1 Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated special stains slide processor that can be used as a stand-alone system or operated in a networked configuration using a local area network (LAN) or a Laboratory Information System (LIS). The system uses special stains kits and accessory reagents packaged in patented cartridges that provide precision dispensing of reagents for optimal staining quality. ArtisanLink utilizes a patented reaction chamber and can handle multiple tasks in a laboratory in a single run. With a broad menu of special stains, the system can optimize laboratory workflow through automation dramatically improving productivity. This article provides tips on hardware and software usage as well as some recommendations regarding the use of slides and reagents to get the most out of the ArtisanLink Staining System 1. of each session, a waste valve rinse should be performed to prevent contamination of the tubing and bottles. The user should inspect the inner orange waste gasket when replacing the cap to check that this has not been torn or ripped 2 (Fig.1). Hardware Tips The following is a four-point inspection to keep ArtisanLink running smoothly: Waste management: The ArtisanLink Staining System utilizes a decrement system to track waste accumulation and capacity requirements. A special aspiration system allows the instrument to separate waste into four categories: metals, alcohol, combined dyes and water solubles. Depending on the laboratory workload volume, 2 L or 4 L bottles can be configured in the software. Upon completion 1 This article provides tips for experienced users. 2 A torn or ripped orange waste gasket may be the result of excessive torque applied when screwing/unscrewing the cap. Figure 1. Waste bottle showing the orange waste gasket. 222 Connection 2010

2 Keeping software and bottle inventory evenly matched at all times will prevent overflow and consequential aborted runs. Each time a bottle is removed, perform the Rinse valves (Fig. 2) from the ArtisanLink bulk fluid screen to enable optimal aspiration. Bulk fluid maintenance: Bulk liquids consist of liquids used in a staining run for rinsing or diluting. Dako Wash Solution, 100% ethanol and 95% ethanol, is required for use on the ArtisanLink Staining System. Dako Wash Solution requires diluting and mixing prior to use. For bulk fluid maintenance, rotate the six bulk positions with adjacent bottles. When the alcohol bottle is rotated through each valve position routinely, it reduces the overall manual maintenance that will be needed outside slide sessions. Always unsnap the bottle tubing from the valve before unscrewing the cap to alleviate torque and pressure on the tubing connectors. When the caps are off, inspect the black o-ring inside the cap and return the cap with the notch to align and secure the cap tightly (Fig. 3). This will provide an even pressure and remove waste fluid during staining. Be sure to match software to bottle inventory. After reloading the bottle, prime three or more times to return the fluid into the tubes or lines. Figure 2. Fluid screen showing how to access the Rinse valves button referring specifically to waste valves. Connection

3 Figure 3. The inside of a bulk bottle cap showing the black o-ring. The red circle denotes the groove that should be aligned to the bottle neck notch. Reagent handling: At the end of each staining session, it is recommended to remove all reagent packs from the reagent carousel. Removing hazardous chemicals routinely will help maintain the life of the instrument. Slide platform maintenance: To ensure proper contact between the slides and the slide heater plates, the user should wipe the back of each slide with a dry gauze pad before placing it on the slide carousel. This will not only ensure a good slide clip seal, but will also optimize heating performance. Likewise, wiping residue from the slide platform upon removal of the slide and clip will prevent chemical erosion. For the same reasons, it is also recommended to regularly clean the spill tray and drip ring. Software Tips The DakoLink software allows review of workload information based on customized settings. Additionally, the software can track information previously tracked manually by the histotechnologist. It can measure and separate waste, track fluid inventory, maintain history of fluid lot numbers and expirations, provide customized audio alerts, perform automatic maintenance and, most importantly, identify slides. The making the hardware and software work together along with other parameters such as using the right slides, reagents/reagent packs and chemistry is the first step toward mastering the automated staining process. handheld 2D barcode scanner can retrieve information about slides and reagents utilized during any given session. The reports generated can be exported while the stainer is still in use. Upon installation, the following suggestions may be helpful in maximizing the use of the DakoLink software: Customize user privileges allowing login access to inventory, report fields, system settings and procedure customizations. These are found under Administration/Configuration/USERS and GROUPS (Fig. 5). The ArtisanLink software is part of the DakoLink software family. 224 Connection 2010

