1 Chapter 10 Immunofluorescence J. Paul Robinson PhD, Jennifer Sturgis BS and George L. Kumar PhD Immunofluorescence (IF) is a common laboratory technique used in almost all aspects of biology. This technique based on pioneering work by Coons and Kaplan (1, 2), and later by Mary Osborne (3), has been widely used both in research and clinical diagnostics. Applications include the evaluation of cells in suspension, cultured cells, tissue, beads and microarrays for the detection of specific proteins. IF techniques can be used on both fresh and fixed samples. In IF techniques, antibodies are chemically conjugated to fluorescent dyes such as fluorescein isothiocyanate (FITC) or tetramethyl rhodamine isothiocyanate (TRITC). These labeled antibodies bind (directly or indirectly) to the antigen of interest which allows for antigen detection through fluorescence techniques. The fluorescence can then be quantified using a flow cytometer, array scanner or automated imaging instrument, or visualized using fluorescence or confocal microscopy. The two main methods of immunofluorescent labeling are direct and indirect. Less frequently used is direct immunofluorescence whereby the antibody against the molecule of interest is chemically conjugated to a fluorescent dye. In indirect immunofluorescence, the antibody specific for the molecule of interest (called the primary antibody) is unlabeled, and a second anti-immunoglobulin antibody directed toward the constant portion of the first antibody (called the secondary antibody) is tagged with the fluorescent dye (Figure 1). Advantages of direct immunofluorescence include shorter sample staining times and simpler dual and triple labeling procedures. In cases where one has multiple antibodies raised in the same species, for example two mouse monoclonals, a direct labeling may be necessary. Disadvantages of direct immunofluorescence include lower signal, generally higher cost, less flexibility and difficulties with the labeling procedure when commercially labeled direct conjugates are unavailable. Advantages of indirect immunofluorescence include greater sensitivity than direct immunofluorescence. There is amplification of the signal in indirect immunofluorescence because more than one secondary antibody can attach to each primary (see Figure 1). Commercially produced secondary antibodies are relatively inexpensive, available in an array of colors, and quality controlled. Disadvantages of indirect immunofluorescence include the potential for cross-reactivity and the need to find primary antibodies that are not raised in the same species or of different isotypes when performing multiple-labeling experiments. Samples with endogenous immunoglobulin may exhibit a high background. Primary Antibody Direct Immunofluorescence Fluorochrome Indirect Immunofluorescence Secondary Antibody Figure 1. Schematic of direct and indirect immunofluorescence. Figure 2. Cultured pulmonary artery endothelial cells stained for tubulin (red), actin (green) and DNA (blue). The dual immunofluorescence procedure used rabbit anti-actin IgG and mouse anti-alpha tubulin IgG as primary antibodies. The secondary antibodies used were Texas Red-conjugated goat, anti-rabbit IgG and FITC-conjugated goat, anti-mouse IgG. The sample was also stained with the DNA-specific dye Hoechst Scale bar is equal to 20 microns. IHC Staining Methods, Fifth Edition 61
2 Principle of Fluorescence Fluorescence and phosphorescence are both types of luminescence. When molecules with luminescent properties absorb light, they emit light of a different wavelength. With fluorescence the emission of light occurs extremely rapidly after the absorption of excitation light, whereas with phosphorescence emission continues for milliseconds to minutes after the energy source has been removed. Fluorescent materials give off light because of their atomic structure. Electrons are arranged in discrete energy levels surrounding the atom s nucleus with each level having a predetermined amount of energy. When an electron absorbs the energy from a photon of light (Figure 3) it becomes excited and jumps to a higher, less stable energy level. The excited state does not last long. The half-life of the excited state is generally less than 10 (8) seconds. The electron loses a small amount of energy as heat and the remainder of the extra energy is given off in the form of a photon. The emitted fluorescence has a lower energy than the absorbed light, so the wavelength of the emitted light is longer than that of the excitation light (except in the case of multiphoton