Zenit G Sight System Experience G. MOROZZI

Size: px
Start display at page:

Download "Zenit G Sight System Experience G. MOROZZI"

Transcription

1 Zenit G Sight System Experience G. MOROZZI Università degli Studi di Siena

2 Scenario Immunofluorescent antinuclear antibody test is still the "gold standard" in the diagnosis of CTD BUT: Standardization is a critical issue due to: subjective interpretation Microscope. Zenit G Sight The novel automated system for image acquisition and interpretation of cell/tissue IFF tests

3 Hardware PC LED Xeon quadcore 2W Power 4GB RAM B2A filter 4x500GB RAID 10 Ex: nm Monitor DM: 505 nm LCD Eizo 24 16/9 Color Camera Screen Res. : 1920x1200 QiCam QImaging 12bit Objective Lens 4x 40x Motorized Revolver

4 Main Advantages No Fluorescence Decay Objective evaluation Analysis and pattern recognition for ANA HEp-2 assays Digitization for IIF substrates wells Navigable wells as a virtual microscope Workflow management File storage and easy retrieval of scanned wells Data and Result exchange with other LABs

5 Block diagram of a scanning process Scanning Process The scanner transforms each well of a glass slide in a digital file Digital Slide Viewing The digital wells are viewed with the use of a virtual microscope

6 Well Image Generation Each tile of the area is acquired by the camera to create a unique well image

7 Navigate the wells of the slides as a Virtual Microscope with the PC monitor

8 Virtual Well Viewing 10x 20x 40x Wells can be navigated Resolution of well viewing can be changed

9 Scanning Advantages (1/2) After that a well is digitalized it is always mantained without fluorescence decay effects The specific fluorescence level of a well should be observed also after a long time LED illumination does not photobleach signal as fast as mercury lamps do

10 Scanning Advantages (2/2) Positive and negative tiles are scanned without out-of-focus No additional filters (DAPI) are required to search focus on negatives Evans Blue doesn t effect the discrimation capabilities of software Scan well area can be reduced in order to save time during acquisition (5-1 ) Positive/Negative - Patterns Discrimination are performed while the system is scanning

11 Workflow Management Woklist loading (automatic or manual) The worklist contains the number of slides to process, the test to be carried out for each slide and sample data for each well Scanning and automatic analysis of samples The system automatically acquires images and analyze them Validation The operator examine every digital image and navigate the virtual slides. The system diagnosis can be conformed or modified Refertation (to print or export file) The system allows report printing and creation of an output worklist file

12 AVAILABLE IN JUNE

13 IN PROGRESS

14 IN PROGRESS

15 Our experience..

16 593 Sera examined in collaboration with UDINE lab. 518 sera for ANA test 75 selected for particular IIF patterns or specificities

17 Samples selected SSA SSB DNA SP100 Cenp-B AMA-M2 Gp210 RNP Scl70 Jo1 PCNA PMscl Sm P-ribosomiale

18 ANALYSIS TRADITIONAL MICROSCOPY: G-SIGHT Hep-2000 Analysis by 2 expert operators Pattern and titre Hep-2000 automated system for image acquisition and Interpretation of pos/neg and patterns

19 Pattern Recognition System has be trained to detect and recognize five patterns: Nucleolar Speckled Homogenuos Centromere Mitocondrial..this topic is in progress

20 TITRE vs SCORE SCORE: a measure of probability is given as value of confidence that analyzed well is positive AT THE MOMENT NO CORRELATION WITH IIF TITRE!! (in progress) SCORE : Algorithm response is: negative-certain (N); negative-uncertain (N?); Positive (P) System has been trained in order to avoid that positive samples are classified as negative (HIGHT SENSITIVITY)

21 Hep-2000 G-SIGHT(1) SCORE (1) G-SIGHT (2) SCORE (2) ALTRO pattern titolo OMOG 5120 P 100 P 99 NEG N N? 10 NEG N N 1 NUCLEOL 160 N 1.48 N? 20 NEG N N? 2,5 FIN PUNT + NUCLEOL 320 P 82 P 84 FIN PUNT E NUCLEOL 80 N 4.25 N? 29 FIN PUNT 160 N? 37 P 58 FIN PUNT 80 N 6.83 N? 33 NUCLEOL 160 N? N? 44 FIN PUNT E OMOG 640 P 99 P 98 PCNA 160 N? 46.7 P 64 MATR NUCL 320 P 98 P 95 GRGR 160 N? 15.2 N? 43 SP100 CENTR GRGR AMA 1280 P 99 P 97 SP100 M2 MEMBRANA NUCL 640 P P 77 anti-gp210 PCNA 640 N? P 59 PCNA OMOG E CITOPL 2560 P P 98 P-RIBOSOMI FUSO MITOTICO 320 N 0.16 N? 8

