TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions

Size: px
Start display at page:

Download "TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions"

Transcription

1 Elemental solution standards can be produced that are almost clones of NIST SRMs. TRACEABILITY of Single-Element Calibration Solutions Commercially available single-element solutions are used to calibrate the instruments with which millions of elemental determinations are performed each day. Atomic spectrometric techniques rely on calibration against these solution standards to determine the tens of elements in hundreds of samples in thousands of labs. The solutions are, or are prepared from, certified reference materials (CRMs), which are the basis for the accuracy of the measurement results. A certificate that asserts a claim of traceability usually accompanies these CRMs. The claim of traceability aims to ensure that the certified value of the CRM is linked to appropriate national and international references, which are presumably stable and accurate and of higher metrological order. Properly established, this linkage couples the information content of the traceable value to the information content of the reference, and it can add immensely to the acceptance of the CRM. This linkage also connects the measurement results calibrated from the traceable value of a CRM to the larger network of measurements, which ultimately stems from the International System of Units (SI). The maintenance of the base and derived units of SI, together with traceability to SI, underpins contemporary metrology. SI provides a uniform, stable measurement scale, and traceability of measurement results to SI ensures their comparability over time and space. To provide suitable references for the millions of atomic spectrometric determinations made daily, the National Institute of Standards and Technology (NIST) disseminates elemental solution standards through its standard reference materials (SRMs) program; SRMs are CRMs distributed by NIST. We will describe the concept of traceability and one approach by which traceability to these references can be achieved by directly comparing solutions against them. The approach we describe is based on work from our laboratory using a method we refer to as high-performance inductively coupled plasma-optical emission spectrometry (HP-ICP-OES; 1 4). Traceability The concept of traceability predates our contemporary global metrology system. Originally, the king s foot was a measure of length; then, traceable artifacts the same length as his foot were disseminated to permit measurements to be made with ease, thus saving the king s foot a lot of wear and tear. The International Vocabulary of Basic and General Terms in Metrology defines traceability as the property of the result of a measurement or the value 136 A ANALYTICAL CHEMISTRY / APRIL 1, AMERICAN CHEMICAL SOCIETY

2 Marc L. Salit Gregory C. Turk National Institute of Standards and Technology of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties (5). This definition, together with agreed-upon units of measure, ensures that a traceable result (a value and its associated uncertainty) measured in the absence of bias can be compared with other traceable results, also measured in the absence of bias. Such results may arise from comparison to the value of the same stable artifact in different places at different times, or from comparison to the values of different artifacts, whose values are also traceable. Comparability of the results is ensured by traceability. The ability to compare a result to other results, expectations, or specifications is essential. A measurement result that cannot be compared is of no value. Even if results are to be compared only within a given laboratory, those results are useful only by virtue of their values relative to other results. The extent of space and time over which results can be compared depends on the scope of the reference to which they are linked or to which they are traceable. Local traceability ensures local comparability. A measurement result traceable to a reference held in a particular laboratory can be meaningfully compared only to other measurement results traceable to that reference. If that reference is itself traceable to an external reference (to which other references may also be traceable), then traceability to that external reference may be established. In this manner, measurement results can then be compared among laboratories. This network rises from standards held at individual laboratories through those held at national and international levels. Comparability between national and regional measurement systems is being established in part through the 1999 Comité International des Poids et Mesures (International Committee for Weights and Measures; CIPM) multilateral mutual recognition arrangement. This agreement, signed by the 38 member nations of the Meter Convention, was a response to the need for an open, transparent, and comprehensive scheme to give users reliable quantitative information on the comparability of national metrology systems ( mra.html). Through this horizontal comparability between nations, the vertical traceability of a measurement to a national reference may link it to international comparability, which is potentially important for international trade, commerce, and regulatory affairs, as well as global science. Of increasing relevance to chemical analysis is laboratory accreditation based on International Organization for Standard- APRIL 1, 2005 / ANALYTICAL CHEMISTRY 137 A