4 Set SYSTEM SETTINGS to Automatically perform waste valve rinse after runs and Require confirmation before run begins. This will automatically start the run and perform routine maintenance (Fig. 4). Customize the NEW SLIDES tab to fit the patient information that is pertinent to your Quality Control (QC) program. Slide accessioning will be easier, and a simple 2D scan of the slide label (see Fig. 6) will yield valuable information. Set up slide labels to print customizable text. Maximize the text by utilizing Slide notes as a field option or placing Hospital institution as Case fields. Customizing slide labels using the Case fields option will ease slide accessioning and ultimately save time in the QC process that follows. Populate the bulk fluid lot numbers in the fluid inventory screen to ensure QC history reports are easily obtained upon demand. The assignment of bulk liquids automatically sets a 30-day reminder to rotate the fluids. The fluids are shown in red font when they have exceeded the 30-day reminder. Tips on Slides and Reagents Slides The type of glass slide used and how it is clipped into position can affect the staining procedure. Some tips to ensure a consistent staining result are: Avoid using slides with etched or painted surfaces. The uneven surface can prevent a consistent horizontal spreading of the fluid during sessions. Only use chemically resistant slide labels, e.g., Dako s S3393 (DL213), 0.75 x 0.75 mm labels. Larger labels will prevent the gasket from sealing tightly and as a consequence may leak fluid during the staining session. Slides, whether charged or uncharged, can be used for automation UNLESS mirroring occurs (see Reference). In this case, lowering the heat exposure to the slide or switching from ionically-coated slides to uncharged slides will reduce the phenomenon. While an instrument is staining, the user can set up the next session. Retrieve live updates on the reagent inventory utilizing the ADMINISTRATION report features. Multitasking by pulling out colder reagents needed for the next run can save valuable setup time. Use clean slides: Wipe the back of the slide before placing it into position. This will not only ensure a good slide clip seal, but will also optimize heating performance. Make sure the slide is completely flat to seal the gasket firmly into place. This will ensure an even staining. Fill the slide clip with 2-3 ml of Dako Wash Solution (AR102) and presoak the tissue for a minimum of 5 minutes. This will condition the slide surface to aid in fluid spreading during the session. Store the slide clips in a clean and dry place to ensure a well-sealed gasket every time. Figure 4. SYSTEM SETTINGS screen depicting automatic maintenance settings upon installation (default status). Connection

5 Figure 5. SYSTEM SETTINGS screen depicting possible privilege settings for User groups. 226 Connection 2010

6 Reagents Every reagent pack specifies a storage temperature on the label. Each reagent kit comes with a package insert that includes the specifications for use. These include priming instructions, recommended control tissues, and intended use. Following the instructions given in the package insert is important for a good and consistent staining of the tissue. In other words, the staining quality of the last slide should be as good as the first one. Some tips for using reagents/reagent packs are: Reagent packs stored in a cold room should be brought to ambient room temperature before priming for staining use. This will usually take minutes. If packs are cold at session startup, the chemicals continue to reach ambient temperature throughout the session. This can yield a varying temperature dispense from first to last slide. Reagent packs should not be shaken, but gently mixed prior to each session by inverting the pack several times. Some chemicals foam or release gases during motion which can damage plastic parts causing erroneous dispense behaviors. Figure 6. A 2D slide label demonstrating customized information for quick identification. Each reagent pack should be primed until a steady stream is obtained. A good technique is to point the tip away from you into an approved waste container. This is done by holding the pack at a slight angle, tapping the side gently and priming with a plunge of the dispenser. Priming ensures that no air is left in the bottle neck during the dispense. NOTE: A steady prime will equal 250 µl. Dako provides extra reagent for each test pack to accommodate priming. Reagent packs should be placed on the carousel immediately after priming while maintaining them in a tip-down orientation. This will ensure a first full dispense and that no air is left. Reagent packs should be stored at their labeled temperature in a tip-down orientation, using Dako provided reagent storage trays, AR409. It is advised not to leave reagent packs on the carousel when the reagents are not in use. Proper storage will maintain manufacture label shelf life. Refrain from writing on the spine of the reagent packet labels as this may interfere with the barcode scanning. If the pack will not scan, the session will not begin. NOTE: The handheld 2D barcode scanner can be used at any time to verify inventory. In summary, making the hardware and software work together along with other parameters such as using the right slides, reagents/reagent packs and chemistry is the first step toward mastering the automated staining process. With the tips described in this article, you should be able to obtain consistent quality staining results without making extensive changes to the system or using additional time in the laboratory. Reference Brown RW (Ed.). Histologic Preparations: Common Problems and Their Solutions. College of American Pathologists, Northfield IL, 2009, pp Connection