22 RESULTS N = 593 Neg POS N? G-SIGHT* 86 (14.5%) 2 ev 161 (27%) 396 (67%) 1 ev 130 (22%) 345 (58%) 65 (11%) Conv. microscopy 397 (67%) 196 (33%) * 1 well UD

23 Pos P+N? Pos P Neg N False pos N? (pos) pos ** 86* 311* : 291 (N?) 20 (as P) 54 ** 0 (neg) *IIF neg **IIF pos

24 LOW TITRE Conv. 1:80 1:160 mycroscopy (58%) No positive samples have been classified as negative Negative score: 0 2

25 Results in term of positivity ( P) hom speck nucl centr Nucl. memb 52/77 78% 94/125 75% 21/37 57% 9/10 90% 3/5 60% Nucl matrix 3 /4 75% PCNA cit mit 2 /2 100% 6/ 6 100% 2/2 100% SSA/ SSB 24/26 92% Scl70 CENP Sm RNP 7/7 100% 4/4 100% PCNA/ simil PM/ Scl GP210 Sp100 GrGr M2 Jo1 5/5 2/3 1/1 4/6 1/10 4/4 0/1* * IIF neg

26 Accuratezza pattern singoli e misti (media 96%) 0 Sconosciuto = 87.5% 1 Centromerico = 100.0% 2 Mitocondriale = 100.0% 3 Nucleolare = 100.0% 4 Omogeneo = 100.0% 5 Granulare = 100.0% Omogeneo-Granulare = 97.7% Nucleolare-Citoplasm. Omogeneo-Citoplasm. = 100.0% Granulare-GRGR = 100.0% Granulare-Nucleolare = 94.1% = 0.0% (1 campione analizzato)

27 In conclusion, my questions: By automatic systems. Is it useful to define ANA pattern?... (in my opinion ) Is it useful to define the fluorescence titre?... (in my opinion ) Is it useful to to avoid that positive samples are classified as negative? (in my opinion. On the other hands.higher sensitivity or specificity?

28

Quality in noninfectious

Quality in noninfectious Quality in noninfectious serology Willy Vandersteen o Pre-analysis o Method o Pipetting and incubation station o Reagents o Microscopy/Signal detection o People o Informatics o Post-analysis o Murphy?

More information

Testing for RA. The Ideal Lab Test. William M. Wason, MD, PhD 9/24/2010. Confusion Abounds

Testing for RA. The Ideal Lab Test. William M. Wason, MD, PhD 9/24/2010. Confusion Abounds Confusion Abounds Rheumatoid arthritis: ulnar deviation and muscle artrophy, hands Poor sensitivity and specificity Hepatitis C causes lots of false + tests Changing technology in how tests are done Historic

More information

The ANA Test: All You Need to Know Department of Family and Community Medicine Family Medicine Update April 25, 2014

The ANA Test: All You Need to Know Department of Family and Community Medicine Family Medicine Update April 25, 2014 The ANA Test: All You Need to Know Department of Family and Community Medicine Family Medicine Update April 25, 2014 Celso R. Velázquez MD Division of Rheumatology University of Missouri velazquezc@health.missouri.edu

More information

Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides

Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides The Integrated Lab Connected Welcome to the fully integrated autoimmune laboratory. Saves time by reducing hands-on functions NOVA Lite Barcoded IFA Slides NOVA View Reduces costs with automation efficiencies

More information

Fluorescent Array Imaging Reader

Fluorescent Array Imaging Reader Fluorescent Array Imaging Reader Multiplex Enabled For 2-Color Fluorescent Microarrays Extended Dynamic Range Automated Spot Analysis Compact and Affordable Microarray Analysis With FLAIR FLAIR Sensovation

More information

MIRAX SCAN The new way of looking at pathology

MIRAX SCAN The new way of looking at pathology Microscopy from Carl Zeiss MIRAX SCAN The new way of looking at pathology Greater reliability. Greater efficiency. Virtual microscopy for research & analysis. Better. More efficient. Quality as the basis

More information

Servicio de Salud de Castilla-La Mancha

Servicio de Salud de Castilla-La Mancha Servicio de Salud de Castilla-La Mancha Telepathology network of Castilla-La Mancha MARCIAL GARCÍA ROJO Chief. Pathology Department Hospital General de Ciudad Real marcial@cim.es Servicio de Anatomía Patológica

More information

Contents. Company overview 2. Ophthalmology aladdin 6 Ca-800 8 dv-3 10 CC-100XP 12 CC-100 13 CC-90SW 14 CV-tablet 15

Contents. Company overview 2. Ophthalmology aladdin 6 Ca-800 8 dv-3 10 CC-100XP 12 CC-100 13 CC-90SW 14 CV-tablet 15 your digital eye Contents Company overview 2 Ophthalmology aladdin 6 Ca-800 8 dv-3 10 CC-100XP 12 CC-100 13 CC-90SW 14 CV-tablet 15 Autoimmunity Zenit g Sight for Zenit amidot 16 Zenit g Sight 18 Cell

More information

Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available!

Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available! Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available! Arrayit SpotLight Two-Color Microarray Scanners provide the market s most affordable

More information

VENTANA Digital Pathology. Reliable. Efficient. Comprehensive.

VENTANA Digital Pathology. Reliable. Efficient. Comprehensive. VENTANA Digital Pathology Reliable. Efficient. Comprehensive. In the quest to improve the lives of all patients afflicted with cancer, every second counts. Fast, informed clinical decisions are essential

More information

Thermo Scientific ArrayScan XTI High Content Analysis Reader. revolutionizing cell biology with the power of high content

Thermo Scientific ArrayScan XTI High Content Analysis Reader. revolutionizing cell biology with the power of high content Thermo Scientific ArrayScan XTI High Content Analysis Reader revolutionizing cell biology with the power of high content learn more about your cells using high content technology Thermo Scientific High

More information

Hardware Requirements

Hardware Requirements Hardware Requirements This guide details the critical points of installing the IntraVet software from specific technology requirements to explicit hardware recommendations. communication organization practice

More information

Computer Vision for Quality Control in Latin American Food Industry, A Case Study

Computer Vision for Quality Control in Latin American Food Industry, A Case Study Computer Vision for Quality Control in Latin American Food Industry, A Case Study J.M. Aguilera A1, A. Cipriano A1, M. Eraña A2, I. Lillo A1, D. Mery A1, and A. Soto A1 e-mail: [jmaguile,aciprian,dmery,asoto,]@ing.puc.cl

More information

Chapter 12 Filters for FISH Imaging

Chapter 12 Filters for FISH Imaging Chapter 12 Filters for FISH Imaging Dan Osborn The application of in situ hybridization (ISH) has advanced from short lived, non-specific isotopic methods, to very specific, long lived, multiple color

More information

NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12

NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12 Envisions confirmed Content NyONE - Cell imaging in a bird s eye view 4 From cells to numbers 6 NyONE...resolution matters! 8 Features & benefits 10 Fluorescence excitation channels 12 Technical specifications

More information

Imaging and Bioinformatics Software

Imaging and Bioinformatics Software Imaging and Bioinformatics Software Software Overview 242 Gel Analysis Software 243 Ordering Information 246 Software Overview Software Overview See Also Imaging systems: pages 232 237. Bio-Plex Manager

More information

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie marc.lavoie@cavehill.uwi.edu MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment

More information

Introduction to Pattern Recognition

Introduction to Pattern Recognition Introduction to Pattern Recognition Selim Aksoy Department of Computer Engineering Bilkent University saksoy@cs.bilkent.edu.tr CS 551, Spring 2009 CS 551, Spring 2009 c 2009, Selim Aksoy (Bilkent University)

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

Bode Collection Point Electronic DNA Sample Information Program Technical Specifications

Bode Collection Point Electronic DNA Sample Information Program Technical Specifications Bode Collection Point Electronic DNA Sample Information Program Technical Specifications Product Definition Bode Collection Point software is used to collect and report personal information and other meta

More information

Network Guidelines and Hardware Requirements

Network Guidelines and Hardware Requirements Network Guidelines and Hardware Requirements To optimize performance in the IntraVet program we suggest that your network and computers follow the recommendations listed below. Doing so will help ensure

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

a measurable difference

a measurable difference Thermo Scientific Cell Analysis Software and Informatics Products a measurable difference for everyone Thermo Scientific HCS Studio Cell Analysis Software Thermo Scientific Store Image and Database Management

More information

How To Use Trackeye

How To Use Trackeye Product information Image Systems AB Main office: Ågatan 40, SE-582 22 Linköping Phone +46 13 200 100, fax +46 13 200 150 info@imagesystems.se, Introduction TrackEye is the world leading system for motion

More information

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation

More information

Upgrade or Not? Agenda

Upgrade or Not? Agenda Upgrade or Not? An open discussion on the merits of upgrading to GP 2013 R2 or GP 2015 Agenda Key Features & Benefits of GP 2013 R2 Key Features & Benefits of GP 2015 Deployment Options System Requirements

More information

ANA testing can now be ordered in several ways, depending on the clinical circumstances:

ANA testing can now be ordered in several ways, depending on the clinical circumstances: LAB TEST CONNECT Multiplex ANA Screen Dr. Joseph Schappert, M.D.,Medical Director Chief Medical Offi cer ANA has been the primary screening test for connective tissue diseases (CTD s) for many years. While

More information

Capture. Structure. Access. Any EHR.

