Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications



Similar documents
StackFlowMaster FPD580 TORBAR probe Stack gas metering systems. Measurement made easy. Introduction. For more information

Data Sheet 10/ EN. Contrac power electronics Customer specific settings

Predictive emission monitoring systems (PEMS) Deploying software-based emission monitoring systems for refining processes

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

StackFlowMaster FPD581, FPD583 and FPD585 Stack gas metering systems

Predictive emission monitoring systems The power of software analyzers

VOLU-probe/SS. Stainless Steel Pitot Airflow Traverse Probes. Accurate airflow measurement for demanding applications

Using meters to measure steam flow

EBS852 (Contrac) Power Electronic Unit

CalMaster2 Field and Software Tools. The world's first battery- and mains-powered in situ flow verification system

Air Quality Performance Test Guidelines. Kansas Department of Health and Environment Bureau of Air and Radiation

Monitoring manufacturing, production and storage environments in the pharmaceutical industry

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Monitoring the autoclaving process in the pharmaceutical industry

Advances in Thermal Dispersion Mass Flow Meter Accuracy

CO MPa (abs) 20 C

ScreenMaster RVG200 Paperless recorder FDA-approved record keeping. Measurement made easy

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Drives and motors. A guide to using variable speed drives and motors in data centres Meeting your Carbon Reduction Commitment (CRC)

Averaging Pitot Tubes; Fact and Fiction

*Note: This document is an electronic (scanned) version of the original paper report and formatting may vary somewhat from the original.

Process gas chromatographs. Measurement made easy. Compliance Solutions for 40CFR60 Subpart Ja

Stack Sampling Stack sampling or source sampling

Service catalog Flow products

Drives and motors. A guide to using variable-speed drives and motors in car parks

Measurement & Analytics. VIS Multiphase Flow Meter High gas content applications solved with no radioactive source - Measurement made easy

Flow Measurement Options for Pipeline and Open Channel Flow

THE RELEVANCE OF TWO DIFFERENT PATH LAYOUTS FOR DIAGNOSTIC PURPOSES IN ONE ULTRASONIC METER

IOMaster FPD510 Compact integral orifice flowmeter. Low-cost measurement of small flow rates

SUMMARY of PRODUCTS. HVAC Systems Control Airflow, Space, & Building Pressurization. Accurate airflow measurement for demanding applications

The World Market for Thermal Flowmeters

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES

Chapter 3.5: Fans and Blowers

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

ABB Flowmeters Review of industrial processing flowmeters

Understanding Pressure and Pressure Measurement

Certificate: / 20 August 2013

AquaMaster 3 flow measurement Saving every drop of energy and cost naturally!

Drives and motors. A guide to using variable-speed drives and motors in data centres

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

TEIP11 I/P signal converter for standard signals. Measurement made easy

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector

Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested

F ox W hi t e Paper. Reduce Energy Costs and Enhance Emissions Monitoring Systems

Improved fluidized bed drying through precise air control using Contrac electrical control actuators

Gas Custody Transfer Calibration

Fan Performance By: Mark Stevens AMCA International Deputy Executive Director Technical Affairs

Experiment 3 Pipe Friction

FLUID FLOW Introduction General Description

Thermal Dispersion Mass Flow

The Unique Accelabar Flow Meter

Exhaust Calculation Booklet

Take total control of your HVAC system with cost-saving flow and thermal energy meters

Flow distribution and turbulent heat transfer in a hexagonal rod bundle experiment

Sulfur Tail Gas Thermal Oxidizer Systems By Peter Pickard

VariTrane Duct Designer An in depth look...

4.What is the appropriate dimensionless parameter to use in comparing flow types? YOUR ANSWER: The Reynolds Number, Re.

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

INSITU TESTS! Shear Vanes! Shear Vanes! Shear Vane Test! Sensitive Soils! Insitu testing is used for two reasons:!

