Exploring nsight Imaging a totally new architecture for premium ultrasound
|
|
|
- Felix Jones
- 9 years ago
- Views:
Transcription
1 Exploring nsight Imaging a totally new architecture for premium ultrasound Philips EPIQ ultrasound system Karl Thiele, Principal Scientist James Jago, Principal Scientist Robert Entrekin, Technical Fellow Roy Peterson, Director Systems Engineering For the past 30 years, ultrasound has experienced dramatic changes as manufacturers strive to develop new solutions to meet the needs of clinicians. From the pioneering days of static B-scan to color Doppler and real-time imaging, each brought major dynamics in diagnostic ultrasound. New innovative beamforming technologies are emerging that promise a dramatic increase in imaging performance. These exciting developments elevate ultrasound into a more definitive modality that will potentially expand ultrasound in more diverse roles as well as improve clinical decision support. nsight Imaging at a glance New precision beamforming Processes an enormous and massive parallel processing amount of acoustic data allowing the system to focus down to the pixel level all in real time Reconstructs virtually perfect Users can now experience beams with fewer transmit both highly detailed operations, breaking the ultrasound images traditional compromise of and extraordinary architectures of the past temporal resolution Dynamically calculates and Achieve superb tissue reconstructs the optimal uniformity all the way up transmit and receive focusing to the skin line without the continually at all depths compromise of conventional down to the pixel level transmit focus limitations Table of contents Conventional beamforming 1 Multi-line beamforming 2 Synthetic focus beamforming 3 nsight Imaging explored 4 nsight Imaging clinical breakthroughs 7 Ultra-wide dynamic range and unique beam reconstruction provides exceptional tissue information at greater depths with less noise Visualize extraordinary levels of detail and contrast resolution with exceptional penetration at higher frequencies even on difficult patients
2 Philips EPIQ ultrasound system Conventional ultrasound beamforming In the early days of real-time ultrasound, a fixed focus single element transducer was mechanically swept back and forth over the image field to acquire images. The focus of the transducer was defined by an acoustic lens and was fixed for both wave transmission and echo reception, providing the best image quality only around the fixed focal depth. The 1980s brought the advent of solid-state phased array transducers. Array-based transducers brought electronic beamforming technology to ultrasound, introducing new levels of clinical versatility and imaging performance. The beamformer enabled functions such as transmit beam steering, transmit focusing, and receive focusing. The programmable digital beamformer enabled the echo signals to be continually focused as they were received by the array transducer and the received beams were maintained in constant focus. This was done by constantly updating the beamformer focus delays as the echoes returned from ever-increasing depths of field. This only solved half of the focusing problem, however. The transmit beam was still only focused at one depth of field and optimal resolution was only attained around the transmit focal region (Figure 1). The round-trip beam profile, a product of both the transmit beam profile and the receive beam profile, still had room for improvement. Multi-line beamforming The frame rate of ultrasound images is governed by the time required to scan the complete image field, which in turn is a function of the number of transmit-receive cycles needed to scan the image. In the simplest form of beamforming a single transmit will result in a single receive data line used to build the image (Figure 2). Variable time delays Shifted transmit pulses Wave fronts Single line receive generation Single transmit beam Focal region Array elements Figure 1 Simple transmit focusing Figure 1 Simple transmit focusing Figure 2 Single line beamformer Figure 2 Single line beamformer 2
3 Efforts were made in the 1990s to reduce the number of transmit-receive cycles in two ways. One was to increase the spacing between transmitted and received scanlines, then interpolate synthetic scanlines between the actual received lines. While providing an immediate increase in frame rate, interpolation had its own limitations. Anatomical structures tended to be less resolved since the interpolated scanlines were averaged from the actual ultrasound data on either side. The extent to which the actual scanline spacing can be increased is limited by the need to adequately spatially sample the image field with transmitted and received ultrasound. While providing an improvement in temporal resolution, interpolation also required a compromise in image resolution. The second effort toward improved temporal resolution was to prove more meaningful. Parallel receive beam processing appeared commercially in the form of multi-line beam formation. Improvements in processing power now made it possible to insonify a broad area encompassing several scanlines, then to receive and form beams at multiple scanline locations simultaneously. Instead of obtaining four scanlines with four transmit-receive cycles, the four scanlines could be obtained in response to a single transmit event (Figure 3). The time to scan a full image field and hence the frame rate of display improved by a factor of four in this example. But higher levels of multi-line requires the broadening of the transmit beam to insonify the greater expanse of multiple scanline locations. Multi-line receive generation Single transmit beam The transmit beam was even less focused than in the past, leading to the development of image artifacts. However, even with artifacts, multi-line beam formation was an important step in improving frame rate. Synthetic focus beamforming More recently, synthetic focus beamforming (also known as zonal focusing or plane wave imaging) has been available commercially. Synthetic focus beamforming has been known since the mid-1970s, but remained only of academic interest until recent years when higher density data storage and increased computational capability of microprocessors enabled the development of practical implementations. Synthetic focusing attempts to overcome some of the limitations of conventional beamforming by insonifying all or most of the image field during each transmit event essentially reducing the influence of transmit focusing on the final image (Figure 4). Image data Multi-line receive Software synthetic focus Figure 4 Synthetic focus/plane wave beamforming Single unfocused transmit beam This allows each transducer element to receive energy from most of the target region, so that all of the round-trip focusing required to generate an image can be performed on receive thus allowing for the possibility of much higher frame rates, with reasonable spatial resolution. Figure 3 Multi-line beamforming 3
4 However, synthetic focus beamformers have limitations of their own. Since synthetic focusing uses low intensity, unfocused plane waves on transmit, focusing occurs only on receive. The low transmit intensity results in poor signal to noise and a loss in penetration. If one attempts to recover signal to noise by averaging images from multiple transmit events, it reduces the frame rate, making it poorly suited for cardiac imaging other scenarios involving motion. Tissue harmonic imaging in particular requires higher amplitude ultrasound waves to be generated, and this normally requires some amount of transmit focusing which is again incompatible with synthetic focusing. Figure 5 Conventional focused transmit beam Breaking the rules of conventional ultrasound Through all of these incremental improvements in beam focusing, image resolution, and frame rate, the transmit beam has remained the same. It was still focused only at its programmed transmit focal point and its profile is still that of an hourglass (Figure 5). Spatial resolution Figure 6 Conventional architecture linked imaging attributes With conventional beamforming, imaging attributes such as spatial resolution, temporal resolution and tissue uniformity are linked meaning if one attribute is improved another attribute is adversely affected (Figure 6). One must always compromise one or more of the imaging attributes to achieve performance gains for another. For example, when users want to image with high frame rates perhaps to visualize a rapidly moving structure or to allow the transducer to be swept across the patient when searching for pathology they must accept a compromise in image quality, typically spatial resolution. Temporal resolution Tissue uniformity Figure 7 nsight Imaging transmit beam reconstruction This conventional wisdom has now been dashed by Philips nsight Imaging a totally new way to form images without the compromise of conventional architectures. nsight Imaging reconstructs virtually perfect transmit beams throughout the depth of field. The round-trip beam profile is now no longer the product of an hourglass shape on transmit and a pencil shape on receive, but the product of two sharply defined pencil profiles (Figure 7). 4
5 As a result, dramatic improvements in contrast resolution, anatomical detail resolution and depth of field are realized, particularly in the very near and very far fields. With nsight Imaging spatial resolution, temporal resolution and tissue uniformity are unlinked, allowing the user to improve all aspects of imaging performance without compromising each other (Figure 8). Spatial resolution Figure 8 nsight Imaging attributes are unlinked nsight Imaging supports a much higher degree of parallel beamforming than conventional multi-line approaches, allowing a much higher opportunity for frame rate improvements. The technique of combining multiple receive beams with other receive beams obtained from different transmit events essentially a form of inverse filtering automatically corrects for broadened transmit beam profiles and eliminates multi-line artifacts. Inverse filtering also continuously adjusts with depth as the transmit beam profile changes (due to transmit focusing), thus increasing the depth of field of each transmit focus. Temporal resolution Tissue uniformity Since multiple transmit beams are combined to generate each round-trip beam, and each transmit beam has a different noise pattern, the noise sums incoherently while the signal sums coherently, thus improving the signal-to-noise ratio and hence penetration. Exploring nsight Imaging nsight Imaging is a unique combination of a new precision beamformer and massive parallel processing. This innovative architecture allows coherent beam reconstruction in real time, incorporating the best features of multi-line beamforming with synthetic focusing techniques (Figure 9). To understand the details of nsight Imaging further, let s dissect the ultrasound beams of a conventional transmit-receive cycle (Figure 10). The red focus line is the transmit beam center and the blue line is the aligned receive beam focus. When the transmit beam is launched from the transducer array, it is timed to have a curved wave front which will cause it to converge at a predetermined focal point. Reconstructed image data High order multi-line receive Focused transmit beams Transmit wave front diverging away from focus Precision beamforming Massive parallel processing Accoustic data memory T 1 T 2 T 3 T 4 Receive Focus Figure 9 Philips nsight Imaging Transmit wave front converging toward focus Figure Conventional transmit beam beam 5
