Automatic compression measurement using network analyzers



Similar documents
Power Amplifier Gain Compression Measurements

RF Network Analyzer Basics

Understanding Mixers Terms Defined, and Measuring Performance

Michael Hiebel. Fundamentals of Vector Network Analysis

1. The Slotted Line. ECE 584 Microwave Engineering Laboratory Experiments. Introduction:

Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer

Basic of Load Pull Measurements Active and Passive load pull & Harmonic load pull testbench

0HDVXULQJWKHHOHFWULFDOSHUIRUPDQFH FKDUDFWHULVWLFVRI5),)DQGPLFURZDYHVLJQDO SURFHVVLQJFRPSRQHQWV

Performing Amplifier Measurements with the Vector Network Analyzer ZVB

A Guide to Calibrating Your Spectrum Analyzer

Agilent PN RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note

Understanding Power Splitters

Visual System Simulator White Paper

Impedance 50 (75 connectors via adapters)

Optimizing IP3 and ACPR Measurements

Improving Network Analyzer Measurements of Frequency-translating Devices Application Note

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

Application Note Noise Frequently Asked Questions

Understanding Power Splitters

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

MATRIX TECHNICAL NOTES

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

One Port Network Analyzer

Network analyzer and spectrum analyzer two in one

Field Calibration Software

Making Spectrum Measurements with Rohde & Schwarz Network Analyzers

EE 186 LAB 2 FALL Network Analyzer Fundamentals and Two Tone Linearity

Agilent 8720 Family Microwave Vector Network Analyzers

RF Communication System. EE 172 Systems Group Presentation

Noise Figure Measurement without a Noise Source on a Vector Network Analyzer

Tx/Rx A high-performance FM receiver for audio and digital applicatons

A Low Frequency Adapter for your Vector Network Analyzer (VNA)

RF System Design and Analysis Software Enhances RF Architectural Planning

Network Analyzer Operation

F = S i /N i S o /N o

High Power Amplifier Measurements Using Agilent s Nonlinear Vector Network Analyzer

AN Application Note: FCC Regulations for ISM Band Devices: MHz. FCC Regulations for ISM Band Devices: MHz

Keysight Technologies Understanding the Fundamental Principles of Vector Network Analysis. Application Note

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Experiment 7: Familiarization with the Network Analyzer

UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN )

HF Receiver Testing. Issues & Advances. (also presented at APDXC 2014, Osaka, Japan, November 2014)

Agilent 10 Hints for Making Better Network Analyzer Measurements. Application Note B

Cable Impedance and Structural Return Loss Measurement Methodologies

Agilent AN Improving Throughput in Network Analyzer Applications. Application Note

Introduction to Network Analyzer Measurements

Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range. Application Note

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

R&S FSW signal and spectrum analyzer: best in class now up to 50 GHz

ELEMENTS OF CABLE TELEVISION

MAINTENANCE & ADJUSTMENT

CAS: Network Analysis (Part 2)

PLANAR 304/1 Vector Network Analyzer. Operating Manual

Fast and Accurate Test of Mobile Phone Boards

Scalar Network Analyzer

ELECRAFT KX3 EXTENDED VFO TEMPERATURE COMPENSATION PROCEDURE Copyright 2012 Elecraft LLC Rev. A8, October 27, 2012

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz

Power meter. Caratterizzazione di amplificatori a RF. Amplificatori a RF. Incertezza Corso di Laboratorio di misure ad alta frequenza

Agilent 8753ET/8753ES Network Analyzers

IVIS IVIS - Indoor Video Insertion System

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

R3765/67 CG Network Analyzer

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

Conquering Noise for Accurate RF and Microwave Signal Measurements. Presented by: Ernie Jackson

Broadband Push-Pull Power Amplifier Design at Microwave Frequencies

Understanding the DriveRack PA. The diagram below shows the DriveRack PA in the signal chain of a typical sound system.

Analysis of the DF0MTL 144 MHz Signal

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Cable Analysis and Fault Detection using the Bode 100

SR2000 FREQUENCY MONITOR

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)

Tuning Subwoofers - Calibrating Subwoofers

Agilent FieldFox RF Vector Network Analyzer N9923A Quick Reference Guide

Microwave. 6200B series RF and Microwave Test sets.

