1 Introduction. Fast Gradient Echo Sequences Including Balanced SSFP Brian Hargreaves Department of Radiology, Stanford University

Size: px
Start display at page:

Download "1 Introduction. Fast Gradient Echo Sequences Including Balanced SSFP Brian Hargreaves Department of Radiology, Stanford University"

Transcription

1 Fast Gradient Echo Sequences Including Balanced SSFP Brian Hargreaves Department of Radiology, Stanford University 1 Introduction Many common magnetic resonance imaging sequences are classified either as spin echo or gradient echo techniques. Generally, a gradient echo occurs when spins are coherent regardless of their position within the sample. This occurs whenever the net gradient area since a previous echo is zero. In a spin echo, spins are coherent regardless of their resonance frequency [1]. Here we will focus on gradient echoes in more detail, including how they evolve, and the so-called gradient-echo pulse sequences. These include gradient spoiled sequences, spoiled sequences and balanced steady-state free precession (SSFP) sequences. 1.1 General Rapid Gradient Echo Sequences Rapid gradient echo sequences consist of a single radio frequency () excitation, imaging gradients and acquisition, and spoiler gradients. Figure 1 shows a basic example of the and spoiling gradients. The pulse has some flip angle α and some phase angle, φ. For simplicity, we omit imaging and slice-select gradients, and assume that the spoiler gradients are along only one axis. i i+1 i time Four basic gradient-echo sequences can be derived from Fig. 1. Balanced SSFP (steady state free precession) sequences have no spoiler gradients. Gradientspoiled sequences have a spoiler at the end of the repetition (Fig. 1, solid gradient.). Reversed gradient-spoiled sequences have a spoiler gradient immediately after the pulse, before imaging. (Fig. 1, dotted line). Finally, -spoiled sequences use a gradient spoiler at the end of the repetition and an additional quadratic phase increment for φ. By first examining the dynamics of balanced SSFP, we can build up the other three sequences. 1.2 Spin Dynamics The dynamics of spins in an MRI system with radio frequency () and gradient coils are accurately described by the Bloch equation. Given and gradient waveforms, as well as the resonance frequency, relaxation times and position of a spin, the (3x1) magnetization vector can be easily modeled using a matrix formalism [2], where an pulse is represented by a series of rotation matrices about an arbitrary axis in the transverse (M x M y ) plane. Precession, due to the resonance frequency as well as the effect of gradients is also represented by a series of rotation matrices about the longitudinal (z) axis. Relaxation is represented as a matrix multiplication, with an addition of a recovery term. Using a time step that is small compared with the rate of any of these processes, these three matrices can be applied sequentially and still provide a very good model of the spin dynamics. 1.3 Magnetization Steady States Gradient time When the same sequence of rotations, precession and relaxation is repeated, a steady state forms, where the magnetization at some point in the sequence is the same from one repetition to the next. There are three cases: Figure 1: Gradient echo sequences generally consist of repeated pulses, and some type of spoiling gradients. The flip angle (α) is usually constant, while the phase angle (φ) can be constant, or can increment linearly or quadratically with the repetition number. Spoiling gradients usually occur at the end of the repetition (solid lines), but may also occur at the beginning (dotted lines), or not at all. Three time points of interest, numbered 1-3, are used in analyzing the signal levels. T R > 2T 1 and T R > 2T 2. Here the magnetization at the start of the sequence relaxes completely to the equilibrium magnetization, M. The steady state is just the equilibrium magnetization. T R > 2T 2 but T R < 2T 1. Here all transverse magnetization will be zero at the start of the sequence. However, the longitudinal magnetization reaches a steady state. 1

2 (a) M z M x (c) Magnitude.2.1!4!2 2 4 Precession per ( ) (b) M y M x Phase ( ) 2 1!1!2!4!2 2 4 Precession per ( ) Figure 2: Magnetization dynamics in balanced SSFP. (a) and (b) show the path of magnetization in the M x M z and M x M y planes, while (c) shows the signal (M xy -magnetization) magnitude and phase as a function of resonance frequency. Colored lines represent single resonance frequencies. The, and + symbols represent magnetization at time points 1, 2 and 3 in Fig. 1. Immediately after the pulse, steady-state magnetization is distributed along an ellipse ( symbols and dotted black line in a,b). Spins precession over T R (colored lines in b), aligning along the M y axis at T R/2 ( symbols). Immediately before the pulse, magnetization is again distributed along an ellipse (+ symbols and dashed black line). T R < 2T 1 and T R < 2T 2. Both longitudinal and transverse magnetization reach a non-zero steady state. We will concentrate on the third of these cases, which applies to most rapid gradient echo sequences. 2 Balanced SSFP Dynamics Balanced SSFP consists simply of pulses that are repeated every T R, with no spoiling gradients [3]. The dynamics of other rapid gradient echo sequences can be derived from those of balanced SSFP. First we assume a basic sequence with the following dynamics: rotation of an angle α about the y axis. Precession θ due to resonance frequency offset. T 1 and T 2 relaxation. After numerous repetitions, the magnetization reaches a steady state, that can be calculated analytically by propagating the above dynamics through a sequence repetition [4 7]. However, it is important to try to intuitively understand the dynamics [8 1]. For moderately large α, there are two cases of interest (shown in Fig. 2): 1. When the amount of precession is very small, (θ close to ), the magnetization mostly rotates around the y axis. A steady state forms the effects of relaxation and rotation perfectly cancel, which, for large α requires the magnetization length to be very small (Fig 2, red line). 2. When precession is significant, it almost completely cancels nutation. The magnetization follows the path(s) shown in Fig. 2. The signal is refocused midway through the sequence along either the positive or negative y-axis, depending on the direction and amount of precession. The magnetization length is such that the direction of relaxation is orthogonal to the magnetization vec- 2

