1 (2004) 33, q 2004 The British Institute of Radiology RESEARCH Using a flat-panel detector in high resolution cone beam CT for dental imaging R Baba*,, K Ueda 2 and M Okabe 2 Central Research Laboratory, Hitachi, Ltd., Tokyo, Japan; 2 Research & Development Center, Hitachi Medical Corporation, Chiba, Japan Objectives: Cone beam CT (CBCT) requires a two-dimensional X-ray detector. In the several CBCT systems developed for dental imaging, detection has been by the combination of an X-ray image intensifier and charge-coupled device (CCD) camera. In this paper, we propose a new CBCT system in which the detector is of the flat-panel type and evaluate its performance in dental imaging. Methods: We developed a prototype CBCT that has a flat-panel-type detector. The detector consists of a CsI scintillator screen and a photosensor array. First, the flat panel detector and image intensifier detector were compared in terms of the signal-to-noise ratio (SNR) of projected images. We then used these data and a theoretical formula to evaluate noise in reconstructed images. Second, reconstructed images of a bar pattern phantom were obtained as a way of evaluating the spatial resolution. Then, reconstructed images of a skull phantom were obtained. Results: The SNR of the developed system was.6 times as high as that of a system with an image intensifier detector of equal detector pitch. The system was capable of resolving a 0.35 mm pattern and its field of view almost completely encompassed that of an image intensifier detector which is used in dentomaxillofacial imaging. The fine spatial resolution of the detector led to images in which the structural details of a skull phantom were clearly visible. Conclusions: The system s isotropically fine resolution will lead to improved precision in dental diagnosis and surgery. The next stage of our research will be the development of a flat panel detector system with a high frame acquisition rate. (2004) 33, doi: 0.259/dmfr/ Keywords: imaging, three-dimensional; tomography, X-ray computed; surgery, oral Introduction Cone beam computed tomography (CBCT) requires a twodimensional (2D) X-ray detector and a conical or pyramidal X-ray beam. Along with the 2D fluoroscopic and radiographic functions of the basic hardware, the acquisition of a full set (i.e. a set that covers 3608) of projection images during a single rotation scan plus computer processing allows us to reconstruct isotropically fine spatial resolution three-dimensional (3D) images. The pioneering CBCT systems were developed for application in angiography. 4 A system in which an X-ray image intensifier and charge-coupled device (CCD) television camera act as the *Correspondence to: Rika Baba, -280, Higashi-koigakubo, Kokubunji, Tokyo , Japan; [email protected] Received 22 April 2003; revised 2 August 2004; accepted 2 August 2004 detector has been developed; this has been shown to be an effective aid for diagnosis, surgical planning and explanation of orthopaedic and angiographic procedures to patients. 5 7 More recently, a system in which the patient sits on a chair and the X-ray tube and detector revolve in a horizontal plane around the patient have been developed for dentomaxillofacial imaging. 8,9 The system can produce thin slice images, maximum intensity projection (MIP) images, rendered images and panoramic images from an internal 3D representation, thus providing a comprehensive range of high quality dental images. In the several CBCT systems that have been developed for dental imaging, detection has been achieved by the combination of an X-ray image intensifier and CCD camera. 8 0 Recently, a flat-panel-type detector has begun to be used.,2
2 286 Flat-panel detector in cone beam CT RBabaet al In this paper, we propose a new CBCT system in which the detector is of the flat panel type and compare the performance of the two detector types in dental imaging. 3 Methods We developed two experimental setups for comparative evaluation of the two detector types in CBCT. Projected images of the subject on each detector were magnified by about.5 times. In the flat panel detector system, the X-ray source and detector are stationary and the subject is rotated in the horizontal plane by a turntable (Figure a). In the image intensifier detector system, the subject is stationary, and the X-ray source and detector revolve around the subject in the horizontal plane (Figure b). Equivalent images are obtained by both methods. The projected images are obtained during a 3608 rotation scan. Reconstruction processing is performed by using the Feldkamp algorithm with the standard Shepp and Logan filter. 