excitation). Emission Excitation will be produced. This optimal wavelength is called the excitation peak. Similarly, the light produced by fluorochromes has a range of wavelengths. The emission of light from fluorescein ranges from 490 nm to 630 nm, and the emission peak is approximately 515 nm. Since the phenomenon of fluorescence was first explained by a British scientist, Sir George Stokes, in 1852, the shift in wavelength from short to long during fluorescence is called Stokes shift (Figure 4). Peak ex. = 495 nm Stokes Shift FITC ex. FITC em. FITC em. at 470 nm Peak em. = 520 nm Figure 4. Excitation and emission spectrum of fluorescein. When fluorescein is excited at a wavelength other than its peak excitation (470 nm in this example), the shape of the emission curve (darker green) remains the same, but the relative intensity is reduced. The efficiency of the excitation at 470 nm is 45% of peak excitation. Figure 3. Principle of fluorescence. Excitation spectrum: Blue Emission spectrum: Green Nucleus A range of wavelengths of light can excite the electrons of a fluorochrome. For example, fluorescein will fluoresce when hit by light with any wavelength between 450 nm and 520 nm. However, the closer the excitation wavelength is to 495 nm, the more fluorescence Some fluorochromes have a small Stokes shift while other fluorescent compounds have large Stokes shifts. For example, the fluorochrome fluorescein can be excited by blue-green light, and its Stokes shift is only about 20 nm, which means that the light emitted is green. This contrasts with another fluorochrome, phycoerythrin, which also can be excited by blue-green light, but has a large Stokes shift. Thus, the light emitted is yellow-orange. In immunofluorescence, a single wavelength can be used to excite several fluorochromes with different Stokes shifts and thereby produce a variety of fluorescent colors as shown in Figure IHC Staining Methods, Fifth Edition
3 The ideal fluorochrome would be a molecule with the following properties: An absorption peak at an excitation wavelength available on the fluorescence detection instrument (large extinction coefficient at the wavelength of excitation) Bright fluorescence (high quantum yield) A narrow emission spectrum that falls within one of the instrument s detection bands Good photostability and Fluorescence properties that are not significantly altered by conjugation to an antibody or by the local environment of the sample Figure 5. Excitation of three spectrally distinct fluorochromes using a single laser line. The example in Figure 5 shows a single wavelength at 488 nm (blue line) exciting three different fluorochromes identified by their absorption curves on the left of the figure (blue line). Each fluorochrome is excited at a different efficiency and, therefore, the resulting emission will be at different intensities for equivalent fluorochrome concentrations. Knowing the excitation and emission properties of fluorescent compounds makes it possible to select combinations of fluorochromes that will work together. However, for a fluorochrome to be useful in a biological application it must attach to or be contained within a structure of biological significance. Fluorochromes can be attached to antibodies which will then bind to specific chemical structures on or inside cells. Many other chemical and physical properties of fluorochromes determine when and where these dyes are useful in various biological assays. For example, some of the fluorochromes that bind to DNA, such as Hoechst 33342, can get into living cells, but most DNA-binding fluorochromes cannot get past the cell membrane. Those fluorescent dyes that cannot get past an intact cell membrane, such as propidium iodide (PI), are often used to distinguish live from dead and dying cells. Limitations of IF Techniques Photobleaching As with most fluorescence techniques, a significant problem with immunofluorescence is photobleaching (and phototoxicity). Photobleaching is the photochemical destruction of a fluorophore due to the generation of reactive oxygen species in the specimen as a byproduct of fluorescence excitation (Figure 6). Although the exact mechanism of photobleaching is unknown, it is thought that the primary causative mechanism appears to be photosensitization of singlet oxygen ( 1 O 2 ) (see glossary for details) generation by the dye triplet-excited state and reference (4) for details). Photobleaching can be minimized by: (a) decreasing the excitation light in both intensity and duration, (b) reducing the availability of singlet oxygen ( 1 O 2 ) by the addition of singlet oxygen scavengers (= antifade reagents), and (c) using a low concentration of a fluorochrome with highquantum efficiency. IHC Staining Methods, Fifth Edition 63
