Instrumental neutron activation analysis (INAA) for steel analysis and certification
|
|
|
- Branden Anderson
- 10 years ago
- Views:
Transcription
1 NUKLEONIKA 2011;56(4): ORIGINAL PAPER Instrumental neutron activation analysis (INAA) for steel analysis and certification Halina Polkowska-Motrenko, Ewelina Chajduk, Bożena Danko Abstract. Instrumental neutron activation analysis (INAA) procedure of elemental steel analysis has been elaborated. The reliability of results obtained by the described procedure has been proved in the frame of different proficiency testing/interlaboratory comparison (PT/ILC) programmes, in which the Institute of Nuclear Chemistry and Technology (INCT, Warsaw) took part since This work summarizes participation of INCT in these exercises and demonstrates the usefulness of INAA to the certification of steel reference materials. Key words: steel analysis neutron activation analysis (NAA) proficiency testing/interlaboratory comparison (PT/ILC) certification of steel reference materials Introduction H. Polkowska-Motrenko, E. Chajduk, B. Danko Laboratory of Nuclear Analytical Methods, Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., Warsaw, Poland, Tel.: , Fax: , Received: 24 January 2011 Accepted: 9 May 2011 Addition of trace elements (intended or as impurities) in steel or alloys can completely change the properties of materials; thus, there is a necessity to have reliable analytical techniques to determine major and trace components in this type of material. Also, to solving problems with quality control/quality assurance (QC/QA) in this subject, appropriate certified reference materials should be available in the wide range. Several methods have been proposed for quantitative analysis of steel samples: inductively coupled plasma (ICP-OES) [1, 4], inductively coupled plasma mass spectrometry (ICP-MS) [5, 8], atomic absorption spectrometry (AAS) [11], X-ray fluorescence (XRF) [1, 6], and neutron activation analysis (NAA) [1, 2, 7, 10]. As a rule, all of these techniques possess some limitation like required sample dissolution or appropriate sample mass. These requirements like preliminary separation get worse the accuracy. For accuracy reason, it is advisable to apply more than one method for analysis. INAA is one of the preferred methods because it provides information on a large number of elements simultaneously. Another important feature of INAA is that it is free from the problems connected with digestion of difficult matrices and potential contamination of a sample from reagents. Moreover, by changing irradiation and cooling parameters as well as measurement time, the effect of the interfering radionuclides on the results of analysis of particular elements can
2 312 H. Polkowska-Motrenko, E. Chajduk, B. Danko be decreased. Because of these advantages, INAA is used as a microanalytical method in metallic material analysis [1, 2, 16]. The position of NAA among the analytical techniques is still strong despite some weaknesses like the necessity of having access to a nuclear reactor, long analysis time, etc. and NAA is, with no doubt, a valuable method in inorganic trace analysis, especially important in certification of reference materials [9, 12]. During recent years, it has been proved that instrumental NAA (INAA), because of its unique possibilities, can meet Consultative Committee for Amount of Substance (CCQM) criteria for a primary method of measurement (PMM) [14 17]. INAA procedure including irradiation conditions, decay and measurement times, interpretation of the results for the elemental analysis of metallurgical samples has been described. Elaborated procedure gives accurate results for several elements in a different concentration range with very low combined uncertainty, what has been confirmed by participation in proficiency testing/interlaboratory comparison (PT/ILC) exercises concerning elemental analysis of steel material by neutron activation analysis. Experimental Sample irradiation Different certified steel reference materials were applied for analysis. All used stock standards solutions were obtained from appropriate amounts of special purity elements or oxides by weighing an appropriate amount, dissolving in aqua regia or nitric acid, and weighing the solution obtained. Standard solutions for irradiation were prepared by weighing aliquots of a standard solution in a polyethylene (PE) capsule and air dried in a laminar flow box before encapsulation. The package containing samples sandwiched between standards, certified reference materials (CRMs) and blanks, was irradiated in the reactor MARIA (Otwock/ Świerk, Poland) for 5 min at a thermal neutron flux density of cm 2 s 1. The irradiation, cooling and counting times