3 ization Guide standards that require a clear demonstration of traceability. In the United States, government regulations or laboratory quality systems often require that the results of an analysis be traceable to NIST, an ill-considered phrase that, more properly stated, means traceable to the values on NIST measurements or standards. NIST policy on traceability explicitly states: Other organizations are responsible for establishing the traceability of their own results or values to those of NIST or other stated references ( Additionally, the policy states that NIST [a]sserts that providing support for a claim of traceability of the result of a measurement or value of a standard is the responsibility of the provider of that result or value, whether that provider is NIST or another organization, and that assessing the validity of such a claim is the responsibility of the user of that result or value. This caveat warns that the users of materials that claim traceability must be able to evaluate those claims, and thus it gives tremendous importance to the transparency of a traceability claim. The analyst must recognize that traceability only requires two things: a suitable reference to compare to and a measurement method for the comparison. Of course, an essential part of a traceable result is its uncertainty, which must consist of the uncertainty of the value of the reference combined with the uncertainty of the comparison. An assessment of traceability evaluates the validity and suitability of the reference, the appropriateness and reliability of the comparison, and whether the uncertainty of the traceable value is valid. Elemental solution standards In the United States, designating a calibration solution for elemental analysis as traceable usually means that the certified value of that solution is traceable to the certified value of the matching elemental solution from the NIST 3100 series of SRMs. These 69 single-element solution SRMs serve as national standards for elemental solution mass fraction. These SRMs are prepared in bulk and are batch-certified for the mass fraction of the constituent element (typically 10 mg/g). The target range for the relative uncertainty in the certified mass fraction for these SRMs is ±0.3%. As part of our process to certify the 3100 series, candidate SRM materials are compared against well-characterized primary materials. Through this comparison, traceability of the SRM values to the SI is established. HP-ICP-OES has been our comparison method of choice for this process since This method uses unmodified, readily available, mature, reliable commercial instrumentation; it has been demonstrated to have very small comparison uncertainty, ruggedness with respect to bias (small changes in conditions do not bias the results), economy, sufficient linear dynamic range, and a high degree of inter-element selectivity. The HP-ICP-OES method has been applied to the analysis of single-element solutions for 64 elements (3), the certification of the major constituents of a high-temperature alloy (4), and the characterization of an LiAlO 2 ceramic material (2). The method was used in an international key comparison (K8: Monoelemental Calibration Solutions) of the Comité Consultatif pour la Quantité de Matière (Consultative Committee for the Amount of Substance) of the CIPM, with excellent results that validated it against a number of other analytical techniques (6). A similar approach using ICPMS for the multielement analysis of environmental CRMs has been reported (7 ). Pairing of the 3100 series with a rugged and practical comparison method enables the manufacture of traceable calibration solutions. This traceability linkage couples the information content of the SRM to the traceable solution. Thus, a manufacturer of a CRM could leverage this approach, relying on comparison against the SRM, and avoid duplication of the work done at NIST. The 3100 series solutions are themselves traceable to the mole, the SI unit for chemistry. The well-characterized primary materials, against which the SRMs are compared, are realizations of the mole for the elements. Comparison against the SRM can yield a value for a CRM, also now traceable to the mole. Production of a rhodium solution CRM is a good example. Rhodium metal is extremely difficult to dissolve, and soluble salts of rhodium are difficult to assay with small uncertainty. A commercial or in-house CRM solution could be prepared from a soluble salt, and a traceable value derived by comparison against SRM 3144, the rhodium standard solution SRM (8, 9). Unquestionably, the traceable mass fraction of analyte in the CRM could not have smaller uncertainty than the SRM. The 3100 series solutions are optimized to have stable mass fractions with small uncertainties. However, the quality of a calibration solution is derived from a number of parameters, not only the uncertainty of the mass fraction. Solution purity, stability, packaging, and concomitant anions all play a role in fitness for purpose, or quality. A traceable CRM could be of either higher or lower quality than the SRM. For example, many of the 3100 series solutions are packaged in glass ampoules that enhance stability by limiting transpiration but could induce contamination. A CRM could have a certified value traceable to the SRM and have higher purity, making it more suitable for multielement analysis. Or, a solution CRM might be prepared from a poor grade of material containing many impurities. The traceable certified value might be excellent accurate, with small uncertainty yet the material may be inappropriate for its application and of low quality. How to do it HP-ICP-OES exploits two strengths of array-detector-based spectrometers: simultaneous measurement of the analyte and internal standard, and time-correlated off-line background correction (3). Moderately high analyte mass fractions and signal levels provide practical immunity from blank and spectral interferences, 138 A ANALYTICAL CHEMISTRY / APRIL 1, 2005