Basic User Guide. Document Number D09665 Revision 01 July, 2012. For In Vitro Diagnostic Use

Basic User Guide. Document Number D09665 Revision 01 July, 2012. For In Vitro Diagnostic Use Basic User Guide Document Number D09665 Revision 01 July, 2012 For In Vitro Diagnostic Use Copyright 2012 Dako North America, Inc. All rights reserved This document may not be copied in whole or in part

More information

Quick Start Guide See Inside for Use and Safety Information

Quick Start Guide See Inside for Use and Safety Information 3 rd Generation Personal 3D Printer Quick Start Guide See Inside for Use and Safety Information The USB Mass Storage Device Contains the User Guide and Quick Start Guide in other Languages Congratulations

More information

Instructions for Use

Instructions for Use 1 Instructions for Use HUMULIN R U-500 KwikPen insulin human injection U-500 (500 units/ml, 3 ml pen) Important: Know your dose of HUMULIN R U-500 insulin. The Pen delivers your dose in insulin units.

More information

KwikPen Insulin delivery device

KwikPen Insulin delivery device USER MANUAL KwikPen Insulin delivery device PLEASE READ THESE INSTRUCTIONS BEFORE USE Lilly Introduction The KwikPen is designed for ease of use. It is a disposable pen containing 3 ml (300 units) of U-

More information

Disposable Insulin Delivery Device User Manual

Disposable Insulin Delivery Device User Manual PV 3734 AMP Disposable Insulin Delivery Device User Manual Instructions for Use Read and follow all of these instructions carefully. If you do not follow these instructions completely, you may get too

More information

Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin)

Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin) 1 Instructions for Use HUMALOG KwikPen insulin lispro injection (rdna origin) Read the Instructions for Use before you start taking HUMALOG and each time you get a refill. There may be new information.

More information

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance Step 1 Action Perform the Manually Clean the ISE Mix Bar, Liquid Level Sensors, Sample Pot and Sample Pot Tubing procedure AU680 Weekly (Analyzer

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Milwaukee USER MANUAL

Milwaukee USER MANUAL Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the

More information

Coating and Extraction of Honeycomb Denuders

Coating and Extraction of Honeycomb Denuders QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and

More information

Information for patients and nurses

Information for patients and nurses Information for patients and nurses Rocket IPC Pleural Catheter Indwelling Catheter Rocket Indwelling Pleural Catheter (IPC) Contents Contact Information...03 What s in the Rocket Dressing Pack and Bottle

More information

Private Water Supplies Sampling Manual. A Field Guide

Private Water Supplies Sampling Manual. A Field Guide Private Water Supplies Sampling Manual A Field Guide Foreword This sampling manual details standard procedures for the collection, storage and transportation of samples at private water supplies in accordance

More information

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01 Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery

More information

PA 9117 FSAMP. Lilly. Prefilled Insulin Delivery Device User Manual

PA 9117 FSAMP. Lilly. Prefilled Insulin Delivery Device User Manual 1 Lilly PA 9117 FSAMP Prefilled Insulin Delivery Device User Manual Instructions for Use Read and follow all of these instructions carefully. If you do not follow these instructions completely, you may

More information

AxyPrep TM Mag PCR Clean-up Protocol

AxyPrep TM Mag PCR Clean-up Protocol AxyPrep TM Mag PCR Clean-up Protocol Intro The AxyPrep Mag PCR Clean-up kit utilizes a unique paramagnetic bead technology for rapid, high-throughput purification of PCR amplicons. Using this kit, PCR

More information

Protocol Micro Lab. K:\Microlab_protocols\Protocols_active\03 Instrument manuals\ml03_001_001 DGGE.doc. Preparation of gels

Protocol Micro Lab. K:\Microlab_protocols\Protocols_active\03 Instrument manuals\ml03_001_001 DGGE.doc. Preparation of gels Preparation of gels Cautions: Wear gloves all times when handling acryl amide and form amide solutions and all work with acryl amide and form amide has to be done in a flow hood. Acryl amide is very toxic.

More information

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com.

