Assessing the Complementarities of MALDI and ESI for Protein Identification in Complex Mixtures

Size: px
Start display at page:

Download "Assessing the Complementarities of MALDI and ESI for Protein Identification in Complex Mixtures"

Transcription

1 Assessing the Complementarities of MALDI and ESI for Protein Identification in Complex Mixtures Comparing the AB SCIEX TOF/TOF 5800 Analyzer with the ESI Orbitrap using ProteinPilot Software Sean Seymour 1, Aaron Booy 2, Rebekah Gundry 3, Jennifer Van Eyk 3, Christie Hunter 1 1 Applied Biosystems, USA, 2 MDS Analytical Technologies, Canada, 3 The Johns Hopkins University, USA There are two dominant soft ionization strategies in proteomics today for biomarker discovery and protein identification in complex mixtures. On-line LCMS using electrospray ionization (ESI) is commonly used with mass spectrometers such as ion traps, quadrupole time of flight, and Orbitrap instruments. Matrix Assisted Laser Desorption Ionization (MALDI) is the other main soft ionization technique and is normally coupled with Time of Flight (TOF) detection systems. With the advent of LC MALDI 1, it has also become widely used in off-line LCMS strategies. While LC MALDI has much more flexibility in acquisition strategies due to the separation of LC from MS analysis, it has traditionally been speed limited relative to the ESI approach. With the faster laser repetition rate possible on the TOF/TOF 5800 System and the intelligent DynamicExit Algorithm for efficient acquisition, the speed of LC MALDI is now 10-15X faster 2 then previous instruments. As LC MALDI begins to approach the speed of on-line ESI strategies, it is worthwhile investigating the complementarities of the techniques to understand how to best leverage the strengths of each and get the most information from complex proteomics samples. To do the study properly, it was important to assess both the differences between instruments and the differences within multiple acquisitions on each instrument. The strategy used was to measure three replicates of a complex biological sample (E.coli cell lysate) at three different sample loadings on column and assess the differences. Reverse database searching with ProteinPilot Software 3.0 was used to measure the numbers of proteins and peptides detected at a fixed false discovery rate (FDR) from each instrument. Figure 1. Summary of ESI and MALDI Acquisitions Protein Level. Proteins identified at 1% global FDR for each of the three replicate injections at each sample loading is shown. Replicate acquisitions were performed on both the ESI-Orbitrap platform (red) and the TOF/TOF 5800 System (blue) to enable proper comparison of these orthogonal acquisitions strategies.

2 Figure 2. Summary of ESI and MALDI Acquisitions Peptide Level. Peptides identified at 1% global FDR for each of the three replicate injections at each sample loading is shown. Replicate acquisitions were performed on both the ESI-Orbitrap platform and the TOF/TOF 5800 System to enable proper comparison of these orthogonal acquisitions strategies. Materials and Methods Sample Preparation: E.coli lysate was re-suspended in 50% trifluoroethanol/50 mm ammonium biocarbonate buffer at 1.35 mg/ml (by weight). After heat denaturation, reduction (dithiothreitol) and alkylation (iodoacetamide) were performed, the sample was digested with trypsin overnight at 37 C. Sample was dried using vacuum centrifugation, then re-suspended at 1 mg/ml in 2% acetonitrile / 0.1% formic acid. Sample loadings of 20, 100 and 500 ngs were run in three replicates for each loading. LC MALDI Acquisition: Peptides were separated on a Monolithic CapRod column (100 µm ID x 15cm, Merck, Germany) with a 90 minute linear gradient of 5-35% acetonitrile at 2 µl/min (in 0.1% TFA) using the Tempo LC MALDI Spotting System. 720 spots were collected per LC run with online mixing of α-cyano-4-hydroxycinnamic acid matrix (Agilent). MS/MS spectra were acquired on the 25 most intense precursors per spot using a laser frequency of 1000 Hz and DynamicExit Algorithm on the TOF/TOF 5800 System (for more acquisition details 2 ). LC ESI Acquisition: Peptides were separated on a selfpacked column (75 µm x 10cm, Magic C 18 AQ (Michrom) in a PicoTip emitter (New Objective)) using a 90 min linear gradient of 5-35% acetonitrile at 300 nl/min (Agilent 1200 LC system). The eluent was directed into an LTQ Orbitrap (ThermoFisherScientific). MS scans were acquired in the Orbitrap and the top 7 precursor ions above a threshold of 8000 counts were selected for MS/MS in the LTQ.Data acquisition parameters were optimized according to the vendor s recommendations 3, and included monoisotopic precursor selection, 2+, 3+ charge state selection, and a dynamic exclusion repeat count of 1. Database searching: Each dataset was searched with ProteinPilot Software 3.0 against a database of all E.coli proteins in UniProt as of Jan. 27, 2009 (4379) plus 77 common contaminant proteins using the Paragon Algorithm in thorough mode with biological modifications and substitutions 4. False discovery rate (FDR) analysis was done by on-the-fly analysis of the reversed sequences using the embedded PSPEP tool 5. For processing of the Orbitrap data, an mgf file was generated using ReAdw4Mascot2.exe, a NIST derivative of the ISB tool, ReAdW. Protein and Peptide ID Results Three different sample loadings of E.coli were analyzed by ESI Orbitrap and LC MALDI 5800 system and compared using the protein and peptide identifications at the 1% global FDR. The processing of the three data acquisition sets is summarized in Figure 1 and 2. At the protein level (Figure 1), the results at the 20 ng level were quite comparable. At the higher sample loading of 100 and 500 ngs, a higher number of proteins were routinely seen using the LC MALDI strategy. The same observation was seen at the peptide level (Figure 2), with the LC MALDI approach finding ~2x more peptides at the highest loading amounts.