3 tor. This means that as the magnetization vector direction approaches the M x M y plane, the vector length becomes smaller. The motion in the steady state is a function of several parameters. First, the flip angle, α, and phase, φ, determine the direction and amount of rotation. The resonance frequency and repetition time, T R, give the amount of precession between pulses. The relaxation times, T 1 and T 2, together with T R determine the amount of magnetization decay and recovery. The signal magnitude varies periodically with resonant precession with a period of 36. Immediately before, and immediately after the pulse, the transverse magnetization component is confined to a half plane. The signal is also refocused by the pulse [11], resulting in piecewise constant phase midway between pulses. 2.1 Phase Cycling If the and receiver phase are incremented by φ on each repetition, the magnetization dynamics of Fig. 2 will be altered. A constant phase increment can be viewed as a shift in the center frequency of φ/t R [4, 12]. The result is that the signal profile in Fig. 2c is shifted along the precession axis by φ, and an additional phase of φ/2 is added to the signal phase. Magnitude Phase ( ) !25!2!15!1! !1 Precession per ( ) =9!2!25!2!15!1! Precession per ( ) = =18 =9 = =18 Figure 3: Balanced SSFP signal magnitude and phase as a function of resonant precession ( ) for φ =, φ = 18, and φ = 9. The magnitude and phase profiles shift along the precession axis by φ, and the phase changes by φ/ Frequency Modulation The and receiver phase can also be modulated using quadratic phase increments, where φ is increasing by a constant factor on each repetition. This results in the center frequency shifting with time, so the signal profile of Fig. 2 slowly shifts along the horizontal axis. If the quadratic phase increment is small, there is little distortion to the signal profile, and spins can track the frequency change [13]. For larger phase increments, the profile becomes somewhat distorted. Either way, spins will form a periodic steady state with a period equal to the time for the frequency shift to equal 1/T R. 3 Pulse Sequences The previous section described the steady-state that forms when a periodic sequence of rotations, precession, and decay are applied. We now describe several different pulse sequences using these concepts to understand the signal formation with each sequence. In all cases, we use the x, y and z axes as the readout, phaseencode and slice-select axes respectively. 3.1 Balanced SSFP Balanced SSFP sequences (TrueFISP, FIESTA, Balanced-FFE, True SSFP) have the signal characteristics described already. Normally, the pulse sign alternates, which is equivalent to a phase increment of 18. This shifts the signal profile so that on-resonant spins produce high signal (Fig. 3, dashed line). Figure 4 shows a balanced SSFP pulse sequence. Note that all gradient waveforms are balanced, that is they have zero net area over a full repetition. Balanced SSFP sequences give the highest signal level of the rapid gradient echo sequences [14, 15]. The signal shape is highly sensitive to resonance frequency so minimizing precession, and thus T R, is important for avoiding banding artifacts or signal variations across images [16]. However, the minimum time required to achieve adequate spatial resolution or to limit power absorption limit the minimum T R that can be achieved. Furthermore, susceptibility variations make balanced SSFP more difficult at higher field strengths. 3.2 Gradient-Spoiled Echo The gradient-spoiled sequence, (FE, GRASS, FISP, FAST), uses a spoiler gradient at the end of the sequence 3

4 Balanced SSFP Sequence Gradient Spoiled Sequence Figure 4: Balanced SSFP pulse sequence. All gradient waveforms are fully rewound, or balanced. The pulse sign usually alternates, so that a high signal is produced for on-resonant spins. The signal echo location is shown on the waveform. Figure 5: Gradient spoiled pulse sequence. In this example, a spoiler gradient is included on both the readout (x) and sliceselect (z) axes. The signal echo location is shown on the waveform. as shown in Fig. 5 [7,17]. Each spin now sees a positiondependent rotation in addition to that due to free precession. Note, however that the steady state for each spin at time points 1 and 3 (Fig. 1) is the same as that for balanced SSFP, when the gradient-induced rotation is included in the precession per T R. The spoiler gradient area is chosen to induce rotations that range over at least 36 within each voxel, and it assumed that there is a uniform distribution of precession values within a voxel. Thus the signal immediately after the pulse is the average transverse component of all precession values, i.e. the average of the balanced SSFP signal on the dotted ellipse in Fig 2b. Between the pulse and the echo time, some dephasing of the signal due to frequency variations can occur. This results in T 2 signal loss. There are many variations of gradient-spoiled echo sequences, depending on the direction and size of the spoiler gradient. However, if the assumption can be made that the unbalanced gradient induces numerous rotations within a single voxel, the signal is similar for all variations. 3.3 Reversed Gradient-Spoiled Echo The reversed gradient-spoiled echo sequence (CE-FAST, SSFP, T2-FFE and PSIF) instead uses a gradient spoiler at the beginning of the sequence, as shown in Fig. 6 [17]. As with the gradient-spoiled sequence, the steady state magnetization for any single spin is the same as for balanced SSFP when the precession due to the gradient spoiler is included. However, imaging occurs after the spoiler gradient, at time point 3 in Fig. 1. The signal is the average transverse component of the balanced SSFP magnetization, or the average signal on the dashed ellipse in Fig 2b. Again, T2 signal loss occurs, but should be calculated in reverse from time point 3 to the point of the imaging echo. Reversed Gradient Spoiled Sequence Figure 6: Reversed gradient spoiled pulse sequence. This is precisely the reverse of the gradient spoiled sequence. Note that it is important to rephase the readout gradient and defocus the slice-select gradient. 3.4 Fast Acquisition Double Echo (FADE) The fast acquisition double echo (FADE) sequence acquires both the gradient-spoiled echo and the reversed gradient-spoiled echo in a single repetition [18]. This is achieved by placing the gradient spoiling in the middle of 4