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. 2 0 0 5 D U K E A I R X C / S L / R A C E S E R V I C E G U I D E Information

More information

1-888-448-7250 GENERAL CARE PLUMBING & VALVE INFORMATION TUB & SHOWER DOOR INFORMATION FOR QUESTIONS CALL OUR SERVICE DEPARTMENT

1-888-448-7250 GENERAL CARE PLUMBING & VALVE INFORMATION TUB & SHOWER DOOR INFORMATION FOR QUESTIONS CALL OUR SERVICE DEPARTMENT GENERAL CARE PLUMBING & VALVE INFORMATION TUB & SHOWER DOOR INFORMATION FOR QUESTIONS CALL OUR SERVICE DEPARTMENT 1-888-448-7250 or email service@bathsaver.com caring for your new bath fitter unit Thank

More information

HUMULIN 70/30 KwikPen

HUMULIN 70/30 KwikPen 1 Instructions for Use HUMULIN 70/30 KwikPen (70% human insulin isophane suspension 30% human insulin injection [rdna origin]) Read the Instructions for Use before you start taking HUMULIN 70/30 and each

More information

OxySense Gen III Operations Manual

OxySense Gen III Operations Manual OxySense Gen III Operations Manual OxySense GEN III 300 Series OxySense GEN III 5000 Series COMPLIES WITH ASTM 2714-08 OxySense, Inc. 13111 North Central Expressway, Suite 440 Dallas, Texas 75243 Phone

More information

PA 9114 FSAMP. Lilly Disposable Insulin Delivery Device User Manual

PA 9114 FSAMP. Lilly Disposable Insulin Delivery Device User Manual 1 Lilly Disposable Insulin Delivery Device User Manual PA 9114 FSAMP Instructions for Use Read and follow all of these instructions carefully. If you do not follow these instructions completely, you may

More information

not appearance. ADHESIVE SYSTEMS CORIAN As an integral part of its sheet and shape products, DuPont has developed a special adhesive.

not appearance. ADHESIVE SYSTEMS CORIAN As an integral part of its sheet and shape products, DuPont has developed a special adhesive. CORIAN JOINT ADHESIVE SYSTEM 8.1 CORIAN JOINT ADHESIVE SYSTEM As an integral part of its sheet and shape products, DuPont has developed a special adhesive. The DuPont Corian covers seam performance, not

More information

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE 4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement

More information

Western Blotting For Protein Analysis

Western Blotting For Protein Analysis Western Blotting For Protein Analysis Part 1: Laemmli Gel Electrophoresis Using Mini-PROTEAN II Electrophoresis Cell Note: Powder-free gloves should be worn throughout the entire procedure. A. Preparing

More information

Make sure oven is OFF. Never remove parts or touch the fan unless the oven is OFF.

Make sure oven is OFF. Never remove parts or touch the fan unless the oven is OFF. Morning Cleaning The following start-up procedures MUST be completed before using the oven. The procedures may be completed at night after the oven has completed the clean cycle, or in the morning BEFORE

More information

Triac Printed Circuit Board Replacement

Triac Printed Circuit Board Replacement Technical Service Bulletin: Triac Printed Circuit Board Replacement TRONIC 5000C Pro Models: WH17, WH27, WH36 Introduction Fig. 1 ELECTRICITY IS EXTREMELY DANGEROUS. TAKE EXTRA PRECAUTIONS AND ENSURE ALL

More information

2003-2004 SID REAR SERVICE GUIDE

2003-2004 SID REAR SERVICE GUIDE 2003-2004 SID REAR SERVICE GUIDE For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. Contact your local distributor or visit

More information

Infectious Waste Management Plan

Infectious Waste Management Plan Infectious Waste Management Plan Infectious Waste Management Plan USC Health & Safety Programs Unit 777-5269 POLICY: A. In keeping with the University of South Carolina's policy of providing protection

More information

Fixed and Adjustable Spacing Electronic Pipettes

Fixed and Adjustable Spacing Electronic Pipettes viaflo Fixed and Adjustable Spacing Electronic Pipettes viaflo Electronic Pipettes All viaflo pipettes function as repeaters, mixers, serial dilutors, manual pipettes, handheld sample processors and more!