4 In the flat panel detector (provided by Varian Medical Systems, Inc., Palo Alto, CA), the sensor elements are produced in a thin film of amorphous silicon (a-si) and a large detector is thus realisable. The flat panel detector consists of a CsI scintillator screen and a photosensor array (Figure 2a). The photosensor array consists of arrayed photodiodes and switching devices. The scintillator converts an X-ray beam into an optical signal, which the photodiode converts to an electrical signal, which is in turn read out by the switching device array. The flat panel detector does not generate veiling glare or distortion in the image, and has a smaller detector pitch than an image intensifier detector. 5 7 The flat panel detector provides us with a square field of view (FOV). However, the flat panel detector is unlike the image intensifier detector in that it does not have an optical iris as a direct means for adjustment; this makes it difficult to prevent saturation of the pixels. Correction for saturated pixels, defective pixels, offsets, and the sensitivity of the detector is applied in the new system and reduces the intensity of streak and ring artefacts in reconstructed images. The flat panel detector has three operating modes: a non-binning mode, a 2 2 binning mode, and a 4 4 binning mode (Table ). Figure Experimental set-ups (top view): (a) flat panel detector system; (b) image intensifier detector system Figure 2 Components of the detectors: (a) flat panel detector; (b) image intensifier detector All modes provide the same FOV, but the detector pitch of the 2 2 and 4 4 binning modes is twice and four times, respectively, the pitch in the non-binning mode. The image intensifier detector consists of a 9 00 X-ray image intensifier, optics and a CCD television camera (Figure 2b). The X-ray beam is converted to an optical signal by the input phosphor screen; this signal is in turn converted to electrons by the photocathode screen. Electrons are accelerated by the electric field inside the image intensifier and converted back to an optical signal at the output phosphor screen. The intensity of the optical signal is adjusted by the optical iris; the signal is then detected by the CCD. The read-out image includes geometrical distortion and a blurring component of veiling glare generated by the image intensifier. This detector s FOV is circular. In this system, correction for the offset and sensitivity of the sensors, and the distortion of the detector is applied to reduce the intensity of artefacts in the reconstructed image. The image intensifier detector has three imaging modes: a mode, 7 00 mode and 9 00 mode (Table ). The FOV differs with the mode. The detector pitch in the mode is the minimum, i.e. fits the minimum FOV, while that in the 9 00 mode is the maximum, i.e. fits the maximum FOV. The flat panel detector in the 2 2 binning mode has a mm detector pitch and acquires projected images that consist of bit pixels at 5 frames s 2.The image intensifier detector in the mode has a mm detector pitch and acquires projected images that consist of bit pixels at 30 frames s 2. The pitch of this detector in the 9 00 mode is mm pitch; acquisition is as in the mode. The flat panel detector has a larger FOV than the image intensifier detector in the mode or 7 00 mode; its FOV is, in fact, almost equal to that of the image intensifier detector in the 9 00 mode (Figure 3). The flat panel detector was evaluated in the 2 2 binning mode. This mode was selected because the pitch of the detector is then almost equal to that of the image intensifier detector in mode and its FOV is almost equal to that of the image intensifier detector in 9 00 mode. Thus, this mode provides the combination of small detector pitch and large FOV that is suitable for dentomaxillofacial imaging. First, the two detectors were compared in terms of the signal-to-noise ratio (SNR) of projected images. We then
3 Flat-panel detector in cone beam CT RBabaet al 287 Table Parameters of the detectors Flat panel detector Image intensifier detector Mode no binning 2 2 binning 4 4 binning Detector pitch (mm) Measured image matrix (pixel) Detector field of view (mm) w22 w76 w224 ADC bit number (bit) Frame rate (frame s 2 ) ADC, apparent diffusion coefficient Figure 3 Field of view of the detectors used these data and a theoretical formula to evaluate noise in reconstructed images. Second, reconstructed images of a bar pattern phantom were obtained as a way of evaluating the spatial resolution. Here, the image obtained by the flat panel detector was compared with that obtained by the image intensifier detector in mode, because this mode provides the smallest pitch available to the image intensifier