Clinica Chimica Acta 365 (2006) 50 57 Review Present and future of the autoimmunity laboratory José M. González-Buitrago a, *, Concepción González b a Servicio de Bioquímica, Hospital Universitario, Salamanca, Spain and Departamento de Bioquímica y Biología Molecular, Universidad de Salamanca, 37007 Salamanca, Spain b Servicio de Bioquímica Clínica, Hospital Universitario Virgen Macarena, Sevilla, Spain Received 30 May 2005; received in revised form 4 July 2005; accepted 5 July 2005 Available online 26 August 2005 www.elsevier.com/locate/clinchim Abstract At present, autoimmunity laboratories are very dynamic owing to the constant and increasing availability of new tests, mainly due to the detection of new autoantibodies. The main characteristic of the autoimmunity laboratory and the one that differentiates it from other laboratories that use immunoassays as basic techniques is that it determines antibodies (autoantibodies) and not antigens. For this reason, immunoassay techniques must employ antigens as reagents. Indirect immunofluorescence has and continues to be a basic technique in autoimmunity studies. However, over the last few years, a significant trend at autoimmunity laboratories has been the gradual replacement of immunofluorescence microscopy by immunoassay. Of the several different forms of immunoassay, the enzyme-linked immunosorbent assay (ELISA) format is the one most used in autoimmunity laboratories. Recombinant DNA technology has allowed the production of large quantities of antigens for autoantibody analysis. Flow cytometry for the analysis of microsphere-based immunoassays allows the simultaneous measurement of several autoantibodies. Likewise, autoantigen microarrays provide a practical means to analyse biological fluids in the search for a high number of autoantibodies. We are now at the beginning of an era of multiplexed analysis, with a high capacity of autoantibody specificities. Future trends in this field include immunoassays with greater analytical sensitivity, simultaneous multiplexed capability, the use of protein microarrays, and the use of other technologies such as microfluidics. D 2005 Elsevier B.V. All rights reserved. Keywords: Autoimmune diseases; Autoimmunity laboratory; Enzyme immunoassay; Flow cytometry; Indirect immunofluorescence; Protein microarray Contents 1. Introduction............................................................ 51 2. The role of the autoimmunity laboratory in autoimmune disease................................ 51 3. Techniques used for the study of autoantibodies......................................... 51 4. Indirect immunofluorescene................................................... 52 5. Enzyme immunoassay...................................................... 53 5.1. Solid phases........................................................ 53 5.2. Antigens.......................................................... 53 5.3. Standardization...................................................... 54 5.4. Applications........................................................ 54 Abbreviations: AMA, anti-mitochondrial antibodies; ANA, Antinuclear antibodies; ANCA, anti-neutrophil cytoplasmic antibodies; ARA, American College of Rheumatology; DNA, deoxyribonucleic acid; EIA, Enzyme immunoassay; ELISA, Enzyme-linked immunosorbent assay; IIF, Indirect immunofluorescence; MPO, myeloperoxidase; mrna, messenger RNA; SLE, Systemic Lupus Erythematosus; SS, Sjögren s syndrome; TPO, thyroid peroxidase. * Corresponding author. E-mail address: buitrago@usal.es (J.M. González-Buitrago). 0009-8981/$ - see front matter D 2005 Elsevier B.V. All rights reserved. doi:10.1016/j.cca.2005.07.023
J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 51 6. Flow cytometry.......................................................... 54 7. Microarrays............................................................ 55 8. Conclusion............................................................ 55 References............................................................... 56 1. Introduction Autoimmune diseases form a heterogeneous group of illnesses characterized by humoral or cell-mediated immune reactions against one or more of the body_s own constituents. Clinicians classify autoimmune diseases as systemic or organspecific. Although this classification is clinically useful, it does not indicate the cause or causes of the disease [1,2]. Systemic autoimmune diseases display autoantibodies directed against nuclear or cytoplasmic molecules that participate in DNA replication, DNA transcription, and the translation of messenger RNA. Organ-specific autoimmune diseases exhibit autoantibodies directed against an organ or related organs. In several systemic autoimmune diseases the presence of certain autoantibodies in serum is one of the diagnostic criteria used for their classification. Antinuclear antibodies (ANA) are included among the classification criteria for both systemic lupus erythematosus (SLE) and Sjögren s syndrome (SS). In organ-specific autoimmune diseases, the detection of autoantibodies is a fundamental datum for diagnosing them as autoimmune diseases. In the case of some autoantibodies, apart from their usefulness for diagnosis they are important for establishing a prognosis. In any case, autoantibodies should be considered to be no more than markers of disease. They are commonly found in normal individuals in the absence of any definable disease and with increasing prevalence in ageing populations. Furthermore, the presence of autoantibodies may be detected several years before the onset of disease, and in this context they are markers of future disease in presently healthy individuals [3]. Such identification might allow immunological treatment to prevent disease, and when disease cannot be prevented, life-threatening but treatable conditions could be avoided [3]. 