5 the sequence as shown in Fig. 7. The echo characteristics are unchanged from the descriptions in 3.2 and 3.3. Fast Acquisition Double Echo (FADE) Sequence Figure 7: FADE pulse sequence, which acquires both the gradient-spoiled echo and the reversed gradient-spoiled echo (Shown on waveform). A spoiler gradient is included on both the readout (x) and slice-select (z) axes, midway between pulses. 3.5 Spoiling Gradient spoiled, reversed gradient spoiled and balanced SSFP sequences all have a signal with T 2 /T 1 contrast. The T 2 dependence can be virtually eliminated by spoiling [19 21]. spoiling quadratically increments the phase of the pulse with the result that residual transverse magnetization before the pulse is not refocused and can be neglected. The sequence looks exactly like the gradient-spoiled sequence (Fig. 5 The signal in spoiled sequences (SPGR, FLASH, T1-FFE) can be calculated by numerical simulation, but is well approximated by simply neglecting residual transverse magnetization prior to each pulse (case 2 described in 1.3). 4 Imaging Considerations In this section, we briefly discuss the signal considerations with the rapid gradient echo pulse sequences, including relative signals and contrast of different sequences. 4.1 s and Contrast The contrast behavior of rapid gradient-echo sequences is highly dependent on flip angle. T R is usually kept as short as possible to minimize scan time or to reduce banding artifacts in balanced SSFP. Figure 8 shows the flipangle dependence of four sequences: balanced SSFP with and without alternating the pulse sign, gradient spoiled sequences, and spoiled sequences. For certain T 1 and T 2 times, each sequence has an optimal flip angle that maximizes SNR. An interesting point to note is that at the Ernst angle, cos 1 (e T R/T 1 ), which maximizes signal for -spoiled sequences, the signal is identical for all sequences. [5]. The flip angle is usually chosen to maximize a combination of SNR as well as contrast between the tissues of interest. (a) (b) Blood Muscle Flip Angle ( ) Flip Angle ( ) b!ssfp (,!,,!,...) Gradient Spoiled b!ssfp (,,,,...) Spoiled Figure 8: as a function of flip angle for (a) blood and (b) muscle. Shown are the signals from balanced SSFP ( Hz, alternating sign), gradient spoiled, balanced SSFP ( Hz, non-alternating sign), and spoiled sequences. The Ernst angle, cos 1 (e T R/T 1 ), signal for all sequences, and the peak signal for spoiled sequences [5] Figure 9 compares the signal in balanced SSFP, gradient-spoiled, reversed gradient-spoiled and spoiled sequences with identical T R and flip angle. Four tissues are compared: blood, liver, fat and muscle. The balanced SSFP signal is very sensitive to flip angle, and the shape also depends on the T 1 /T 2 ratio. At low T 1 /T 2, the signal is M-shaped (fat, blood). At higher T 1 /T 2 (muscle, liver) or higher flip angles (not shown) the signal shape is more smooth. Balanced SSFP gives the highest signal. Gradient-spoiled and reversed gradientspoiled sequences give similar signal levels, the latter being more T 2 -weighted. Finally, -spoiled sequences are purely T 1 -weighted, and give the lowest signal of these four sequences. 5

6 (a) Blood (b) Liver 5 Summary (c) Fat (d) Muscle Figure 9: Comparison of signals in different tissue with balanced SSFP, gradient spoiled, reversed gradient spoiled, and spoiled sequences. (a) Blood, T 1 /T 2 = 1/2 ms. (b) Liver, T 1 /T 2 = 5/43 ms. (c) Fat, T 1 /T 2 = 27/85 ms. (d) Muscle, T 1 /T 2 = 85/47 ms. Other sequence parameters are T R=4 ms, flip angle=3. T E= ms for gradient spoiled and spoiled, T E=2 ms for balanced SSFP, and T E=T R for reversed gradient echo. Note that the balanced SSFP signal is highest, but also is sensitive to frequency variations. The reversed gradient spoiled sequence produces more T 2 weighting than the gradient spoiled sequence. Finally, spoiled sequences produce the lowest signal, but it is purely T 1 -weighted to-noise Ratio The signal-to-noise ratio (SNR) is one of the most important considerations in selecting imaging parameters for MRI. SNR is proportional to voxel size, the square root of the total acquisition time, and the signal level for the given sequence and parameters. The sequences shown here typically achieve good SNR by using a short T R so that the signal is averaged over many repetitions, often with 3D phase-encoding. However, as T R is shortened, a greater proportion of time is taken by pulses and preparatory gradients such as phase-encoding, dephasing and rephasing. This means that the acquisition duty cycle or the portion of T R spent acquiring data is critical. Maximizing SNR becomes a trade-off between keeping T R short to minimize T2 loss or banding artifacts (in balanced SSFP) and keeping the readout duty cycle long. Rapid gradient echo sequences with very short repetition times are commonly used in MRI. The signal behavior of these sequences depends strongly on the residual magnetization prior to each excitation, which can be altered by using combinations of gradient spoiling and spoiling. After many sequence repetitions, the magnetization reaches a steady-state, where it varies periodically. The steady state calculation for the fully-balanced SSFP sequence can be used to understand the signal formation in gradient spoiled and spoiled sequences. Imaging parameters must then be chosen to optimize SNR and contrast for specific applications. References [1] Hahn EL. Spin echoes. Phys Rev 195; 8: [2] Jaynes E. Matrix treatment of nuclear induction. Phys Rev 1955; 98: [3] Carr HY. Steady-state free precession in nuclear magnetic resonance. Phys Rev 1958; 112: [4] Zur Y, et al. An analysis of fast imaging sequences with steady-state transverse magnetization refocusing. Magn Reson Med 1988; 6: [5] Buxton RB, et al. intensity in fast NMR imaging with short repetition times. J Magn Reson 1989; 83: [6] van der Meulen P, et al. Fast field echo imaging: An overview and contrast calculations. Magn Reson Imaging 1988; 6: [7] Sekihara K. Steady-state magnetizations in rapid NMR imaging using small flip angles and short repetition intervals. IEEE Trans Med Imaging 1987; 6: [8] Scheffler K. A pictorial description of steady-states in rapid magnetic resonance imaging. Concepts Magn Reson 1999; 11: [9] Dharmakumar R, Wright GA. Understanding steadystate free precession: A geometric perspective. Concepts Magn Reson 25; 26A:1 1. [1] Schmitt P, et al. A simple geometrical description of the TrueFISP ideal transient and steady-state signal. Magn Reson Med 26; 55. [11] Scheffler K, Hennig J. Is TrueFISP a gradient-echo or a spin-echo sequence? Magn Reson Med 23; 49: [12] Hinshaw WS. Image formation by nuclear magnetic 6

