Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays



Similar documents
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

ProteinPilot Report for ProteinPilot Software

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Research-grade Targeted Proteomics Assay Development: PRMs for PTM Studies with Skyline or, How I learned to ditch the triple quad and love the QE

A Streamlined Workflow for Untargeted Metabolomics

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report

Tutorial 9: SWATH data analysis in Skyline

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

MarkerView Software for Metabolomic and Biomarker Profiling Analysis

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Lipid analysis by multiple "omic" mass spectrometry platforms Where we ve been, where we are, and where we re going

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Mass Spectrometry Based Proteomics

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Retrospective Analysis of a Host Cell Protein Perfect Storm: Identifying Immunogenic Proteins and Fixing the Problem

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit

La Protéomique : Etat de l art et perspectives

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

High-Throughput 3-D Chromatography Through Ion Exchange SPE

using ms based proteomics

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

NUVISAN Pharma Services

Introduction to Proteomics 1.0

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Thermo Scientific Mass Spectrometric Immunoassay (MSIA)

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Quantitative proteomics background

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Mass Spectrometry Signal Calibration for Protein Quantitation

Alignment and Preprocessing for Data Analysis

Analyst 1.6 Software. Software Reference Guide

WATERS QUANTITATIVE ANALYSIS solutions

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

A Quadrupole-Orbitrap Hybrid Mass Spectrometer Offers Highest Benchtop Performance for In-Depth Analysis of Complex Proteomes

A Hybrid Microchip/Capillary Electrophoresis Mass Spectrometry Platform for Rapid and Ultrasensitive Bioanalysis

ProSightPC 3.0 Quick Start Guide

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Chapter 14. Modeling Experimental Design for Proteomics. Jan Eriksson and David Fenyö. Abstract. 1. Introduction

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Pinpointing phosphorylation sites using Selected Reaction Monitoring and Skyline

Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

RapidFire High-throughput MS technology. Enhancing Drug Discovery Turning Mass Specs into Plate Readers

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Analyst Software

LC-MS/MS for Chromatographers

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Quantification of Multiple Therapeutic mabs in Serum Using microlc-esi-q-tof Mass Spectrometry

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Proteomic Analysis using Accurate Mass Tags. Gordon Anderson PNNL January 4-5, 2005

Analyst Software. Advanced User Guide

itraq Tips and Tricks

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Proteomics in Practice

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Transcription:

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative proteomics has become a key tool in proteomics research and is driven by the well-established sensitivity and selectivity attributes of multiple reaction monitoring (MRM) 1. As more extensive protein panels need to be monitored in a targeted way across multiple samples, higher MRM multiplexing is becoming essential for throughput. With the Scheduled MRM Algorithm (introduced with Analyst Software 1.5), significantly more MRM transitions can be monitored in a single acquisition while maintaining the quantitative reproducibility by time scheduling the acquisition of MRM transitions 2. The speed and sensitivity of MS/MS acquisition on the TripleTOF Systems have enabled a high resolution targeted quantification strategy. This strategy (termed MRM HR workflow) enables a degree of selectivity using high resolution that cannot be reached using standard triple quadrupole based In this technical note, the Scheduled MRM HR workflow (Figure 1) instruments. 3,4 that is now available in Analyst Software 1.6 TF will be assessed for the degree of multiplexing possible and its impact on quantitative reproducibility. Scheduling the acquisition of high resolution MS/MS on the target peptide in a short time window around its elution time allows more accumulation time to be used per peptide while still maintaining an optimized cycle time. In this way more peptides can be monitored in a single LCMS run with highest quantitation quality. Key Features of Scheduled MRM HR Workflow on TripleTOF Systems Using the precursor mass and retention time of each peptide, large numbers of peptides can be analyzed with high resolution and high quality data in a single run Scheduled MRM HR workflow enables longer accumulation times to be used during acquisition to provide good analytical precision at high multiplexing Figure 1. Scheduled MRM HR Workflow. Using knowledge of the elution time of each peptide, each precursor mass is monitored only during a short retention time window and the full scan high resolution MS/MS is acquired. This allows many more peptides to be monitored in a single LC run with higher accumulation time and an optimized cycle time for the highest quantitation quality. Full scan, high resolution data from Scheduled MRM HR workflow can be searched using ProteinPilot Software for confirmation of detected peptides Scheduled MRM HR Workflow is available on all TripleTOF Systems using Analyst Software 1.6 or higher. p 1

