Advanced Inertial Test Laboratory: Improving Low-Noise Testing of High-Accuracy Instruments

Size: px
Start display at page:

Download "Advanced Inertial Test Laboratory: Improving Low-Noise Testing of High-Accuracy Instruments"

Transcription

1 U.S. Air Force T&E Days February 2009, Albuquerque, New Mexico AIAA Advanced Inertial Test Laboratory: Improving Low-Noise Testing of High-Accuracy Instruments Cleon H. Barker 1 and William G. Rock 2 746th Test Squadron, Holloman Air Force Base, New Mexico The 746th Test Squadron, Central Inertial GPS Test Facility, has a long legacy of testing high-accuracy inertial guidance and navigation components and systems. Located in the Tularosa Basin of New Mexico, away from urban and marine-generated seismic contamination, makes it an ideal location for performing low-noise characterization testing of high-accuracy instruments. The Advanced Inertial Test Laboratory (AITL) was designed and built in the late 1960s for testing high accuracy inertial components and systems for the Department of Defense Strategic Missile Development. In recent years, there has been an increased interest within the Department of Defense for development of high-accuracy inertial and pointing systems. In anticipation of meeting the demanding requirements for low-noise characterization of these systems, the 746th Test Squadron began an initiative to characterize the background seismic environment on the test pads in the AITL facility and to make improvements in the seismic environment by reducing the cultural contamination generated by the environmental support equipment necessary for creating a laboratory environment. To date, the seismic environment on the test pads has been characterized, and improvements have begun to reduce the seismic contamination. This paper will discuss the characterization that has been completed, the improvements that have been completed and the future efforts that are anticipated for making further improvements. Nomenclature g = normative acceleration of gravity on Earth ( m/s 2 ) I. Introduction he 746th Test Squadron, also known as the Central Inertial and GPS Test Facility (CIGTF), at Holloman Air T Force Base, New Mexico is home to the Advanced Inertial Test Laboratory (AITL), figure 1. AITL is the most seismically-quiet test facility in the United States. The seismic characteristics of AITL are due, in part, to its location in the Tularosa Basin, a naturally seismically-quiet part of the nation away from urban and oceanic-generated disturbances. AITL was designed and built in the late 1960s for testing high-accuracy inertial components and systems for the Department of Defense s intercontinental ballistic missile programs. This testing was conducted throughout the 1970s and into the mid-1980s. Since the mid-80s, AITL was utilized to test inertial systems which were placed on the Hubble Space Telescope. When AITL was constructed, the facility s environmental support equipment (ESE) were mounted on vibration isolators and housed in a separate building, the Mechanical Building (see figure 1), in order to minimize the effects its operation had on the test environment. A seismic study conducted in Dec 05 revealed considerable contamination of the seismic environment on the test pads inside the AITL facility. Further investigation revealed much of the contamination was caused by degradation of some of the isolators, the lack of isolators under new replacement ESE and modifications to hardware mounting structures resulting in parallel paths for transmissibility of vibrations around the isolators. Since the Dec 05 study, a new refrigerated air conditioning water chiller and a new, larger air compressor for the test cells thermostats and pneumatic actuators was installed. Because of the results of the Dec 05 study, the installation of the new ESE and the expectation of future testing of high accuracy inertial and pointing systems for the Department of Defense, the decision was made in the spring of 2007 to conduct a new study to 1 Consultant, Strategic Development, 1644 Vandergrift Road Holloman Air Force Base NM Program Manager, Avionics Element, 1644 Vandergrift Road Holloman Air Force Base NM Copyright 2009 by the American Institute of Aeronautics and Astronautics, Inc. All rights reserved.

2 determine the most offensive ESE and the effectiveness of the passive isolation employed on each of the ESE. This study was conducted from Aug 07 to Mar 08. Figure 1. The Advanced Inertial Test Laboratory (AITL). During the course of the study, it became evident that the sensors, which were being used to record the seismic environment on the quietest test pad in AITL, were problematic. It was determined that the bandwidths of the same model sensor were not identical. Consequently, the outputs of the seismic sensors did not agree across the frequency band. This issue was somewhat mitigated by pairing up two sensors with similar bandwidths and disregarding the data at the higher frequencies where the bandwidths were different. While this provided more usable data in the context of the study, it was decided that a more robust solution was desirable. After the conclusion of the study, an effort was begun to determine the actual bandwidth for each seismic sensor and develop a mathematical model to compensate the scale factor for each of the sensors. II. Study Methods and Results The study was carried out in three phases. The first phase focused on determining the vibration generated by the newly installed air conditioning chiller and the air compressor, the second on the attenuation being provided by the passive isolators and, finally, measuring the seismic disturbance on the isolation pad in AITL from the transmitted vibrations through the ground. Two clusters of single-axis PCB Piezotronics 393B05 seismic accelerometers were used to record the vibrations generated by the chiller and air compressor and the vibration on the floor underneath the ESE. Our most sensitive seismometers could not be used for this recording, because they were physically saturated by the magnitude of the vibrations. Each cluster was comprised of three seismic accelerometers mounted orthogonally on an aluminum cube (figure 2). With only the chiller or air compressor operating and all other ESE turned off, data were recorded with one cluster placed on the ESE and the other on the concrete slab underneath. The recorded data from each sensor were converted into the frequency domain through a fast-fourier transform to produce a power spectral density (PSD) plot. A coherence plot was generated to determine the frequency content of the signals that were common to both sensor clusters. Next, a transfer function was plotted by dividing the data from the sensor under the unit by its parallel sensor mounted on the unit in the frequency domain. This then provided, in the frequency domain, a general assessment of how much attenuation was being provided by the passive isolators. Finally, the data were converted into units of acceleration (g) by taking the square root of the sum of the acceleration squared energy in the PSD. One plot displayed the cumulative acceleration recorded on the chiller or air compressor, and the other displayed the cumulative acceleration being felt on the floor underneath. The difference in the total cumulative accelerations was calculated in order to serve as a baseline for measuring the effectiveness of any future improvements that are made in isolation. This same procedure was repeated for each individual piece of ESE supporting the AITL facility.

