DENSITY. Density. Principles of Imaging Science II (RAD 120)

Size: px
Start display at page:

Download "DENSITY. Density. Principles of Imaging Science II (RAD 120)"

Transcription

1 Principles of Imaging Science II (RAD 120) Radiographic Density 1 DENSITY One of the photographic properties that determines visibility of detail Overall blackness or darkness of the entire radiographic image or a specific area When evaluating an image for proper radiographic density, the density of the entire image is considered 2 Density Only the anatomical parts should be evaluated for proper density The collimated border surrounding the structure represents x-ray beam attenuation in air The radiologist evaluates the image for anatomic or pathologic changes that cause a change in normal density 3 1

2 DENSITY CONTROLLING FACTOR: mas (Quantity) mas = ma X time (sec) 25 mas = 100 ma X 0.25 sec (250 ms) 25 mas = 200 ma X sec (125 ms) 25 mas = 400 ma X sec (62.5 ms) ma, time inversely proportional to maintain the same mas or density Why change ma, time when a desired mas is known? 4 Density Controlling Factors ma = mas/time Calculate the ma, given 50 mas, 100 ms 500 ma Calculate the ma, given 100 mas, 100 ms 1000 ma time = mas/ma Calculate the time, given 20 mas, 200 ma 0.1 sec, 100 ms Calculate the time, given 20 mas, 400 ma 0.5 sec, 50 ms 5 Density Evaluation Your initial reaction should determine if the radiograph demonstrates proper density If it is not correct, then it is too dark (overexposed), or too light (underexposed) You must determine the extent to which it is to dark or too light. 6 2

3 Density Evaluation Over and underexposure is generally expressed as either 50% or 100% incorrect If a radiograph is very light or dark, it is 100% over or underexposed Underexposure - correct by multiplying the mas by 2 (40 mas vs 80 mas) Overexposure correct by multiplying the mas by 0.5 (40 mas vs 20 mas) 7 Density Evaluation If a radiograph is minimally light or dark and warrants a repeat, it is 50% over or underexposed Underexposure - correct by multiplying the mas by 1.5 (40 mas vs 60 mas) Overexposure correct by multiplying the mas by 0.75 (40 mas vs 30 mas) Remember that at least a 30% change in mas is needed to demonstrate a visible difference in density. Known as the 30% rule. mas is the factor of choice for density changes 8 mas 9 3

4 Density: Influencing Factors Kilovoltage Controls the energy, penetrability or quality of the x-ray beam. As kvp is increased or decreased, the quantity of x-ray photons produced increases or decreases 15% rule application 15% change in kvp results in 100% change in density 7.5% change in kvp results in 50% change in density 10 Density Evaluation When evaluating density determine if the structure is properly penetrated. This applies to underexposed images. An image cannot be overexposed if it is not adequately penetrated Look for structures behind another If it is penetrated, change the mas; if not change the kvp according to the 15% rule 11 Density: INFLUENCING FACTORS kilovoltage I 1: I 2 : kvp 1: I 1 kvp 1 2 == I 2 kvp 2 2 Beginning Intensity New Intensity Beginning kilovoltage kvp 2 : New kilovoltage 12 4

5 Density: kilovoltage calculations 13 Density: Influencing Factors Source-Image Distance (SID) Distance between the x-ray source(fs) and the image receptor Affects the intensity of the x-ray beam according to the inverse square law If the SID is changed, and density should be maintained, apply the Density Maintenance Formula 14 Density: INFLUENCING FACTORS Source - Image Distance mas 1 D 1 2 == mas 2 D 2 2 mas 2 = mas 1 (D 22 ) (D 12 ) mas 1: mas 2 : D 1: D 2 : Beginning mas New mas Beginning distance New distance 15 5

6 Density: Distance calculations. 16 Density: INFLUENCING FACTORS There is a direct relationship between RSS and density Common RSS range from As RSS increases, density increases As RSS decreases, density decreases Based upon no compensatory changes in mas A radiograph will be dark if you intended to use an extremity cassette and you incorrectly selected a high speed cassette. 17 Density: INFLUENCING FACTORS Film/Screen Combination mas 1 RSS 2 == mas 2 RSS 1 mas 2 = mas 1 (RSS 1 ) (RSS 2 ) mas 1: mas 2 : RSS 1: RSS 2 : Beginning mas New mas Beginning Film/Screen Speed New Film/Screen Speed 18 6

