SPECIFICATION FOR HALT-HASS TEST CHAMBER

Size: px
Start display at page:

Download "SPECIFICATION FOR HALT-HASS TEST CHAMBER"

Transcription

1 SPECIFICATION FOR HALT-HASS TEST CHAMBER Sl: No. Description / Spec Required by LRDE Quantity Specification offered by Vendor Remarks / Deviation To be entered in Price Bid only Unit Price Total Price 1. Dimensions Internal : 1350 mm 1350 mm 1200 mm (W x D x H) 2. Chamber Volume Approx 2000 litres 3. Interior 4. Exterior 5. Insulation 6. Doors and viewing windows The Interior of chamber shall be constructed with stainless steel for assuring rustproof and need to be sealed to minimise moisture migration in to the chamber. The chamber shall be of structural steel frame work and 16 gauge sheet metal panels. Removable access panels to be provided where appropriate for servicing the components and subsystems of the chamber. All walls and the chamber doors to be insulated with a preferably rigid fibreglass board and staggered layers of fibreglass insulation to provide optimum isolation between chamber interior and exterior. System shall have split-open type doors on front with two numbers of viewing window of size 500 x 500 mm on each door 7. Port holes 2 Nos. Of 100mm 300mm with sealing plugs 8. Circulation motors Circulation motors with sealed bearing (no greasing) 9. Power Supply System shall work with V, 3 Ø, 50 Hz power supply

2 Vibration system 1. Dimension of Vibration Table (W D) 2. Number of Actuators Min Control Frequency Range 5 Hz Hz 4. Direction and type of excitation 6 DOF Broad Band excitation 5 Test load capacity Minimum 400 Kg inclusive of fixture 6. Min Vibration range amplitude 7. Vibration table position >50g rms (as an average of X, Y, Z axis) System shall have two different positions to enable testing for small & large specimens 8. Inserts on the vibration table Threaded M10 SS inserts at 100mm grid 9. Fixture Kit 10. Compressor 11. Pressure Regulators Temperature system with LN2 Fixture Kit containing an assortment of aluminium extrusions and connectors, threads rods and quick threading split nuts. A suitable capacity of air compressor to operate vibration system Shall be supplied. Precision Electronic (Digital) pressure regulators for supply and control of compressed air to vibration table actuators 1. System temperature range -100ºC to 200ºC (with LN2 cooling) 2. Ramp Up/Down rate 70ºC/min average (with LN2 cooling) 3. Temperature stability ±1ºC (Steady State) 4. Control accuracy ± 0.1ºC to 1ºC

3 Portable LN2 tanks and control and operational accessories Manifold LN2 leakage 8. Removal of the gas from the pipeline Four numbers of portable 250 lts capacity of LN2 cylinders and control and operational accessories shall be supplied and installed. SS Vacuum jacketed piping along with necessary couplers shall be supplied and installed between the LN2 tanks/cylinders and the system. No flanges shall be used in the pipelines System shall be provided with vacuum jackets (VJ) manifold with redundant vacuum jackets valves with ½ female bayonet connector. The male bayonet connection from the tank supply shall be fixed with proper welding. LN2 tanks and pipeline installation shall be done properly to eliminate LN2 leakage either at the tank or in the pipeline. Keepful / Keep cool devices shall be fixed in the LN2 pipeline to remove the gas if any from the pipeline to ensure LN2 in liquid form is made available to the system 9. LN2 Pressure LN2 pressure shall be maintained at / not more than 50 PSI in the LN2 tanks and LN2 pipeline Temperature system with refrigeration(cooling system) HALT-HASS chamber shall also be designed to perform with mechanical refrigeration system. This should be of separate unit, shall be connected to HALT-HASS chamber through suitable low thermal loss/insulated flexible ducts/hose (length of hose shall be maximum 5 metres) and quick attachable/removable clamping mechanism. This system should also be usable as an independent temperature cyclic chamber. 1. Internal dimension 1 mtr x 1 mtr x 1 mtr 2. Interior The Interior of chamber shall be constructed with stainless steel for assuring rustproof and withstanding floor load up to 150 Kgs. 3. Exterior The chamber shall be of structural steel frame work and 16 gauge sheet metal panels. Removable access panels to be provided where appropriate for servicing the components and subsystems of the chamber. Unit shall be preferably fitted with heavy duty caster/nylon wheels and locking mechanism. 4. Temperature range -60ºC to 125ºC, ±1ºC 5. Temperature gradient 5ºC/min linear from -40ºC to 85ºC in the empty chamber 6. Refrigeration system Cascaded refrigeration system with ozone friendly refrigerants.

4 7. Chiller Suitable capacity of chillier shall be provided to cool the condenser of 1 st stage compressor. This shall be part of the cooling system. 8. Port holes 2 No s. Of dia 100mm with sealing plugs 9. Controller Programmable temperature control unit with touch screen LCD colour monitor and relative operating software. 10. Power Supply System shall work with V, 3 Ø, 50 Hz power supply Controller and operating software System Controller Suitable controller for the HALT-HASS chamber with 15 LCD colour monitor with multi stage password protection shall be provided. 4 channel input module with low pass filter, signal conditioner and 4 Nos. Of suitable accelerometers with mounting blocks and 12 Nos. Accelerometers cables of 5 meters length shall be provided 16 channel temperature monitoring device with input card, operating software and 16 numbers suitable thermocouples of 2 metre long, connectors and sockets shall be provided. Software should have the features like: a) Various alarm / Interlocks, Set Points, Heater control status b) Graphical display of % of heat, cooling, & vibration levels c) Test Profile, number of loops remaining, current sequence number and elapsed time. d) Auto tune facility to aid in the tuning of PID parameters of the chamber. e) Doors Lock/open automatic control feature Computer of HP 8000, Intel core i5 650, 3.2 GHz, 4 MB L3 Cache, Intel Q57 or better on OEM Mother Board, 8 GB RAM DDR MHz, 1 TB 7200 rmp serial ATA HDD, DVD writer, 29 LED display, Logitech Multimedia Keyboard and Optical scroll mouse, Six USB Ports, one parallel Port, Ethernet Connection, Preloaded Windows 7 Professional and MS Office 7 with media and documentation and certificate of authenticity, McAfee Antivirus with media and documentation (latest version included in case of windows 7) shall be provided (Mother Board drivers soft copy for Windows 7 & XP Professional).