3 Figure 2. PCB Piezotronics 393B05 Seismic Accelerometer Clusters. The second phase of the study focused on the effect that the ESE had on the seismic environment on the test pad in AITL s Room 13. The Room 13 test pad was chosen as the test area, because, of all the test pads in AITL, it provided the most isolation and was the location for the Hubble Space Telescope low-noise characterization gyro tests. For this part of the study, two different models of seismometers were used. The first model was a Geotech KS-2000 three axis seismometer. The KS-2000 is a single instrument with three single-axis seismometers mounted orthogonally inside its case. The second model was a Geotech S-510 single-axis seismometer. Using these instruments, two seismometer clusters were created. Each cluster consisted of one KS-2000 and three S-510s mounted to an aluminum plate. The S-510s were mounted orthogonally to each other and aligned with the seismometers inside the KS During testing, the two clusters were placed side by side on the test pad (figure 3). Figure 3. Two KS-2000 and S-510 Seismometer Clusters. Data from the Room 13 test pad were recorded under differing seismic conditions. First, data were collected with all of the ESE operating, followed by all of the ESE turned off. These measurements provided a baseline which indicated the best and worst possible seismic conditions on the test pad. The next series of measurements were conducted with a selected piece of ESE operating while all others were turned off. The individual pieces of ESE operated during this portion of the study were: an air compressor, a chilled water pump and several air handlers. Also, in addition to recording the disturbance generated by each individual air handler alone, all

4 air handlers were turned on simultaneously to record their combined contribution to the disturbance. The effects of the operation of air conditioning chiller on the isolation pad in Room 13 were also recorded. As before, the data collected were converted into the frequency domain through a fast-fourier transform and plotted as a PSD. Next, the coherence between two parallel, same model, seismometers was calculated in the frequency domain. A coherence of approximately one in a frequency range indicated the two sensors recorded true motion. A coherence of approximately zero indicates the data recorded in a particular frequency band was mostly random noise which is incoherent. Using the coherence function, the coherent power was calculated by taking the PSD from one seismometer and multiplying it by the coherence between the two parallel sensors. The resulting plot in the frequency domain then represented the true motion sensed by the seismometers. Because the units of PSD and Coherent Power (g 2 /Hz) are sometimes difficult to relate to, the power across the frequency band was summed and the square root taken of the sum yielding units in acceleration, g. The third phase, as mentioned previously, focused on measuring the vibration generated by the remaining ESE in the Mechanical Building and determining the attenuation provided by their respective passive isolators. The sensors, test methods, and analysis techniques were identical to those described in the first phase. The pieces of ESE evaluated in this manner were: each individual air handler, chilled water pump, ultra-clean air supply, and main commercial power transformer feeding the electrical requirements of the AITL complex. Once the study was completed, each ESE was ranked according to its contribution to the seismic disturbances recorded on the isolation pad in Room 13. Also from the study, it was determined that the disturbances generated by the ESE lie in the frequency band between 5 and 100 Hz. Additional information, concerning this study, can be found in the Seismic Characterization Study of the Advanced Inertial Test Laboratory (AITL) Report (CIGTF-SR ), dated 8 December III. Seismometer Calibration Method Vibrations important to inertial testing lie in the frequency band between 0.1 to 100 Hz. Motion below 0.1 Hz that concerns inertial testing is tilt or changes in the gravity vector. A different class of sensors is used to measure this motion and is readily available commercially. Seismometers with sensitivities down to 1 nano-g and an upper bandwidth of 100 Hz are not commercially available. Vibration sensors that operate between 1 and few thousand Hz have threshold sensitivity in the 1 micro-g range. Consequently, replacement with commercially-available seismometers was not an option. An attempt was made to have the seismometers in our inventory independently calibrated. The result of this effort was a certification that the manufacturers provided scale factors were accurate to within 10%. The manufacturer-supplied scale factor was a constant value for converting the signal voltage out of the sensor to values of acceleration, g, across its operating frequency band. What we required was a compensation algorithm to compensate for the high frequency roll off and, thereby, extend the upper frequency operating range. For the above reasons the 746th Test Squadron started an effort to develop a compensation model to extend the bandwidths of the KS-2000 and S-510 seismometers up to 100 Hz by performing our own calibration in house. The results from the study revealed the KS-2000s seismometers had good agreement below 40 Hz and the S-510s below 80 Hz. Therefore, our challenge was to generate a white random noise vibration between 20 and 100 Hz. To further complicate the effort, the vibration had to be less than 100 micro-g in magnitude to prevent the seismometers from physically saturating above 100 micro-g. The squadron has several shaker vibration systems used for environmental-type testing scenarios. These systems can generate precise controlled vibrations in the frequency range of 1 and 1 khz and accelerations from 0.1 to 85 g. One such shaker vibration system is the Ling B335 shaker (figure 4).