7 Density: RSS calculations 19 Density: INFLUENCING FACTORS Grids are used to absorb scattered photon emitted from the patient before they interact with the image receptor (IR) Placed between the patient and IR When a grid is used, you must adjust mas to compensate for the loss of density that would be produced by scatter 20 Density: INFLUENCING FACTORS Inverse relationship between grid use and density An image would be more dense if no grid were used compared to using a grid Using a grid requires higher mas according to the following: Grid Ratio GCF 5:1 2X 6:1 3X 8:1 4X 10:1 4X 12:1 5X 16:1 6X Non-grid 1X 21 7

8 Density: INFLUENCING FACTORS Grids mas 1 GCF 1 == mas 2 GCF 2 mas 2 = mas 1 (GCF 2 ) (GCF 1 ) mas 1: mas 2 : GCF 1: GCT 2 : Beginning mas New mas Beginning grid conversion factor New grid conversion factor 22 Density: Grid calculations 23 Density: String Formula mas 2 = mas 1 (D 2 ) 2 (GCF 2 ) (RSS 1 ) (D 1 ) 2 (GCF 1 ) (RSS 2 ) 24 8

9 Density: String Formula 25 Density: String Formula 26 Density: String Formula mas 2 = mas 1 (D 2 ) 2 (GCF 2 ) (RSS 1 ) (D 1 ) 2 (GCF 1 ) (RSS 2 ) 27 9

10 Density: Influencing Factors Generator Type An AP projection of the hip is taken using a three phase, 12 pulse high frequency generator. Technical factors are 70 kvp, 24 mas, 400 RSS, 40 SID, 10:1 grid. Using the correct conversion factor(2.0), calculate the mas needed when using a single phase, 2 pulse unit. To maintain image density use the formula mas 2 = mas 1 X conversion factor mas 2 = Generator Type Conversions & Voltage Ripple Voltage waveforms resulting from various power supplies. The ripple of the kilovoltage is indicated as a percentage for each waveform. 29 Density Influencing Factors Beam Restriction (Collimation) Less scatter radiation is produced and density is decreased 14 X 17 to 8 X 10, increase mas 50-60% 14 X 17 to 10 X 12, increase mas 30-40% When collimation is increased, the amount of irradiated tissue decreases 30 10

11 Density: Influencing Factors Filtration Indirect relationship with density Increased filtration = Decreased density Film Processing Direct relationship with density for developer time, temperature, and replenishment rates Focal Spot Size, Anode Heel Effect Minimal effect 31 Anode Heel Effect 32 Density: Influencing Factors Anatomical Part Thickness or Tissue Indirect relationship with density Larger patients attenuate more x-ray photons Higher atomic # tissues attenuate more x-ray photons Pathology Additive Disease - Density decreases Destructive Disease Density increases Contrast Media Positive Contrast Density decreases Negative Contrast Density increases Refer to density chart in Carlton text (mas adjustments are not made) 34 11

12 Radiographic Density Review 85 kvp to 60 kvp A. Increased density B. Decreased density C. No change in density 100 milliseconds to 50 milliseconds 800 ma, 0.04 seconds to 600 ma, 53 ms 40 SID to 100 cm SID Table-top radiography to a 10:1 bucky grid 2.5 mm focal spot size to 1.2 mm focal spot size 90 sec film processing to 3 min film processing 200 ma to 1000 ma Normal tissue to additive pathologic condition 15 cm abdomen patient to 28 cm abdomen patient 35 Greatest Density Calculation Which of the following sets of technical factors will demonstrate the greatest density? ma time SID RSS Grid A ms :1 B ms :1 C ms :1 D ms :1 36 Greatest Density Calculation Practice Which of the following sets of technical factors will demonstrate the greatest density? mas SID RSS Grid A :1 B :1 C :1 D :

13 Digital Image Receptor Systems Density and contrast Controlled through post-processing Still, the image receptor must receive proper exposure Extremely high or low exposures do not produce acceptable image quality 38 Digital Image Receptor Systems kvp based on desired contrast Distance selected based on desired beam geometry mas is adjusted to meet exposure needs 39 Concept Point Check Part II: Matching Identify the effect on radiographic density when a single change is made. A. Radiographic density will increase B. Radiographic density will decrease C. No change in density will result 40 SID to 60 SID 400 ma 20 milliseconds to 800 ma 30 milliseconds 1.0 mm Focal Spot Size to 1.2 mm Focal Spot Size 500 RSS to 100 RSS 50 kvp to 65 kvp 5:1 ratio grid to 16:1 ratio grid 40 13

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

Energy. Mechanical Energy

Energy. Mechanical Energy Principles of Imaging Science I (RAD119) Electromagnetic Radiation Energy Definition of energy Ability to do work Physicist s definition of work Work = force x distance Force acting upon object over distance

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

Best Practices in Digital Radiography

Best Practices in Digital Radiography WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,

More information

33-10-06-03. General requirements. 1. Administrative Controls.