4 and a drift-correction approach mitigates low-frequency noise. These strategies typically yield results with variability <0.1%, which is on the order of that associated with sample handling. Sample-handling variability is quantified with an experiment that includes the analysis of replicate preparations. The HP-ICP-OES method, or one like it using a different measurement tool for comparison, might be used by those wishing to prepare solutions traceable to NIST SRMs in particular, by commercial CRM manufacturers whose products are designed to ensure the accuracy and comparability of most elemental analysis results. We have developed a spreadsheet tool that leads the analyst through the HP-ICP-OES comparison to establish traceability of the value of a CRM to the value of a NIST 3100 series SRM. The essential elements of this comparison include a fixed experimental design; a method development experiment that establishes the sensitivities for the analyte and an internal standard; a gravimetric sample preparation scheme; a fixed set of emission intensity measurements of both elements; and the calculation of the traceable value and its uncertainty. The spreadsheet also includes useful diagnostic charts that permit visualization of the instrument drift and the noise correlation between the analyte and internal standard. The spreadsheet tool, an example data set, and instructions are available in Supporting Information. Four separate preparations are made of the desired element from the 3100 series SRM and the single-element solution to be compared (the test sample). The 4 preparations of the SRM are designed to consume ~10 g of solution (many of the 3100 series solutions are deployed in packages of 5 single-use 10-g glass ampoules). Ideally, the four preparations of the test sample are selected from different portions of the batch, so the population of results from the test sample will include a measure of any heterogeneity. This replication permits evaluation of several uncertainty components associated with the preparation of the solutions and acts to validate the method and its results. The analyst must take proper care to develop the instrumental method (wavelengths, spectral-background-correction approach and parameters, and integration times) such that the measured emission intensities have high S/ N and selectivity. A useful reference for method development is the study reporting HP-ICP-OES analyses for 64 different elements (3). Sample prep. Analyte and internal standard mass fractions are selected to yield an emission intensity ratio near unity. For some ICP-OES instruments, this facilitates simultaneous integration, in which time is used to extend the dynamic range of a spectrometer, and limits the impact of deviation from linear intensity response. Because the analyte and internal-standard spectral lines are likely to have different sensitivities, a preliminary experiment is performed to roughly determine the sensitivity ratio. Solutions of the analyte and internal standard, whose mass fractions are approximately known (10% uncertainty is adequate), are measured under the selected operating conditions, and the emission intensities are A measurement result that cannot be compared to other results, expectations, or specifications is of no value. recorded. These solutions should be well within reasonable operating levels for the instrument; for most elements, 10 µg/g is appropriate. Once the sensitivity ratio is established, a mass ratio (mass of analyte per mass of internal standard) that yields unity intensity ratio can be calculated. Then, an internal standard spike solution of appropriate mass fraction can be prepared. Once the spike is added to the analyte, the mass ratio is established. This is the only sample preparation data value used in calculating the result, and it is unaffected by any subsequent dilutions. Typically, the spiked samples are diluted when introduced to the plasma, in order to yield an appropriate signal level that is comfortably within the working linear response range of the instrument and sufficiently larger than any anticipated blank or interference level. This dilution can be performed by eye using a digital pipette, and it is not necessary to record any values associated with this dilution. All masses whose uncertainties propagate to the results are >2.5 g. This amount permits the use of common top-loading balances, which typically have 1.5-mg-repeatability standard deviations. This keeps the relative uncertainty of weighing to <0.1%, which is on the order of our ability to prepare duplicate preparations. Aliquots of ~2.5 g are used for both the SRM and the test samples. The scheme assumes that the nominal mass fractions of the SRM and the test samples are within an order of magnitude of each other. Aliquot masses are adjusted to deliver the same mass of analyte from each solution. An internal standard solution is prepared so that the spike masses are ~20 g. The mass fraction of this solution need not be accurately known, as only the relative masses of spike weighed into the reference and test solutions appear in the calculation of results. The mass used need not be precisely 20 g, and the solution is typically delivered with a graduated cylinder. The 20 g of spike solution is easily weighed with negligible uncertainty and typically dilutes the high-mass-fraction (10-mg/g) SRM. When two solutions with high mass fractions are mixed, a precipitate might form; dilution helps mitigate this. Final dilution from the spiked sample is also facilitated by this initial dilution. Moreover, the lower the mass fraction of the spike, the less the relative loss (from missed droplets, evaporation, or aerosolization) during solution handling. Intensity measurements. Intensity measurements are performed sequentially on the eight samples, alternating the SRM and test samples (the run order is SRM sample 1, test sample 1, SRM sample 2, test sample 2, etc.). This sequence of measurements is repeated five times. Each is typically the mean of several replicate intensity measurements. The short-term noise of the intensities, expressed as the relative standard deviation of the replicate measurements, is a good diagnostic measure of the sample input performance. The short-term noise of the ratios, and its improvement over the intensity noise, helps to differentiate among the sources of measurement noise; the correlation of an- APRIL 1, 2005 / ANALYTICAL CHEMISTRY 139 A

