Malware Protection White Paper



Similar documents
Malware Protection II White Paper Windows 7

Remote Monitoring and Control of the R&S RTO with a Web Browser Application Note

Remote Monitoring and Control of the R&S FSV with a Web Browser

Remote Monitoring and Control of the R&S FSL with a Web Browser

Sophos Anti-Virus standalone startup guide. For Windows and Mac OS X

Sophos for Microsoft SharePoint startup guide

Sophos Anti-Virus for NetApp Storage Systems startup guide. Runs on Windows 2000 and later

Sophos Anti-Virus for NetApp Storage Systems startup guide

SecuraLive ULTIMATE SECURITY

Kaseya Server Instal ation User Guide June 6, 2008

Getting Started. Symantec Client Security. About Symantec Client Security. How to get started

Apple ipad Remote Control for Broadcasting T&M Instruments Application Note

AVG 8.5 Anti-Virus Network Edition

Step-by-Step Guide to Securing Windows XP Professional with Service Pack 2 in Small and Medium Businesses

Agilent Technologies Electronic Measurements Group Computer Virus Control Program

User Manual. Onsight Management Suite Version 5.1. Another Innovation by Librestream

Getting Started with Symantec Endpoint Protection

Magaya Software Installation Guide

Core Protection for Virtual Machines 1

Avira Endpoint and Security. HowTo

DocuPrint C3290 FS Features Setup Guide

Avira Small Business Security Suite Avira Endpoint Security. Quick Guide

Maintenance Guide. Outpost Firewall 4.0. Personal Firewall Software from. Agnitum

GE Healthcare Life Sciences UNICORN Administration and Technical Manual

Sophos Endpoint Security and Control standalone startup guide

Data and Network Security

Reporting for Contact Center Setup and Operations Guide. BCM Contact Center

Legal Notes. Regarding Trademarks KYOCERA Document Solutions Inc.

XStream Remote Control: Configuring DCOM Connectivity

Topaz Installation Sheet

Avira Small Business Security Suite. HowTo

Managing Software Updates with System Center 2012 R2 Configuration Manager

Installation Instruction STATISTICA Enterprise Server

Portions of this product were created using LEADTOOLS LEAD Technologies, Inc. ALL RIGHTS RESERVED.

Team Foundation Server 2010, Visual Studio Ultimate 2010, Team Build 2010, & Lab Management Beta 2 Installation Guide

Table of Contents. CHAPTER 1 About This Guide CHAPTER 2 Introduction CHAPTER 3 Database Backup and Restoration... 15

How To Install Avira Small Business Security Suite (Small Business) On A Microsoft Microsoft Server (Small Bserver) For A Small Business (Small) Computer (Small Server)

Trend ScanMail. for Microsoft Exchange. Quick Start Guide

QUANTIFY INSTALLATION GUIDE

Net Protector Admin Console

Get Started Guide - PC Tools Internet Security

Diamond II v2.3 Service Pack 4 Installation Manual

Windows Firewall Configuration with Group Policy for SyAM System Client Installation

Using. Microsoft Virtual PC. Page 1

Sharp Remote Device Manager (SRDM) Server Software Setup Guide

Virtual CD v10. Network Management Server Manual. H+H Software GmbH

Getting started. Symantec AntiVirus Corporate Edition 8.1 for Workstations and Network Servers

Universal Management Service 2015

Configuring NXT Hardware with Mercury Inside into Doors.NET TM Application Note

Yale Software Library

Troubleshooting Sprint Mobile Broadband USB Modem by Novatel Wireless TM (Ovation TM U727)

Getting started. Symantec AntiVirus Corporate Edition. About Symantec AntiVirus. How to get started

CentreWare Internet Services Setup and User Guide. Version 2.0

Networking Best Practices Guide. Version 6.5

Using TS-ACCESS for Remote Desktop Access

Symantec Endpoint Protection Getting Started Guide

Remote Control of ROMES on Windows 7. Application Note

VRC 7900/8900 Avalanche Enabler User s Manual

Legal Notes. Regarding Trademarks KYOCERA Document Solutions Inc.

Internet Security Software

Setting Up a Unisphere Management Station for the VNX Series P/N Revision A01 January 5, 2010

Immotec Systems, Inc. SQL Server 2005 Installation Document

Network Setup Guide. Introduction. Setting up for use over LAN

Creating client-server setup with multiple clients

Personal Call Manager User Guide. BCM Business Communications Manager

MGC WebCommander Web Server Manager

ReadyNAS Setup Manual

GUARD1 PLUS SE Administrator's Manual

PC User s Guide PC User s Guide Muratec America, Inc.

