1 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique image receptors to capture an analog image and sophisticated readers to process the image and convert the information into a digital format. Since Computed Radiography (CR) utilizes digital technology, images may be electronically manipulated to adjust contrast and density. The computer-generated images obtained with CR can be displayed electronically and/or on film. Historical Perspective Electronic imaging technologies and image management systems offer the capabilities to attain needed efficiencies. Many digital imaging technologies are presently utilized such as Computed tomography (CT) and diagnostic medical sonography, however, general radiography (a predominantly analog process), represents 60-80% of all diagnostic radiography examinations. To meet this need, Computed Radiography was introduced in 1981 by FUJIFILM at the International Congress of Radiography in Brussels and at the Radiological Society of North America in Chicago. The FCR 101 was the first available clinical system and was released in However, widespread adoption of Computed Radiography lagged behind other computer-driven modalities, primarily because digitization of analog data produced reductions in the quality and quantity of information. As such, technological improvements were needed to meet the quality standards essential for the performance of highly detailed, general radiographic examinations. Subsequently, Fuji focused on CR system improvements in key clinical areas: Improved processing speed, higher spatial resolution, reducing patient exposure, and the augmentation of data storage and transmission to provide timely retrieval and communication through the use of industry standards such as DICOM. What is a Digital Image? Digital images are described in terms of numeric values displayed in an array of rows and columns or Pixels (picture elements); this array of pixels is termed a Matrix. The pixel is the smallest component of the matrix and the quantity of information in each pixel is determined by the number of bits (binary digits) of information per pixel. Corresponding to a specific anatomical location, each pixel represents the x-ray intensity by a numeric value corresponding to a particular area on the image. Effective Date: 08/2007

2 Page: 2 of 6 Introduction to Fuji Computed Radiography (CR) Computed Radiography (CR) refers to the use of Photostimulable Storage Phosphors (PSP) for projection of X-ray imaging. In lieu of conventional screen/film, an imaging plate (IP) is used within a specially designed cassette. Imaging Plate The IP is comprised of multiple layers. The top layer is a protective layer and is a thin, transparent layer used to guard against damage caused by handling. The photostimulable phosphor layer serves as the functional layer of the IP by storing energy absorbed from the x-ray exposure. Information obtained through the initial stimulus or primary excitation (x-ray exposure) is stored in the phosphor and later read out when light or a secondary stimulus is applied to the phosphor layer. This phenomenon is called a photostimulable luminescence and makes energy storage possible within the crystal. The phosphor layer is comprised of barium fluorohalide in combination with a small amount of europium. The europium acts as an activator replacing some of the barium atoms and creating a luminescence-center in the crystal that becomes ionized by the incident radiation and is the site where energy (information) is stored as a latent image. The latent image is formed within the crystal following the transfer of energy by a photoelectric interaction. During image processing, a secondary stimulation created by a laser beam scan of the IP causes the crystal to emit blue-violet light proportionate in intensity to the x-ray energy absorbed. The electroconductive layer provides a certain amount of flexible rigidity to the IP; it also protects the phosphor layer from extraneous forces and shocks. There are additional support and light shielding layers added for additional protection and to inhibit reflection of the laser light. For some plate types, a backing layer is employed, and a barcode label is attached to identify each imaging plate. For Dual Side reading FCR devices, the shielding and backing layers are replaced with a transparent backing layer that allows for the collection of PSL from both sides of the IP simultaneously. Imaging Plates synopsis: The IP is used to capture the image It is very important to note that the IP is more sensitive to all types of radiation than conventional intensifying screen phosphors. o Direct o Scatter o Natural Background Radiation Ambient radiation can increase mottle, lower contrast, and have an affect on image quality. Any IP not used within 48 hours should be erased before use. The IP is housed within a special cassette that looks very similar to a conventional screen/film cassette. Latches Barcode Window A green stripe is on the tube-side as reference guide for image orientation.