4 Photobleaching is the irreversible decomposition of the fluorescent molecules in the excited state because of their interaction with molecular oxygen prior to emission Energy Absorption (blue) Intersystem green Crossing fluorescence emission Ground Energy State Phosphoresence Figure 6. Illustration of how a singlet-excited state can convert to a tripletexcited state. Autofluorescence Biological autofluorescence in mammalian cells due to flavin coenzymes (FAD and FMN: absorption, 450 nm; emission, 515 nm) and reduced pyridine nucleotides (NADH: absorption, 340 nm; emission, 460 nm) can be problematic in the detection of fluorescence probes in tissues and cells. Fixation with aldehydes, particularly glutaraldehyde, can result in high levels of autofluorescence. This can be minimized in fixed cells by washing with 0.1% sodium borohydride in phosphate-buffered saline (5) prior to antibody incubation. Problems due to autofluorescence can be minimized by selecting probes and optical filters that maximize the fluorescence signal relative to the autofluorescence. Other factors that limit IF include the performance of the detection instrument (i.e. how well the microscope has been calibrated and set), the specificity of the antibodies, and the specimen preparation. Figure 7. NADH autofluorescence in a human colon carcinoma cell line (HCT116). Ultra-violet excitation at 363 nm was used and the emitted fluorescence greater than 440 nm was collected. Scale bar is 10 microns. Credit: Giselle M. Knudsen, Department of Medicinal Chemistry and Molecular Pharmacology, Purdue University, West Lafayette, IN, USA. Fluorescence Overlap One of the problems that must be dealt with when measuring fluorescence of more than one color is the possibility that the emission signals overlap. It is necessary to remove the overlapping signal or it will give a false level for one or more colors. For example, as shown in figure 8, there is significant overlap when using FITC and PE. A range of wavelengths will be collected for each detection channel. In the figure, these are identified as the fluorescein detector bandwidth and the PE detector bandwidth. These band-pass optical filters will allow photons within this wavelength range to reach the detector. However, as can be seen in figure 8, there is a very small amount of PE fluorescence which is within the FITC band and similarly a small amount of FITC fluorescence within the PE band. These must be electronically removed or the measurement for each detector will overestimate the actual signal. This process is called fluorescence 64 IHC Staining Methods, Fifth Edition
5 compensation and can be automatically calculated in many detection systems using single positive controls. FITC detector bandwidth PE detector bandwidth PE ex. PE em. FITC ex. FITC em. immunofluorescence are numerous, and the technique has proven to be a powerful tool for determining the cellular distribution of known antigens in frozen tissues or in the localization of specific DNA sequences on chromosomes. The method has achieved the status of combining high sensitivity with high resolution in the visualization of antigens and will be a major tool for many years to come that any pathologist studying cells or molecules cannot afford to ignore. Figure 8. Fluorescence overlap of FITC and PE. Overlap Applications of Immunofluorescence in Pathology Some practical applications of immunofluorescence in diagnostic pathology are: Analysis of antigens in fresh, frozen or fixed tissues; sub-cellular localization of antigens in tissue culture monolayers; observation of bacterial or parasitic specimens; Detection and localization of the presence or absence of specific DNA sequences on chromosomes; and For a methodology article on immunofluorescence labeling of formalin-fixed, paraffin-embedded tissue, see reference 7. References 1. Coons AH, et al. Immunological properties of an antibody containing a fluorescent group. Proc Soc Exp Biol Med 1941;47, Coons AH and Kaplan MH () Localization of antigen in tissue cells. II. Improvements in a method for the detection of antigen by means of fluorescent antibody. Journal of Exp Med 1950; 91,1. 