Step-by-Step Guide for Creating and Testing Connection Manager Profiles in a Test Lab

R&S AFQ100A, R&S AFQ100B I/Q Modulation Generator Supplement

How To Install Safari Antivirus On A Dv8000 Dv Recorder On A Pc Or Macbook Or Ipad (For A Pc) On A Microsoft Dv8 (For Macbook) On An Ipad Or Ipa (

What Is Ad-Aware Update Server?

Administration guide. Océ LF Systems. Connectivity information for Scan-to-File

Symantec AntiVirus Corporate Edition Patch Update

HoneyBOT User Guide A Windows based honeypot solution

Network Storage Link

Frequently Asked Questions: Xplornet Internet Security Suite

Portions of this product were created using LEADTOOLS LEAD Technologies, Inc. ALL RIGHTS RESERVED.

Table of Contents. Introduction...9. Installation Program Tour The Program Components...10 Main Program Features...11

Quickstart Guide. First Edition, Published September Remote Administrator / NOD32 Antivirus 4 Business Edition

TecLocal 4.0 MultiUser Database

Internet and Intranet Calling with Polycom PVX 8.0.1

Allworx OfficeSafe Operations Guide Release 6.0

AVG 9.0 Internet Security Business Edition

Installation Instruction STATISTICA Enterprise Small Business

Setup and Configuration Guide for Pathways Mobile Estimating

FortKnox Personal Firewall

Software Version 5.1 November, Xerox Device Agent User Guide

HP MediaSmart Server Software Upgrade from v.2 to v.3

Using XP Service Pack 2 with Cameleon

Aspera Connect User Guide

Partner. Sage Pastel. Accounting. Installation Guide

K7 Business Lite User Manual

Team Foundation Server 2013 Installation Guide

Getting started. Symantec AntiVirus Business Pack. About Symantec AntiVirus. Where to find information

This document details the following four steps in setting up a Web Server (aka Internet Information Services -IIS) on Windows XP:

(1) Network Camera

InduSoft Web Studio + Windows XP SP2. Introduction. Initial Considerations. Affected Features. Configuring the Windows Firewall

Basic Setup Guide. Remote Administrator 4 NOD32 Antivirus 4 Business Edition Smart Security 4 Business Edition

Transcription:

Malware Protection White Paper Rohde & Schwarz recognizes the potential risk of computer virus infection when connecting Windows -based test instrumentation to other computers via local area networks (LANs), or using removable storage devices. This white paper introduces measures to minimize malware threats and discusses ways to mitigate risks while insuring that instrument performance is not compromised. The paper discusses the use of anti-virus software, with recommended configuration settings. It also outlines how to keep the Windows XP operating system properly updated through regular installation of OS patches. Malware Protection White Paper Jochen Wolle 06.2010-1EF73_0e

Windows -Based Instruments Table of Contents 1 Windows -Based Instruments... 4 1.1 Overview...4 1.2 Computer Virus Control Program...4 1.3 Preventative Maintenance Considerations...4 1.4 User / Admin Account...5 2 Firewall Settings... 6 2.1 Firewall Port Configuration...6 2.2 Changing Firewall Settings...7 3 USB Devices... 8 3.1 Disable USB Autorun Function...8 3.2 Scan USB Devices...9 4 Anti-Virus Software... 10 4.1 Norton AntiVirus 2010...11 4.1.1 Installation...11 4.1.2 Requirements...11 4.1.3 Deactivate Automatic Updates and Virus Scans...12 4.1.4 Update Virus Signatures and Scan for Viruses on Demand...14 4.2 Kaspersky Anti-Virus 2010...16 4.2.1 Installation...16 4.2.2 Requirements...16 4.2.3 Deactivate Automatic Updates and Virus Scans...17 4.2.4 Update Virus Signatures and Scan for Viruses on Demand...19 4.3 Microsoft Security Essentials...20 4.3.1 Installation...20 4.3.2 Requirements...20 4.3.3 Deactivate Automatic Virus Scans...21 4.3.4 Update Virus Signatures and Scan for Viruses on Demand...22 4.4 Scanning from a USB thumb drive...23 4.5 Scanning Instruments from another PC...23 4.5.1 Share Drives of the Instrument...23 4.5.2 Mapping Drives and Scanning for Viruses...25 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 2