2. The role of the autoimmunity laboratory in autoimmune diseases The clinical laboratory offers a very important tool for clinicians to diagnose and treat autoimmune diseases. Autoimmunity laboratories analyse and measure an increasing number of autoantibodies, employing a broad spectrum of techniques and methods. The main characteristic of the autoimmunity laboratory, and indeed the one that differentiates it from other laboratories that use immunoassays as basic techniques, is that it determines antibodies (autoantibodies) and not antigens. For this reason, immunoassay techniques must employ antigens as reagents. At present, autoimmunity laboratories are in a very dynamic situation owing to the constant and increasing availability of new tests, mainly due to the detection of new autoantibodies and demonstration of their clinical usefulness. Improved biochemical and molecular methods have allowed rapid dissection of the autoantigens associated with specific autoimmune diseases. Likewise, the spectacular development of serology analyses of autoantibodies over the last two decades can be attributed directly to the explosive growth of molecular biology. As has been pointed out recently, new serologic technologies and assays for autoimmune diseases are currently being developed by laboratory scientists and commercial industries at a very fast pace and may eventually lead to better reproducibility in confirming diagnosis and estimating prognosis, ultimately improving the quality of clinical care [4]. Inadequate use of laboratory tests is one of the most frequent problems in autoimmunity, leading to incorrect diagnoses, inadequate treatment and unnecessary costs. Accordingly, different committees and task forces from the scientific community have developed guidelines for the adequate use of laboratory tests for the study of autoimmune diseases. Such guidelines are the result of the analysis and evaluation by these task forces of published results. One of the most active societies is the American College of Rheumatology (ACR). An introductory document on the guidelines for immunological laboratory tests for rheumatic diseases was published in 2002 [5]. The principal objective of these guidelines is to improve patient care through rational use of laboratory tests. 3. Techniques used for the study of autoantibodies The main analytical techniques used in the autoimmunity laboratory are shown in Table 1. Over the last few years, the Table 1 Analytical techniques used in the autoimmunity laboratory Agglutination Immunoprecipitation Turbidimetry Nephelometry Double immunodiffusion Counterimmunoelectrophoresis Immunofluorescence Immunoassays ELISA IRMA Immunoblot Western blot Dot blot
52 J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 most significant trend at the autoimmunity laboratory has been the gradual replacement of microscopy (indirect immunofluorescence, IIF) and other manual methods such as double immunodiffusion and counterimmunoelectrophoresis by immunoassay, mainly enzyme immunoassay (EIA), which requires less skilled operators, is more objective, and can easily be automated. 4. Indirect immunofluorescence Indirect immunofluorescence has and continues to be a basic technique in autoimmunity studies. In the first applications of this technique, slices of mouse or rat tissue were employed. For an assay, tissues are fixed by appropriate methods to slides. Then, they are incubated with serum samples and the binding of serum autoantibodies to their corresponding antigens in the tissues is detected by using fluorescent-labelled anti-immunoglobulin antiserum. The most common fluorochromes used for IIF in the autoimmunity laboratory are fluorescein isothiocyanate, yielding green fluorescence, and rhodamine B, which produces orange fluorescence. Autoantibodies give characteristic fluorescence images called fluorescence patterns. The staining pattern provides some indication of the specificity of the antibodies in the sample. Background staining is a major drawback for the interpretation of IIF patterns. For most IIF assays, the laboratory should select a screening dilution, because undiluted serum gives a background staining due to the non-specific binding of clinically non-significant levels of circulating autoantibodies. The choice of the screening dilution depends on several factors, among them the recommendations of the reagent manufacturer, its actual performance in the laboratory, the population analysed, and the needs of clinicians. Moreover, the screening dilution performs a critical function in determining assay sensitivity and specificity. Increasing the screening dilution produces a less sensitive and more specific assay. The possibility of discriminating between several patterns and the fact that it is a good screening method are the main advantages of IIF and mean that it is still an indispensable Table 2 Fluorescence patterns and clinical associations of antinuclear antibodies (ANA) Fluorescence pattern Target Clinical association Homogeneous DNA, histones SLE, RA, JRA Speckled Coarse Fine Proteins that