Three-dimensional ultrasound imaging

Size: px
Start display at page:

Download "Three-dimensional ultrasound imaging"

Transcription

1 INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) R67 R99 PII: S (01) TOPICAL REVIEW Three-dimensional ultrasound imaging Aaron Fenster 1,Dónal B Downey and H Neale Cardinal The John P Robarts Research Institute, London, Canada and the Department of Diagnostic Radiology and Nuclear Medicine, The University of Western Ontario, London, Canada [email protected] Received 6 November 2000 Abstract Ultrasound is an inexpensive and widely used imaging modality for the diagnosis and staging of a number of diseases. In the past two decades, it has benefited from major advances in technology and has become an indispensable imaging modality, due to its flexibility and non-invasive character. In the last decade, research investigators and commercial companies have further advanced ultrasound imaging with the development of 3D ultrasound. This new imaging approach is rapidly achieving widespread use with numerous applications. The major reason for the increase in the use of 3D ultrasound is related to the limitations of 2D viewing of 3D anatomy, using conventional ultrasound. This occurs because: (a) Conventional ultrasound images are 2D, yet the anatomy is 3D, hence the diagnostician must integrate multiple images in his mind. This practice is inefficient, and may lead to variability and incorrect diagnoses. (b) The 2D ultrasound image represents a thin plane at some arbitrary angle in the body. It is difficult to localize the image plane and reproduce it at a later time for follow-up studies. In this review article we describe how 3D ultrasound imaging overcomes these limitations. Specifically, we describe the developments of a number of 3D ultrasound imaging systems using mechanical, free-hand and 2D array scanning techniques. Reconstruction and viewing methods of the 3D images are described with specific examples. Since 3D ultrasound is used to quantify the volume of organs and pathology, the sources of errors in the reconstruction techniques as well as formulae relating design specification to geometric errors are provided. Finally, methods to measure organ volume from the 3D ultrasound images and sources of errors are described. 1 Address for correspondence: The J P Robarts Research Institute, Imaging Research Laboratories, 100 Perth Drive, London, Ontario, Canada N6A 5K /01/ $ IOP Publishing Ltd Printed in the UK R67

2 R68 A Fenster et al 1. Introduction For most of the last century, medical imaging has involved visualizing the interior structures of the human body with two-dimensional (2D) x-ray images. Although 2D x-ray imaging remains an indispensable imaging modality, it has the disadvantage that almost all three-dimensional (3D) information about the interior of the human body is lost to the physician. In the early 1970s, the introduction of computed tomography (CT) revolutionized diagnostic radiology by providing, for the first time, true 3D anatomical information to the physician. At first this information was simply presented as a series of contiguous 2D image slices of the body. However, the advent of 3D images from CT, and later magnetic resonance imaging (MRI), provided a major impetus to the field of 3D visualization, which has lead to the development of a wide variety of applications in diagnostic medicine (Fishman et al 1991, Robb and Barillot 1989, Vannier et al 1984). Although the origins of medical ultrasound imaging can be traced back half a century to the pioneering work of Wild and Reid (1952), progress was slow for the first three decades, and, outside of research laboratories, medical ultrasound was largely limited to applications in cardiology, obstetrics, and gynaecology. However, in the past two decades, medical ultrasound has benefited from major advances in technology and has become an indispensable imaging modality, due to its flexibility and non-invasive character. Moreover, due to continuous improvements in image quality, and the availability of blood flow information via the Doppler effect, ultrasonography is progressively achieving a greater role in radiology and cardiology, and is finding new roles in image-guided surgery and therapy. In the last decade, research investigators and commercial companies have further advanced ultrasound imaging by combining these advances in ultrasound image quality with recent advances in 3D image visualization, to develop 3D ultrasound imaging. This review article discusses this new imaging modality, which is rapidly achieving widespread application in medical imaging. 2. Limitations of ultrasonography addressed by 3D imaging In a conventional ultrasound examination, the ultrasound transducer is manipulated to obtain a series of 2D ultrasound images, which are mentally combined by the operator to form a subjective impression of the 3D anatomy and pathology. The success of the diagnostic or interventional procedure is thus largely dependent on the skill and experience of the operator in performing these tasks. This is a suboptimal approach, due to the following limitations: Mentally transforming multiple 2D images to form a 3D impression of the anatomy and pathology is not only time-consuming and inefficient, but is also more importantly, variable and subjective, which can lead to incorrect decisions in diagnosis, and in the planning and delivery of therapy. Diagnostic (e.g. obstetric) and therapeutic (e.g. staging and planning) decisions often require accurate estimation of organ or tumour volume. Conventional 2D ultrasound techniques calculate volume from simple measurements of height, width and length in two orthogonal views, by assuming an idealized (e.g. ellipsoidal) shape. This practice can potentially lead to low accuracy, high variability and operator dependency. Conventional 2D ultrasound is suboptimal for monitoring therapeutic procedures, or for performing quantitative prospective or follow-up studies, due to the difficulty in adjusting the transducer position so that the 2D image plane is at the same anatomical site and in the same orientation as in a previous examination.

3 Three-dimensional ultrasound imaging R69 Because the location and orientation of conventional 2D ultrasound images are determined by those of the transducer, some views are impossible to achieve because of restrictions imposed by the patient s anatomy or position. In a 3D ultrasound examination, the 2D ultrasound images are combined by a computer to form an objective 3D image of the anatomy and pathology. This image can then be viewed, manipulated and measured in 3D by the physician on the same or another computer. Also, a 2D cross-sectional image can be generated in any orientation, without restriction, at any anatomical site, which may easily be registered with a previous or subsequent 3D image. Thus, 3D ultrasound imaging promises to overcome the limitations of 2D ultrasound imaging described above. Moreover, unlike CT and MR imaging, in which 2D images are usually acquired at a slow rate as a stack of parallel slices, in a fixed orientation, ultrasound provides tomographic images at a high rate (15 60 s 1 ), and in arbitrary orientations. The high acquisition rate and arbitrary orientation of the images provide unique problems to overcome and opportunities to exploit in extending 2D ultrasound imaging to 3D and dynamic 3D (i.e. 4D) visualization. Over the past two decades, many investigators have attempted to develop 3D ultrasound imaging techniques (Brinkley et al 1982, Fenster and Downey 1996, Ghosh et al 1982, Greenleaf et al 1993, King et al 1993, Nelson and Pretorius 1992, Rankin et al 1993). However, progress has been slow due to the enormous computational requirements which must be met in order to acquire, reconstruct and view 3D information in near real time on low-cost systems. Advances in low-cost computer technology and visualization techniques in the past few years have made 3D ultrasound imaging viable. A number of laboratories have led the way in demonstrating the feasibility of 3D ultrasound imaging and, more recently, commercial companies have begun to provide these options on their ultrasound equipment. A few review articles and two books describing the development of 3D ultrasound imaging have been published in the past few years (Fenster and Downey 1996, Greenleaf et al 1993, Rankin et al 1993, Belohlavek et al 1993, Fenster 1998, Ofili and Nanda 1994, Nelson et al 1999, Nelson and Pretorius 1998, Baba and Jurkovic 1997, De Castro et al 1998). In this review we summarize the various approaches that have been used in the development of 3D ultrasound imaging systems, with particular emphasis on the quantitative aspects of geometric accuracy in viewing the anatomy in 3D and in measuring anatomical volumes with 3D ultrasound imaging. 3. 3D ultrasound scanning techniques Most 3D ultrasound imaging systems make use of conventional 1D ultrasound transducers to acquire a series of 2D ultrasound images, and differ only in the method used to determine the position and orientation of these 2D images within the 3D image volume being examined. The production of 3D images without distortions requires that three factors be optimized: The scanning technique must be either rapid or gated, to avoid image artefacts due to involuntary, respiratory or cardiac motion. The locations and orientations of the acquired 2D images must be accurately known, to avoid geometric distortions in the 3D image that would lead to measurement errors. The scanning apparatus must be simple and convenient to use, so that the scan is not complicated or awkward to perform and hence easily included in the examination procedure. Over the past two decades, four different 3D ultrasound imaging approaches have been pursued: mechanical scanners, free-hand techniques with position sensing, free-hand

4 R70 A Fenster et al Table 1. Summary of 3D scanning methods, acquisition methods of images used to produce the 3D image, and the major disadvantages of the scanning methods. Scanning method Image acquisition method Disadvantages Mechanical Linear Acquired images are parallel to Bulky device each other with equal spacing Tilt Acquired images are fan-like with Resolution degrades with depth equal angular spacing Rotational Acquired images are propeller-like Motion of axis of rotation with equal angular spacing results in artefacts Free-Hand Acoustic Measure time-of-flight of sound Line of sight required and sound from spark gaps on transducer to velocity varies with humidity microphones above patient Articulated arms Measure angulation between Scanning volume limited, flexing movable arms of arms Magnetic sensor Measure magnetic field generated Ferrous metals distort magnetic by transmitter beside the patient field with receiver on transducer Image correlation Measure speckle decorrelation Special computer processor between adjacent images required, compound motion is difficult to track No position sensing Distance or angle between images Cannot measure distance is assumed 2D arrays 2D phased array transmits a System cost and signal/noise diverging pyramidal beam and returned echoes are displayed in real time as multiple planes techniques without position sensing and 2D arrays. These approaches are summarized in table 1 and are discussed below Mechanical scanners In this approach, the anatomy is scanned by using a motorized mechanical apparatus to translate, tilt or rotate a conventional transducer as it rapidly acquires a series of 2D ultrasound images spanning the volume of interest, which are recorded by a computer. Because the scanning protocol is predefined and precisely controlled, the relative position and orientation of each 2D image can be known accurately. Using current computer technology, the acquired 2D images may either be stored in their original digital format in the ultrasound system s computer memory, or the analogue video output from the ultrasound machine may be digitized and stored in an external computer memory. The reconstruction and viewing of the 3D image can then be carried out either in the ultrasound machine s computer or in an external computer, using the predefined geometric parameters that describe the orientation and position of each 2D image within the 3D image volume. Usually, the angular or spatial interval between successive 2D images is made adjustable, so that it can be optimized to minimize the scanning time while still adequately sampling the volume (Smith and Fenster 2000). Various kinds of scanning apparatus have been developed to translate or rotate the conventional transducer in the required manner. These vary in size from small integrated 3D

5 Three-dimensional ultrasound imaging R71 Figure 1. Schematic diagrams showing the four mechanical scanning approaches. (a) Linear scanning approach, in which a series of parallel 2D images are collected and used to reconstruct the 3D image. (b) Tilt scanning approach, in which a series of 2D images are collected as the transducer is tilted and then reconstructed into a 3D image. (c) Tilt scanning approach used with a side-firing transrectal (TRUS) transducer to produce 3D images of the prostate. (d) Rotational scanning approach used with a end-firing endocavity transducer in gynaecological and urological imaging. probes that house the scanning mechanism within the transducer housing, to external fixtures that mechanically hold the housing of a conventional ultrasound probe. Although the integrated 3D probes are larger and heavier than conventional probes, they are usually easy for the operator to use. However, they require the purchase of a special ultrasound machine that can interface with them. On the other hand, while external fixtures are generally bulkier than integrated probes they can be adapted to hold the transducer of any conventional ultrasound machine, obviating the need to purchase a special-purpose 3D ultrasound machine. The image quality and imaging techniques offered by any conventional ultrasound machine can thus also be achieved in 3D. Three basic types of mechanical scanners have been developed, as shown schematically in figure 1: linear, tilt and rotational Linear 3D scanning. In this approach a motorized drive mechanism is used to linearly translate the transducer across the patient s skin, while the 2D images are acquired at regular spatial intervals so that they are uniformly spaced and parallel (figure 1(a)). The ability to vary the translation speed and sampling interval is necessary in order to match the sampling rate to the frame rate of the ultrasound machine and to match the sampling interval to (half) the elevational resolution of the transducer (Smith and Fenster 2000). The simple predefined geometry of the acquired 2D image planes allows for a fast 3D image reconstruction. Thus, using this approach a 3D image can be obtained immediately after performing the scan, and its resolution can be optimized. Because the 3D image is produced from the series of conventional 2D images, its resolution will not be isotropic. In the direction parallel to the acquired 2D image planes, it will equal that of the original 2D images, but in the perpendicular direction it will be equal to the elevational resolution of the transducer. Thus, the 3D image resolution will generally be the worst in the 3D scanning direction, and for optimal results a transducer with good elevational resolution should be used. The linear scanning approach has been successfully implemented in many vascular imaging applications, using B-mode and colour Doppler images of the carotid arteries (Downey

6 R72 A Fenster et al (a) (b) Figure 2. 3D ultrasound images of the carotid arteries of a patient with carotid atherosclerotic disease. The 3D images were obtained using the mechanical linear scanning approach and have been oriented and sliced to reveal the details of the plaque. (a) Transverse slice revealing the complex plaque. (b) Longitudinal slice revealing the extent of the plaque. and Fenster 1995a, Fenster et al 1995, Guo and Fenster 1996, Dicot et al 1991, 1993, Pretorius and Nelson 1992), tumour vascularization (Bamber et al 1992, Carson et al 1993, Downey and Fenster 1995b, King et al 1991), test phantoms (Guo and Fenster 1996, Guo et al 1995, Dabrowski et al 2001) and power Doppler images (Downey and Fenster 1995a, b, Guo and Fenster 1996). An example of a linearly scanned 3D ultrasound image made with an external fixture is given in figure 2, which shows a 3D image of carotid arteries with plaques. In addition to radiological applications, the utility of linear scanning in echocardiology has also been demonstrated, using pullback of a transesophageal (TE) probe whose imaging plane is horizontal (i.e. perpendicular to the probe axis) (Pandian et al 1992, Ross et al 1993, Wollschlager et al 1993).

