1 MIDDLE EAST TECHNICAL UNIVERSITY DEPARTMENT OF AEROSPACE ENGINEERING AE 547 EXPERIMENTAL AERODYNAMICS LABORATORY PROCEDURE VENTURI METER (H5) ANIL GÜÇLÜ

2 TABLE OF CONTENT 1. INTRODUCTION VENTURI METER TEST SETUP INFORMATION... 3 a. Venturi Meter Apparatus... 3 b. Volumetric Hydraulic Bench INSTALLATION AND ASSEMBLY THEORETICAL INFORMATION EXPERIMENTS... 9 a. Discharge Rate Calculations... 9 b. Pressure Calculations

3 1. INTRODUCTION Venturi meters have been used to measure the volume flow along a pipe for many years. The fluid flowing in the pipe passes through a section which is called as throat. The throat has a smaller cross sectional area than other parts of the pipe. Assuming that the volume flow speed is constant, the velocity of the fluid through the throat is higher than the other sides of the pipe. There is a relation of the velocity of the flow and the pressure at the walls. As the velocity increases, the pressure falls proportional to the velocity. So, the flow rate can be calculated with the measured pressure and pipe parameters. 2. VENTURI METER TEST SETUP INFORMATION Venturi meter test setup composed of two main parts, which are Venturi meter apparatus and volumetric hydraulic bench. Detailed information about these components is given in this chapter. a. Venturi Meter Apparatus With using venture meter, which is procured from TecQuipment Company, it is possible to see and measure the complete static head distribution along a horizontal venture tube, Figure 1. Figure 1 Venturi Meter (H5) The Venturi meter has three main holes. These holes can be named as fluid inlet, outlet, and air valve. As seen from Figure 1, fluid is supplied from the inlet (from 3

4 supply) and goes out to the measurement tank from the outlet. Flow velocity is adjusted with a control valve. The order of measured pressure can be changed with changing the air pressure in the manometer. Air valve hole is used to change and set the air pressure of the manometers. Technical details of the Venturi meter can be seen in Table 1. Table 1 Technical Detail of Venturi Meter Item Details Dimensions and Weight 720 mm high x 650 mm wide x 300 mm front to back, 15 kg Nominal Venturi Dimensions Inlet pipe diameter 26 mm Throat diameter 16 mm Water Manometers 11 off 0 to 40 mm water The Venturi tube has two heads. One of them is placed at the cross-sectional area at the upstream section; another one is at the throat section. With the help of eleven manometers, pressure distribution along the meter can be observed. The ideal pressure distribution and the measured pressure distribution can be compared. Coefficients of the discharge for the meter parameter can be identified with using the outcomes of the experiments. b. Volumetric Hydraulic Bench The Volumetric Hydraulic Bench has a mobile water source and a water level indicator. There is a water pool inside the bench. The water is generally is colored as blue. With the water pump, which is placed inside of the bench, the water is pumped from the reservoir, to another test apparatus. It is important that the pump should not be worked when there is not enough water on the reservoir. The water, which goes out of the test apparatus, which is placed, on the bench, goes to the pool again. There are some test apparatus, which is recommended, however the bench can be used for any other suitable test apparatus. The volumetric hydraulic bench can be seen in Figure 2. 4

5 Figure 2 H1D Volumetric Hydraulic Bench There is a valve, which controls the flow of water to an outlet pipe that passes through a hole and connects to another test apparatus. Inside of the volumetric hydraulic test bench can be seen in Figure 3. Figure 3 The Drain Valve There are also some holes on the volumetric tank to prevent forgetting the drain the tank. Inside of the bench can be seen in Figure 4. Figure 4 Inside of the Hydraulic Bench There is an On/Off switch, a fill level indicator, the stack pipe and a flow control valve mounted to the bench, which is shown in Figure 5. 5

6 Figure 5 Tools of the Bench Technical details of the hydraulic test bench are shown in Table 2. Table 2 Technical Details of the Hydraulic Test Bench Item Dimensions and weight Sump tank capacity Volumetric tank capacity Maximum accurate measuring rate Pump Electrical supply Circuit protection Details Net: 1200 mm long x 760 mm wide x 1100 mm high and 80 kg Approximately 160 Liters 35 Liters 2 L/sec 60L/min 220 to 240 VAC 50Hz 1A or 110 to 120 VAC 60Hz 2A or 220 VAC 60Hz 2.5A Thermal overload and under voltage sensor built into on/off switch. 3. INSTALLATION AND ASSEMBLY For measuring the flow rate with Venturi meter test bench, it has to be mounted on the hydraulic test bench. After the Venturi meter is placed on the hydraulic test bench, test can be operated after the below steps are followed. Steps 1-5 are related with the hydraulic test bench and 6-14 is related with the Venturi meter apparatus. 6