6 As the drawing shows, the transmitted converging wave front appears convex when viewed from the array. After the wave front converges at the focus it begins to diverge, with a concave appearance as viewed from the array. This process of convergence, focus, and divergence gives the beam the classic hourglass beam profile shown on either side of the beam center. The receive beam, being continually focused as it is received, has a narrow, pencil-like beam profile. The resultant round-trip beam profile (that defines the lateral resolution) is the product of both transmit and receive beam profiles. So even if the transmit beam is not as well focused as the receive beam, the round-trip beam profile should still be sharper (narrower). However, if both beams are pencil thin, then the product of the two would give us the sharpest round-trip beam possible, especially away from the original transmit focus. This is the promise of nsight Imaging. Multiple transmit-receive cycles nsight Imaging technology applies multiple transmit-receive cycles with spatially different transmit beam profiles. For example, Figure 11 illustrates three spatially adjacent transmit beams, T1, T2, and T3. All three transmit beams have the same converging and diverging wave fronts as shown in the drawing and all three are focused at the same depth as indicated by the vertical red line in the center. On receive, a beam is formed in response to each transmit beam which is at the same spatial location indicated by the blue receive beam R. In each case, the receive beamforming is the same, so three dynamically focused receive beams are received along the blue arrow R. While the three transmit beams are all tightly focused at the focal depth, it is seen that their wave fronts are out of phase both before and after focal convergence. R Figure 11 Multiple transmit-receive cycles Consider just the near field shown in Figure 12. The single center beam has the correct phase location, yet the outer two beams (top and bottom) have the incorrect phase, and that nsight will align them with the center beam reference. Figure 12 T 1 T 2 T 3 6
7 nsight Imaging adjusts for this phase misalignment by bringing the out-of-phase wave front into alignment with the center wave fronts, as shown in Figure 13. Destructive interference Constructive reinforcement cancellation. This is akin to the principle of frame averaging in ultrasound. But instead of combining one entire frame image with another, the signals received from an image field are adjusted and combined individually, point-by-point, throughout the entire image field. As a result, images are clearer and sharper than ever before. In the near field example of Figure 13, this means that detailed near field structures like muscular striations, highly superficial breast lesions, and liver capsule nodules can be resolved and viewed like never seen before. Figure 13 Alignment of wave fronts It does not do this by any adjustment of the transmit waveform itself because, as stated above, once the transmit wave front is launched into the body it cannot be altered by the ultrasound system. The phase alignment is done by operating on the stored received signals which are affected by the round-trip beam profile and system delays, which causes the round-trip signals to appear as if they experienced dynamic transmit focusing. Multiple improvements in resolution and detail It is seen in Figure 13 that several beneficial effects are produced by nsight Imaging. First, the three transmit wave fronts are all in phase on the scanline as shown by the yellow circle. Thus, the signals from all three beams constructively reinforce each other and will additively combine, producing a clearer, stronger signal precisely on the scanline being formed by the beamformer. On either side of the scanline it is seen that the three wave fronts are out of phase with each other. Consequently they will destructively interfere with each other away from the scanline axis. As a result, the desired scanline signals will be enhanced by the positive reinforcement while unwanted adjacent signals and noise experience Penetration also improves nsight Imaging brings improvement in penetration in the far field as shown in Figure 14. In the far field, the diverging wave fronts are similarly brought into phase coherence at the point indicated by the yellow circle. Since the signals from three transmit-receive events are being combined, there is inherent pulse averaging. Weak signals from the far field are enhanced multiple times when the signals are combined, providing increased clarity and resolution at the greater depths of field. As in the near field example, off-axis signals are out of phase which results in cancellation. Noise is averaged out and greater penetration with improved resolution results. A consequence of this far field performance is that far field artifacts common with previous ultrasound such as ventricular clutter are greatly diminished with nsight Technology. Enhanced penetration with inherent pulse averaging Figure 14 Enhanced penetration with pulse averaging 7
8 The dynamically refocused reconstructed transmit beam is now precisely focused throughout the full depth of field and has the same narrow, uniform beam pattern exhibited by dynamically focused receive beams. The product of the two beam profiles is an expected pencil-like round-trip beam profile as shown in Figure 7. Another benefit of nsight Imaging is that critical placement of a transmit focal depth is no longer required. The image will be more fully focused over the entire depth of field and less dependent on the set focal point. Complete image focusing attained in the past with multiple focal zones in zone focusing is achieved automatically with nsight Imaging without suffering the significant drop in frame rate experienced by multiple transmit zone focusing techniques. The concerns involved in focal zone setting are now relegated to the prior history of conventional diagnostic ultrasound. No loss in temporal resolution If the previous processing were done with three conventional transmit-receive cycles as Figure 11 illustrates, frame rate would decrease by a factor of three. This deleterious effect does not occur with nsight Imaging through its use of high-order multi-line transmission and reception as illustrated by Figure 15. This example shows four transmit-receive cycles with four transmit beams T1, T2, T3, and T4. The transmit beam centers are shifted along the array from one event to another as shown in the drawing. T 1 T 2 T 3 T 4 Figure 15 High order multi-line 8