7 Three-dimensional ultrasound imaging R Tilt 3D scanning. In this approach, a motorized drive mechanism is used to tilt a conventional transducer about an axis parallel to the transducer face while the 2D images are acquired at regular angular intervals, so that they form a fan of images radial to the axis as shown in figure 1(b). This type of motion is possible with an integrated 3D probe as well as with an external fixture. In either case, the probe housing remains at a single location on the patient s skin. Because of the fan-like geometry of the acquired 2D images, a large region of interest can be swept out with an angular separation (typically 0.5 to 1.0 ) that can be adjustably predefined to yield high quality images in a variety of applications (Fenster et al 1995, Delabays et al 1995, Downey et al 1995a, b, Gilja et al 1994, Sohn et al 1992). This approach lends itself to compact designs, both for integrated 3D probes and for external fixtures. Special 3D ultrasound systems with integrated 3D probes, well-suited for abdominal and obstetric imaging, have been successfully demonstrated by Kretztechnik (Zipf, Austria) and Aloka (Korea). This compactness allows for easy hand-held positioning and manipulation. Also, with suitable a total scanning angle, angular interval and ultrasound frame rate, the scanning time can be made short. For example, with a total scanning angle of 90, an angular interval of 1 and a 30 Hz acquisition rate, the scanning time is only 3s. As with linear scanning, the simple predefined geometry of the 2D image planes allows for a fast 3D image reconstruction, which can therefore be performed immediately after the scan is complete. Also, the 3D image resolution will again not be isotropic. Typically, the resolution will degrade away from the axis of rotation, for two reasons. First, because of the fan-like geometry of the acquired 2D images, the linear distance between acquired image planes increases with distance from the axis, resulting in decreased spatial sampling and lower spatial resolution. Second, because of beam spreading in the elevational (or scan) direction, as well as within the acquired image plane, the resolution will degrade with distance from the transducer, which is located near the axis (Blake et al 2000). In a variation of tilt scanning, an external fixture is used to rotate a cylindrical endocavity probe, such as a transoesophageal (TE) or transrectal ultrasound (TRUS) probe, about its axis to perform the scan. The 2D images are acquired by a side-firing linear transducer array parallel to the probe axis, and they again form a fan of images radial to the axis, as shown in figure 1(c), which span a total scanning angle of about 80 to 110 (Tong et al 1996, 1998). This approach has been successfully applied to prostate imaging (Downey and Fenster 1995a, b, Fenster et al 1995, Tong et al 1996, Elliot et al 1996), as illustrated in figure 3, and to 3D ultrasound guided cryosurgery (Downey et al 1995c, Chin et al 1998, 1999, Onik et al 1996), and it has been commercialized by Life Imaging Systems Inc. (London, Canada). Also, an application of this approach to echocardiography has been demonstrated by TomTec Inc. (Munich, Germany) using a TE probe whose imaging plane is vertical (i.e. parallel to the probe axis) (Belohlavek et al 1993, Martin and Bashein 1989) Rotational 3D scanning. In this approach, a motorized mechanism rotates the transducer array about a fixed axis that perpendicularly bisects the transducer array through an angle of at least 180, while the 2D images are acquired. Thus, in this scanning geometry, the acquired images sweep out a conical volume about the axis, in a propeller-like pattern, as shown in figure 1(d). This approach has also been implemented with both an integrated 3D probe and an external fixture (e.g. attached to an end-firing TRUS probe). As with the other scanning approaches, the 3D image resolution will not be isotropic. Because the acquired 2D images all intersect along the axis, spatial sampling will be highest near the axis and poorest away from it. Also, both axial and elevational resolution in the acquired 2D images will decrease with distance from the transducer. The combination of these

8 R74 A Fenster et al (a) (b) Figure 3. 3D images of two prostates with cancer. Both images were obtained with the tilt scanning approach using a side-firing transrectal transducer. The image has been sliced to reveal a tumour (arrow pointing to dark lesion) and its relationship to the seminal vesicle and rectal wall. Image (b) is courtesy of Life Imaging Systems Inc. effects will cause the 3D image resolution to vary in a complicated manner, being highest near the transducer and near the axis, and degrading with distance from either. This technique is particularly sensitive to operator or patient motion. Because the acquired 2D images all intersect along the rotational axis at the centre of the 3D image, any motion of the axis or the patient during the scan will cause the axial pixels to have different values in different 2D images. For example, images acquired 180 apart will then not be mirror images of each other. These inconsistencies will result in image artefacts or resolution loss on the axis, depending on the method of reconstruction. Similarly, artefacts will occur if the axis of rotation is not accurately known, or if it is not within the 2D image plane of the transducer. This approach has successfully been used for transrectal imaging of the prostate and for endovaginal imaging, as illustrated in figure 4. For imaging the prostate, Downey and others

9 Three-dimensional ultrasound imaging R75 Figure 4. 3D image of a pregnant uterus with twins. The image has been sliced to reveal the two gestational sacs. This 3D image was obtained using the rotational scanning approach using an end-firing endovaginal transducer. acquired 200 images spanning 200 in 13 s (Downey and Fenster 1995b, Tong et al 1996, Elliot et al 1996). Also, rapid imaging of the heart with a multiplane transoesophageal transducer has been demonstrated by other investigators (Ghosh et al 1982, McCann et al 1987, Roelandt et al 1994) Free-hand scanning with position sensing The mechanical scanning approaches described above offer short imaging times, high-quality 3D images and fast reconstruction times. However, the bulkiness and weight of the scanning apparatus sometimes make it inconvenient to use, and large structures are difficult to scan. To overcome this problem, free-hand scanning techniques that do not require a motorized fixture have been developed by many investigators. In these approaches, a sensor is attached to the transducer to measure its position and orientation. Thus, the transducer can be held by the operator and be manipulated in the usual manner over the anatomy to be imaged. While the transducer is being manipulated, the acquired 2D images are stored by a computer together with their positions and orientations. This information is then used to reconstruct the 3D image. Since the relative locations of the acquired 2D images are not predefined, the operator must ensure that the spatial sampling is appropriate and that the set of 2D images has no significant gaps. Several free-hand scanning approaches have been developed, making use of four different sensing approaches: articulated arms, acoustic sensing, magnetic field sensing and image-based sensing Articulated arm 3D scanning. Position and orientation sensing can be achieved by mounting the ultrasound transducer on a multiple-jointed mechanical arm system.

10 R76 A Fenster et al Figure 5. Schematic diagram showing two free-hand scanning approaches: (a) acoustic tracking, (b) magnetic field sensing. The arrangement of the acquired 2D images has no regular geometry. For accurate 3D reconstruction, the position and orientation of each plane must be known accurately. Potentiometers located at the joints of the movable arms provide the information necessary to calculate the relative position and orientation of the acquired 2D images. This arrangement allows the operator to manipulate the transducer while the computer records the 2D images and the relative angles of all the arms. Using this information, the position and orientation of the transducer is calculated for each acquired 2D image, allowing the 3D image to be reconstructed. To avoid 3D image distortions and inaccuracies, the arms must not flex. Improved performance can be achieved by keeping each arm as short as possible and reducing the number of movable joints. Thus, increased 3D image quality can be achieved by reducing the scanning flexibility and the maximum size of the scanned volume (Geiser et al 1982) Acoustic free-hand 3D scanning. In this approach, an array of three sound-emitting devices, such as spark gaps, are mounted on the transducer, and an array of fixed microphones are mounted above the patient, as illustrated in figure 5(a). The microphones continuously receive sound pulses from the emitters while the transducer is being manipulated and the 2D images are being acquired. The position and orientation of the transducer for each acquired 2D image can then be determined from knowledge of the speed of sound in air, the measured timesof-flight from each emitter to each microphone and the fixed locations of the microphones. The microphones must be placed over the patient, to provide unobstructed lines of sight to the emitters, and must be close enough that the sound pulses can be detected with a good signal-to-noise ratio (Brinkley et al 1982, 1984, King et al 1990, 1992, Levine et al 1989, Moritz et al 1983, Rivera et al 1994, Weiss et al 1983) Free-hand 3D scanning with magnetic field sensors. The most popular free-hand scanning approach makes use of a magnetic field sensor with six degrees of freedom. In this approach, a transmitter is used to produce a spatially varying magnetic field and a small receiver containing three orthogonal coils senses the magnetic field strength, as illustrated in figure 5(b). By measuring the strength of three components of the local magnetic field, the position and orientation of the transducer can be calculated each time a 2D image is acquired. Typically, the transmitter is placed beside the patient and the receiver is mounted on the hand-held transducer. Magnetic field sensors are small and unobtrusive, allowing the transducer to be tracked with less constraint than the previously described approaches. However, electromagnetic

11 Three-dimensional ultrasound imaging R77 (a) (b) Figure 6. 3D B-mode image of the carotid arteries obtained with the free-hand scanning approach using a magnetic position and orientation measurement (POM) device. (a) The 3D image has been sliced to reveal an atherosclerotic plaque at the entrance of the internal carotid artery. The plaque is heterogeneous, with calcified regions casting a shadow on the common carotid artery. (b) The same image has been volume rendered to obtain a view of the vessel wall and the surface of the plaque. interference from sources such as CRT monitors, ac power cables and electrical signals from the transducer itself can compromise the tracking accuracy. Geometric distortions in the final 3D image can occur if ferrous or highly conductive metals are located nearby. In addition, errors in the determination of the location of the moving transducer will occur if the magnetic field sampling rate is low and the transducer is moved too quickly. This results in an image lag artefact, which can be avoided by using a sampling rate of about 100 Hz or higher. By ensuring that the environment of the scanned volume is free of metals and electrical interference, and using an adequate sampling rate, high-quality 3D images can be obtained, as illustrated in figure 6.

12 R78 A Fenster et al Magnetic position and orientation measurement (POM) devices of sufficient quality for 3D ultrasound imaging are currently being produced by two companies: the Fastrack by Polhemus, and Flock-of-Birds by Ascension Technologies. These devices have been used successfully in many diagnostic applications, including echocardiography, obstetrics, and vascular imaging (Fenster et al 1995, Nelson and Elvins 1993, Bonilla-Musoles et al 1995, Detmer et al 1994, Ganapathy and Kaufman 1992, Hodges et al 1994, Hughes et al 1996, Leotta et al 1997, Gilja et al 1997, Nelson and Pretorius 1995, Ohbucki et al 1992, Pretorius and Nelson 1994, Raab et al 1979, Riccabona et al 1995) D tracking by speckle decorrelation. The free-hand scanning techniques described above require a POM device to record the position and orientation of each acquired image. An alternative technique, not requiring any device, uses the acquired images themselves to extract their relative positions. This can be accomplished using the well-known phenomenon of speckle decorrelation. When a source of coherent energy interacts with scatterers, the reflected spatial energy pattern will vary, due to interference, and appear as a speckle pattern. Ultrasound images are characterized by image speckle, which can be used to make velocity images of moving blood (Friemel et al 1998). In this situation if the red blood cells remain stationary then two sequential signals scattered by these cells will be correlated and the speckle pattern will be identical. However, if the blood cells move, the two sequential signals will become decorrelated, with the degree of decorrelation being proportional to the distance moved. The same principle can be used to determine the spacing between two adjacent 2D images. If two images are acquired from the same location, then the speckle pattern will be the same, so that there will be no decorrelation. However, if one of the images is moved with respect to the first, then the degree of decorrelation will be proportional to the distance moved, the exact relationship depending on the beam width in the direction of the motion (Tuthill et al 1998). Speckle decorrelation techniques for 3D ultrasound imaging are complicated by the fact that a pair of adjacent 2D images may not necessarily be parallel to each other. Thus, in order to determine whether the transducer has been tilted or rotated, the acquired images are subdivided into smaller regions, and similar regions in adjacent images are cross-correlated. In this manner, a pattern of decorrelation values is generated, which in turn generates a pattern of distance vectors. These are then analysed to determine the relative position and orientation of the two 2D images Free-hand scanning without position sensing An alternative 3D scanning approach, without any position sensing, involves manipulating the transducer over the patient while acquiring 2D images, and then reconstructing the 3D image by assuming a predefined scanning geometry. Because no geometrical information is recorded during the transducer s motion, the operator must be careful to move the transducer at a constant linear or angular velocity so that the acquired 2D images are obtained with a regular spacing. With uniform motion and knowledge of the approximate total distance or total angle scanned, very good 3D images may be reconstructed, as illustrated by figure 7 (Downey and Fenster 1995a). However, because this approach offers no guarantee that the 3D images are geometrically accurate, they should not be used for any measurements D Arrays for dynamic 3D ultrasound The mechanical and free-hand scanning approaches for 3D ultrasound images described above all require the 2D ultrasound image produced by a conventional transducer, with a 1D array,

13 Three-dimensional ultrasound imaging R79 Figure 7. Maximum intensity projection (MIP) of a 3D power Doppler image of a kidney. The image was obtained with the free-hand scanning approach without any positioning sensing. The image has been sliced to demonstrate that excellent 3D images of vascular structures can be obtained. to be mechanically or manually swept across the anatomy. However, by using a transducer with a 2D array, the transducer could remain stationary, and electronic scanning could be used to sweep the broad ultrasound beam over the entire volume under examination. Although a number of 2D array designs have been developed, the most advanced is the one at Duke University, which was developed for real-time 3D echocardiography and has been used for clinical imaging (Shattuck et al 1984, Smith et al 1991, 1992, Snyder et al 1986, Turnbull and Foster 1991, von Ramm and Smith 1990, von Ramm et al 1991). In this approach, a 2D phased array of transducer elements is used to transmit a broad beam of ultrasound diverging away from the array, sweeping out a volume shaped like a truncated pyramid. The returned echoes detected by the 2D array are processed to display multiple planes from this volume in real time. These planes can be manipulated interactively to allow the user to explore the whole volume under investigation. Although this approach has been successfully used in echocardiology, some problems must be overcome before its use can become widespread in radiology. These problems are related to high cost of the 2D transducer arrays, which results from the low yield of the manufacturing process, which requires many electronic leads to be properly connected to the numerous small elements in the array. 4. Reconstruction of 3D ultrasound images Image reconstruction refers to the process of generating a 3D representation of the anatomy by first placing the acquired 2D images in their correct relative positions and orientations in the 3D image volume, and then using their pixel values to determine the voxel values in the 3D image. Two distinct reconstruction methods have been used: feature-based and voxel-based reconstruction.