3. Weber K, et al. Specific visualization of tubulin-containing structures in tissue culture cells by immunofluorescence. Cytoplasmic microtubules, vinblastine-induced paracrystals, and mitotic figures. Exp Cell Res 1975;95, Johnson ID. Practical considerations in the selection and application of fluorescent probes. In: Handbook of biological confocal microscopy, 3rd ed. (J.B. Pawley. Ed), Plenum Press. New York pp Bacallao R, et al. Guiding principles of specimen preservation for confocal fluorescence microscopy. In: Handbook of biological confocal microscopy,3rd ed. (J.B. Pawley. Ed), Plenum Press. New York pp Jennette JC and Wick MR. Immunohischemical techniques. In Immunohistology in diagnostic pathology, Ed. Jennette JC. CRC Press. Boca Raton, FL. USA Pp Robertson D, et al. Multiple immunofluorescence labelling of formaqlinfixed paraffin-embedded (FFPE) tissue. BMC Cell Biology 2008; 9:13. Defining the spatial-temporal patterns of gene expression within cells/tissues. The most widely used method of IF in pathology is indirect IF. However, in some very specialized applications direct IF has been used for localization of IgG in immune complexes along the dermal-epidermal junction of skin biopsies from patients suffering from systemic lupus erythematosus (6). In summary, immunofluorescence is the visualization of antigens within cells using antibodies as fluorescent probes. The benefits of IHC Staining Methods, Fifth Edition 65
IMMUNOFLUORESCENCE MODULE 62.1 INTRODUCTION. Notes
Immunofluorescence MODULE 62 IMMUNOFLUORESCENCE 62.1 INTRODUCTION Immunofluorescence (IF) is one of the very common laboratory techniques used in almost all disciplines of Biology including Medicine for
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
Labeling and Detection. The best and brightest Alexa Fluor dyes
The best and brightest Alexa Fluor dyes Superior alternatives to standard dyes Brighter conjugate fluorescence Superior photostability Ideal spectral match for all of the popular filters Now you have a
Chapter 12 Filters for FISH Imaging
Chapter 12 Filters for FISH Imaging Dan Osborn The application of in situ hybridization (ISH) has advanced from short lived, non-specific isotopic methods, to very specific, long lived, multiple color
FluoProbes dyes. FluoProbes dyes directly conjugated to secondary antibodies
FluoProbes dyes Interchim's FluoProbes dyes exhibit optimal fluorescence intensity combined with excellent photostability. Activated dyes are available independent (please request information), conjugated
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions
Chapter 6 Antigen-Antibody Interactions: Principles and Applications Antigen-Antibody Properties You must remember antibody affinity (single) VS avidity (multiple) High affinity: bound tightly and longer!
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
ArC Amine Reactive Compensation Bead Kit
ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting
Technical Guide SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting Principles of SmartFlare technology RNA detection traditionally requires transfection, laborious sample prep,
MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
These particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
Principles of Flowcytometry
Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal
Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose
Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:
Annexin V-EGFP Apoptosis Detection Kit
ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended
Technical Note. Roche Applied Science. No. LC 18/2004. Assay Formats for Use in Real-Time PCR
Roche Applied Science Technical Note No. LC 18/2004 Purpose of this Note Assay Formats for Use in Real-Time PCR The LightCycler Instrument uses several detection channels to monitor the amplification of
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating
PROTOCOL. Immunocytochemistry (ICC) MATERIALS AND EQUIPMENT REQUIRED
PROTOCOL Immunocytochemistry (ICC) 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 11-07 MATERIALS AND EQUIPMENT REQUIRED Materials: MitoSciences primary monoclonal antibody/antibodies Fluorophore-conjugated
Advantages of the Alexa Fluor 488 secondary antibody conjugates
Secondary Antibodies Looking on the bright side with Alexa Fluor secondary antibody conjugates Alexa Fluor 488 dye a superior alternative to FITC You have a choice when it comes to green-fluorescent dye
Three Rules for Compensation Controls
From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for
Zecotek S Light Projection Network Marketing
White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company
WESTERN BLOT DETECTION KIT Buffers and detection reagents for up to ten 10 x 10 cm 2 blots. Fluorescent detection via: Goat anti-mouse SureLight P3