5 Figure 4. B335 Vibration System Mounted on Seismic Mass. The shaker was mounted on top of a concrete seismic mass which was 10 ft long by 5 ft wide and 3 ft thick. The concrete mass was supported on six isolating air bags for attenuating cultural disturbances that could potentially contaminate the test article preventing vibrations, generated by the shaker, from contaminating other test beds within the facility. The six air bags were connected in such a way that two air bags were connected to the same manifold that was pressurized to lift the seismic mass to a preset height. This configuration then created a three legged system (figure 5). In figure 5, the air bags are represented by the blue circles, the shaker by the orange and the mass by the gray. Figure 5. Top View Configuration of Shaker, Seismic Mass, & Air Bags. To create vertical vibrations of less than 100 micro-g, we hypothesized we would place the sensors on the concrete seismic mass supporting the vibration shaker and sense the vibrations that would be coupled into the concrete mass. To create horizontal vibrations, the shaker would be rotated about its trunion axis such that the shaker movement would be along a horizontal axis, thus generating horizontal vibrations in the seismic mass. For a reference standard to quantify the vibration profile of the concrete mass, we decided to use the PCB 393B05 seismic accelerometers mentioned above and shown in figure 2. The bandwidths of the seismic accelerometers were from 2 to several khz. The threshold sensitivities of a few micro g required the vibrations coupled into the seismic mass to be greater than 10 micro-g. The squadron has significant experience using the PCB seismic accelerometers and has built up a great deal of confidence in their ability to accurately record low-level seismic vibrations. The PCB used for this test effort was first calibrated against a calibrated accelerometer that was traceable to the National

6 10 0 PSD 10-2 S510 (V 2 /Hz) V 2 /Hz g 2 /Hz PSD PCB (g 2 /Hz) Coherence 1.0 Coherence Frequency (hz) Figure 6. PSD and Coherence of Vibrations on Seismic Mass. Institute of Standards and Technology. The scale factor was checked and the linearity across the frequency band of 1 to 100 Hz was confirmed. Through experimentation, it was determined a 0.5 g, 5 and 100 Hz, white random noise excitation of the shaker, generated vibrations of 10 and 100 micro-g on the seismic mass. However, the motion on the seismic mass was no longer white, because of resonances and anti-resonances in the shaker mount, concrete seismic mass and air bag suspension (figure 6). This was not expected to be a problem, as there was excellent coherence between the PCB seismic accelerometer and the seismometer under test. We then generated transfer functions between the PCB seismic accelerometer and the KS-2000 or S-510 seismometers. Sample transfer functions are represented in both figures 7 and 8. Figure 7 is a typical transfer function with the gain displayed logarithmically. The output was seismometer data in volts, and the input was PCB data in g. Because the logarithmic gain plot didn t provide adequate resolution for our purpose, we re-plotted the transfer function with linear gain (figure 8). When the gain was viewed in linear units of volts/g, it was easier to observe the differences in the two sensor outputs.

7 10 4 Gain log gain Phase 100 Degrees Coherence 1.0 Coherence Frequency (hz) Figure 7. Transfer Function (Gain), Phase and Coherence.

8 PSD 10 0 PCB (g 2 /Hz) S510 (V 2 /Hz) 10-5 Power/Hz Transfer Function (Linear Gain) 500 Scale Factor (v/g) Frequency (hz) Figure 8. PSDs of S510, PCB and Linear Gain. Next a model was generated from each of the linear transfer functions between each seismometer and the seismic accelerometer using the method of least squares. Figure 9 presents the models for both SN1402 and SN1405 seismometers. The red trace is the recorded data used to calculate the model and the black trace is the best fit quartic model plotted on a logarithmic frequency scale. Both generated models appeared to fit the recorded data satisfactorily.

9 Model Data Sensor 1402 gain Model Data Sensor 1405 gain Frequency (Hz) Figure 9. Linear Gain and Model. Seismic data were collected on the seismic test pier in room 13 of AITL. The uncorrected seismic data using the manufacturers provided scale factors are presented in figure 10. The top traces are the coherent power from each seismometer, the center plot is the coherence between the two seismometers, and the bottom traces are the coherent cumulative acceleration from each seismometer. The coherence between the two sensors became very poor above 75 Hz. The cause of the loose of coherence could indicate a resonance in the seismometer mounting fixture or loss of significant seismic motion above this frequency.

10 g 2 /Hz Coherent Power 10-8 Sensor Sensor Coherence 0.8 g 2 /Hz g 0.0 (x10-6 ) Sensor 1402 Sensor 1405 Root Sum Acceleration 10 0 Frequency (hz) Figure 10. Uncorrected Coherent Power, Coherence, and Cumulative Root Sum Acceleration. The generated model, unique to each seismometer, was then applied to the recorded seismic data, and a determination of the improvement in cumulative sum acceleration agreement between two parallel seismometers was made. An example of this investigation is shown in figures 11, and 12. Figure 11 shows the same data after applying the proper correction model as a scaling factor to the coherent power and cumulative sum acceleration of the two sensors. If the correction model achieved the desired level of improvement, the two cumulative sum accelerations would overlay. Observing the two cumulative sum acceleration traces in both figures 10 and 11 doesn t provide the fidelity necessary to make a clear determination. Therefore, the uncorrected cumulative sum accelerations were differenced as were the corrected cumulative sum accelerations.