bind nucleic acids and ribonucleoproteins (U-RNP, Sm) Proteins that bind nucleic acids and ribonucleaproteins (SS-A, SSB) SLE, MCTD Nucleolar Nucleolar components Systemic sclerosis Centromeric Centromeric proteins Limited systemic sclerosis SS technique in the autoimmunity laboratory. Subjective interpretation, the wide variability of results, high intra-laboratory and inter-laboratory variance, semiquantitative results, and the fact that the technique cannot be completely automated are the main disadvantages. Use of HEp-2 cells instead of tissue slides for the assay of antinuclear antibodies in IIF represented a significant improvement in the study of autoantibodies [6]. HEp-2 are cells from a human larynx epithelioma cell line that have larger nuclei and nucleoli than rodent tissue cells. This facilitates their microscopic observation and provides greater sensitivity in the detection of small quantities of antibodies and a better resolution of some antigens. Moreover, cells in different phases of mitosis or the cell cycle allow the observation of antigens whose expression varies in the different stages of cell cycle. HEp-2 cells allow the recognition of more than 30 different nuclear and cytoplasmic patterns, which are given by upwards of 100 different autoantibodies [7,8]. However, the greater sensitivity of ANA assays with HEp-2 cells as compared to rodent tissues is associated with lower specificity. Thus, more patients with diseases other than systemic lupus erythematosus and healthy people show positive results. To overcome this problem a 1/ 160 titer has been recommended as the most acceptable cutoff to separate normal from abnormal sera [9]. Historically, clinicians have relied on fluorescence patterns as an aid in diagnosis. The main ANA-fluorescence patterns observed are homogeneous, speckled, nucleolar and centromeric [7]. The most frequent ANA-fluorescence patterns and clinical associations are shown in Table 2. With the advent of specific autoantibody tests, the clinical usefulness of traditional fluorescence patterns has been reduced significantly, except perhaps for anti-centromere and anti-nucleolar patterns. Fluorescence patterns are now used mainly to help guide which specific autoantibody is to be measured. IIF has also been used to measure anti-dsdna antibodies with Crithidia luciliae [10,11]. The Crithidia assay detects the binding of anti-dsdna to the kinetoplast of the organism, which contains circular dsdna unassociated with histone proteins. It can be used to detect IgG anti-dsdna, IgM antidsdna, or all isotypes of anti-dsdna. However, at present the use of Crithidia assays in the autoimmunity laboratory has been reduced and replaced by enzyme immunoassay. Antineutrophil cytoplasmic antibodies (ANCA) are autoantibodies directed against antigens found in cytoplasmic granules of neutrophils and monocytes [12]. ANCAs are closely associated with Wegener granulomatosis, microscopic polyangitis, and Churg Strauss syndrome [13]. Determination of ANCA by indirect immunofluorescence using neutrophil preparations yields three patterns: cytoplasmic (C-ANCA), perinuclear (P-ANCA) and atypical (A- ANCA) [14,15]. C-ANCA is associated with anti-proteinase 3 (PR3) antibodies, and P-ANCA with anti-myeloperoxidase (MPO) antibodies. A-ANCA is a perinuclear pattern obtained with ethanol-fixed neutrophils, but negative with formalin-fixed neutrophils. In A-ANCA, the main target
J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 53 antigens are lactoferrin, elastase and cathepsin G. The A- ANCA pattern is mainly observed in non-vasculitic diseases, such as rheumatoid arthritis, intestinal bowel diseases and infectious bacterial or viral diseases [14]. Other autoantibodies measured by IIF are anti-mitochondrial antibodies (AMA), anti-smooth muscle antibodies, anti-liver and kidney microsomal (LKM) antibodies, antiparietal cell antibodies, anti-adrenal cell antibodies and other organ-specific autoantibodies. As noted above, two of the objections to IIF are its subjectivity and the fact that it cannot be completely automated. Recently, a computer-assisted system for the classification of interphase HEp-2 cell immunofluorescence patterns for autoimmune diagnoses has been described [16]. Designed as a system for assistance, representative patterns are acquired by the operator with a digital microscope camera and transferred to a computer. By use of a software package based on image analysis, feature extraction, and machine learning algorithms, relevant characteristics describing the patterns can be detected. The system can be used to discriminate between positive and negative sera for a previous classification of the most important patterns [16]. As regards automation, in the last few years robot systems for microplate management have incorporated the capacity to prepare slides with cell or tissue preparations. These systems automatically add serum samples, incubate, wash, and add fluorescent conjugates. Thus, the slides are already prepared for microscopic observation. 5. Enzyme immunoassay Today, enzyme immunoassays offer a basic technique in clinical laboratories. Of the several different forms of enzyme immunoassay, the enzyme-linked immunosorbent assay (ELISA) format is the one most used. Two of the main components of an ELISA for autoantibodies are the solid phase and the antigen. 