WESTERN BLOT DETECTION KIT Buffers and detection reagents for up to ten 10 x 10 cm 2 blots Fluorescent detection via: Goat anti-mouse SureLight P3 Cat. #: WK-P112 6440 Dobbin Road, Suite D Phone (443)
March 19, 2014. Dear Dr. Duvall, Dr. Hambrick, and Ms. Smith,
Dr. Daniel Duvall, Medical Officer Center for Medicare, Hospital and Ambulatory Policy Group Centers for Medicare and Medicaid Services 7500 Security Boulevard Baltimore, Maryland 21244 Dr. Edith Hambrick,
Annexin V-FITC Apoptosis Detection Kit
ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research
KMS-Specialist & Customized Biosimilar Service
KMS-Specialist & Customized Biosimilar Service 1. Polyclonal Antibody Development Service KMS offering a variety of Polyclonal Antibody Services to fit your research and production needs. we develop polyclonal
Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers
An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,
Fluorescent dyes for use with the
Detection of Multiple Reporter Dyes in Real-time, On-line PCR Analysis with the LightCycler System Gregor Sagner, Cornelia Goldstein, and Rob van Miltenburg Roche Molecular Biochemicals, Penzberg, Germany
DNA Detection. Chapter 13
DNA Detection Chapter 13 Detecting DNA molecules Once you have your DNA separated by size Now you need to be able to visualize the DNA on the gel somehow Original techniques: Radioactive label, silver
Goat Anti Mouse IgG Antibodies
Goat Anti Mouse IgG Antibodies Table 1 Contents and storage Material Amount Concentration Storage Stability Whole antibodies 0.5 ml F(ab ) 2 Fragments 250 µl 2 mg/ml in 0.1 M sodium phosphate, 0.1 M NaCl,
Annexin V-FITC Apoptosis Detection Kit
ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research
LIVE/DEAD Fixable Dead Cell Stain Kits
USER GUIDE LIVE/DEAD Fixable Dead Cell Stain Kits Pub. No. MAN0002416 (MP34955) Rev. A.0 Table 1. Contents and storage Material Amount Storage Stability Individual Kits: Blue, violet, aqua, yellow-, green,
EdU Flow Cytometry Kit. User Manual
User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg
Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation
Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap
QDot Nanocrystals Insights
QDot Nanocrystals Insights Invitrogen Corporation Eugene, OR. HeLa cell labeled with three Qdot conjugates for nucleus, Golgi, and tubulin What are quantum dots? Highly fluorescent, nanometer-size, crystals
1.Gene Synthesis. 2.Peptide & Phospho-P. Assembly PCR. Design & Synthesis. Advantages. Specifications. Advantages
1.Gene Synthesis Assembly PCR Looking for a cdna for your research but could not fish out the gene through traditional cloning methods or a supplier? Abnova provides a gene synthesis service via assembly
Chapter 6: Antigen-Antibody Interactions
Chapter 6: Antigen-Antibody Interactions I. Strength of Ag-Ab interactions A. Antibody Affinity - strength of total noncovalent interactions between single Ag-binding site on an Ab and a single epitope
Flow Cytometry. flow cytometer DNA apoptosis ph
Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal
Principles of Immunohistochemistry Queen s Laboratory For Molecular Pathology
Principles of Immunohistochemistry Queen s Laboratory For Molecular Pathology Table of Contents POLYCLONAL VERSUS MONOCLONAL ANTIBODIES...PAGE 3 DIRECT & INDIRECT ASSAYS.PAGE 4 LABELS..PAGE 5 DETECTION...PAGE
Investigating the role of a Cryptosporidium parum apyrase in infection
Investigating the role of a Cryptosporidium parum apyrase in infection David Riccardi and Patricio Manque Abstract This project attempted to characterize the function of a Cryptosporidium parvum apyrase
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
Standardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
Outline. 1. Experiment. 2. Sample analysis and storage. 3. Image analysis and presenting data. 4. Probemaker
Tips and tricks Note: this is just an informative document with general recommendations. Please contact [email protected] should you have any queries. Document last reviewed 2011-11-17 Outline 1. Experiment
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
Guide to Purification of Polyclonal Antibodies
Guide to Purification of Polyclonal Antibodies When choosing a polyclonal antibody, either as a primary or secondary antibody in an immunoassay, researchers are often inundated with an array of antibody
Flow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
Topic: Serological reactions: the purpose and a principle of reactions. Agglutination test. Precipitation test. CFT, IFT, ELISA, RIA.