11 Coherent Power 10-8 Sensor Sensor 1405 g 2 /Hz (x10-6 ) Root Sum Acceleration 40 Sensor Sensor 1405 g Frequency (Hz) Figure 11. Corrected Coherent Power and Cumulative Room Sum Acceleration. Figure 12 then displays the two differenced cumulative sum accelerations overlaid on the same chart. The black trace is the uncorrected difference and the red trace is the corrected difference. If the correction models were ideal, the corrected difference would have been zero across the entire frequency band. As the result in figure 12 shows, the correction model employed improved the data but did not remove all of the error present.

12 (x10-6 ) 2 Corrected Uncorrected Cumulative Sum Difference 0-2 g Frequency (Hz) Figure 12. Uncorrected and Corrected Differenced Cumulative Root Sum Accelerations. In conclusion our attempts in modeling the transfer functions to calculate a correction model for each of the seismometers have been a limited success. While the data were improved with the correction models, the level of improvement falls short of our requirements and expectations. With the poor coherence above 75 Hz in the seismic data, we could not expect close agreement between the two parallel sensors above 75 Hz. Future efforts will include testing the plate the seismometers are mounted on to determine if there are resonances above 75 Hz which are compromising the coherence. If there are resonances, we will redesign the seismometer plate to remove the resonances. Also, we anticipate retesting in the vibration laboratory with the calibrated PCB. These new tests will be conducted with 5 to 150 Hz white noise excitation to the shaker. By extending the frequency to 150 Hz we hope to better define the transfer function at the higher frequencies. We may also attempt to perform a calibration test with the shaker control accelerometer mounted on the seismic mass the shaker is anchored to. If the shaker can be controlled with its signal sensor mounted on the seismic mass, the disturbances felt on the seismic mass should be considerably whitened. Future efforts will undoubtedly produce unexpected results and present new challenges for us to overcome.

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication Thomas Reilly Data Physics Corporation 1741 Technology Drive, Suite 260 San Jose, CA 95110 (408) 216-8440 This paper

More information

Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs

Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs Dan Gregory Sandia National Laboratories Albuquerque NM 87185 (505) 844-9743 Fernando Bitsie Sandia

More information

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA ABSTRACT Random vibration is becoming increasingly recognized as the most realistic method of simulating the dynamic environment of military

More information

Active Vibration Isolation of an Unbalanced Machine Spindle

Active Vibration Isolation of an Unbalanced Machine Spindle UCRL-CONF-206108 Active Vibration Isolation of an Unbalanced Machine Spindle D. J. Hopkins, P. Geraghty August 18, 2004 American Society of Precision Engineering Annual Conference Orlando, FL, United States

More information

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER W. Li, J. Vandewege Department of Information Technology (INTEC) University of Gent, St.Pietersnieuwstaat 41, B-9000, Gent, Belgium Abstract: Precision

More information

Building a simple seismometer

Building a simple seismometer Building a simple seismometer Seismometers operate on the principle of inertia, i.e. a body at rest will tend to remain that way unless a force is applied to make it move. An ideal seismometer would be

More information

Vibrations can have an adverse effect on the accuracy of the end effector of a

Vibrations can have an adverse effect on the accuracy of the end effector of a EGR 315 Design Project - 1 - Executive Summary Vibrations can have an adverse effect on the accuracy of the end effector of a multiple-link robot. The ability of the machine to move to precise points scattered

More information

Application Note Noise Frequently Asked Questions

Application Note Noise Frequently Asked Questions : What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random

More information

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1 WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's

More information

Subminiature Load Cell Model 8417

Subminiature Load Cell Model 8417 w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...

More information

Protecting data centers from seismic activity Understanding enclosure standards is critical to protecting equipment

Protecting data centers from seismic activity Understanding enclosure standards is critical to protecting equipment Protecting data centers from seismic activity Understanding enclosure standards is critical to protecting equipment Author: Joel Young, engineer at Crenlo Abstract In the event of an earthquake, the type

More information

Vibration Isolation in Data Centers

Vibration Isolation in Data Centers Vibration Isolation in Data Centers Vibrations in Data Centers Vibrations in Data Centers can be produced by nearby construction works, heavy traffic, railways or even the own cooling units inside or next

More information

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force

More information

The Calculation of G rms

The Calculation of G rms The Calculation of G rms QualMark Corp. Neill Doertenbach The metric of G rms is typically used to specify and compare the energy in repetitive shock vibration systems. However, the method of arriving

More information

Cell Phone Vibration Experiment

Cell Phone Vibration Experiment Objective Cell Phone Vibration Experiment Most cell phones are designed to vibrate. But at what frequency do they vibrate? With an accelerometer, data acquisition and signal analysis the vibration frequency

More information

Jitter Measurements in Serial Data Signals

Jitter Measurements in Serial Data Signals Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring

More information

Whitepaper. Image stabilization improving camera usability

Whitepaper. Image stabilization improving camera usability Whitepaper Image stabilization improving camera usability Table of contents 1. Introduction 3 2. Vibration Impact on Video Output 3 3. Image Stabilization Techniques 3 3.1 Optical Image Stabilization 3