5 Safety and interlocks Electrical circuit protection Interlocks Component control feature High temperature limiter Vibration level limiter Pressure differential switch Door closure and safety features Safety air solenoid LN2 feed control and safety features Air purge kit An electrical circuit protection against burned contacts or a faulty instrument to be de-energized if any normal safety device opens The circulators and heaters to be interlocked. If the circulator switch is off, or if a limit is tripped, the heating and cooling systems shall become inoperative. Provision shall be made limit the components at an ambient temperature in order to continue the temperature stress, testing beyond the point. Non-adjustable high temperature limits switch to de-energise chamber heaters if chamber exceeds maximum design operating temperature. System shall be equipped with a user selectable Min/Max vibration level safety device. Should protect the heaters in the event of extremely large amount of heat due to disrupted airflow in the chamber. Pressure switch should monitor airflow and shut it down if it detects a problem. System shall have pneumatic door closures for ease with built in interlocks with a finger pinch sensor that will prevent door closing if anybody gets finger trapped in between the doors by accident shall be provided. Additional safety, a plunger mechanism shall be provided inside the chamber when pressed should activate the pneumatic control to release the door open A safety air solenoid valve shall be provided in the system to cut off compressor air supply to vibration table actuators in case of vibration run over A proportional valve and safety valve shall be provided in the system for LN2 feed control As an additional safety, a pressure leak valve shall also be provided in the system to bleed LN2 in case of excessive LN2 pressure occur. System shall be supplied with air purge kit to ensure complete evacuation of nitrogen before doors are opened.

6 Oxygen sensor General Wall mount type Oxygen sensor of CE certified with LCD display shall be provided Installation and acceptance Training Operation and maintenance manual Warranty The complete HALT-HASS system with all its accessories, vibration table, LN2 piping, cooling system and controller shall be installed at LRDE, Bangalore and shall be demonstrated for all its functional parameters. System will be accepted by LRDE after satisfactory completion of installation, demonstration and necessary operation and maintenance training for LRDE engineers and staffs. Detailed operating and maintenance training shall be given to LRDE engineers and staffs for 5 days at LRDE with free of cost. Two numbers hardcopy and a soft copy of detailed Operation and maintenance manual shall be provided. 36 months (In case of repair during warranty all charges for parts and labour including freight charge if any are to be borne by supplier) from the date of installation and acceptance at LRDE, Bangalore. Delivery period 6 months from receipt of purchase order DELIVERABLES Sl: No. DESCRIPTION QUANTITY 1. Complete HALT HASS system with vibration system, accessories including four numbers 1 set of portable 250 litres LN2 filled cylinders, 2. Cooling system with controller and connecting ducts 1 set 3. LN2 pipes and accessories 1 lot 4. Air compressor 1 No. 5. System calibration kit consisting of vibration meters, accelerometers and thermocouples 1 set 6. Spares as listed 1 set 7. Computer with preloaded software 1 No. 8. Fixture Kit 1 set 8 Calibration and FAT reports 1 set 9. Installation, operating and maintenance manual, drawings and schematic diagrams 2 sets & 1 soft copy

7 SPARES (Please Quote as optional) Sl: No. DESCRIPTION QUANTITY 1. Circuit breakers used in control panel 2 Nos. Each of various capacities used 2. Fuses (if any used in system) 2 Nos. Each type 3. PLC, control cards 1 No. each 3. Solid state relays and contactors 2 Nos. Each type 4. Control relays 2 Nos. Each 4. Limit switches 2 Nos. 5. Actuators 2 Nos. 6. Timers 2 Nos. 7. Thermocouples used in system 12 Nos. 8 Accelometers 4 Nos. each type/size 8. Solenoid valves 6 Nos. 9. Pressure /oil cut outs 2 Nos. Each 10. Controller 1 No. Each 11 Refrigeration compressors 1 No. each type 12 Expansion valves 1 each 13 Any other critical parts lot The vendor should indicate point-to-point compliance for the full specification. Only original equipment manufacturers or authorised representatives with certificate from OEM should be quoted. The vendor should furnish the details of similar item supplied earlier with evidence. Note: 1. The above format to be submitted in duplicate after duly filling but price information to be given in only one copy. 2. The copy of above format without price information should be enclosed to techno commercial bid 3. The copy with price information must be enclosed to price bid only 4. Disclosure of price information in techno commercial bid will lead to rejection of the quote VR.PRAKASH BABU SCIENTIST D

TECHNICAL SPECIFICATION FOR PROJECTION WELDING MACHINE (Qty. 2 Nos.)