9 Instead of receiving a single receive beam as in Figure 11, sixteen (or more) receive beams are acquired in response to each transmit beam. The black arrow shows that these four transmit-receive events will result in a scanline formed from four transmit-receive events, which is both transmit-focused and receive-focused. It can also be seen in Figure 15 that three other fully focused beams can also be formed at this time, one in the scanline position above the black arrow and the two below. Thus, for each incremental transmit event, four fully focused scanlines will be formed with significantly faster frame rates than conventional single beam acquisition and with significantly greater resolution and detail than that provided by the three-cycle example of Figures This means that frame rate and temporal resolution are improved four-fold over the conventional approach. The need to trade off temporal resolution for spatial resolution and image detail is relegated to the past with Philips patented nsight Imaging. Massive parallel processing comes to ultrasound The precision beam reconstruction method of nsight Imaging also requires the addition of powerful processing. In order to perform virtually instantaneous beam reconstruction, a new architecture needed to be developed. The result is proprietary hardware and software able to perform massive parallel processing of data. The unique nsight Imaging architecture is able to perform 450 x bit multiply-accumulates per second. To put this into perspective, this would equate to: ~ 5000x Cray-1 supercomputers working in unison ($5-8M each in 1980) ~ 75x high-end multicore DSPs working in unison (3 cores, 2 32-bit MACs per core, 1 GHz) ~ 25x high-end gamer desktop PCs working in unison (dual socket, 8 cores per socket, 2 GHz) 9
10 Reaching its full potential A powerful new beamformer architecture capable of resolving fine detail throughout an entire image field cannot realize its full potential unless the system signal path preceding the beamformer delivers the highest quality signal information. Hence, Philips has designed the front end of the EPIQ system to enable the reception and delivery of signals with greater image information content than in the past. The initial acquisition signal path has been redesigned with new front-end analog circuitry to provide increased acoustic signal bandwidth, which is especially important for maximizing sensitivity and penetration at the highest imaging frequencies. This signal path has also been redesigned with increased dynamic range, which enables detection of a wider range of acoustic signal amplitudes without distortion for greater resolution of fine detail. With a new signal path providing signals of increased signal bandwidth and dynamic range, nsight Imaging attains unrivaled levels of performance. Clinical breakthroughs Philips nsight Imaging produces truly extraordinary clinical results across multiple applications. Improvements in all aspects of imaging can be realized as seen in the table below. Technical feature Precision real time beam reconstruction Fewer transmit operations with increased frame rate Ultra-wide system bandwidth and exceptional signal-to-noise ratio Clinical impact Superb spatial resolution Enhanced contrast resolution Outstanding tissue uniformity Fewer artifacts and reduced image clutter Facilitates 2D survey scan with less artifacts Excellent B-mode IQ in color flow modes Enhanced SonoCT and harmonic modes Outstanding CEUS performance Superb 3D/4D IQ and frame rate Outstanding CEUS performance Superb Doppler performance Exceptional high frequency performance Enhanced penetration and contrast resolution 10
11 nsight Imaging seeing is believing Philips EPIQ ultrasound system with nsight Imaging truly represents the next era in premium ultrasound with a new level of diagnostic certainty and clinical confidence. Along with the introduction of the most powerful architecture ever created for ultrasound, Philips EPIQ overshadows every premium ultrasound system that has come before with exceptional image quality, an unmatched user experience, and unheard of levels of adaptive intelligence. C9-2 imaging of a 14-week fetus reveals superb contrast resolution and delineation of structural detail. Outstanding spatial resolution and stunning uniformity seen with the C9-2 PureWave curved array transducer. High frequency imaging of the thyroid gland with the L18-5 linear array demonstrates superb detail and contrast resolution. This image of the carotid artery reveals a subtle dissection abnormality. Breast imaging with the L18-5 linear array transducer shows a small cyst with microcalcification. PureWave C10-3v transducer demonstrates excellent detail and contrast resolution with superb frame rate. X5-1 xmatrix transducer shows superb myocardial and valve definition. 11
12 Philips Healthcare is part of Royal Philips How to reach us Asia Europe, Middle East, Africa Latin America North America (toll free, US only) Please visit Koninklijke Philips N.V. All rights are reserved. Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication. Printed in The Netherlands * Jun 2013
Reliable versatility. Philips HD5 ultrasound system
Reliable versatility Philips HD5 ultrasound system Affordability and features in one Designed to perform as you need it Every day, your patients come to you for high-quality care. Now, there s an ultrasound
Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography
Philips CX50 CompactXtreme System A New Class of Compact Echocardiography Insights from compact ultrasound users Diagnostic confidence is paramount Ultrasound labs are busy and workflow efficiency is crucial
LOGIQ E9. Ultrasound agility has arrived.