3 Figure 3. Intersection of Proteins and Peptides Identified by ESI vs. MALDI. Left: The best replicate acquisitions from the 500 ng sample loading experiments were compared to assess the intersection between ESI and MALDI strategies, at the protein level (1% global FDR) and the peptide level (5% local FDR). Right: Correlating the Unused Prot Score (UPS) per protein between instruments clearly shows that more peptide identifications using the 5800 system yields higher UPS and therefore higher protein confidence. Protein and Peptide Level Intersection Using the 500 ng sample loading level, the intersection between the best ESI-Orbitrap acquisition was compared to the best LC MALDI 5800 acquisition at the protein and peptide level (Figure 3, left), using the 1% global FDR values. Out of a total of 778 proteins identified, 466 proteins (60%) were shared with the 5800 system finding about 30% more proteins. Even more significant differences were observed when comparing the peptide level intersection, where only 25% of peptides were shared between acquisitions. Here, LC MALDI on the 5800 system found about 108% more peptides than the ESI-Orbitrap system, meaning more proteins and more sequence coverage for the detected proteins. The value of higher numbers of peptides for the purpose of protein identification can be determined by aligning the protein level identifications and comparing the Unused Prot Score (UPS - a measure of protein confidence in ProteinPilot Software) from each acquisition strategy. As a rough approximation, each 2 units of UPS correspond to one high confidence peptide sequence segment (excluding multiple modification states). The additional peptide coverage by LC MALDI clearly translates into more confident protein identifications (Figure 3, right, log scaled) as seen by the higher proportion of data above the diagonal. Assessing the Peptide Level Intersection A comparison using only one replicate is insufficient and provides only limited insight. It is important to assess both intra- and inter-instrument reproducibility. The degree of similarity between each pair of replicates both within and across instruments was calculated as the ratio of the intersection to the union of identifications for the replicate pair (Figure 4). Figure 4. Peptide-Level Intersection between All Acquisitions at 500 ng Sample Loading Level. Each cell shows the fraction of all peptides found in each pair of replicates that are common to both. Formatting from red to green indicates increasing similarity. The average intersection at the peptide level between replicates on the TOF/TOF 5800 system at the 500 ng loading was high at 64% (Figure 4, Samples 1-3). The average intersection between the Orbitrap replicates was also high at 61% (Samples 4-6).However, if the peptide comparison is done between platforms, the average intersection decreases to only 25% (Figure 4, red boxes). Comparison of replicates within each instrument clearly show more similarity than comparisons of replicates from different instruments.the source of this difference is an interesting investigation.