Methods Sample Preparation: Protein digests of Bovine Serum Albumin (Michrom BioResources), Myoglobin, (Michrom BioResources) and Beta-Galactosidase (AB SCIEX) were mixed together to create a test sample and loaded at 10 fmol on column. E.coli protein digest (Waters) was analyzed at 250 ng on column. Chromatography: Separation of protein digests was performed on an Eksigent NanoLC-Ultra 2D System and chiplc system (Eksigent, USA) in serial column mode. For each injection, the sample was desalted on a 75 µm x 15 mm analytical column, then eluted onto a second analytical column to create a 30cm column length for separation 5. Both column chips were filled with ChromXP C18-CL 3μm 120Å phase (Eksigent). Peptides were separated using a linear gradient formed by A (2% ACN, 0.1% FA) and B (98% ACN, 0.1% FA), from 12 32% of B over 60 minutes at a flow rate of 250 nl/min. Each injection was performed using a full loop injection with a 1 µl sample loop. Figure 2. Assay Development Process for Scheduled MRM HR Workflow. Experimental MS/MS data from IDA acquisition on TripleTOF Systems or MS/MS data from other sources can be imported for initial assay development. MRMHR workflow methods can be built for target proteins and peptides and used for assay refinement and retention time determination. Final Scheduled MRM HR workflow methods are built and used to run biological study. At the same time, a MultiQuant Software quantitation method is exported for use in final data processing. Mass Spectrometry: The MS analysis was performed on a TripleTOF 5600 system (AB SCIEX) using both the MRM HR Workflow 3,4 and Scheduled MRM HR Workflow. MS spectra were acquired in high resolution mode (>30,000) using 250 ms accumulation time per spectrum. Full scan MS/MS was acquired in high sensitivity mode with an accumulation time optimized per cycle. Collision energy (CE) was set using rolling collision energy with a collision energy spread (CES) of 5 V. Retention time window used for most datasets was 2.5 minutes, except for the 800 peptide method where a window of 2.0 minutes was used. Figure 3. High Resolution XIC Generation from a Scheduled MRM HR Workflow Demonstrates Sufficient Points Across an LC Peak. To quantify the peptide DGEDPGYTLYDLSER (m/z 577.2566) from Betagluatmate decarboxylase in E.coli cell lysate, a looped MS/MS spectrum is acquired across the LC peak at high resolution (>15,000) (top). High resolution XICs are generated for the peptide post-acquisition (bottom). Scheduling the acquisition of the MS/MS spectra around the elution time for the target peptide enables higher multiplexing while maintaining data quality. Data Processing: Peptide identification on the MS/MS collected on the E.coli sample was performed using ProteinPilot Software 4.0 and an xml file was created for import into Skyline Software (University of Washington, USA). Proteins and peptides were selected for analysis and MRM HR workflow data was acquired and processed to determine peptide retention times. Final Scheduled MRM HR workflow acquisition methods and MultiQuant Software quantitation methods were created simultaneously by Skyline. Final data processing was performed using MultiQuant Software. p 2

Assay Development Workflow Assay development is simplified for MRM HR and Scheduled MRM HR workflow as full scan MS/MS is always acquired and fragment ion selection is done post-acquisition. Only the peptide precursor ion m/z and the collision energy are set up-front. This process of developing Scheduled MRM HR workflow assays has been implemented within the Skyline Software and consists of four simple steps (Figure 2). 1. Previously acquired MS/MS data from a database search result, such as ProteinPilot Software, can be brought into the Skyline software to provide protein, peptide and MS/MS information. Assays can also be developed from protein sequence information. 2. Initial MRM HR workflow acquisition methods are constructed using the intensity information from MS/MS data. 3. These acquisition methods are run against the biological sample of interest and the MRM HR data is evaluated. Peptide and fragment ion selections can be refined to develop the highest quality assay (Figure 3). Retention times of each peptide are also determined. 4. The final Scheduled MRM HR workflow acquisition method is built and will be used for data collection on the biological sample of interest. In addition, a MultiQuant Software quantitation method is created to be used for final data processing. Figure 4. Assessing the Quantitative Accuracy of Higher Multiplexing using Scheduled MRM HR Workflow. The effect of higher numbers of peptides on the reproducibility of 30 peptides across 5 or more replicate injections on the TripleTOF 5600 system was assessed. Because the data was acquired using the Scheduled MRM HR Workflow, the reproducibility of up to 800 peptides showed very good quantitative accuracy. Figure 5. Quantitative Reproducibility for Peptides in E.coli Cell Lysate Digest. 139 peptides from proteins across the abundance range in an E.coli digest were monitored using the Scheduled MRM HR workflow. Replicate analysis was performed and the reproducibility at both the XIC and peptide level was assessed. Extremely low variance was found at the transition XIC level (blue trace) across replicates with 90% of data having CVs of less than 20% and over 95% of data at the peptide level (orange trace). Assessing Quantitative Reproducibility at High Multiplexing An experiment was designed to assess the effects of higher multiplexing on analytical reproducibility on the TripleTOF 5600 system. Thirty peptides from Beta-Galactosidase (BG), Horse Myoglobin (Mb) and Bovine Serum Albumin (BSA) protein digests were monitored using the MRM HR Workflow with a cycle time of 1.8 seconds. Random generated Q1 m/z values and retention times were then added to the real peptides and these were used to build Scheduled MRM HR workflow methods with the same cycle time of 1.8 seconds and a retention time window of 2.5 minutes. Each acquisition method was run on the protein digests in 5 or more replicates by NanoLC. The reproducibility of fragment ion XICs for the 30 real peptides from each method across the replicate injections was used as a measure of the analytical reproducibility (Figure 4). The total number of XICs (in %) at a specific % coefficient of variation (%CV) was computed and plotted in Figure 4 (top), the median %CV was also computed (Figure 4, bottom). This cumulative reproducibility plot shows the majority of XICs achieved %CV of less than 20%, even at the highest peptide multiplexing in this experiment. Scheduled MRM HR workflow for 200 and 400 peptides (orange and blue) showed reproducibility similar or better to the MRM HR workflow data (green) on 30 peptides as the MS/MS accumulation time was greater than 40 msec in those methods. For 600 peptides (red line), there was a small drop in data quality as the accumulation time decreased to 30 msec. To run 800 peptides (purple line), the retention time window was reduced from 2.5 minutes to 2 minutes to maintain an accumulation time p 3