33-10-06-03. General requirements. 1. Administrative Controls. 33-10-06-03. General requirements. 1. Administrative Controls. a. Registrant. The registrant shall be responsible for directing the operation of the x-ray systems which have been registered with the department.

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR MAMMOGRAPHY

EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR MAMMOGRAPHY AAPM REPORT NO. 29 EQUIPMENT REQUIREMENTS AND QUALITY CONTROL FOR MAMMOGRAPHY Published for the American Association of Physicists in Medicine by the American Institute of Physics AAPM REPORT NO. 29 EQUIPMENT

More information

Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH)

Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH) Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH) By S.O ANIKOH Dept. of Physics, University of Jos, Jos. HABILA

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Waves - Transverse and Longitudinal Waves

Waves - Transverse and Longitudinal Waves Waves - Transverse and Longitudinal Waves wave may be defined as a periodic disturbance in a medium that carries energy from one point to another. ll waves require a source and a medium of propagation.

More information

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice Professional Curriculum Clinical Practice Procedural Performance Clinical Competency Basic Principles of Digital Radiography Image Acquisition Image Acquisition Errors Fundamental Principles of Exposure

More information

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY AAPM REPORT NO. 74 QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY Report of Task Group #12 Diagnostic X-ray Imaging Committee Members S. Jeff Shepard, Chairman Pei-Jan Paul Lin, Co-Chairman John M. Boone Dianna

More information

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

Digital radiography conquers the veterinary world

Digital radiography conquers the veterinary world Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing

More information

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects. RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

X-ray (Radiography), Chest

X-ray (Radiography), Chest X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012

Running head: OBESITY AND MEDICAL IMAGING 1. Obesity and Medical Imaging. November 13, 2012 Running head: OBESITY AND MEDICAL IMAGING 1 Obesity and Medical Imaging November 13, 2012 OBESITY AND MEDICAL IMAGING 2 Abstract Obesity is a growing problem in the United States, and it affects the health

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581 X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly

More information

R12-2-602. Recognized Certificate-granting Bodies R12-2-104. Approval of Radiologic Technology Schools

R12-2-602. Recognized Certificate-granting Bodies R12-2-104. Approval of Radiologic Technology Schools Arizona Administrative Code Title 12, Ch. 2 TITLE 12. NATURAL RESOURCES CHAPTER 2. RADIATION REGULATORY AGENCY MEDICAL RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS Authority: A.R.S. 32-2803 et seq. ARTICLE

More information

Digital Radiography Image Quality: Image Acquisition

Digital Radiography Image Quality: Image Acquisition Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly

More information

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,

More information

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Annual MQSA Inspection Questions NMQAAC April 19, 2004 1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0

More information

mathematics at work Health Care

mathematics at work Health Care mathematics at work Health Care Mathematics in the Health Care Sector The health care industry is one of the fastest-growing sectors of our economy, and all signs point to this trend continuing as advances

More information

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power Parameters used to completely characterize a sound wave Describing Sound Waves Chapter 3 Period Frequency Amplitude Power Intensity Speed Wave Length Period Defined as the time it take one wave vibrate

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

X-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions.

X-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions. DATA SHEET INDUSTRIAL RADIOGRAPHIC SYSTEMS RADIOGRAPHIC FILM IX20, IX25, IX50, IX80, IX, IX, IX29 AND IX59 Fujifilm industrial X-ray films feature revolutionary new film technology. The combination of

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Preadmission Information

Preadmission Information Preadmission Information British Columbia Institute of Technology 3700 Willingdon Avenue Burnaby, BC V5G 3H2 Program Head: 604 451 6918 Administrative Coordinator: 604 456 1257 medrad/web/medrad_medrad_preadmission_info

More information

Diagnostic X-ray facilities as per quality control performances in Tanzania

Diagnostic X-ray facilities as per quality control performances in Tanzania JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 4, FALL 2006 Diagnostic X-ray facilities as per quality control performances in Tanzania Yesaya Y. Sungita, 1 Simon S.L. Mdoe, 1 and Peter

More information

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come

More information

X-ray (Radiography) - Chest

X-ray (Radiography) - Chest Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL 788405/11 LAT LAT SIN DEX. kv kv. mas. mas CTL

PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT TECHNICAL MANUAL 788405/11 LAT LAT SIN DEX. kv kv. mas. mas CTL LAT mas 8 3 2 1 LAT mas PLANMED SOPHIE MAMMOGRAPHY X-RAY UNIT OBL CC OBL SIN DEX kv kv TECHNICAL MANUAL En 788405/11 TABLE OF CONTENTS Chapter A Chapter B Chapter C GENERAL & TECHNICAL DATA 1 WARNINGS...............................................