7 1. Lift the Drain Valve from its hole in the Volumetric Tank. 2. Pour clean water into Volumetric Tank. 3. Connect the electrical supply. 4. Press the ON button of the on/off switch to start the pump and check the leaks. 5. Switch off the pumps. 6. Put the apparatus on the top of the Hydraulic Bench. 7. Connect the bench supply hose to the inlet of the Venturi meter. 8. Connect the outlet of the Venturi meter to the plastic tube. 9. Set both apparatus flow control and bench supply valve to approximately one third fully open positions. 10. Check that the air valve on the other manifold is tightly closed. 11. Switch on the bench supply and allow water to flow. To clear air from the manometer tubes, you may tilt the apparatus. 12. Shut the apparatus flow control valve. Air now will be trapped in the upper parts of the manometer tubing and manifold. 13. Open air valve just enough to allow water to rise approximately halfway up the manometer scale. 14. Watch the values on the manometers and note it down. 15. You will record 3 data sets while turning the valve. 4. THEORETICAL INFORMATION There are some notations which are used during the theoretical calculations. Table 3 Notation Symbol Meaning Units u, u1, u2 and un Velocity of flow m.s -1 D Diameter of the pipe m h, h1, h2 and hn Head m of water Δh Head Differential (h1-h2) m of water Q Volume flow or discharge m 3.s -1 a, a1 and a2 Cross-sectional areas m 2 C Flow Coefficient - G Acceleration due to gravity 9.81m.s -2 The relation among the velocity at the throat (u2), the volume flow, and the cross sectional area is the given in equation 1. u 2 = Q a 2 (1) 7

8 Figure 6 Ideal Conditions in a Venturimeter For an ideal venturimeter, Bernoulli s theorem states that; Constant flow along the pipe can be found as; Substituting the equation 2 from equation 3 yields; u h 2g 1 = u 2 + h 2g 2 = u n + h 2g n (2) Q = u 1 a 1 = u 2 a 2 = u n a n (3) u 2 2 2g (a 2 After solving the equations for u2 leads; ) 2 + h a 1 = u 2 1 2g + h 2 (4) u 2 = 2g(h 1 h 2 ) Lastly, the discharge rate (volume flow) can be found as; 1 ( a 2 a1 )2 (5) Q = a 2 2g(h 1 h 2 ) Ideal dimensionless pressure can be obtained from; 1 ( a 2 a1 )2 (6) ( a 2 a 1 ) 2 ( a 2 a n ) 2 (7) 8

9 5. EXPERIMENTS After the experiment setup is prepared for test, the pressure tap values on the venturi meter manometers are observed and recorded with rotating the valve. Then, you must have ten different data sets at different flow rates. According to the data set, you will do some calculations about discharge rate and the pressure. a. Discharge Rate Calculations During the experiment, write down the heights of the manometer bars in the cells in Table 4. Note: Q1 flow rate is zero which means there is no flow (valve is closed), Q3 is the fastest flow rate among Q1-Q3. Table 4 Experimental Data Sets Tapping Heads (h) (mm) Q1 Q2 Q3 A (Upstream) (h1) B C D (Throat) (h2) E F G H J K L 9

10 Manometer Heights (mm) Draw the heights of the manometers in the following plot at different discharge rates (Q1-Q3) Q1 Q2 20 Q A B C D E F G H J K L Venturi Meter Sections Figure 7 Measured Data vs. Venturi Meter Sections

11 According to the heights data at 3 different discharge rates, calculate the each discharge rate Q1-Q3 and write down in Error! Reference source not found.. Table 5 Discharge Rate Values ( x 10-4 m 3 /s) Q1 Q2 Q3 b. Pressure Calculations In pressure calculations, you will calculate and compare the ideal and actual dimensionless pressure values. For ideal case, the venturi shape parameters will be used (Figure 8). Figure 8 Venturi Meter Parameters 11

12 Fill the Table 6 with your calculations. Table 6 Ideal Pressure Distribution Throat Area (a 2) / Upstream Area (a 1) = (a 2/a 1) 2 = Tapping Distance along Venturi (mm) Diameter (m) Area (m 2 ) Throat Area/Area (a 2/a n) (a 2/a n) 2 Ideal Dimensionless Pressure ( a 2 ) a 1 A (Upstream)(a 1) 0 0 B(a 3) C(a 4) D(Throat)(a 2) 1 E(a 5) F(a 6) G(a 7) H(a 8) J(a 9) 2 2 ( a 2 a n ) K(a 10) L(a 11) 12

13 The actual pressures will be calculated in Table 7. As stated before, there are ten measurement data sets. Fill the actual pressure distribution table, Table 7, according to the flow rate, Q3 which you calculated before. Table 7 Actual Pressure Distribution Flow Q 3 = 2 u 2 2g = Tapping Distance along Venturi (mm) h n (m) h n-h 1 (m) Actual Dimensionless Pressure h n h 1 (u 2 2 /2g) A (Upstream)(h 1) B(h 3) C(h 4) D(Throat)(h 2) E(h 5) F(h 6) G(h 7) H(h 8) J(h 9) K(h 10) L(h 11) 13

14 Calculate the pressure values at each section of flow discharge rates Q1-3 (x 10-4 m 3 /s) are given in Table 8. Table 8 Actual Pressure Distribution for all Qs Tapping Q1 Q2 Q3 A(Upstream)(h 1) B(h 3) C(h 4) D(Throat)(h 2) E(h 5) F(h 6) G(h 7) H(h 8) J(h 9) K(h 10) L(h 11) 14

15 Dimensionless Pressure After calculating the actual pressure values, compare with the ideal pressure values. The data graph of the pressure values are given in Figure Ideal Q1 Q2 0.4 Q A B C D E F G H J K L Distance along venturi (mm) Figure 9 Pressure Distribution Results 15

Experiment 3 Pipe Friction

Experiment 3 Pipe Friction EML 316L Experiment 3 Pipe Friction Laboratory Manual Mechanical and Materials Engineering Department College of Engineering FLORIDA INTERNATIONAL UNIVERSITY Nomenclature Symbol Description Unit A cross-sectional