14 R80 A Fenster et al 4.1. Feature-based reconstruction In this method, desired features or surfaces of anatomical structures are first determined and then reconstructed into a 3D image. For example, in echocardiographic or obstetric imaging, the ventricles or foetal structures may be outlined (either manually or automatically by computer) in the 2D images, and only the resulting boundary surfaces presented to the viewer in 3D. The surfaces of different structures may be assigned different colours and shaded, and some structures may be eliminated to enhance the visibility of others. This approach has been used extensively in 3D echocardiography to identify the surfaces of ventricles (Ofili and Nanda 1994, Nadkarni et al 2000a, Coppini et al 1995, Wang et al 1994). Then, using ventricle surfaces from 3D images obtained at different phases of the cardiac cycle, the complex motion of the heart can be viewed with a computer workstation, as a cine loop. A similar approach has been also used to reconstruct 3D images of the vascular lumen from intravascular ultrasound (IVUS) images (Reid et al 1995). Because this approach reduces the 3D image content to the surfaces of a few anatomical structures, the contrast of these structures can be optimized. Also, by using multiple (e.g. cardiac gated) images, their dynamic behaviour can be easily appreciated. In addition, it also enables the 3D image to be manipulated efficiently using inexpensive computer display hardware (e.g. to produce fly-through views). However, this approach also has major disadvantages. Because it represents anatomical structures by simple boundary surfaces, important image information, such as subtle anatomical features and tissue texture, is lost at the initial stage of the reconstruction process. Moreover, the artificial contrast between structures may misrepresent subtle features of the anatomy or pathology. Furthermore, the boundary identification step is tedious and time-consuming if done manually, and subject to errors if done automatically by a computer segmentation algorithm Voxel-based reconstruction The more popular approach to 3D ultrasound image reconstruction is based on using a set of acquired 2D images to build a voxel-based image, i.e. a regular Cartesian grid of volume elements in three dimensions. Reconstruction is accomplished in two steps: first, the acquired images are embedded in the image volume, by placing each image pixel at its correct 3D coordinates (x, y, z), based on the 2D coordinates (x, y ) of that pixel in its 2D image, and the position and orientation of that image with respect to the 3D coordinate axes. Then, for each 3D image point, the voxel value (colour or grey-scale image intensity) is calculated by interpolation, as the weighted average of the pixel values of its nearest neighbours among the embedded 2D image pixels. For mechanically scanned images, the interpolation weights may be precomputed and placed in a look-up table, allowing the 3D image to be rapidly reconstructed (Tong et al 1996). This voxel-based reconstruction approach preserves all the information originally present in the acquired 2D images. Thus, by taking suitable cross sections of the 3D image volume, the original 2D images can be recovered. Moreover, new views not found in the original set of images can be generated. However, if the scanning process does not sample the object volume adequately, so that there are gaps between acquired 2D images which are greater than about half the elevational resolution, then the interpolated voxel values will not depict the true anatomy in the gap. In this situation, spurious information is introduced, and the resolution of the reconstructed 3D image is degraded. To avoid this situation, the volume must be well sampled. This generates large data files which require efficient 3D viewing software. Typical image files can range

15 Three-dimensional ultrasound imaging R81 from 16 MB to 96 MB in size, depending on the application. For example, 96 MB images have been reported in 3D TRUS imaging of the prostate for cryosurgical guidance (Downey et al 1995c, Chin et al 1996). Moreover, because all the original image information is preserved, the 3D image can be processed repeatedly using a variety of rendering techniques. For example, the operator can scan through the data (review the complete 3D image) and then choose the technique which displays the features of interest to best advantage. The operator may then apply segmentation and classification algorithms to segment boundaries, measure volumes or perform various volume-based rendering operations. Then, if the selected process does not achieve the desired results, the operator may return to the original 3D image and try a different procedure Freehand 3D scanning. Here it is assumed that the transducer housing is rigidly attached to the sensor of a position and orientation measurement (POM) device (such as a magnetic field sensor). This device determines the position and orientation of the sensor relative to a source in a fixed remote location (such as a magnetic field generator). Suppose now that a 2D image pixel has coordinates (x, y ) relative to the transducer, 3D coordinates (x,y,z ) relative to the sensor, 3D coordinates (x,y,z ) relative to the source and 3D voxel coordinates (x,y,z) in the reconstructed image volume. Suppose further that the x y plane would have to be sequentially rotated through the three angles (α, β, γ ) about three (given) axes and translated by the vector (X, Y, Z) in order to coincide with the x y plane; that the x y plane would similarly have to be rotated through (α,β,γ ) and translated by (X,Y,Z ) in order to coincide with the x y plane; and, finally, that the x y plane would have to be rotated through (α,β,γ ) and translated by (X,Y,Z ) in order to coincide with the x y plane. Then the embedding transformation from 2D pixel coordinates (x, y ) to 3D voxel coordinates (x,y,z) can be written as the product of three homogeneous linear transformations, so that: x y = T T T z 1 x y 0 1 (1) where, in terms of the 3D rotation matrix R ij = r ij (α, β, γ ) R 11 R 12 R 13 X R T = 21 R 22 R 23 Y (2) R 31 R 32 R 33 Z with similar definitions for T in terms of R ij = r ij (α,β,γ ) and T in terms of R ij = r ij (α,β,γ ). Because T is defined by the user and T is given by the output of the POM device, only T needs to be determined in order to fully specify the embedding transformation. This is done by calibrating the rotation angles (α, β, γ ) and the translation vector components (X, Y, Z). Methods for doing this are described by Detmer et al (1994) and Prager et al (1998). Using commercially available magnetometer-based POM devices, Prager et al (1998) found a root-mean-square (rms) precision of mm in each 3D coordinate, depending on the calibration method, and a mean accuracy of better than 0.25 mm for each of the methods tested. These are similar to results obtained earlier by Detmer et al (1994) and Leotta et al (1997) for specific calibration methods. Also, the precision and accuracy of various in vitro and in vivo volume measurements has been reported by Hodges et al (1994) and Hughes et al (1996).

16 R82 A Fenster et al Mechanical 3D scanning. Here, the transducer housing is rigidly mounted on a mechanical device, often hand-held, which then uses a stepper motor to scan the object by translating, rotating or tilting the transducer according to a predetermined protocol, often under microcomputer control. The embedding transformations for mechanical linear and tilt scanners are described in connection with the next section. 5. Effects of errors in the reconstruction Except for systems using 2D arrays, 3D ultrasound images are reconstructed from multiple 2D images, using knowledge of their relative positions and orientations. Thus, any errors in the position or orientation of the 2D images will cause geometric distortion in the reconstructed 3D image, resulting in errors in the measured lengths, areas and volumes of anatomical features in these images. Although other sources of error such as tissue motion during the scan (possibly caused by the scanning procedure itself) or image segmentation errors can also affect these measurements, they overlie this fundamental geometric image distortion. In the literature, analyses of these geometric distortions have been performed for linear and tilt mechanical scanners Linear 3D scanning The analysis was performed by Cardinal et al (2000) for the general case of an ultrasound transducer oriented as shown in figure 8 and scanned parallel to the z-axis. Here, the transducer array is initially oriented along the x-axis, and aimed in the y direction. After being tilted by an angle θ about the x-axis, and then swiveled by an angle φ about the y-axis, it is translated in the z direction, in steps of size d, to acquire a series of parallel 2D images. Thus, for the 2D image acquired after n steps, the pixel at coordinates (x, y ) has 3D Cartesian coordinates (x,y,z) given by the linear transformation [ ] [ ][ x cos φ sin φ sin θ 0 x ] y = 0 cos θ 0 y. (3) z sin φ cos φ sin θ d n Moreover, the embedded 2D image planes are all normal to the unit vector U = (sin φ cos θ, sin θ,cos φ cos θ) (4) and have a normal spacing D = d cos φ cos θ>0 (5) assuming (without loss of generality) that θ and φ are each less than π/2 (90 ) in magnitude. When φ = 0, the embedded 2D image planes are all parallel to the x-axis, and hence the 3D image interpolation reduces to a 2D interpolation in the y z plane, which is identical for each value of x. Thus, when φ = 0, the 3D image may be efficiently reconstructed by computing a single set of 2D interpolation coefficients and applying it to each x plane in succession. The error analysis assumes that the 3D image is reconstructed with the nominal parameter values (d,θ,φ), but that the 3D object was scanned with the actual parameter values (d 0,θ 0,φ 0 ). Because the embedding transformation is linear, straight lines are always imaged as straight lines. However, the errors d = d d 0, θ = θ θ 0 and φ = φ φ 0 still cause relative errors L/L, A/A and V /V in image length, area and volume. If the error vector E is now defined as: E = ( φ cos φ cos θ, θ ) cos θ, d d cos φ cos θ (6)

17 Three-dimensional ultrasound imaging R83 Figure 8. Schematic diagram showing the scanning geometry for linear scanning with a conventional transducer. The transducer array is initially located on the object s surface, along the x-axis, and aimed perpendicularly into the object, in the y direction. The transducer is then tilted by an angle θ about the x-axis and subsequently swiveled by an angle φ about the y-axis, before being translated across the object s surface in the z direction (in steps of size d, not shown) to acquire a series of parallel 2D images. then, assuming that d/d 1, θ 1 and φ 1, it is found that all three relative errors are well described by the linear approximations K K = E C K K = L, A, V (7) where L (U L) (U L)L C L = U = (8a) L 2 which is the component of U parallel to the image line vector L (U A)A A (U A) C A = U = (8b) A 2 A 2 which is the component of U perpendicular to the image area vector A, and C V = U. (8c) Hence, because C K U =1, within these linear approximations K K C K E E K = L, A, V. (9) For volume, the linear approximation is explicitly V V = d tan θ θ tan φ φ (10) d which is insensitive to φ when φ = 0. Also, the exact ratio of image volume V to object volume V 0 is given by the formula V d cos φ cos θ = = D (11) V 0 d 0 cos φ 0 cos θ 0 D 0 because the determinant of the embedding linear transformation is D. L 2

18 R84 A Fenster et al Figure 9. Schematic diagram illustrating the scanning geometry for tilt scanning with a side-firing transrectal ultrasound (TRUS) probe. The acquired 2D images form a radial fan about the probe axis, with an inner radius R 0, and spanning a total scan angle. For the case φ 0 φ = 0, θ 0 = 32.7, the exact volume error formula was verified experimentally, using phantom images, within an experimental uncertainty of less than 0.5%. Also, it was verified numerically that all three linear approximations are perfectly adequate for parameter errors of any likely magnitude (Cardinal et al 2000) Tilt 3D scanning The analysis was performed by Tong et al (1998) for the case of a side-firing TRUS probe rotated about its axis, as shown in figure 9. Here, the scanned images comprise a radial fan of 2D images, each of size (X, Y ), whose inner edges all lie at a distance R 0 from the axis, which forms the z-axis. By convention, the X and Y directions are respectively chosen to be parallel (lateral) and perpendicular (axial) to the probe axis. Thus, for the 2D image at angle θ, the pixel at (x, y ) has 3D cylindrical coordinates (r,θ,z) = (R 0 + y,θ,x ) and hence 3D Cartesian coordinates (x,y,z)= (r cos θ,r sin θ,z). Hence, in this case, the 3D interpolation reduces to a 2D interpolation in the x y plane, which is identical for each value of z. The 3D image may therefore be efficiently reconstructed by computing a single set of 2D interpolation coefficients and applying it to each z plane in succession. The error analysis assumes that the angular interval between acquired 2D images is uniform, and proportional to the total scan angle (Tong et al 1998). However, it applies equally to the case where the angular interval is a variable number of steps, provided the step size remains proportional to. Errors R 0 in R 0 and in will cause relative errors L/L, A/A and V /V in image length, area and volume, as well as a transverse deviation h from linearity in the image of a line of length L (defined so that h vanishes at the image endpoints), due to the nonlinearity of the embedding transformation. It is found that all three relative errors, as well as h/l, vary closely as K K ( = P R0 R ) + Q ( ) K = L, A, V (12) where P 1, Q 1 and R is the average radius of the object from the axis, which is usually well approximated by the radius of the object s centroid. Hence, within these linear approximations K K R 0 R + K = L, A, V. (13)