LOGIQ E9 Ultrasound agility has arrived. We started with a blank sheet of paper. Nothing compromised the imaginative thinking that went into designing the new LOGIQ E9 premium ultrasound system. The sheet
SAMSUNG ULTRASOUND RS80A
Samsung Medison is a global leading medical device company. Founded in 1985, the company sells cutting-edge diagnostic ultrasound devices around the world in various medical fields. The company has attracted
A new era in premium vascular ultrasound. Philips EPIQ 5 ultrasound system
A new era in premium vascular ultrasound Philips EPIQ 5 ultrasound system The new challenges in global healthcare Unprecedented advances in premium ultrasound performance can help address the strains on
Ultrasound. Your versatile performer to depend on. Philips ClearVue ultrasound family for women s health care
Ultrasound ClearVue Your versatile performer to depend on Philips ClearVue ultrasound family for women s health care Over 70 countries More than 5,000 systems Touching 25 million lives every year For one
Improve your image. ClearVue 350 ultrasound system. The print quality of this copy is not an accurate representation of the original.
Improve your image ClearVue 350 ultrasound system Not yet available in the U.S. Smart and intuitive The system looks sleek from the outside, but it s what s on the inside that you ll value most. ClearVue
Exploring the value of EPIQ premium ultrasound
Exploring the value of EPIQ premium ultrasound Reducing cost of care through more definitive diagnoses Philips EPIQ offers a new level of premium performance ultrasound, and with it the opportunity to
A new era in premium ultrasound. Philips EPIQ 7 ultrasound system
A new era in premium ultrasound Philips EPIQ 7 ultrasound system The new challenges in global healthcare Unprecedented advances in premium ultrasound performance can help address the strains on overburdened
Company profile. Quote from the analyst of. Company Milestone. Feel the Difference. Tel: +86-755-26722860 Fax: +86-755-26722850
Company profile Since the establishment, SonoScape has committed to provide high standard medical equipments for healthcare of human beings. SonoScape specializes in the development and production of diagnostic
Improve your image. ClearVue 650 ultrasound system
Improve your image ClearVue 650 ultrasound system Introducing the ClearVue 650 The Philips ClearVue 650 ultrasound system is as versatile as it is elegant. Use it in a variety of settings for many types
A new way of thinking. Philips Sparq regional anesthesia and pain medicine ultrasound system
A new way of thinking Philips Sparq regional anesthesia and pain medicine ultrasound system Introducing Philips Sparq Breakthrough technology should not make your life more complex. Quite the opposite.
A new era in premium ultrasound. Philips EPIQ 5 ultrasound system
A new era in premium ultrasound Philips EPIQ 5 ultrasound system The new challenges in global healthcare Unprecedented advances in premium ultrasound performance can help address the strains on overburdened
Premium performance compact ultrasound. Philips CX50 CompactXtreme ultrasound system
Premium performance compact ultrasound Philips CX50 CompactXtreme ultrasound system Philips CX50 CompactXtreme system Now you can be confident in the data from your exams, including your most technically
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
Not just a new ultrasound, a new vision. Philips VISIQ ultrasound
Not just a new ultrasound, a new vision Philips VISIQ ultrasound Ultra mobile Wherever you need to deliver high-quality Ob/Gyn or abdominal ultrasound imaging, whether in traditional or remote locations,
The promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality
Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL
High-quality/High-resolution Digital Ultrasound Diagnostic Scanner
Hitachi Review Vol. 52 (2003), No. 4 207 High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Tsuyoshi Otake Toshihiko Kawano Takashi Sugiyama Tsuyoshi Mitake Shinichiro Umemura, Ph. D. OVERVIEW:
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare
GE Healthcare Vivid S6 Cardiovascular ultrasound system imagination at work Taking care of everyone. Strong performance and innovative miniaturization make the Vivid S6 a highly desirable system for a
Choose versatility. Philips HD11 XE ultrasound system
Choose versatility Philips HD11 XE ultrasound system Fully equipped The HD11 XE is ready to expand with your clinical needs. It is a complete, digital imaging system that delivers high-definition imaging
Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University
Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process
Ultrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System