5 Table 1. Sensitivity test. Target Sensitivity working solution Measured mass mass emission Sensitivity fraction Sensitivity fraction (µg/g) intensity [(µg/g) 1 ] (µg/g) ratio Cobalt 10 1,670, , Scandium 5 22,042,376 4,408, alyte and internal standard signals is an indicator of performance and sets the expectation for uncertainty. Calculations. The straightforward calculations are based on Equation 1 in Reference 6. Intensity ratios are calculated for each repeat measure of each solution; these ratios are drift-corrected and summarized for each sample. Mass ratios are calculated from the preparation data, and the ratio of the intensity ratio to the mass ratio is calculated for each sample. A single calibration factor is determined from the summary of the SRM measurements and is used to calculate the traceable mass fraction of the test sample. The uncertainty budget includes components for dispersion of measurements of the SRM and the test sample (due to variability of the intensity ratio measurements and preparation and weighing) and for the uncertainty of the certified value of the SRM. A coverage factor is selected to expand the standard uncertainty to a 95% confidence interval. Test experiment Two ampoules of cobalt SRM 3113 (lot , certified mass fraction ± mg/g) solution were selected to demonstrate the approach. One of these ampoules was diluted by ~0.5% (45.76 mg of 2% volume fraction HNO 3 was added to g of SRM solution) to prepare the synthetic test sample. The test sample mass fraction was ± mg/g. Cobalt mass fraction was determined by HP-ICP-OES comparison of the diluted sample against the unadulterated SRM. Details of the experiment can be found elsewhere (3). Table 1 reports the results of the sensitivity test: the mass fractions of the sensitivity solutions, measured intensities, sensitivities, RSD working mass fractions (nominal mass fractions of analyte and internal standard to be introduced to the plasma), and sensitivity ratio. The working mass fractions were calculated to yield an operating signal level of 10 6 counts/s. The target and actual weights for the eight solutions prepared can be found in Supporting Information. The target weights are based upon the sensitivity test, an SRM aliquot of 2.5 g, and a minimum spike weight of 20 g. Target weights are nominal values. Variations of as much as tens of percent are well within expected norms when the solutions are prepared. Typical variability is observed in these data. Approximately 60 g of solution were prepared for analysis. Based upon the working mass fraction, the nominal weight of spiked solution to dilute to 60 g is calculated as 0.32 g. The 5 replicate measurements of the 8 samples took place in <2 h. Five repeat integrations of 8 s each were measured for each replicate measurement of each sample. Total photon integration time was 26 min 40 s. At every sample change (40 of them), 90 s elapsed to permit the new solution to traverse to and equilibrate in the spray chamber and reach the plasma. The analyte and internal standard signals were highly correlated. A sevenfold reduction in the noise in the ratio compared to the analyte signal can be observed in Figure 1. A straight-line calibration model is fitted and forced through zero for the four SRM solutions. The slope of this line is estimated as the mean of the slopes calculated for each of the four solutions, resulting in a single calibration factor. Mass fractions for all solutions (treating each SRM solution as an unknown) are then calculated from this calibration factor; Figure 2 presents results before and after drift-correction and reference lines for the mean mass fraction and uncertainty of the SRM. The center of the gray reference bar is the reference value for our synthetic test sample, and the thickness is its uncertainty. The uncertainty of the SRM is excluded from all results graphed for the test sample to allow only the comparison to be assessed. The dashed black lines represent the mean and standard uncertainty of the HP-ICP-OES results (9.950 ± mg/g before drift-correction and ± mg/g thereafter). Several observations can be made directly. When the data for the individual solutions are drift-corrected, the dispersion is smaller. However, the dispersion among the results of the four solutions (for either the SRM or the test sample) is essentially unaffected by drift-correction. The dispersion among the four preparations of a solution dominates the dispersion of the measurements. This preparation effect has been documented and arises from uncertainty in solution handling and weighing, limitations of the balance, and evaporation of the solution while it is Intensity Intensity ratio FIGURE 1. S/N improvement expressed as relative standard deviation with time-correlated internal standardization. 140 A ANALYTICAL CHEMISTRY / APRIL 1, 2005