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

Title Protection of Environment; Revised as of July 1, 1991 CHAPTER I -- ENVIRONMENTAL PROTECTION AGENCY SUBCHAPTER C -- AIR PROGRAMS

CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG March 24, 2010

Accurate Air Flow Measurement in Electronics Cooling

Electromagnetic Flow Meter

CALDON LEFM 380Ci Gas Ultrasonic Flow Meter. Integrating experience, proven technology and innovation

PROCESS PARTICLE COUNTER (PPC) SENSOR/CONTROLLER FOR OPTIMIZING POWER RECOVERY EXPANDER AND GAS TURBINE PERFORMANCE

Understanding the Flo-Dar Flow Measuring System

Instruction Manual IM/LM_3. Field IT. Data Capture Utility. LogMaster

UPM Figure 1 MPFM installed on well pad

Fluid Flow Instrumentation

Analog signal converters CC-E I/I Current / current isolators

Phoenix Process Engineering, Inc. Project Experience Helping Clients Achieve MACT Compliance

(a) Method 1 Sample and Velocity Traverses for Stationary Sources.

HYDRAULICS. H91.8D/C - Computerized Open Surface Tilting Flow Channel - 10, 12.5, 15 and 20 m long

Boiler Control Solution. Boiler Control Solution. Instruments and Solution for Automatic Boiler Control LF3BUSS02-00EN.

1.0 What Are the Purpose and Applicability of Performance Specification 11?

Testing Protocol for Differential Pressure Measurement Devices API MPMS Chapter 22.2

ABB i-bus KNX Security Solutions with KNX Security Terminal Product Information

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

Power Generation Service Life Cycle Management for Power Plants

AOE 3104 Aircraft Performance Problem Sheet 2 (ans) Find the Pressure ratio in a constant temperature atmosphere:

Analyze IT Explorer Asset management for analyzers and analyzer systems

The Behaviour Of Vertical Jet Fires Under Sonic And Subsonic Regimes

Making Better Medical Devices with Multisensor Metrology

Energy Systems Engineering Technology

AT500 Magnetostrictive Level Transmitter. Compact magnetostrictive liquid level transmitter for direct insertion K-TEK Products

LLB-OCMFM 2000 PORTABLE DOPPLER AREA-VELOCITY FLOWMETER

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER

Brochure. ABB Customer Training Program for Motors and Generators Sharing knowledge and creating value

Temperature Transmitter TTX300

High Speed Aerodynamics Prof. K. P. Sinhamahapatra Department of Aerospace Engineering Indian Institute of Technology, Kharagpur

CHAPTER 2 BIAS DUE TO PROBE LOCATION AND STRATIFICATION

Service Power Technologies. Retrofitting A future-oriented concept in lifecycle management

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING

PLATE HEAT EXCHANGER. Installation Manual. Customer Name: Serial number: Purchase order number: Project:

Type K current limiting fuse canister Technical information and installation guide

Mass Flow Meter Product Guide

CAMS. Combustion Airflow Management System. Proven solutions for a tough industry

Transcription:

Article AT/FLOW/002 EN Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications Measurement made easy Specifying the most appropriate pitot flowmeter design in stack applications Introduction When metering stack gas flowrates within CEMS applications, a number of factors can influence a supplier or end user to consider the use of a pitot flow probe that spans only part of the stack diameter. This article discusses these factors and considers the circumstances under which partial insertion probes are an acceptable alternative plus (as importantly) where they may not be acceptable by discussing the advantages and limitations of the technique. For more information Further details of ABB Measurement Products are available for free download from www.abb.com/measurement-products or by scanning this code:

Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications How pitot tubes work A simple pitot tube is typically a probe inserted into the process pipe with the open end pointing upstream into the flow. This approach is used on aircraft for basic airspeed indication. The process stream impacts on the opening (port) in the probe and creates an elevated pressure depending, among other factors, on the velocity and density of the fluid. The static pressure is measured by a separate port that does not face directly into the flow stream. The two ports are connected to (for example) a DP transmitter that measures the difference between the two pressures; that itself is a measure of the flowrate. Faster flow One disadvantage of this 'single point' pitot is that it measures only the velocity (and therefore the flowrate) at a single point in the pipe, at whatever depth into the pipe it is inserted into. Slower flow Fig. 2: Asymmetrical flow profile (not fully-developed) However, if the flow profile is not fully-developed (for example as shown in Fig. 2), then the relationship between the velocity measured (at that point) and the average pipe velocity (that is what is required to be determined) cannot be predicted. Consequently any flow measurement produced using a single point device is poor. Fig. 1: Fully-developed flow profile If the flow velocity profile across the pipe is symmetrical and fully developed (see Fig. 1), the average flowing velocity (and hence the average flowrate) can be established from this single reading, if the probe is inserted to a predicted specific depth. However, this typically occurs only if the pitot is used following long, straight, uninterrupted lengths of pipe. To improve the measurement of flow with such a device in these cases, it is necessary to establish the actual velocity profile across the pipe by a process known as 'traversing' or 'flow profiling', where the probe is inserted at different points across the pipe and the velocity measured at each point; all while maintaining a constant flowrate. The insertion point (depth) at which the point velocity has the same value as the average velocity is established from the data and the probe can then be installed to that insertion depth. However it is important to note that this is valid only if the flow profile has the same shape and characteristics at all application flowrates. In practice this may well not be the case. 2 AT/FLOW/002 EN Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications

The optimal solution is a pitot that inherently measures the mean fluid velocity without the need for traversing. Multipoint or averaging pitot tubes (Figs 3 and 4) have multiple upstream holes at specific points across the whole pipe diameter. These holes measure the flow at several specific points across the pipe diameter and flow profile. The impact pressures across the pipe are combined within the probe to give an average upstream impact pressure. A separate hole, not facing into the flow stream, measures the static pressure alone and the difference between the impact (high) and static (low) pressures is proportional to the flow rate. This construction overcomes the main disadvantage of a simple pitot tube as it averages the velocities across the pipe diameter, instead of just taking a single point measurement. This removes the requirement to know the existing flow profile because effectively the unit continuously infers an average of the flow profile. In the ABB Torbar the static pressure sensing is at a point on the back of the sensor, removing the need to create a separate hole in the process pipe. Fig. 4: Typical ABB StackFlowMaster full-insertion averaging pitot tube Fig. 3: Averaging pitot tube Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications AT/FLOW/002 EN 3

Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications Full-insertion averaging pitot tubes Full-insertion probes span the whole diameter of the stack and incorporate strategically spaced pitot sensing holes to measure a number of point velocities simultaneously. Subsequently the flow average is taken across the complete stack diameter. The probe tip is often supported on the 'far' side of the stack, especially with large diameter stacks. The ABB approach with CEMS flow meters is to offer full-insertion probes wherever possible, primarily for performance-related reasons, whereas the drivers for the use of partial-insertion probes are economic and / or for reasons of installation constraints. The key advantage of full-insertion probes is that they provide acceptable flow measurement performance even with flow profiles that are not fully developed, which is why ABB prefer them. The limitations are: probe can be expensive in large diameter stacks probe can be difficult to install (long and heavy) a platform may need to give access to the complete stack circumference for access to the end support mounting Partial-insertion averaging pitot tubes Partial-insertion flowmeter pitot tubes are a shortened version of the full-insertion tube and typically span only approximately 1 /3 to 1 /2 of the full stack diameter. The number of velocity sensing holes is reduced to take only an average velocity across that portion of the stack diameter. Where allowed by the relevant legislative authority, partial insertion probes can be used and offer the following advantages in purchase cost: lower cost in large diameter stacks easier to install (shorter and less weight) platform can be smaller as it only needs to give access to part of the stack circumference, for access to the single mounting assembly. However, they do have a serious limitation in that they are likely to give unacceptable readings for non-fully developed flows. The question often asked as a result of this is: How do I know if a partial-insertion is suitable? 4 AT/FLOW/002 EN Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications

The answer to this depends on the flow profile at the measurement point. If the flow is fully developed and the flow profile is therefore uniform across the full diameter of the stack, then measuring across almost half of the stack diameter gives an acceptable reading that is representative of the whole stack diameter. However, if the flow profile is not uniform, then it is not acceptable to measure part of the flow profile only. The status of the flow profile depends on a number of parameters such as the pipe layout at the bottom of the stack and the straight lengths available upstream and downstream. The profiles below are clearly very different in the 2 applications shown above (Figs 5 and 6). Consequently it is only possible to establish the shape of the flow profile by carrying out a test, either in situ (for example, a 'traverse') or by computation Fig. 6: Stack with many upstream straight stack diameters available Fig. 5: Stack with few upstream straight stack diameters available Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications AT/FLOW/002 EN 5

Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications What the standards say European Community Standards are in place governing the application and practice of flow metering and gas analysis in CEMS applications within the EC. Other countries often rely on these standards for guidance. These standards are: EN 14181 Stationary source emissions. Quality assurance of automated measuring systems EN 15259 Measurement of stationary source emissions. Requirements for measurement sections and sites and for the measurement objective, plan and report EN15267 Certification of automated measuring systems. Part 1 General principles Part 2 Initial assessment of the AMS manufacturer's quality management system and post certification surveillance for the manufacturing process Part 3 Performance criteria and test procedures for automated measuring systems for monitoring emissions from stationary sources For this pre-investigation (traversing / flow profiling) to be deemed unnecessary, a number of onerous conditions must be satisfied: The meter must be located at a distance of at least 25 stack diameters downstream of any flow disturbance There must be no movable dampers in the stack run There must not be any multiple feeds into the stack Table 1: Condition-specific requirements Suitable methods for performing the pre-investigation are described in Section 8 of EN16911. Note that part of the pre-investigation process is identical to the investigation required to meet EN 15259 and could be combined with that to minimize costs. The pre-investigation establishes the flow profile that, if found to be fully developed, may permit the use of a partial insertion meter. However, it should be noted that the final decision on this rests with the competent legislative authority, not with the end user. EN16911 Stationary source emissions manual and automatic determination of velocity and volume flow rate in ducts: Part 1 Manual Reference Method Part 2 Automated Measuring Systems EN16911 is a new standard that includes advice on whether a 'probe AMS' (that is how the standard includes reference to a partial-insertion probe) can be used. Section 7 of the standard recommends that, unless the AMS can be positioned at a point where the flow profile is both fully developed and does not change over the complete operational flow range, a pre-investigation should be performed in order to characterize the flow. 6 AT/FLOW/002 EN Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications

Recommendation Under EN16911, ABB strongly recommend that, unless the specific conditions described in Table 1, page 6 are all satisfied for the application, a pre-investigation should be carried out to establish the flow profile in the stack. This profile should then be included in the submission to the competent legislative authority for their decision as to the type of instrument that can be used and its positioning on the stack. If the flow profile is sufficiently fully developed, the legislative authority may approve a partial insertion probe in the selected position. If such a pre-investigation is not carried out and the specific conditions do not exist, there is a serious risk that when the inspection of the installed equipment is performed, the installation may be deemed unacceptable. Therefore ABB will normally offer full insertion probes, unless we are advised that partial insertions are acceptable. Providing there is a statement from the customer confirming that the flow profile is fully developed over the full flow range at the proposed installation, ABB can propose a partial insertion flow probe. Flue gas measurement and control Partial-insertion pitot tubes in CEMS applications AT/FLOW/002 EN 7

Contact us ABB Limited Process Automation Salterbeck Trading Estate Workington, Cumbria CA14 5DS UK Tel: +44 (0)1946 830 611 Fax: +44 (0)1946 832 661 ABB Inc. Process Automation 125 E. County Line Road Warminster PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. Copyright 2014 ABB All rights reserved AT/FLOW/002 EN 01.2014 www.abb.com Sales Service Software