GE Healthcare Vivid 7 Dimension Cardiovascular Ultrasound System New dime Introducing multi-dimensional and 4D imaging. The ability to incorporate multi-dimensional and 4D imaging into your clinical routine
Scanning Acoustic Microscopy Training
Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed
application note Directional Microphone Applications Introduction Directional Hearing Aids
APPLICATION NOTE AN-4 Directional Microphone Applications Introduction The inability to understand speech in noisy environments is a significant problem for hearing impaired individuals. An omnidirectional
Prepared by: Paul Lee ON Semiconductor http://onsemi.com
Introduction to Analog Video Prepared by: Paul Lee ON Semiconductor APPLICATION NOTE Introduction Eventually all video signals being broadcasted or transmitted will be digital, but until then analog video
Gage Applied Technologies
Copyright 2004 As seen in Evaluation Engineering/September 2004 1 Use of a PC-Based Digitizer in an Ultrasonic Flaw-Detection System Written by: Andrew Dawson, Ph.D. ABSTRACT: Use of modern high-speed
Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010
1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound
More than a Portable System
More than a Portable System General Imaging More than Mobile Setting a new standard: the Premium-performance Mobile A new class of systems: the Gold Standard in terms of quality and solutions meets the
A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)
chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy
Whitepaper. Image stabilization improving camera usability
Whitepaper Image stabilization improving camera usability Table of contents 1. Introduction 3 2. Vibration Impact on Video Output 3 3. Image Stabilization Techniques 3 3.1 Optical Image Stabilization 3
Always in touch. Philips PageWriter TC50 cardiograph
Always in touch Philips PageWriter TC50 cardiograph Intuitive and efficient You need to fit a lot into every day. That s why we ve fit a lot into the Philips PageWriter TC50 cardiograph. It s an affordable
The new generation in ECG interpretation
The new generation in ECG interpretation Philips DXL ECG Algorithm, Release PH100B The Philips DXL ECG Algorithm, developed by the Advanced Algorithm Research Center, uses sophisticated analytical methods
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Compactly Powerful. Ultrasound System UGEO PT60A CT-UGEO PT60A-CWW-CMI-0705-EN
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer, in 110
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging.
Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging. Ira O. Wygant a, Mustafa Karaman b, Ömer Oralkana, and Butrus T. Khuri-Yakub
Omni Antenna vs. Directional Antenna
Omni Antenna vs. Directional Antenna Document ID: 82068 Contents Introduction Prerequisites Requirements Components Used Conventions Basic Definitions and Antenna Concepts Indoor Effects Omni Antenna Pros
5 Factors Affecting the Signal-to-Noise Ratio
5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed
A New Radar Technology Broadband Radar Explained
A New Radar Technology Broadband Radar Explained by Bill Johnson The landscape of small boat radar has just changed. Until two months ago, all the radars for the leisure marine market worked in pretty
Three Dimensional Ultrasound Imaging
Three Dimensional Ultrasound Imaging Hans Torp/ Sevald Berg/Kjell Kristoffersen m/flere Department of circulation and medical imaging NTNU Hans Torp NTNU, Norway Acquisition Reconstruction Filtering Collecting
Expand your view. IntelliVue MX600, MX700, and MX800 bedside patient monitors
Expand your view IntelliVue MX600, MX700, and MX800 bedside patient monitors Trusted by more, for more Philips IntelliVue patient monitors are already the world s best selling, providing the accurate,
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
You ll love the Vue. Philips IntelliVue Information Center ix
You ll love the Vue Philips IntelliVue Information Center ix IT Director It has to it into our IT infrastructure and integrate easily with our EMR and HIS. Clinical Engineering Make it easy to support.
Leading the New Standards
Samsung Merdison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer, in 110
White paper. HDTV (High Definition Television) and video surveillance
White paper HDTV (High Definition Television) and video surveillance Table of contents Introduction 3 1. HDTV impact on video surveillance market 3 2. Development of HDTV 3 3. How HDTV works 4 4. HDTV
The ultrasound system that sets new standards
Samsung Medison is one of the world's leading researchers, developers and manufacturers of ultrasound and other medical imaging products. Founded in 1985, Samsung Medison was acquired by Samsung Electronics
FUNDAMENTALS OF MODERN SPECTRAL ANALYSIS. Matthew T. Hunter, Ph.D.