detector. Reconstructed images of a skull phantom were then obtained. Here, images obtained by the flat panel detector were compared with those obtained by the image intensifier detector in 9 00 mode, since this mode provides the largest FOV available to the image intensifier detector, and is capable covering the whole skull. Results Noise in reconstructed images Projected images from the flat panel detector in 2 2 binning mode (0.254 mm detector pitch, 5 frames s 2 frame rate) and the image intensifier detector in mode (0.239 mm detector pitch, 30 frames s 2 frame rate) were evaluated for noise. The SNR was measured at a 20 kv X-ray tube voltage (Figure 4). Measured SNR values for the former and latter detectors are plotted as O and, respectively. The horizontal axis shows X-ray dose per pixel as measured at the surface of the detector. These measured data approximate the equation for the SNR of each detector. The equation consists of a first term that is proportional to the X-ray dose incident on the detector and a second term proportional to the square of this X-ray dose. The SNR of the flat panel detector calculated by using the equation is plotted as a solid line. The SNR of the image intensifier detector as calculated by using the equation is plotted as a dashed line. The lines fit the measured data well. Figure 4 shows that the projected image from the flat Figure 4 panel detector has an SNR.72 times that of the image intensifier detector. According to the Feldkamp algorithm, the noise of a reconstructed image is given by Equation (). vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi 8 a X sd ¼ 2 n X <! 9 u = t 2 wðkþ n : 2 c SNR i ; i¼ k¼2 vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi! u a X ¼ 2 2 t wðkþ 2 Xn c ðþ n SNR k¼2 i¼ i Here, sd means the standard deviation of the voxels of the reconstructed image, a is the detector pitch as projected on the subject (mm), n the number of projected images, SNR i the signal-to-noise ratio of the ith projected image, w(k) the convolution filter and c is the coefficient for transformation from X-ray absorption coefficient (mm 2 ) to CT number (HU). When the Shepp and Logan filter with linear interpolation is in use, the convolution filter is as given by Equation (2). X k¼2 Signal-to-noise ratio (SNR) of projected images wðkþ 2 ¼ ð a Wðf Þ 2 df 2 ¼ ð a 2 a 2 2 a a 2 ðp a f Þ 4 sin6 ðp a f Þdf ¼ 0:440 2 a 4 ð2þ
4 288 Flat-panel detector in cone beam CT RBabaet al Here, is the coefficient obtained for this filter. Substituting Equation (2) into Equation () gives us Equation (3). vffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi a 2 sd ¼ n 0:440 2a 4! u 2 t n Xn c ð3þ SNR i¼ i When the subject is assumed to be a cylinder, the SNR of the projected image is as given by Equation (4).! 2 n Xn 2 ¼ ð4þ SNR i¼ i SNR Here, SNR indicates the signal-to-noise ratio of the projected image. With Equation (4), Equation (3) becomes Equation (5). rffiffiffiffiffiffiffiffi 0:440 c sd ¼ ð5þ 2 n a SNR The noise of the reconstructed image is given by Equation (5), i.e. is inversely proportional to the detector pitch and SNR of the projected image. Table shows that the detector pitch of the flat panel detector is.06 times that of the image intensifier detector. In that case, Figure 4 shows that the projected images of the flat panel detector have SNR values.72 times as high as use of the image intensifier detector. As the SNR is proportional to the detector pitch, the SNR of the flat panel detector when it has the same pitch as the image intensifier detector will be.62 times that of the image intensifier detector. Thus, given equal detector pitch, the reconstructed image of the flat panel detector is presumed to have a noise level.62 times lower than that of the image intensifier detector. Spatial resolution of reconstructed images We evaluated the spatial resolution of reconstructed images obtained by the flat-panel detector in 2 2-binning mode (0.254 mm detector pitch, 5 frames s 2 frame rate) and the image intensifier detector in mode (0.239 mm detector pitch, 30 frames s 2 frame rate). The spatial resolution was evaluated by measuring a high-contrast bar pattern in a 65 mm diameter cylindrical water phantom. The pattern was set in a plane perpendicular to the axis of rotation for the evaluation of axial images and in a plane parallel to the same axis for the evaluation of coronal images. When the detector pitch of the flat panel detector was mm, a 0.35 mm bar pattern was resolved in both axial and coronal images extracted from the respective reconstructed images (20 kv tube voltage, mas total dose, 576 frame projection, 0.69 mm voxel pitch) (Figure 5a, b). The 0.35 mm pattern