19 Three-dimensional ultrasound imaging R85 For line images, the coefficients P and Q for L/L and h/l are simple functions of the line s orientation and size, and the relative location of the point at which h is to be calculated. For area and volume, P = Q = 1, and the exact ratio of image size K to object size K 0 is given by the formula K K 0 = ( 1+ R )( 0 1+ ) R K = A, V. (14) Moreover, this formula applies to cross-sectional areas in any orientation. Tong et al (1998) also describe methods for using simple phantoms to calibrate R 0 and accurately. Moreover, the linear length, area and volume formulae were verified numerically, using simulated images, and experimentally, using phantom images (and the calibrated values of R 0 and ), within an experimental uncertainty of less than 0.5%. 6. 3D ultrasound image display Over the past decade, many algorithms and software utilities have been developed to manipulate 3D images interactively using a variety of visualization tools (Udupa 1999, Udupa et al 1991, Robb 1995). Although the quality of the 3D image depends critically on the method of image acquisition and the fidelity of the 3D image reconstruction, the viewing technique used to display the 3D image often plays a dominant role in determining the information transmitted to the operator. There are many approaches for displaying 3D images, which differ in the details of the display implementation. Many of these techniques have been investigated for use with 3D ultrasound images, and these can be divided into three broad classes: surface rendering (SR), multiplanar reformatting (MPR) and volume rendering (VR) Surface rendering (SR) A common 3D display technique used in medical imaging is based on visualizing the surfaces of organs. Because the 3D image must first be reduced to a description of surfaces, image classification and segmentation steps precede the rendering. In the first step, each voxel (or voxel group) in the 3D image must be either manually or automatically classified as to which structure it (or they) belongs (Bezdek et al 1993). Organ boundaries are then identified using manual contouring (Coppini et al 1995, Neveu et al 1994, Lobregt and Viergever 1995), or computer-based segmentation techniques (Gill et al 2000). Once the organs have been classified and segmented, the boundary surfaces are each represented by a mesh and then texture mapped with a colour and texture which appropriately represents the corresponding anatomical structure. Because early 3D ultrasound imaging techniques did not have the benefit of the high computational speed now available in inexpensive computers, only simple wire-frame rendering techniques were used at first. These early approaches were used for displaying the foetus (Brinkley et al 1984, Sohn et al 1989, Sohn and Rudofsky 1989, Brinkley et al 1982), various abdominal structures (Sohn and Grotepass 1989, Sohn et al 1988, Sohn 1989), and the endocardial and epicardial surfaces of the heart (Moritz et al 1983, Coppini et al 1995, Fine et al 1988, Nixon et al 1983, Martin et al 1990, Linker et al 1986, Sawada et al 1983). More recent approaches use more complex surface representations, which are texturemapped, shaded and illuminated, so that both the surface topography and its 3D geometry are more easily comprehended. These approaches typically allow user-controlled motion for viewing the anatomy from various perspectives. This approach has been used successfully by many investigators in the rendering of 3D echocardiographic (Greenleaf et al 1993, Ofili

20 R86 A Fenster et al and Nanda 1994, Ross et al 1993, Delabays et al 1995, Wang et al 1994, McCann et al 1998, Belohlavek et al 1994, Delabays et al 1995) and obstetric images (Nelson and Pretorius 1992, Lees 1992) Multiplanar reformatting (MPR) Rather than displaying only the surfaces of structures, as in the SR technique, the MPR technique provides the viewer with planar cross-sectional images extracted from the 3D image. Thus, this approach requires that either a voxel-based 3D image be first reconstructed, or that an algorithm be used that can extract an arbitrarily oriented plane from the original set of acquired 2D images (Robb 1995). Because the extracted plane will generally not coincide with one of these original image planes, interpolation is required to provide an image similar in appearance to an acquired 2D image. Thus, the extracted image resembles a conventional 2D ultrasound image, which is already familiar to the operator; the operator can then easily position and orient it to the optimal image plane for the examination. This technique is easy to learn and is therefore preferred by radiologists. Two MPR techniques have been used to view 3D ultrasound images: orthogonal planes and cube-view Orthogonal planes. In this approach, computer user interface tools provide three perpendicular planes which are displayed on the screen simultaneously, with graphical cues indicating their relative orientations. The operator can then select any single or multiple planes and move them within the 3D image volume to provide a cross-sectional view at any desired location or orientation, including oblique views (Nelson and Pretorius 1992, Zosmer et al 1996, Gerscovich et al 1994, Kuo et al 1992, Kirbach and Whttingham 1994) Cube-view. In this approach, illustrated by figures 2, 3, 4, 6(a) and 10(a), the 3D image is presented as a polyhedron which represents the boundaries of the reconstructed volume. Each face of the multifaceted polyhedron is rendered with the appropriate ultrasound image for that plane, using a texture-mapping technique (Fenster et al 1995, Tong et al 1996, Robb 1995). With the computer user interface tools provided, the operator can select any face, and move it in or out, parallel to the original face or reorient it obliquely, at any angle to the original face, while the appropriate ultrasound data are continuously texture-mapped in real time onto the new face. In addition, the entire polyhedron may be arbitrarily rotated in any direction to obtain the desired orientation of the 3D image. In this way, the operator always has 3D image-based cues relating the plane being manipulated to the anatomy being viewed (Rankin et al 1993, Fenster et al 1995, Picot et al 1993, Blake et al 2000) Volume rendering (VR) The volume rendering technique presents a display of the entire 3D image after it has been projected onto a 2D plane, with the image projection typically being accomplished via raycasting techniques (Tuy and Tuy 1984, Levoy 1990a, b). The 3D image voxels that intersect each ray are weighted and summed to achieve the desired result in the rendered image. Although many VR algorithms have been developed, 3D ultrasound imaging currently makes use of two main approaches: maximum (minimum) intensity projection and translucency rendering Maximum (minimum) intensity projection (MIP). A common approach is to display only the maximum (minimum) voxel intensity along each ray. This approach is easy to

21 Three-dimensional ultrasound imaging R87 Figure 10. 3D image of a foetal face, displayed using (a) the multiplanar reformatting cube-view approach, and (b) the translucency rendering approach, done with ray-casting. In (a) the 3D image is presented as a polyhedron with the appropriate 2D image texture-mapped onto each face. This polyhedron can then be arbitrarily sliced and rotated to reveal the profile of the foetal face. In (b), the amniotic fluid has been rendered as transparent, and tissues have been rendered as translucent to opaque, as a function of their voxel intensity. implement and fast to compute, allowing real-time manipulation of the MIP image on inexpensive computers. Excellent results are typically achieved when the image information is sparse, such as it is in 3D power Doppler images (see figure 7) (Guo and Fenster 1996, Downey and Fenster 1995b) Translucency rendering. This is the VR approach most commonly used in 3D ultrasound imaging. In this approach, the accumulated luminance C for a given ray is calculated by tracing the ray through the 3D image and adding the contribution from each voxel along the ray. Tracing the ray forwards, towards the projected image plane, the ray-tracing equation is C = C[1 α(i)]+c(i)α(i) (15) where C and C are respectively the accumulated luminance not including and including the contribution from the ith voxel, α(i) is the opacity (1 minus the transmittance) of the ith voxel, and c(i) is its luminance. The parameters α(i) and c(i) are generally functions of the ith voxel intensity, which are chosen to control the specific type of rendered image desired. Thus, if α(i) = 0, the ith voxel is transparent, while if α(i) = 1, then it is opaque or luminescent, depending on the value of c(i). However, the contribution of voxels far from the projected image plane is typically negligible, due to the high total opacity (low total transmittance) of the intervening voxels. Thus, it is often preferred to trace the ray backwards, away from the projected image plane. In this case, the ray-tracing equation becomes C = C + c(i)α(i)[1 A(i)] (16)

22 R88 A Fenster et al where C and C are respectively the accumulated luminance not including and including the contribution from the ith voxel (now numbered in reverse order), and A(i) is the accumulated opacity at the ith voxel, given by [1 A(i)] = j<i[1 α(j)] (17) so that [1 A(i)] represents the total transmittance between the ith voxel and the projected image plane. In practice, the backwards ray-tracing equation is iterated until [1 A(i)] is less than some predefined level, at which point C is output as the display luminance. Because VR preserves much of the information in the 3D image, and projects it onto a 2D image plane after nonlinear processing, the projected image is, at times, difficult to interpret. Thus, this approach is best suited for imaging simple anatomical structures, in which clutter is either absent or has been removed, and to which depth cues and shading have been added. Excellent results have been achieved in displaying foetal (Nelson and Pretorius 1992, 1995, Nelson and Elvins 1993, Pretorius and Nelson 1994, 1995, Nelson et al 1996) and vascular anatomy (Downey and Fenster 1995a, Fine et al 1991), as demonstrated by figure 10(b). 7. Measurement accuracy with 3D ultrasound An important advantage of 3D ultrasound imaging is the ability to measure the length, area or volume of organs or lesions in arbitrary orientations. While a qualitative visual assessment is often valuable, quantitative data provide a more accurate basis for decision making and comparisons against previous studies or reference data-bases (Elliott et al 1996, Rivera et al 1994, Hodges et al 1994, Hughes et al 1996, Riccabona et al 1995, 1996a, Nadkarni et al 2000a, Favre et al 1995, Gilja et al 1995, Sivan et al 1997). In many applications, the goal is the measurement of the volume of the organ or lesion, rather than its diameter or perimeter. Since conventional ultrasound imaging provides only 2D images, volume must be estimated from measurements of diameter in two or three userselected planes and by assuming some idealized shape. This approach results in inaccuracy and variability. Since 3D ultrasound images provide a complete view of the structure, they should provide a basis for more accurate and precise estimates of volume, especially for complex structures Distance measurement With a 3D image, distance measurements need not be made in the plane of an acquired 2D image, but instead can be made in any orientation. Thus, for this measurement to be accurate, the reconstruction process must not distort the anatomical geometry. With proper calibration and reconstruction, accurate distance measurements can be made. Tong et al (1996) have evaluated the accuracy of distance measurements using a 3D wire phantom composed of four layers of surgical wires, with a separation of mm between adjacent wires within each layer, and a separation of mm between adjacent layers. The wire phantom, immersed in a 7% glycerol solution, was imaged with a 3D transrectal ultrasound (3D TRUS) system, which uses the mechanical tilt-scanning approach described by figure 1(c) and in more detail by figure 9. The results of this study showed that with proper calibration of the parameters R 0 and, which specify the scanning geometry, as shown in figure 9, this 3D ultrasound imaging system can provide distance measurements with an error of less than about 1.0%.

23 Three-dimensional ultrasound imaging R Measurement of cross-sectional area using 3D ultrasound Using the multiplanar reformatting (MPR) technique, the 3D image can be sliced in any orientation, to obtain the optimal cross section for the organ measurement. The desired area can then be outlined either manually, using planimetric techniques (Tong et al 1996, Elliot et al 1996), or automatically, using segmentation algorithms (Gill et al 2000, Mao et al 2000, Ladak et al 2000). The area of the region is then found by counting the pixels in the outlined region and multiplying by the pixel area Volume measurement using 3D ultrasound The availability of 3D images allows the measurement of organ volume using either manual or algorithmic techniques. Automated and semiautomated segmentation algorithms are described below in section 8, and are still under development in many laboratories. The alternative manual technique (manual planimetry) makes use of MPR viewing to slice the 3D image into a series of uniformly spaced parallel 2D images. In each 2D image, the cross-sectional area of the organ is manually outlined on the computer screen, and the areas are then summed and multiplied by the interslice distance to obtain an estimate of the organ volume (Elliot et al 1996) In vitro measurements of volume accuracy. Tong et al (1996) used a 3D TRUS system (described in figures 1(c) and 9) to evaluate the accuracy of volume measurements in vitro. Five balloons filled with various known volumes of 7% glycerol solution were imaged, and their volumes were measured using the manual planimetry method. A comparison of measured to true volumes yielded an rms error of 0.9% Volume measurements of prostates in vitro. Rather than using idealized volume shapes such as balloons, Elliot et al (1996) compared prostate volumes measured using manual planimetry of 3D TRUS images with their true volumes. For this study, six prostates were harvested from fresh cadavers, fixed and immersed in a 7% glycerol solution for imaging. The MPR technique was then used to slice the reconstructed prostate images into 20 to 30 transaxial slices 2 5 mm apart, depending on the prostate size. A comparison of measured to true volumes yielded an rms error of 2.6% Volume measurements in vivo. A number of studies have been published demonstrating that 3D ultrasound can be used to accurately measure volumes for a variety of organs (Nelson et al 1999). Brunner et al (1995) showed an increased accuracy over 2D techniques in follicular volume measurements. Chang et al (1997) showed accurate volume measurement results for foetal heart volumes across a range of gestational ages. Riccabona et al (1996b) showed accurate volume measurements by comparing voided urine volumes and measured bladder volumes before voiding. Hughes et al (1996) measured a variety of organ volumes, and found that the errors ranged from 2% to 6%. Favre et al (1995) reported that in using 3D ultrasound for estimation of foetal weight by measuring foetal arm and thigh circumference, the mean error was 1.4% for macrosomic foetuses Effect of interslice distance on planimetric volume measurements. Volume measurement by manual planimetry requires that the 3D image be sliced into individual parallel slices, and the boundary of the anatomical structure be outlined in every slice. Thus, while a small interslice distance increases accuracy, it also increases the number of slices that

24 R90 A Fenster et al must undergo the time-consuming and tedious process of manual planimetry. It is therefore important to determine the maximum interslice distance that still provides an accurate volume estimate. Although irregular or complex structures require the use of small interslice distances, using an interslice distance of less than the elevational resolution will not result in increased accuracy. Thus, the optimal interslice distance will depend on the elevational resolution of the transducer and the complexity of the organ. Elliot et al (1996) have reported on a study examining the effect of the interslice distance on measurement of prostate volume. In this study, prostate volume was measured in a 3D TRUS image by usual manual planimetry with interslice distances from 1 mm to 15 mm. The results showed that the measured prostate volume was constant up to an interslice distance of 8 mm, but underestimated for greater interslice distances. Other reports addressing this issue have recommended interslice distances ranging from 2 mm to 5 mm for measuring prostate volume using 2D ultrasound and manual planimetry (Terris and Stamey 1991) Theoretical variability of planimetric volume measurements. Because volume measurements in 3D images are usually performed via manual planimetry, Nadkarni et al (2000b) have derived a simple formula for theoretically estimating the variance of a planimetric volume measurement, and compared it with experimental variances of volume measurements. In their approach, the volume variance was estimated by summing the variances of individual measurements of the cross-sectional area of an object, multiplied by the interslice distance. For simplicity, it is assumed that n cross-sectional slices are taken perpendicular to the axis of a cylindrically symmetric object of length a. Thus, each slice is a circle, with a radius r k, k = 1,...,n, and an interslice distance x = a/n. Hence, each slice is a circular disk of volume V k = πrk 2 x, k = 1,...,n. Then the planimetric estimate of the object volume V is given by a n n V = πr 2 (x) dx πrk 2 x = V k. (18) 0 k=1 Now, assuming that each radius r k has the same variance σr 2, the variance σ k 2 in V k is given by ( ) 2 σk 2 = Vk σr 2 r = (2πr k x) 2 σr 2 = (4π x)v kσr 2,k= 1,...,n. (19) k Hence, because each slice is measured independently, the variance σv 2 in V is given by σ 2 V = n n σk 2 = (4π x) V k σr 2 k=1 k=1 k=1 = 4πaV n σ 2 r. (20) Therefore, the fractional variability of a planimetrically measured volume V can be estimated as: σ V 4πaσ V = 2 r nv. (21) Thus, the fractional variability in volume measurement using the planimetric method can be reduced by increasing the number n of independent 2D image slices used to make the measurement. In addition, the fractional variability is proportional to the standard deviation σ r in the radial placement of the boundary contour in each slice, which depends, in part, on the image resolution in the 2D image slices. As an example, for an object of volume V = 100 cm 3 and length a = 5 cm, whose volume has been planimetrically measured with n = 10 slices, a distance x = a/n = 5 mm apart, with a boundary position uncertainty of σ r = 1 mm, we estimate from the formula (21) that the fractional variability in the volume measurement will be 2.5%.