More information

FLUID FLOW Introduction General Description

FLUID FLOW Introduction General Description FLUID FLOW Introduction Fluid flow is an important part of many processes, including transporting materials from one point to another, mixing of materials, and chemical reactions. In this experiment, you

More information

The University of Toledo Soil Mechanics Laboratory

The University of Toledo Soil Mechanics Laboratory The University of Toledo Soil Mechanics Laboratory Permeability Testing - 1 Constant and Falling Head Tests Introduction In 1856 the French engineer Henri D arcy demonstrated by experiment that it is possible

More information

Equipment for Engineering Education

Equipment for Engineering Education Equipment for Engineering Education Instruction Manual Venturi Flume G.U.N.T. Gerätebau GmbH Fahrenberg 4 D-885 Barsbüttel Germany Phone: ++49 (40) 670854.0 Fax: ++49 (40) 670854.4 E-mail: sales@gunt.de

More information

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness

More information

Flow Measurement Options for Pipeline and Open Channel Flow

Flow Measurement Options for Pipeline and Open Channel Flow Flow Measurement Options for Pipeline and Open Channel Flow October 2013 Presented by Molly Skorpik - 2013 Montana Association of Dam and Canal Systems Conference Irrigation Training and Research Center

More information

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Technical Information sheet 0001/07/Rev01 ENG FUNCTION Series S001 and S002 circulation units are applied to the primary circuit of solar

More information

Power consumption (kw)

Power consumption (kw) OS Series Series description LIT OS systems are applied to wastewater disinfection. The models of this series were developed on base of high-efficiency low-pressure amalgam lamps and have all components

More information

Experiment (13): Flow channel

Experiment (13): Flow channel Introduction: An open channel is a duct in which the liquid flows with a free surface exposed to atmospheric pressure. Along the length of the duct, the pressure at the surface is therefore constant and

More information

Experiment # 3: Pipe Flow

Experiment # 3: Pipe Flow ME 05 Mechanical Engineering Lab Page ME 05 Mechanical Engineering Laboratory Spring Quarter 00 Experiment # 3: Pipe Flow Objectives: a) Calibrate a pressure transducer and two different flowmeters (paddlewheel

More information

1-Way Chilled Water Cassettes KCO 60, 90 & 120. Technical Brochure. TM KCO-W.1GB Date : November 2005 Supersedes : None

1-Way Chilled Water Cassettes KCO 60, 90 & 120. Technical Brochure. TM KCO-W.1GB Date : November 2005 Supersedes : None 1-Way Chilled Water Cassettes KCO 60, 90 & 120 Technical Brochure TM KCO-W.1GB Date : November 2005 Supersedes : None General Features Control valve (optional) Air vent In-tube electric heating elements

More information

CEE 370 Fall 2015. Laboratory #3 Open Channel Flow

CEE 370 Fall 2015. Laboratory #3 Open Channel Flow CEE 70 Fall 015 Laboratory # Open Channel Flow Objective: The objective of this experiment is to measure the flow of fluid through open channels using a V-notch weir and a hydraulic jump. Introduction:

More information

HYDRAULICS. H91.8D/C - Computerized Open Surface Tilting Flow Channel - 10, 12.5, 15 and 20 m long

HYDRAULICS. H91.8D/C - Computerized Open Surface Tilting Flow Channel - 10, 12.5, 15 and 20 m long HYDRAULICS H91.8D/C - Computerized Open Surface Tilting Flow Channel - 10, 12.5, 15 and 20 m long 1. General The series of channels H91.8D has been designed by Didacta Italia to study the hydrodynamic

More information

Please also see 10 golden rules for correct Saniflo installation on www.saniflo.co.uk

Please also see 10 golden rules for correct Saniflo installation on www.saniflo.co.uk 1. Description Sanicom is a lifting pump for wastewater used in commercial applications (restaurants, hairdressing salons, the catering industry etc.). The unit operates automatically, has a high performance

More information

Flow Over Weirs. By John Fuller. Fluid Mechanics Lab. Wednesday(1-345pm) Group member: Abdur Rahaman

Flow Over Weirs. By John Fuller. Fluid Mechanics Lab. Wednesday(1-345pm) Group member: Abdur Rahaman Flow Over Weirs By John Fuller Fluid Mechanics Lab Wednesday(1-345pm) Group member: Abdur Rahaman Abstract The objective of this lab is to determine the characteristics of open-channel flow over, firstly,

More information

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction

More information

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector Magne3000 PLUS Smart Electromagnetic Flowmeter Open channel Flowmeter Detector Model NNK150/951 OVERVIE The Magne3000 PLUS Electromagnetic Flowmeter is submersible type of flowmeter mainly used for flow

More information

Maxi Pressurisation Units (Maxi, Maxi Plus & HP)

Maxi Pressurisation Units (Maxi, Maxi Plus & HP) Maxi Pressurisation Units (Maxi, Maxi Plus & HP) Installation and Maintenance Instructions Contents 1 Dimensions 3 1.1 Installation 4 1.1 Mains/boosted cold water inlet 4 1.2 Connection to the system 5

More information

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection. www.swagelok.com

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection. www.swagelok.com www.swagelok.com Valve Sizing Te chnic al Bulletin Scope Valve size often is described by the nominal size of the end connections, but a more important measure is the flow that the valve can provide. And