More information

NIH Clinical Center Patient Education Materials Giving a subcutaneous injection

NIH Clinical Center Patient Education Materials Giving a subcutaneous injection NIH Clinical Center Patient Education Materials What is a subcutaenous injection? A subcutaneous injection is given in the fatty layer of tissue just under the skin. A subcutaneous injection into the fatty

More information

Using the Surface Zeta Potential Cell Kit

Using the Surface Zeta Potential Cell Kit Using the Surface Zeta Potential Cell Kit Note: Malvern has a video that goes through all of these steps in detail. There is a link on the NBTC Zetasizer information page and it is bookmarked on the Zetasizer

More information

Agilent High Sensitivity DNA Kit Guide

Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Agilent Technologies Notices Agilent Technologies, Inc. 2009, 2013 No part of this manual may be reproduced in any form or by any means

More information

Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless

More information

Conservation of Momentum Greg Kifer

Conservation of Momentum Greg Kifer SCIENCE EXPERIMENTS ON FILE Revised Edition 6.7-1 Conservation of Momentum Greg Kifer Topic Conservation of momentum Time 1 hour! Safety Please click on the safety icon to view the safety precautions.

More information

Tablet PC Tutorial for Windows 7

Tablet PC Tutorial for Windows 7 VIRGINIA TECH Tablet PC Tutorial for Windows 7 Getting Started Guide Instructional Technology Team, College of Engineering Last Updated: Fall 2013 Email tabletteam@vt.edu if you need additional assistance

More information

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient

More information

INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section:

INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section: HOSPITAL NAME INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section: TITLE/DESCRIPTION POLICY NUMBER HANDLING OF H AZARDOUS DRUGS EFFECTIVE DATE REVIEW DUE REPLACES NUMBER NO. OF PAGES APPROVED

More information

EXPRESS STILL WARNING It is legal in New Zealand to own and operate a still for the purpose of producing alcohol for your own consumption.

EXPRESS STILL WARNING It is legal in New Zealand to own and operate a still for the purpose of producing alcohol for your own consumption. THE EXPRESS STILL This still represents the next generation in home distillation equipment. It eliminates the problem of having to spend all day processing a 25 litre wash. Taking 1 hour to heat up, then

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

2 (. 3 4 5 . -,.. -,. 3. : 1 2 3 > 4, :!! 1,, 17 Lt MAX 2 3 4,, > 5.,,, 6 :, :1)! 2)! 7,, 1, :,. 2,, Bottom of steam generator 7 8 9 10 11 12 automatically and this phenomenon will continue for

More information

Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)

Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models) TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models) WARNING - HIGH VOLTAGE AC electrical circuits are connected to this heater. Do not attempt any service work on the

More information

COMMON LABORATORY APPARATUS

COMMON LABORATORY APPARATUS COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen

More information

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3

More information

Support Cleaning Apparatus. Operation, Cleaning, & Safety Manual

Support Cleaning Apparatus. Operation, Cleaning, & Safety Manual Support Cleaning Apparatus Operation, Cleaning, & Safety Manual Revision 1.1 31 May 2014 WARNING: The instructions and precautionary statements contained herein are intended for the Support Cleaning Apparatus

More information

LBP-1210 Laser Printer

LBP-1210 Laser Printer LBP-1210 Laser Printer Getting Started Guide IMPORTANT: Read this manual carefully before using your printer. Save this manual for future reference. Copyright Copyright 2002 Canon Inc. All information

More information

2006 JUDY SERVICE GUIDE

2006 JUDY SERVICE GUIDE 2006 JUDY SERVICE GUIDE For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. Information contained in this publication is subject

More information

Achieving Independence

Achieving Independence Bard: Intermittent Self-Catheterization A Guide to Self-Catheterization Achieving Independence Introduction This brochure is provided by Bard, a leading provider of urology products since 1907. The best

More information

Performance Qualification of an Automated Plate Washer Using the Artel MVS

Performance Qualification of an Automated Plate Washer Using the Artel MVS Performance Qualification of an Automated Plate Washer Using the Artel MVS Kevin Khovananth, Tanya Knaide, Keith Albert ARTEL Abstract The purpose of this application note is to demonstrate the ability

More information

4RU High Density Shelf Installation Instructions

4RU High Density Shelf Installation Instructions 4RU High Density Shelf Installation Instructions Table of Contents General Product Information... 1.0 Safety Precautions... 2.0 Tools Required... 3.0 Package Contents... 4.0 Installing the Product Unpacking...