5.1. Solid phases Microplates, beads, and magnetic particles are the main solid phases employed in ELISA. Of these three, microplates are the most used. This solid support is very useful for developing the techniques and runs well with a number of Table 3 Sources of antigenic material Native Human tissues and blood products Animal tissues Recombinant Prokaryotic (E coli) Eukaryotic Mammalian cells Yeasts Insect cells/baculovirus samples that is not too high. Microplate handling of ELISA tests has been automated with programmed Cartesian robots. In recent years, systems that use beads or magnetic particles as the solid phase have been introduced for autoantibody measurements [17 21]. 5.2. Antigens The specificity of ELISAs for autoantibody measurements is strongly dependent on the quality of the antigens used. In ELISAs for autoantibody measurements, it is of the utmost importance that the antigen should have exactly the same sequence, conformation and post-translational modifications as the human antigen. The main sources of antigenic material are shown in Table 3. In the production of antigens, one crucial step is their purification, which is accomplished by affinity chromatography of either human or animal tissues. During the purification process, the post-translational modifications must be maintained as well as the tertiary and quaternary structures of the antigens. Likewise, conformational epitopes must be conserved. If during the purification process a contaminant antigen is present, cross-reactivity may occur during the later ELISA assay. Recombinant technology has allowed the production of large quantities of defined proteins. In the field of autoimmunity, recombinant DNA technology has been used for antigen production in bacteria and eukaryotic cell cultures [22,23]. The advantages of recombinant antigens are that the immunodominant antigens are chosen selectively; they combine optimally in the solid phase coating; they provide excellent reproducibility, with good precision and quantification, and they provide highly sensitive assays. To obtain complementary DNA (cdna) that will code for an autoantigen, several approaches have been developed. All the procedures begin with the isolation of messenger RNA (mrna) from human tissue, the target of the autoantibodies. Total mrnas are converted into cdnas by reverse transcriptase. The cdnas are then inserted into bacteriophage vectors and a cdna library is constructed. The cdnas are cultured and allowed to express the corresponding proteins. These are purified and their antigenicity is analysed using antibodies. The system most frequently used in recombinant technology is E. coli. However, for antigen production in autoantibody analysis bacteria are not very suitable. The size of the proteins expressed is very limited; there are no post-translational modifications (glycosylation, phosphorylation, acylation), which may severely disturb immunological recognition by the autoantibody. Finally, foreign proteins expressed in bacteria tend to form insoluble aggregates. Alternatives to bacteria are yeast and insect cell/baculovirus systems [22,23]. The main disadvantages of yeast are the problems encountered when extracting cellular proteins, and the additional carbohydrate attached to the secreted proteins, which reduces their immunoreactivity [23].
54 J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 During the past decade, the insect cell/baculovirus system has allowed the expression of many human autoantigens [23]. When developing an ELISA for autoantibody measurement, for the antigen that will be bound to solid phases approaches involving cell extracts, purified antigens from animal tissues, recombinant antigens, mixtures of cell extracts and purified antigens, mixtures of cell extracts and recombinant antigens, and mixtures of purified and recombinant antigens can be used. All of these approaches have been used in commercial kits for autoantibody determination. 5.3. Standardization One of the main problems with commercial EIA kits for autoantibody measurements is that of standardization. In 1982 the Arthritis Foundation (AF) in collaboration with the Centers for Disease Control (CDC) prepared a panel of five AF-CDC reference sera that includes specificities for ANA, dsdna, SS-B, RNP, and Sm antibodies [24]. The AF-CDC standards were later reevaluated to define their usefulness [25]. The AF-CDC serum panel was used to evaluate the performance of EIA kits from different manufacturers [26]. Likewise, the Association of Medical Laboratory Immunologists has developed a panel of antinuclear and anticytoplasmic antibody consensus sera that could be useful in the evaluation of newly available EIA methods for the detection of autoantibodies [27]. 5.4. Applications In the late 1980s, commercial enzyme immunoassay kits were developed for ANA screening to overcome the drawbacks of IIF [28]. These kits used HEp-2 cell extracts, attempting to mimic the IIF technique. However, they did not gain popularity because they were unable to produce results equivalent to those possible with ANA-IIF [29]. In recent years, new commercial ANA-EIA kits that use HEp-2 nuclear extracts plus purified or recombinant antigens have been developed. The objective is to replace ANA-IIF by ANA-EIA as a screening test. Comparative studies between ANA-IIF and ANA-EIA have been performed, but with discordant results [17,20,26,30 38]. In general, ANA-ELISAs show good sensitivity and a high negative predictive value, so they can be used to eliminate samples without ANA, although because of their low