Windows -Based Instruments 5 Windows Patches and Updates... 27 5.1 Installation and Configuration of Windows Update Agent...28 5.2 Configuring Automatic Updates...29 5.3 Instruments connected to a Windows Update Server...30 5.4 Configuring Automatic Updates...31 5.5 Viewing installed Updates...31 6 Related Documents and Links... 32 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 3

Windows -Based Instruments 1 Windows -Based Instruments 1.1 Overview Rohde & Schwarz is dedicated to ensuring that all R&S products are shipped virusfree. Instruments that run Windows operating systems should be protected from malware just like any other PC. Users are strongly advised to take measures to protect their instruments such as using anti-virus software and installing OS patches and updates on a regular basis. It is highly recommended that you work closely with your IT department or system administrator to ensure compliance with your company policies when connecting instruments to your company s network. 1.2 Computer Virus Control Program Rohde & Schwarz recognizes the potential risk of computer virus infections on Windows-based instrumentation which are connected to local area networks (LANs). Rohde & Schwarz has established processes within the company to take all reasonable precautions to prevent the spread of viruses from instruments to our customers computers and networks: All computers used within Rohde & Schwarz that may be connected to instruments destined for customers are equipped with centrally managed firewall and anti-virus software and maintain the latest virus definitions. Computers and removable storage devices are scanned regularly to prevent the spread of computer viruses. Strict virus control protocols have been established in manufacturing, service, support, sales, distribution and demonstration environments. This includes the use of isolated LANs, scanning of instruments and removable storage devices and/or re-imaging hard drives, as appropriate depending upon instrument configuration. Procedures have been established for all Rohde & Schwarz employees who come in contact with customer instruments to reinforce anti-virus security protocols. This includes all personnel from manufacturing, service, support, sales and distribution. 1.3 Preventative Maintenance Considerations The steps described above help to guarantee that any instrument from Rohde & Schwarz will be virus-free when delivered to the customer. From that point on it is the user's responsibility to ensure the security of the instrument. Before connecting the instrument to your company's network, please consult with your IT department or system administrator to determine what specific policies apply. Remember that the instrument appears to be a standard computer to the network. Follow your company s policies with regards to computer security and virus protection. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 4

Windows -Based Instruments It is also important to update both the virus definitions and operating system regularly. Rohde & Schwarz recommends checking both virus definitions and operating system updates, in addition to scanning the instrument for any malware, at least once per week. Be sure to always update the OS and anti-virus definitions if advised to do so by your IT department or system administrator. The following steps should be taken to ensure the instrument s operating system is protected: Use the Internet firewall on the instrument. Scan all removable storage devices (e.g. USB thumb drives) that are used with an instrument regularly and deactivate the Autorun / Autoplay function to prevent inadvertent execution of malicious code from these devices. Install the latest Windows patches and updates on the instrument. Scan the instrument regularly with anti-virus software, and keep virus definition files updated. It is NOT recommended to run anti-virus software in the background ( on-access mode) as this will impact instrument performance significantly. 1.4 User / Admin Account Windows requires that users identify themselves by entering a user name and password in a login window. In general, R&S instruments provide a factory-installed autologin function, i.e. login is carried out automatically during the startup of the instrument. The factory default for this auto-login function has administrator rights with unrestricted access, so that printer installation and network configuration are possible. For many instruments you can set up two types of user account, either an administrator account with unrestricted access to the instrument OS or a standard user account with limited access. You can manage the accounts via Windows Start Control Panel User Accounts. Refer to the instrument user manuals for more information on how to change or add new users and on how to de-activate the automatic login. Note: Changing firewall settings, installing and configuring Anti-Virus software and Windows updates require unrestricted administrator rights. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 5