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Lecture 2 Macroscopic Interactions. 22.106 Neutron Interactions and Applications Spring 2010

Lecture 2 Macroscopic Interactions. 22.106 Neutron Interactions and Applications Spring 2010 Lecture 2 Macroscopic Interactions 22.106 Neutron Interactions and Applications Spring 2010 Objectives Macroscopic Interactions Atom Density Mean Free Path Moderation in Bulk Matter Neutron Shielding Effective

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

Integrated Management & Educational Consultancy

Integrated Management & Educational Consultancy Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:admin@imecs.freeserve.co.uk

More information

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode Swedish Laser-Medical Society The dose is the most important parameter in laser phototherapy. At a first glance, the dose seem very

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

X-ray Imaging System. X-Ray Circuit. Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry

X-ray Imaging System. X-Ray Circuit. Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry Principles of Imaging Science II (RAD 120) X-ray Imaging System Circuitry X-ray Imaging System Operating console Set x-ray tube current (quantity) and voltage (quality) Controls line compensation, kvp,

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Take away concepts. What is Energy? Solar Energy. EM Radiation. Properties of waves. Solar Radiation Emission and Absorption

Take away concepts. What is Energy? Solar Energy. EM Radiation. Properties of waves. Solar Radiation Emission and Absorption Take away concepts Solar Radiation Emission and Absorption 1. 2. 3. 4. 5. 6. Conservation of energy. Black body radiation principle Emission wavelength and temperature (Wein s Law). Radiation vs. distance

More information

X-ray (Radiography) - Abdomen

X-ray (Radiography) - Abdomen Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,

More information

Basic Radiographic Procedures

Basic Radiographic Procedures Basic Radiographic Procedures The GPT Method G. Patrick Thomas, Jr., DC, DACBR Copyright 2002 G. Patrick Thomas, Jr., DC, DACBR PO Box 1000 Blue Springs, MO 64013-1000 gptdacbr@juno.com Table of Contents

More information

RADIATION SAFETY STANDARD

RADIATION SAFETY STANDARD Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession

More information

Diagnostic x-ray imaging relies on the attenuation of

Diagnostic x-ray imaging relies on the attenuation of X-Ray Imaging Physics for Nuclear Medicine Technologists. Part 2: X-Ray Interactions and Image Formation* J. Anthony Seibert, PhD; and John M. Boone, PhD Department of Radiology, University of California

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

Digital radiography has become the norm in U.S.

Digital radiography has become the norm in U.S. Expanded content at www.asrt.org/publications. Image Gently: Using Exposure Indicators To Improve Pediatric Digital Radiography Quentin T Moore, MPH, R.T.(R)(T)(QM) Steven Don, MD Marilyn J Goske, MD,

More information

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS

SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS AERB SAFETY CODE NO. AERB/SC/MED-2 (Rev. 1) SAFETY CODE FOR MEDICAL DIAGNOSTIC X-RAY EQUIPMENT AND INSTALLATIONS Approved by the Board on October 5, 2001 This document is subject to review, after a period

More information

XRS-3 X-RAY SOURCE OPERATOR S MANUAL

XRS-3 X-RAY SOURCE OPERATOR S MANUAL XRS-3 X-RAY SOURCE OPERATOR S MANUAL CONTENTS ITEM PAGE 1.0 INTRODUCTION... 2 2.0 WARNINGS... 2 EXCLUSION ZONE DIAGRAM. 4 3.0 PHYSICAL DESCRIPTION... 5 3.1 HIGH VOLTAGE PULSER/TUBEHEAD... 5 3.2 BASE...

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY AAPM REPORT No. 4 BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY American Association of Physicists in Medicine FOREWORD This document, Basic Quality Control in Diagnostic Radiology, is the fourth in a

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

Tuesday 20 May 2014 Morning

Tuesday 20 May 2014 Morning Tuesday 20 May 2014 Morning AS GCE PHYSICS B (ADVANCING PHYSICS) G491/01 Physics in Action *1203458796* Candidates answer on the Question Paper. OCR supplied materials: Data, Formulae and Relationships

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

Lab 4: Magnetic Force on Electrons

Lab 4: Magnetic Force on Electrons Lab 4: Magnetic Force on Electrons Introduction: Forces on particles are not limited to gravity and electricity. Magnetic forces also exist. This magnetic force is known as the Lorentz force and it is

More information

AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES

AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES AAPM REPORT NO. 14 PERFORMANCE SPECIFICATIONS AND ACCEPTANCE TESTING FOR X-RAY GENERATORS AND AUTOMATIC EXPOSURE CONTROL DEVICES Published for the American Association of Physics in Medicine by the American

More information