25 Three-dimensional ultrasound imaging R Intra- and interobserver variability of volume measurements. Volume measurement with 2D ultrasound requires that the diameter of the organ be measured in at two least images, and that a particular shape must be assumed. For example, using measurements of the height H, width W and length L of a prostate from two orthogonal views, and assuming an ellipsoid shape, the volume may be estimated from the formula V = (π/6)h WL (Terris and Stamey 1991). However, organs are not typically well-described by simple geometric shapes, and selection of the appropriate orthogonal views and chords to measure H, W and L is largely dependent on observer preference, leading to high intra- and inter-observer variability in volumes estimated by this 2D method. Because manual planimetry of 3D ultrasound images uses the entire organ for volume measurement, the intra- and interobserver variability in volumes estimated by this 3D method should be lower. To compare these two methods, Tong et al (1998) measured their intra- and interobserver variability. Eight observers (four experienced radiologists and four technicians or graduate students) measured 15 prostate volumes twice with each method. An analysis of variance (ANOVA) was then used to determine the intra- and interobserver standard errors of measurement SEM intra and SEM inter of the prostate volume measurements made using each method (Eliasziw et al 1994, Mitchell et al 1996). The ANOVA results shown in table 2 demonstrate that the 3D method reduces SEM intra by about a factor of three and SEM inter by about a factor of two (Tong et al 1998). Often SEM values are interpreted in terms of the minimum volume change V detectable at the 95% level of confidence, which is given by V = 2.77 SEM. These values for V are also given in table 1. For a single observer, they show that the minimum confidently detectable volume change is 43% for the 2D method but only 14% for the 3D method. Table 2. Results of the intra- and interobserver study comparing prostate volume measurements using 3D ultrasound and the conventional 2D method. V is the minimum relative prostate volume change that can be detected between successive prostate volume measurements at the 95% level of confidence ( V = 2.77 SEM). SEM/volume (%) V /volume (%) 3D 2D 3D 2D Intraobserver Interobserver Also, Chang et al (1997) have compared the reproducibility of the two methods for the assessment of foetal liver volume, and report that the intra- and interobserver reproducibility are improved with the 3D method. Their volume measurements had an intraobserver standard deviation of 8.46 cm 3 for the 2D method and 2.15 cm 3 for 3D method. 8. Segmentation of 3D ultrasound images Because the use of manual planimetry for identifying organs and measuring their volumes is time-consuming and tedious, investigators are developing automated or semiautomated segmentation techniques. Segmentation is the process of partitioning a digital image into separate structures. In general, for a technique to be useful, it must be accurate, reproducible and fast. However, because 3D ultrasound images suffer from artefacts such as shadowing, speckle and poor contrast, fully automated segmentation may result in unacceptable errors. Thus, semiautomated approaches may be better, since they allow image-specific initialization and editing of the results (Ladak et al 2000).

26 R92 A Fenster et al Successful segmentation approaches in 3D ultrasound imaging have been based on deformable models, which make use of geometric objects (1D contours, 2D surfaces or 3D solids) that deform to fit the desired image features (Lobregt et al 1995, Gill et al 2000, Ladak et al 2000, McInerney and Terzopoulos 1996). External forces are used to deform the model, while internal forces within the model provide the desired degree of stiffness, or resistance to deformation. Gill et al (2000) have used this approach to segment the carotid arteries in 3D B-mode images. Their approach comprises three major steps: (a) interactive placement of an initial balloon model inside the lumen; (b) automatic inflation of the model towards the arterial wall and (c) automatic localization of the arterial wall. The model shape is described by a triangular mesh. The equation of motion for each vertex on the mesh is given by m i ẍ i (t) + v i ẋ i (t) + g(x i (t)) = f(x i (t)) (22) where x i (t), ẋ i (t) and ẍ i (t) are respectively the vertex 3D position, velocity and acceleration at time t respectively, m i is its mass, v i is its damping coefficient, g(x i (t)) is the resultant surface tension at the vertex and f(x i (t)) is a driving force. Equilibrium is attained when both ẋ i (t) and ẍ i (t) become zero, which can take a very long time. However, by setting m i = 0 and v i = 1 for the system, the equation of motion reduces to ẋ i (t) = f(x i (t)) g(x i (t)). (23) Equilibrium is then attained when ẋ i (t) becomes zero, which can occur quickly, because the system does not possess inertia. To compute the equilibrium position of each vertex, and hence the deformed shape of the model, the model is inflated by iteratively updating x i (t) via the formula x i (t + t) x i (t) + ẋ i (t) t (24) until ẋ i (t) is sufficiently small at all vertices. When the model inflation stops, the mesh provides an approximation of the vessel boundary. At this stage, the driving force f(x i (t)) is replaced by an image-based force f edge (x i (t)), that locally modifies the mesh to conform to fine details in the 3D image. This force is defined in terms of the potential function P(x i (t)) defined by (Gill et al 2000): 1 P(x i (t)) = (25) (G σ I) + ε where (G σ I) represents the convolution of a 3D Gaussian smoothing filter of characteristic width σ with the image intensity I(x i (t)), denotes the gradient operator and ε is a small positive constant to prevent division by zero. Minima of the potential function coincide with the arterial wall. This potential function produces the force field that is used to deform the model: f edge (x i (t)) = k edge P(x i (t)). (26) The strength of the force is controlled by the coefficient k edge, the value of which must be optimized to allow the mesh to conform to the vessel surface. 9. Trends and future developments Advances in the past 5 years now allow real-time or fast 3D acquisition, real-time reconstruction, and 3D visualization with real-time image manipulation. The availability of clinically useful 3D ultrasound systems is allowing users to focus on demonstrating their clinical utility. Also, advanced visualization and measurement tools are allowing the examination of complex anatomical structures and the accurate measurement of

27 Three-dimensional ultrasound imaging R93 (a) (b) Figure 11. 3D image of a prostate after a brachytherapy procedure. The radioactive seeds appear as bright structures. Panel (a) shows a transverse view from a 3D ultrasound image, and panel (b) shows the same view from a CT image. complex volumes. These have been particularly important in obstetrics, cardiology and urology (Nelson et al 1999, Baba and Jurkovic 1997). Although 3D ultrasound has been shown to be useful, progress still needs to be made for this technique to become a routine tool. The following are current trends and needed developments in 3D ultrasonography The use of 3D ultrasound in brachytherapy Current practice involves acquiring 2D ultrasound images of the prostate at 5 mm intervals. With these, the dose plan is made and then the radioactive seeds are implanted. The availability of 3D ultrasonography may allow the development of an intraoperative technique (Blake et al 2000). Using 3D ultrasound, the image of the prostate could be obtained quickly and a dose plan made with a 3D image of the prostate. After implantation, the brachytherapy seeds

28 R94 A Fenster et al could be accurately detected and their locations determined. Views of a patient s prostate after brachytherapy seed implantation are shown in figure 11. Although the seeds are difficult to distinguish, image processing might improve their contrast, which would allow them to be automatically segmented The need for improved free-hand scanning and 2D arrays Although free-hand scanning techniques are already in use, they are still subject to potential artefacts and inaccuracies. Further progress is needed for these techniques to become routine tools producing high-quality images under all circumstances. In addition, improvements in 2D arrays are still required to produce 3D images whose quality is similar to those currently being produced with conventional 1D arrays. Nevertheless, 2D arrays producing 3D images directly are already useful when real-time 3D imaging is required, as in echocardiography The need for improved visualization tools Current 3D ultrasound imaging systems already make use of real-time visualization tools. However, these tools are generally difficult to use, and require complicated user interfaces. For 3D ultrasound imaging to become widely accepted, intuitive tools are required to manipulate the 3D image so that the user may view any section of the anatomy in relation to any other sections. In addition, intuitive real-time tools are needed to produce volume-rendered images. Currently, the production of volume-rendered images requires multiple parameters to be manipulated. Techniques are needed that provide immediate optimal rendering, based on both image data and reference data for the organ being viewed, without significant user intervention Conclusions Clinical experience with 3D ultrasound has already demonstrated the clear advantages it offers both for the diagnosis of disease and in providing image guidance for minimally invasive therapy. Technical advances will improve 3D imaging even further, making it a routine tool. Although advances are required in both 3D visualization software and hardware, the potential for innovation is clearly present. Based on the pace of development over the past 5 years, many major advances can be anticipated over the next 5 years, leading to improvements in existing applications and to the development of new applications in a wide variety of medical disciplines. References Baba K and Jurkovic D 1997 Three-Dimensional Ultrasound in Obstetrics and Gynecology (New York: Parthenon) Bamber J C, Eckersley R J, Hubregtse P, Bush N L, Bell D S and Crawford D C 1992 Data processing for 3-D ultrasound visualization of tumour anatomy and blood flow Proc. SPIE Belohlavek M, Foley D A, Gerber T C, Greenleaf J F and Seward J B 1994 Three-dimensional reconstruction of color Doppler jets in the human heart J. Am. Soc. Echocardiogr Belohlavek M, Foley D A, Gerber T C, Kinter T M, Greenleaf J F and Seward J B 1993 Three- and four-dimensional cardiovascular ultrasound imaging: a new era for echocardiography Mayo Clin. Proc Bezdek J C, Hall L O and Clarke L P 1993 Review of MR image segmentation techniques using pattern recognition Med. Phys Blake C, Elliot T, Slomka P, Downey D and Fenster A 2000 Variability and accuracy of measurements of prostate brachytherapy seed position in vitro using three dimensional ultrasound: an intra- and inter-observer study Med. Phys Bonilla-Musoles F, Raga F, Osborne N G and Blanes J 1995 Use of three-dimensional ultrasonography for the study of normal and pathologic morphology of the human embryo and fetus: preliminary report J. Ultrasound Med

29 Three-dimensional ultrasound imaging R95 Brinkley J F, McCallum W D, Muramatsu S K and Liu D Y 1982 Fetal weight estimation from ultrasonic threedimensional head and trunk reconstructions: evaluation in vitro Am. J. Obstet. Gynecol Fetal weight estimation from lengths and volumes found by three-dimensional ultrasonic measurements J. Ultrasound Med Brinkley J F, Muramatsu S K, McCallum W D and Popp R L 1982 In vitro evaluation of an ultrasonic three-dimensional imaging and volume system Ultrason Imaging Brunner M, Obruca A, Bauer P and Feichtinger W 1995 Clinical application of volume estimation based on threedimensional ultrasonography Ultrasound Obstet. Gynecol Cardinal H N, Gill J and Fenster A 2000 Analysis of geometrical distortion and statistical variance in length, area, and volume in a linearly scanned 3D ultrasound image IEEE Trans. Med. Imaging Carson P L, Li X, Pallister J, Moskalik A, Rubin J M and Fowlkes J B 1993 Approximate quantification of detected fractional blood volume and perfusion from 3-D color flow and Doppler power signal imaging Ultrasonics Symp. Proc. (Piscataway, NJ: IEEE) pp Chang F M, Hsu K F, Ko H C, Yao B L, Chang C H, Yu C H, Liang R I and Chen H Y 1997 Fetal heart volume assessment by three-dimensional ultrasound Ultrasound Obstet. Gynecol Chin J L, Downey D B, Mulligan M and Fenster A 1998 Three-dimensional transrectal ultrasound guided cryoablation for localized prostate cancer in nonsurgical candidates: a feasibility study and report of early results J. Urol Chin J L, Downey D B, Onik G and Fenster A 1996 Three-dimensional prostate ultrasound and its application to cryosurgery Tech. Urol Chin J L, Downey D B, Tong S, Mclean C A, Elliot T, Fourtier M and Fenster A 1999 Three dimensional transrectal ultrasound imaging of the prostate: clinical validation Can. J. Urol Coppini G, Poli R and Valli G 1995 Recovery of the 3-D shape of the left ventricle from echocardiographic images IEEE Trans. Med. Imaging Dabrowski W, Dumore-Buyze J, Cardinal H N and Fenster A 2001 A real vessel phantom for flow imaging: 3D Doppler ultrasound of steady flow Ultrasound Med. Biol. at press De Castro S, Yao J and Pandian N G 1998 Three-dimensional echocardiography: clinical relevance and application Am. J. Cardiol G 102G Delabays A, Pandian N G, Cao Q L, Sugeng L, Marx G, Ludomirski A and Schwartz S L 1995 Transthoracic realtime three-dimensional echocardiography using a fan-like scanning approach for data acquisition: methods, strengths, problems, and initial clinical experience Echocardiography Delabays A, Sugeng L, Pandian N G, Hsu T L, Ho S J, Chen C H, Marx G, Schwartz S L and Cao Q L 1995 Dynamic three-dimensional echocardiographic assessment of intracardiac blood flow jets Am. J. Cardiol Detmer P R, Bashein G, Hodges T, Beach K W, Filer E P, Burns D H and StrandnessDEJr1994 3D ultrasonic image feature localization based on magnetic scanhead tracking: in vitro calibration and validation Ultrasound Med. Biol Downey D B, Chin J L and Fenster A 1995c Three-dimensional US-guided cryosurgery Radiology 197(P) 539 Downey D B and Fenster A 1995a Vascular imaging with a three-dimensional power Doppler system Am. J. Roentgenol b Three-dimensional power Doppler detection of prostatic cancer Am. J. Roentgenol Downey D B, Nicolle D A and Fenster A 1995a Three-dimensional orbital ultrasonography Can. J. Ophthalmol b Three-dimensional ultrasound of the eye Administrative Radiol. J Eliasziw M, Young S L, Woodbury M G and Fryday-Field K 1994 Statistical methodology for the concurrent assessment of interrater and intrarater reliability: using goniometric measurements as an example Phys. Ther Elliot T L, Downey D B, Tong S, McLean C A and Fenster A 1996 Accuracy of prostate volume measurements in vitro using three-dimensional ultrasound Acad. Radiol Favre R, Bader A M and Nisand G 1995 Prospective study on fetal weight estimation using limb circumferences obtained by three-dimensional ultrasound Ultrasound Obstet. Gynecol Fenster A D sonography technical aspects Echoenhancers and Transcranial Color Duplex Sonography ed U Bogdahn, G Becker and F Schlachetzki (Berlin: Blackwell Science) pp Fenster A and Downey D B D ultrasound imaging: A review IEEE Eng. Med. Biol Fenster A, Tong S, Sherebrin S, Downey D B and Rankin R N 1995 Three-dimensional ultrasound imaging Proc. SPIE Fine D, Perring S, Herbetko J, Hacking C N, Fleming J S and Dewbury K C 1991 Three-dimensional (3D) ultrasound imaging of the gallbladder and dilated biliary tree: reconstruction from real-time B-scans Br. J. Radiol