4 Figure 5. Differences in Peptide Detection by ESI and MALDI. Of the peptides detected at 500ng loading, the proportion of each terminating in Lys, Arg or other residues was determined (left). MALDI finds a higher proportion of Arginine-terminating peptides than ESI. The peptide populations found by a single technique (right) shows a more distinctive trend. Because of the bias of MALDI towards finding more Arginine-terminating peptides, there were many more Arg peptides found only in the MALDI strategy, and Lysine-terminating peptides in the ESI strategy. Differences in Peptide Populations Previous work has suggested that there is a slight bias towards specific types of peptides detected between ESI and MALDI experiments. However, without proper assessment of the random nature of acquisition, no real trends can be assessed. Here, using three replicate acquisitions as the basis for our analysis, we can assess real acquisition based differences. As shown in Figure 5, there are higher proportions of Arginine (Arg)-terminating peptides detected in LC MALDI and this is consistent across replicates. In the ESI acquisition, the distribution of peptides between Lysine (Lys) and Arg is roughly equal. Therefore, when the unique peptides found by only LC MALDI were analyzed, there was a very high proportion of Arg peptides found (~67%) vs only ~33% by ESI (Figure 5, right). The mass of the peptides found were also analyzed, to understand if there were any real differences in peptide size. Figure 6 shows the distribution of precursors found across all replicates. No strong differences are seen between acquisition strategies, although there are small differences that appear consistent across replicates. Figure 6. Mass Distributions of Replicates. The analysis of mass distributions reveals only small but consistent differences in the masses of peptides acquired by each instrument.

5 Conclusions By analyzing both the intra- and inter-instrument acquisition replicates, a much clearer picture into the complementary nature of ESI and MALDI can be ascertained. The LC MALDI using the AB SCIEX TOF/TOF 5800 system reproducibly obtains more proteins and peptides per acquisition over the ESI-Orbitrap acquisition at the higher samples loadings. At lower amounts on column, the acquisition results are comparable. The analysis of the peptide intersection demonstrates that distinct peptide populations are detected when using LC MALDI vs. ESI acquisition (Figure 3). This suggests that a complementary strategy combining both ESI and MALDI analysis could provide deeper proteome coverage. Consideration of combinations of replicates (Figure 7, 500ng) shows that multiple replicate acquisitions on an ESI-Orbitrap does not equal the peptide or protein yield from a single LC MALDI run. This comparison is of interest because of the acquisition time consideration, 3 LCMS runs takes ~7.5 hours of instrument time and 1 LC MALDI run takes ~7 hours 2. Another interesting comparison is that the combination of two 5800 system runs is nearly equivalent to a combination of 1 ESI-Orbitrap run and 1 LC MALDI run. Since one ESI and one MALDI run can be done in less time and the differences in the peptide populations detected, the combined use of these two platforms might be a valuable strategy for deep proteome coverage. References 1. Hattan et al. J. Prot. Res. (2005) 4, Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests. AB SCIEX Technical Note Advantage of the LTQ Orbitrap for Protein Identification in Complex Digests, ThermoFisher Scientific App note: 386, Shilov IV et al. (2007) Mol. Cell Prot. 6.9, Tang WH et al. (2008) J Prot. Res. 7, Figure 7. Relative Benefits of Repeat Acquisitions. The plot of the ESI-Orbitrap data (red) and the LC MALDI 5800 system data (blue) at both the protein (top) and peptide level (bottom) demonstrates similar relative gains from increasing replicates although the absolute numbers significantly favor the 5800 system. The purple line assesses the scenario of combining equal numbers of replicates from ESI and MALDI acquisitions. For example, the combination of system run with 1 Orbitrap run (purple point at 2 replicates) shows more peptide identifications over a single 5800 system run (blue point at 1 replicate). 115TN19-01 Trademarks of Life Technologies and it affiliated companies: Applied Biosystems and AB (Design). Trademarks of Applied Biosystems / MDS Analytical Technologies (a joint venture between Applied Biosystems and MDS, Inc.): MALDI TOF/TOF, TOF/TOF, DynamicExit, ProteinPilot, Paragon, and OptiBeam. All other trademarks are the sole property of their respective owners Applied Biosystems and MDS Inc. Joint Owners. All rights reserved. Information subject to change without notice. Headquarters 850 Lincoln Centre Drive Foster City, CA USA Phone Toll Free International Sales For our office locations please call the division headquarters or refer to our Website at