Firewall Settings 2 Firewall Settings With Windows XP SP2 and later versions, a firewall can be used to protect a computer or instrument against attacks from the network. R&S instruments are shipped with the Windows firewall enabled and preconfigured. Having the firewall activated on the instruments is helpful even when you use the instruments within your company s protected network. With the number of worms, viruses and other malware circulating on the Internet today, it is inevitable that something will penetrate the enterprise firewall. Instrument firewalls not only help protect against threats inside the perimeter, but they can also prevent the spread of many viruses and worms. If you have additional requirements for security and protection please contact your IT department or system administrator to ensure conformity with your company s security policy. 2.1 Firewall Port Configuration R&S instruments are preconfigured in such a way that all ports and connections for remote control are enabled. See the following table for details: Ports Service Description 21 tcp FTP Instrument web server - FTP port 80 tcp (HTTP) Web server Instrument web server (LXI) 111 tcp, 111 udp Portmapper Portmapper service for VXI-11 / LXI 161 udp 162 udp 705 tcp (AgentX) SNMP Standard ports for SNMP agent 319 tcp udp 320 tcp udp 1588 PTP LXI Class B/A IEEE1588 PTP (Precision Time Protocol) 2525 tcp RSIB R&S SCPI socket connection 4880 tcp HiSLIP High Speed LAN Interface Protocol 5025 (data) 5125 (abort) TCP Socket Raw SCPI socket connection 5044 tcp udp LXI Class B LXI LAN messages and events Multicast address udp: 224.0.23.159 5800 tcp 5900 tcp VNC Instrument soft front panel via web server (Browser interface) 13217 tcp udp RS Installer R&S Software distributor service 14142-16383 tcp udp (dynamic assignment) ONC-RPC Sun ONC-RPC protocol VXI-11 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 6

Firewall Settings 2.2 Changing Firewall Settings Rohde & Schwarz highly recommends the use of the firewall on your instrument. Note that changing firewall settings requires administrator rights. You can manage the firewall settings via Windows Start Control Panel Windows Firewall: Problems that are related to the default firewall configuration appear in two ways: Client programs may not receive data from the instrument. Server programs that are running on the instrument may not respond to client requests. If a program is being blocked, you may receive the following Windows Firewall Security Alert: To unblock the program, click Unblock in the Security Alert dialog box. You can find a detailed description for firewall setup and configuration at: http://support.microsoft.com/kb/875357/en-us 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 7

USB Devices 3 USB Devices USB thumb drives and removable hard drives are now common throughout the workplace, as they have considerable storage capacity and can be used to store instrument settings, measurement results, hardcopies etc. in a very convenient way. However, they also introduce new problems; a large number of viruses, trojans and other malware infect computers via USB storage devices. Once an infected USB drive is plugged into an instrument, the malware on it can spread through the whole network. 3.1 Disable USB Autorun Function Generally, viruses that propagate via USB drives use the autorun function of Windows, as it does not require any user confirmation and runs silently in the background. R&S instruments are preconfigured with the Autorun / Autoplay function disabled. This prevents any malware from automatically executing itself from a USB drive. You can control or change the settings using the Group Policy editor. If the instrument is used on a corporate network, and is a member of the network domain, then Group Policy settings can be configured centrally by your IT department or system administrator. Click Windows Start Run and then enter gpedit.msc to open the group policy settings. Go to Computer Configuration Administrative Templates System, scroll down and double-click on Turn off Autoplay to start the settings dialog: Click on the Enabled radio button, then from the Turn off Autoplay on drop down list select All drives to prevent any program from automatically executing from any USB drive or other removable media. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 8

USB Devices Note: If System is not listed, a settings template needs to be added. Rightclick Administrative Templates and choose Add/Remove Templates. In the dialog, click Add, and select system.adm. Click Open and Close to return to the main window. You can find a detailed description of the autorun function, if required, at: http://support.microsoft.com/kb/967715/en-us 3.2 Scan USB Devices Rohde & Schwarz recommends scanning USB thumb drives and removable hard drives with anti-virus software on a regular basis to keep them free from malware. Use your computer and your anti-virus software to scan the USB storage devices before inserting them into a R&S instrument. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 9

4 Anti-Virus Software As with personal and business computers, users must take appropriate steps to protect their instruments from infection. Beside the use of strong firewall settings and regularly scanning any removable storage device used with a R&S instrument, it is also recommended that anti-virus software be installed on the instrument. While Rohde & Schwarz does NOT recommend running anti-virus software in the background ( onaccess mode) on Windows-based instruments, due to potentially degrading instrument performance, it does recommend running it during non-critical hours at least once per week. Today s anti-virus software requires a significant amount of system resources (both hard drive space and memory consumption). Therefore some instruments may not be capable of installing or running anti-virus software due to limited resources. Other options in that case are to scan these instruments with software run from a USB thumb drive, or to mount these instruments as a drive on the network and scan them from another computer with anti-virus software. These options will be detailed later. Note: The following sections are intended to highlight recommendations for anti-virus software, using as examples a few commonly used programs. It is recognized that there are other capable programs; the ones used in the following sections serve as general examples and the principles apply to other programs which may be used by your IT department or system administrator. Installing, configuring and using anti-virus software requires administrator rights. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 10