30 R96 A Fenster et al Fine D G, Sapoznikov D, Mosseri M and Gotsman M S 1988 Three-dimensional echocardiographic reconstruction: qualitative and quantitative evaluation of ventricular function Comput Methods Programs Biomed Fishman E K, Magid D, Ney D R, Chaney E L, Pizer S M, Rosenman J G, Levin D N, Vannier M W, Kuhlman J E and Robertson D D 1991 Three-dimensional imaging Radiology Friemel B H, Bohs L N, Nightingale K R and Trahey G E 1998 Speckle decorrelation due to two-dimensional flow gradients IEEE Trans. Ultrason. Ferroelectr. Frequency Control Ganapathy U and Kaufman A D acquisition and visualization of ultrasound data Proc. SPIE Geiser E A, ChristieLGJr,Conetta D A, Conti C R and Gossman G S 1982 A mechanical arm for spatial registration of two-dimensional echocardiographic sections Cathet. Cardiovasc. Diagn Gerscovich E O, Greenspan A, Cronan M S, Karol L A and McGahan J P 1994 Three-dimensional sonographic evaluation of developmental dysplasia of the hip: preliminary findings Radiology Ghosh A, Nanda N C and Maurer G 1982 Three-dimensional reconstruction of echo-cardiographic images using the rotation method Ultrasound Med. Biol Gilja O H, Detmer P R, Jong J M, Leotta D F, Li X N, Beach K W, Martin R and Strandness D E Jr 1997 Intragastric distribution and gastric emptying assessed by three-dimensional ultrasonography Gastroenterology Gilja O H, Smievoll A I, Thune N, Matre K, Hausken T, Odegaard S and Berstad A 1995 In vivo comparison of 3D ultrasonography and magnetic resonance imaging in volume estimation of human kidneys Ultrasound Med. Biol Gilja O H, Thune N, Matre K, Hausken T, Odegaard S and Berstad A 1994 In vitro evaluation of three-dimensional ultrasonography in volume estimation of abdominal organs Ultrasound Med. Biol Gill J, Ladak H, Steinman D and Fenster A 2000 Accuracy and variability assessment of semi-automatic technique for segmentation of the carotid arteries from 3D ultrasound images Med. Phys Greenleaf J F, Belohlavek M, Gerber T C, Foley D A and Seward J B 1993 Multidimensional visualization in echocardiography: an introduction Mayo Clin. Proc Guo Z and Fenster A 1996 Three-dimensional power Doppler imaging: a phantom study to quantify vessel stenosis Ultrasound Med. Biol Guo Z, Moreau M, Rickey D W, Picot P A and Fenster A 1995 Quantitative investigation of in vitro flow using three-dimensional colour Doppler ultrasound Ultrasound Med. Biol Hodges T C, Detmer P R, Burns D H, Beach K W and Strandness D E Jr 1994 Ultrasonic three-dimensional reconstruction: in vitro and in vivo volume and area measurement Ultrasound Med. Biol Hughes S W, D Arcy T J, Maxwell D J, Chiu W, Milner A, Saunders J E and Sheppard R J 1996 Volume estimation from multiplanar 2D ultrasound images using a remote electromagnetic position and orientation sensor Ultrasound Med. Biol King D L, Gopal A S, Sapin P M, Schroder K M and Demaria A N 1993 Three-dimensional echocardiography Am. J. Card. Imaging King D L, Harrison MR, King D L Jr, Gopal A S, Kwan O L and DeMaria A N 1992 Ultrasound beam orientation during standard two-dimensional imaging: assessment by three-dimensional echocardiography J. Am. Soc. Echocardiogr KingDL,KingDLJrandShao M Y 1990 Three-dimensional spatial registration and interactive display of position and orientation of real-time ultrasound images J. Ultrasound Med Evaluation of in vitro measurement accuracy of a three-dimensional ultrasound scanner J. Ultrasound Med Kirbach D and Whittingham T A D ultrasound the kretztechnik voluson approach Eur. J. Ultrasound Kuo H C, Chang F M, Wu C H, Yao B L and Liu C H 1992 The primary application of three-dimensional ultrasonography in obstetrics Am. J. Obstet. Gynecol Ladak H, Mao F, Wang Y, Downey D B, Steinman D and Fenster A 2000 Prostate boundary segmentation from 2D ultrasound images Med. Phys Lees W A D ultrasound images optimize fetal review Diagn. Imaging Leotta D F, Detmer P R and Martin R W 1997 Performance of a miniature magnetic position sensor for threedimensional ultrasound imaging Ultrasound Med. Biol Levine R A, Handschumacher M D, Sanfilippo A J, Hagege A A, Harrigan P, Marshall J E and Weyman A E 1989 Three-dimensional echocardiographic reconstruction of the mitral valve, with implications for the diagnosis of mitral valve prolapse Circulation Levoy M 1990a Volume rendering, a hybrid ray tracer for rendering polygon and volume data IEEE Comput. Graphics Appl b Efficient ray tracing of volume data ACM Trans. Graphics Linker D T, Moritz W E and Pearlman A S 1986 A new three-dimensional echocardiographic method of right ventricular volume measurement: in vitro validation J. Am. Coll. Cardiol

31 Three-dimensional ultrasound imaging R97 Lobregt S and Viergever M A 1995 A discrete dynamic contour model IEEE Trans. Med. Imaging Mao F, Gill J, Downey D B and Fenster A 2000 Segmentation of carotid artery in ultrasound images: method development and evaluation technique Med. Phys Martin R W and Bashein G 1989 Measurement of stroke volume with three-dimensional transesophageal ultrasonic scanning: comparison with thermodilution measurement Anesthesiology Martin R W, Bashein G, Detmer P R and Moritz W E 1990 Ventricular volume measurement from a multiplanar transesophageal ultrasonic imaging system: an in vitro study IEEE Trans. Biomed. Eng. 3: McCann H A, Chandrasekaran K, Hoffman E A, Sinak L J, Kinter T M and Greenleaf J F 1987 A method for three-dimensional ultrasonic imaging of the heart in vivo Dynamic Cardiovasc. Imaging McCann H A, Sharp J C, Kinter T M, McEwan C N, Barillot C and Greenleaf J F 1988 Multidimensional ultrasonic imaging for cardiology Proc. IEEE McInerney T and Terzopoulos D 1996 Deformable models in medical image analysis: a survey Med. Image Anal Mitchell J R, Karlik S J, Lee D H, Eliasziw M, Rice G P and Fenster A 1996 The variability of manual and computer assisted quantification of multiple sclerosis lesion volumes Med. Phys Moritz W E, Pearlman A S, McCabe D H, Medema D K, Ainsworth M E and Boles M S 1983 An ultrasonic technique for imaging the ventricle in three dimensions and calculating its volume IEEE Trans. Biomed. Eng Nadkarni S K, Boughner D R, Drangova M and Fenster A 2000a Three-dimensional echocardiography: assessment of inter- and intra-operator variability and accuracy in the measurement of left ventricular cavity volume and myocardial mass Phys. Med. Biol b Three-dimensional echocardiography: Assessment of inter- and intra-operator variability and accuracy in the measurement of left ventricular cavity volumes and myocardial masses Phys. Med. Biol Nelson T R, Downey D B, Pretorius D H and Fenster A 1999 Three-Dimensional Ultrasound (Philadelphia, PA: Lippincott Williams & Wilkins) Nelson T R and Elvins T T 1993 Visualization of 3D ultrasound data IEEE Comput. Graphics Appl Nelson T R and Pretorius D H 1992 Three-dimensional ultrasound of fetal surface features Ultrasound Obstet. Gynecol Visualization of the fetal thoracic skeleton with three-dimensional sonography: a preliminary report Am. J. Roentgenol Three-dimensional ultrasound imaging Ultrasound Med. Biol Nelson T R, Pretorius D H, Sklansky M and Hagen-Ansert S 1996 Three-dimensional echocardiographic evaluation of fetal heart anatomy and function: acquisition, analysis, and display J. Ultrasound Med Neveu M, Faudot D and Derdouri B 1994 Recovery of 3D deformable models from echocardiographic images Proc. SPIE Nixon J V, Saffer S I, Lipscomb K and Blomqvist C G 1983 Three-dimensional echoventriculography Am. Heart. J Ofili E O and Nanda N C 1994 Three-dimensional and four-dimensional echocardiography Ultrasound Med. Biol Ohbuchi R, Chen D and Fuchs H 1992 Incremental volume reconstruction and rendering for 3D ultrasound imaging Proc. SPIE Onik G M, Downey D B and Fenster A 1996 Three-dimensional sonographically monitored cryosurgery in a prostate phantom J. Ultrasound Med Pandian N G, Nanda N C, Schwartz S L, Fan P, Cao Q L, Sanyal R, Hsu T L, Mumm B, Wollschlager H and Weintraub A 1992 Three-dimensional and four-dimensional transesophageal echocardiographic imaging of the heart and aorta in humans using a computed tomographic imaging probe Echocardiography Picot P A, Rickey D W, Mitchell R, Rankin R N and Fenster A 1991 Three-dimensional colour Doppler imaging of the carotid artery Proc. SPIE Three-dimensional colour Doppler imaging Ultrasound Med. Biol Prager R W, Gee A and Berman L 1998 Stradx: real-time acquisition and visualization of freehand three-dimensional ultrasound Med. Image Anal Pretorius D H and Nelson T R 1994 Prenatal visualization of cranial sutures and fontanelles with three-dimensional ultrasonography J. Ultrasound Med Fetal face visualization using three-dimensional ultrasonography J. Ultrasound Med Pretorius D H, Nelson T R and Jaffe J S dimensional sonographic analysis based on color flow Doppler and gray scale image data: a preliminary report J. Ultrasound Med Raab F H, Blood E B, Steiner T O and Jones H R 1979 Magnetic position and orientation tracking system IEEE Trans. Aerospace Electron. Syst