TECHNICAL SPECIFICATION FOR PROJECTION WELDING MACHINE (Qty. 2 Nos.) TECHNICAL SPECIFICATION FOR PROJECTION WELDING MACHINE (Qty. 2 Nos.) 1. Requirement: Two (02) Nos. DC Type, 3 phase, 3 wire Projection Welding Machine is required for welding of ventilation M.S. / Stainless

More information

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR)

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) I. Demonstrate safety skills in typical HVACR work situations to NATE Core Installer Knowledge Areas for Technician Excellence for Safety

More information

BB-18 Black Body High Vacuum System Technical Description

BB-18 Black Body High Vacuum System Technical Description BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The

More information

Material test chamber with mechanical convection

Material test chamber with mechanical convection FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and

More information

HERZ-Thermal Actuators

HERZ-Thermal Actuators HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01

More information

SELECTION, APPLICATION AND MAINTENANCE

SELECTION, APPLICATION AND MAINTENANCE DIESEL PROTECTION SYSTEMS Automatic Diesel Engine Shut Down System for Safe Area Applications SELECTION, APPLICATION AND MAINTENANCE Series 300 Series 310 SYSTEM DESCRIPTION Suitable for attended engine

More information

SYLLABUS For BASIC TRADE COURSE (360 Hours) On REFRIGERATION AND AIR CONDITIONING

SYLLABUS For BASIC TRADE COURSE (360 Hours) On REFRIGERATION AND AIR CONDITIONING SYLLABUS For BASIC TRADE COURSE (360 Hours) On REFRIGERATION AND AIR CONDITIONING INDEX 1. Introduction 3 2. Objectives 3 3. Employment Opportunity 4 4. Course Structure 4 5. Entry Qualification 4 6. List

More information

Why and How we Use Capacity Control

Why and How we Use Capacity Control Why and How we Use Capacity Control On refrigeration and air conditioning applications where the load may vary over a wide range, due to lighting, occupancy, product loading, ambient weather variations,

More information

National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook

National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook Trainee:... Training Provider:... Year:... Trainee Contact no:... Competency Based Skill Training Logbook Introduction

More information

Cold Room & Refrigeration System (Casserole Assembly) No of units: 1

Cold Room & Refrigeration System (Casserole Assembly) No of units: 1 Technical Specifications Cold Room & Refrigeration System (Casserole Assembly) No of units: 1 For chiller storage of all sorts of items Chiller storage temperature: 2-5 C Provides excellent thermal efficiency

More information

MAINTENANCE INSTRUCTIONS. Thermia Robust heat pump

MAINTENANCE INSTRUCTIONS. Thermia Robust heat pump MAINTENANCE INSTRUCTIONS Thermia Robust heat pump 9 6 8 0-5 4 7 4 5 0 0 1 R e v. 3 Table of contents 1 Important information.................. 2 1.1 Product description....................... 2 1.2 General................................

More information

543-0032-00, 943-0032-00. User s Manual

543-0032-00, 943-0032-00. User s Manual 543-0032-00, 943-0032-00 User s Manual 1 Comfort Alert Diagnostics Faster Service And Improved Accuracy The Comfort Alert diagnostics module is a breakthrough innovation for troubleshooting heat pump and

More information

CoolTop. A unique water cooled air-conditioning unit for server room cooling from the top AC-TOPx-CW-240/60

CoolTop. A unique water cooled air-conditioning unit for server room cooling from the top AC-TOPx-CW-240/60 CoolTop A unique water cooled air-conditioning unit for server room cooling from the top AC-TOPx-CW-240/60 Application CoolTop is a new, unique air conditioning unit especially designed for datacenters.

More information

Medical Air Systems: Scroll

Medical Air Systems: Scroll GENERAL The Chemetron skid mount scroll medical air system is designed to provide medical breathing air for hospital and medical institutions. This system meets NFPA 99 requirements for Level 1 breathing

More information

heat treatment equipment & accessories

heat treatment equipment & accessories heat treatment equipment & accessories Dear Customer, ElectroHeat Sweden AB is an innovative company based in Gothenburg, and has been operating since 1975. We manufacture industrial furnaces and heat

More information

FOXAIR 60/50 ALL MODELS

FOXAIR 60/50 ALL MODELS FOXAIR 60/50 ALL MODELS OPERATION AND SERVICE MANUAL UNIT WEIGHT - ALL MODELS APPROX. 300 POUNDS FOXTRONICS, INC. 38 L x 28 W x 24 H Love Field-Dallas TX MADE IN THE USA 3448 West Mockingbird Lane Dallas,

More information

10 tips for getting more reliability from your air-conditioning system [reliability]

10 tips for getting more reliability from your air-conditioning system [reliability] 10 tips for getting more reliability from your air-conditioning system [reliability] A guide for installers, specifiers, contractors and architects February 2010 10 tips for getting more reliability from

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

KNC Model 3666 Automatic Pressure Calibration System

KNC Model 3666 Automatic Pressure Calibration System KNC Model 3666 Automatic Pressure Calibration System Product Description The Model 3666 Automatic Pressure Calibration (APC) System by King Nutronics Corporation is a portable secondary standards laboratory

More information

SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT

SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT PART 1 - GENERAL 1.1 DESCRIPTION A. The Work specified in this Section consists of designing, furnishing and installing factory-assembled

More information

Computer Controlled Capacity Control Methods in Refrigeration, with SCADA and PID Control THARA2C/1

Computer Controlled Capacity Control Methods in Refrigeration, with SCADA and PID Control THARA2C/1 Technical Teaching Equipment Computer Controlled Capacity Control Methods in Refrigeration, with SCADA and PID Control THARA2C/1 EDIBON SCADA System and PID CONTROL included Teaching Technique used 5 6

More information

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS

Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Technical Information sheet 0001/07/Rev01 ENG FUNCTION Series S001 and S002 circulation units are applied to the primary circuit of solar

More information

Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832)

Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Z-Plus Series Temperature/Humidity Chambers Reliability You Expect Features You Want Value You Need Z-Plus chambers