4.1 Norton AntiVirus 2010 This section describes the installation, configuration and use of the anti-virus software Norton AntiVirus 2010 on R&S instruments. 4.1.1 Installation Install the Norton AntiVirus 2010 software on the instrument as described in the vendor s manual. Deactivate the control check box I want to join the fight... on the installation welcome page and start the installation with Agree & Install: After installation completes, Norton AntiVirus 2010 tries to connect to the Symantec server to get the latest virus signatures and program updates (a process called Live- Update). 4.1.2 Requirements Norton AntiVirus 2010 has the following requirements: 200 MB Free space on the instrument s hard drive 256 MB Memory Windows XP SP2 or later Make sure that as a minimum Windows XP SP2 is installed on your R&S instrument. Refer to the instrument s user manual for how to check the current OS version. If it s based on an older version, contact your R&S representative for update possibilities. For many instruments R&S provides an instrument recovery DVD with the latest OS version for re-imaging the instrument s hard drive. During LiveUpdate or a virus scan, two processes (both named ccsvchst.exe) run on the instrument and occupy up to 270 MB of memory. Therefore, Rohde & Schwarz recommends that the firmware on the instrument be stopped before starting LiveUpdate or a virus scan. Refer to the instrument s manual for how to stop execution of the instrument s firmware. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 11

4.1.3 Deactivate Automatic Updates and Virus Scans Symantec LiveUpdate needs an Internet connection and administrator rights in order to be executed. The updates are downloaded from the Symantec server or from a proxy server in your company. Contact your IT department or system administrator for details on your company s policy. Configure LiveUpdate and Scans to be executed on demand to avoid degradation of the instrument s performance. Double-click on the Norton AntiVirus icon in the system tray to bring up the main dialog: Click on Settings to configure LiveUpdate and scans: Deactivate both Automatic LiveUpdate and Pulse Updates in the Computer Settings dialog: Save the settings with OK. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 12

The final step of the configuration is to deactivate automatic virus scans. Bring up the main dialog as in the steps above and select Run Custom Scan. Select Scan scheduled to modify the list of scheduled virus scans: Delete entries in the schedule dialog until the drop down box is empty. This deactivates any automatic virus scans: 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 13

4.1.4 Update Virus Signatures and Scan for Viruses on Demand To start LiveUpdate for the virus signature database and the anti-virus software on the instrument, you need an Internet connection. Double-click on the Norton Antivirus icon in the system tray to bring up the main dialog: Start the update process with Run LiveUpdate: When LiveUpdate is finished, press the OK button. Once the virus signature database is updated you can start the virus scan process. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 14

To start the virus scan process press Scan Now in the main dialog: In the following dialog you can choose between the three scan options: You can either scan commonly infected areas (Quick Scan), scan your entire instrument (Full System Scan), or run a custom scan of drives, folders and files (Custom Scan). When the scan process is complete click on Finish to close the dialog. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 15

4.2 Kaspersky Anti-Virus 2010 This section describes the installation, configuration and use of the anti-virus software Kaspersky Anti-Virus 2010 on R&S instruments. 4.2.1 Installation Install the Kaspersky Anti-Virus 2010 software on the instrument as described in the vendor s manual. After installation completes, you should start Kaspersky Anti-Virus 2010 to connect to the Kaspersky server to get the latest virus signatures and program updates. 4.2.2 Requirements Kaspersky Anti-Virus 2010 has the following requirements: 300 MB Free space on the instrument s hard drive 256 MB Memory Windows XP SP2 or later Make sure that as a minimum Windows XP SP2 is installed on your R&S instrument. Refer to the instrument s manual for how to check the current OS version. If it s based on an older version contact your R&S representative for update possibilities. For many instruments R&S provides an instrument recovery DVD with the latest OS version to re-image the instrument s hard drive. During a virus signature / program update or a virus scan, two processes (both named avp.exe) run on the instrument and occupy up to 320 MB of memory. Therefore, Rohde & Schwarz recommends that the firmware on the instrument be stopped before starting updates or a virus scan. Refer to the instrument s manual for how to stop execution of the instrument s firmware. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 16