More information

ph Measurements of Common Substances

ph Measurements of Common Substances Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the

More information

What You Will Learn in This Module: Testing for Pullorum-Typhoid Disease

What You Will Learn in This Module: Testing for Pullorum-Typhoid Disease Module 5-B: Testing Poultry for Pullorum-Typhoid Disease This module applies only to individuals who have attended and successfully completed an Indiana State Poultry Association NPIP Blood Testing School

More information

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page

Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide

More information

CHROME FRONT BRAKE MASTER CYLINDER KIT

CHROME FRONT BRAKE MASTER CYLINDER KIT -J075 REV. 009-0-0 CHROME FRONT BRAKE MASTER CYLINDER KIT GENERAL Kit Number 5-99D, 5-99D Models These Chrome Master Cylinder Kits are designed to replace the original equipment front brake master cylinder

More information

How to safely collect blood samples from persons suspected to be infected with highly infectious blood-borne pathogens (e.g.

How to safely collect blood samples from persons suspected to be infected with highly infectious blood-borne pathogens (e.g. How to safely collect blood samples from persons suspected to be infected with highly infectious blood-borne pathogens (e.g. Ebola) Step 1: Before entering patient room, assemble all equipment (1 st part)

More information

SPILLS & SPILL KITS. Spills -General Guidelines:

SPILLS & SPILL KITS. Spills -General Guidelines: Spills -General Guidelines: If handled properly, a spill may be nothing more than a nuisance. If handled improperly, a spill can seriously disrupt your activities and the work of your colleagues. At worst,

More information

Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools

Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools APPENDIX II Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools (Prepared by Environmental Protection Department) 1. Introduction The Waste Disposal (Chemical

More information

Mitos P-Pump. product datasheet

Mitos P-Pump. product datasheet product datasheet page Product description 2 Main benefits 3 Why choose the Mitos P-Pump? 3 Closed-loop flow control 4 Product specifications 5 Accessories - overview 6 Mitos P-Pump Basic 7 Mitos P-Pump

More information

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com.

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. Information contained in this publication is subject to change at anytime

More information

PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129

PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129 PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: The information included in this Splendide Repair Manual may change without notice. Please see our web site www.splendide.com/service/docs.html for updates,

More information

INSTRUCTIONS FOR USE. Read this information before you start. Important things to know about your SmartJect autoinjector. Important things to remember

INSTRUCTIONS FOR USE. Read this information before you start. Important things to know about your SmartJect autoinjector. Important things to remember INSTRUCTIONS FOR USE SIMPONI (SIM-po-nee) (golimumab) SmartJect autoinjector If your doctor decides that you or a caregiver may be able to give your injections of SIMPONI at home, you should receive training

More information

Achieving Independence. A Guide to Self-Catheterization with the Bard Touchless Plus Intermittent Catheter System

Achieving Independence. A Guide to Self-Catheterization with the Bard Touchless Plus Intermittent Catheter System Bard: Intermittent Self-Catheterization A Guide to Self-Catheterization with the Bard Touchless Plus Intermittent Catheter System Achieving Independence Introducing the Bard Touchless Plus Catheter One

More information

A discussion of condensate removal systems for clarifier and thickener drives for water and wastewater facilities.

A discussion of condensate removal systems for clarifier and thickener drives for water and wastewater facilities. A discussion of condensate removal systems for clarifier and thickener drives for water and wastewater facilities. Richard L. Dornfeld, BSME, P.E. Staff Engineer September 25, 2014 Moisture can be a problem

More information

Artisan Thermostatic basin mixer. Fitting Instructions & Contents List

Artisan Thermostatic basin mixer. Fitting Instructions & Contents List Artisan Thermostatic basin mixer Fitting Instructions & Contents List Please keep these instructions for future reference and request of replacement parts 1. Introduction Thank you for choosing Bristan,

More information

Depending on which elastic support you have you can secure the V2 power unit to the following seat bracket tubes: EXTERNAL DIAMETER 27.

