Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE



Similar documents
Waters 2695 Separations Module

[ Care and Use Manual ]

Successful HPLC Operation. A Troubleshooting Guide Version 1.1

ChromQuest 5.0 Chromatography Data System

ÄKTA system Training Guide

Agilent 1260 Infinity Binary Pump


23 The Thermal Conductivity Detector

Protocol: HPLC (amino acids)

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

ORGANIC SAMPLE PREPARATION

TECHNICAL DATA. Dry 122a. November 12, 2013

INSTRUCTIONS Edition AC

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Analysis of Liquid Samples on the Agilent GC-MS

Maintaining Your Agilent 1100 Series HPLC System. Optimize performance and maximize efficiency in your lab. Learn essential HPLC maintenance tips

The On-Board Refueling Vapor Recovery (ORVR) Evaporative Emission (EVAP) system.

Thermal Conductivity Detector

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

Standard Hydraulic Piping Flushing Procedure

After the reactor has been running and the air in the chamber has been purged the co2 can be introduced and the dial in process can begin.

Trouble Shooting. Pump

Maintaining Your Agilent LC and LC/MS Systems. Our measure is your success.

Chromatograph Maintenance and Troubleshooting Class # 5050

Turbocharger system components, servicing

Amino Acid Analyzer L-8900

[ Care and Use Manual ]

Cryogenic Vacuum Pumps

Stainless Steel Single and Dual Circulation Kits

FPLC standard operating procedure AKTA system (GE)

The Electron Capture Detector (ECD)

Hydraulic Trouble Shooting

TROUBLESHOOTING FLOW CHART

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47)

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

How To Test For Contamination In Large Volume Water

M V Intake Manifold INSTALLATION INSTRUCTIONS

Thermo Top - Troubleshooting Tree

Powers Controls TH 192 HC Heating/Cooling Room Thermostat

KUSTER ELECTRONIC GAUGES

X-TYPE. Fuel System Drivability/Starting Issues Customer Complaints Diagnostic Procedures

Pump Therapy MADE FOR YOU. Pump User Guide

ESS Method 130.1: General Auto Analyzer Procedures

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

Purification of reaction mixtures using flash chromatography.

Micromass LCT User s Guide

System Saver 318 Air Compressor for Mack E-Tech and ASET Engines

Volkswagen Jetta, Golf, GTI 1999, Liter VR6 2V Engine Mechanical, Engine Code(s): AFP 17 Engine-Lubrication (Page GR-17)

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

CARING FOR YOUR WATER HEATER

Water Priming Mode Purge air from plumbing system. How do I solve ER-3 WATER PRIME: TIPS ON FILLING SPA IMPORTANT NOTE WARNING

IMPORTANT SAFETY RULES TO FOLLOW

Installation Instructions

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Agilent High Sensitivity DNA Kit Guide

ArtisanLink Staining System is an automated special stains slide

Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat

Trace Dissolved Oxygen Sensor

Agilent 1100 Series Well-plate Sampler & Micro Well-plate Sampler. Reference Manual

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Care and use guide. Contents. a handy solution made easy. page

Tuning & Mass Calibration

BACKPACK SPRAYERS. MODEL NOS: KSP16 & KSP20 Part Nos: & OPERATING & MAINTENANCE INSTRUCTIONS GC04/12

DTC P0401, DTC P1491,

Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)

PS3 Peptide Synthesizer QUICK START GUIDE

Operator s Manual. Pressure Injection Cell Model PC8500. Congratulations! Contents

STAINLESS STEEL TANK TABLE OF CONTENTS TROUBLESHOOTING... 1 TESTING WATER TEMPERATURE... 3 TANK & JACKET ASSEMBLY... 4.

Dionex IonPac AS19-4µm

Thermostatic Valve Replacement Cartridge and Rebuild Kit

Coolflo Safe Touch Shower Mixing Valve ST20015CP

ACQUITY UPLC H-Class System Guide

Background Information

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

3.3. NGC Chromatography Systems and ChromLab Software Instrument Guide Version 3.3

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Rheodyne Manual Sample Injector Valves

ENGINE FUEL FUEL FILTER... FUEL HEATER... INJECTOR... SUPPLY PUMP... COMMON RAIL... FUEL PRESSURE LIMITTER...