Capture. Structure. Access. Any EHR. ImagIN Health Solutions Unstructured Intelligence Available at the Point of Care. Is your practice drowning in paperwork? ImagIN Health Solutions is a set of software and services designed to help you

More information

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution Product Information Version 1.0 ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution ZEISS Axiocam 506 color Sensor Model Sensor Pixel

More information

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images InterView FUSION InterView FUSION is a new visualization and evaluation software developed by Mediso built on state of

More information

HISTO SPOT SSO System

HISTO SPOT SSO System HISTO SPOT SSO System for HLA Typing HISTO SPOT Kits MR.SPOT Processor HISTO MATCH Software Complete system certified for IvD use HISTO SPOT Features The HISTO SPOT SSO assay in combination with the MR.SPOT

More information

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both

More information

How To Use Bodescan For 3D Imaging Of The Human Body

How To Use Bodescan For 3D Imaging Of The Human Body «Bodescan» THE ULTIMATE 3D SCANNER FOR DATA ACQUISITION OF HUMAN BODY A COMPREHENSIVE SYSTEM INCLUDING HARDWARE AND SOFTWARE TOOLS Page 2 of 9 Bodescan HUMAN BODY3D SCANNING HARDWARE AND SOFTWARE INCLUDED

More information

EHR Implementation: What you need to know to have a successful project: Part 2. Bruce Kleaveland President Kleaveland Consulting, Inc.

EHR Implementation: What you need to know to have a successful project: Part 2. Bruce Kleaveland President Kleaveland Consulting, Inc. EHR Implementation: What you need to know to have a successful project: Part 2. Bruce Kleaveland President Kleaveland Consulting, Inc. Learning Objectives: Recognize key EHR hardware & network components

More information

RiMONITOR. Monitoring Software. for RIEGL VZ-Line Laser Scanners. Ri Software. visit our website www.riegl.com. Preliminary Data Sheet

RiMONITOR. Monitoring Software. for RIEGL VZ-Line Laser Scanners. Ri Software. visit our website www.riegl.com. Preliminary Data Sheet Monitoring Software RiMONITOR for RIEGL VZ-Line Laser Scanners for stand-alone monitoring applications by autonomous operation of all RIEGL VZ-Line Laser Scanners adaptable configuration of data acquisition

More information

NextStar with Ribbon Scanning The Film Conversion Approach of the 21 st Century

NextStar with Ribbon Scanning The Film Conversion Approach of the 21 st Century NextStar with Ribbon Scanning The Film Conversion Approach of the 21 st Century By Kurt Breish CTO, nextscan Inc. Abstract An innovative film conversion processing approach, called Ribbon Scanning, is

More information

FDA Regulation of Whole Slide Imaging (WSI) Devices: Current Thoughts

FDA Regulation of Whole Slide Imaging (WSI) Devices: Current Thoughts FDA Regulation of Whole Slide Imaging (WSI) Devices: Current Thoughts Clinical Laboratory Improvement Advisory Committee Meeting Centers for Disease Control and Prevention February 15, 2012 Tremel A. Faison,

More information

Using Web Services for scanning on your network (Windows Vista SP2 or greater, Windows 7 and Windows 8)

Using Web Services for scanning on your network (Windows Vista SP2 or greater, Windows 7 and Windows 8) Using Web Services for scanning on your network (Windows Vista SP2 or greater, Windows 7 and Windows 8) The Web Services protocol allows Windows Vista (SP2 or greater), Windows 7 and Windows 8 users to

More information

Microsoft Dynamics CRM 2011 Guide to features and requirements

Microsoft Dynamics CRM 2011 Guide to features and requirements Guide to features and requirements New or existing Dynamics CRM Users, here s what you need to know about CRM 2011! This guide explains what new features are available and what hardware and software requirements

More information

Digital Imaging and Communications in Medicine (DICOM) Supplement 145: Whole Slide Microscopic Image IOD and SOP Classes

Digital Imaging and Communications in Medicine (DICOM) Supplement 145: Whole Slide Microscopic Image IOD and SOP Classes 5 Digital Imaging and Communications in Medicine (DICOM) Supplement 145: Whole Slide Microscopic Image IOD and SOP Classes 10 15 20 Prepared by: DICOM Standards Committee, Working Group 26, Pathology 1300

More information

Prof. Dr. M. H. Assal

Prof. Dr. M. H. Assal Prof. Dr. M. H. Assal AS 26/10/2014 Computer hardware is the collection of physical elements that comprise a computer system Computer hardware can be classified as following: o Input devices o Output devices