More information

Unit 24: Applications of Pneumatics and Hydraulics

Unit 24: Applications of Pneumatics and Hydraulics Unit 24: Applications of Pneumatics and Hydraulics Unit code: J/601/1496 QCF level: 4 Credit value: 15 OUTCOME 2 TUTORIAL 3 HYDRAULIC AND PNEUMATIC MOTORS The material needed for outcome 2 is very extensive

More information

Model 1210C Battery Powered Pump Shown. Description

Model 1210C Battery Powered Pump Shown. Description 12 Volt DC Rotary Vane Pump Series 1200C Model 1210C Battery Powered Pump Shown Description of Included Models Model Number FR1205C FR1210C FR1211C FR2410C FR2411C Description Basic 12 volt DC pump with

More information

1 DESCRIPTION OF THE APPLIANCE

1 DESCRIPTION OF THE APPLIANCE 1 DESCRIPTION OF THE APPLIANCE 1.1 INTRODUCTION The cast iron SF boilers are a valid solution for the present energetic problems, since they can run with solid fuels: wood and coal. These series of boilers

More information

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to:

Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: I. OBJECTIVE OF THE EXPERIMENT. Swissmetro travels at high speeds through a tunnel at low pressure. It will therefore undergo friction that can be due to: 1) Viscosity of gas (cf. "Viscosity of gas" experiment)

More information

Installation and Commissioning Manual Solar Stations FlowCon Max Evolution II

Installation and Commissioning Manual Solar Stations FlowCon Max Evolution II Installation and Commissioning Manual Solar Stations FlowCon Max Evolution II 1 For your safety Table of Contents 1 For your safety... 3 1.1 About this manual... 3 1.2 Designated use... 3 1.3 Qualification

More information

METHOD OF TEST FOR DETERMINATION OF PERMEABILITY OF GRANULAR SOILS

METHOD OF TEST FOR DETERMINATION OF PERMEABILITY OF GRANULAR SOILS Laboratory Testing Manual Date: 99 06 21 Page 1 of 7 METHOD OF TEST FOR DETERMINATION OF PERMEABILITY OF GRANULAR SOILS 1. SCOPE 1.1 This method covers the determination of the coefficient of permeability

More information

Electrical supply to the pump motor is via a switch and RCD mounted in the recess on the front of the bench.

Electrical supply to the pump motor is via a switch and RCD mounted in the recess on the front of the bench. FLUID MECHANICS LAB EQUIPMENTS 1. HYDRAULIC BENCH The hydraulics bench consists of a moulded plastic sump tank which supports GRP bench top incorporating a flow channel and volumetric measuring tank. A

More information

Conlift1, Conlift2, Conlift2 ph+

Conlift1, Conlift2, Conlift2 ph+ GRUNDFOS DATA BOOKLET Conlift1, Conlift2, Conlift2 ph+ Small lifting stations 50 Hz CONLIFT1, CONLIFT2, CONLIFT2 ph+ Table of contents 1. Product overview 3 Conlift for condensate applications 3 Applications

More information

SPARE PARTS EXPLODED VIEW GAS WALL BOILERS Model GENUS PREMIUM FS CUENOD. R8216470-01EN - Edition 01-26/10/2009

SPARE PARTS EXPLODED VIEW GAS WALL BOILERS Model GENUS PREMIUM FS CUENOD. R8216470-01EN - Edition 01-26/10/2009 Instantaneous Gas Water Heaters Storage Gas Water Heaters SPARE PARTS EXPLODED VIEW GAS WALL BOILERS Model R8216470-01EN - Edition 01-26/10/2009 Boilers CUENOD Renewable 3121488 12 EAA1...... 3124527 25

More information

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING Toshiharu Kagawa 1, Yukako Saisu 2, Riki Nishimura 3 and Chongho Youn 4 ABSTRACT In this paper, we developed a new laminar flow

More information

oil liquid water water liquid Answer, Key Homework 2 David McIntyre 1

oil liquid water water liquid Answer, Key Homework 2 David McIntyre 1 Answer, Key Homework 2 David McIntyre 1 This print-out should have 14 questions, check that it is complete. Multiple-choice questions may continue on the next column or page: find all choices before making

More information

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any Chapter 10 Flow Measurements Material from Theory and Design for Mechanical Measurements; Figliola, Third Edition Flow Rate Flow rate can be expressed in terms of volume flow rate (volume/time) or mass

More information

Designed for multi-injector endurance testing, the

Designed for multi-injector endurance testing, the Designed for multi-injector endurance testing, the ETB Test Bench can optionally provide shot-to-shot mass measurement. The injectors are precisely subject to variable controls and hydraulic pressures

More information

ME 315 - Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS

ME 315 - Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS ME 315 - Heat Transfer Laboratory Nomenclature Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS A heat exchange area, m 2 C max maximum specific heat rate, J/(s

More information

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5

FREE FALL. Introduction. Reference Young and Freedman, University Physics, 12 th Edition: Chapter 2, section 2.5 Physics 161 FREE FALL Introduction This experiment is designed to study the motion of an object that is accelerated by the force of gravity. It also serves as an introduction to the data analysis capabilities

More information

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump Heat Thermodynamic cycle Heat pump LD Physics Leaflets P2.6.3.2 Investigating the function of the expansion valve of the heat pump Objects of the experiment g To study the operational components of the

More information

Extremely compact in size to allow direct flange-mounting on vehicle engine or gearbox PTOs.