More information

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System

TECHNICAL SPECIFICATION. StirLIN-1 Economy Liquid Nitrogen System TECHNICAL SPECIFICATION StirLIN-1 Economy Liquid Nitrogen System Reference 80 8120_01 Issue Date August 1, 2015 1. INTRODUCTION Since more than sixty years Stirling Cryogenics has designed and manufactured

More information

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de

LABOKLAV. Autoclaves Made in Germany. www.shp-steriltechnik.de LABOKLAV Autoclaves Made in Germany www.shp-steriltechnik.de 2 LABOKLAV STEAM STERILIZERS Steam sterilization is the sterilization method with the widest range of applications efficient, safe, reproducible

More information

SPECIFICATION TECHNICAL AIR-CONDITIONING UNITS SINGLE AND SPLIT (DUCTED) (AIR COOLED PACKAGED UNIT)

SPECIFICATION TECHNICAL AIR-CONDITIONING UNITS SINGLE AND SPLIT (DUCTED) (AIR COOLED PACKAGED UNIT) SPECIFICATION TECHNICAL AIR-CONDITIONING UNITS SINGLE AND SPLIT (DUCTED) (AIR COOLED PACKAGED UNIT) MAINTENANCE TYPE: STATUTORY 1.0 SCOPE OF SERVICES The Contractor shall provide maintenance Services for

More information

Failure code manual. content

Failure code manual. content Failure code manual content 一 wall split AC series 2 二 floor standing AC series. 4 三 portable AC series.. 5 四 dehumidifer 6 五 DC inverter single split series...7 六 DC inverter multi-split series 10 1 一

More information

Conlift1, Conlift2, Conlift2 ph+

Conlift1, Conlift2, Conlift2 ph+ GRUNDFOS DATA BOOKLET Conlift1, Conlift2, Conlift2 ph+ Small lifting stations 50 Hz CONLIFT1, CONLIFT2, CONLIFT2 ph+ Table of contents 1. Product overview 3 Conlift for condensate applications 3 Applications

More information

AIR-COOLED CHILLERS APPLIED SYSTEMS. www.daikin.eu EUWA*5-24KAZW - EUWY*5-24KAZW -

AIR-COOLED CHILLERS APPLIED SYSTEMS. www.daikin.eu EUWA*5-24KAZW - EUWY*5-24KAZW - AIR-COOLED CHILLERS APPLIED SYSTEMS www.daikin.eu Cooling Only Heat pump EUWA*5-24KAZW - EUWY*5-24KAZW - Environmental Awareness Daikin and the Environment In recent years, motivated by a global awareness

More information

Electronic Flow Switches

Electronic Flow Switches Industries Electronic Flow Switches Petrochemical Refining Oil Production Water Treatment Pharmaceutical Food and Beverage Pulp and Paper Power Production Gas Processing Mining Biotechnology Semiconductor

More information

Visions become real. Temperature control and cooling systems. The innovative and reliable solution

Visions become real. Temperature control and cooling systems. The innovative and reliable solution Temperature control and cooling systems The innovative and reliable solution With our PulseTemp equipment you have the option of up to 72 cooling circuits in the impulse cooling, temperature control, temperature

More information

VDL 53 - Vacuum drying oven for flammable solvents

VDL 53 - Vacuum drying oven for flammable solvents Vacuum drying ovens VDL 53 - Vacuum drying oven for flammable solvents The VDL series ensures maximum safety when drying organic solvents in accordance with TÜV/GS guidelines. The interior chamber is designed

More information

Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier

Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier Procurement Specification for a High Security Lift Arm Barrier (HSLAB) EB950 CR Armstrong High Security Lift Arm Barrier A. Requirement This document is to be used to specify the physical and operational

More information

FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS

FVA THIRD-GENERATION VERTICAL SATURATED-STEAM STERILIZERS The key difference between a small vertical or bench-top sterilizer and industrial sterilizer is performance. Small saturated-steam sterilizers with vertical or horizontal chambers were devised many years

More information

Freezers - Ultrafreezers. Platilab & Irilab Series

Freezers - Ultrafreezers. Platilab & Irilab Series Freezers - Ultrafreezers Platilab & Irilab Series PLATILAB: top security and performances with reliable temperature control. IRILAB: combining the high reliability of PLATILAB with the possibility to store

More information

Vacuum drying oven for non-flammable solvents

Vacuum drying oven for non-flammable solvents VD series 53 Vacuum drying ovens Vacuum drying oven for non-flammable solvents A BINDER vacuum drying oven of the VD series is impressive while gently drying with its homogeneous temperature distribution.

More information

The PACE5000 Modular Pressure Controller/Indicator

The PACE5000 Modular Pressure Controller/Indicator GE Sensing The PACE5000 Modular Pressure Controller/Indicator First in a new generation of modular, high precision Druck pressure controllers/indicators, designed for ATE and test bench applications. The

More information

Split-type Air-Conditioner INSTALLATION MANUAL CONTENTS FOR INSTALLER MXZ-3A30NA MXZ-4A36NA ATTENTION. English. Français. Español

Split-type Air-Conditioner INSTALLATION MANUAL CONTENTS FOR INSTALLER MXZ-3A30NA MXZ-4A36NA ATTENTION. English. Français. Español Split-type Air-Conditioner MXZ-3A30NA MXZ-4A36NA INSTALLATION MANUAL Refer to the installation manual of each indoor unit for indoor unit installation. English Français Español ATTENTION This manual mentions

More information

data sheet DONATELLO Rev. I emiss - 25/10/2013 52PRO03 code Automatic tissue processor DONATELLO

data sheet DONATELLO Rev. I emiss - 25/10/2013 52PRO03 code Automatic tissue processor DONATELLO Rev. I emiss - /0/0 DONATELLO Automatic floor closed cycle processor for histological specimens Automatic tissue processor DONATELLO code PRO0 General characteristics DONATELLO is a closed cycle processor

More information

Table V. Troubleshooting Checklist for Refrigeration Systems. Air or non-condensable gas in system. Inlet water warm.