were chosen according to half-lives of the determined elements. Table 1 presents irradiation and measurements details. Apparatus To prepare samples, CRMs and standards for irradiation, calibrated analytical and micro-analytical balances Sartorius MC5 and Sartorius BP221S were used. To perform gamma-ray spectroscopic measurements the following spectrometer was used: a 255 cm 3 high purity germanium (HPGe) well type detector (Canberra) with associated electronics (resolution 2.15 kev for 1332 kev 60 Co line, efficiency ca. 55%), coupled to a multichannel analyzer and spectroscopy software Gennie-2000 (Canberra). Results and discussion Table 1. Irradiation and measurements details for stainless steel analysis by INAA In the INCT instrumental neutron activation analysis is mostly used for the elemental analysis of various unique types of materials like tissues, meteorites etc. Due to lack of access to a pneumatic transfer system (rabbit) in the nuclear reactor, elements forming short-lived isotopes like: Al, Ti, V have not been determined. To show usefulness of INAA in steel analysis, the INCT took part in interlaboratory comparisons (ILC) and proficiency tests (PT) in the frame of INAA European Laboratories Network [2]. The first exercise was conducted in 2004 by the Swedish Institute for Corrosion and Metals Research (SIMR), where five INAA laboratories from four European countries: Poland, Romania, Belgium and Hungary participated in the evaluation of elemental composition of high alloy steel (SS1) [1, 2, 13]. Figure 1 illustrates the results obtained in the INCT in comparison to the assigned values calculated by the providers from results obtained by atomic spectrometry techniques, and to the arithmetic mean of all INAA results. As can be seen, results obtained in the INCT are in good agreement with both values, however a higher difference between the Laboratory results and the assigned value is observed. Only the difference in antimony determination by atomic techniques and NAA is observed. INAA demonstrates a very good sensitivity and selectivity for antimony and it is often required for measuring Sb in materials in which the antimony level is very low. Non-destructive methods are not only advantageous because of reduced sample handling, but also because they are independent of the efficiency of the digestion or extraction procedure. Hence, it can be assumed that the results provided by NAA laboratories are more accurate than those obtained by other methods. The second exercise was performed in 2008 by the Horia Hulubei National Institute for Physics and Nuclear Engineering, Romania. The analysed material was EURONORM-CRM no High Alloy Stainless Steel HAAS-1. Table 2 demonstrates the results obtained by the INCT in comparison with the assigned value. For all elements, the results agree very well, however INAA assures the lower relative standard uncertainty u (k = 1) for all the determined elements. In this type of material, due to a high content of boron in the matrix (mean B concentration was ± %) correction for neutron self-absorption has been made. Nuclide used Thermal neutron flux, irradiation time Cooling time Measurement time 56 Mn 24 h 100 s cm 2 s 1 76, As, 64 Cu, 187 W h s 5 min 51 Cr, 59 Fe, 60 Co, 122 Sb, 124 Sb, 99 Mo, 58 Ni (via 58 Co) 2 weeks 2000 s
3 Instrumental neutron activation analysis (INAA) for steel analysis and certification 313 Fig. 1. Comparison of the results obtained by the INCT with the assigned values calculated by the providers from results obtained by atomic spectrometry techniques, and with the arithmetic mean of all INAA results. In PT exercise conducted in the frame of NAA Euro-Network, six NAA laboratories from Europe participated in As test material EURONORM- -CRM no Duplex Stainless Steel was chosen. The Table 2. Results of analysis of High Alloy Stainless Steel HAAS-1 by INCT Element Assigned value NAA spectrum of a steel sample after 4 days cooling is presented in Fig. 2. In Table 3 the certified and information values for CRM no [3] are compared to the results obtained in the INCT. As can be seen, High Alloy Stainless Steel HAAS-1 INCT results by INAA Wt.% u (k = 1)% Wt.% u (k = 1)% As Co Cr Cu Mn Mo Ni Sb W