Extremely compact in size to allow direct flange-mounting on vehicle engine or gearbox PTOs. TXV - Presentation pumps with Load Sensing control variable displacement piston pumps ADVANTAGES pumps are variable displacement with pressure-flow control called Load Sensing. They self-regulate to give

More information

Fluids and Solids: Fundamentals

Fluids and Solids: Fundamentals Fluids and Solids: Fundamentals We normally recognize three states of matter: solid; liquid and gas. However, liquid and gas are both fluids: in contrast to solids they lack the ability to resist deformation.

More information

Replacement parts. WM97+ gas-fired water boiler Boiler Manual. OBTAIN PARTS ONLY THROUGH WEIL-McLAIN THE BOILER CONTAINS CERAMIC FIBER MATERIALS

Replacement parts. WM97+ gas-fired water boiler Boiler Manual. OBTAIN PARTS ONLY THROUGH WEIL-McLAIN THE BOILER CONTAINS CERAMIC FIBER MATERIALS Replacement parts DO NOT SERVICE THE BOILER WITHOUT A WM97+ MAINTENANCE KIT AVAILABLE Failure to adhere to these guidelines can result in severe personal injury, death or substantial property damage. The

More information

ECOCIAT. Domestic hot water heat recovery unit

ECOCIAT. Domestic hot water heat recovery unit Heat recovery unit Domestic hot water High energy efficiency with R410A Compact and quiet Scroll compressors Brazed-plate heat exchangers Heating Heat recovery ENVIRONMENTALLY HFC R410A PROTECTION DE FRIENDLY

More information

KESSEL KTP 500 Submersible Pump with lateral and vertical outlet connection Order Numbers 28710, 28810, 28850

KESSEL KTP 500 Submersible Pump with lateral and vertical outlet connection Order Numbers 28710, 28810, 28850 INSTALLATION AND OPERATING INSTRUCTIONS KESSEL KTP 500 Submersible Pump with lateral and vertical outlet connection Order Numbers 28710, 28810, 28850 Product advantages Automatic float switch as water

More information

Unit 24: Applications of Pneumatics and Hydraulics

Unit 24: Applications of Pneumatics and Hydraulics Unit 24: Applications of Pneumatics and Hydraulics Unit code: J/601/1496 QCF level: 4 Credit value: 15 OUTCOME 2 TUTORIAL 2 HYDRAULIC AND PNEUMATIC CYLINDERS The material needed for outcome 2 is very extensive

More information

Lift and Drag on an Airfoil ME 123: Mechanical Engineering Laboratory II: Fluids

Lift and Drag on an Airfoil ME 123: Mechanical Engineering Laboratory II: Fluids Lift and Drag on an Airfoil ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction In this lab the characteristics of airfoil lift, drag,

More information

CONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions

CONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions Mounting and operating instructions CONTOIL DFM 8EDM Table of contents Safety instructions 2 - Designed use - Installation, commissioning and operation - Operational safety - Return of the instruments

More information

PART 2 FORKLIFT HYDRAULIC SYSTEM

PART 2 FORKLIFT HYDRAULIC SYSTEM PART 2 FORKLIFT HYDRAULIC SYSTEM Chapter 1 Description and Operation Component Locations & Circuit Layouts 1 Hydraulic Pump 11 Control Valve 14 Valve Section Oil Flows 15 Anti-Cavitation Valve 22 Velocity

More information

WASTE OIL DRAINERS 99. Waste oil drainers

WASTE OIL DRAINERS 99. Waste oil drainers WASTE OIL DRAINERS 99 8 100 WASTE OIL DRAINERS wheel-mounted for drums 20/220 kg Art. 42050 (20/60 kg) Art. 42000 (180/220 kg) Art. 42050 Waste oil drainer for tanks or drums composed of: art. 80050 Trolley

More information

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER FLOW MEASUREMENT 200 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER Dr D G Stewart, NEL Dr M Reader-Harris, NEL Dr R J W Peters, McCrometer Inc INTRODUCTION The V-Cone

More information

WATER MEASUREMENT USING TWO INCH (50 mm) DRAIN TESTS

WATER MEASUREMENT USING TWO INCH (50 mm) DRAIN TESTS GAP.14.1.2.2 A Publication of Global Asset Protection Services LLC WATER MEASUREMENT USING TWO INCH (50 mm) DRAIN TESTS INTRODUCTION A hydrant or other large-volume flow test is necessary for proper water

More information

! WARNING. McDonnell & Miller Installation & Maintenance Instructions MM-217(I) Series 150S and 157S (Snap Switch, All Models except 157S-RB-P)

! WARNING. McDonnell & Miller Installation & Maintenance Instructions MM-217(I) Series 150S and 157S (Snap Switch, All Models except 157S-RB-P) Series 150S and 157S (Snap Switch, All Models except 157S-RB-P) Low Water Cut-Off/Pump Controllers For Steam Boilers and Other Level Control Applications McDonnell & Miller Installation & Maintenance Instructions

More information

PRO-CLEAN. Technical Description. Specifications. Space-saving design with minimum heat loss (melamine resin insulation)

PRO-CLEAN. Technical Description. Specifications. Space-saving design with minimum heat loss (melamine resin insulation) PRO-CLEAN Product description Specifications Technical Description Max. backup operating pressure Max. service water operating pressure 3 bar 6 bar Hot water connection " female thread* V4A (no..4) Cold