Topic: Serological reactions: the purpose and a principle of reactions. Agglutination test. Precipitation test. CFT, IFT, ELISA, RIA. Serology is the study and use of immunological tests to diagnose and
Goat Anti Rabbit IgG Antibodies
Goat Anti Rabbit IgG Antibodies Table 1 Contents and storage Material Amount Concentration Storage Upon Receipt Stability Whole antibodies 0.5 ml F(ab') 2 fragments 250 µl 2 mg/ml in 0.1 M sodium phosphate,
Islet Viability Assessment by Single Cell Flow Cytometry
Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing
Predict Reactivity Note Chicken (100%), Sheep (100%), Rhesus Monkey (100%), Chimpanzee (100%), Bovine (100%), Guinea pig (100%)
Datasheet GeneTex, Inc : Toll Free 1-877-GeneTex (1-877-436-3839) Fax:1-949-309-2888 [email protected] GeneTex International Corporation : Tel:886-3-6208988 Fax:886-3-6208989 [email protected] Date :
FITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry
FITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry Catalog no. V13242 Table 1. Contents and storage information. Material Amount Composition Storage* Stability FITC annexin
Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation
Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F:
protocol handbook 3D cell culture mimsys G hydrogel
handbook 3D cell culture mimsys G hydrogel supporting real discovery handbook Index 01 Cell encapsulation in hydrogels 02 Cell viability by MTS assay 03 Live/Dead assay to assess cell viability 04 Fluorescent
Uses of Flow Cytometry
Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7
The immune response Antibodies Antigens Epitopes (antigenic determinants) the part of a protein antigen recognized by an antibody Haptens small
The immune response Antibodies Antigens Epitopes (antigenic determinants) the part of a protein antigen recognized by an antibody Haptens small molecules that can elicit an immune response when linked
Compensation in Flow Cytometry
Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome
Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model
A p p l i c a t i o n N o t e Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model Brad Larson and Peter Banks, BioTek Instruments, Inc., Winooski, VT Grant Cameron, TAP Biosystems
Cell Cycle Phase Determination Kit
Cell Cycle Phase Determination Kit Item No. 10009349 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Safety
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
TECHNICAL BULLETIN. FluoroTag FITC Conjugation Kit. Product Number FITC1 Storage Temperature 2 8 C
FluoroTag FITC Conjugation Kit Product Number FITC1 Storage Temperature 2 8 C TECHNICAL BULLETIN Product Description The FluoroTag FITC Conjugation Kit is suitable for the conjugation of polyclonal and
Chapter 18: Applications of Immunology
Chapter 18: Applications of Immunology 1. Vaccinations 2. Monoclonal vs Polyclonal Ab 3. Diagnostic Immunology 1. Vaccinations What is Vaccination? A method of inducing artificial immunity by exposing
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
INSIDE THE BLACK BOX
FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Two-photon FCS Tutorial. Berland Lab Department of Physics Emory University
Two-photon FCS Tutorial Berland Lab Department of Physics Emory University What is FCS? FCS : Fluorescence Correlation Spectroscopy FCS is a technique for acquiring dynamical information from spontaneous
竞 争 性 分 析 Epitope Mapping 实 验 方 法
竞 争 性 分 析 Epitope Mapping 实 验 方 法 ABSTRACT The simplest way to determine whether two monoclonal antibodies bind to distinct sites on a protein antigen is to carry out a competition assay. The assay can
Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.
Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off
STANDING Tall CUSTOM PRODUCTS AND SERVICES
STANDING Tall CUSTOM PRODUCTS AND SERVICES COME GROW WITH US! Contact KPL and let us know how we can support your OEM needs. Quality antibodies, substrates, and immunoassay support reagents Large scale
Electromagnetic Radiation (EMR) and Remote Sensing
Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through
7.1 A Wide Variety of Protein Conjugates
7.1 A Wide Variety of Protein Conjugates Overview of Molecular Probes Protein Conjugates Antibody and Avidin Conjugates The quality of a conjugate depends to a large degree on the quality of the protein
INTERPRETATION INFORMATION SHEET
Creative Testing Solutions 2424 West Erie Dr. 2205 Highway 121 10100 Martin Luther King Jr. St. No. Tempe, AZ 85282 Bedford, TX 76021 St. Petersburg, FL 33716 INTERPRETATION INFORMATION SHEET Human Immunodeficiency
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)
Cell Viability Measurement
Cell Viability Measurement Viability is a measure of the metabolic state of a cell population which is indicative of the potential for growth Most common method: membrane integrity Trypan blue exclusion:
Chapter 15 Multi-Staining Immunohistochemistry
Chapter 15 Multi-Staining Immunohistochemistry Nanna K. Christensen PhD and Lars Winther PhD Immunohistochemistry (IHC) has become an established tool for both research and diagnostic purposes. However,
THE His Tag Antibody, mab, Mouse
THE His Tag Antibody, mab, Mouse Cat. No. A00186 Technical Manual No. TM0243 Update date 01052011 I Description.... 1 II Key Features. 2 III Storage 2 IV Applications.... 2 V Examples - ELISA..... 2 VI
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Really Good Antibodies. Secondary Antibodies. Over 670 Anti-Mouse, Rabbit, Goat and Rat Secondary Antibodies. bethyl.com
Really Good Antibodies Secondary Antibodies Over 670 Anti-Mouse, Rabbit, Goat and Rat Secondary Antibodies bethyl.com Bethyl Laboratories, Inc. has been dedicated to supporting scientific discovery through
HER2 FISH pharmdx. Assay Kit
PATHOLOGY HER2 FISH pharmdx Assay Kit HER2 FISH pharmdx Clarity You Can Count On Reliable Results at a Glance The robust HER2 FISH pharmdx assay offers bright and distinct signals for easy and fast reading
Hypoxyprobe -1 Plus Kit Kit contents:
Updated 2015 1 PRODUCT INSERT Hypoxyprobe, Inc 121 Middlesex Turnpike Burlington, MA 01803 USA www.hypoxyprobe.com Hypoxyprobe -1 Plus Kit Kit contents: Solid pimonidazole HCl (Hypoxyprobe -1) FITC conjugated
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
Fluorescence Compensation. Jennifer Wilshire, PhD [email protected]
Fluorescence Compensation Jennifer Wilshire, PhD [email protected] Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due
Chapter 2 Antibodies. Contents. Introduction
Chapter 2 Antibodies Keywords Immunohistochemistry Antibody labeling Fluorescence microscopy Fluorescent immunocytochemistry Fluorescent immunohistochemistry Indirect immunocytochemistry Immunostaining
Reactive oxygen species in leaves and how to catch these
Reactive oxygen species in leaves and how to catch these Éva Hideg Molecular Stress- & Photobiology Group, BRC Szeged EPPN Summer School 2013 Outline How to prove that ROS are involved? The central role
SCIENCE WHERE YOU ARE
Newsletter. November 2013 SCIENCE WHERE YOU ARE Dealing with BioNordika includes a number of first class benefits: Next day delivery.... and some times even faster! Free delivery. Focus on your application.
Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available!
Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available! Arrayit SpotLight Two-Color Microarray Scanners provide the market s most affordable
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows
INSTRUCTION Probemaker
INSTRUCTION Probemaker Instructions for Duolink In Situ Probemaker PLUS (Art. no. 92009-0020) and Duolink In Situ Probemaker MINUS (Art. no. 92010-0020) Table of content 1. Introduction 4 2. Applications
Fluorescence Microscopy for an NMR- Biosensor Project
Fluorescence Microscopy for an NMR- Biosensor Project Ole Hirsch Physikalisch-Technische Bundesanstalt Medical Optics Abbestr. -1, 10587 Berlin, Germany Overview NMR Sensor Project Dimensions in biological
Physics 1230: Light and Color
Physics 1230: Light and Color Instructor: Joseph Maclennan TOPIC 3 - Resonance and the Generation of Light http://www.colorado.edu/physics/phys1230 How do we generate light? How do we detect light? Concept
Flow Cytometry A Basic Overview
Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,