positive predictive value, positive samples must be analysed by IIF to confirm the presence of ANA and to determine the final result. To conclude our discussion on ANA, it should be noted that new horizons are continually being found for EIA in the autoimmunity laboratory, raising the question of whether this new era of ANA screening might mean farewell to the microscope [39]. ELISA is also used to measure anti-dsdna antibodies [26,40,41]. These assays use purified dsdna as antigen. However, during the physical process of dsdna binding to the solid phase, the double-helix structure may become distorted, exposing hidden epitopes, and modifying conformational epitopes, which could contribute to the discrepancies observed on comparing commercial ELISAs for antidsdna. In general, anti-dsdna ELISAs have very good diagnostic sensitivity, but low diagnostic specificity, because they detect high- and low-avidity antibodies; the latter considered of low clinical relevance. When using ELISA as a screening method for SLE, another test should be used to increase diagnostic specificity, such as IIF and the Farr assay, which only detect high-avidity antibodies. Nucleosomes have been considered a major autoantigen in SLE and the measurement of anti-nucleosome antibodies has been proposed as more sensitive and specific than that of antidsdna antibodies for SLE and drug-induced lupus [42,43]. Antinucleosome autoantibodies react with epitopes comprised of the native DNA structure found in chromatin, the native histone epitopes exposed in chromatin, and epitopes made up of the histone DNA complex found in chromatin. Specifically excluded are reactivities to non-histone proteins such as centromere and DNA topoisomerase I [43]. However, in our experience anti-nucleosome antibodies and antidsdna antibodies provide similar information in established SLE [44]. ELISA is also currently being used for the measurement of anticentromere antibodies using recombinant antigen [45]. Several ELISA methods have been described for the detection of anti-neutrophil cytoplasm antibodies, using purified and recombinant antigens. Attempts to obtain recombinant PR3 have not been successful for obtaining a protein reactive to human antibodies. Immobilisation of native PR3 by coating plastic plates can provide partial denaturation, with changes to conformational epitopes, which can produce a loss of reactivity with autoantibodies. This phenomenon has been observed in some patients treated with anti-pr3 antibodies and could be overcome by using a capture monoclonal antibody to immobilize PR3 protein [46]. 6. Flow cytometry Flow cytometry for the analysis of microsphere-based immunoassays has been developed for the simultaneous determination of different substances [47,48]. Among the applications of this technology is the simultaneous measurement of several autoantibodies. The system uses polystyrene microspheres labelled internally with different ratios of two different fluorochromes. Each fluorochrome can have any of 10 possible levels of fluorescence intensity, thereby creating a family of 100 spectrally addressed bead sets. The antigens are bound to the microspheres. Each of the 100 spectrally addressed bead sets can contain an antigen specific for a unique target antibody. In brief, the assay is as follows. Serum is incubated with the microspheres for around 30 min at room temperature in a filter-membrane microtitre plate. After incubation, the microspheres are washed and a solution of antihuman IgG
J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 55 conjugated with another fluorochrome is added and the microspheres are incubated again for 30 min at room temperature. The reactions are then analysed directly by the flow cytometer, which measures the fluorescence of the internal fluorochromes, allowing the classification of microspheres in an array of the 100 sets. Simultaneously, the fluorescence associated with autoantibody binding is measured. At present, several commercial kits have been introduced for the simultaneous measurement of different autoantibodies by flow cytometry. The evaluation of a kit (Biomedical Diagnostics, Marne la Vallée, France) for the simultaneous quantitative determination in the same sample of 9 antinuclear antibody specificities (dsdna, SSA, SSB, Sm, Sm/RNP, Scl70, Jo-1, ribosome and centromere B) has yielded good results; the concordances with other established methods were between 99.1% and 100.0% [49]. The system is reliable (coefficient of variation<10%) and accurate (correlation coefficient with ELISA between 0.90 and 0.97) over a large measurement range. Other kits for the simultaneous determination of antinuclear antibody specificities have also been evaluated, excellent correlation with ELISA methods being reported [50 52]. Thus, the particle-based flow cytometric assay of autoantibodies has great potential. The multiplexing ability and reasonable cost suggest that this technique will have a significant impact in the autoimmunity laboratory in coming years. 