Depending on which elastic support you have you can secure the V2 power unit to the following seat bracket tubes: EXTERNAL DIAMETER 27. ASSEMBLY 1 - EPS V2 POWER UNIT (SOLUTION 5) 1.1 - POSITIONING INSIDE THE SEAT TUBE WITH ELASTIC SUPPORT IN THE SEAT BRACKET TUBE Depending on which elastic support you have you can secure the V2 power

More information

globe reactions made easy Process perfection

globe reactions made easy Process perfection globe reactions made easy Process perfection Benefits More reactions in less time: The Globe oil drain unit and quick clamp allow single hand vessel changes in seconds Space saving: 50ml, 100ml, 250ml,

More information

Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit

Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Catalog No. MBS702810 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of mouse KLH(IgM)antibody concentrations

More information

Installation Instructions 4662.003

Installation Instructions 4662.003 Installation Instructions.00 SINGLE CONTROL KITCHEN FAUCET Thank you for selecting American-Standard...the benchmark of fine quality for over 00 years. To ensure that your installation proceeds smoothly--please

More information

Powers Controls TH 192 HC Heating/Cooling Room Thermostat

Powers Controls TH 192 HC Heating/Cooling Room Thermostat Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional

More information

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter INL/EXT-08-15056 Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter J. E. Daw November 2008 The INL is a U.S. Department of Energy National Laboratory operated by Battelle Energy

More information

Biosafety Level 2 (BSL-2) Safety Guidelines

Biosafety Level 2 (BSL-2) Safety Guidelines BLS-4 Biosafety Level 2 (BSL-2) Safety Guidelines BSL-3 BSL-2 BSL-1 BSL-2 builds upon BSL-1. If you work in a lab that is designated a BSL-2, the microbes used pose moderate hazards to laboratory staff

More information

Ultrasonic support removal tank

Ultrasonic support removal tank Ultrasonic support removal tank User guide and safety manual See inside for use and safety information. The Ultrasonic support removal tank automatically removes PLA support material from a printed ABS

More information

ROTOR LOADER OWNER S MANUAL

ROTOR LOADER OWNER S MANUAL ROTOR LOADER OWNER S MANUAL ROTOR LOADER OWNER S MANUAL WARNING IMPORTANT SAFETY INSTRUCTIONS AND GUIDELINES. Misuse of paintball equipment may cause serious injury or death. QUICK SET-UP GUIDE BATTERY

More information

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting

More information

ATS Overhead Table Shelf System INSTRUCTION MANUAL

ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship

More information

Laboratory Hazardous Waste Disposal Guidelines

Laboratory Hazardous Waste Disposal Guidelines Satellite Accumulation Area (SAA) For safety and environmental reasons, hazardous waste must be stored in a designated "Satellite Accumulation Area." These areas must be inspected weekly for container

More information

GenStore Archive Cabinet

GenStore Archive Cabinet GenStore Archive Cabinet Setup and Maintenance Manual Version A April 2014 Two GenStore Archives shown mounted on optional table stand. 2 Contents Getting Started... 4 Using this Manual...4 Safety Precautions...4

More information

Installation Instructions 6028.801

Installation Instructions 6028.801 DAZZLE Installation Instructions 08.80 Spread Lavatory Faucet with Speed Connect Drain* Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American

More information

Manual 20.2.2015 v1 Page 1 of 17. Manual. Laboratory autoclave

Manual 20.2.2015 v1 Page 1 of 17. Manual. Laboratory autoclave Manual 20.2.2015 v1 Page 1 of 17 Manual Laboratory autoclave Autoclave using steam for the treatment of material and laboratory processes e.g. sterilization, as an aid in research and development CertoClav

More information

Cell Sorting. S3e Cell Sorter Quick Guide

Cell Sorting. S3e Cell Sorter Quick Guide Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be

More information

THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494

THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494 THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494 1. BEFORE YOU BEGIN IMPORTANT INSTRUCTIONS READ AND SAVE FOR THE CONSUMER WARNING: Risk of scalding or other severe injury. Before completing installation,

More information

Installation Instructions / Warranty

Installation Instructions / Warranty Installation Instructions / Warranty Vertical Spray Bidet Fitting Rough 1322X181 Citterio 39224xx0 Citterio 39225xx0 Starck 10220xx0 Steel 35220800 Uno 38220xx0 Terrano 37224xx0 Terrano 37225xx0 Carlton

More information

Sikaflex Sealant Installation. Sika Corporation

Sikaflex Sealant Installation. Sika Corporation Sikaflex Sealant Installation 1 Critical Success Factors Substrate preparation If done properly would probably eliminate 95% of all call backs Most common mode of sealant failure Must remove all weak material

More information

1.1.2 Polypropylene The polypropylene must be a white opaque film, 1 2 mil thick. Biaxial orientation is preferred.

1.1.2 Polypropylene The polypropylene must be a white opaque film, 1 2 mil thick. Biaxial orientation is preferred. Library of Congress Preservation Directorate Specification Number 700 704 09 Specifications for Pressure Sensitive Adhesive Labels For Application to Single Paper Sheets and Text Pages of Bound Books For

More information

Lean Strategies Used to Optimize Automation. Why Lean Six Sigma? Laboratory Goals. Decreased TAT. Accurate Results. LEAN Goals.