3 Page: 3 of 6 CR Imaging Process 1. Using standard X-ray equipment, the IP (within the cassette) is exposed. The x-ray photons pass through the subject and strike the IP to form the latent image. 2. The IP is typically bar-coded at an Image and Information Processing (IIP) station to ensure proper processing. 3. The Cassette is then inserted into the CR Reading Unit and the IP is mechanically removed for processing. 4. The IP is transported through the system where it is scanned by a laser beam. The stored energy is released from the IP in the form of light, known as Photostimulable Luminescence (PSL). 5. The PSL is collected and passed through a photomultiplier, then converted to a digital signal. 6. Data recognition processes occur to meet the specified diagnostic need. The signals are reconstructed for output with a specified set of image processing parameters applied. The image can be sent to another Computer and/or Laser printer. 7. After the readout and image processing is completed, the IP is exposed to high intensity light erasing any latent image. 8. The IP is then returned to the cassette, ready to be used for another image acquisition. CR Reading System As a laser scans the IP, light is emitted where X-ray stimulated the phosphor during exposure OPTICAL SCANNER PHOTOMULTIPLIER TUBE LASER BEAM A/D Converter IMAGING PLATE Motor EDR (Exposure Data Recognizer) EDR is an automatic image adjustment function that provides optimum density and contrast even when the exposure conditions vary. A histogram is created based upon menu selection, centering/positioning, and collimation/scatter. The size and shape of the histogram will determine the appearance of the image. The proper EDR mode should be selected based upon the imaging conditions as described in the following information. AUTO Mode is the default mode and will work for the majority of the exams. The EDR mode can be selected/modified at the IIP after the menu has been selected and study started, prior to inserting the cassette into the CR Reader Unit.

4 Page: 4 of 6 EDR AUTO MODE (Centering/positioning important) Density and Contrast will be automatically applied based upon the Default processing parameters for the menu selected. AUTO Collimation recognition Complete Histogram analysis EDR SEMI MODE (Use for tightly collimated images) Centering must be precise. Density only will be automatically Applied. Fixed L value Small reading area No collimation detection done Proper kvp must be used to maintain subject contrast EDR-SEMI-X (Use when centering is not possible) Fixed L value Area of interest is Selected Same precautions as Semi-Mode Must know where the Green stripe is in relation to AOI. Helpful in cross-table exams EDR FIXED MODE The Fixed Mode will give the image a light or dark appearance base on the amount of exposure used. This is similar to screen/film system when a manual technique is used. The Fixed Mode is best used for the problem image, for example: X-Table Hips C7-T1 Laterals Any Image with a lot of Metal Hardware Images that cannot be centered properly

5 Page: 5 of 6 Fuji Computed Radiography FAQs Question: Why are we switching to Computed Radiography (CR)? Answer: CR has been shown to improve department efficiency and workflow, reduce image retakes due to exposure errors, and is a key technology enabling the adoption of PACS. Question: What is the System Speed of FCR? Answer: FCR should be considered a multi-speed system. The imaging plate employs a linear capture of exposure exceeding the capabilities of film/screen systems. FCR is not density or contrast limited but is noise limited. Question: Can I use the same techniques as before? Answer: For some exams the exposure factors will be the same. This is highly dependent on the currently deployed film/screen system speed. Question: Can I continue to use phototiming with FCR? Answer: Yes. A requirement for the use of phototiming with FCR is that calibration must be completed before using CR Cassettes and Imaging Plates. When a phototimer is used, the resulting mas can change depending upon the type of cassette and detector (due to the absorption and backscatter characteristics of the cassette/detector). As the density of the image is controlled by the FCR processing, this calibration must be completed as described in the Fuji CR Users Guide Question: Can I acquire multiple images on the same Cassette/IP? Answer: Yes*. Collimated borders are detected as part of the Exposure Data Recognizer (EDR) processing. For best results, the collimated borders should be sharp and well defined. This ensures that unnecessary information, such as scatter; outside the collimated edges will be eliminated from the analysis. This process is often referred to as PRIEF (Pattern Recognizer for Irradiated Exposure fields). See the CR Users Guide for guidelines in exposure patterns. Note* when delivering images to PACS, multiple images obtained on a single cassette will limit the individual control of image enhancements and must be taken into consideration. For that reason, many facilities that have employed PACS prefer that only one image is acquired per cassette/ip. Question: When should multiple views not be done on the same cassette? Answer: When the thickness of the anatomy varies such as an AP and Lateral. The CR Reader Unit will average both images and poor quality will result. Question: What is the significance of the green stripe on my CR cassettes? Answer: The green stripe is used as a reference for image orientation and should be placed at the cephalic/top of the image in the portrait orientation. For landscape orientation (crosswise) the green stripe should be toward the patient s right side. Incorrect image orientation should be corrected prior to sending images to PACS or the Laser Printer. Question: What is the significance of the S Value? Answer: The Sensitivity number, so-called S Value is an indicator of the Photostimulable Luminescence (PSL) given off by the imaging plate (IP) while being scanned by the laser. The values are inversely proportional to the amount of radiation that strikes the IP.