FUNDAMENTALS OF MODERN SPECTRAL ANALYSIS Matthew T. Hunter, Ph.D. AGENDA Introduction Spectrum Analyzer Architecture Dynamic Range Instantaneous Bandwidth The Importance of Image Rejection and Anti-Aliasing
Interference. Physics 102 Workshop #3. General Instructions
Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by
primary SURVEILLANCE 3D RADAR
AIR TRAFFIC MANAGEMENT AIRport & ROUTE primary SURVEILLANCE 3D RADAR Supplying ATM systems around the world for more than 90 years indracompany.com AIRport & ROUTE primary SURVEILLANCE 3D RADAR Latest
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Note monitors controlled by analog signals CRT monitors are controlled by analog voltage. i. e. the level of analog signal delivered through the
DVI Interface The outline: The reasons for digital interface of a monitor the transfer from VGA to DVI. DVI v. analog interface. The principles of LCD control through DVI interface. The link between DVI
White paper. CCD and CMOS sensor technology Technical white paper
White paper CCD and CMOS sensor technology Technical white paper Table of contents 1. Introduction to image sensors 3 2. CCD technology 4 3. CMOS technology 5 4. HDTV and megapixel sensors 6 5. Main differences
SentryScope. Achieving Ultra-high Resolution Video Surveillance through Linescan Camera Technology. Spectrum San Diego, Inc. 10907 Technology Place
SentryScope Achieving Ultra-high Resolution Video Surveillance through Linescan Camera Technology Spectrum San Diego, Inc. 10907 Technology Place San Diego, CA 92127 858 676-5382 www.sentryscope.com Introduction
Samsung Ultrasound H60
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer around
Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect
Objectives: PS-7.1 Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Illustrate ways that the energy of waves is transferred by interaction with
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative
BIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
A comparison of radio direction-finding technologies. Paul Denisowski, Applications Engineer Rohde & Schwarz
A comparison of radio direction-finding technologies Paul Denisowski, Applications Engineer Rohde & Schwarz Topics General introduction to radiolocation Manual DF techniques Doppler DF Time difference
Technical Information. Digital Signals. 1 bit. Part 1 Fundamentals
Technical Information Digital Signals 1 1 bit Part 1 Fundamentals t Technical Information Part 1: Fundamentals Part 2: Self-operated Regulators Part 3: Control Valves Part 4: Communication Part 5: Building
What is the difference between an equivalent time sampling oscilloscope and a real-time oscilloscope?
What is the difference between an equivalent time sampling oscilloscope and a real-time oscilloscope? Application Note In the past, deciding between an equivalent time sampling oscilloscope and a real
Direct and Reflected: Understanding the Truth with Y-S 3
Direct and Reflected: Understanding the Truth with Y-S 3 -Speaker System Design Guide- December 2008 2008 Yamaha Corporation 1 Introduction Y-S 3 is a speaker system design software application. It is
PCM Encoding and Decoding:
PCM Encoding and Decoding: Aim: Introduction to PCM encoding and decoding. Introduction: PCM Encoding: The input to the PCM ENCODER module is an analog message. This must be constrained to a defined bandwidth
Evolution of Satellite Communication Systems
Mathieu DERVIN Brussels, 6th May 2015 Brussels, May 2015 Agenda I. From Sputnik to wideband satellite services: The key technological evolutions II. Increase the satellite system capacity: A global system
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements. Application Note 1304-6
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements Application Note 1304-6 Abstract Time domain measurements are only as accurate as the trigger signal used to acquire them. Often
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION Introduction The outputs from sensors and communications receivers are analogue signals that have continuously varying amplitudes. In many systems
Laser-Optimized Fiber
FIBER FAQs Laser-Optimized Fiber Technical Resource: Tony Irujo Manager, Customer Technical Support FIBER FAQs Laser-Optimized Fiber As transmission speeds over optical fiber networks in the enterprise
Antenna Diversity in Wireless Local Area Network Devices
Antenna Diversity in Wireless Local Area Network Devices Frank M. Caimi, Ph.D. Kerry L. Greer Jason M. Hendler January 2002 Introduction Antenna diversity has been used in wireless communication systems
IntelliVue to go. IntelliVue MX40 wearable patient monitor
IntelliVue to go IntelliVue MX40 wearable patient monitor Freedom of movement Real time, real mobility Get ready to go. You ve never seen anything like the Philips IntelliVue MX40, which combines the benefits
STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI
STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI Institute of Fundamental Technological Research, Department
TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS
TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of
Realtime FFT processing in Rohde & Schwarz receivers
Realtime FFT in Rohde & Schwarz receivers Radiomonitoring & Radiolocation Application Brochure 01.00 Realtime FFT in Rohde & Schwarz receivers Introduction This application brochure describes the sophisticated
A Note on Rim Width, Tire Sensitivity, and Rim Depth in High Performance Bicycle Wheels.