6 being weighed. The magnitude of this effect is ~0.03% (3). The result for the synthetic test sample from the comparison is consistent with the reference value. The uncertainty budget before and after drift-correction for the comparison is presented in Table 2. This simple yet rigorous budget includes components of uncertainty for the measurement of the test sample and SRM These standard uncertainties are evaluated as the standard deviation of the mean of the results from the four preparations. The variability of the intensity Mass fraction (mg/g) Drift-corrected Reference value for test sample Drift-corrected SRM SRM Test sample Test sample FIGURE 2. Results of test experiment comparison. ratios is convolved into the dispersion of the replicate preparations, as is the variability of preparation. This method-validation approach to uncertainty evaluation is a useful feature of the experiment, bundling together the evaluation of individual uncertainty components into evaluation via replication. The dominant uncertainty in the traceable value calculated from the comparison arises from the uncertainty of the certified value of the SRM. This value is ~4 larger than the uncertainty due to the comparison. It is a general rule of thumb that uncertainty components 3 smaller than the largest components can be regarded as insignificant (11). This rule of thumb applies here: We observe an insignificant difference between the uncertainty of the certified value of the SRM and the uncertainty of the result for the test solution. The results of the test experiment demonstrate that establishing traceability can be done without inflating uncertainty and in less than half a workday. HP-ICP- OES is an effective tool to link the value of a test solution, commercial CRM, or in-house standard to global comparability. When traceable materials are used to calibrate a valid method, the results of analyses using that method can be compared with other results, thereby giving them real value. Marc L. Salit is a research chemist and Gregory C. Turk heads the spectrochemical methods group at the National Institute of Standards and Technology. Address correspondence about this article to Turk at gregory.turk@nist.gov. Table 2. Results and uncertainty budget. Before drift-correction After drift-correction Degrees of Degrees of mg/g freedom mg/g freedom Observed value of test solution Uncertainty sources SRM Test sample Combined uncertainty of measurements only Uncertainty of certified value of SRM Combined standard uncertainty Coverage factor Expanded uncertainty References (1) Salit, M. L.; Turk, G. C. Anal. Chem. 1998, 70, (2) Salit, M. L.; Vocke, R. D., Jr.; Kelly, W. R. Anal. Chem. 2000, 72, (3) Salit, M. L.; et al. Anal. Chem. 2001, 73, (4) National Institute of Standards and Technology. SRM 1775/2175 Certificate of Analysis. NIST: Gaithersburg, MD. (5) International Organization for Standardization. International Vocabulary of Basic and General Terms in Metrology, Section 6.10, ISO: Geneva, Switzerland, (6) Metrologia 2002, 39, Tech. Suppl (7) Turk, G. C.; et al. Fresenius J. Anal. Chem. 2001, 370, (8) Beck, C. M., II; et al. Anal. Chem. 1993, 65, (9) Gordon, C. L.; Schlecht, W. G.; Wichers, E. J. Res. Natl. Bur. Stand. (U.S.) 1944, 33, (10) Taylor, B. N.; Kuyatt, C. E. Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results; NIST Technical Note 1297, 2nd ed.; National Institute of Standards and Technology: Gaithersburg, MD, 1994; (11) Eurachem/CITAC Guide: Quantifying Uncertainty in Analytical Measurement, 2nd ed.; Laboratory of the Government Chemist: London, APRIL 1, 2005 / ANALYTICAL CHEMISTRY 141 A