7. Microarrays Microarray technology allows the simultaneous analysis of thousands of molecular parameters. Microarrays are made using either on-chip synthesis strategies or with an arrayer based on contact-printing or ink-jet technology. Antigens (proteins, nucleic acid fragments) are deposited as microspots at defined positions on the chip. These capture molecules, termed ligands or probes, are immobilized in rows and columns on the solid support. The chips are then exposed to samples containing autoantibodies, which bind to their corresponding immobilized antigens in the microarray. The concentration of the analyte in the sample can be determined by detecting the resulting antigen antibody complex in the microspots of the chip. The recognition of autoantibodies may be achieved using fluorescence or radioactive labelling, chemiluminescence, mass spectrometry, or electrochemical methods [53]. Several reports employing protein microarray technology for the analysis of autoantibodies have been published. The first group describing the development of antigen arrays for the specific purpose of analysing autoantibodies was that of Joos et al. [53]. This group used microarrays for the simultaneous determination of 18 known autoantigens used as serologic markers in prominent autoimmune diseases such as autoimmuno thyroiditis, systemic rheumatic diseases, systemic vasculitis and antiphospholipid syndrome. The antigen corresponding to autoimmuno thyroiditis is thyroid peroxidase (TPO), and the antigen to systemic vasculitis is myeloperoxidase (MPO). Antinuclear antigens are important markers for systemic rheumatic diseases: dsdna, Sm, SmB histone, U1snRNP A y C for SLE; U1snRNP 68 kda for MCTD, SsA and SSB for Sjögren s syndrome, Scl-70 for scleroderma; CENP B for Crest syndrome, Jo1 for polymyositis, Pm-Scl 100 for an overlapping syndrome, and h2- glycoprotein I for the antiphospholipid syndrome. Microarrays contain serial dilutions of the various antigens, thereby allowing accurate determination of autoantibody titres but using minimum amounts of serum. For detection, rabbit anti-human peroxidase-conjugated antibody was used [53]. Detection was accomplished using luminol as substrate, and chemiluminescence measurements were performed. Images were acquired with a charge-coupled device (CCD) camera. The authors reported that the assay was very sensitive and highly specific, and the signal intensities observed from serial dilutions of the immobilized antigen correlated well with serial dilutions of autoimmune sera [53]. Another study used microarrays with 196 different biomolecules, representing major autoantigens targeted by autoantibodies from patients with autoimmune rheumatic diseases [54]. The authors used a robotic arrayer to print 1152-feature arrays containing 196 different biomolecules. The arrays were incubated with human sera followed by secondary anti-human antibody conjugated to the Cy-3 fluorochrome. The arrays were scanned and the images were analysed to determine the level of autoantibody binding. The authors reported that their antigen arrays were four- to eight-fold more sensitive than conventional ELISAs. They concluded that autoantigen microarrays represent a powerful tool for studying the specificity and pathogenesis of autoantibody responses, and for identifying and defining relevant autoantigens in human autoimmune diseases. More recently, another study used microarrays with 15 autoantigens for the detection of autoantibodies in autoimmune rheumatic diseases [55]. The authors indicated that the use of microarrays combined with artificial intelligence analysis might provide additional improvements in throughput, cost-effectiveness, and accuracy for the molecular diagnosis of autoimmune diseases. Several reviews and opinion analyses and theoretical developments concerning the use of microarrays for the study of autoimmune diseases have been published [56 60]. As yet, arrays for autoantibodies have not been commercialised, but in the coming future this technology should be accessible to clinical autoimmunity laboratories. 8. Conclusions Present-day clinical laboratories have at their disposal immunoassay techniques and methods that are more and more sensitive and specific for the measurement of an increasing number of different autoantibodies. To a large extent, this improvement has been due to the production of recombinant antigens. Flow cytometry for the analysis of
56 J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 microsphere-based immunoassays allows the simultaneous measurement of several autoantibodies. Likewise, autoantigen microarrays provide a practical means to analyse biological fluids in the search for autoantibodies directed against thousands of different autoantigens with a low-cost and low-volume format. We are now at the beginning of an era of multiplexed analysis, with a high capacity of autoantibody specificities. Future trends in this field include immunoassays with greater analytical sensitivity, simultaneous multiplex analysis, the use of protein microarrays and the use of other nanotechnologies such as microfluidics. References [1] Davidson A, Diamond B. Autoimmune diseases. N Engl J Med 2001;345:340 50. [2] Marrack P, Kapler J, Kotzin BL. Autoimmune diseases: why and where it occurs. Nat Med 2001;7:899 905. [3] Scofield RH. Autoantibodies as predictors of disease. Lancet 2004;363:1544 6. [4] Wiik AS, Gordon TP, Kavanaugh AF, et al. Cutting edge diagnostics in rheumatology: the role of patients, clinicians, and laboratory scientists in optimizing the use of autoimmune serology. Arthritis Care Res 2004;51:291 8. [5] American College of Rheumatology. Guidelines for immunologic laboratory testing in the rheumatic diseases: an introduction. Arthritis Care Res 2002;47:429 33. [6] Humbel RL. Detection of antinuclear antibodies by immunofluorescence. In: Van Venrooij WJ, Maini RN, editors. Manual of biological markers