of around 30 msec and good quantitative reproducibility was observed. In addition, the effect of multiplexing on quantitative reproducibility for a set of peptides from proteins across the abundance range in an E.coli digest was studied. A total number of 139 peptides from 32 proteins were analyzed in a single LC/MS run using Scheduled MRM HR workflow and 795 fragment ions were quantified. Replicate analysis was performed and the reproducibility was assessed at both the transition XIC area and peptide level (multiple fragment ion XIC areas were summed to obtain a peptide area). Figure 5 shows extremely low variance at the transition XIC level (blue trace) and across replicates with 90% of data having CVs of less than 20% and over 95% of data at the peptide level (orange trace). The reproducibility of the retention times for the peptides across the replicate injections was an average of 0.2% RSD. To evaluate the XIC peak area quality and reproducibility further, each individual data point was plotted based on its intensity and %CV (Figure 6). The variance in the data increases at the lower intensity regimes as expected. The red line shows that only at the very lowest intensity bins does the %CV exceed 10%. Increased Specificity on a TripleTOF Systems When targeted quantification is done on a triple quadrupole based instrument, quadrupoles are typically set at unit resolution, transmitting a 0.7 Da wide window in both Q1 and Q3. The difference with the MRM HR Workflow is that the TOF analyzer replaces the Q3 fragment ion detection with its inherent advantages of detecting all ions, at high resolution and high mass accuracy. To illustrate the advantage of high resolution, multiple fragment ions for the peptide VNAIAAPTR from an uncharacterized protein yrar were extracted at variable extraction widths (Figure 7). At the narrow extraction widths possible from the high resolution TOF data, good specificity is observed with minimal interfering peaks (Figure 7, top). When the extraction width is widened to simulate monitoring using a quadrupole, an interference is observed for two of the transitions (Figure 7, bottom). Figure 6. Reproducibility Assessment of Transition Peak Area versus %CV. The blue dots represent the individual transition peak areas and are plotted based on their intensity and observed variance. The red line describes that 90% out of the total transitions area are below the shown %CV. This demonstrates the very high data reproducibility that was obtained by using the Scheduled MRM HR workflow on the TripleTOF 5600 System. Figure 7. Removing Interferences or Background through High Resolution Fragment Ion Extractions. Multiple fragment ions for the peptide VNAIAAPTR from an uncharacterized protein yrar were extracted at variable extraction widths. High resolution extraction of fragment ions can provide higher specificity in complex matrices when specific interferences are present. p 4

Conclusions High resolution targeted assays can provide higher specificity for some peptides in complex biological samples and now with the ability to time-scheduled acquisition, high multiplexing can be achieved. TripleTOF Systems have the MS/MS sensitivity and speed to perform MRM-like analysis. Post-acquisition extraction of fragment ions from the high resolution TOF MS/MS data allows for high specificity MRM-like data (MRM HR Workflow) to be obtained. Scheduled MRM HR Workflow extends the multiplexing possible with this high resolution workflow, maintaining quantitative robustness. In the multiplexing test on standard protein digests, up to 800 precursors were analyzed in a single method and high reproducibility was observed. This reproducibility translated into the complex biological sample test, where 130 peptides were monitored in E.coli with 90% of the transition peak areas showing reproducibility better than 20%. References 1. Anderson L and Hunter CL (2006) MCP 5.4, 573-588. 2. The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring - Delivering up to 4000 MRM Transitions per LC Run. AB SCIEX Technical Note 0921010-03. 3. Impact of High-Resolution MRM Analysis on Complex Proteomes - Targeted Protein Quantification using MRM HR Workflow on the TripleTOF 5600 System. AB SCIEX Technical Note 4040211-01. 4. Increasing LCMS Assay Robustness through Increased Specificity using High Resolution MRM-like Analysis - MRM HR Analysis using the TripleTOF 5600 System. AB SCIEX Technical Note 2780411-01. 5. Increasing Depth of Coverage using Serial Two- Column Workflows - Powerful Proteomics Workflows Enabled by the Eksigent chiplc -nanoflex with nanolc-ultra System. AB SCIEX Technical Note 1870211-02. For Research Use Only. Not for use in diagnostic procedures. 2012 AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: 6260212-01 p 5