4.2.3 Deactivate Automatic Updates and Virus Scans Kaspersky Anti-Virus 2010 needs an Internet connection and administrator rights in order to be executed. The updates are downloaded from the Kaspersky server or from a proxy server in your company. Contact your IT department or system administrator for details on your company s policy. Configure virus definition updates and virus scans to be executed on demand to avoid degradation of the instrument s performance. Double-click on the Kaspersky Anti-Virus icon in the system tray to bring up the main dialog: Click on My Update Center. To deactivate automatic virus / program updates select Manually under Start update Run mode: To configure automatic virus scans select Settings in the upper right of the main dialog: 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 17

Select Full Scan in the left hand navigation pane and then Settings to configure the Run Mode: Select Manually under Schedule and confirm with OK to deactivate automatic virus scans: 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 18

4.2.4 Update Virus Signatures and Scan for Viruses on Demand To start updates for the virus signature database and the anti-virus software on the instrument, you need an Internet connection. Double-click on the Kaspersky Anti-Virus icon in the system tray to bring up the main dialog: To start the update process select My Update Center on the left hand tabs in the main dialog and then Start Update: To start a virus scan select Scan My Computer on the left hand tabs in the main dialog and then Start Full Scan: Other options for virus scanning are Quick Scan or Objects Scan. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 19

4.3 Microsoft Security Essentials This section describes the installation, configuration and usage of the anti-virus software Microsoft Security Essentials on R&S instruments. 4.3.1 Installation Install the Microsoft Security Essentials anti-virus software on the instrument as described in the vendor s manual. No Internet connection is necessary to complete the installation. After installation completes, Microsoft Security Essentials tries to connect to the Microsoft server to get the latest virus signatures and program updates. To prevent this, deactivate the control box Scan my computer for potential threats and select Finish to complete the installation. 4.3.2 Requirements Microsoft Security Essentials has the following requirements: 300 MB Free space on the instrument s hard drive 256 MB Memory Windows XP SP2 or later Make sure that as a minimum Windows XP SP2 is installed on your R&S instrument. Refer to the instrument s manual for how to check the current OS version. If it s based on an older version contact your R&S representative for update possibilities. For many instruments R&S provides an instrument recovery DVD with the latest OS version to re-image the hard drive of the instrument. During a virus signature / program update or a virus scan, two processes (named MsMpEng.exe and msseces.exe) run on the instrument and occupy up to 110 MB of memory. Therefore, Rohde & Schwarz recommends that the firmware on the instrument be stopped before starting a virus scan. Refer to the instrument s manual for how to stop execution of the instrument s firmware. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 20

4.3.3 Deactivate Automatic Virus Scans Microsoft Security Essentials needs an Internet connection and administrator rights to be executed. Note: The virus signature updates are downloaded from the Microsoft server automatically if the virus signature database is older than 24h. There is no configuration to disable this automatic update process. Also, Microsoft Security Essentials can not be configured to use a proxy server in your company. Configure virus scans to be executed on demand to avoid degradation of the instrument s performance. Double-click on the Microsoft Security Essentials icon in the system tray to bring up the main dialog: Select the Settings tab and Scheduled scan on the left hand navigation pane. Deselect Run a scheduled scan to deactivate automatic virus scans. Store the configuration with Save changes. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 21

4.3.4 Update Virus Signatures and Scan for Viruses on Demand To start updates for the virus signature database and the anti-virus software on the instrument you need an Internet connection. Double-click on the Microsoft Security Essentials icon in the system tray to bring up the main dialog: Select the Update tab in the main dialog and then press Update to start the update process: To start a virus scan select Full Scan in the main dialog and then Scan now: Other options for virus scanning are Quick Scan or Objects Scan. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 22

4.4 Scanning from a USB thumb drive There may be instruments that do not have the resources to have anti-virus software installed. For these instruments, scanning can be done from a USB thumb drive. 4.5 Scanning Instruments from another PC Before scanning with anti-virus software from another computer the instrument has to be mounted as a drive on the network. Note: Scanning instrument hard drives remotely has some limitations and should only be used if the other options are not available: only visible files can be scanned, memory and processes will not be scanned, and a rootkit can completely hide itself. 4.5.1 Share Drives of the Instrument Connect the instrument to the network. Check for the instrument s computer name and Workgroup. (This information is needed later on to scan this specific instrument from your PC). To view these settings use Windows Start Control Panel System and select the Computer Name tab in the dialog: In this case the instrument s computer name is FSx-000000 and is part of the workgroup WORKGROUP. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 23