7 resonance: the sensitive-point method. J Appl Phys 1976; 47: [13] Foxall DL. Frequency-modulated steady-state free precession imaging. Magn Reson Med 22; 48: [14] Oppelt A, et al. FISP a new fast MRI sequence. Electromedica 1986; 54: [15] Duerk JL, et al. Remember true FISP? a high SNR near 1-second imaging method for T 2 -like contrast in interventional MRI at.2 T. J Magn Reson Imaging 1998; 8: [16] Freeman R, Hill HDW. Phase and intensity anomalies in Fourier transform NMR. J Magn Reson 1971; 4: [17] Gyngell ML. The application of steady-state free precession in rapid 2DFT NMR imaging: FAST and CE-FAST sequences. Magn Reson Imaging 1988; 6: [18] Redpath TW, Jones RA. FADE - a new fast imaging sequence. Magn Reson Med 1988; 6: [19] Zur Y, et al. Spoiling of transverse magnetization in steady-state sequences. Magn Reson Med 1991; 21: [2] Frahm J, et al. Transverse coherence in rapid FLASH NMR imaging. J Magn Reson 1987; 72: [21] Haase A. Snapshot FLASH MRI. applications to T 1, T 2, and chemical-shift imaging. Magn Reson Med 199; 13:

MRI SEQUENCES. 1 Gradient Echo Sequence

MRI SEQUENCES. 1 Gradient Echo Sequence 5 An MRI sequence is an ordered combination of RF and gradient pulses designed to acquire the data to form the image. In this chapter I will describe the basic gradient echo, spin echo and inversion recovery

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Basic Principles of Magnetic Resonance

Basic Principles of Magnetic Resonance Basic Principles of Magnetic Resonance Contents: Jorge Jovicich jovicich@mit.edu I) Historical Background II) An MR experiment - Overview - Can we scan the subject? - The subject goes into the magnet -

More information

NMR Techniques Applied to Mineral Oil, Water, and Ethanol

NMR Techniques Applied to Mineral Oil, Water, and Ethanol NMR Techniques Applied to Mineral Oil, Water, and Ethanol L. Bianchini and L. Coffey Physics Department, Brandeis University, MA, 02453 (Dated: February 24, 2010) Using a TeachSpin PS1-A pulsed NMR device,

More information

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

Chapter 2 NMR in Inhomogeneous Fields

Chapter 2 NMR in Inhomogeneous Fields Chapter 2 NMR in Inhomogeneous Fields Federico Casanova and Juan Perlo 2.1 Introduction Besides the redesigning of the hardware to excite and detect NMR signals from sample volumes external to the sensor,

More information

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi Diffusione e perfusione in risonanza magnetica E. Pagani, M. Filippi DW-MRI DIFFUSION-WEIGHTED MRI Principles Diffusion results from a microspic random motion known as Brownian motion THE RANDOM WALK How

More information

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 2016 Electronic Supplementary Information Achieving High Resolution and Controlling

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance Author: James Dragan Lab Partner: Stefan Evans Physics Department, Stony Brook University, Stony Brook, NY 794. (Dated: December 5, 23) We study the principles behind Nuclear

More information

MRI Physics for Radiologists

MRI Physics for Radiologists Alfred L. Horowitz MRI Physics for Radiologists A Visual Approach Second Edition With 94 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest Alfred

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium

Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium Xia Lee and Albert Tsai June 15, 2006 1 1 Introduction Nuclear magnetic resonance (NMR) is a spectroscopic

More information

Generation and Detection of NMR Signals

Generation and Detection of NMR Signals Generation and Detection of NMR Signals Hanudatta S. Atreya NMR Research Centre Indian Institute of Science NMR Spectroscopy Spin (I)=1/2h B 0 Energy 0 = B 0 Classical picture (B 0 ) Quantum Mechanical

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance Spectroscopy Introduction NMR is the most powerful tool available for organic structure determination. It is used to study a wide variety of nuclei: 1 H 13 C 15 N 19 F 31 P 2

More information

Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy

Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy Journal of Magnetic Resonance 164 (2003) 321 328 www.elsevier.com/locate/jmr Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy Nikolaus M. Loening, a, * Michael J. Thrippleton,

More information

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory 1 Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory Foreword Pulsed nuclear magnetic resonance is fascinating in its own right, and is also an incredibly important tool for

More information

When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid.