6 Page: 6 of 6 Question: What is the significance of the L value? Answer: The Latitude value, so called L value is as important as the S Value when critiquing a CR image. For most clinical imaging, L values typically range between 1.6 and 2.3. When L values are outside this range, the result can also be an abnormal S Value. Question: Is radiation exposure the only factor that affects the S Value? Answer: No. The following factors can affect the S Value: Scatter Positioning and centering. Distance SID and OFD Proper use of Grids Collimation Improper Exam selection at the IIP Failure to erase IP if it has not been used for 48 hours or longer Delay in processing from time of exposure Question: If my S number is out of the recommended range, should I change it to be within the recommended range? Answer: In most cases it is NOT recommended. The S Value should not be altered unless the image data is compromised due to an over correction of the Exposure Data Recognizer (EDR). For example, this can occur when the S Value goes below 25 and L Value is greater than 2.0. These images typically appear completely whitened out after the EDR is applied; however, the image details were present during the initial plate reading. Question: What are the differences between the NETWORK list, TODAY list, ALL list, DELIVERED list, and the LOCAL WL? Answer: Network list contains the patients to be done. This will link to the procedure ordered at the RIS/HIS system. Today list shows patients completed today. All list shows all the patients completed, including today and prior to today. Delivered list shows all the patients completed with the study delivered to any destination such as a printer and/or PACS. Local list shows patients that have been registered at the IIP with menus picked and suspended or reserved waiting for the exam to be performed.

Digital radiography conquers the veterinary world

Digital radiography conquers the veterinary world Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Digital Mammography Update: Design and Characteristics of Current Systems

Digital Mammography Update: Design and Characteristics of Current Systems : Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen

More information

Computed Radiography: Acceptance Testing and Quality Control

Computed Radiography: Acceptance Testing and Quality Control Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means

More information

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

Interfacing Fuji CR in a DICOM Network Environment

Interfacing Fuji CR in a DICOM Network Environment Interfacing Fuji CR in a DICOM Network Environment Wayne T. DeJarnette, Ph.D., Valmore Albrecht, RT, Alan Orth, RT, Jon Sayer, M.S. DeJarnette Research Systems, Inc. 401 Washington Avenue Towson, MD 21204

More information

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture Course Information: Course title: Advanced Imaging Equipment and Patient Care Course Code: RAD 184 Credits: 2.0 (30 lecture hours) Pre-requisites: RAD 276 - Imaging Equipment RAD 281 - Radiography II RAD

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer X-rays were discovered by W.C. Roentgen because of their ability to cause fluorescence. X-ray photons cannot be seen. The image produced

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Best Practices in Digital Radiography

Best Practices in Digital Radiography WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

Digital Radiography Image Quality: Image Acquisition

Digital Radiography Image Quality: Image Acquisition Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

Economic Evaluation of Conventional Radiography with Film and Computed Radiography: Applied at BMC

Economic Evaluation of Conventional Radiography with Film and Computed Radiography: Applied at BMC Advances in Computed Tomography, 2012, 1, 23-29 http://dx.doi.org/10.4236/act.2012.13006 Published Online December 2012 (http://www.scirp.org/journal/act) Economic Evaluation of Conventional Radiography

More information

Overview of Digital Detector Technology

Overview of Digital Detector Technology Overview of Digital Detector Technology J. Anthony Seibert, Ph.D. Department of Radiology University of California, Davis Disclosure Member (uncompensated) Barco-Voxar Medical Advisory Board ALARA (CR

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

Patient Prep Information

Patient Prep Information Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Guidelines on education and training for digital radiology Author:

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

MEDICAL IMAGING. 90/30 Remote Controlled RF System

MEDICAL IMAGING. 90/30 Remote Controlled RF System MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice

Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice Professional Curriculum Clinical Practice Procedural Performance Clinical Competency Basic Principles of Digital Radiography Image Acquisition Image Acquisition Errors Fundamental Principles of Exposure

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

Using Automatic Exposure Control in Digital Radiography

Using Automatic Exposure Control in Digital Radiography Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems

Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems AAPM REPORT NO. 93 Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems Report of AAPM Task Group 10 October 2006 DISCLAIMER: This publication is based on sources

More information

Checklist for the Qualification of Digital Detector Array Systems

Checklist for the Qualification of Digital Detector Array Systems Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California d.bernard@dage-group.com ABSTRACT The x-ray inspection of PCB assembly processes

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

Hardware Guide for the Classic and Elite CR Systems. 2011-08-04 PN 9G9637 Version 1.0

Hardware Guide for the Classic and Elite CR Systems. 2011-08-04 PN 9G9637 Version 1.0 Hardware Guide for the Classic and Elite CR Systems 2011-08-04 PN 9G9637 Version 1.0 Use of the Content Carestream Health CR Systems are designed to meet international safety and performance standards.