More information

GigaSPEED X10D Solution How

GigaSPEED X10D Solution How SYSTIMAX Solutions GigaSPEED X10D Solution How White Paper June 2009 www.commscope.com Contents Modal Decomposition Modeling and the Revolutionary 1 SYSTIMAX GigaSPEED X10D Performance MDM An Introduction

More information

Basic Principles of Inertial Navigation. Seminar on inertial navigation systems Tampere University of Technology

Basic Principles of Inertial Navigation. Seminar on inertial navigation systems Tampere University of Technology Basic Principles of Inertial Navigation Seminar on inertial navigation systems Tampere University of Technology 1 The five basic forms of navigation Pilotage, which essentially relies on recognizing landmarks

More information

Decoding an Accelerometer Specification. What Sensor Manufacturer s Don t Tell You! David Lally VP Engineering PCB Piezotronics, Inc.

Decoding an Accelerometer Specification. What Sensor Manufacturer s Don t Tell You! David Lally VP Engineering PCB Piezotronics, Inc. Decoding an Accelerometer Specification Sheet What Sensor Manufacturer s Don t Tell You! David Lally VP Engineering PCB Piezotronics, Inc. IMAC-XXVII February 2008 1 Overview Specification Sheet Provides

More information

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz

AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2

More information

Introduction to Accuracy and Repeatability in Linear Motion Systems

Introduction to Accuracy and Repeatability in Linear Motion Systems Introduction to accuracy and repeatability in linear motion systems By Gary Rosengren, Director of Engineering Tolomatic, Inc. About the Author Gary Rosengren is Director of Engineering at Tolomatic and

More information

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is: 4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing

More information

Frequency Response of Filters

Frequency Response of Filters School of Engineering Department of Electrical and Computer Engineering 332:224 Principles of Electrical Engineering II Laboratory Experiment 2 Frequency Response of Filters 1 Introduction Objectives To

More information

Lecture 07: Work and Kinetic Energy. Physics 2210 Fall Semester 2014

Lecture 07: Work and Kinetic Energy. Physics 2210 Fall Semester 2014 Lecture 07: Work and Kinetic Energy Physics 2210 Fall Semester 2014 Announcements Schedule next few weeks: 9/08 Unit 3 9/10 Unit 4 9/15 Unit 5 (guest lecturer) 9/17 Unit 6 (guest lecturer) 9/22 Unit 7,

More information

E6211.02-113-11-R0 ACOUSTICAL PERFORMANCE TEST REPORT ASTM E90. Rendered to: TREX COMPANY, LLC. Series/Model: Trex Seclusion

E6211.02-113-11-R0 ACOUSTICAL PERFORMANCE TEST REPORT ASTM E90. Rendered to: TREX COMPANY, LLC. Series/Model: Trex Seclusion ACOUSTICAL PERFORMANCE TEST REPORT ASTM E90 Rendered to: TREX COMPANY, LLC Series/Model: Trex Seclusion Type: Sound Barrier Fence Summary of Test Results Data File No. Description STC OITC E6211.01 Trex

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Address for Correspondence

Address for Correspondence International Journal of Advanced Engineering Technology E-ISSN 0976-3945 Research Paper DEVELOPMENT OF LOW COST SHAKE TABLES AND INSTRUMENTATION SETUP FOR EARTHQUAKE ENGINEERING LABORATORY C. S. Sanghvi

More information

SEISMIC MONITORING & PROTECTION SYSTEMS

SEISMIC MONITORING & PROTECTION SYSTEMS SEISMIC MONITORING & PROTECTION SYSTEMS PRODUCT GUIDE Seismic Transducers & Switches Nuclear Infrastructure Protection Structural & Environmental Monitoring Data Capture & Analysis 25 YEARS 1978-2003 PROTECTING

More information

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies

Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies Soonwook Hong, Ph. D. Michael Zuercher Martinson Harmonics and Noise in Photovoltaic (PV) Inverter and the Mitigation Strategies 1. Introduction PV inverters use semiconductor devices to transform the

More information

How To Control Gimbal

How To Control Gimbal Tarot 2-Axis Brushless Gimbal for Gopro User Manual V1.0 1. Introduction Tarot T-2D gimbal is designed for the Gopro Hero3, which is widely used in film, television productions, advertising aerial photography,

More information

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5 Physics 161 FREE FALL Introduction This experiment is designed to study the motion of an object that is accelerated by the force of gravity. It also serves as an introduction to the data analysis capabilities

More information

Impedance 50 (75 connectors via adapters)

Impedance 50 (75 connectors via adapters) VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:

More information

Experiment 7: Familiarization with the Network Analyzer

Experiment 7: Familiarization with the Network Analyzer Experiment 7: Familiarization with the Network Analyzer Measurements to characterize networks at high frequencies (RF and microwave frequencies) are usually done in terms of scattering parameters (S parameters).