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

Elemental Analyses by ICP-AES

Elemental Analyses by ICP-AES Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

Performance and advantages of qnmr measurements

Performance and advantages of qnmr measurements Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended

More information

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based

More information

Quantifying Uncertainty in Analytical Measurement

Quantifying Uncertainty in Analytical Measurement EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Procedure for Equipment Calibration and Maintenance

Procedure for Equipment Calibration and Maintenance Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory

More information

HS 1003 Part 2 HS 1003 Heavy Metals Test

HS 1003 Part 2 HS 1003 Heavy Metals Test HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:

More information

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12

More information

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA ABSTRACT Comparison of Activity Determination of Radium 226 in FUSRAP Soil using Various Energy Lines - 12299 Brian Tucker*, Jough Donakowski**, David Hays*** *Shaw Environmental & Infrastructure, Stoughton,

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 18 Traceability for Life Sciences Testing and (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications

The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications Kevin Gates, Ning Chang, Isil Dilek, Huahua Jian, Sherri

More information

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

Sample Analysis Design Isotope Dilution

Sample Analysis Design Isotope Dilution Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with

More information

A commitment to quality and continuous improvement

A commitment to quality and continuous improvement Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009 Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production

More information

Overview of ISO 17511 for 'Biologicals'

Overview of ISO 17511 for 'Biologicals' JCTLM Symposium on Reference Measurement Systems for Biologicals International Bureau of Weights and Measures Pavillon de Breteuil, Sèvres, FR 2004-12-15 Overview of ISO 17511 for 'Biologicals' R. Dybkaer

More information

How To Use The Nexion 300

How To Use The Nexion 300 APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of

More information

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

Calibration of Volumetric Glassware

Calibration of Volumetric Glassware Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

IAEA-TECDOC-1350. Development and use of reference materials and quality control materials

IAEA-TECDOC-1350. Development and use of reference materials and quality control materials IAEA-TECDOC-1350 Development and use of reference materials and quality control materials April 2003 The originating Section of this publication in the IAEA was: Industrial Applications and Chemistry Section

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters. Environment Agency July 2015 Version 2

Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters. Environment Agency July 2015 Version 2 Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters Environment Agency July 2015 Version 2 Foreword We set up our Monitoring Certification Scheme

More information

Stimuli to the Revision Process

Stimuli to the Revision Process 2 STIMULI TO THE REVISION PROCESS of the USPC or the USP Council of Experts Vol. 33(6) [Nov. Dec. 2007] The Application of Uncertainty to USP s Compendial Reference Standards Program: Certified Reference

More information

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant

Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth

More information

UNCERTAINTY AND CONFIDENCE IN MEASUREMENT

UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ελληνικό Στατιστικό Ινστιτούτο Πρακτικά 8 ου Πανελληνίου Συνεδρίου Στατιστικής (2005) σελ.44-449 UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ioannis Kechagioglou Dept. of Applied Informatics, University

More information

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 1 TABLE OF CONTENTS FOREWORD... 1 1.0 SCOPE... 1 1.5 Safety Requirements... 1 4.0 MANAGEMENT REQUIREMENTS... 1 4.1

More information

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development

More information

ALACC Frequently Asked Questions (FAQs)

ALACC Frequently Asked Questions (FAQs) Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC

More information

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology

More information

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often

More information

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective

More information

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS

PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS Maria Belli Istituto Superiore per la Protezione e la Ricerca Ambientale (former APAT) ISPRA activities production and characterization

More information

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

Validation of measurement procedures

Validation of measurement procedures Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation

More information

and Drug Administration/Office of Regulatory Affairs (FDA/ORA) laboratories.

and Drug Administration/Office of Regulatory Affairs (FDA/ORA) laboratories. Page 1 of 13 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6.C.4. added or electronically ) 7. Definitions 8. Records

More information

QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES

QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,

More information

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.