Start the Windows Explorer on the instrument and expand the folder My Computer to see all the drives. Right-click on Drive C: to open the context menu and select Sharing and Security: In the dialog which opens up select New Share to enter a name e.g. "DriveC", and confirm with OK. The symbol for Drive C: should now have changed to the symbol for a shared drive: Repeat the procedure for any other drives (e.g. drive D: and E: of your instrument). This enables a remote virus scan to access all drives of the instrument. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 24

4.5.2 Mapping Drives and Scanning for Viruses Open the Windows Explorer on your PC and expand the folders My Network Places Entire Network Microsoft Windows Network Workgroup. Note that Workgroup might be a different name if you used a different name in the workgroup configuration. Click on the computer name of the instrument you want to scan, e.g. Fsx-000000 in this example. You will be prompted for a user name and password. Enter User Name and Password - see the instrument s manual for these settings. The folders of the instrument will appear in the right hand window. Select Tools in the menu bar and then Map Network Drive 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 25

In the Map Network Drive dialog, map the first shared drive (e.g. DriveC of the instrument) as a network drive (e.g. N: on the control PC). You can use Browse to find the complete network name of the shared drive (e.g. \\FSX-000000\DriveC ) in the network tree. Click Finish to complete the network drive mapping. In this example the instrument drive C: is now mapped to drive N: on the control PC. Repeat these steps for any other instrument hard drives and map them to free drives on the control PC. To scan the instrument s hard drive, start the anti-virus software on the control PC. Select one of the mapped drives of the instrument and run a virus scan. Please refer to the anti-virus software s user manual for how to scan a network drive. In order to return the instrument to its original state, the drive sharing on the instrument has to be removed: Start the Windows Explorer and expand folder My Computer to see all drives. Right-click on Drive C: to open the context menu. Select Sharing in the properties dialog. Expand the Share name list and select DriveC then click on Remove Share or on the radio button Do not share this folder. Finally click OK to remove the drive sharing. Repeat these steps for any other shared drives, if applicable. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 26

Windows Patches and Updates 5 Windows Patches and Updates Microsoft regularly creates security updates and other patches to protect Windowsbased operating systems. These are released through the Microsoft Update website and associated update server. Instruments using Windows, especially those that connect to a network, should be updated regularly. Note that Microsoft Update supersedes Windows Update, which was for Windowsbased products only. The following sections describe the installation of the Windows Update Agent and its configuration. This enables the instrument to download and install the latest Windows patches and updates. Make sure that as a minimum Windows XP SP2 is installed on your R&S instrument. Refer to the instrument s manual for how to check the current OS version. If it s based on an older version contact your R&S representative for update possibilities. For many instruments R&S provides an instrument recovery DVD with the latest OS version to re-image the hard drive of the instrument. Note: It s NOT recommended to upgrade an instrument from SP2 to SP3 with the Microsoft Update service, or by manual installation of a standalone service pack executable. For most instruments re-imaging of the OS is necessary. In general, there are two scenarios for instruments using the Microsoft Update service: The instruments are permitted access to the Internet, and download updates directly from the Microsoft Update server. The instruments download updates from an update server in your company. In the second scenario, system administrators set up a server running Windows Server Update Services (WSUS) inside the corporate firewall, which synchronizes content directly with Microsoft Update and distributes updates to client computers and instruments. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 27

Windows Patches and Updates 5.1 Installation and Configuration of Windows Update Agent Most R&S instruments are based on Windows XP Embedded which is a customizable version of Windows XP Professional. The OS is scaled and optimized to the requirements of the specific instruments. Therefore, in many cases, the Windows update service has to be separately installed on the instruments. Download the Windows Update Agent installer WindowsUpdateAgent30-x86.exe from the Microsoft web site http://go.microsoft.com/fwlink/?linkid=100334 and copy it onto a USB thumb drive. The installation is straightforward and does not present critical installation options. The Windows Update Agent installation steps are listed below: Press CTRL + ESC or click Start to bring up the Windows Start menu and then start the Windows Explorer. Select the directory on the USB thumb drive where the Windows Update Agent installer is located. Start the installation by double-clicking on the EXE file. Read and accept the license agreement by pressing the Next Button. Follow the installation wizard to finish the installation. To configure the Windows Update Agent settings select Windows Start Control Panel and then Administrative Tools Services and double-click on Security Center to bring up the settings dialog: 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 28