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein

More information

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics

More information

ProteinPilot Report for ProteinPilot Software

ProteinPilot Report for ProteinPilot Software ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Application Note: 386 Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Rosa Viner, Terry Zhang, Scott Peterman, and Vlad Zabrouskov, Thermo Fisher Scientific, San Jose, CA,

More information

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF

More information

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples

More information

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,

More information

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes Investigating Biological Variation of Liver Enzymes in Human Hepatocytes MS/MS ALL with SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 AB SCIEX, USA, 2 Pfizer, USA

More information

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry Thermo Scientific Pierce Controls and Standards for Mass Spectrometry calibration tools for mass spectrometry Ensure confidence in instrument performance with Thermo Scientific Pierce Calibration Solutions

More information

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can

More information

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:

More information

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn

Aiping Lu. Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn Aiping Lu Key Laboratory of System Biology Chinese Academic Society APLV@sibs.ac.cn Proteome and Proteomics PROTEin complement expressed by genome Marc Wilkins Electrophoresis. 1995. 16(7):1090-4. proteomics

More information

Mass Spectrometry Based Proteomics

Mass Spectrometry Based Proteomics Mass Spectrometry Based Proteomics Proteomics Shared Research Oregon Health & Science University Portland, Oregon This document is designed to give a brief overview of Mass Spectrometry Based Proteomics

More information

Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance?

Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance? Optimization 1 Choices, choices, choices... Which sequence database? Which modifications? What mass tolerance? Where to begin? 2 Sequence Databases Swiss-prot MSDB, NCBI nr dbest Species specific ORFS

More information

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis

MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis Overview MarkerView software is a novel program designed for metabolomics applications and biomarker profiling workflows 1. Using

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories. Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine

More information

Proteomics in Practice

Proteomics in Practice Reiner Westermeier, Torn Naven Hans-Rudolf Höpker Proteomics in Practice A Guide to Successful Experimental Design 2008 Wiley-VCH Verlag- Weinheim 978-3-527-31941-1 Preface Foreword XI XIII Abbreviations,

More information

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application

More information

ProSightPC 3.0 Quick Start Guide

ProSightPC 3.0 Quick Start Guide ProSightPC 3.0 Quick Start Guide The Thermo ProSightPC 3.0 application is the only proteomics software suite that effectively supports high-mass-accuracy MS/MS experiments performed on LTQ FT and LTQ Orbitrap

More information

Supplementary Materials and Methods (Metabolomics analysis)

Supplementary Materials and Methods (Metabolomics analysis) Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in

More information

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry

ProteinScape. Innovation with Integrity. Proteomics Data Analysis & Management. Mass Spectrometry ProteinScape Proteomics Data Analysis & Management Innovation with Integrity Mass Spectrometry ProteinScape a Virtual Environment for Successful Proteomics To overcome the growing complexity of proteomics

More information

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction

More information

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,

More information

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap

Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap Proteomic data analysis for Orbitrap datasets using Resources available at MSI. September 28 th 2011 Pratik Jagtap The Minnesota http://www.mass.msi.umn.edu/ Proteomics workflow Trypsin Protein Peptides

More information

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080 Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien allis@stanford.edu 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

Introduction to Proteomics 1.0

Introduction to Proteomics 1.0 Introduction to Proteomics 1.0 CMSP Workshop Tim Griffin Associate Professor, BMBB Faculty Director, CMSP Objectives Why are we here? For participants: Learn basics of MS-based proteomics Learn what s

More information

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate

More information

Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report

Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

MASCOT Search Results Interpretation

MASCOT Search Results Interpretation The Mascot protein identification program (Matrix Science, Ltd.) uses statistical methods to assess the validity of a match. MS/MS data is not ideal. That is, there are unassignable peaks (noise) and usually

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008

Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008 Bruker Daltonics Application Note # MT-90 MALDI-TDS: A Coherent MALDI Top-Down-Sequencing Approach Applied to the ABRF-Protein Research Group Study 2008 In the ABRF-PRG study 2008 [*] the ability to characterize