More information

Definition of the Measurand: CRP

Definition of the Measurand: CRP A Reference Measurement System for C-reactive Protein David M. Bunk, Ph.D. Chemical Science and Technology Laboratory National Institute of Standards and Technology Definition of the Measurand: Human C-reactive

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail inab@inab.ie Web

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting. Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description

More information

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data Low cost ICP-OES analysis high quality data The NEW Thermo Scientific icap 7000 Series ICP-OES provides low cost multi-element analysis for measuring trace elements in a diverse sample range. The instrument

More information

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis QA/QC To identify, measure and keep systematic and random errors under control in planning, sampling, analysis and reporting of data. Quality assurance = how samples and data are collected and managed.

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

Certificate Standard Reference Material 1866b

Certificate Standard Reference Material 1866b National Institute of Standards & Technology Certificate Standard Reference Material 1866b Common Commercial Asbestos This standard reference material (SRM) is comprised of three commercial-grade asbestos

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

1.0 What Are the Purpose and Applicability of Performance Specification 11?

1.0 What Are the Purpose and Applicability of Performance Specification 11? While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

Certifying Reference Materials to ISO Guide 34

Certifying Reference Materials to ISO Guide 34 Certifying Reference Materials to ISO Guide 34 Gill Holcombe Head of Reference Materials Production Environmental Laboratory Quality Day 17 November 2011 4005 Outline LGC background Quality systems - ISO

More information

LAB 37 Edition 3 June 2013

LAB 37 Edition 3 June 2013 LAB 37 Edition 3 June 2013 Accreditation Requirements for Sampling and Testing in Accordance with the Drinking Water Testing Specification (DWTS) CONTENTS SECTION PAGE 1 Introduction 2 2 Regulatory requirements

More information

EUROLAB 25 years. The role of metrology and testing to characterize materials and products. EUROLAB 25th Anniversary Seminar

EUROLAB 25 years. The role of metrology and testing to characterize materials and products. EUROLAB 25th Anniversary Seminar EUROLAB 25 years The role of metrology and testing to characterize materials and products April 9, 2015, Odense Danmark 1 Origin of Conformity Assessment The New Approach for the European Economic Community

More information

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly

More information

Process Measurement Assurance Program For U.S. State Metrology Laboratories 1

Process Measurement Assurance Program For U.S. State Metrology Laboratories 1 Process Measurement Assurance Program For U.S. State Metrology Laboratories 1 Jerry L. Everhart JTI Systems, Inc. And Georgia L. Harris NIST Office of Weights and Measures Abstract This paper describes

More information

VHG Labs, Inc. Standards and Supplies for Spectrochemical Oil Analysis

VHG Labs, Inc. Standards and Supplies for Spectrochemical Oil Analysis VHG Labs, Inc. and Supplies for Spectrochemical Oil Analysis Section 1 Multi-Element Coolant Standard Coolant Standard The following standard is intended for use in the analysis of metals in ethylene glycol

More information

Chapter 4. Evaluating Analytical Data. When using an analytical method we make three separate evaluations of experimental error.

Chapter 4. Evaluating Analytical Data. When using an analytical method we make three separate evaluations of experimental error. Chapter 4 Evaluating Analytical Data Chapter Overview 4A Characterizing Measurements and Results 4B Characterizing Experimental Errors 4C Propagation of Uncertainty 4D The Distribution of Measurements

More information

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon

More information

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

Maria Belli. (former APAT) IX. Congress of clinical chemistry & laboratory medicine Prague 20-22 September 2009

Maria Belli. (former APAT) IX. Congress of clinical chemistry & laboratory medicine Prague 20-22 September 2009 IX. Congress of clinical chemistry & laboratory medicine Prague 20-22 September 2009 Maria Belli Istituto Superiore per la Protezione e la Ricerca Ambientale (former APAT) production and characterization

More information

Experis Gas Mixtures. When accuracy, stability and reliability are a must

Experis Gas Mixtures. When accuracy, stability and reliability are a must Experis Gas Mixtures When accuracy, stability and reliability are a must Making the Difference For over 70 years, Air Products has enabled customers to become more productive, energy efficient and sustainable.

More information

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014 Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information