More information

9.7" Microscope Tablet Camera (REALPAD 5.5) Introduction. Feature

9.7 Microscope Tablet Camera (REALPAD 5.5) Introduction. Feature 9.7" Microscope Tablet Camera (REALPAD 5.5) Introduction Brand new high performance and highly cost-effective, super reliable 9.7inches smart digital microscope tablet camera! Equipped with 5MP color COMS

More information

IMAGING SOFTWARE. Image-Pro Insight Image Analysis Made Easy. Capture, Process, Measure, and Share

IMAGING SOFTWARE. Image-Pro Insight Image Analysis Made Easy. Capture, Process, Measure, and Share IMAGING SOFTWARE Image-Pro Insight Image Analysis Made Easy Capture, Process, Measure, and Share Image-Pro Insight Image Analysis Made Easy Capture, Process, Measure, and Share Image-Pro Insight, the latest

More information

AxioCam HR The Camera that Challenges your Microscope

AxioCam HR The Camera that Challenges your Microscope Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,

More information

DICOM in Pathology. 2005 DICOM International Conference Budapest, Hungary

DICOM in Pathology. 2005 DICOM International Conference Budapest, Hungary DICOM in Pathology 2005 DICOM International Conference Budapest, Hungary Bruce Beckwith, MD Department of Pathology Beth Israel Deaconess Medical Center Harvard Medical School Boston, Massachusetts, USA

More information

How To Use An Edge 3.1 Scientific Cmmos Camera

How To Use An Edge 3.1 Scientific Cmmos Camera edge 3.1 scientific CMOS camera high resolution 2048 x 1536 pixel low noise 1.1 electrons global shutter USB 3.0 small form factor high dynamic range 27 000:1 high speed 50 fps high quantum efficiency

More information

The How Do Guide to AxioVision. Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc.

The How Do Guide to AxioVision. Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc. The How Do Guide to AxioVision Author: Brian Svedberg Image Analysis Specialist Carl Zeiss MicroImaging, Inc. How do I grab a single image? How do I grab a single image? Click on the Camera icon to switch

More information

Overview Image Acquisition of Microscopic Slides via Web Camera

Overview Image Acquisition of Microscopic Slides via Web Camera MAX-PLANCK-INSTITUT FÜR MARINE MIKROBIOLOGIE ABTEILUNG MOLEKULARE ÖKOLOGIE Overview Image Acquisition of Microscopic Slides via Web Camera Andreas Ellrott and Michael Zeder Max Planck Institute for Marine

More information

Installation Guide ZEN 2012 SP2

Installation Guide ZEN 2012 SP2 Installation Guide ZEN 2012 SP2 Carl Zeiss Microscopy GmbH Carl-Zeiss-Promenade 10 07745 Jena, Germany microscopy@zeiss.com www.zeiss.com/microscopy Carl Zeiss Microscopy GmbH Königsallee 9-21 37081 Göttingen

More information

Maschinelles Lernen mit MATLAB

Maschinelles Lernen mit MATLAB Maschinelles Lernen mit MATLAB Jérémy Huard Applikationsingenieur The MathWorks GmbH 2015 The MathWorks, Inc. 1 Machine Learning is Everywhere Image Recognition Speech Recognition Stock Prediction Medical

More information

Image Analysis Using the Aperio ScanScope

Image Analysis Using the Aperio ScanScope Image Analysis Using the Aperio ScanScope Allen H. Olson, PhD Algorithm Development Engineer Aperio Technologies INTRODUCTION Why should I choose the Aperio ScanScope over competing systems for image analysis?

More information

How do I Check if My Computer is Compatible with Windows 7

How do I Check if My Computer is Compatible with Windows 7 How do I Check if My Computer is Compatible with Windows 7 Enterprise Computing & Service Management 1 Follow this link to download the Windows 7 Upgrade Advisor http://www.microsoft.com/windows/windows-7/get/upgrade-advisor.aspx

More information

Microsoft Office 2010 system requirements

Microsoft Office 2010 system requirements Microsoft Office 2010 system requirements When upgrading from Microsoft Office 2007 to the 2010 release you should not need to upgrade your hardware, although you might need to upgrade to a supported operating

More information

LOCAL SURFACE PATCH BASED TIME ATTENDANCE SYSTEM USING FACE. indhubatchvsa@gmail.com

LOCAL SURFACE PATCH BASED TIME ATTENDANCE SYSTEM USING FACE. indhubatchvsa@gmail.com LOCAL SURFACE PATCH BASED TIME ATTENDANCE SYSTEM USING FACE 1 S.Manikandan, 2 S.Abirami, 2 R.Indumathi, 2 R.Nandhini, 2 T.Nanthini 1 Assistant Professor, VSA group of institution, Salem. 2 BE(ECE), VSA

More information

1 PERSONAL COMPUTERS

1 PERSONAL COMPUTERS PERSONAL COMPUTERS 1 2 Personal computer a desktop computer a laptop a tablet PC or a handheld PC Software applications for personal computers include word processing spreadsheets databases web browsers