More information

Oildyne. 165 Series Hydraulic Power Units. Pressures to 241 bar (3500 psi) Flows to 5.4 lpm (1.4 gpm)

Oildyne. 165 Series Hydraulic Power Units. Pressures to 241 bar (3500 psi) Flows to 5.4 lpm (1.4 gpm) Oildyne Hydraulic Power Units Pressures to 241 bar (3500 psi) Flows to 5.4 lpm (1.4 gpm) 11 Power Unit Features We are pleased to introduce our new power units. The power units let you put more power where

More information

Pressure in Fluids. Introduction

Pressure in Fluids. Introduction Pressure in Fluids Introduction In this laboratory we begin to study another important physical quantity associated with fluids: pressure. For the time being we will concentrate on static pressure: pressure

More information

HERZ-Thermal Actuators

HERZ-Thermal Actuators HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01

More information

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES 191: Calibration and Standards in Measurement Richard Paton National Engineering Laboratory, Scotland, UK 1 General Principles 1 2 Gravimetric Calibration of Liquid meters 2 3 Volumetric Calibration of

More information

Max primary circuit temperature 90ºC Max primary circuit temp. 90ºC Max secondary circuit temperature 45ºC Max secondary circuit temp.

Max primary circuit temperature 90ºC Max primary circuit temp. 90ºC Max secondary circuit temperature 45ºC Max secondary circuit temp. EGLISH 1 Product description exchanger equipped with an electronic control unit and circulation pump for the primary circuit. All Aqua-Mex variants can be ordered with an interior coil of either titanium

More information

The Unique Accelabar Flow Meter

The Unique Accelabar Flow Meter The Unique Accelabar Flow Meter The Accelabar is a new and unique flow meter that combines two differential pressure technologies to produce operating ranges never before attainable in a single flow meter.

More information

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS W. M. Torres, P. E. Umbehaun, D. A. Andrade and J. A. B. Souza Centro de Engenharia Nuclear Instituto de Pesquisas Energéticas e Nucleares

More information

Understanding the Flo-Dar Flow Measuring System

Understanding the Flo-Dar Flow Measuring System Understanding the Flo-Dar Flow Measuring System Independent tests verify non-contact flowmeter is highly accurate under both open channel and surcharge conditions TM Flo-Dar is the only non-contact open

More information

Water hammering in fire fighting installation

Water hammering in fire fighting installation Water hammering in fire fighting installation Forward One of major problems raised in the fire fighting network installed at Pioneer company for pharmaceutical industry /Sulaymania was the high water hammering

More information

This unit is a pump that will clear the waste from a domestic shower and washbasin, bath, washing machine, glass washer and a sink.

This unit is a pump that will clear the waste from a domestic shower and washbasin, bath, washing machine, glass washer and a sink. 1. Description This unit is a pump that will clear the waste from a domestic shower and washbasin, bath, washing machine, glass washer and a sink. The unit will start automatically. Please pay particular

More information

What is the most obvious difference between pipe flow and open channel flow????????????? (in terms of flow conditions and energy situation)

What is the most obvious difference between pipe flow and open channel flow????????????? (in terms of flow conditions and energy situation) OPEN CHANNEL FLOW 1 3 Question What is the most obvious difference between pipe flow and open channel flow????????????? (in terms of flow conditions and energy situation) Typical open channel shapes Figure

More information

HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT

HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT HIGH PRESSURE TECHNOLOGY HYDRAULICS PNEUMATICS TESTING EQUIPMENT MAXIMATOR GmbH 2 MAXIMATOR has an extensive know-how concerning concept, development, construction and manufacturing of test benches and

More information

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS 1. Define an accumulator and explain its function A hydraulic accumulator is a device that stores the potential energy of an incompressible fluid

More information

SCP 012-130 ISO. Other advantages: SCP 012-130 ISO is a series of piston pumps with a fixed displacement for mobile and stationary hydraulics.

SCP 012-130 ISO. Other advantages: SCP 012-130 ISO is a series of piston pumps with a fixed displacement for mobile and stationary hydraulics. SCP 012-130 ISO is a series of piston pumps with a fixed displacement for mobile and stationary hydraulics. SCP 012-130 ISO covers the entire displacement range 12-130 cmᶟ rev. at a maximum pressure of

More information

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme Grant Agreement No. 228296 SFERA Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME Capacities Specific Programme Research Infrastructures Integrating Activity - Combination of

More information

Hydraulics Laboratory Experiment Report

Hydraulics Laboratory Experiment Report Hydraulics Laboratory Experiment Report Name: Ahmed Essam Mansour Section: "1", Monday 2-5 pm Title: Flow in open channel Date: 13 November-2006 Objectives: Calculate the Chezy and Manning coefficients

More information

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS Section A.1. Flow Coefficients Definition The flow coefficient or pressure loss coefficient is used to relate the pressure loss

More information

PERMEABILITY TEST. To determine the coefficient of permeability of a soil using constant head method.

PERMEABILITY TEST. To determine the coefficient of permeability of a soil using constant head method. PERMEABILITY TEST A. CONSTANT HEAD OBJECTIVE To determine the coefficient of permeability of a soil using constant head method. need and Scope The knowledge of this property is much useful in solving problems

More information

CENTRIFUGAL PUMP OVERVIEW Presented by Matt Prosoli Of Pumps Plus Inc.