4 314 H. Polkowska-Motrenko, E. Chajduk, B. Danko Fig. 2. The NAA spectrum of a steel sample 4 days after cooling, measurement time: 1500 s. the INAA values were near the center of the certified values. On the basis of the INCT results and estimated standard uncertainties, zeta-score was calculated (Fig. 3). All the obtained values are in the range 1 < zeta < 1 and this means that the procedure applied in the INCT for standard uncertainty evaluation is correct. The mean values of the experimental data given by all participants using the INAA method were as follows (Cx ± SD), (%, w/w): As: ± , Co: ± 0.005, Cr: 24.2 ± 1.2, Cu: ± 0.025, Fe: ± 3.96, Mn: 0.39 ± 0.02, Mo: 3.68 ± 0.30, Ni: 7.35 ± 0.33, Sb: ± , W: ± Zeta-score calculated for all the above-mentioned values is between 1.0 and 1.0 (Fig. 4). This range delivers proof that the neutron activation analysis gives accurate results. Also, looking at the antimony determination, the delivered data (mean Sb concentration for five laboratories 8.1 ± 0.4 mg kg 1 ) are in excellent agreement with data listed in the Certificate (mean value: 7.4 ± 0.8 mg kg 1 ). This confirms that the previous analysis of Sb in SS1 steel were carried out correctly by NAA laboratories. Looking at the obtained INAA data for steel certified reference materials analysis, it can be noted that this method gives accurate results for the major and trace elements. A comparison of the results from NAA method with the certified/assigned values gives an independent proof that the applied certification/assignment procedure is properly performed. In the case of some elements like antimony, when atomic spectroscopic reference methods [2, 13] give questionable results, INAA can meet requirements of the reference method. Conclusions The elaborated INAA procedure has proved its reliability for steel analysis. This can be stated on the basis of results of the participation of INCT in proficiency tests dedicated for European NAA Laboratories. It can be emphasized that INAA results obtained by INCT in the frames of PT/ILC exercises in the years Table 3. Results of analysis of ECISS a for CRM no Duplex Stainless Steel Element ECISS Certificate Duplex Stainless Steel INCT results by INAA Wt.% u (k = 1) Wt.% u (k = 1) As c Co Cr Cu Fe Mn Mo Ni Sb b W c a ECISS European Committee for Iron and Steel Standardization. b The mean concentration for Sb calculated from the data in the Certificate [3]. c Information value. Fig. 3. Zeta-score for INCT results for Duplex Stainless Steel. Fig. 4. Zeta-score calculated for the arithmetic mean of INAA results delivered by participant laboratories.
5 Instrumental neutron activation analysis (INAA) for steel analysis and certification agree very well with the consensus/certified values of the investigated steel test materials. In the case of this type of material, INAA can be used for the determination of both major and trace elements which are determined with similar relative standard uncertainty. Obviously, in this type of materials (metal, steel, alloy), characterized by relatively high density or addition elements with high absorption cross sections, the gamma-ray self-absorption occurs and appropriate corrections should be calculated [14, 15]. Nevertheless, INAA is one of the preferred multi-elemental methods for difficult samples analysis, because it is free from problems connected with digestion of difficult matrices and potential contamination of a sample from reagents. Unique properties of NAA, i.e. the fact that this is a method based on nuclear properties of individual isotopes of an element and good accuracy at low concentration levels, are the explanation why this technique is often used for certification of the candidate reference materials. For some difficult elements, NAA should be taken into consideration as an alternative reference technique instead of atomic spectrometry techniques for certification of CRMs. Acknowledgment. The authors would like to thank Dr. E. Cincu from IFIN-HH, Bucharest-Magurele, Romania for organization of PTs and providing steel samples. References 1. Cincu E, Craciun L, Manera-Grigore I et al. (2009) Application of the INAA technique for elemental analysis of metallic biomaterials used in dentistry. Appl Radiat Isot 67: Cincu E, Manera I, Manu V et al. (2007) Comparative performance of INAA and other spectroscopy techniques in the elemental analysis of stainless steel materials. J Radioanal Nucl Chem 274: European Committee for Iron and Steel Standardization (2002) Certificate of chemical analysis EURONORM- -CRM no Duplex Stainless Steel. Swedish Institute for Metals Research, Stockholm 4. Gervasio APG, Luca GC, Menegario AA, Reis BF, Filho HB (2000) On-line electrolytic dissolution of alloys in flow injection analysis: determination of iron, tungsten, molybdenum, vanadium and chromium in tool steels by inductively coupled plasma atomic emission spectrometry. Anal Chim Acta 405: Granfors M, Gustavsson I (2001) Determination of trace and ultra-trace elements in wet-digested steel using inductively coupled plasma time of flight mass spectrometry (ICP-TOFMS). J Anal At Spectrom 16: Ida H, Kawai J (2005) An X-ray fluorescence spectrometer with X-ray generator and a secondary target for the determination