More information

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be 86 Chapter 4 Accelerometer Testing 4.1 Introduction The accelerometer designed and realized so far is intended for an aerospace application. Detailed testing and analysis needs to be conducted to qualify

More information

Understanding Dynamic Range in Acceleration Measurement Systems. February 2013 By: Bruce Lent

Understanding Dynamic Range in Acceleration Measurement Systems. February 2013 By: Bruce Lent in Acceleration Measurement Systems February 2013 By: Bruce Lent Topics to discuss Definition of dynamic range The effective range Making full use of the high level Using filters to improve dynamic range

More information

Email: tjohn@mail.nplindia.ernet.in

Email: tjohn@mail.nplindia.ernet.in USE OF VIRTUAL INSTRUMENTS IN RADIO AND ATMOSPHERIC EXPERIMENTS P.N. VIJAYAKUMAR, THOMAS JOHN AND S.C. GARG RADIO AND ATMOSPHERIC SCIENCE DIVISION, NATIONAL PHYSICAL LABORATORY, NEW DELHI 110012, INDIA

More information

The Sonometer The Resonant String and Timbre Change after plucking

The Sonometer The Resonant String and Timbre Change after plucking The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes

More information

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes

More information

Fiber Optic High Density ACC Sensor System Faser Optisches Beschleunigungs-Messsystem

Fiber Optic High Density ACC Sensor System Faser Optisches Beschleunigungs-Messsystem End Winding Remote unit PNR: 9600.30.180 Remote unit and 6 x One-Axis optical, MIC-Accelerometer Frequency response 1 1000 Hz Easy installation on the end-winding High voltage electrical insulation between

More information

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION 1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.

More information

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in

More information

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative

More information

Harmonic oscillations of spiral springs Springs linked in parallel and in series

Harmonic oscillations of spiral springs Springs linked in parallel and in series .3.26 Related topics Spring constant, Hooke s Law, oscillations, limit of elasticity, parallel springs, serial springs, use of an interface. Principle and task The spring constant D is determined for different

More information

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

Bandwidth-dependent transformation of noise data from frequency into time domain and vice versa

Bandwidth-dependent transformation of noise data from frequency into time domain and vice versa Topic Bandwidth-dependent transformation of noise data from frequency into time domain and vice versa Authors Peter Bormann (formerly GeoForschungsZentrum Potsdam, Telegrafenberg, D-14473 Potsdam, Germany),

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena

More information

Product Overview 9100D

Product Overview 9100D Product Overview 9100D Portable Shaker Vibration Calibrator The Modal Shop 3149 East Kemper Road Cincinnati, OH 45241 Phone: +1.513.351.9919 Fax: 513.458.2171 info@modalshop.com Brief Overview We've taken

More information

200350 and 200355 Accelerometers

200350 and 200355 Accelerometers and Accelerometers Bently Nevada Asset Condition Monitoring Description The and Accelerometers are general purpose, casemounted seismic transducers designed for use with Trendmaster Pro Constant Current

More information

Measuring Impedance and Frequency Response of Guitar Pickups

Measuring Impedance and Frequency Response of Guitar Pickups Measuring Impedance and Frequency Response of Guitar Pickups Peter D. Hiscocks Syscomp Electronic Design Limited phiscock@ee.ryerson.ca www.syscompdesign.com April 30, 2011 Introduction The CircuitGear

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

PRODUCT DATASHEET. J1939 Vehicle Inertia Monitor. Advanced Vehicle Inertial Measurement and Vibration Monitoring Device. fleet-genius.

PRODUCT DATASHEET. J1939 Vehicle Inertia Monitor. Advanced Vehicle Inertial Measurement and Vibration Monitoring Device. fleet-genius. PRODUCT DATASHEET fleet-genius.com J1939 Vehicle Inertia Monitor Advanced Vehicle Inertial Measurement and Vibration Monitoring Device Prova s J1939 Vehicle Inertia Monitor (VIM) formulates moving vehicle

More information

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows LS-296 Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows T. L. Smith ASD / RF Group Advanced Photon Source Argonne National Laboratory June 26, 2002 Table of Contents 1) Introduction

More information

Figure 1. The Ball and Beam System.

Figure 1. The Ball and Beam System. BALL AND BEAM : Basics Peter Wellstead: control systems principles.co.uk ABSTRACT: This is one of a series of white papers on systems modelling, analysis and control, prepared by Control Systems Principles.co.uk

More information

Lateral Acceleration. Chris Garner

Lateral Acceleration. Chris Garner Chris Garner Forward Acceleration Forward acceleration is easy to quantify and understand. Forward acceleration is simply the rate of change in speed. In car terms, the quicker the car accelerates, the

More information

Force/position control of a robotic system for transcranial magnetic stimulation

Force/position control of a robotic system for transcranial magnetic stimulation Force/position control of a robotic system for transcranial magnetic stimulation W.N. Wan Zakaria School of Mechanical and System Engineering Newcastle University Abstract To develop a force control scheme

More information

CHAPTER 3 MODAL ANALYSIS OF A PRINTED CIRCUIT BOARD

CHAPTER 3 MODAL ANALYSIS OF A PRINTED CIRCUIT BOARD 45 CHAPTER 3 MODAL ANALYSIS OF A PRINTED CIRCUIT BOARD 3.1 INTRODUCTION This chapter describes the methodology for performing the modal analysis of a printed circuit board used in a hand held electronic

More information

Acceleration of Gravity Lab Basic Version

Acceleration of Gravity Lab Basic Version Acceleration of Gravity Lab Basic Version In this lab you will explore the motion of falling objects. As an object begins to fall, it moves faster and faster (its velocity increases) due to the acceleration