When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid. Fluid Statics When the fluid velocity is zero, called the hydrostatic condition, the pressure variation is due only to the weight of the fluid. Consider a small wedge of fluid at rest of size Δx, Δz, Δs

More information

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference. Pulsed Fourier Transform NR The rotating frame of reference The NR Eperiment. The Rotating Frame of Reference. When we perform a NR eperiment we disturb the equilibrium state of the sstem and then monitor

More information

Kap 8 Image quality, signal, contrast and noise

Kap 8 Image quality, signal, contrast and noise 4/5/ FYS-KJM 474 contrast SNR MR-teori og medisinsk diagnostikk Kap 8 Image qualit, signal, contrast and noise resolution vailable MRparameters speed Main source of noise in MRI: Noise generated within

More information

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents Preface Symbols and fundamental constants 1. What is spectroscopy? A semiclassical description of spectroscopy Damped harmonics Quantum oscillators The spectroscopic experiment Ensembles and coherence

More information

The Hydrogen Atom Is a Magnet. http://www.seed.slb.com/en/scictr/watch/gashydrates/detecting.htm

The Hydrogen Atom Is a Magnet. http://www.seed.slb.com/en/scictr/watch/gashydrates/detecting.htm The Hydrogen Atom Is a Magnet Nuclear Magnetic Resonance Spectroscopy (NMR) Proton NMR A hydrogen nucleus can be viewed as a proton, which can be viewed as a spinning charge. As with any spinning charge,

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance Practical Course M I. Physikalisches Institut Universität zu Köln May 15, 2014 Abstract Nuclear magnetic resonance (NMR) techniques are widely used in physics, chemistry, and

More information

Lecture L3 - Vectors, Matrices and Coordinate Transformations

Lecture L3 - Vectors, Matrices and Coordinate Transformations S. Widnall 16.07 Dynamics Fall 2009 Lecture notes based on J. Peraire Version 2.0 Lecture L3 - Vectors, Matrices and Coordinate Transformations By using vectors and defining appropriate operations between

More information

Nuclear Magnetic Resonance and Its Application in Condensed Matter Physics

Nuclear Magnetic Resonance and Its Application in Condensed Matter Physics Nuclear Magnetic Resonance and Its Application in Condensed Matter Physics Kangbo Hao 1. Introduction Nuclear Magnetic Resonance (NMR) is a physics phenomenon first observed by Isidor Rabi in 1938. [1]

More information

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging V. Prof. Dr.

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging V. Prof. Dr. 1 12/9/2008 Page 1 Master s Program in Medical Physics Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging V Chair in Faculty of Medicine Mannheim University of Heidelberg Theodor-Kutzer-Ufer

More information

runl I IUI%I/\L Magnetic Resonance Imaging

runl I IUI%I/\L Magnetic Resonance Imaging runl I IUI%I/\L Magnetic Resonance Imaging SECOND EDITION Scott A. HuetteS Brain Imaging and Analysis Center, Duke University Allen W. Song Brain Imaging and Analysis Center, Duke University Gregory McCarthy

More information

Rotation: Moment of Inertia and Torque

Rotation: Moment of Inertia and Torque Rotation: Moment of Inertia and Torque Every time we push a door open or tighten a bolt using a wrench, we apply a force that results in a rotational motion about a fixed axis. Through experience we learn

More information

32-Channel Head Coil Imaging at 3T

32-Channel Head Coil Imaging at 3T 32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

More information

GE Medical Systems Training in Partnership. Module 12: Spin Echo

GE Medical Systems Training in Partnership. Module 12: Spin Echo Module : Spin Echo Spin Echo Objectives Review the SE PSD. Review the concepts of T, T, and T*. Spin Echo PSD RF Gz Gy 90 80 Gx Spin Echo - SE Spin echo is a standard pulse sequence on Signa MRi/LX and

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine G16.4426/EL5823/BE6203 Medical Imaging MRI Instrumentation, Data Acquisition, Image Reconstruction Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department

More information

A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS

A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The

More information

FDA Guidelines for Magnetic Resonance Equipment Safety

FDA Guidelines for Magnetic Resonance Equipment Safety FDA Guidelines for Magnetic Resonance Equipment Safety Loren A. Zaremba, Ph.D. Center for Devices and Radiological Health Food and Drug Administration Outline I. Introduction II. Static Magnetic Field

More information

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law. 260 17-1 I. THEORY EXPERIMENT 17 QUALITATIVE STUDY OF INDUCED EMF Along the extended central axis of a bar magnet, the magnetic field vector B r, on the side nearer the North pole, points away from this

More information

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters Overview Optimizing MRI Protocols Clinical Practice & Compromises Geoffrey D. Clarke, Radiology Department University of Texas Health Science Center at San Antonio Pulse timing parameters for adjusting

More information

The MRI Study Guide for Technologists

The MRI Study Guide for Technologists The MRI Study Guide for Technologists Kenneth S. Meacham The MRI Study Guide for Technologists With 51 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest

More information

Lecture - 4 Diode Rectifier Circuits

Lecture - 4 Diode Rectifier Circuits Basic Electronics (Module 1 Semiconductor Diodes) Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati Lecture - 4 Diode Rectifier Circuits