of disease, vol. A2. Kluwer Academic Publishers; 1993. p. 1 16. [7] Bradwell AR, Stokes RP, Johnson GD. Atlas of HEp-2 patterns. The binding site. 1995. [8] Krapf AR, von Muhlen CR, Krapf EE, Nakamura RN, Tan EM. Atlas of immunofluorescence autoantibodies. Urban and Schwarzenberg; 1996. [9] Tan EM, Feltkamp TE, Smolen JS, et al. Range of antinuclear antibodies in healthy individuals. Arthritis Rheum 1997;40:1601 11. [10] Aarden L, de Groot E, Feltkamp T. Immunology of DNA. III. Crithidia luciliae, a simple substrate for the determination of antidsdna with immunofluorescent technique. Ann N Y Acad Sci 1975;254:505 8. [11] Huber O, Greenberg ML, Huber J. Complement-fixing anti-doublestranded DNA with the Crithidia method: a better indicator of active SLE than anti-dna with the Farr method. J Lab Clin Med 1979;93:32 9. [12] van der Woude FJ, Daha MR, van Es LA. The current status of neutrophil cytoplasmic antibodies. Clin Exp Immunol 1989;78:143 8. [13] Schmitt WH, van der Woude FJ. Clinical applications of antineutrophil cytoplasmic antibody testing. Curr Opin Rheumatol 2004;16:9 17. [14] Savige JA, Paspaliaris B, Silvestrini R, et al. A review of immunofluorescence patterns associated with antineutrophil cytoplasmic antibodies (ANCA) and their differentiation from other antibodies. J Clin Pathol 1998;51:568 75. [15] Wiik A. Rational use of ANCA in the diagnosis of vasculitis. Rheumatology 2002;41:481 3. [16] Sack U, Knoechner S, Warschkan H, Pigla U, Emmrich F, Kamprad M. Computer-assisted classification of HEp-2 immunofluorescence patterns in autoimmune diagnostics. Autoimmun Rev 2003;2:298 304. [17] Bayer PM, Bauerfeind S, Bienvenu J, et al. Multicenter evaluation study of a new HEp-2 ANA screening enzyme immune assay. J Autoimmun 1999;13:89 93. [18] Bossuyt X. Evaluation of two automated enzyme immunoassays for detection of antinuclear antibodies. Clin Chem Lab Med 2000;38: 1033 7. [19] Hayashi N, Kawamoto T, Mukai M, et al. Detection of antinuclear antibodies by use of an enzyme immunoassay with nuclear HEp-2 cell extract and recombinant antigens: comparison with immunofluorescence assay in 307 patients. Clin Chem 2001;47:1649 59. [20] González C, Guevara P, Alarcón I, Hernando M, Navajo JA, González-Buitrago JM. Antinuclear antibodies (ANA) screening by enzyme immunoassay with nuclear HEp-2 cell extract and recombinant antigens: analytical and clinical evaluation. Clin Biochem 2002;35:463 9. [21] González C, García-Berrocal, Pérez M, Navajo JA, Herráez O, González-Buitrago JM. Laboratory screening of connective tissue diseases by a new automated ENA screening assay (EliA Symphony) in clinically defined patients. Clin Chim Acta 2005;359:109 14. [22] Schmitt J, Papisch W. Recombinant autoantigens. Autoimmun Rev 2002;1:79 88. [23] Schmitt J. Recombinant autoantigens for diagnosis and therapy of autoimmune diseases. Biomed Pharmacol 2003;57:261 8. [24] Tan EM, Fritzler RJ, McDougal JS, et al. Reference sera for antinuclear antibodies. I. Antibodies to native DNA, Sm, nuclear RNP, and SS-B/La. Arthritis Rheum 1982;25:1003 5. [25] Smolen JS, Butcher B, Fritzler MJ, et al. Reference sera for antinuclear antibodies. II. Further definition of antibody specificities in international antinuclear antibody reference sera by immunofluorescence and Western blotting. Arthritis Rheum 1997;40:413 8. [26] Tan EM, Smolen JS, McDougal JS, Klippel JH, Frizler MJ, Gordon TP. A critical evaluation of enzyme immunoassays for detection of antinuclear antibodies of defined specificities. Arthritis Rheum 1999;42:455 64. [27] James K, Carpenter AB, Cook L, Marchand R, Nakamura RM. Development of the antinuclear and anticytoplasmic antibody consensus panel by the Association of Medical Laboratory Immunologists. Clin Diagn Lab Immunol 2000;7:436 43. [28] Moncé NH, Boguski TR, Cappel NN. An enzyme immunoassay screening test for the detection of total antinuclear antibodies. J Clin Lab Anal 1991;5:439 42. [29] Jaskowski TD, Schroeder C, Martins TB, et al. Screening for antinuclear antibodies by enzyme immunoassay. Am J Clin Pathol 1996;105:468 73. [30] Emlen W, O Neill L. Clinical significance of antinuclear antibodies. Arthritis Rheum 1997;40:1612 8. [31] Gniewek RA, Stites DP, McHugh TM, Hilton JF, Nakagawa M. Comparison of antinuclear antibody testing methods: immunofluorescence assay versus enzyme immunoassay. Clin Diagn Lab Immunol 1997;4:185 8. [32] Gniewek RA, Sandbulte C, Fox PC. Comparison of antinuclear antibody testing methods by ROC analysis with reference to disease diagnosis. Clin Chem 1997;43:1987 9. [33] Homburger HA, Cahen YD, Griffiths J, Jacob GL. Detection of antinuclear antibodies: comparative evaluation of enzyme immunoassay and indirect immunofluorescence methods. Arch Pathol Lab Med 1998;122:993 9. [34] Reisner BS, DiBlasi J, Goel N. Comparison of an enzyme immunoassay to an indirect fluorescent immunoassay for the detection of antinuclear antibodies. Am J Clin Pathol 1999;111:503 6. [35] Olaussen O, Rekvig OP. Screening test for entinuclear antibodies (ANA): selective use of central nuclear antigens as a rational basis for screening by ELISA. J Autoimmun 1999;13:95 102. [36] Kern P, Kron M, Hiesche K. Measurement of antinuclear antibodies: assessment of different test systems. Clin Diagn Lab Immunol 2000;7:72 8. [37] Ulvestad E, Kanestrom A, Madland TM, Thomassen E, Haga HJ, Vollset SE. Evaluation of diagnostics tests for antinuclear antibodies in rheumatological practice. Scand J Immunol 2000;52: 309 15.