CDS TROUBLESHOOTING SECTION I. VACUUM Weak vacuum at wand. Gauge reads normal (10hg to 14hg)

POSEIDON 2-29, 2-25 & 2-22 POSEIDON 2-29, 2-25 & 2-22 XT

Determination of Calcium and Magnesium in Brine

2006 HEADSHOK Service Video #1

Diagnosing Hydraulic Problems

Oasis HLB Cartridges and 96-Well Plates

Installation Operation Maintenance Troubleshooting

A.Y. McDonald Mfg. Co. Troubleshooting Submersible and Jet Pumps

Fleck Service Manual INSTALLATION AND START-UP PROCEDURE TABLE OF CONTENTS JOB SPECIFICATION SHEET

FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL

SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE

HPLC Basics, Equipment, and Troubleshooting

E - THEORY/OPERATION

National- Spencer Inc.

ACQUITY UPLC H-Class System Guide

Constantemp Double Wall Low pressure steam-water Heater F-340LDW,F-640LDW, F-940LDW and F-1240LDW

B150 Series Commercial Brewer. Direct Plumb Kit

Installation Instructions for Water Methanol Stage 1

PC1130 Electric Air Compressor

REPAIR INSTRUCTIONS UNDER BODY CYLINDER (UM, UL)

Dissolved Oxygen Sensors

Transcription:

Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2

2690/5 Theory 2690/5 Solvent Management Flow Path 3

2690/5 Theory 2690/5 Solvent Management Delivery Theory The solvent management system components work together to deliver solvent in the following sequence. 1. The accumulator plunger (left plunger) moves forward delivering solvent, while the primary plunger (right plunger) moves backwards filling with solvent. 2. While the accumulator continues to deliver solvent, the primary plunger waits. The primary pressure transducer then measures the pressure and matches it to the system pressure by moving the primary plunger forward, pre-compressing the solvent. 3. When the accumulator plunger reaches the end of its stroke, it slows down and reverses direction. Simultaneously, the primary plunger moves forward delivering solvent. 4. While the accumulator plunger moves backwards, the primary plunger moves forward slightly faster which allows the primary plunger stroke to both fill the accumulator piston chamber and to deliver solvent to the rest of the system This sequence enables the solvent management system to maintain a consistent and pulse-free manner. 4

2690/5 Theory 2690/5 Sample Management System Theory 2690/5 Injection Cycle The injection cycle is completed in three steps. 1. Isolation of the loop. 2. Sample withdrawal. 3. Sample injection. Valve 1 closes. Needle wash pump turns on. Valve 4 opens. Needle moves into the bottom seal. Valve 3,2 open venting the loop to atmospheric pressure. Valve 3 closes Needle wash pump turns off. Valve 4 closes. Needle moves into the vial. Syringe moves withdrawing sample. Needle moves to the bottom seal. Valve 3 opens. Valve 2 closes. Needle moves into the stream. Valve 1 opens. Sample is injected. Syringe moves home. Valve 3 closes 5

2690/5 Theory 2690/5 Sample Management System Purge & Compression Check The purge cycle consists of three parts. 1. Purge: Fills the sample loop with new composition of mobile phase replacing the previous mobile phase and any air bubbles. 2. Flush syringe: Fills the syringe with new composition of mobile phase replacing the previous mobile phase and any air bubbles. 3. Compression check: The mobile phase in the injector assembly is compressed to check for leaks, the presence of air bubbles and undegassed mobile phase. 6