More information

Agilent 8510-13 Measuring Noninsertable Devices

Agilent 8510-13 Measuring Noninsertable Devices Agilent 8510-13 Measuring Noninsertable Devices Product Note A new technique for measuring components using the 8510C Network Analyzer Introduction The majority of devices used in real-world microwave

More information

MODULATION Systems (part 1)

MODULATION Systems (part 1) Technologies and Services on Digital Broadcasting (8) MODULATION Systems (part ) "Technologies and Services of Digital Broadcasting" (in Japanese, ISBN4-339-62-2) is published by CORONA publishing co.,

More information

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display

More information

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE T-Series industrial inclinometers are compact high performance sensors used to determine inclination in roll and pitch axes with excellent precision and

More information

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

Direct and Reflected: Understanding the Truth with Y-S 3

Direct and Reflected: Understanding the Truth with Y-S 3 Direct and Reflected: Understanding the Truth with Y-S 3 -Speaker System Design Guide- December 2008 2008 Yamaha Corporation 1 Introduction Y-S 3 is a speaker system design software application. It is

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2009 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

The Time Constant of an RC Circuit

The Time Constant of an RC Circuit The Time Constant of an RC Circuit 1 Objectives 1. To determine the time constant of an RC Circuit, and 2. To determine the capacitance of an unknown capacitor. 2 Introduction What the heck is a capacitor?

More information

RF Network Analyzer Basics

RF Network Analyzer Basics RF Network Analyzer Basics A tutorial, information and overview about the basics of the RF Network Analyzer. What is a Network Analyzer and how to use them, to include the Scalar Network Analyzer (SNA),

More information

MXD7202G/H/M/N. Low Cost, Low Noise ±2 g Dual Axis Accelerometer with Digital Outputs

MXD7202G/H/M/N. Low Cost, Low Noise ±2 g Dual Axis Accelerometer with Digital Outputs Low Cost, Low Noise ±2 g Dual Axis Accelerometer with Digital Outputs MXD7202G/H/M/N FEATURES Low cost Resolution better than 1 milli-g Dual axis accelerometer fabricated on a monolithic CMOS IC On chip

More information

State Newton's second law of motion for a particle, defining carefully each term used.

State Newton's second law of motion for a particle, defining carefully each term used. 5 Question 1. [Marks 20] An unmarked police car P is, travelling at the legal speed limit, v P, on a straight section of highway. At time t = 0, the police car is overtaken by a car C, which is speeding

More information

E190Q Lecture 5 Autonomous Robot Navigation

E190Q Lecture 5 Autonomous Robot Navigation E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator

More information

Automated Contact Resistance Tester CR-2601

Automated Contact Resistance Tester CR-2601 Automated Contact Resistance Tester CR-2601 What s New What s New Summary of Hardware Improvements: The middle Stiffener has been improved and no longer comes in direct contact with the main board thus

More information

Building Design for Advanced Technology Instruments Sensitive to Acoustical Noise

Building Design for Advanced Technology Instruments Sensitive to Acoustical Noise Building Design for Advanced Technology Instruments Sensitive to Acoustic Noise Michael Gendreau Colin Gordon & Associates Presentation Outline! High technology research and manufacturing instruments respond

More information

Weight Measurement Technology

Weight Measurement Technology Kistler-Morse (KM) introduced bolt-on weight measuring systems three decades ago. These devices featured Walter Kistler s invention, the Microcell. Over the years, many improvements were made to the Microcell

More information

IR-Zimmerman Lifting Equipment Product Overview

IR-Zimmerman Lifting Equipment Product Overview IR-Zimmerman Lifting Equipment Product Overview Handling Devices Torque Reaction Products Irax Spring Balancers Quantum Electric Chain Hoists End Effectors The real workhorse of material handling equipment

More information

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range Application Note Table of Contents 3 3 3 4 4 4 5 6 7 7 7 7 9 10 10 11 11 12 12 13 13 14 15 1. Introduction What is dynamic range?

More information

Performance Evaluation of a UWB-RFID System for Potential Space Applications Abstract

Performance Evaluation of a UWB-RFID System for Potential Space Applications Abstract Performance Evaluation of a UWB-RFID System for Potential Space Applications Abstract This talk presents a brief overview of the ultra-wideband (UWB) RFID system with emphasis on the performance evaluation

More information

Hands On ECG. Sean Hubber and Crystal Lu

Hands On ECG. Sean Hubber and Crystal Lu Hands On ECG Sean Hubber and Crystal Lu The device. The black box contains the circuit and microcontroller, the mini tv is set on top, the bars on the sides are for holding it and reading hand voltage,

More information

Setting up the Physical Environment Monitoring System for Advanced LIGO: SURF Final Report

Setting up the Physical Environment Monitoring System for Advanced LIGO: SURF Final Report Setting up the Physical Environment Monitoring System for Advanced LIGO: SURF Final Report Maggie Tse Mentors: Robert Schofield and Daniel Sigg August 9, 01 1 Introduction Environmental influences contribute

More information

The process components and related data characteristics addressed in this document are:

The process components and related data characteristics addressed in this document are: TM Tech Notes Certainty 3D November 1, 2012 To: General Release From: Ted Knaak Certainty 3D, LLC Re: Structural Wall Monitoring (#1017) rev: A Introduction TopoDOT offers several tools designed specifically

More information

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil

Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Physics 8.02 Spring 2006 Experiment 3: Magnetic Fields of a Bar Magnet and Helmholtz Coil OBJECTIVES 1. To learn how to visualize magnetic field lines

More information

Small, Low Power, 3-Axis ±3 g Accelerometer ADXL335

Small, Low Power, 3-Axis ±3 g Accelerometer ADXL335 Small, Low Power, 3-Axis ±3 g Accelerometer ADXL335 FEATURES 3-axis sensing Small, low profile package 4 mm 4 mm 1.45 mm LFCSP Low power : 35 μa (typical) Single-supply operation: 1.8 V to 3.6 V 1, g shock

More information

Using angular speed measurement with Hall effect sensors to observe grinding operation with flexible robot.