J.M. González-Buitrago, C. González / Clinica Chimica Acta 365 (2006) 50 57 57 [38] Fenger M, Wiik A, Hoier-Madsen M, et al. Detection of antinuclear antibodies by solid-phase immunoassays and immunofluorescence analysis. Clin Chem 2004;50:2141 7. [39] Nossent H, Rekvig OP. Antinuclear antibody screening in this new millenium: farewell to the microscope? Scand J Rheumatol 2001;30:123 6. [40] Hernando M, González C, Sánchez A. Clinical evaluation of an automated anti-dsdna fluorescent immunoassay. Clin Chem Lab Med 2002;40:1056 60. [41] González C, Guevara P, García-Berrocal B, Navajo JA, González- Buitrago JM. Clinical evaluation of Cobas Core anti-dsdna EIA Quant. J Clin Lab Anal 2004;18:200 5. [42] Burlingame RW, Cervera R. Anti-chromatin (anti-nucleosome) autoantibodies. Autoimmun Rev 2002;1:321 8. [43] Burlingame RW. Recent advances in understanding the clinical utility and underlying cause of antinucleosome (antichromatin) autoantibodies. Clin Appl Immunol Rev 2004;4:351 66. [44] González C, García-Berrocal B, Herráez O, Navajo JA, González- Buitrago JM. Anti-nucleosome, anti-chromatin, anti-dsdna and antihistone antibody reactivity in systemic lupus erythematosus. Clin Chem Lab Med 2004;42:266 72. [45] González-Buitrago JM, González C, Hernando M, et al. Antibodies to centromere antigens measured by an automated enzyme immunoassay. Clin Chim Acta 2003;328:135 8. [46] Sun J, Fass DN, Hudson JA, Vis MA, Homburger HA, Specks U. Capture-ELISA based on recombinant proteinase 3 (Pr3) is sensitive for Pr3-ANCA/Pr 3 immune complexes. J Immunol Methods 1998;221:111 23. [47] Fulton RJ, McDade RL, Smith PL, Kienker LJ, Kettman JR. Advanced multiplexed analysis with the FlowMetrix system. Clin Chem 1997;43:1749 56. [48] Vignali DAA. Multiplexed particle-based flow cytometric assays. J Immunol Methods 2000;243:243 55. [49] Rouquette AM, Desgruelles C, Laroche P. Evaluation of the new multiplexed immunoassay, FIDIS, for simultaneous quantitative determination of antinuclear antibodies and comparison with conventional methods. Am J Clin Pathol 2003;120:676 81. [50] Martins TB, Burlingame RW, von Muhlen CA, Jaskowski TD, Litwin CH, Hill HR. Evaluation of multiplexed fluorescent microsphere immunoassay for detection of autoantibodies to nuclear antigens. Clin Diagn Lab Immunol 2004;11:1054 9. [51] Gilburd B, Abu-Shakra M, Shoenfeld Y, et al. Autoantibodies profile in the sera of patients with Sjögren s syndrome: the ANA-evaluationa homogeneous, multiplexed system. Clin Dev Immunol 2004;11: 53 6. [52] Shovman O, Gilburd B, Zandman-Goddard, Yehiely A, Langevitz P, Shoenfeld Y. Multiplexed AtheNa multi-lyte immunoassay for ANA screening in autoimmune diseases. Autoimmunity 2005;38:105 9. [53] Joos T, Schrenk M, Höpfl P, et al. A microarray enzyme-linked immunosorbent assay for autoimmune diagnostics. Electrophoresis 2000;21:2641 50. [54] Robinson WH, DiGennaro C, Hueber W, et al. Autoantigen microarrays for multiplex characterization of autoantibody responses. Nat Med 2002;8:295 301. [55] Feng Y, Ke X, Ma R, Chen Y, Hu G, Liu F. Parallel detection of autoantibodies with microarrays in rheumatoid diseases. Clin Chem 2004;50:416 22. [56] Robinson WH, Steinman L, Utz PJ. Proteomics technologies for the study of autoimmune disease. Arthritis Rheum 2002;46:855 93. [57] Hueber W, Utz PJ, Steinman L, Robinson WH. Autoantibody profiling for the study and treatment of autoimmune disease. Arthritis Res 2002;4:290 5. [58] Robinson WH, Steinman L, Utz PJ. Protein and peptide array analysis of autoimmune disease. Biotechniques 2002;33:S66 9. [59] Robinson WH, Steinman L, Utz PJ. Protein arrays for autoantibody profiling and fine-specificity mapping. Proteomics 2003;3:2077 84. [60] Robinson WH, Utz PJ, Steinman L. Genomic and proteomic analysis of multiple sclerosis. Curr Opin Immunol 2003;15:660 7.