More information

Managing Relion IEDs Protection and control IED manager PCM600

Managing Relion IEDs Protection and control IED manager PCM600 Managing Relion IEDs Protection and control IED manager PCM600 Efficient management of Relion protection and control IEDs The protection and control IED manager PCM600 tool provides versatile functionalities

More information

STI Hardware Specifications for PCs

STI Hardware Specifications for PCs Local School Fileserver with STIOffice & STIClassroom WIN 1GHz or Higher 512MB 1GB Free Space Windows 2000/2003 Server, Novell Netware v4.11 or greater. Local School Fileserver with STIOffice, STIClassroom

More information

KS732 Family. Rev. 1.01 KS732. 1/3 CCD Color USB Camera

KS732 Family. Rev. 1.01 KS732. 1/3 CCD Color USB Camera KS732 Family 1/3 CCD Color USB Camera The KS732 is a product line of 1/3" compact cameras which is available as an USB version. It is designed to be used in a wide variety of professional industrial, security

More information

HARDWARE AND SOFTWARE REQUIREMENTS

HARDWARE AND SOFTWARE REQUIREMENTS HARDWARE AND SOFTWARE REQUIREMENTS WINDOWS SPECIFICATIONS File Server Pentium 4 64 - Bit Dual Core, 3.2 GHZ or better (dual/quad Xeon processors recommended) 4GB of RAM or better Two 72GB or better fast

More information

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope. MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize

More information

icr3600 icr3600m VertX idr Film Digitizers Clarity PACS Innovative Medical Imaging

icr3600 icr3600m VertX idr Film Digitizers Clarity PACS Innovative Medical Imaging icr3600 icr3600m VertX idr Film Digitizers Clarity PACS Innovative Medical Imaging Twenty Years of Medical Imaging Innovation The Future of Digital Radiography is at icrco. Founded in 1990 by Stephen Neushul,

More information

SECTION 25 15 16 INTEGRATED AUTOMATION SOFTWARE FOR CONTROL OF MONITORING NETWORKS

SECTION 25 15 16 INTEGRATED AUTOMATION SOFTWARE FOR CONTROL OF MONITORING NETWORKS SECTION 25 15 16 INTEGRATED AUTOMATION SOFTWARE FOR CONTROL OF MONITORING NETWORKS Specifier: The Specifier/Design Professional is responsible for the accuracy of all project specifications, including

More information

s y s t e m r e q u i r e m e n t s

s y s t e m r e q u i r e m e n t s s y s t e m r e q u i r e m e n t s System Requirements describe minimum and recommended standards for using Dentrix G4. Exceeding the minimum standards The may result in better system performance. Note:

More information

How to recover a failed Storage Spaces

How to recover a failed Storage Spaces www.storage-spaces-recovery.com How to recover a failed Storage Spaces ReclaiMe Storage Spaces Recovery User Manual 2013 www.storage-spaces-recovery.com Contents Overview... 4 Storage Spaces concepts and

More information

INSTALLATION MINIMUM REQUIREMENTS. Visit us on the Web www.docstar.com

INSTALLATION MINIMUM REQUIREMENTS. Visit us on the Web www.docstar.com INSTALLATION MINIMUM REQUIREMENTS Visit us on the Web www.docstar.com Clients This section details minimum requirements for client workstations that use. Workstation (Scan) Windows 7 (32/64 bit); Windows

More information

QUARTZ FA-LIMS. Laboratory Information Management System for Failure Analysis THE ONLY LIMS SYSTEM DESIGNED SPECIFICALLY FOR FAILURE ANALYSIS

QUARTZ FA-LIMS. Laboratory Information Management System for Failure Analysis THE ONLY LIMS SYSTEM DESIGNED SPECIFICALLY FOR FAILURE ANALYSIS QUARTZ FA-LIMS Laboratory Information Management System for Failure Analysis THE ONLY LIMS SYSTEM DESIGNED SPECIFICALLY FOR FAILURE ANALYSIS FROM JOB REQUEST TO FINAL REPORT, THE COMPLETE, CUSTOMIZED ANALYTICAL

More information

A System for Capturing High Resolution Images

A System for Capturing High Resolution Images A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: pitas@zeus.csd.auth.gr

More information

DBSWIN the digital image centre for the surgery

DBSWIN the digital image centre for the surgery DBSWIN the digital image centre for the surgery Compressed air Suction Imaging Dental care Hygiene Digital X-ray diagnostics with VistaScan and VistaRay The powerful DBSWIN imaging software works with

More information

Komet 7. Acquisition and Analysis Software for the Comet Assay - Research and GLP

Komet 7. Acquisition and Analysis Software for the Comet Assay - Research and GLP Komet 7 Acquisition and Analysis Software for the Comet Assay - Research and GLP Komet 7 Now with Zyla 5.5 USB 3.0 support. Score comets faster than ever, less scanning more scoring. Komet software allows

More information

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen

More information

Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University

Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University Chapter 4 Microscopy, Staining, and Classification 2012 Pearson Education Inc. Lecture prepared by Mindy Miller-Kittrell North Carolina State University Microscopy and Staining 2012 Pearson Education Inc.