More information

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a BASIC PRINCIPLES INTRODUCTION TO NUCLEAR MAGNETIC RESONANCE (NMR) 1. The nuclei of certain atoms with odd atomic number, and/or odd mass behave as spinning charges. The nucleus is the center of positive

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Applied Spectroscopy Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Banwell and McCash Chapter 7 Skoog, Holler Nieman Chapter 19 Atkins, Chapter 18 Relaxation processes We need

More information

An Adaptive Signal Processing Approach to Dynamic Magnetic Resonance Imaging

An Adaptive Signal Processing Approach to Dynamic Magnetic Resonance Imaging An Adaptive Signal Processing Approach to Dynamic Magnetic Resonance Imaging A Thesis Presented by William Scott Hoge to The Department of Electrical and Computer Engineering in partial fulfillment of

More information

Drive circuit basics + V. τ e. Industrial Circuits Application Note. Winding resistance and inductance

Drive circuit basics + V. τ e. Industrial Circuits Application Note. Winding resistance and inductance ndustrial Circuits Application Note Drive circuit basics For a given size of a stepper motor, a limited space is available for the windings. n the process of optimizing a stepper motor drive system, an

More information

Algebra 2 Chapter 1 Vocabulary. identity - A statement that equates two equivalent expressions.

Algebra 2 Chapter 1 Vocabulary. identity - A statement that equates two equivalent expressions. Chapter 1 Vocabulary identity - A statement that equates two equivalent expressions. verbal model- A word equation that represents a real-life problem. algebraic expression - An expression with variables.

More information

Candidate Number. General Certificate of Education Advanced Level Examination June 2014

Candidate Number. General Certificate of Education Advanced Level Examination June 2014 entre Number andidate Number Surname Other Names andidate Signature General ertificate of Education dvanced Level Examination June 214 Physics PHY4/1 Unit 4 Fields and Further Mechanics Section Wednesday

More information

Subjects: Fourteen Princeton undergraduate and graduate students were recruited to

Subjects: Fourteen Princeton undergraduate and graduate students were recruited to Supplementary Methods Subjects: Fourteen Princeton undergraduate and graduate students were recruited to participate in the study, including 9 females and 5 males. The mean age was 21.4 years, with standard

More information

POWER SYSTEM HARMONICS. A Reference Guide to Causes, Effects and Corrective Measures AN ALLEN-BRADLEY SERIES OF ISSUES AND ANSWERS

POWER SYSTEM HARMONICS. A Reference Guide to Causes, Effects and Corrective Measures AN ALLEN-BRADLEY SERIES OF ISSUES AND ANSWERS A Reference Guide to Causes, Effects and Corrective Measures AN ALLEN-BRADLEY SERIES OF ISSUES AND ANSWERS By: Robert G. Ellis, P. Eng., Rockwell Automation Medium Voltage Business CONTENTS INTRODUCTION...

More information

How To Understand The Measurement Process

How To Understand The Measurement Process April 24, 2015 Exam #3: Solution Key online now! Graded exams by Monday! Final Exam Monday, May 4 th, 10:30 a.m. Room: Perkins 107 1 A Classical Perspective A classical view will help us understand the

More information

Supporting Information

Supporting Information S1 Supporting Information GFT NMR, a New Approach to Rapidly Obtain Precise High Dimensional NMR Spectral Information Seho Kim and Thomas Szyperski * Department of Chemistry, University at Buffalo, The

More information

Relaxation Can T, Be Longer Than T,?

Relaxation Can T, Be Longer Than T,? Concepts in Magnetic Resonance, 1991, 3, 171-177 Relaxation Can T, Be Longer Than T,? Daniel D. Traficante Departments of Chemistry and Medicinal Chemistry and NMR Concepts UniVersily of Rhode Island Kingston,

More information

Super-Resolution Reconstruction in MRI: Better Images Faster?

Super-Resolution Reconstruction in MRI: Better Images Faster? Super-Resolution Reconstruction in MRI: Better Images Faster? Esben Plenge 1,, Dirk H. J. Poot 1, Monique Bernsen 2, Gyula Kotek 2, Gavin Houston 2, Piotr Wielopolski 2, Louise van der Weerd 3, Wiro J.

More information

Lecture #7 (2D NMR) Utility of Resonance Assignments

Lecture #7 (2D NMR) Utility of Resonance Assignments Lecture #7 (2D NMR) Basics of multidimensional NMR (2D NMR) 2D NOESY, COSY and TOCSY 2/23/15 Utility of Resonance Assignments Resonance Assignments: Assignment of frequency positions of resonances (peaks)

More information

Shaft. Application of full spectrum to rotating machinery diagnostics. Centerlines. Paul Goldman, Ph.D. and Agnes Muszynska, Ph.D.

Shaft. Application of full spectrum to rotating machinery diagnostics. Centerlines. Paul Goldman, Ph.D. and Agnes Muszynska, Ph.D. Shaft Centerlines Application of full spectrum to rotating machinery diagnostics Benefits of full spectrum plots Before we answer these questions, we d like to start with the following observation: The

More information

Nuclear Magnetic Resonance (NMR) Spectroscopy

Nuclear Magnetic Resonance (NMR) Spectroscopy April 28, 2016 Exam #3: Graded exams on Tuesday! Final Exam Tuesday, May 10 th, 10:30 a.m. Room: Votey 207 (tentative) Review Session: Sunday, May 8 th, 4 pm, Kalkin 325 (tentative) Office Hours Next week:

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging What are the uses of MRI? To begin, not only are there a variety of scanning methodologies available, but there are also a variety of MRI methodologies available which provide