of Cr in steel. Spectrochim Acta B 60: Kimura A, Toh Y, Oshima M, Hatsukawa Y, Goto J (2007) Determination of As and Sb in iron and steel by neutron activation analysis with multiple gamma-ray detection. J Radioanal Nucl Chem 271: Liu H, Jiang SJ (2003) Dynamic reaction cell inductively coupled plasma mass spectrometry for determination of silicon in steel. Spectrochim Acta Part B 58: Mackey EA, Becker DA, Spatz RO et al. (2004) Certification of NIST Standard Reference Material 1575a Pine Needles and Results of an International Laboratory Comparison. NIST Special Publication NIST, Washington 10. Moreira EG, Vasconcellos MBA, Saiki M (2005) Instrumental neutron activation analysis applied to the determination of the chemical composition of metallic materials with study of interferences. J Radioanal Nucl Chem 264: Oss Giacomelli MB, Borba da Silva JB, Curtius AJ (1998) Determination of Mo and Bi in steels by electrothermal atomic absorption spectrometry after complexation and sorption on activated carbon. Talanta 47: Polkowska-Motrenko H, Dybczyński RS, Chajduk E (2010) Certification of reference materials for inorganic trace analysis: the INCT approach. Accredit Qual Assur 15: SIMR (2005) Report on Interlaboratory Comparison of elemental composition in Stainless Steel SS1 determined by laboratories from Belgium, Hungary, Poland and Romania using the INAA analytical technique on the Stainless Steel SS1 received from SIMR within the PT Exercise Swedish Institute for Corrosion and Metals Research, Stockholm Tian W, Ni B, Wang P, Cao L, Zhang Y (2001) Metro- logical role of neutron activation analysis. IA. Inherent characteristics of relative INAA as a primary ratio method of measurement. Accredit Qual Assur 6: Tian W, Ni B, Wang P, Cao L, Zhang Y (2002) Metro- logical role of neutron activation analysis. IB. Inherent characteristics of relative INAA as a primary ratio method of measurement. Accredit Qual Assur 7:7 12 Zeisler R, Greenberg RR, Mackey EA, Murphy KE, Spatz RO, Tomlin BE (2009) Challenges and success in the use of neutron activation analysis procedures for value assignment of animal serum and bovine liver Standard Reference Materials. J Radioanal Nucl Chem 281:11 15 Zeisler R, Lindstrom RM, Greenberg RR (2005) Instru- mental neutron activation analysis: a valuable link in chemical metrology. J Radioanal Nucl Chem 263:
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
HS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser
Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser V.Nassisi #, G.Caretto #, A. Lorusso #, D.Manno %, L.Famà %, G.Buccolieri %, A.Buccolieri %, U.Mastromatteo* # Laboratory of Applied
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
Determination of elements by ICP-AES and ICP-MS
September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL
h e l p s y o u C O N T R O L
contamination analysis for compound semiconductors ANALYTICAL SERVICES B u r i e d d e f e c t s, E v a n s A n a l y t i c a l g r o u p h e l p s y o u C O N T R O L C O N T A M I N A T I O N Contamination
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol
Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS INTRODUCTION Each REACH registrant is required to file its
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
Certificate of Analysis
The Japan Society for Analytical Chemistry Certificate of Analysis Certified Reference Material JSAC 0602-3 Plastics (tip form) for Chemical Analysis of Hazardous Metals This certified reference material
Elemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE
Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications
NIST Special Publication 260-172
NIST Special Publication 260-172 Certification of Three NIST Renewal Soil Standard Reference Materials for Element Content: SRM 2709a San Joaquin Soil, SRM 2710a Montana Soil I, and SRM 2711a Montana Soil
ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
MILESTONE CLEAN CHEMISTRY LINE DUOPUR TRACECLEAN ULTRA-TRACE INSERTS. A Breakthrough in Reduction and Control of the Analytical Blank MILESTONE
MILESTONE CLEAN CHEMISTRY LINE A Breakthrough in Reduction and Control of the Analytical Blank DUOPUR TRACECLEAN ULTRA-TRACE INSERTS MILESTONE MILESTONE CLEAN CHEMISTRY LINE An innovative and complete
Universal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence
A Digital Spectrometer Approach to Obtaining Multiple Time-Resolved Gamma-Ray. Spectra for Pulsed Spectroscopy. W. K. Warburton a
1 IRRMA6 #21 A Digital Spectrometer Approach to Obtaining Multiple Time-Resolved Gamma-Ray Spectra for Pulsed Spectroscopy H. Tan a,, S. Mitra b, A. Fallu-Labruyere a, W. Hennig a, Y. X. Chu a, L. Wielopolski
There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.