More information

Medisoft Clinical v17 System Requirements

Medisoft Clinical v17 System Requirements Medisoft Clinical v17 System Requirements Table of Contents Glossary... 3 Thin Client Workstation... 3 Application Server... 3 Thin Client Server... 3 Standard Workstation... 3 Application Folder... 3

More information

LR120 LoadRunner 12.0 Essentials

LR120 LoadRunner 12.0 Essentials LR120 LoadRunner 12.0 Essentials Overview This five-day course introduces students to HP LoadRunner 12.0, including the usage of Virtual User Generator (VuGen), Controller and Analysis tools. This course

More information

Alert Technology Group 46 Empey St, Brantford ON Office 519.750.1516 Toll Free 1.888.750.1516 www.alerttechnologygroup.ca

Alert Technology Group 46 Empey St, Brantford ON Office 519.750.1516 Toll Free 1.888.750.1516 www.alerttechnologygroup.ca Alert Technology Group 46 Empey St, Brantford ON Office 519.750.1516 Toll Free 1.888.750.1516 www.alerttechnologygroup.ca Alert Technology Group is the leader in providing technologically advanced security

More information

Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model

Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model A p p l i c a t i o n N o t e Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model Brad Larson and Peter Banks, BioTek Instruments, Inc., Winooski, VT Grant Cameron, TAP Biosystems

More information

Thermo Scientific CellInsight CX5 HCS Platform. discover more uncover the possibilities

Thermo Scientific CellInsight CX5 HCS Platform. discover more uncover the possibilities Thermo Scientific CellInsight CX5 HCS Platform discover more uncover the possibilities high content screening Driving your Productivity with Innovated Technology What is High Content Screening (HCS)? High

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

1. Open the battery compartment as shown in the image.

1. Open the battery compartment as shown in the image. This Quick User Guide helps you get started with the IRIScan Book 3 scanner. This scanner is supplied with the software applications Readiris Pro 12, IRIScan Direct and IRISCompressor. Corresponding Quick

More information

Toolbox 4.3. System Requirements

Toolbox 4.3. System Requirements Toolbox 4.3 February 2015 Contents Introduction... 2 Requirements for Toolbox 4.3... 3 Toolbox Applications... 3 Installing on Multiple Computers... 3 Concurrent Loading, Importing, Processing... 4 Client...

More information

Speed Performance Improvement of Vehicle Blob Tracking System

Speed Performance Improvement of Vehicle Blob Tracking System Speed Performance Improvement of Vehicle Blob Tracking System Sung Chun Lee and Ram Nevatia University of Southern California, Los Angeles, CA 90089, USA sungchun@usc.edu, nevatia@usc.edu Abstract. A speed

More information

inform Advanced Image Analysis Software For Accurately Quantifying Biomarkers in Tissue Sections

inform Advanced Image Analysis Software For Accurately Quantifying Biomarkers in Tissue Sections P R O D U C T N O T E inform Advanced Image Analysis Key Features Quantitative analysis of biomarker epression in tissue sections and TMAs Separation of weakly epressing and overlapping markers Cellular

More information

Each ievo reader needs its own network cable, as each reader has its own IP address. Ensure this has been taken into consideration

Each ievo reader needs its own network cable, as each reader has its own IP address. Ensure this has been taken into consideration WELCOME ievo would like to thank you for purchasing our product. ievo advise the use of a. Ensure you have calculated the correct power supply amp rating if you are connecting the ievo board and other

More information

Displays ready for Signage

Displays ready for Signage Displays ready for Signage Display Forum 2007 November 20th, 2007 Silvio Cerato List of Content Introduction The Digital Signage Market Technology Aspects Optimization Hardware & Solution 1 History of

More information

// SUPER technology for precise web viewing: SUPER HANDYScan 4000. Be inspired. Move forward.