Fundamentals of Vector Network Analysis Fundamentals of Vector Network Analysis Primer

A Network Analyzer For Active Components

How To Measure Two Tone, Third Order Intermodulation Distortion

Understanding the Fundamental Principles of Vector Network Analysis. Application Note Table of Contents. Page

Frequency Response of Filters

Achieving New Levels of Channel Density in Downstream Cable Transmitter Systems: RF DACs Deliver Smaller Size and Lower Power Consumption

Agilent Compatibility of the U2000 Series USB Power Sensors with Agilent Instruments. Application Note

W-band vector network analyzer based on an audio lock-in amplifier * Abstract

Calibration Guide. Agilent Technologies ESG Vector Signal Generator

Design of Bidirectional Coupling Circuit for Broadband Power-Line Communications

Group and Phase Delay Measurements with Vector Network Analyzer ZVR

Distance-To-Fault (DTF) Measurement in 75 Ohm Systems with the R&S FSH3-TV

Objectives The purpose of this lab is build and analyze Differential amplifiers based on NPN transistors (or NMOS transistors).

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE

Site Master Cable and Antenna Analyzer with Spectrum Analyzer

ATSv5 AUTOMATED TUNER SYSTEM SOFTWARE MT993 Series Introduction

Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer

Measuring RF Parameters of Networks Bill Leonard NØCU

Transcription:

Automatic compression measurement using network analyzers Introduction The dynamic range of an amplifier is determined by noise figure and compression. In multi carrier applications third order intercept point is also essential. Conventional compression measurements are error prone and tedious. Multichannel and power sweep capability of modern network analyzers can be used to perform automated compression measurements. Power sweep feature also enables compression measurement of any arbitrary value. This is very useful in multistage amplifier design and comparison of amplifiers with similar output powers. One powerful feature of this automated measurements is, that amplifiers can now be tuned for compression by semi-skilled operators. The ease of measurements can enhance the quality and increase production. l db Compression Compression is a measure of the linearity of a device. 1 db compression is normally used as a standard of this measure. For an amplifier it is defined as the output power at which the gain decreases by 1 db from the small signal gain. Fig. 1 Definition of 1 db compression 1 db compression is universally specified by amplifier vendors. This seemingly simple parameter is very tedious to measure and is error prone. Conventional Compression Measurement Conventional compression measurements are performed by switching a 10 db attenuator from input

to output and noting the gain change. The output power at which the gain differs by 1 db is called output power at 1 db compression. Fig. 2 Conventional compression measurement This set up eliminates the need for computing gains, but the input power has to be adjusted till the gains differ by 1 db. This measurement needs few iterations and also involves twice as many connect, disconnect operations. When measurement over a band of frequencies are required, it takes innumerable iterations and connect, disconnect operations and the final results are basically a reflection of operators skill or fatigue. Modern Compression Measurements Thanks to advances in instrumentation, now power sweep is available in modern network analyzers and sweep oscillators. This feature enables compression measurements much more faster and accurate. Fig. 3 Compression measurement using power sweep. As shown in Figure 3, input power to the DUT is swept at a fixed frequency. In the linear range of the device, the output power rises linearly and starts deviating from linearity at compression. It is difficult to determine compression parameters in this mode. If the network analyzer is normalized with a through, then linear gain and compression can be seen clearly, as shown in Figure 4. But the troubles are far from over. One has to determine the output power at 1 db compression. This is normally done by placing a marker at the compression point and returning the network analyzer to absolute power measurement mode and determine the output power.