was also resolved in the reconstructed image of the image intensifier detector with a detector pitch of mm (Figure 5c). The respective reconstructed images had roughly equal spatial resolution. CBCT using the flat panel detector had isotropic spatial resolution equal to that of the image intensifier detector with the smallest detector pitch, i.e. in the imaging mode that is used for dentomaxillofacial imaging. Skull phantom imaging A skull phantom in water was imaged by the flat panel detector system in 2 2-binning mode (0.254 mm detector pitch, 5 frames s 2 frame rate) and by the image intensifier detector system in 9 00 mode (0.438 mm detector pitch, 30 frames s 2 frame rate). The reconstructed image obtained with the flat panel detector system had a cylindrical FOV with 0.25 mm voxel pitch. The reconstructed image obtained with the image intensifier detector system had a globular FOV with mm voxel pitch. Reconstructed images covered the skull. In the flat panel detector system, correction for defective or saturated pixels and the offset, and sensitivity of detector elements was applied to the projected images, so that neither streak nor ring artefacts were found in the reconstructed images (20 kv tube voltage, 24.8 mas total dose, 288 frame projection) (Figure 6a). The details of the skull s structures were more clearly depicted by the flat panel detector system than by the image intensifier detector system (Figure 6b, c). The reconstructed images obtained by the flat panel detector system had high resolution on all planes (axial, coronal and sagittal, Figure 6a and Figure 7a, b). The surface of the bone appears smooth in the volume rendered image because this system had a high spatial resolution and little blurring (Figure 7c). Even the thin bone in the nasal cavity is distinguishable. Conclusions and discussion The evaluation showed that, given equal detector pitch, reconstructed images obtained with the flat panel detector Figure 5 Three-dimensional images of a high-contrast pattern: (a) coronal image by flat panel detector (2 2-binning mode); (b) axial image by flat panel detector (2 2-binning mode); (c) axial image by image intensifier detector ( mode)
5 Flat-panel detector in cone beam CT RBabaet al 289 Figure 6 Axial images of a skull phantom: (a) flat panel detector system (2 2-binning mode); (b) flat-panel detector system (2 2-binning mode); (c) image intensifier detector system (9 00 mode) Figure 7 Three-dimensional images of a skull phantom obtained by the flat panel detector system: (a) coronal image; (b) sagittal image; (c) volume rendered image have less noise than those obtained with the image intensifier detector. The flat panel detector system we developed had spatial resolution equal to that of the image intensifier detector system in its smallest detector pitch mode, and our system s FOV was almost equal to that of the image intensifier detector system at its largest FOV, that is, in the operating mode used for dentomaxillofacial imaging. The high quality of reconstructed images obtained with the CBCT system using a flat panel detector offers improved precision in dentistry, for both diagnosis and surgery. The next stage of our research will be the development of a flat panel detector CBCT with a high frameacquisition rate. We also intend to improve fineness of contrast resolution as this will be important if we are to expand the application of CBCT. We are currently evaluating a new prototype CBCT in which the detector is of this type.
6 290 Flat-panel detector in cone beam CT RBabaet al References. Ritman EL, Kinsey JH, Robb RA, Gilbert BK, Harris LD, Wood EH. Three-dimensional imaging of heart, lungs, and circulation. Science 980; 20: Ning R, Kruger RA. Computer simulation of image intensifier-based computed tomography detector: vascular application. Med Phys 988; 5: Saint-Felix D, Picard C, Ponchut C, Romeas R, Rougee A, Troussert Y. Three dimensional X-ray angiography: first in vivo results with a new system. Proc SPIE 993; 897: Ning R, Wang X, Conover DL, Tang X. An image intensifier-based volume tomographic angiography imaging system. Proc SPIE 997; 3032: Baba R, Ueda K, Ueki H, Umetani K. Ellipsoid scan: chest cone-beam CT with a large ellipsoidal view field using a 6-inch X-ray image intensifier. Proc SPIE 999; 3032: Baba R, Ueda K, Kawai H, Kuba A, Takagi H. Development of threedimensional imaging method with cone-beam X-ray. MEDIX 999; 3: Baba R, Ueda K, Kuba A, Kohda E, Shiraga N, Sanmiya T. Development of a subject-standing-type cone-beam computed tomography for chest and orthopedic imaging. Front Med Biol Eng 200; : Seo K, Yamamoto K, Ueno K, Tanaka K, Matsuoka M, Okabe K, et al. Development of dentomaxillofocial conebeam X-ray CT system model CB MercuRay. MEDIX 2002; 37: Maki K, Usui T, Kubot M, Nakano H, Shibasaki Y, Araki K, et al. Application of cone-beam X-ray CT in dento-maxillofacial region. Proc CARS 2002; Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of a compact computed tomographic apparatus for dental use. Dentmaxillofac Radiol 999; 28: Sukovic P, Brooks S, Perez L, Clinthorne NH. DentoCATe A novel design of a cone beam CT scanner for dentomaxillofacial imaging: introduction and preliminary results. Proc CARS 200; Sukovic P. Cone beam computed tomography in craniofacial imaging. COAST 2002; 9 (Abstr). 