For availability in your area, please contact the appropriate Dräger office from those listed below. 90 69 857 05.15 Communications & Sales Marketing YY SK Printed in the UK Chlorine-free environmentally compatible Subject to modifications 2015 CORPORATE HEADQUARTERS www.draeger.com Manufacturer: As of August 2015: Dräger Medical GmbH changes to. REGION EUROPE CENTRAL AND EUROPE NORTH Tel +49 451 882 0 Fax +49 451 882 2080 info@draeger.com REGION EUROPE SOUTH Dräger Médical S.A.S. Parc de Haute Technologie d Antony 2 25, rue Georges Besse 92182 Antony Cedex, France Tel +33 1 46 11 56 00 Fax +33 1 40 96 97 20 dlmfr-contact@draeger.com REGION MIDDLE EAST, AFRICA Branch Office P.O. Box 505108 Dubai, United Arab Emirates Tel +971 4 4294 600 Fax +971 4 4294 699 contactuae@draeger.com REGION ASIA / PACIFIC Draeger Medical South East Asia Pte Ltd. 25 International Business Park #04-27/29 German Centre Singapore 609916, Singapore Tel +65 6572 4388 Fax +65 6572 4399 asia.pacific@draeger.com REGION CENTRAL AND SOUTH AMERICA Dräger Panama Comercial S. de R.L. Complejo Business Park, V tower, 10th floor Panama City Tel +507 377 9100 Fax +507 377 9130 contactcsa@draeger.com