0(752/2*
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
Thème Analytical Chemistry ECUE composant l UE
Thème Analytical Chemistry ECUE composant l UE UE 7.1 Performing analyses in the environment and complex systems UE 7.2 Heterogeneity and complexity UE 7.3 Processes in aquatic systems and at geo-interfaces
Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination
Analytical Techniques for Trace Elemental Analyses on Wafer Surfaces for Monitoring and Controlling Contamination J. Wang, M. K. Balazs Balazs Analytical Laboratory, Sunnyvale, CA 94089 P. Pianetta, K.
RESULTS OBTAINED IN MEASUREMENTS OF Rn-222 WITH THE ROMANIAN STANDARD SYSTEM*
NUCLEAR PHYSICS RESULTS OBTAINED IN MEASUREMENTS OF Rn-222 WITH THE ROMANIAN STANDARD SYSTEM* M. SAHAGIA 1, **, A. LUCA 1, A. C. WATJEN 1, A. ANTOHE 1, C. IVAN 1, D. STANGA 1, C.VARLAM 2, I. FAURESCU 2,
WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA
Upgrade and Certification of the Los Alamos National Laboratory SHENC 2011-12270 ABSTRACT S.B. Stanfield 1, M. Villani 1, P. T. Barton 1, C. Gerlock 1, D. Nakazawa 1, R. C. Baumann 1, R. Mowry 1, J.P Harvill
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
RoHS Test Methods. Presented by: Bruce Peterson
RoHS Test Methods Presented by: Bruce Peterson Motivation for Presentation The EU RoHS directive takes effect July 1, 2006 Customer confusion about - Scope of directive How much testing needs to be performed
CATALOGUE REFERENCE MATERIALS
IPT Institute for Technological Research CATALOGUE REFERENCE MATERIALS 2014 Laboratory of Metrological References IPT s Reference Materials www.ipt.br/nmr.htm Av. Prof. Almeida Prado n0 532 Predio 31 Cidade
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
THREE MAIN SOLIDIFICATION REACTIONS OF VANADIUM MODIFIED T1 TUNGSTEN HIGH SPEED TOOL STEEL. Hossam Halfa
THREE MAIN SOLIDIFICATION REACTIONS OF VANADIUM MODIFIED T1 TUNGSTEN HIGH SPEED TOOL STEEL Hossam Halfa Steel Technology Department, Central Metallurgical R&D Institute (CMRDI), Helwan, Egypt, [email protected];
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature
ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility
Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology
Quantitation of Drugs in Dried Bloodstains. Thomas Meyer Anna Gomenyuk Nadiah Lester
Quantitation of Drugs in Dried Bloodstains Thomas Meyer Anna Gomenyuk Nadiah Lester Objective To find the quantitative value of drugs in dried bloodstains Currently Cloth With Dried Bloodstain Containing
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
All answers must use the correct number of significant figures, and must show units!
CHEM 10113, Quiz 2 September 7, 2011 Name (please print) All answers must use the correct number of significant figures, and must show units! IA Periodic Table of the Elements VIIIA (1) (18) 1 2 1 H IIA
Potassium Aluminium Silicate (Mica) Chemical and Technical Assessment (CTA) Prepared by Daniel E. Folmer, Ph.D., and reviewed by Madduri V. Rao, Ph.D.