3. Baba R, Ueda K, Kadomura T. High-resolution FPD cone-beam CT for dental and orthopedic surgery. Radiology 2002; 225(P): 94 (Abstr). 4. Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am A 984; : Colbeth RE, Allen MJ, Day DJ, Gilblom DL, Harris R, Job ID, et al. Flat panel imaging system for fluoroscopy applications. Proc SPIE 998; 3336: Yamada S, Umazaki H, Takahashi A, Honda M, Shiraishi K, Rudin S, et al. Image quality evaluation of a selenium-based flat-panel digital X-ray detector system based on animal studies. Proc SPIE 2000; 3977: Granfors PR, Aufrichtig R. Performance of a 4 4-cm 2 amorphous silicon flat panel X-ray detector for radiographic imaging applications. Med Phys 2000; 27:
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
DENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
3D Accuitomo FPD XYZ Slice View Tomograph. Super-High Resolution Images of Region of Interest
3D Accuitomo FPD XYZ Slice View Tomograph. Super-High Resolution Images of Region of Interest Thinking ahead. Focused on life. Thinking ahead. Focused on life. The new era of user-friendly, 3-dimensional
ENGLISH. Planmeca ProMax 3D concept
ENGLISH Planmeca ProMax 3D concept Profound understanding of anatomy Planmeca ProMax 3D concept is an intelligent and multipurpose X-ray unit series designed to obtain complete information on patient anatomy
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis
Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of
Experimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
Modern Fluoroscopy Imaging Systems
Modern Fluoroscopy Imaging Systems Eric Gingold, PhD Thomas Jefferson University, Philadelphia, PA SUMMARY Fluoroscopy, or real-time projection X-ray imaging, has been in clinical use since shortly after
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
X-RAY IMAGING Emerging Digital Technology - CMOS Detectors
Application Note Case Study Technology Primer White Paper X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Image Sensors X-Ray DETECTORS Scanners Image Processing Custom Solutions In all domains
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
How To Improve Your Ct Image Quality
Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board
Dental Radiography Core Subject. Digital Radiography
Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of
Users Group Meeting 2011 Northeastern Society of Orthodontists 89th Annual Meeting November 11-14, 2010 Hilton Montréal Bonaventure Montréal, Quebec, Canada 14 Lecture Schedule Saturday continued w Doctor
HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950
HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple
I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
APPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS
APPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS V.A. Popovich 1, W. Verwaal 2, M. Janssen 1, I. J. Bennett 3, I.M.Richardson 1, 1. Delft University of Technology, Department of Materials
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Introduction to Medical Imaging. Lecture 11: Cone-Beam CT Theory. Introduction. Available cone-beam reconstruction methods: Our discussion:
Introduction Introduction to Medical Imaging Lecture 11: Cone-Beam CT Theory Klaus Mueller Available cone-beam reconstruction methods: exact approximate algebraic Our discussion: exact (now) approximate
Development of High Speed Oblique X-ray CT System for Printed Circuit Board
計 測 自 動 制 御 学 会 産 業 論 文 集 Vol. 6, No. 9, 72/77 (2007) Development of High Speed Oblique X-ray CT System for Printed Circuit Board Atsushi TERAMOTO*, Takayuki MURAKOSHI*, Masatoshi TSUZAKA**, and Hiroshi
Digital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
Signal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
Scanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK
vii LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK LIST OF CONTENTS LIST OF TABLES LIST OF FIGURES LIST OF NOTATIONS LIST OF ABBREVIATIONS LIST OF APPENDICES
Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?