More information

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100)

Standard Mixture. TOF/TOF Calibration Mixture. Calibration Mixture 1 (Cal Mix 1, 1:10) Calibration Mixture 1 (Cal Mix 1, 1:100) Mass Standards Kit for Calibration of AB SCIEX TOF/TOF Instruments Protocol 1 Product Description The Mass Standards Kit includes reagents needed to test instrument function, optimize instrument parameters,

More information

Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS

Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS Application note Explosives by Direct Injection LC/MS/MS Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS API 5000 LC/MS/MS System

More information

High resolution mass spectrometry (HRMS*) in Graz

High resolution mass spectrometry (HRMS*) in Graz High resolution mass spectrometry (HRMS*) in Graz 10.2.2012 * Instruments with high resolution where exact mass measurements can be performed 1 HRMS in Graz s and Ionization Techniques 2 6 3 5 4 1 TU Graz

More information

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;

More information

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature

TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the

More information

Master course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam

Master course KEMM03 Principles of Mass Spectrometric Protein Characterization. Exam Exam Master course KEMM03 Principles of Mass Spectrometric Protein Characterization 2010-10-29 kl 08.15-13.00 Use a new paper for answering each question! Write your name on each paper! Aids: Mini calculator,

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

Protein Prospector and Ways of Calculating Expectation Values

Protein Prospector and Ways of Calculating Expectation Values Protein Prospector and Ways of Calculating Expectation Values 1/16 Aenoch J. Lynn; Robert J. Chalkley; Peter R. Baker; Mark R. Segal; and Alma L. Burlingame University of California, San Francisco, San

More information

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping For Conclusive Characterization of Biologics Deep Protein Characterization Is Crucial Pharmaceuticals have historically been small

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2

More information

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.

More information

A Hybrid Microchip/Capillary Electrophoresis Mass Spectrometry Platform for Rapid and Ultrasensitive Bioanalysis

A Hybrid Microchip/Capillary Electrophoresis Mass Spectrometry Platform for Rapid and Ultrasensitive Bioanalysis A Hybrid Microchip/Capillary Electrophoresis Mass Spectrometry Platform for Rapid and Ultrasensitive Bioanalysis Ryan Kelly 1, Chenchen Wang, Cheng S. Lee, Richard D. Smith, Keqi Tang 1 Environmental Molecular

More information

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs

More information

Tutorial for proteome data analysis using the Perseus software platform

Tutorial for proteome data analysis using the Perseus software platform Tutorial for proteome data analysis using the Perseus software platform Laboratory of Mass Spectrometry, LNBio, CNPEM Tutorial version 1.0, January 2014. Note: This tutorial was written based on the information

More information

An In-Gel Digestion Protocol

An In-Gel Digestion Protocol An In-Gel Digestion Protocol This protocol describes the digestion of a protein present in an SDS-PAGE gel band with trypsin. The band can be taken from either a 1D or 2D electrophoresis gel. Reagents

More information

Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench

Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench Agilent G2721AA/G2733AA Spectrum Mill MS Proteomics Workbench Application Guide Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any

More information

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF

Tutorial for Proteomics Data Submission. Katalin F. Medzihradszky Robert J. Chalkley UCSF Tutorial for Proteomics Data Submission Katalin F. Medzihradszky Robert J. Chalkley UCSF Why Have Guidelines? Large-scale proteomics studies create huge amounts of data. It is impossible/impractical to

More information

25-4-2007 MALDI-MS Product overview 1 Company Instrument Benchtop (j/n)

25-4-2007 MALDI-MS Product overview 1 Company Instrument Benchtop (j/n) 25-4-2007 MALDI-MS Product overview 1 Voyager TM DE- PRO ja ja

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

Proteomic Analysis using Accurate Mass Tags. Gordon Anderson PNNL January 4-5, 2005

Proteomic Analysis using Accurate Mass Tags. Gordon Anderson PNNL January 4-5, 2005 Proteomic Analysis using Accurate Mass Tags Gordon Anderson PNNL January 4-5, 2005 Outline Accurate Mass and Time Tag (AMT) based proteomics Instrumentation Data analysis Data management Challenges 2 Approach

More information

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,

More information

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment

More information

A Streamlined Workflow for Untargeted Metabolomics

A Streamlined Workflow for Untargeted Metabolomics A Streamlined Workflow for Untargeted Metabolomics Employing XCMS plus, a Simultaneous Data Processing and Metabolite Identification Software Package for Rapid Untargeted Metabolite Screening Baljit K.