SETUP PROCEDURES Priming a New or Dry system using undegassed solvents 1. Fill the solvent bottles with Mobile Phase. 2. From the Status Page turn the degasser on. * 3. Perform a dry prime until solvent flows from each of the four lines. 4. Starting with the organic solvent, perform a 3 minute Wet Prime on each line separately at a flow rate of 7.5 ml/min ** 5. Once all lines are primed, set the flow rate to 0mL/min with the degasser on for 2 minutes. 6. Start the flow with the initial conditions of your chromatographic method. Equilibrate with 10 column volumes of solvent. 7. Check the Pressure Ripple on page two of the Status Page. The ripple should be less than 2% of the system pressure. (For system pressures above 1000psi, repeat Wet Prime if the ripple is too high) 8. After the column is equilibrated, Purge the autosampler. (initial conditions) 9. Run samples. Priming an idle system using existing mobile phase 1. From the Status Page turn the degasser on. * 2. For each line that will be used, starting with the organic solvent, perform a 1.0 minute Wet Prime at a flow rate of 7.5 ml/min. ** 3. Once all lines are primed, set flow rate to 0mL/min with degasser on for 2 minutes. 4. Start the flow with the initial conditions of your chromatographic method. Equilibrate with 10 column volumes of solvent. 5. Check the Pressure Ripple on page two of the Status Page. The ripple should be less than 2% of the system pressure. (for system pressures above 1000psi, repeat Wet Prime if the ripple is too high) 6. After the column is equilibrated, Purge the autosampler. (initial conditions) 7. Run Samples. 7

SETUP PROCEDURES Priming an idle system changing the mobile phase 1. Fill all of the solvent bottles with mobile phase. 2. For each line that will be used, starting with the organic solvent, perform a 3 minute Wet Prime at a flow rate of 7.5 ml/min. ** 3. Once all of the lines are primed, set the flow rate to 0 ml/min with the degasser on for 2 minutes. 4. Start the flow with the initial conditions of your chromatographic method. Equilibrate with 10 column volumes of solvent. 5. Check the Pressure Ripple on page two of the Status Page. The ripple should be less than 2% of the system pressure. (for system pressures above 1000psi, repeat wet prime if the pressure is too high) 6. After the column is equilibrated, Purge the autosampler. (initial conditions) 7. Run Samples. Replacing the Needle Wash Frit (as needed or 6 months) Remove all carousels and power off the 2690/5. Reach into the sample compartment and unscrew the lower frit retainer which located under the injector assembly. Remove the frit with a small diameter pick and discard. Wet the replacement frit with methanol and seat it completely within the retainer. Reinstall the lower retainer using a clockwise motion. Screw the retainer in by hand until it is firmly seated. * The degasser should never be left on with no flow for long periods of time. (i.e. overnight) ** Caution: To avoid precipitating salts in the Separations Module use an intermediate solvent such as water when you change from buffers to high-organic content solvents. The volume of each solvent line including the vacuum degasser chamber is 7.5 ml. This volume must be considered when priming the system and changing to a different mobile phase. The degasser can be turned off, the flow rate stopped and the lamp turned off (if controlling the detector through the Alliance) in a Shutdown Method. 8

SETUP PROCEDURES Priming the Seal Wash Pump (when needed) 1. Press the Diag screen key in the Main screen. The Diagnostic screen appears. 2. Attach the syringe adapter (from the Startup Kit) to the syringe. 3. Remove the solvent filter from the plunger seal wash inlet line or remove the plunger seal wash waste line from the waste container. (Either pushing solvent through it or drawing solvent through it can be used to prime the plunger seal wash pump.) 4. Fill the syringe with the plunger seal wash solvent* and attach the syringe adapter to the end of the plunger seal wash line or attach the syringe to the end of the plunger seal wash waste line. 5. Press the Prime Sealwash Screen key and then press the Start screen key. The plunger seal wash procedure begins. Push on the syringe to push seal wash solution through the system or draw on the syringe to pull solvent through the plunger seal wash pump. 6. When solvent flows out of the plunger seal wash line or into the syringe (the pump will run quieter when it is primed) press the Halt screen key. 7. Reinstall the solvent filter and place the seal wash inlet line back into the plunger seal wash reservoir or place the plunger seal wash waste line back into the waste container. To finish the priming push the Start screen key again. 8. Press Close to return to the diagnostics screen. * Choose a Seal Wash solution that has high solubility with salts and/or buffers used in the mobile phase. (Usually a solution of 80-90% aqueous, 20-10% methanol.) Priming the Needle Wash Pump (daily or weekly) 1. Make sure Needle Wash line (Green Line) is in its appropriate bottle. 2. Press the Diag screen key in the Main screen. The Diagnostics screen appears. 3. Press the Prime NdlWsh screen key. The 30-second needle wash procedure begins. If solvent does not flow out of the waste line (the yellow line) press the Start Again screen key. * Choose a Needle Wash solution that your sample is soluble in. Buffered solutions or a mobile phase modifier such as TFA are not recommended. If these solutions are necessary they must be flushed from the instrument. 9