Disclosures Understanding IAC CT Accreditation Serve as one of two AAPM representatives to IAC CT Board of Directors, Serve as IAC representative to AAPM for focus group on accreditations Provide 3-hour
Resolution Enhancement of Photogrammetric Digital Images
DICTA2002: Digital Image Computing Techniques and Applications, 21--22 January 2002, Melbourne, Australia 1 Resolution Enhancement of Photogrammetric Digital Images John G. FRYER and Gabriele SCARMANA
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
http://dx.doi.org/10.1117/12.906346
Stephanie Fullerton ; Keith Bennett ; Eiji Toda and Teruo Takahashi "Camera simulation engine enables efficient system optimization for super-resolution imaging", Proc. SPIE 8228, Single Molecule Spectroscopy
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
A Grangeat-type half-scan algorithm for cone-beam CT
A Grangeat-type half-scan algorithm for cone-beam CT Seung Wook Lee a) CT/Micro-CT Lab., Department of Radiology, University of Iowa, Iowa City, Iowa 52242 and Department of Nuclear and Quantum Engineering,
5 Factors Affecting the Signal-to-Noise Ratio
5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Endoscopy & X-Ray Image Sensors Markets
Endoscopy & X-Ray Image Sensors Markets IMAP December 2012 2012 Le Quartz - 75 cours Emile Zola - 69100 LYON-Villeurbanne Tel: +33 472 83 01 80 - Fax: +33 472 83 01 83 Web: www.yole.fr Agenda Overview
Polarization of Light
Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights
Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT
MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT C. Reinhart, C. Poliwoda, T. Guenther, W. Roemer, S. Maass, C. Gosch all Volume Graphics GmbH, Heidelberg, Germany Abstract:
Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected] X-Ray
Endoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1
Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Comparison of Amorphous Selenium and Amorphous Silicon Flat-Panel Detectors November 20, 2012 COMPARISON OF FLAT-PANEL DETECTORS 2 Abstract
Digital Fluoroscopic Imaging: Acquisition, Processing & Display
Digital Fluoroscopic Imaging: Acquisition, Processing & Display J. Anthony Seibert, Ph.D. University of California Davis Medical Center Sacramento, California Outline of presentation Introduction to digital
Shape Measurement of a Sewer Pipe. Using a Mobile Robot with Computer Vision
International Journal of Advanced Robotic Systems ARTICLE Shape Measurement of a Sewer Pipe Using a Mobile Robot with Computer Vision Regular Paper Kikuhito Kawasue 1,* and Takayuki Komatsu 1 1 Department
Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected].
Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected] Outline Motivation why do we need GPUs? Past - how was GPU programming
Video Camera Image Quality in Physical Electronic Security Systems
Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems In the second decade of the 21st century, annual revenue for the global
ENGLISH. Planmeca ProMax 3D Mid
ENGLISH Planmeca ProMax 3D Mid Genuine all-in-one unit Planmeca ProMax 3D Mid is a genuine all-in-one CBVT (Cone Beam Volumetric Tomography) unit including 3D imaging, digital panoramic, digital cephalometric
Industrial X-ray for Nondestructive Testing Unrestricted Siemens AG 2014. All rights reserved
Overview of X-ray Technology and Competence offered by Corporate Technology Industrial X-ray for Nondestructive Testing Nondestructive Testing (NDT) with X-rays: Our offer at a glance High-tech X-ray lab
European Academy of DentoMaxilloFacial Radiology
European Academy of DentoMaxilloFacial Radiology Framework for Specialist Training in Dental and Maxillofacial Radiology Background The scope of DentoMaxilloFacial Radiology DMFR (Dental and Maxillofacial
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology
product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905
The phantom portion of the American College of Radiology ACR Computed Tomography CT accreditation program: Practical tips, artifact examples, and pitfalls to avoid a Cynthia H. McCollough b) and Michael
High Quality Image Magnification using Cross-Scale Self-Similarity
High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg
WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter?