More information

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische

More information

SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays

SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays Protocol SpikeTides TM Peptides for relative and absolute quantification in SRM and MRM Assays Contact us: InfoLine: +49-30-6392-7878 Order per fax: +49-30-6392-7888 or e-mail: www: peptide@jpt.com www.jpt.com

More information

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3

Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3 Fast and Automatic Mapping of Disulfide Bonds in a Monoclonal Antibody using SYNAPT G2 HDMS and BiopharmaLynx 1.3 Hongwei Xie and Weibin Chen Waters Corporation, Milford, MA, U.S. A P P L I C AT ION B

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Real-time PCR: Understanding C t

Real-time PCR: Understanding C t APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence

More information

NUVISAN Pharma Services

NUVISAN Pharma Services NUVISAN Pharma Services CESI MS Now available! 1st CRO in Europe! At the highest levels of quality. LABORATORY SERVICES Equipment update STATE OF THE ART AT NUVISAN CESI MS Now available! 1st CRO in Europe!

More information

La Protéomique : Etat de l art et perspectives

La Protéomique : Etat de l art et perspectives La Protéomique : Etat de l art et perspectives Odile Schiltz Institut de Pharmacologie et de Biologie Structurale CNRS, Université de Toulouse, Odile.Schiltz@ipbs.fr Protéomique et Spectrométrie de Masse

More information

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone

More information

Session 1. Course Presentation: Mass spectrometry-based proteomics for molecular and cellular biologists

Session 1. Course Presentation: Mass spectrometry-based proteomics for molecular and cellular biologists Program Overview Session 1. Course Presentation: Mass spectrometry-based proteomics for molecular and cellular biologists Session 2. Principles of Mass Spectrometry Session 3. Mass spectrometry based proteomics

More information

Software for protein identification and quantitative analysis of mass spectrometry data used for protein characterization and proteomics Version 5.

Software for protein identification and quantitative analysis of mass spectrometry data used for protein characterization and proteomics Version 5. Release Notes ProteinPilot Software for protein identification and quantitative analysis of mass spectrometry data used for protein characterization and proteomics Version 5.0 Introduction Thank you for

More information

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik

Laboration 1. Identifiering av proteiner med Mass Spektrometri. Klinisk Kemisk Diagnostik Laboration 1 Identifiering av proteiner med Mass Spektrometri Klinisk Kemisk Diagnostik Sven Kjellström 2014 kjellstrom.sven@gmail.com 0702-935060 Laboration 1 Klinisk Kemisk Diagnostik Identifiering av

More information

360 Master Mix. , and a supplementary 360 GC Enhancer.

360 Master Mix. , and a supplementary 360 GC Enhancer. Product Bulletin AmpliTaq Gold 360 Master Mix and 360 DNA Polymerase AmpliTaq Gold 360 Master Mix AmpliTaq Gold 360 DNA Polymerase 360 Coverage for a Full Range of Targets AmpliTaq Gold 360 Master Mix

More information

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird

More information

Quantitative proteomics background

Quantitative proteomics background Proteomics data analysis seminar Quantitative proteomics and transcriptomics of anaerobic and aerobic yeast cultures reveals post transcriptional regulation of key cellular processes de Groot, M., Daran

More information

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics

泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics 泛 用 蛋 白 質 體 學 之 質 譜 儀 資 料 分 析 平 台 的 建 立 與 應 用 Universal Mass Spectrometry Data Analysis Platform for Quantitative and Qualitative Proteomics 2014 Training Course Wei-Hung Chang ( 張 瑋 宏 ) ABRC, Academia

More information

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access

More information

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data

SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data SimGlycan Software*: A New Predictive Carbohydrate Analysis Tool for MS/MS Data Automated Data Interpretation for Glycan Characterization Jenny Albanese1, Matthias Glueckmann2 and Christof Lenz2 1Applied

More information

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation Craig Aurand 1, David Bell 1, Anders Fridstrom 2, and Klaus Buckendahl 2 1 Supelco, Div.