Using angular speed measurement with Hall effect sensors to observe grinding operation with flexible robot. Using angular speed measurement with Hall effect sensors to observe grinding operation with flexible robot. François Girardin 1, Farzad Rafieian 1, Zhaoheng Liu 1, Marc Thomas 1 and Bruce Hazel 2 1 Laboratoire

More information

Measuring Laser Power and Energy Output

Measuring Laser Power and Energy Output Measuring Laser Power and Energy Output Introduction The most fundamental method of checking the performance of a laser is to measure its power or energy output. Laser output directly affects a laser s

More information

Field Application Note

Field Application Note Field Application Note Reverse Dial Indicator Alignment RDIA Mis-alignment can be the most usual cause for unacceptable operation and high vibration levels. New facilities or new equipment installations

More information

On Cables and Connections A discussion by Dr. J. Kramer

On Cables and Connections A discussion by Dr. J. Kramer KRAMER ELECTRONICS LTD. On Cables and Connections A discussion by Dr. J. Kramer We are frequently asked - "what length of cable can I use for a specific application?" Seemingly a simple question, but the

More information

Laboratory Report Scoring and Cover Sheet

Laboratory Report Scoring and Cover Sheet Laboratory Report Scoring and Cover Sheet Title of Lab _Newton s Laws Course and Lab Section Number: PHY 1103-100 Date _23 Sept 2014 Principle Investigator _Thomas Edison Co-Investigator _Nikola Tesla

More information

Filter Comparison. Match #1: Analog vs. Digital Filters

Filter Comparison. Match #1: Analog vs. Digital Filters CHAPTER 21 Filter Comparison Decisions, decisions, decisions! With all these filters to choose from, how do you know which to use? This chapter is a head-to-head competition between filters; we'll select

More information

Determination of source parameters from seismic spectra

Determination of source parameters from seismic spectra Topic Determination of source parameters from seismic spectra Authors Michael Baumbach, and Peter Bormann (formerly GeoForschungsZentrum Potsdam, Telegrafenberg, D-14473 Potsdam, Germany); E-mail: pb65@gmx.net

More information

Design Validation and Improvement Study of HVAC Plumbing Line Assembly under Random Loading Condition

Design Validation and Improvement Study of HVAC Plumbing Line Assembly under Random Loading Condition Design Validation and Improvement Study of HVAC Plumbing Line Assembly under Random Loading Condition Rakesh Jakhwal Senior Engineer Chrysler Group LLC RMZ Millenia II, Perungudi Chennai 600096, India

More information

Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP

Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP Department of Electrical and Computer Engineering Ben-Gurion University of the Negev LAB 1 - Introduction to USRP - 1-1 Introduction In this lab you will use software reconfigurable RF hardware from National

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

Command-induced Tracking Jitter Study I D. Clark November 24, 2009

Command-induced Tracking Jitter Study I D. Clark November 24, 2009 Command-induced Tracking Jitter Study I D. Clark November 24, 2009 Introduction Reports of excessive tracking jitter on the MMT elevation axis have lately been theorized to be caused by the input command

More information

State Newton's second law of motion for a particle, defining carefully each term used.

State Newton's second law of motion for a particle, defining carefully each term used. 5 Question 1. [Marks 28] An unmarked police car P is, travelling at the legal speed limit, v P, on a straight section of highway. At time t = 0, the police car is overtaken by a car C, which is speeding

More information

Capacitive Touch Sensor Project:

Capacitive Touch Sensor Project: NOTE: This project does not include a complete parts list. In particular, the IC described here does not come in a dual-inline-package (DIP), and so a gull-wing package has to be soldered to an adaptor

More information

TELESCOPE Motorized Lighting Suspension

TELESCOPE Motorized Lighting Suspension TELESCOPE Motorized Lighting Suspension Mod. 8400 FEATURES The De Sisti TELESCOPE is a single point suspension specifically designed to provide rigidity and stabilization of all moving parts. It can extend

More information

PdM Overview. Predictive Maintenance Services

PdM Overview. Predictive Maintenance Services PdM Overview Predictive Maintenance Services Objective Assessments Maximize Uptime, Lower Costs Predictive Maintenance (PdM) solutions from Rexnord Industrial Services help manage the condition of your

More information

8. Potential Energy and Conservation of Energy Potential Energy: When an object has potential to have work done on it, it is said to have potential

8. Potential Energy and Conservation of Energy Potential Energy: When an object has potential to have work done on it, it is said to have potential 8. Potential Energy and Conservation of Energy Potential Energy: When an object has potential to have work done on it, it is said to have potential energy, e.g. a ball in your hand has more potential energy

More information

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER Latest revision: October 1999 Introduction A vector network analyzer (VNA) is a device capable of measuring both the magnitude and phase of a sinusoidal

More information