Table V. Troubleshooting Checklist for Refrigeration Systems. Air or non-condensable gas in system. Inlet water warm. Table V Troubleshooting Checklist for Refrigeration Systems TROUBLE POSSIBLE CAUSE CORRECTIVE MEASURE High condensing pressure. Low condensing pressure. Air or non-condensable gas in system. Inlet water

More information

STERILIZERS, LABORATORY DRYING OVENS

STERILIZERS, LABORATORY DRYING OVENS TS9026 TS9053 TS9135 TS9430 TS 9000 SERIES: STERILIZERS, LABORATORY DRYING OVENS Series TS9000 consists of four different cabinets available in sizes from 26 litres to 430 litres. Interior housing and

More information

Technical data. Danfoss DHP-A

Technical data. Danfoss DHP-A Technical data Danfoss DHP-A An air heat pump which produces both heat and hot water Can operate efficiently down to -0 C Danfoss TWS tank gives plenty of hot water quickly and with low operating costs

More information

Condensing unit type EPSILON-ECHOS LE Technical information manual

Condensing unit type EPSILON-ECHOS LE Technical information manual Condensing unit type EPSILON-ECHOS LE Technical information manual 2 > EPSILON ECHOS Water chiller > EPSILON ECHOS /HP Reversible heat pump > EPSILON ECHOS /ST Water chiller with storage tank and pumps

More information

REFRIGERATED TYPE COMPRESSED AIR DRYERS INSTRUCTION MANUAL. For Sales & Service please contact:

REFRIGERATED TYPE COMPRESSED AIR DRYERS INSTRUCTION MANUAL. For Sales & Service please contact: 7610.478.16 8/99 INSTRUCTION MANUAL Models 8005, 8010, 8015 For Sales & Service please contact: CENTRAIR Air Systems & Supplies Phone: 705-722-5747 Fax: 705-722-5458 Email: sales@centrair.ca Website: www.centrair.ca

More information

Test Code: 8297 / Version 1

Test Code: 8297 / Version 1 Blueprint HVAC Maintenance Technology PA Test Code: 8297 / Version 1 Copyright 2014. All Rights Reserved. General Assessment Information HVAC Maintenance Technology PA Blueprint Contents General Assessment

More information

Air Conditioner Water Heater - A Product of HotSpot Energy LLC

Air Conditioner Water Heater - A Product of HotSpot Energy LLC Air Conditioner Water Heater - A Product of HotSpot Energy LLC PLEASE READ THIS BEFORE YOU INSTALL THE UNIT 1. This air conditioner must be installed and/or repaired by a qualified technician. If you perform

More information

VTS³ and VCS³ Stress Screening Systems. More Power for Maximum Quality Assurance

VTS³ and VCS³ Stress Screening Systems. More Power for Maximum Quality Assurance VTS³ and VCS³ Stress Screening Systems More Power for Maximum Quality Assurance Tailor-made for your product: Our proven testing methods Thermal and climatic conditions can affect the production, storage

More information

WARDEN 4 ELECTRONIC WATER MANAGEMENT SYSTEM FEATURING:

WARDEN 4 ELECTRONIC WATER MANAGEMENT SYSTEM FEATURING: WARDEN 4 ELECTRONIC WATER MANAGEMENT SYSTEM FEATURING: New Zealand made multi purpose plumbing control for Prison cells Vandal resistant controls Water saving fittings DESCRIPTION Warden 4 is a multi purpose

More information

PRODUCT RELEASE CKMTA12 / 24. Copyright 2011 by ARB Corporation Limited

PRODUCT RELEASE CKMTA12 / 24. Copyright 2011 by ARB Corporation Limited PRODUCT RELEASE CKMTA12 / 24 Copyright 2011 by ARB Corporation Limited ARB 4x4 ACCESSORIES 720 SW 34Th Street Tel: (425) 264-1391 Renton, Washington 98057 Fax: (425) 264-1392 USA North and South American

More information

ECOCIAT. Domestic hot water heat recovery unit

ECOCIAT. Domestic hot water heat recovery unit Heat recovery unit Domestic hot water High energy efficiency with R410A Compact and quiet Scroll compressors Brazed-plate heat exchangers Heating Heat recovery ENVIRONMENTALLY HFC R410A PROTECTION DE FRIENDLY

More information

Product Data. Split System Cooling 22-1904-1F-EN. March 2015

Product Data. Split System Cooling 22-1904-1F-EN. March 2015 Product Data Split System Cooling 4TTR4018L1000A 4TTR4024L1000A 4TTR4025L1000A 4TTR4030L1000A 4TTR4031L1000A 4TTR4036L1000A 4TTR4037L1000A 4TTR4042L1000A 4TTR4043L1000A 4TTR4048L1000A 4TTR4060L1000A Note:

More information

Training Syllabus to Instruct/Prepare for the ASE Transit Bus HVAC Test

Training Syllabus to Instruct/Prepare for the ASE Transit Bus HVAC Test APTA STANDARDS DEVELOPMENT PROGRAM RECOMMENDED PRACTICE American Public Transportation Association 1666 K Street, NW, Washington, DC, 20006-1215 APTA BTS-BMT-RP-001-10 Approved October, 2010 APTA Bus Maintenance

More information

OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608

OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608 Electric Water Heater OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608 If your water Heater is Damaged or you have questions

More information

THERMO KING TRUCK & TRAILER UNIT ALARM CODES THIS DOCUMENT SHOWS ALL CURRENT ALARM CODES FOR THERMO KING TRUCK AND TRAILER UNITS.