Lean Strategies Used to Optimize Automation. Why Lean Six Sigma? Laboratory Goals. Decreased TAT. Accurate Results. LEAN Goals. Lean Strategies Used to Optimize Automation Linda Stubbs, ASQ CSSBB Senior Workflow Specialist Why LEAN Six Sigma? Why Lean Six Sigma? Laboratory Goals Decreased TAT Accurate Results LEAN Goals Increase

More information

ACCESSORIES OXYGEN-THERAPY

ACCESSORIES OXYGEN-THERAPY OXYGEN-THERAPY English version 11/ 2010 HUMIDIFIERS The humidifier is used to humidify the oxygen or the medical air delivered to the patient through the respiratory tract. Screwed to the outlet of a flowmeter

More information

History of Innovation

History of Innovation 1 History of Innovation Only Immucor consistently delivers the next-generation of blood bank automation to labs worldwide ABS HV Galileo NEO 1998 1999 2004 2007 2010 ABS2000 ABS Precis Echo 2 Innovation

More information

site monitoring Kit Site Monitoring Kit User Manual we prove it.

site monitoring Kit Site Monitoring Kit User Manual we prove it. site monitoring Kit 1 User Manual Site Monitoring Kit User Manual we prove it. 2 site monitoring Kit Content 1. Introduction 3 2. Content of the Site Monitoring Kit 4 3. Preparation 6 1. Positioning of

More information

Care of Your Hickman Catheter

Care of Your Hickman Catheter Care of Your Hickman Catheter Johns Hopkins Kimmel Cancer Center, Revised 7/11 Contents What is a Hickman Catheter? Page 3 Does the Catheter Limit My Activities? Page 4 How Do I Care for My Catheter? Page

More information

Human Free Testosterone(F-TESTO) ELISA Kit

Human Free Testosterone(F-TESTO) ELISA Kit Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be

More information

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172

USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172 Model # 900172: Page 1 USER MANUAL Bottom Loading Bottled Water Dispenser Model # 900172 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING

More information

www.morphyrichards.co.uk

www.morphyrichards.co.uk 43922 rev2:x 06/01/2010 15:06 Page 1 Upon registration with Morphy Richards (UK & ROI only) Register online for your 2 year guarantee. See back of this instruction book for details. (UK and Ireland customers

More information

SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies

SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies SDS-PAGE Protocol Mutated from the SDS-PAGE protocol written by the Lord of the Flies Pouring the resolving gel 1. Clean glass plates with soap and water, then with ethanol. Assemble the glass plates and

More information

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588). User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,

More information

GE Wireless Devices Battery Replacement

GE Wireless Devices Battery Replacement 60-506-319.5 Crystal Smoke Detector Two 9V Duracell 9V 1. Twist counter-clockwise until detector become loose from base. 2. Replace batteries observing correct polarity. 3. Replace detector by twisting

More information

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE

INSTRUCTIONS FOR USE HUMIRA 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE INSTRUCTIONS FOR USE HUMIRA (Hu-MARE-ah) (adalimumab) 40 MG/0.8 ML, 20 MG/0.4 ML AND 10 MG/0.2 ML SINGLE-USE PREFILLED SYRINGE Do not try to inject HUMIRA yourself until you have been shown the right way

More information

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam

More information

Cytotoxic and Biotherapies Credentialing Programme Module 6

Cytotoxic and Biotherapies Credentialing Programme Module 6 Cytotoxic and Biotherapies Credentialing Programme Module 6 1. Safe Handling and Administration of Cytotoxic and Biotherapies 2. Waste and Spill Management At the completion of this module the RN will

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

Bottom Loading Water Dispenser

Bottom Loading Water Dispenser Bottom Loading Water Dispenser Model # 601000 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING DISPENSER. SAVE THIS MANUAL FOR FUTURE

More information