TROUBLE SHOOTING THE 2690/5 2690/5 System Error Messages MESSAGE ACTION STEPS Accumulator over pressure Locate restriction... guard column / column / inline filter / blocked tubing. Bubble found on intake Carousel malfunction Carousel Z axis malfunction Compression check failed Degasser pressure fault Lost prime Missing vial Needle malfunction Plunger homing fault Primary transducer range (volts) Sample Wet prime and allow system to sit at zero flow with the degasser in continuous mode for approx. 10 min. prior to starting the run. If the flow rate is greater than 1 ml/min. install extended flow inlet tube. Cycle power. If message persists call Waters Service Representative. Cycle power. If message persists call Waters Service Representative. Solvent not degassed properly. Check for bubble on syringe plunger. Test seal pack and valve 3. Perform appropriate maintenance. Replace faulty degasser chamber or vacuum pump. Call Waters Service Representative. Solvent not degassed properly. Change check valves. If the flow rate is greater than 1 ml/min. install extended flow inlet tube. The missing vial should be placed in the correct location in the tray. Disable the vial sensor [Main>>Config>>Highlight] Verify vial presence" press clear. Call Waters Service Representative. Check needle wash reservoir. Prime needle wash. Cycle power. If message persists call Waters Service Representative. Cycle power. Change inline filter. If message persists call Waters Service Representative. Call Waters Service Representative. 10

TROUBLE SHOOTING THE 2690/5 2690/5 System Error Messages MESSAGE Solvent delivery h/w fault Solvent delivery motor lost sync Syringe Movement malfunction ACTION STEPS Cycle power. If message persists call Waters Service Representative. Cycle power. Change inline filter. If message persists call Waters Service Representative. Cycle power. If message persists call Waters Service Representative. System over pressure Locate restriction... guard column / column / inline filter / blocked tubing. System under pressure Connect column. Locate leak. 50 psi is needed to run the system. Chromatographic Errors SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTI0N Erratic retention times Air bubble Wet prime the unit. Degas all solvents. Malfunctioning check valves Solvent leak Chemistry Clogged solvent filters Temperature Clean/ replace check valves. Check all fittings. Replace seals. Check mobile phase, column, Gradient Proportioning Valve. Replace filters. Use column oven. Increased retention times Incorrect flow rate Malfunctioning check valve. Solvent leak. Incorrect Mobile Phase Column Contaminated. Change solvent. Clean/ replace the column. 11

TROUBLE SHOOTING THE 2690/5 Chromatographic Errors SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTI0N Reduced retention times Incorrect flow rate Change flow rate. Incorrect solvent composition High column temp. Column contamination Solvent not degassed Change composition. Check gradient proportioning valve. Change column temp. Clean/ replace column. Degas solvent. Reproducibility errors Solvent not degassed Degas solvent. Air in syringe Incompatible vial septa Sample management system Purge solvent management system. Switch to approved vial and septa. Contact Waters Service Representative. Baseline drift Column not equilibrated Equilibrate column. Detector not equilibrated Solvent contaminated Solvent not properly degassed Flow/ Pressure fluctuations Temperature fluctuations Leaking flow cell Dirty flow cell Allow detector to warm up. Time will vary with wavelength and sensitivity. Use fresh solvent. Degas solvent. Prime system. Replace check valves, pump seals. Stabilize operating environment. Repair leak. Clean flow cell. Baseline noise Flow fluctuations Malfunctioning check valve. Leaks, Seals, fittings. Radio frequency noise Air in detector Solvent not degassed Eliminate interference. Purge detector. Degas solvent. 12