CENTRIFUGAL PUMP OVERVIEW Presented by Matt Prosoli Of Pumps Plus Inc. CENTRIFUGAL PUMP OVERVIEW Presented by Matt Prosoli Of Pumps Plus Inc. 1 Centrifugal Pump- Definition Centrifugal Pump can be defined as a mechanical device used to transfer liquid of various types. As

More information

80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500

80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500 Release R.2.0 Uk 80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500 Installation User and Service Manual www.state-waterheaters.co.uk your installer Warning Read this manual carefully before first

More information

Pre-installation Manual. CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP

Pre-installation Manual. CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP Pre-installation Manual CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP Pre-installation Manual CONCEPT Processor 305 DW / 405 DW / 505 DW / 505 DW XP T11336 Edition AA, April 2013 This book has

More information

Activity P13: Buoyant Force (Force Sensor)

Activity P13: Buoyant Force (Force Sensor) July 21 Buoyant Force 1 Activity P13: Buoyant Force (Force Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Archimedes Principle P13 Buoyant Force.DS P18 Buoyant Force P18_BUOY.SWS

More information

Installation, operation and maintenance manual TX 35A

Installation, operation and maintenance manual TX 35A Installation, operation and maintenance manual TX 35A Rev.10 may 2012 Page 1 of 18 1.0.0 Table of contents INSTALLATION, OPERATION AND MAINTENANCE MANUAL...1 1.0.0 TABLE OF CONTENTS...2 2.0.0 ILLUSTRATIONS...2

More information

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER Valve size often is described by the nominal size of the end connections but a more important measure is the flow that

More information

C. starting positive displacement pumps with the discharge valve closed.

C. starting positive displacement pumps with the discharge valve closed. KNOWLEDGE: K1.04 [3.4/3.6] P78 The possibility of water hammer in a liquid system is minimized by... A. maintaining temperature above the saturation temperature. B. starting centrifugal pumps with the

More information

ME 305 Fluid Mechanics I. Part 8 Viscous Flow in Pipes and Ducts

ME 305 Fluid Mechanics I. Part 8 Viscous Flow in Pipes and Ducts ME 305 Fluid Mechanics I Part 8 Viscous Flow in Pipes and Ducts These presentations are prepared by Dr. Cüneyt Sert Mechanical Engineering Department Middle East Technical University Ankara, Turkey csert@metu.edu.tr

More information

A10VSO Nominal size 18 see RD 92712 A10VSO Nominal size 28...140 see RD 92711

A10VSO Nominal size 18 see RD 92712 A10VSO Nominal size 28...140 see RD 92711 A10VO Nominal size 18 see RD 92712 A10VO Nominal size 28...140 see RD 92711 The variable displacement axial piston pump A10VO in swashplate design was designed for hydrostatic drives in open circuits.

More information

CE 6303 MECHANICS OF FLUIDS L T P C QUESTION BANK PART - A

CE 6303 MECHANICS OF FLUIDS L T P C QUESTION BANK PART - A CE 6303 MECHANICS OF FLUIDS L T P C QUESTION BANK 3 0 0 3 UNIT I FLUID PROPERTIES AND FLUID STATICS PART - A 1. Define fluid and fluid mechanics. 2. Define real and ideal fluids. 3. Define mass density

More information

Lab 1a Wind Tunnel Testing Principles & Lift and Drag Coefficients on an Airfoil

Lab 1a Wind Tunnel Testing Principles & Lift and Drag Coefficients on an Airfoil Lab 1a Wind Tunnel Testing Principles & Lift and Drag Coefficients on an Airfoil OBJECTIVES - Calibrate the RPM/wind speed relation of the wind tunnel. - Measure the drag and lift coefficients of an airfoil

More information

MFP14, MFP14S and MFP14SS Automatic Pumps

MFP14, MFP14S and MFP14SS Automatic Pumps Page 1 of 6 TI-P136-02 ST Issue 12 Cert. No. LRQ 0963008 ISO 9001, S and SS utomatic Pumps Description The Spirax Sarco automatic pump is a displacement receiver operated by steam or compressed air. It

More information

Builder Specifications for Solar Ready Homes

Builder Specifications for Solar Ready Homes Builder Specifications for Solar Ready Homes This document specifies the essential elements required to make a new home ready for the future installation of roof-mounted solar domestic hot water and photovoltaic

More information

Troubleshooting Guide for Jacks Down LED Lights

Troubleshooting Guide for Jacks Down LED Lights Troubleshooting Guide for Jacks Down LED Lights Equalizer Systems Auto-Level systems manufactured after 2005 feature a pressure switch system that monitors the retracted state of leveling jacks and any

More information

Fuel Cell Characterization Lab

Fuel Cell Characterization Lab SJSU E 10 Introduction to Engineering Fuel Cell Characterization Lab What is a Fuel Cell? Before we answer that question, let s first review the process of electrolysis. In the process of electrolysis,

More information

How To Use A Corona Mdi 110

How To Use A Corona Mdi 110 CORONA MDI 110 APPLICATION Domestic water meter FEATURES 4 Modular multi-jet domestic meter with inductive scanning (without magnetic influence) 4 System capability, with standard applicable pulse output

More information

SANITARY SEWER SPECIFICATIONS

SANITARY SEWER SPECIFICATIONS SANITARY SEWER SPECIFICATIONS OCTOBER 2003 HARVEST-MONROVIA WATER, SEWER, AND FIRE PROTECTION AUTHORITY SECTION 1.00 1.10 Purpose The purpose of this document is to assemble the sewer specifications, policies,

More information

Experimental investigation of the filling and draining processes of the drainback system (Part 1)

Experimental investigation of the filling and draining processes of the drainback system (Part 1) Available online at www.sciencedirect.com Energy Procedia 00 (013) 000 000 www.elsevier.com/locate/procedia Abstract 013 ISES Solar World Congress Experimental investigation of the filling and draining

More information

Chapter 15. FLUIDS. 15.1. What volume does 0.4 kg of alcohol occupy? What is the weight of this volume? m m 0.4 kg. ρ = = ; ρ = 5.