More information

Primer on Magentic Resonance Imaging Tomography c 2004

Primer on Magentic Resonance Imaging Tomography c 2004 Primer on Magentic Resonance Imaging Tomography c 2004 Timothy Chupp and Scott Swanson University of Michigan Departments of Physics, Biomedical Engineering, and Radiology March 1, 2004 Contents 1 Introduction

More information

4 Coherence Selection: Phase Cycling and Gradient Pulses

4 Coherence Selection: Phase Cycling and Gradient Pulses 4 Coherence Selection: Phase Cycling and Gradient Pulses,QWURGXFWLRQ A multiple-pulse NMR experiment is designed to manipulate the spins in a certain carefully defined way so as to produce a particular

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications Safety Issue 3.0 T MRI Pro 3.0 T MRI Con Immediate fringe field surrounding magnet A ferromagnetic object inadvertently brought into the scan room will experience a sharp increase in attraction toward

More information

Incorporating Internal Gradient and Restricted Diffusion Effects in Nuclear Magnetic Resonance Log Interpretation

Incorporating Internal Gradient and Restricted Diffusion Effects in Nuclear Magnetic Resonance Log Interpretation The Open-Access Journal for the Basic Principles of Diffusion Theory, Experiment and Application Incorporating Internal Gradient and Restricted Diffusion Effects in Nuclear Magnetic Resonance Log Interpretation

More information

LONI De-Identification Policy

LONI De-Identification Policy The following defines how different file formats are de-identified with LONI tools. Each metadata attribute of each file can have the following operations performed: Operation keep Remove Replace Description

More information

Spin-lattice and spin-spin relaxation

Spin-lattice and spin-spin relaxation Spin-lattice and spin-spin relaation Sequence of events in the NMR eperiment: (i) application of a 90 pulse alters the population ratios, and creates transverse magnetic field components (M () ); (ii)

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3 Practical Aspects of Quantitative NMR Experiments This discussion presumes that you already have an understanding of the basic theory of NMR. There are a number of issues that should be considered when

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Command-induced Tracking Jitter Study I D. Clark November 24, 2009

Command-induced Tracking Jitter Study I D. Clark November 24, 2009 Command-induced Tracking Jitter Study I D. Clark November 24, 2009 Introduction Reports of excessive tracking jitter on the MMT elevation axis have lately been theorized to be caused by the input command

More information

Flip-angle-optimized fast dynamic T 1 mapping with a 3D gradient-echo sequence

Flip-angle-optimized fast dynamic T 1 mapping with a 3D gradient-echo sequence Flip-angle-optimized fast dynamic T 1 mapping with a 3D gradient-echo sequence Olaf Dietrich 1, Maximilian Freiermuth 1, Linus Willerding 2, Maximilian F. Reiser 1, Michael Peller 1 1 Josef Lissner Laboratory

More information

Switch Mode Power Supply Topologies

Switch Mode Power Supply Topologies Switch Mode Power Supply Topologies The Buck Converter 2008 Microchip Technology Incorporated. All Rights Reserved. WebSeminar Title Slide 1 Welcome to this Web seminar on Switch Mode Power Supply Topologies.

More information

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9 Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9 Dick J. Drost a) Nuclear Medicine and MRI Department, St. Joseph s Health Centre, London, Ontario N6A 4L6, Canada William

More information

NMR Pulse Spectrometer PS 15. experimental manual

NMR Pulse Spectrometer PS 15. experimental manual NMR Pulse Spectrometer PS 15 experimental manual NMR Pulse Spectrometer PS 15 Experimental Manual for MS Windows For: MS Windows software Winner Format: MS Word 2002 File: PS15 Experimental Manual 1.5.1.doc

More information

TWO-DIMENSIONAL TRANSFORMATION

TWO-DIMENSIONAL TRANSFORMATION CHAPTER 2 TWO-DIMENSIONAL TRANSFORMATION 2.1 Introduction As stated earlier, Computer Aided Design consists of three components, namely, Design (Geometric Modeling), Analysis (FEA, etc), and Visualization

More information

A look at the utility of pulsed NMR

A look at the utility of pulsed NMR 1 A look at the utility of pulsed NMR Katherine Magat and Vasudev Mandyam Physics 173, Spring 2004, Prof. Kleinfeld Introduction Pulsed nuclear magnetic resonance (NMR) was first introduced in the 1940

More information

Let s first see how precession works in quantitative detail. The system is illustrated below: ...

Let s first see how precession works in quantitative detail. The system is illustrated below: ... lecture 20 Topics: Precession of tops Nutation Vectors in the body frame The free symmetric top in the body frame Euler s equations The free symmetric top ala Euler s The tennis racket theorem As you know,

More information

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1 WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's

More information

Magnetic Field Modeling of Halbach Permanent Magnet Array

Magnetic Field Modeling of Halbach Permanent Magnet Array Magnetic Field Modeling of Halbach Permanent Magnet Array Shengguo Zhang *1, Kai Wang 1, Xiaoping Dang 2 School of Electrical Engineering, Northwest University for Nationalities, Lanzhou, China School

More information

A GENERALIZED FINITE SIZED DIPOLE MODEL FOR RADAR AND MEDICAL IMAGING PART II: NEAR FIELD FORMULATION FOR MAGNETIC RESONANCE IMAGING

A GENERALIZED FINITE SIZED DIPOLE MODEL FOR RADAR AND MEDICAL IMAGING PART II: NEAR FIELD FORMULATION FOR MAGNETIC RESONANCE IMAGING Progress In Electromagnetics Research, PIER 24, 227 256, 1999 A GENERALIZED FINITE SIZED DIPOLE MODEL FOR RADAR AND MEDICAL IMAGING PART II: NEAR FIELD FORMULATION FOR MAGNETIC RESONANCE IMAGING T. S.