Potassium Aluminium Silicate (Mica) Chemical and Technical Assessment (CTA) Prepared by Daniel E. Folmer, Ph.D., and reviewed by Madduri V. Rao, Ph.D. 1. Summary Potassium aluminium silicate (PAS) was
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
Perspectives of a method for measuring soil-gas radon by an opened counting vial
Perspectives of a method for measuring soil-gas radon by an opened counting vial Hugo López del Río Juan Carlos Quiroga Cifuentes Unidad Académica de Estudios Nucleares Universidad Autónoma de Zacatecas
Chapter 1: Moles and equations. Learning outcomes. you should be able to:
Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including
Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions
Elemental solution standards can be produced that are almost clones of NIST SRMs. TRACEABILITY of Single-Element Calibration Solutions Commercially available single-element solutions are used to calibrate
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
Coating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
ANALYTICAL REPORT LABORATORY DETAILS. Telephone Facsimile Email. SGS Reference Report Number Date Reported Date Received
ANALYTICAL REPORT CLIENT DETAILS Contact Client Address Matthew Behre SGS Metallurgical Services 431 Victoria Street Malaga MALAGA WA 6090 LABORATORY DETAILS Manager Laboratory Address Said Hirad SGS Newburn
Performance and advantages of qnmr measurements
Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Hannu Teisala a, Mikko Tuominen a, Mikko Aromaa b, Jyrki M. Mäkelä b, Milena Stepien c, Jarkko J. Saarinen c, Martti Toivakka c
SUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536
SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet
AUSTRALIAN ATOMIC ENERGY COMMISSION
^ "^ UJ O UJ AAEC/E615 AUSTRALIAN ATOMIC ENERGY COMMISSION RESEARCH ESTABLISHMENT LUCAS HEIGHTS RESEARCH LABORATORIES DEVELOPMENT OF A PILOT PLANT FOR THE REMOVAL OF RHENIUM FROM MOLYBDENUM TRIOXIDE by
Start of introducing ICP MS method for analyzing water samples for Strontium. November 27, 2014 Tokyo Electric Power Company
Start of introducing ICP MS method for analyzing water samples for Strontium November 27, 2014 Tokyo Electric Power Company 1. Current Strontium analysis and introduction of various methods Strontium analysis
Paper No. 4071 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION
Paper No. 471 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION Yang Yang, Bruce Brown and Srdjan Nešić Institute for Corrosion and Multiphase Technology, Department of Chemical and Biomolecular Engineering
CH3 Stoichiometry. The violent chemical reaction of bromine and phosphorus. P.76
CH3 Stoichiometry The violent chemical reaction of bromine and phosphorus. P.76 Contents 3.1 Counting by Weighing 3.2 Atomic Masses 3.3 The Mole 3.4 Molar Mass 3.5 Percent Composition of Compounds 3.6
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
ATI Wah Chang Laboratory and Technical Services: Analytical Laboratory
ATI Wah Chang Laboratory and Technical Services: Analytical Laboratory TECHNICAL DATA SHEET INTRODUCTION For over 50 years, ATI Wah Chang s laboratory personnel have been producing high quality inorganic
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
APPENDIX B: EXERCISES
BUILDING CHEMISTRY LABORATORY SESSIONS APPENDIX B: EXERCISES Molecular mass, the mole, and mass percent Relative atomic and molecular mass Relative atomic mass (A r ) is a constant that expresses the ratio
NON-DESTRUCTIVE PHYSICAL MEASUREMENTS FOR THE STUDY OF MUSEUM ENVIRONMENTS AND CHARACTERIZATION OF ARCHAEOLOGICAL ARTIFACTS
NON-DESTRUCTIVE PHYSICAL MEASUREMENTS FOR THE STUDY OF MUSEUM ENVIRONMENTS AND CHARACTERIZATION OF ARCHAEOLOGICAL ARTIFACTS M. F. Alberghina 1,2, M. Brai 1,2, D. Fontana 1,2*, S. Schiavone 3, L. Tranchina
Prentice Hall. Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition. High School. High School
Prentice Hall Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition High School C O R R E L A T E D T O High School C-1.1 Apply established rules for significant digits,
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
Fuel Cell Activities at TU Graz
Fuel Cell Activities at TU Graz Viktor Hacker Institute of Chemical Engineering and Environmental Technology Graz University of Technology IEA Workshop TU Graz September 1 st, 2010 Content Lifetime and
EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction
ACTA PHYSICA DEBRECINA XLVI, 25 (2012) EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES L. Csedreki 1 1 Institute of Nuclear Research of the Hungarian Academy of Sciences,
WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA
ABSTRACT Comparison of Activity Determination of Radium 226 in FUSRAP Soil using Various Energy Lines - 12299 Brian Tucker*, Jough Donakowski**, David Hays*** *Shaw Environmental & Infrastructure, Stoughton,
Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015
Introduction to Elemental Analysis by EDXRF Justin Masone Product Specialist 3 June 2015 1 / 9 Shimadzu Corporation Established in 1875. Headquartered in Kyoto, Japan Ranked Top 5 Instrument Providers
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of
Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm
Page 1 of 14 Amount of Substance Key terms in this chapter are: Element Compound Mixture Atom Molecule Ion Relative Atomic Mass Avogadro constant Mole Isotope Relative Isotopic Mass Relative Molecular
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
Radiation Protection s Benefits in the Production of Iodine-131
Radiation Protection s Benefits in the Production of Iodine-131 Daniel M. Cestau *, Ariel N. Novello, Cecilia Bravo, Pablo Cristini, Marcelo Bronca, Eduardo Carranza, Ricardo Bavaro, Julián Cestau and
Elemental Analysis of Ash using X-Ray Fluorescence Technique
Asian Journal of Chemistry Vol. 21, No. 10 (2009), S220224 Elemental Analysis of Ash using XRay Fluorescence Technique PRAKASH CHAND, ASHAVANI KUMAR, ANURAG GAUR and S.K. MAHNA Department of Physics, National
Quality Control on Aerospace Components Using Handheld X-ray Fluorescence (XRF)
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada Quality Control on Aerospace Components Using Handheld X-ray
Mass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
Innovative Technology Verification Report
United States Office of Research and EPA/600-R-03/149 Environmental Protection Development May 2004 Agency Washington DC 20460 Innovative Technology Verification Report Field Measurement Technology for
APPLYING DATA MINING METHODS TO PREDICT DEFECTS ON STEEL SURFACE
APPLYING DATA MINING METHODS TO PREDICT DEFECTS ON STEEL SURFACE 1 SAYED MEHRAN SHARAFI, 2 HAMID REZA ESMAEILY 1 Computer Engineering Department, Islamic Azad University, Najaf Abad Branch, Esfahan, Iran
Nuclear ZPE Tapping. Horace Heffner May 2007
ENERGY FROM UNCERTAINTY The uncertainty of momentum for a particle constrained by distance Δx is given, according to Heisenberg, by: Δmv = h/(2 π Δx) but since KE = (1/2) m v 2 = (1/(2 m) ) (Δmv) 2 ΔKE
Trace Metals in Traditional Chinese Medicine: A Preliminary Study Using ICP-MS for Metal Determination and As Speciation
Trace Metals in Traditional Chinese Medicine: A Preliminary Study Using ICP-MS for Metal Determination and As Speciation Xiaoru Wang, Zhixia Zhuang, Dahai Sun, Jiangxing Hong, and Xihong Wu Chemistry Department,
Chapter 3 Mass Relationships in Chemical Reactions
Chapter 3 Mass Relationships in Chemical Reactions Student: 1. An atom of bromine has a mass about four times greater than that of an atom of neon. Which choice makes the correct comparison of the relative
Chapter Three: STOICHIOMETRY
p70 Chapter Three: STOICHIOMETRY Contents p76 Stoichiometry - The study of quantities of materials consumed and produced in chemical reactions. p70 3-1 Counting by Weighing 3-2 Atomic Masses p78 Mass Mass
WASTE STREAM 2Y51 Analytical Services Process Facilities - North Labs
WASTE STREAM 2Y51 Analytical Services Process Facilities North Labs SITE SITE OWNER WASTE CUSTODIAN WASTE TYPE Sellafield Nuclear Decommissioning Authority Sellafield Limited LLW WASTE VOLUMES Stocks:
Field portable XRF analysis of environmental samples
Journal of Hazardous Materials 83 (2001) 93 122 Field portable XRF analysis of environmental samples Dennis J. Kalnicky a, Raj Singhvi b, a Lockheed Martin Technology Services Group, Environmental Services/REAC,
B) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal
1. The elements on the Periodic Table are arranged in order of increasing A) atomic mass B) atomic number C) molar mass D) oxidation number 2. Which list of elements consists of a metal, a metalloid, and
Using the PDF for material identification using elemental data. from XRF and SEM EDS.
XRF and SEM EDS Using the PDF for material identification using elemental data from XRF and SEM EDS. XRF and SEM EDS What? The Powder Diffraction File contains data on pure solid state compounds of well
Instrumentation. (Figure 2)
X-Ray Fluorescence Lab Report Nydia Esparza Victoria Rangel Physics of XRF XRF is a non destructive analytical technique that is used for elemental and chemical analysis. X-Ray Fluorescence Spectroscopy
Overview of Nuclear Detection Needs for Homeland Security. Abstract
Overview of Nuclear Detection Needs for Homeland Security Timothy E. Valentine * Oak Ridge National Laboratory, P.O. Box 2008, Oak Ridge, Tennessee, 37831, USA Abstract The need for advanced and improved