32 R98 A Fenster et al Rankin R N, Fenster A, Downey D B, Munk P L, Levin M F and Vellet A D 1993 Three-dimensional sonographic reconstruction: techniques and diagnostic applications Am. J. Roentgenol Reid D B, Douglas M and Diethrich E B 1995 The clinical value of three-dimensional intravascular ultrasound imaging J. Endovasc. Surg Riccabona M, Johnson D, Pretorius D H and Nelson T R 1996 Three dimensional ultrasound: display modalities in the fetal spine and thorax Eur. J. Radiol Riccabona M, Nelson T R and Pretorius D H 1996 Three-dimensional ultrasound: accuracy of distance and volume measurements Ultrasound Obstet. Gynecol Riccabona M, Nelson T R, Pretorius D H and Davidson T E 1995 Distance and volume measurement using threedimensional ultrasonography J. Ultrasound Med Rivera J M, Siu S C, Handschumacher M D, Lethor J P, Guerrero J L, Vlahakes G J, Mitchell J D, Weyman A E, King M E and Levine R A 1994 Three-dimensional reconstruction of ventricular septal defects: validation studies and in vivo feasibility J. Am. Coll. Cardiol Robb R A 1995 Three-dimensional Biomedical Imaging: Principles and Practice (New York: VCH) Robb R A and Barillot C 1989 Interactive display and analysis of 3-D medical images IEEE Trans. Med. Imaging Roelandt J R, ten Cate F J, Vletter W B and Taams M A 1994 Ultrasonic dynamic three-dimensional visualization of the heart with a multiplane transesophageal imaging transducer J. Am. Soc. Echocardiogr Ross J J Jr, D Adamo A J, Karalis D G and Chandrasekaran K 1993 Three-dimensional transesophageal echo imaging of the descending thoracic aorta Am. J. Cardiol Sawada H, Fujii J, Kato K, Onoe M and Kuno Y 1983 Three dimensional reconstruction of the left ventricle from multiple cross sectional echocardiograms. Value for measuring left ventricular volume Br. Heart J Shattuck D P, Weinshenker M D, Smith S W and von Ramm O T 1984 Explososcan: a parallel processing technique for high speed ultrasound imaging with linear phased arrays J. Acoust. Soc. Am Sivan E, Chan L, Uerpairojkit B, Chu G P and Reece E A 1997 Growth of the fetal forehead and normative dimensions developed by three-dimensional ultrasonographic technology J. Ultrasound Med Smith W L and Fenster A 2000 Optimum scan spacing for three-dimensional ultrasound by speckle statistics Ultrasound Med. Biol Smith S W, Pavy H G Jr and von Ramm O T 1991 High-speed ultrasound volumetric imaging system. Part I. Transducer design and beam steering IEEE Trans. Ultrason. Ferroelectr. Frequency Control Smith S W, Trahey G E and von Ramm O T 1992 Two-dimensional arrays for medical ultrasound Ultrasonic Imaging Snyder J E, Kisslo J and von Ramm O 1986 Real-time orthogonal mode scanning of the heart. I. System design J. Am. Coll. Cardiol Sohn C 1989 A new diagnostic technique: three-dimensional ultrasound imaging Ultrasonics International 89 Conf. Proc Sohn C and Grotepass J 1989 Representation tridimensionnelle par ultrason Radiologie J. CEPUR (in French) Sohn C, Grotepass J, Schneider W, Funk A, Sohn G, Jensch P, Fendel H, Ameling W and Jung H 1988 Initial studies of 3-dimensional imaging using ultrasound Z. Geburtshilfe Perinatol (in German) Sohn C, Grotepass J and Swobodnik W 1989 Possibilities of 3-dimensional ultrasound imaging Ultraschall. Med (in German) Sohn C and Rudofsky G 1989 Die dreidimensionale ultraschalldiagnostik ein neues verfahren fur die klinische routine? Ultraschall. Klin. Prax (in German) Sohn C H, Stolz W, Kaufmann M and Bastert G 1992 Three-dimensional ultrasound imaging of benign and malignant breast tumors initial clinical experiences Geburts. Frauenheilkd Terris M K and Stamey T A 1991 Determination of prostate volume by transrectal ultrasound J. Urol Tong S, Cardinal H N, Downey D B and Fenster A 1998 Analysis of linear, area and volume distortion in 3D ultrasound imaging Ultrasound Med. Biol Tong S, Cardinal H N, McLoughlin R F, Downey D B and Fenster A 1998 Intra- and inter-observer variability and reliability of prostate volume measurement via two-dimensional and three-dimensional ultrasound imaging Ultrasound Med. Biol Tong S, Downey D B, Cardinal H N and Fenster A 1996 A three-dimensional ultrasound prostate imaging system Ultrasound Med. Biol Turnbull D H and Foster F S 1991 Beam steering with pulsed two-dimensional transducer arrays IEEE Trans. Ultrason. Ferroelectr. Frequency Control Tuthill T A, Krucker J F, Fowlkes J B and Carson P L 1998 Automated three-dimensional US frame positioning computed from elevational speckle decorrelation Radiology

33 Three-dimensional ultrasound imaging R99 Tuy H K and Tuy L T 1984 Direct 2-D display of 3-D objects IEEE Comput. Graphics Appl Udupa J K 1999 Three-dimensional visualization and analysis methodologies: a current perspective Radiographics Udupa J K, Hung H M and Chuang K S 1991 Surface and volume rendering in three-dimensional imaging: a comparison J. Digital Imaging Vannier M W, Marsh J L and Warren J O 1984 Three dimensional CT reconstruction images for craniofacial surgical planning and evaluation Radiology von Ramm O T and Smith S W 1990 Real time volumetric ultrasound imaging system Proc. SPIE von Ramm O T, Smith SWandPavyHGJ1991 High-speed ultrasound volumetric imaging system. Part II. Parallel processing and image display IEEE Trans. Ultrason. Ferroelectr. Frequency Control Wang X F, Li Z A, Cheng T O, Deng Y B, Zheng L H, Hu G and Lu P 1994 Clinical application of three-dimensional transesophageal echocardiography Am. Heart. J Weiss J L, Eaton L W, Kallman C H and Maughan W L 1983 Accuracy of volume determination by two-dimensional echocardiography: defining requirements under controlled conditions in the ejecting canine left ventricle Circulation Wild J J and Reid J M 1952 Application of echo-ranging techniques to the determination of the structure of biological tissues Science Wollschlager H, Zeiher A M, Geibel A, Kasper W and Just H 1993 Transesophageal echo computer tomography of the heart Cardiac Ultrasound edjrtcroelandt et al (London: Churchill Livingstone) pp Zosmer N, Jurkovic D, Jauniaux E, Gruboeck K, Lees C and Campbell S 1996 Selection and identification of standard cardiac views from three-dimensional volume scans of the fetal thorax J. Ultrasound Med

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

Twelve. Figure 12.1: 3D Curved MPR Viewer Window Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from

More information

Company profile. Quote from the analyst of. Company Milestone. Feel the Difference. Tel: +86-755-26722860 Fax: +86-755-26722850

Company profile. Quote from the analyst of. Company Milestone. Feel the Difference. Tel: +86-755-26722860 Fax: +86-755-26722850 Company profile Since the establishment, SonoScape has committed to provide high standard medical equipments for healthcare of human beings. SonoScape specializes in the development and production of diagnostic

More information

5-Axis Test-Piece Influence of Machining Position

5-Axis Test-Piece Influence of Machining Position 5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,

More information

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami

More information

Solving Simultaneous Equations and Matrices

Solving Simultaneous Equations and Matrices Solving Simultaneous Equations and Matrices The following represents a systematic investigation for the steps used to solve two simultaneous linear equations in two unknowns. The motivation for considering

More information

ENG4BF3 Medical Image Processing. Image Visualization

ENG4BF3 Medical Image Processing. Image Visualization ENG4BF3 Medical Image Processing Image Visualization Visualization Methods Visualization of medical images is for the determination of the quantitative information about the properties of anatomic tissues

More information

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - [email protected]

More information

Clinical Training for Visage 7 Cardiac. Visage 7

Clinical Training for Visage 7 Cardiac. Visage 7 Clinical Training for Visage 7 Cardiac Visage 7 Overview Example Usage 3 Cardiac Workflow Examples 4 Remove Chest Wall 5 Edit Chest Wall Removal 6 Object Display Popup 7 Selecting Optimal Phase 8 Thick

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

How To Fuse A Point Cloud With A Laser And Image Data From A Pointcloud

How To Fuse A Point Cloud With A Laser And Image Data From A Pointcloud REAL TIME 3D FUSION OF IMAGERY AND MOBILE LIDAR Paul Mrstik, Vice President Technology Kresimir Kusevic, R&D Engineer Terrapoint Inc. 140-1 Antares Dr. Ottawa, Ontario K2E 8C4 Canada [email protected]

More information

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected].

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt. Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected] Outline Motivation why do we need GPUs? Past - how was GPU programming

More information

SAMSUNG ULTRASOUND RS80A

SAMSUNG ULTRASOUND RS80A Samsung Medison is a global leading medical device company. Founded in 1985, the company sells cutting-edge diagnostic ultrasound devices around the world in various medical fields. The company has attracted

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

NEW HYBRID IMAGING TECHNOLOGY MAY HAVE BIG POTENTIAL FOR IMPROVING DIAGNOSIS OF PROSTATE CANCER

NEW HYBRID IMAGING TECHNOLOGY MAY HAVE BIG POTENTIAL FOR IMPROVING DIAGNOSIS OF PROSTATE CANCER Media Release April 7, 2009 For Immediate Release NEW HYBRID IMAGING TECHNOLOGY MAY HAVE BIG POTENTIAL FOR IMPROVING DIAGNOSIS OF PROSTATE CANCER London, Ontario Improved hybrid imaging techniques developed

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

Volume visualization I Elvins

Volume visualization I Elvins Volume visualization I Elvins 1 surface fitting algorithms marching cubes dividing cubes direct volume rendering algorithms ray casting, integration methods voxel projection, projected tetrahedra, splatting

More information

11.1. Objectives. Component Form of a Vector. Component Form of a Vector. Component Form of a Vector. Vectors and the Geometry of Space

11.1. Objectives. Component Form of a Vector. Component Form of a Vector. Component Form of a Vector. Vectors and the Geometry of Space 11 Vectors and the Geometry of Space 11.1 Vectors in the Plane Copyright Cengage Learning. All rights reserved. Copyright Cengage Learning. All rights reserved. 2 Objectives! Write the component form of

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

Direct Volume Rendering Elvins

Direct Volume Rendering Elvins Direct Volume Rendering Elvins 1 Principle: rendering of scalar volume data with cloud-like, semi-transparent effects forward mapping object-order backward mapping image-order data volume screen voxel

More information

Video Camera Image Quality in Physical Electronic Security Systems

Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems In the second decade of the 21st century, annual revenue for the global

More information

The Olympus stereology system. The Computer Assisted Stereological Toolbox

The Olympus stereology system. The Computer Assisted Stereological Toolbox The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,

More information

We can display an object on a monitor screen in three different computer-model forms: Wireframe model Surface Model Solid model

We can display an object on a monitor screen in three different computer-model forms: Wireframe model Surface Model Solid model CHAPTER 4 CURVES 4.1 Introduction In order to understand the significance of curves, we should look into the types of model representations that are used in geometric modeling. Curves play a very significant

More information

Ultrasound. Sound waves

Ultrasound. Sound waves Ultrasound Basic Idea Send waves into body which are reflected at the interfaces between tissue Return time of the waves tells us of the depth of the reflecting surface History First practical application,

More information

GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System

GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System GE Healthcare Vivid 7 Dimension Cardiovascular Ultrasound System New dime Introducing multi-dimensional and 4D imaging. The ability to incorporate multi-dimensional and 4D imaging into your clinical routine

More information

Three Dimensional Ultrasound Imaging

Three Dimensional Ultrasound Imaging Three Dimensional Ultrasound Imaging Hans Torp/ Sevald Berg/Kjell Kristoffersen m/flere Department of circulation and medical imaging NTNU Hans Torp NTNU, Norway Acquisition Reconstruction Filtering Collecting

More information

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft Calibration of AFM with virtual standards; robust, versatile and accurate Richard Koops VSL Dutch Metrology Institute Delft 19-11-2015 VSL Dutch Metrology Institute VSL is the national metrology institute

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

Polarization of Light

Polarization of Light Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights

More information

Force/position control of a robotic system for transcranial magnetic stimulation

Force/position control of a robotic system for transcranial magnetic stimulation Force/position control of a robotic system for transcranial magnetic stimulation W.N. Wan Zakaria School of Mechanical and System Engineering Newcastle University Abstract To develop a force control scheme

More information

Optical Illusions Essay Angela Wall EMAT 6690

Optical Illusions Essay Angela Wall EMAT 6690 Optical Illusions Essay Angela Wall EMAT 6690! Optical illusions are images that are visually perceived differently than how they actually appear in reality. These images can be very entertaining, but

More information

Physics 202 Problems - Week 8 Worked Problems Chapter 25: 7, 23, 36, 62, 72

Physics 202 Problems - Week 8 Worked Problems Chapter 25: 7, 23, 36, 62, 72 Physics 202 Problems - Week 8 Worked Problems Chapter 25: 7, 23, 36, 62, 72 Problem 25.7) A light beam traveling in the negative z direction has a magnetic field B = (2.32 10 9 T )ˆx + ( 4.02 10 9 T )ŷ

More information

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA.

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. Juan Antonio Martínez Mera Centro Singular de Investigación en Tecnoloxías da Información Universidade de Santiago de Compostela

More information

GRAVE: An Interactive Geometry Construction and Visualization Software System for the TORT Radiation Transport Code

GRAVE: An Interactive Geometry Construction and Visualization Software System for the TORT Radiation Transport Code GRAVE: An Interactive Geometry Construction and Visualization Software System for the TORT Radiation Transport Code E. D. Blakeman Oak Ridge National Laboratory Oak Ridge, TN 37831 [email protected] ABSTRACT

More information

Section 1.1. Introduction to R n

Section 1.1. Introduction to R n The Calculus of Functions of Several Variables Section. Introduction to R n Calculus is the study of functional relationships and how related quantities change with each other. In your first exposure to

More information

discuss how to describe points, lines and planes in 3 space.

discuss how to describe points, lines and planes in 3 space. Chapter 2 3 Space: lines and planes In this chapter we discuss how to describe points, lines and planes in 3 space. introduce the language of vectors. discuss various matters concerning the relative position

More information

Reflection and Refraction

Reflection and Refraction Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,

More information

FETAL HEART ASSESSMENT USING THREE-DIMENSIONAL ULTRASOUND

FETAL HEART ASSESSMENT USING THREE-DIMENSIONAL ULTRASOUND FETAL HEART ASSESSMENT USING THREE-DIMENSIONAL ULTRASOUND 1 2 3 T.R. Nelson, Ph.D., M.S. Sklansky, M.D., D.H. Pretorius, M.D. 1 3 Divisions of Physics and Ultrasound, Department of Radiology 2 Division

More information

The process components and related data characteristics addressed in this document are:

The process components and related data characteristics addressed in this document are: TM Tech Notes Certainty 3D November 1, 2012 To: General Release From: Ted Knaak Certainty 3D, LLC Re: Structural Wall Monitoring (#1017) rev: A Introduction TopoDOT offers several tools designed specifically

More information

Part-Based Recognition

Part-Based Recognition Part-Based Recognition Benedict Brown CS597D, Fall 2003 Princeton University CS 597D, Part-Based Recognition p. 1/32 Introduction Many objects are made up of parts It s presumably easier to identify simple

More information

Making Better Medical Devices with Multisensor Metrology

Making Better Medical Devices with Multisensor Metrology Making Better Medical Devices with Multisensor Metrology by Nate J. Rose, Chief Applications Engineer, Optical Gaging Products (OGP) Multisensor metrology is becoming a preferred quality control technology

More information

Common Core Unit Summary Grades 6 to 8

Common Core Unit Summary Grades 6 to 8 Common Core Unit Summary Grades 6 to 8 Grade 8: Unit 1: Congruence and Similarity- 8G1-8G5 rotations reflections and translations,( RRT=congruence) understand congruence of 2 d figures after RRT Dilations

More information

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION

CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CHAPTER 4 QUALITY ASSURANCE AND TEST VALIDATION CINDY WEILAND AND SANDRA L. KATANICK Continued innovations in noninvasive testing equipment provide skilled sonographers and physicians with the technology

More information

A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter

A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter A Measurement of 3-D Water Velocity Components During ROV Tether Simulations in a Test Tank Using Hydroacoustic Doppler Velocimeter Leszek Kwapisz (*) Marek Narewski Lech A.Rowinski Cezary Zrodowski Faculty

More information

Computer Graphics. Geometric Modeling. Page 1. Copyright Gotsman, Elber, Barequet, Karni, Sheffer Computer Science - Technion. An Example.