More information

DIRECT CURRENT GENERATORS

DIRECT CURRENT GENERATORS DIRECT CURRENT GENERATORS Revision 12:50 14 Nov 05 INTRODUCTION A generator is a machine that converts mechanical energy into electrical energy by using the principle of magnetic induction. This principle

More information

Physics 112 Homework 5 (solutions) (2004 Fall) Solutions to Homework Questions 5

Physics 112 Homework 5 (solutions) (2004 Fall) Solutions to Homework Questions 5 Solutions to Homework Questions 5 Chapt19, Problem-2: (a) Find the direction of the force on a proton (a positively charged particle) moving through the magnetic fields in Figure P19.2, as shown. (b) Repeat

More information

- thus, the total number of atoms per second that absorb a photon is

- thus, the total number of atoms per second that absorb a photon is Stimulated Emission of Radiation - stimulated emission is referring to the emission of radiation (a photon) from one quantum system at its transition frequency induced by the presence of other photons

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance NMR is probably the most useful and powerful technique for identifying and characterizing organic compounds. Felix Bloch and Edward Mills Purcell were awarded the 1952 Nobel

More information

Spin-Lattice Relaxation Times

Spin-Lattice Relaxation Times Spin-Lattice Relaxation Times Reading Assignment: T. D. W. Claridge, High Resolution NMR Techniques in Organic Chemistry, Chapter 2; E. Breitmaier, W. Voelter, Carbon 13 NMR Spectroscopy,3rd Ed., 3.3.2.

More information

RF Measurements Using a Modular Digitizer

RF Measurements Using a Modular Digitizer RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.

More information

Elasticity. I. What is Elasticity?

Elasticity. I. What is Elasticity? Elasticity I. What is Elasticity? The purpose of this section is to develop some general rules about elasticity, which may them be applied to the four different specific types of elasticity discussed in

More information

Direct and Reflected: Understanding the Truth with Y-S 3

Direct and Reflected: Understanding the Truth with Y-S 3 Direct and Reflected: Understanding the Truth with Y-S 3 -Speaker System Design Guide- December 2008 2008 Yamaha Corporation 1 Introduction Y-S 3 is a speaker system design software application. It is

More information

Basic Electrical Technology Dr. L. Umanand Department of Electrical Engineering Indian Institute of Science, Bangalore. Lecture - 33 3 phase System 4

Basic Electrical Technology Dr. L. Umanand Department of Electrical Engineering Indian Institute of Science, Bangalore. Lecture - 33 3 phase System 4 Basic Electrical Technology Dr. L. Umanand Department of Electrical Engineering Indian Institute of Science, Bangalore Lecture - 33 3 phase System 4 Hello everybody. So, in the last class we have been

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

Magnetic Fields and Their Effects

Magnetic Fields and Their Effects Name Date Time to Complete h m Partner Course/ Section / Grade Magnetic Fields and Their Effects This experiment is intended to give you some hands-on experience with the effects of, and in some cases

More information

BASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011

BASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011 AM 5-202 BASIC ELECTRONICS AC CIRCUIT ANALYSIS December 2011 DISTRIBUTION RESTRICTION: Approved for Pubic Release. Distribution is unlimited. DEPARTMENT OF THE ARMY MILITARY AUXILIARY RADIO SYSTEM FORT

More information

High Spatial Resolution EPI Using an Odd Number of Interleaves

High Spatial Resolution EPI Using an Odd Number of Interleaves Magnetic Resonance in Medicine 41:1199 1205 (1999) High Spatial Resolution EPI Using an Odd Number of Interleaves Michael H. Buonocore* and David C. Zhu Ghost artifacts in echoplanar imaging (EPI) arise

More information

Module 3: Floyd, Digital Fundamental

Module 3: Floyd, Digital Fundamental Module 3: Lecturer : Yongsheng Gao Room : Tech - 3.25 Email : yongsheng.gao@griffith.edu.au Structure : 6 lectures 1 Tutorial Assessment: 1 Laboratory (5%) 1 Test (20%) Textbook : Floyd, Digital Fundamental

More information

Eðlisfræði 2, vor 2007

Eðlisfræði 2, vor 2007 [ Assignment View ] [ Pri Eðlisfræði 2, vor 2007 28. Sources of Magnetic Field Assignment is due at 2:00am on Wednesday, March 7, 2007 Credit for problems submitted late will decrease to 0% after the deadline

More information

Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging

Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging Magnetic Resonance Imaging 18 (2000) 479 484 Technical Note Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging J. Baudewig a,b, W. Paulus

More information

Rotation Matrices and Homogeneous Transformations

Rotation Matrices and Homogeneous Transformations Rotation Matrices and Homogeneous Transformations A coordinate frame in an n-dimensional space is defined by n mutually orthogonal unit vectors. In particular, for a two-dimensional (2D) space, i.e., n

More information

Power Quality Paper #3

Power Quality Paper #3 The Effect of Voltage Dips On Induction Motors by: M D McCulloch 1. INTRODUCTION Voltage depressions caused by faults on the system affect the performance of induction motors, in terms of the production

More information

NMR - Basic principles

NMR - Basic principles NMR - Basic principles Subatomic particles like electrons, protons and neutrons are associated with spin - a fundamental property like charge or mass. In the case of nuclei with even number of protons

More information