THERMO KING TRUCK & TRAILER UNIT ALARM CODES THIS DOCUMENT SHOWS ALL CURRENT ALARM CODES FOR THERMO KING TRUCK AND TRAILER UNITS. THERMO KING TRUCK & TRAILER UNIT ALARM CODES THIS DOCUMENT SHOWS ALL CURRENT ALARM CODES FOR THERMO KING TRUCK AND TRAILER UNITS. NOT ALL CODES ARE POSSIBLE ON ANY INDIVIDUAL UNIT. IF THE ALARM APPLIES

More information

COOLSIDE box R410A. conditioned server - racks for direct installation into the room INVERTER E C

COOLSIDE box R410A. conditioned server - racks for direct installation into the room INVERTER E C COOLS conditioned server - racks for direct installation into the room COOLSIDE box INVERTER R410A E C 2 COOLSIDE box GENERAL FEATURES COOLSIDE BOX is a split system consisting of an outdoor motocondensing

More information

Acumen Enterprises, Inc.

Acumen Enterprises, Inc. Hourly rates Discount w/o Rebate Electrician $78.00 10% $70.20 $70.73 HVAC Technician $78.00 10% $70.20 $70.73 Insulator $42.00 10% $37.80 $38.08 Sheet Metal Worker $74.00 10% $66.60 $67.10 Plumber/Pipefitter

More information

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR

USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting

More information

DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS

DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS Electronic thermostatically controlled Low power consumption Compact space saving size Short heat-up times Dehumidification of the cabin Quiet when running Combustion

More information

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version Electronic pressure measurement Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version WIKA data sheet PE 81.66 Applications Machine building

More information

ARB AIR LOCKER Air Operated Locking Differentials

ARB AIR LOCKER Air Operated Locking Differentials CKMTA12 & CKMTA24 On-Board Air Compressor Kit, Maximum Performance Specs: ITEM # QTY DESCRIPTION PART # NOTES 01(a) 2 COMPRESSOR ASSEMBLY (12V) CKM01 1 Voltage 12 Volts (CKMTA12), 24 Volts (CKMTA24) 01(b)

More information

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47)

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47) 47 Brakes - Hydraulic Components (Page GR-47) FS III front brake calipers, servicing Front brake caliper piston, removing and installing FN 3 front brake calipers, servicing Front caliper piston, removing

More information

Heating oven with mechanical convection

Heating oven with mechanical convection FD series 53 Drying ovens / Heating ovens Heating oven with mechanical convection A BINDER FD series heating oven is always used when fast drying and sterilization is required. Thanks to its fully homogeneous

More information

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 E Refrigerant Charging Unit ICOGD 020AH1000 Operating Manual FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 Contents Technical Specifications... 20 Safety... 21 A/C System... 22 Components... 23 Control Panel...

More information

Oil and Coolant Circulating Heating System. Model - OCSM

Oil and Coolant Circulating Heating System. Model - OCSM Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine

More information

Vroom Hardware manual ver. 1.00 Code 114VROOHWE00. Vroom CANBUS USER INTERFACE WITH LCD GRAPHIC DISPLAY AND WITH TEMPERATURE AND HUMIDITY SENSOR

Vroom Hardware manual ver. 1.00 Code 114VROOHWE00. Vroom CANBUS USER INTERFACE WITH LCD GRAPHIC DISPLAY AND WITH TEMPERATURE AND HUMIDITY SENSOR Vroom CANBUS USER INTERFACE WITH LCD GRAPHIC DISPLAY AND WITH TEMPERATURE AND HUMIDITY SENSOR ENGLISH HARDWARE MANUAL ver. 1.00 CODE 114VROOHWE00 page 1 of 22 Important Important Read these instructions

More information

Table of Contents. First Aid for Refrigerant Exposure 35. First Aid for Skin Exposure / First Aid for Eye. Chapter 3:

Table of Contents. First Aid for Refrigerant Exposure 35. First Aid for Skin Exposure / First Aid for Eye. Chapter 3: Table of Contents Chapter 1: Introduction to Automotive Heating, Air Conditioning, and Ventilation 11 Heating, Ventilating, and Air Conditioning (HVAC) Systems 12 Why Is It Called HVAC? Major Heating,

More information

Car and commercial cial air conditioning service solutions

Car and commercial cial air conditioning service solutions Car and commercial cial air conditioning service solutions THE STANDARD IN INFORMATION AND DIAGNOSTICS SYSTEMS INTRODUCTION For the KoolKare range of vehicle air conditioning service equipment we went

More information

Cooling system components, removing and installing

Cooling system components, removing and installing Page 1 of 34 19-1 Cooling system components, removing and installing WARNING! The cooling system is pressurized when the engine is warm. When opening the expansion tank, wear gloves and other appropriate

More information

Heating oven with mechanical control

Heating oven with mechanical control ED series 53 Drying ovens / Heating ovens Heating oven with mechanical control The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C (572 F). Thanks

More information

AIR CONDITIONING LIFE EXTENDER ACX10 FIELD INSTALLATION AND PERFORMANCE EVALUATION

AIR CONDITIONING LIFE EXTENDER ACX10 FIELD INSTALLATION AND PERFORMANCE EVALUATION AIR CONDITIONING LIFE EXTENDER ACX10 FIELD INSTALLATION AND PERFORMANCE EVALUATION Tony Thorne Manager of Electrical Engineering Index Sensors and Controls, Stanwood, WA 98292 USA ABSTRACT Air conditioning