// SUPER technology for precise web viewing: SUPER HANDYScan 4000. Be inspired. Move forward. // SUPER technology for precise web viewing: SUPER HANDYScan 4000. Be inspired. Move forward. // QUALITY: RELIABLY PROTECTED. PRODUCTIVITY: SUSTAINABLY INCREASED. Web viewing that sets standards: BST SUPER

More information

Realization of a UV fisheye hyperspectral camera

Realization of a UV fisheye hyperspectral camera Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical

More information

Multi-Konzeptionelle Verwendung von Low-Cost Hardware in der Lehre Dr.-Ing. Daniel Weida, MathWorks

Multi-Konzeptionelle Verwendung von Low-Cost Hardware in der Lehre Dr.-Ing. Daniel Weida, MathWorks Multi-Konzeptionelle Verwendung von Low-Cost Hardware in der Lehre Dr.-Ing. Daniel Weida, MathWorks 2015 The MathWorks, Inc. 1 Agenda Project-Based Learning in Engineering Education Multipurpose Hardware

More information

Whitepaper Document Solutions

Whitepaper Document Solutions Whitepaper Document Solutions ScannerVision 3 Contents Contents... 2 Introduction... 3 ScannerVision introduction... 4 Concept... 4 Components... 4 Deploying ScannerVision... 5 Supported Operating Systems...

More information

Outline. 1. Experiment. 2. Sample analysis and storage. 3. Image analysis and presenting data. 4. Probemaker

Outline. 1. Experiment. 2. Sample analysis and storage. 3. Image analysis and presenting data. 4. Probemaker Tips and tricks Note: this is just an informative document with general recommendations. Please contact support@olink.com should you have any queries. Document last reviewed 2011-11-17 Outline 1. Experiment

More information

HC-6950. Emission Protected Security Workstation

HC-6950. Emission Protected Security Workstation Emission Protected Crypto AG / 6950-e-2.ppt / DA-HQ / 0004 / 1 PC and Workstation Risks Security Risks: Unauthorized change of data (Sabotage, Viruses) Unauthorized use of data Data loss (deletion) Physical

More information

BenchVue Software Release History

BenchVue Software Release History Release Notes BenchVue Software Release History 3.5 Version Information Released Date: June 21, 2016 Windows 10: 32-bit & 64-bit (Professional, Enterprise, Education, Home versions) Windows 8: 32-bit &

More information

System Requirements. SuccessMaker 5

System Requirements. SuccessMaker 5 System Requirements SuccessMaker 5 System requirements are subject to change. For the latest information on system requirements, go to support.pearsonschool.com. For more information about Digital Learning

More information

Windows 8 Malware Protection Test Report

Windows 8 Malware Protection Test Report Windows 8 Malware Protection Test Report A test commissioned by Kaspersky Lab and performed by AV-Test GmbH Date of the report: January 11 th, 2013, last update: January 11 th, 2013 Executive Summary In

More information

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps 37 500:1 > 70 % pco. low noise high resolution high speed high dynamic range edge 4.2 LT scientific CMOS camera high resolution 2048 x 2048 pixel low noise 0.8 electrons USB 3.0 small form factor high dynamic range 37 500:1 high speed 40 fps high quantum efficiency > 70 % edge

More information

System Requirements for Web Applications

System Requirements for Web Applications for Web Overview This document contains the Web server recommendations and client workstation requirements for Business Portal for Microsoft Dynamics GP 2010, Workflow, Web Services and SQL Reporting Services

More information

LIMS and workflow integration as impulse to establish Virtual Microscopy to daily clinical use

LIMS and workflow integration as impulse to establish Virtual Microscopy to daily clinical use LIMS and workflow integration as impulse to establish Virtual Microscopy to daily clinical use K. Saeger 1, S. Wienert 1, K. Polus 2, D. Emmel 3, P. Hufnagl 4 1 VMscope GmbH, Berlin 2 Nexus/Dis GmbH, Frankfurt

More information

Audit4 Installation Requirements

Audit4 Installation Requirements Audit4 version 8.1+ 2015 Copyright 2012 S4S Pty Ltd Audit4 Support Matrix 2015 The following table provides details on the operating system and database engine support for Audit4 as at March 2015. Operating

More information

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY Adapted from Foundations of Biology I; Lab 6 Introduction to Microscopy Dr. John Robertson, Westminster College Biology Department,

More information

Why Choose Microsoft Project Portfolio Management?

Why Choose Microsoft Project Portfolio Management? Why Choose Microsoft Project Portfolio Management? A comparison between the centralized server solution and Microsoft Project 2013 Client Definition of Microsoft Project Portfolio Management Microsoft

More information

We Believe the Possibilities. Delphic AP

We Believe the Possibilities. Delphic AP We Believe the Possibilities. Delphic AP Designer Software for your pathology workflow. Delphic AP Tracks the specimen process cycle Labels cassettes and slides Enables sophisticated report creation Send

More information

CMAS. Compact Radio Signal Monitoring Solution

CMAS. Compact Radio Signal Monitoring Solution CMAS Compact Radio Signal Monitoring Solution CMAS is a high-performance, automatic radio monitoring solution for multichannel analysing and processing of HF and V/UHF signals. 1, 2, 20 MHz wideband input

More information