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays J. E. Wilhjelm Ørsted TU Technical University of enmark, Bldg. 348, K-2800 Kongens Lyngby, enmark. X-ray and Computed Tomography Contents History and characterization of X-rays Conventional (projection)

More information

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

Integrated Management & Educational Consultancy

Integrated Management & Educational Consultancy Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:admin@imecs.freeserve.co.uk

More information

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Comparison of Amorphous Selenium and Amorphous Silicon Flat-Panel Detectors November 20, 2012 COMPARISON OF FLAT-PANEL DETECTORS 2 Abstract

More information

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.

More information

Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging

Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging Digital Radiography in Dentistry: Moving from Film-based to Digital Imaging Gail F. Williamson, RDH, MS Continuing Education Units: 4 hours Online Course: www.dentalcare.com/en-us/dental-education/continuing-education/ce350/ce350.aspx

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

EPSON SCANNING TIPS AND TROUBLESHOOTING GUIDE Epson Perfection 3170 Scanner

EPSON SCANNING TIPS AND TROUBLESHOOTING GUIDE Epson Perfection 3170 Scanner EPSON SCANNING TIPS AND TROUBLESHOOTING GUIDE Epson Perfection 3170 Scanner SELECT A SUITABLE RESOLUTION The best scanning resolution depends on the purpose of the scan. When you specify a high resolution,

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film. INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

REGIUS History. Touch REGIUS110. REGIUS170 ImagePilot REGIUS150 REGIUS330 CS-1

REGIUS History. Touch REGIUS110. REGIUS170 ImagePilot REGIUS150 REGIUS330 CS-1 Konica Minolta has been a CR technology leader since 1996, consistently designing, developing, and delivering innovative solutions. The REGIUS Σ represents the sum of this experience. REGIUS Σ delivers

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

X-ray (Radiography) - Abdomen

X-ray (Radiography) - Abdomen Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

GE Healthcare. DoseWatch. Gathering Radiation Dose Data GE Healthcare DoseWatch Gathering Radiation Dose Data Gathering Radiation Dose Data After your organization crosses the bridge of Why? for patient radiation dose management either due to regulation, risk

More information

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Application Note Case Study Technology Primer White Paper X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Image Sensors X-Ray DETECTORS Scanners Image Processing Custom Solutions In all domains

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

Preadmission Information

Preadmission Information Preadmission Information British Columbia Institute of Technology 3700 Willingdon Avenue Burnaby, BC V5G 3H2 Program Head: 604 451 6918 Administrative Coordinator: 604 456 1257 medrad/web/medrad_medrad_preadmission_info

More information

Cargo Container X-ray Inspection Systems

Cargo Container X-ray Inspection Systems Hitachi Review Vol. 53 (2004), No. 2 97 Cargo Container X-ray Inspection Systems OVERVIEW: Container inspection is categorized into three stages. In the initial stage, large-size X-ray DR (digital radiography)

More information

GAMMA-RAY SPECTRA REFERENCES

GAMMA-RAY SPECTRA REFERENCES GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.

More information

Physics 441/2: Transmission Electron Microscope

Physics 441/2: Transmission Electron Microscope Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This

More information

portable x-ray survey report

portable x-ray survey report department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

Introduction to CCDs and CCD Data Calibration

Introduction to CCDs and CCD Data Calibration Introduction to CCDs and CCD Data Calibration Dr. William Welsh San Diego State University CCD: charge coupled devices integrated circuit silicon chips that can record optical (and X-ray) light pixel =

More information

Non-Destructive Testing

Non-Destructive Testing Non-Destructive Testing Minimize Product Line Downtime and Product Loss with Radiography Testing for Sorting and Containment Authored by Mike Forbes, Na onal Sales Manager TUV Rheinland Industrial Solu

More information

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY

ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information