More information

Series P215PR/RM Direct- and Remote-Mount Pressure-Actuated Condenser Fan Speed Controllers for Single-Phase Motors

Series P215PR/RM Direct- and Remote-Mount Pressure-Actuated Condenser Fan Speed Controllers for Single-Phase Motors European Refrigeration Controls Catalogue Catalogue Section 4 Product Bulletin P215PR/RM Issue 01/03/2013 Series P215PR/RM Direct- and Remote-Mount Pressure-Actuated Condenser Fan Speed Controllers for

More information

SERVICE MANUAL OUTDOOR UNIT. No. OBH590 REVISED EDITION-B. Models HFC

SERVICE MANUAL OUTDOOR UNIT. No. OBH590 REVISED EDITION-B. Models HFC SPLIT-TYPE AIR CONDITIONERS Revision B: MUZ-EF25/35/42/50VE - E2 and MUZ-EF25/35VEH - E2 have been added. Please void OBH590 REVISED EDITION-A. OUTDOOR UNIT SERVICE MANUAL HFC utilized R410A. OBH590 REVISED

More information

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Esenboğa Yolu 22 km Akyurt - ANKARA / TURKEY Tel : (00 90 312)

More information

Electronic Diesel Control EDC 16

Electronic Diesel Control EDC 16 Service. Self-Study Programme 304 Electronic Diesel Control EDC 16 Design and Function The new EDC 16 engine management system from Bosch has its debut in the V10-TDI- and R5-TDI-engines. Increasing demands

More information

SENTRA SERIES WATER UNITS. 30 to 300 F Process Temperatures. 1/2 to 7.5 HP Centrifugal Pumps (20-100 GPM) 10 to 34 KW Stainless Steel Heaters

SENTRA SERIES WATER UNITS. 30 to 300 F Process Temperatures. 1/2 to 7.5 HP Centrifugal Pumps (20-100 GPM) 10 to 34 KW Stainless Steel Heaters Temperature Control Units Water & Oil - 00 F Portable Chillers Air & Water-Cooled 0-70 F Central Chillers Air & Water-Cooled Packages & Modules 0-70 F Pump Tank Stations Chilled or Tower Water 00-00 gallons

More information

26 3213.13 Diesel Engine Driven Generators Page 1 of 6

26 3213.13 Diesel Engine Driven Generators Page 1 of 6 Last Update: December 8, 2014 A. Description of System Consultant s Handbook Page 1 of 6 1. Provide a diesel engine driven electric generating unit, factory assembled, tested and certified to operate at

More information

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS

NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug

More information

Typical Aluminum GFM Mass Flow Meter

Typical Aluminum GFM Mass Flow Meter MASS FLOW METERS Design Features n Rigid metallic construction. n Maximum pressure of 1000 psig (70 bars). n Leak integrity 1 x 10-9 of helium. n NIST traceable certification. n Built-in tiltable LCD readout.

More information

Instruction manual for Firstline FCS12000CH

Instruction manual for Firstline FCS12000CH Instruction manual for Firstline FCS12000CH Contents Introduction... 2 Safety Awareness... 3 Safety Awareness... 4 Name of Parts... 5 Name of Parts... 6 Remote Controller Preparation... 7 Operation of

More information

ENGINEERING SPECIFICATIONS FOR CO2 EXTINGUISHING AGENT PRE-ENGINEERED FIRE SUPPRESSION SYSTEMS

ENGINEERING SPECIFICATIONS FOR CO2 EXTINGUISHING AGENT PRE-ENGINEERED FIRE SUPPRESSION SYSTEMS ENGINEERING SPECIFICATIONS FOR CO2 EXTINGUISHING AGENT PRE-ENGINEERED FIRE SUPPRESSION SYSTEMS SECTION 1 GENERAL SPECIFICATIONS I. SCOPE This specification outlines the requirements for a pre-engineered

More information

HEATER, AIR CONDITIONING AND VENTILATION

HEATER, AIR CONDITIONING AND VENTILATION 55-1 GROUP 55 HEATER, AIR CONDITIONING AND VENTILATION CONTENTS GENERAL DESCRIPTION 55-2 HEATER AND AIR CONDITIONING SYSTEM 55-4 HEATER CONTROL 55-6 A/C-ECU 55-7 A/C COMPRESSOR 55-9 CONDENSER 55-9 DUCT

More information

SECTION 23 81 03 - PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM

SECTION 23 81 03 - PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM SECTION 23 81 03 - PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Packaged rooftop air conditioning unit (5 tons and smaller). 2. Roof curb. 1.2

More information

COMMISSIONING OF HVAC SYSTEMS

COMMISSIONING OF HVAC SYSTEMS FORM CL-01 SYSTEMS / EQUIPMENT INCLUDED : SYSTEM STARTUP CHECKLIST I - PRE - STARTUP INSPECTION EQUIPMENT MOUNTING & SETTING CONTROL, INTERLOCK & POWER WIRING ALIGNMENT OF MOTORS & DRIVES PIPING, CONNECTIONS

More information

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases

More information

CW-5000/ 5200,1'8675,$/ &+,//(5 86(5 0$18$/

CW-5000/ 5200,1'8675,$/ &+,//(5 86(5 0$18$/ CW-5000/ 5200 Contents Cautions ---------------------------------------------------------------- 3 Parts introduction --------------------------------------------------- 4 Installation ------------------------------------------------------------