TROUBLE SHOOTING THE 2690/5 Chromatographic Errors SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTI0N Baseline noise Solvents contaminated Use fresh solvents. Column contaminated Dirty flow cell Electronic Clean/ replace column. Clean flow cell. Call Waters Service Representative. Sensitivity loss Solvent leak Check all fittings Degraded, contaminated, or improperly prepared sample. Column contaminated Loss of column efficiency Leak in flow cell Change in mobile phase Contaminated solvents Dirty flow cell Old Lamp Use fresh sample. Clean/ replace column. Clean/ replace column. Repair leak. Correct mobile phase. (i.e. ph or ionic composition) Use fresh mobile phase. Clean flow cell. Replace lamp Split peaks Needle wash solvent Prime needle wash solvent. Solvent not appropriate for samples. Column Clean/ Replace column. 13

User maintenance of the 2690/5 The power on the 2690/5 should be cycled once a week to perform self-diagnostic and calibration routine. Replacing an Inlet Check Valve Cartridge 1. From the Menu screen select [ Diagnostics > Other Tests>Turn GPV off]. 2. Remove the compression fitting from the check valve housing using a 5/16 wrench. 3. Using a ½ wrench to remove the check valve housing from the manifold. 4. Tip the check valve housing upside down to remove the old check valve cartridge. 5. Insert the new check valve cartridge into the check valve housing with the arrow or the two holes facing up. 6. Using a ½ wrench to reinstall the check valve housing to the manifold. (do not over-tighten ) 7. Using a 5/16 wrench reconnect the compression tubing to the check valve housing. (to prevent over-tightening of the check valve housing, use a ½ wrench to prevent the housing from moving) 8. Exit out of diagnostics. 9. Prime and degas the system. NOTE: Check valve cartridges may be cleaned. Sonicate check valve cartridges in water for 15 minutes and in an appropriate organic solvent for an additional 15 minutes. 14

User maintenance of the 2690/5 Replacing the In-Line Filter 1. Use a 5/8 wrench and a 5/16 wrench to separate the compression screw on the left side of the inline filter from the in-line filter inlet housing. 2. Use a 5/8 wrench to hold the filter outlet housing while you loosen the inlet housing with another 5/8 wrench. 3. Tip the housing upside down to remove the in-line filter element. 4. Insert the replacement in-line filter element. 5. Reconnect the in-line filter inlet and outlet housing using two 5/8 wrenches. 6. Reconnect the compression screw into the inlet housing. 7. Inspect connections for leaks. In-line Filter ATTENTION: The in-line filter insert should be replaced every 6 months or whenever a significant increase in back pressure is seen on the system transducer. 15

User maintenance of the 2690/5 Replacing Plunger Seals and Face Seals Plunger Seal Replacement 1. Press the Diagnostic screen from the Main screen. 2. Press the Other Test screen key in the Diagnostics screen. 3. Select Head Removal and Replacement from the list. 4. Follow the directions that appear in the Head Removal and Replacement screen. 5. Remove the plunger seal using the Seal removal tool (P/N WAT039803). 6. Wet the seals and the seal opening with methanol. 7. Insert the plunger seal using the seal insertion tool (P/N WAT270969). Face Seal Replacement 1. Follow directions 1-4 from the plunger seal replacement directions. 2. Remove face seals using plastic tweezers. 3. Wet the seals and the seal opening with methanol. 4. Replace the face seals in the seal openings using a clean flat object. 5. Follow the directions that appear in the Head Removal and Replacement screen. 6. Prime the system and allow 10 minutes to degas the mobile phase before running the system. NOTE: Plunger seals and face seals should be replaced at the same time. ATTENTION: To avoid damage to the sealing surfaces do not use sharp tools. 16

2690/5 SPARE PARTS Description Part Number 2500 µl Syringe WAT077342 250 µl Syringe WAT073109 25 µl Syringe WAT077343 Plunger Seals (kit) Face Seals (kit) Check Valve Cartridge (2) High Flow Inlet Tubing WAT270938 WAT270939 WAT270941 WAT270968 2 ml Loop WAT096106 200 µl Loop WAT045584 Vials with Preslit Septa 186000307 Total Recovery Vials with Preslit Septa 186000385 In-Line Filter Insert WAT088084 Needle Wash Frit (5) 700001318 PerformancePLUS Needle 700001326 Optional Solvent Tray 205000329 Alliance Interactive System Support CD 715000116 17