Fig. 4 Gain compression measurement with power sweep. Normally the output of the DUT is padded to prevent analyzer saturation and damage; in which case the attenuation of the pad has to be added to determine output power at 1 db compression. When compression measurements are needed at several frequencies, this procedure is tedious as normalization has to be done at each frequency which again involves connect and disconnect operations. Automated Compression Measurements Power sweep feature of automatic network analyzers can be utilized in a novel way to accomplish accurate, automated compression measurements. Figure 5 shows the setup for automatic compression measurement. In this set up input power to the DUT is sampled and measured at the same time with the output power. If the coupler is pre calibrated, the input to the DUT can be computed in the computer and this enables gain to be computed as a function of the input power. From the computed gain values 1 db compression can be determined. Fig.5 Automatic compression measurement. Where Do You Get A Coupler It is hard to get a coupler which covers the frequency range of the analyzer. Surprisingly, it is not necessary to look for one. All vector analyzers need a sample of the input signal to accomplish phase lock. This is done using resistive splitters and are part of Bridges or S-parameter set ups and are an integral part of the network analyzer. These bridges/s-parameter set ups cover the frequency range of the instrument and have excellent return loss. The sample output is also used in ratio measurements to eliminate error due to power changes.

By pre-calibrating the coupler over the frequency range of measurement input power to the DUT can be determined by measuring power in the R-channel. Effect of Bad Return Loss of DUT Network analyzer couplers have good directivity. When the device return loss is bad, the reflected signal from the DUT gets sampled due to the finite directivity of the coupler and can result in errors of the order of magnitude of few tenths of a db. This comes as error in gain. To increase the accuracy, an input attenuator can be added between the coupler and DUT. This increases the apparent directivity of the coupler by twice the value of the input attenuator or the return loss of the input attenuator, whichever is less. When they are comparable the vectorial sum of the signals reflected from attenuator and that reflected by the DUT and attenuated by the input attenuator arriving the coupler have to be considered. Normally, 6 to 10 db attenuator will do the job. The draw back of this approach is, it decreases the available drive to the DUT. Safe Attenuator At Output With high drives, network analyzers are driven into compression which can result in measurement errors. With excessive drives the analyzer can be damaged. It is a good practice to add a "safe" attenuator at the output of the DUT to avoid these problems. This, even though it is a good practice, adds additional measurement steps. Attenuators have frequency response. To avoid errors with assumed safe attenuator values it is a good practice to measure the frequency response of safe attenuators and store it in computer. Normally this is done at low powers. Care should be taken to choose an attenuator which does not change attenuation value at high power. Calibration Of Network Analyzer Network analyzers are not basically designed for absolute power measurements. So, they have to be calibrated with power meters. It is a good practice to calibrate both B & R channels to avoid errors. Step By Step Measurement Of Compression Figure 6 shows the complete set up for compression measurement of an amplifier using HP8753A automatic vector network analyzer. Step by step measurement of compression is as follows: 1. Calibrate B & R channels with power meter, store the correction factors in computer. 2. Connect B channel to input of Bridge. Keep network analyzer in frequency sweep. The difference in B & R channel readings gives the correction to be applied to R Channel power to obtain input to DUT. If DUT return loss is bad, have input attenuator while performing these measurements. 3. Connect Safe attenuator after input attenuator and measure its frequency response and store it. 4. Connect DUT as shown in figure 6. Set up instrument for CW frequency and power sweep. Apply power sweep. Read R & B channel powers. Apply correction factors to determine input and output powers as described above. Compute gain. Fine the output power at which gain drops by a db from the lowest power. If I db point is not found increase drive level Repeat measurement. If drive levels at low and at 1 db compression point are less than 10 db decrease lower end of drive level, repeat measurement. 5. Repeat step 4 at each frequency of interest.

All the decisions in steps 4 can be done automatically in the computer. These measurements do not involve repeated connect, disconnect operation and hence are very reliable. Fig. 6 Automatic compression measurement set-up of an amplifier. Uses Of Arbitray Compression Measurements The strength of this procedure is compression point of any arbitrary value can be determined without an excessive number of measurements. This is useful in cascaded amplifier design when the contribution of each stage to the overall compression can be accurately determined. Compression measurement at arbitrary values of compression is also useful for system analysis. One important parameter called figure of merit of compression is defined as the ratio of change of power output to change in compression. For example, if the output power is 32 dbm at 0.5 db compression and 33 dbm at 1.0 db compression, then the figure of merit. Low values of F indicates that unit goes rapidly into compression and high values indicate that it goes slowly into compression. Conclusion A novel adoption of the existing components of the network analyzer and the power sweep feature is used to automate compression measurements. Figure of merit is defined for compression which can be used to compare amplifiers of same output power for a particular application Last Updated: 12/17/1999