Computer Graphics. Geometric Modeling. Page 1. Copyright Gotsman, Elber, Barequet, Karni, Sheffer Computer Science - Technion. An Example. An Example 2 3 4 Outline Objective: Develop methods and algorithms to mathematically model shape of real world objects Categories: Wire-Frame Representation Object is represented as as a set of points

More information

Duplication Images in Vascular Sonography

Duplication Images in Vascular Sonography Article Duplication Images in Vascular Sonography Jonathan M. Rubin, MD, PhD, Jing Gao, MD, Keith Hetel, MD, Robert Min, MD Objective. The purpose of this study was to determine the characteristics and

More information

State of Stress at Point

State of Stress at Point State of Stress at Point Einstein Notation The basic idea of Einstein notation is that a covector and a vector can form a scalar: This is typically written as an explicit sum: According to this convention,

More information

INTRODUCTION TO RENDERING TECHNIQUES

INTRODUCTION TO RENDERING TECHNIQUES INTRODUCTION TO RENDERING TECHNIQUES 22 Mar. 212 Yanir Kleiman What is 3D Graphics? Why 3D? Draw one frame at a time Model only once X 24 frames per second Color / texture only once 15, frames for a feature

More information

Diagnostic Ultrasound Performance Testing & Accreditation

Diagnostic Ultrasound Performance Testing & Accreditation Diagnostic Ultrasound Performance Testing & Accreditation Nicholas J Hangiandreou, PhD Associate Professor of Medical Physics Department of Radiology Mayo Clinic and Foundation Rochester, Minnesota [email protected]

More information

A Short Introduction to Computer Graphics

A Short Introduction to Computer Graphics A Short Introduction to Computer Graphics Frédo Durand MIT Laboratory for Computer Science 1 Introduction Chapter I: Basics Although computer graphics is a vast field that encompasses almost any graphical

More information

Computer Animation and Visualisation. Lecture 1. Introduction

Computer Animation and Visualisation. Lecture 1. Introduction Computer Animation and Visualisation Lecture 1 Introduction 1 Today s topics Overview of the lecture Introduction to Computer Animation Introduction to Visualisation 2 Introduction (PhD in Tokyo, 2000,

More information

A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer

A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer Presented at IMAPS 2002 Denver, Colorado September 5, 2002 (Best of Session Award) A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer Jeff Strole*, Scott Corbett*,

More information

E190Q Lecture 5 Autonomous Robot Navigation

E190Q Lecture 5 Autonomous Robot Navigation E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator

More information

THE CONTROL OF A ROBOT END-EFFECTOR USING PHOTOGRAMMETRY

THE CONTROL OF A ROBOT END-EFFECTOR USING PHOTOGRAMMETRY THE CONTROL OF A ROBOT END-EFFECTOR USING PHOTOGRAMMETRY Dr. T. Clarke & Dr. X. Wang Optical Metrology Centre, City University, Northampton Square, London, EC1V 0HB, UK [email protected], [email protected]

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010

Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

Industrial Robotics. Training Objective

Industrial Robotics. Training Objective Training Objective After watching the program and reviewing this printed material, the viewer will learn the basics of industrial robot technology and how robots are used in a variety of manufacturing

More information

Curriculum Map by Block Geometry Mapping for Math Block Testing 2007-2008. August 20 to August 24 Review concepts from previous grades.

Curriculum Map by Block Geometry Mapping for Math Block Testing 2007-2008. August 20 to August 24 Review concepts from previous grades. Curriculum Map by Geometry Mapping for Math Testing 2007-2008 Pre- s 1 August 20 to August 24 Review concepts from previous grades. August 27 to September 28 (Assessment to be completed by September 28)

More information

Index-Velocity Rating Development for Rapidly Changing Flows in an Irrigation Canal Using Broadband StreamPro ADCP and ChannelMaster H-ADCP

Index-Velocity Rating Development for Rapidly Changing Flows in an Irrigation Canal Using Broadband StreamPro ADCP and ChannelMaster H-ADCP Index-Velocity Rating Development for Rapidly Changing Flows in an Irrigation Canal Using Broadband StreamPro ADCP and ChannelMaster H-ADCP HENING HUANG, RD Instruments, 9855 Businesspark Avenue, San Diego,

More information

Graphical displays are generally of two types: vector displays and raster displays. Vector displays

Graphical displays are generally of two types: vector displays and raster displays. Vector displays Display technology Graphical displays are generally of two types: vector displays and raster displays. Vector displays Vector displays generally display lines, specified by their endpoints. Vector display

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

VFI Technology to Change the Way You Work

VFI Technology to Change the Way You Work Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler ultrasound applications.

More information

Multiphysics Software Applications in Reverse Engineering

Multiphysics Software Applications in Reverse Engineering Multiphysics Software Applications in Reverse Engineering *W. Wang 1, K. Genc 2 1 University of Massachusetts Lowell, Lowell, MA, USA 2 Simpleware, Exeter, United Kingdom *Corresponding author: University

More information

ADVANTAGES AND DISADVANTAGES OF THE HOUGH TRANSFORMATION IN THE FRAME OF AUTOMATED BUILDING EXTRACTION

ADVANTAGES AND DISADVANTAGES OF THE HOUGH TRANSFORMATION IN THE FRAME OF AUTOMATED BUILDING EXTRACTION ADVANTAGES AND DISADVANTAGES OF THE HOUGH TRANSFORMATION IN THE FRAME OF AUTOMATED BUILDING EXTRACTION G. Vozikis a,*, J.Jansa b a GEOMET Ltd., Faneromenis 4 & Agamemnonos 11, GR - 15561 Holargos, GREECE

More information

A STRATEGIC PLANNER FOR ROBOT EXCAVATION' by Humberto Romero-Lois, Research Assistant, Department of Civil Engineering

A STRATEGIC PLANNER FOR ROBOT EXCAVATION' by Humberto Romero-Lois, Research Assistant, Department of Civil Engineering A STRATEGIC PLANNER FOR ROBOT EXCAVATION' by Humberto Romero-Lois, Research Assistant, Department of Civil Engineering Chris Hendrickson, Professor, Department of Civil Engineering, and Irving Oppenheim,

More information

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare GE Healthcare Vivid S6 Cardiovascular ultrasound system imagination at work Taking care of everyone. Strong performance and innovative miniaturization make the Vivid S6 a highly desirable system for a

More information

SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION

SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM AND A NEW METHOD OF RESULTS ANALYSIS 1. INTRODUCTION Journal of Machine Engineering, Vol. 15, No.1, 2015 machine tool accuracy, metrology, spindle error motions Krzysztof JEMIELNIAK 1* Jaroslaw CHRZANOWSKI 1 SPINDLE ERROR MOVEMENTS MEASUREMENT ALGORITHM

More information

DESIGN AND EVALUATION OF PROBE WITH THREE DEGREE- OF-FREEDOM FOR NON-DESTRUCTIVE TEST USING THREE- DIMENSIONAL FINITE ELEMENT METHOD

DESIGN AND EVALUATION OF PROBE WITH THREE DEGREE- OF-FREEDOM FOR NON-DESTRUCTIVE TEST USING THREE- DIMENSIONAL FINITE ELEMENT METHOD DESIGN AND EVALUATION OF PROBE WITH THREE DEGREE- OF-FREEDOM FOR NON-DESTRUCTIVE TEST USING THREE- DIMENSIONAL FINITE ELEMENT METHOD Masafumi Aoyanagi Graduate School of Systems and Information Engineering,

More information

Iris Sample Data Set. Basic Visualization Techniques: Charts, Graphs and Maps. Summary Statistics. Frequency and Mode

Iris Sample Data Set. Basic Visualization Techniques: Charts, Graphs and Maps. Summary Statistics. Frequency and Mode Iris Sample Data Set Basic Visualization Techniques: Charts, Graphs and Maps CS598 Information Visualization Spring 2010 Many of the exploratory data techniques are illustrated with the Iris Plant data

More information

Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars

Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars 11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated

More information

AN EXPERT SYSTEM TO ANALYZE HOMOGENEITY IN FUEL ELEMENT PLATES FOR RESEARCH REACTORS

AN EXPERT SYSTEM TO ANALYZE HOMOGENEITY IN FUEL ELEMENT PLATES FOR RESEARCH REACTORS AN EXPERT SYSTEM TO ANALYZE HOMOGENEITY IN FUEL ELEMENT PLATES FOR RESEARCH REACTORS Cativa Tolosa, S. and Marajofsky, A. Comisión Nacional de Energía Atómica Abstract In the manufacturing control of Fuel

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography

Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography Philips CX50 CompactXtreme System A New Class of Compact Echocardiography Insights from compact ultrasound users Diagnostic confidence is paramount Ultrasound labs are busy and workflow efficiency is crucial

More information

NEW MEXICO Grade 6 MATHEMATICS STANDARDS

NEW MEXICO Grade 6 MATHEMATICS STANDARDS PROCESS STANDARDS To help New Mexico students achieve the Content Standards enumerated below, teachers are encouraged to base instruction on the following Process Standards: Problem Solving Build new mathematical

More information

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard

More information

Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008

Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008 Embedded Systems in Healthcare Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008 About the Speaker Working for Philips Research since 1982 Projects

More information

Relating Vanishing Points to Catadioptric Camera Calibration

Relating Vanishing Points to Catadioptric Camera Calibration Relating Vanishing Points to Catadioptric Camera Calibration Wenting Duan* a, Hui Zhang b, Nigel M. Allinson a a Laboratory of Vision Engineering, University of Lincoln, Brayford Pool, Lincoln, U.K. LN6

More information

THREE DIMENSIONAL GEOMETRY

THREE DIMENSIONAL GEOMETRY Chapter 8 THREE DIMENSIONAL GEOMETRY 8.1 Introduction In this chapter we present a vector algebra approach to three dimensional geometry. The aim is to present standard properties of lines and planes,

More information

Robotics. Lecture 3: Sensors. See course website http://www.doc.ic.ac.uk/~ajd/robotics/ for up to date information.

Robotics. Lecture 3: Sensors. See course website http://www.doc.ic.ac.uk/~ajd/robotics/ for up to date information. Robotics Lecture 3: Sensors See course website http://www.doc.ic.ac.uk/~ajd/robotics/ for up to date information. Andrew Davison Department of Computing Imperial College London Review: Locomotion Practical

More information

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely

More information

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

Index-Velocity Rating Development (Calibration) for H-ADCP Real-Time Discharge Monitoring in Open Channels

Index-Velocity Rating Development (Calibration) for H-ADCP Real-Time Discharge Monitoring in Open Channels Index-Velocity Rating Development (Calibration) for H-ADCP Real-Time Discharge Monitoring in Open Channels Hening Huang Teledyne RD Instruments, Inc., 14020 Stowe Drive, Poway, CA. 92064, USA (Tel: 858-842-2600,

More information

HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAYS AND TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAY AND ROAD TUNNEL LININGS.

HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAYS AND TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAY AND ROAD TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAYS AND TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAY AND ROAD TUNNEL LININGS. The vehicle developed by Euroconsult and Pavemetrics and described

More information

Tracking Systems. Kevin Cleary, PhD

Tracking Systems. Kevin Cleary, PhD Tracking Systems Kevin Cleary, PhD Technical Director, Bioengineering Initiative Children's National Medical Center Adjunct Professor, Department of Radiology Medical Center Washington, DC, USA [email protected]

More information

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM

More information

B2.53-R3: COMPUTER GRAPHICS. NOTE: 1. There are TWO PARTS in this Module/Paper. PART ONE contains FOUR questions and PART TWO contains FIVE questions.

B2.53-R3: COMPUTER GRAPHICS. NOTE: 1. There are TWO PARTS in this Module/Paper. PART ONE contains FOUR questions and PART TWO contains FIVE questions. B2.53-R3: COMPUTER GRAPHICS NOTE: 1. There are TWO PARTS in this Module/Paper. PART ONE contains FOUR questions and PART TWO contains FIVE questions. 2. PART ONE is to be answered in the TEAR-OFF ANSWER

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

Voxar 3D TM. A suite of advanced visualization and analysis software tools

Voxar 3D TM. A suite of advanced visualization and analysis software tools Voxar 3D TM A suite of advanced visualization and analysis software tools The power to deliver advanced visualization throughout the enterprise To effectively manage the rapid growth of large volumetric

More information

Health Care Careers in the Field of Imaging. Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist

Health Care Careers in the Field of Imaging. Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist Health Care Careers in the Field of Imaging Shari Workman, MSM,PHR,CIR MultiCare Health System Senior Recruiter/Employment Specialist What is Health Care Imaging? Technologists working with patients, using

More information

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication Thomas Reilly Data Physics Corporation 1741 Technology Drive, Suite 260 San Jose, CA 95110 (408) 216-8440 This paper

More information