A Note on Rim Width, Tire Sensitivity, and Rim Depth in High Performance Bicycle Wheels. History: In the late 1980 s aerodynamic bicycle wheels were beginning to evolve past the revolutionary disc wheels
Email: [email protected]
USE OF VIRTUAL INSTRUMENTS IN RADIO AND ATMOSPHERIC EXPERIMENTS P.N. VIJAYAKUMAR, THOMAS JOHN AND S.C. GARG RADIO AND ATMOSPHERIC SCIENCE DIVISION, NATIONAL PHYSICAL LABORATORY, NEW DELHI 110012, INDIA
Connected Care. Philips HeartStart connectivity and data management for EMS professionals
Connected Care Philips HeartStart connectivity and data management for EMS professionals Flexible and reliable, like Connected Care is the Philips approach to open data management solutions that simplify
Beyond Built-in: Why a Better Webcam Matters
Whitepaper: Beyond Built-in: Why a Better Webcam Matters How to Uplevel Your Ability to Connect, Communicate and Collaborate Using Your Laptop or PC Introduction The ability to virtually communicate and
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
Transformerless UPS systems and the 9900 By: John Steele, EIT Engineering Manager
Transformerless UPS systems and the 9900 By: John Steele, EIT Engineering Manager Introduction There is a growing trend in the UPS industry to create a highly efficient, more lightweight and smaller UPS
Experience true colors. Philips VIDI lighting technology
Experience true colors : A breakthrough in reaching astonishing color performance Digital front projectors are well accepted and widely used in many applications for front projection. Picture performance
Search keywords: Connect, Meeting, Collaboration, Voice over IP, VoIP, Acoustic Magic, audio, web conferencing, microphone, best practices
Title: Acoustic Magic Voice Tracker II array microphone improves operation with VoIP based Adobe Connect Meeting URL: www.acousticmagic.com By: Bob Feingold, President, Acoustic Magic Inc. Search keywords:
Lesson 26: Reflection & Mirror Diagrams
Lesson 26: Reflection & Mirror Diagrams The Law of Reflection There is nothing really mysterious about reflection, but some people try to make it more difficult than it really is. All EMR will reflect
Experiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
Zeiss 780 Training Notes
Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn
PICTURE PERFECT Ultrasound
PICTURE PERFECT Ultrasound Precision, control and an amazingly clear picture Aplio 400 delivers in every way. Ensuring every examination has an optimal result, it provides you with unsurpassed image quality,
Understanding astigmatism Spring 2003
MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest
MITSUBISHI ELECTRIC ANNOUNCES THE SUCCESSFUL DEVELOPMENT OF AN AIRBORNE Ku-BAND ANTENNA SUBSYSTEM FOR SATELLITE COMMUNICATIONS
FOR IMMEDIATE RELEASE No. 2330 Product Inquiries: Media Contact: Yutaka Kamada Oliver Cox Mitsubishi Electric Corporation Mitsubishi Electric Corporation Tel: +81-3-3218-2391 Tel: +81-3-3218-2346 [email protected]
W a d i a D i g i t a l
Wadia Decoding Computer Overview A Definition What is a Decoding Computer? The Wadia Decoding Computer is a small form factor digital-to-analog converter with digital pre-amplifier capabilities. It is
Direct Digital Amplification (DDX ) The Evolution of Digital Amplification
Direct Digital Amplification (DDX ) The Evolution of Digital Amplification Tempo Semiconductor 2013 Table of Contents Table of Contents... 2 Table of Figures... 2 1. DDX Technology Overview... 3 2. Comparison
FREQUENCY RESPONSE ANALYZERS
FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road
RF Measurements Using a Modular Digitizer
RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.
High-Resolution Doppler-Polarimetric FMCW Radar with Dual-Orthogonal Signals
High-Resolution Doppler-Polarimetric FMCW Radar with Dual-Orthogonal Signals Oleg Krasnov, Leo Ligthart, Zhijian Li, Galina Babur, Zongbo Wang, Fred van der Zwan International Research Centre for Telecommunications
B E S T - I N - C L A S S C O M P A C T M U L T I - P U R P O S E U L T R A S O U N D
B E S T - I N - C L A S S C O M P A C T M U L T I - P U R P O S E U L T R A S O U N D The SONOACE X1 Best-in-Class compact multi-purpose ultrasound imaging system from MEDISON. Introducing the SONOACE
Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)
Page 1 Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT) ECC RECOMMENDATION (06)01 Bandwidth measurements using FFT techniques
Digital Camera Imaging Evaluation
Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information
A passion for patients. Philips Patient Monitoring Services
A passion for patients Philips Patient Monitoring Services Our goal is to make sure you re highly satisfied with every part of your Philips service experience from your conversations with our account managers