Windows Patches and Updates Select Automatic as the Startup Type and press Start to start the service: Press OK to finish the configuration. 5.2 Configuring Automatic Updates Windows can be configured to install important updates as they become available by enabling automatic updating. Optional updates are not downloaded or installed automatically. To start the automatic updates go to Windows Start Control Panel Security Center. Open the dialog and select Turn on Automatic Updates. Automatic updates are now activated on the instrument. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 29

Windows Patches and Updates 5.3 Instruments connected to a Windows Update Server Many companies have a Windows update (WSUS) server running on the network. If an instrument is connected to the network, you can configure it to use the WSUS server for Windows updates. Please contact your IT department or system administrator to set up the update configuration of the instrument in compliance with your company policy. You can control or change the WSUS client settings on the instrument via Windows Start Run and then enter gpedit.msc to start the group policy settings. Navigate in the pane to Computer Configuration Administrative Templates Windows Components Windows Updates.Scroll to and double-click on Specify intranet Microsoft update service location to start the settings dialog: First click on Enabled, then specify the server name within the company s network to be used for detecting updates: Note: Make sure that automatic updates are enabled as described in section 5.1. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 30

Windows Patches and Updates 5.4 Configuring Automatic Updates Configuration of automatic updates is very flexible. For example, updates can be scheduled to a specific day and time, notification to the user can be activated, etc. The automatic update settings can be managed via Windows Start Control Panel Automatic Updates: For R&S instruments, Rohde & Schwarz highly recommends the use of the Notify me configuration, where confirmation from the user is required before download and installation. Download of updates and installation can cause performance degradation on the instrument during that time and may require a reboot. Therefore the user should control when the update process runs, so that it does not occur when the instrument is in use. 5.5 Viewing installed Updates Installed updates can be viewed via Windows Start Control Panel Add or Remove Programs: Make sure that the property Show updates is selected in the dialog box. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 31

Related Documents and Links 6 Related Documents and Links NSA Security papers http://www.nsa.gov/ia/guidance/security_configuration_guides/ News about Security threats http://www.securityfocus.com/ Microsoft Windows Update Agent Download Link http://go.microsoft.com/fwlink/?linkid=100334 Microsoft Support: How to disable the Autorun functionality in Windows http://support.microsoft.com/kb/967715/en-us Microsoft Support: Troubleshooting Windows Firewall settings in Windows XP Service Pack 2 for advanced users http://support.microsoft.com/kb/875357/en-us Microsoft, Windows, Windows XP, and Microsoft Security Essentials are U.S. registered trademarks of Microsoft Corporation. Norton and Norton AntiVirus 2010 are U.S. registered trademarks of Symantec Corporation. Kaspersky and Kaspersky Anti-Virus 2010 are U.S. registered trademarks of Kaspersky Lab ZAO. 1EF73_0e Rohde & Schwarz Malware Protection - White Paper 32

About Rohde & Schwarz Rohde & Schwarz is an independent group of companies specializing in electronics. It is a leading supplier of solutions in the fields of test and measurement, broadcasting, radiomonitoring and radiolocation, as well as secure communications. Established 75 years ago, Rohde & Schwarz has a global presence and a dedicated service network in over 70 countries. Company headquarters are in Munich, Germany. Environmental commitment Energy-efficient products Continuous improvement in environmental sustainability ISO 14001-certified environmental management system Regional contact USA & Canada USA: 1-888-TEST-RSA (1-888-837-8772) from outside USA: +1 410 910 7800 CustomerSupport@rohde-schwarz.com East Asia +65 65 13 04 88 CustomerSupport@rohde-schwarz.com Rest of the World +49 89 4129 137 74 CustomerSupport@rohde-schwarz.com This white paper and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde & Schwarz website. R&S is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners. Rohde & Schwarz GmbH & Co. KG Mühldorfstraße 15 D - 81671 München Phone + 49 89 4129-0 Fax + 49 89 4129 13777 www.rohde-schwarz.com