WHITE PAPER www.basler-ipcam.com Are More Pixels Better? The most frequently asked question when buying a new digital security camera is, What resolution does the camera provide? The resolution is indeed
Implementing and Using the EMVA1288 Standard
Implementing and Using the EMVA1288 Standard A. Darmont *, J. Chahiba, J.-F. Lemaitre, M. Pirson, D. Dethier Aphesa, Rue de Lorcé, 39, 4920 Harzé, Belgium ABSTRACT The European Machine Vision Association
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
Experiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
Digital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
Concepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful
Understanding Line Scan Camera Applications
Understanding Line Scan Camera Applications Discover the benefits of line scan cameras, including perfect, high resolution images, and the ability to image large objects. A line scan camera has a single
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Cone Beam Imaging Why all the Fuss about the X-ray Dose? D.A. Miles BA, DDS, MS, FRCD(C) Dip. ABOM, Dip. ABOMR. Introduction
Cone Beam Imaging Why all the Fuss about the X-ray Dose? D.A. Miles BA, DDS, MS, FRCD(C) Dip. ABOM, Dip. ABOMR Introduction The newest imaging modality to become available to dentists is making its impact
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Twelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
An iterative approach to the beam hardening correction in cone beam CT
An iterative approach to the beam hardening correction in cone beam CT Jiang Hsieh a) Applied Science Laboratory, GE Medical Systems, Milwaukee, Wisconsin 53201 Robert C. Molthen Department of Biomedical
A System for Capturing High Resolution Images
A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: [email protected]
Overview of Digital Detector Technology
Overview of Digital Detector Technology J. Anthony Seibert, Ph.D. Department of Radiology University of California, Davis Disclosure Member (uncompensated) Barco-Voxar Medical Advisory Board ALARA (CR
What is Cone-Beam CT and How Does it Work?
Dent Clin N Am 52 (2008) 707 730 What is Cone-Beam CT and How Does it Work? William C. Scarfe, BDS, FRACDS, MS a, *, Allan G. Farman, BDS, PhD, DSc, MBA b a Department of Surgical/Hospital Dentistry, University
Intelligent Measurement and Diagnostic Techniques for Non-destructive Inspections
Intelligent Measurement and Diagnostic Techniques for Non-destructive Inspections 296 Intelligent Measurement and Diagnostic Techniques for Non-destructive Inspections Fuminobu Takahashi, D. Eng. Shigeru
ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS
ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS INFN-Laboratori Nazionali del Sud Via S. Sofia 44, Catania, Italy Patient positioned for treatment System under consideration (experimental
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP
Now Powered by LightField FEATURES BENEFITS Back Illuminated CCD (248B) For highest sensitivity Front illuminated CCD (248F) Affordable technology for moderate light level applications Ultra low noise
MEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal?
The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal? Don Williams Eastman Kodak Company, Rochester, NY, USA, 14650-1925 ABSTRACT Inexpensive and easy-to-use linear
Geometry Notes VOLUME AND SURFACE AREA
Volume and Surface Area Page 1 of 19 VOLUME AND SURFACE AREA Objectives: After completing this section, you should be able to do the following: Calculate the volume of given geometric figures. Calculate
5-Axis Test-Piece Influence of Machining Position
5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
High-resolution Imaging System for Omnidirectional Illuminant Estimation
High-resolution Imaging System for Omnidirectional Illuminant Estimation Shoji Tominaga*, Tsuyoshi Fukuda**, and Akira Kimachi** *Graduate School of Advanced Integration Science, Chiba University, Chiba
Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35
Development of SCENAIA New Version Software Takayuki Kadomura Toshio Sakamoto Kana Tanaka Takatsugu Ito Yuuko Nishimura Naomi Maekawa MEDIX VOL. 62 P.32 P.35 Development of SCENAIA New Version Software
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
Automation of a CT Acquisition: A System-based User-Support. 1. Introduction. Bärbel KRATZ 1, Frank HEROLD 1, Malte KURFISS 1
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Automation of a CT Acquisition: A System-based User-Support More Info at Open Access Database
CT Image Reconstruction. Terry Peters Robarts Research Institute London Canada
CT Image Reconstruction Terry Peters Robarts Research Institute London Canada 1 Standard X-ray Views Standard Radiograph acquires projections of the body, but since structures are overlaid on each other,
ENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
A method of generating free-route walk-through animation using vehicle-borne video image
A method of generating free-route walk-through animation using vehicle-borne video image Jun KUMAGAI* Ryosuke SHIBASAKI* *Graduate School of Frontier Sciences, Shibasaki lab. University of Tokyo 4-6-1