More information

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics

Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics Pep-Miner: A Novel Technology for Mass Spectrometry-Based Proteomics Ilan Beer Haifa Research Lab Dec 10, 2002 Pep-Miner s Location in the Life Sciences World The post-genome era - the age of proteome

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples Mary Lame, Hua Yang, Sherri Naughton, and Erin Chambers Waters Corporation, Milford,

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

High-Throughput 3-D Chromatography Through Ion Exchange SPE

High-Throughput 3-D Chromatography Through Ion Exchange SPE High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample

More information

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS

More information

Supporting Information. Minimum active structure of insulin-like. peptide 5 (INSL5)

Supporting Information. Minimum active structure of insulin-like. peptide 5 (INSL5) Supporting Information Minimum active structure of insulin-like peptide 5 (INSL5) Alessia Belgi 1,2, Ross A.D. Bathgate *1,2,3, Martina Kocan *4, Nitin Patil 1,5, Suode Zhang 1, Geoffrey W. Tregear 1,2,

More information

2.500 Threshold. 2.000 1000e - 001. Threshold. Exponential phase. Cycle Number

2.500 Threshold. 2.000 1000e - 001. Threshold. Exponential phase. Cycle Number application note Real-Time PCR: Understanding C T Real-Time PCR: Understanding C T 4.500 3.500 1000e + 001 4.000 3.000 1000e + 000 3.500 2.500 Threshold 3.000 2.000 1000e - 001 Rn 2500 Rn 1500 Rn 2000

More information

Introduction to Proteomics

Introduction to Proteomics Introduction to Proteomics Why Proteomics? Same Genome Different Proteome Black Swallowtail - larvae and butterfly Biological Complexity Yeast - a simple proteome 6,113 proteins = 344,855 tryptic peptides

More information

How Mascot Integra helps run a Core Lab

How Mascot Integra helps run a Core Lab How Mascot Integra helps run a Core Lab 1 Areas where a database can help a core lab Project, experiment and sample tracking Flexibility in experiment design Role based security Automation Custom results

More information

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen

More information

itraq Tips and Tricks

itraq Tips and Tricks itraq Tips and Tricks Darryl Pappin Cold Spring Harbor Laboratory Patrick Emery Matrix Science Ltd. 1 Good morning. Unfortunately Darryl Pappin is unable to attend the Matrix Science workshop and the ASMS

More information

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural

More information

Increased Power for the Analysis of Label-Free LC- MS/MS Proteomic Data by Combining Spectral Counts and Peptide Peak Attributes

Increased Power for the Analysis of Label-Free LC- MS/MS Proteomic Data by Combining Spectral Counts and Peptide Peak Attributes Increased Power for the Analysis of Label-Free LC- MS/MS Proteomic Data by Combining Spectral Counts and Peptide Peak Attributes Lee Dicker, Xihong Lin, & Alexander R. Ivanov Corresponding author: email,

More information

Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories

Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories Matthew Clabaugh, Market Development AACC Workshop St. Louis, MO September 17,18 2013 Factors Influencing LCMSMS Moving into

More information

using ms based proteomics

using ms based proteomics quantification using ms based proteomics lennart martens Computational Omics and Systems Biology Group Department of Medical Protein Research, VIB Department of Biochemistry, Ghent University Ghent, Belgium

More information

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program TÁMOP-4.1.1.C-13/1/KONV-2014-0001 projekt Az élettudományi-klinikai felsőoktatás gyakorlatorientált és hallgatóbarát korszerűsítése a vidéki képzőhelyek nemzetközi versenyképességének erősítésére program

More information

Error Tolerant Searching of Uninterpreted MS/MS Data

Error Tolerant Searching of Uninterpreted MS/MS Data Error Tolerant Searching of Uninterpreted MS/MS Data 1 In any search of a large LC-MS/MS dataset 2 There are always a number of spectra which get poor scores, or even no match at all. 3 Sometimes, this

More information