Chapter 15. FLUIDS. 15.1. What volume does 0.4 kg of alcohol occupy? What is the weight of this volume? m m 0.4 kg. ρ = = ; ρ = 5. Chapter 15. FLUIDS Density 15.1. What volume does 0.4 kg of alcohol occupy? What is the weight of this volume? m m 0.4 kg ρ = ; = = ; = 5.06 x 10-4 m ρ 790 kg/m W = D = ρg = 790 kg/m )(9.8 m/s )(5.06 x

More information

Practice Problems on Boundary Layers. Answer(s): D = 107 N D = 152 N. C. Wassgren, Purdue University Page 1 of 17 Last Updated: 2010 Nov 22

Practice Problems on Boundary Layers. Answer(s): D = 107 N D = 152 N. C. Wassgren, Purdue University Page 1 of 17 Last Updated: 2010 Nov 22 BL_01 A thin flat plate 55 by 110 cm is immersed in a 6 m/s stream of SAE 10 oil at 20 C. Compute the total skin friction drag if the stream is parallel to (a) the long side and (b) the short side. D =

More information

Hydraulic Jumps and Non-uniform Open Channel Flow, Course #507. Presented by: PDH Enterprises, LLC PO Box 942 Morrisville, NC 27560 www.pdhsite.

Hydraulic Jumps and Non-uniform Open Channel Flow, Course #507. Presented by: PDH Enterprises, LLC PO Box 942 Morrisville, NC 27560 www.pdhsite. Hydraulic Jumps and Non-uniform Open Channel Flow, Course #507 Presented by: PDH Enterprises, LLC PO Box 942 Morrisville, NC 27560 www.pdhsite.com Many examples of open channel flow can be approximated

More information

Chapter 3.5: Fans and Blowers

Chapter 3.5: Fans and Blowers Part I: Objective type questions and answers Chapter 3.5: Fans and Blowers 1. The parameter used by ASME to define fans, blowers and compressors is a) Fan ration b) Specific ratio c) Blade ratio d) Twist

More information

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration

More information

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter 2 Table of Contents Abstract... 3 Pressure Fundamentals... 3 Applications...

More information

2-Way heavy duty flow control, with pressure compensated, solenoid and load sensing controlled priority flow

2-Way heavy duty flow control, with pressure compensated, solenoid and load sensing controlled priority flow 2-Way heavy duty flow control, with pressure compensated, solenoid and load sensing controlled priority flow RE 18309-64/04.10 1/6 Replaces: RE 00171/02.07 A-VRFC2C-VEI-VS-LS 0M.28.03.80 - Y - Z Description

More information

Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV.

Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV. Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV. MULTIPOINT Please read and understand these instructions before starting work. Please leave this leaflet

More information

NUMERICAL ANALYSIS OF THE EFFECTS OF WIND ON BUILDING STRUCTURES

NUMERICAL ANALYSIS OF THE EFFECTS OF WIND ON BUILDING STRUCTURES Vol. XX 2012 No. 4 28 34 J. ŠIMIČEK O. HUBOVÁ NUMERICAL ANALYSIS OF THE EFFECTS OF WIND ON BUILDING STRUCTURES Jozef ŠIMIČEK email: jozef.simicek@stuba.sk Research field: Statics and Dynamics Fluids mechanics

More information

Examples for Heat Exchanger Design

Examples for Heat Exchanger Design for Heat Exchanger Design Lauterbach Verfahrenstechnik GmbH 1 / 2011 Contents Calculation 1 1. Water- Water Heat Exchanger 1 Basics...1 Task...1 1. Start the WTS program...1 2. Selection of basic data...1

More information

Technical data. Danfoss DHP-A

Technical data. Danfoss DHP-A Technical data Danfoss DHP-A An air heat pump which produces both heat and hot water Can operate efficiently down to -0 C Danfoss TWS tank gives plenty of hot water quickly and with low operating costs

More information

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by:

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by: Flow Measurement in Pipes and Ducts, Course #503 Presented by: PDH Enterprises, LLC PO Box 942 Morrisville, NC 27560 www.pdhsite.com This course is about measurement of the flow rate of a fluid flowing

More information

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF Table of Contents Section 1 - General Introduction 1.1 1 Principles Of Operation 1.2 1 Reshaping The Velocity Profile

More information

Gas Unit Heaters Duct Furnaces Horizontal - 60 HZ Form No. LF24-100-400 (10/2011) LF24 Standard

Gas Unit Heaters Duct Furnaces Horizontal - 60 HZ Form No. LF24-100-400 (10/2011) LF24 Standard LF Standard Features Horizontal unit heaters are compact in design and easy to install. Steady State Efficiencies of up to 80.%. Ideal for economical commercial heating systems. Shipped completely assembled

More information