More information

USER S MANUAL HSC-24A

USER S MANUAL HSC-24A AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL

More information

AE21-1319 R6 December 2013. Digital Capacity Control for Copeland Scroll Refrigeration Compressors AE21-1319 R6

AE21-1319 R6 December 2013. Digital Capacity Control for Copeland Scroll Refrigeration Compressors AE21-1319 R6 AE21-1319 R6 December 2013 Digital Capacity Control for Copeland Scroll Refrigeration Compressors TABLE OF CONTENTS Section Page Section Page Safety Safety Instructions...2 Safety Icon Explanation...2

More information

ORC TURBOGENERATOR TYPE CHP - Organic Rankine Cycle Turbogenerator fed by thermal oil, for the combined production of electric energy and heat -

ORC TURBOGENERATOR TYPE CHP - Organic Rankine Cycle Turbogenerator fed by thermal oil, for the combined production of electric energy and heat - Doc. : 08C00031_e Date : 10.02.2009 Page : 1 / 9 ORC TURBOGENERATOR TYPE CHP - Organic Rankine Cycle Turbogenerator fed by thermal oil, for the combined production of electric - (Preliminary) Doc. : 08C00031_e

More information

Cooking at the Speed of light!

Cooking at the Speed of light! Cooking at the Infrared Cooking & Colouring Infrabaker is a modular infrared continuous cooking system developed by Infrabaker International. The machine is designed to cook and/or put colour on a wide

More information

Vario Roof. Model R171. 287 HO 04 Vario Roof (CrullG) 08-17-04

Vario Roof. Model R171. 287 HO 04 Vario Roof (CrullG) 08-17-04 Vario Roof Model R171 287 HO 04 Vario Roof (CrullG) 08-17-04 1 Objectives At the end of this session, you will be able to: 1. Explain how roof can be operated 2. Explain operation of front and rear windows

More information

Laboratory Autoclave SN210/310/510,SQ510/810

Laboratory Autoclave SN210/310/510,SQ510/810 1/5 Laboratory Autoclave SN210/310/510,SQ510/810 SN210/310/510,SQ510/810 The highest temperature of 135deg.C Upwards opening lid SN210 SN310 SN510 Effective Capacity : 20LEffective Capacity : 32L Effective

More information

RODADEACO. Diesel Engine Air Intake Shutoff Valves. Model RDS1 Swing Gate Valve. by AMOT. Overview. Typical Applications. Key Features and Benefits

RODADEACO. Diesel Engine Air Intake Shutoff Valves. Model RDS1 Swing Gate Valve. by AMOT. Overview. Typical Applications. Key Features and Benefits Diesel Engine Air Intake Shutoff Valves Model RDS1 Swing Gate Valve Overview If flammable gas or vapor is present in the atmosphere surrounding a diesel engine, it can be ingested into the engine and become

More information

Electric Flow Boiler

Electric Flow Boiler Electric Flow Boiler EHC offer the most comprehensive range of electric boilers in the market and with our wealth of knowledge and technical expertise we have designed the Slim Jim Electric Flow Boiler

More information

DESIGN STANDARDS SECTION 23 60 00

DESIGN STANDARDS SECTION 23 60 00 PART 1 - GENERAL 1.01 Work included: A. Piping, tubing and fittings. B. Piping specialties. C. Special duty valves. D. Refrigerants. E. Chillers. F. Refrigerant monitors. 1.02 General requirements: A.

More information

A. Martinez, A. L. Klebaner, J. C. Theilacker, B. D. DeGraff, M. White and G. S. Johnson

A. Martinez, A. L. Klebaner, J. C. Theilacker, B. D. DeGraff, M. White and G. S. Johnson FERMILAB-CONF-11-258-AD FERMILAB SRF CRYOMODULE OPERATIONAL EXPERIENCE A. Martinez, A. L. Klebaner, J. C. Theilacker, B. D. DeGraff, M. White and G. S. Johnson Fermi National Accelerator Laboratory Batavia,

More information

Rotary screw compressors kw 5,5 7,5 11 15 18,5 22 30 bar 8 10 13 15 MODULO

Rotary screw compressors kw 5,5 7,5 11 15 18,5 22 30 bar 8 10 13 15 MODULO Rotary screw compressors kw 5,5 7,5 11 15 18,5 22 30 bar 8 10 13 15 MODULO The sum of technology TriAB Air ends The BALMA screw compressors are fitted with the unique TriAB air ends that are designed and

More information

MACHINES RECOVERY SYSTEMS SUPEREVAC CHARGING HOSES VACUUM AND HOSE ADAPTERS, VALVES AND PARTS CHARGING SYSTEMS GAUGES INSTRUMENTS ELECTRONIC

MACHINES RECOVERY SYSTEMS SUPEREVAC CHARGING HOSES VACUUM AND HOSE ADAPTERS, VALVES AND PARTS CHARGING SYSTEMS GAUGES INSTRUMENTS ELECTRONIC RECOVERY SUPEREVAC 29 RECOVERY 1/4" 4-IN-1 BALL VALVE TOOL For vacuum, charge, cores and recovery 18997 18975 18971 18978 18979 Remove and replace Schrader cores without losing charge. Extra 1/4" side

More information

Highly accelerated Life Testing HALT in Barco

Highly accelerated Life Testing HALT in Barco Highly accelerated Life Testing HALT in Barco 2007 Ivan Malfait Topics HALT in Barco Situating HALT in Barco How it started Current situation: The HALT Installation itself Situating HALT in the Design

More information

DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS

DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS DIESEL-FUELED, BLOWN AIR, DUCTABLE HEATERS Electronic thermostatically controlled Low power consumption Compact space saving size Short heat-up times Dehumidification of the cabin Quiet when running Combustion

More information