Evolution for Bone Collimator-Detector Response Compensation in Iterative SPECT Imaging Reconstruction Algorithm
|
|
- Kevin Gibson
- 7 years ago
- Views:
Transcription
1 GE Healthcare Evolution for Bone Collimator-Detector Response Compensation in Iterative SPECT Imaging Reconstruction Algorithm Version 1.0
2 Introduction The quality and quantitative accuracy of SPECT reconstructed images are affected by noise in the projection data, resolution degradation caused by the collimator-detector response (CDR) function of the imaging system, and photon attenuation and scatter in patient s body. Recently, significant progress has been achieved in the development of model-based corrective SPECT image reconstruction methods that include correction for these image quality-degrading factors. The corrective image reconstruction package (also referred to as Resolution Recovery) developed by JHU group provides improved SPECT images by including physical models of the imaging process into iterative image reconstruction algorithm. In particular, it provides accurate 3D models of the collimatordetector response functions for variety of clinical applications. This results in significant enhancement in image quality and potential improvement in diagnostic properties. Clinical application of these techniques are available and can be applied to reduced time SPECT acquisitions with equal or better image characteristics when compared to standard reconstructions (16). Collimator-detector response in SPECT Collimator-detector blur is the main factor affecting the resolution and noise properties of nuclear medicine images. In reconstructed SPECT images, these effects are strongly affected by the applied reconstruction algorithm and its parameters. In SPECT acquisition with parallel collimation, images of a point source located at different distances from the collimator fully describe (CDR) function for specific collimator design, photon energy and energy window parameters, and distance of the imaged object from collimator surface, as demonstrated in figure 1. Figure 1. SPECT acquisition geometry and collimator-detector response. CDR consists of four main components: intrinsic response (system without collimator) and the geometric, septal penetration and septal scatter components of the collimator parameters. The geometric component is the portion of the CDR resulting from photons by passing through the collimator holes without interaction with the collimator. The septal penetration component describes portion of the CDR due to photons reaching the detector after passing through one or more septa without scattering. The scatter component is due to photons that scatter in the septa and are then detected. For the majority of clinically used collimator designs, the last two components are generally significant for only mediumand high-energy imaging (see fig.2). The response functions are spatially invariant in planes parallel to the face of the collimator and that the intrinsic response is approximately spatially invariant in the detection plane.
3 Figure 3. Scheme of iterative reconstruction process. Compensation requires that a table of CDR functions at all relevant distances to collimator surface be pre-calculated and retrieved, or computed on the fly, during reconstruction. Onthe-fly calculations involve using the theoretical predicted collimator geometric response for an equivalent round hole shape [1, 2], and ignore penetration and septal scatter effects. These approximations work - well for conventional hexagonal shaped holes and low-energy isotopes, such as Tl and Tc- 99m. Figure 2. Examples of CDR function and its components. For each combination of acquisition system, radiopharmaceutical and particular acquisition protocol, the collimator detector response function provides the probability that a photon emitted from any point of the imaged object will contribute to a pixel of the resulting image. A CDR compensation technique was developed at the University of North Carolina Chapel Hill (UNC) and Johns Hopkins University (JHU) [3, 4, 5]. In fast rotation based iterative reconstruction implementations, CDR compensation is achieved by convolving the projected ray with corresponding Point Spread Function (PSF) in the course of projection and back-projection operations (Fig.4) Accurate predictions of the geometric response function for various collimator designs were derived [1, 2]. Penetration and scatter components of the collimator detector response function can be obtained and analyzed using Monte-Carlo simulation methods. Compensation for collimatordetector response in iterative reconstruction By including an accurate model of collimator-detector response function in an iterative SPECT reconstruction algorithm, the blurring effect may be included in the iterative reconstruction process, resulting in improved special resolution. Figure 4. Modeling of collimator-detector response in a rotation based projector. Adapted from: E.C. Frey and B.M.W. Tsui, Collimator-detector response compensation in SPECT In order to incorporate CDR into the reconstruction process, the following information is utilized during the course of image reconstruction:
4 Collimator design parameters : holes length and diameter, septa thickness Detector characteristics : intrinsic resolution, crystal thickness, collimator-detector gap Acquisition parameters: center-of-rotation to collimator face distances for every projection view acquired. This information for a variety of different collimators is stored in an additional look-up table which is part of the reconstruction package. Effect of collimator detector response compensation on reconstructed images data when no compensation is applied. On the other hand, image noise level tends to amplify with iteration number. Taking into account the differences in the iterative reconstruction process, with and without CDR, applicationspecific optimization of acquisition and processing parameters is essential to successful utilization of corrective reconstruction method [12, 13, 14, 15 ] (Fig.6) It was demonstrated (Fig.6) that application of corrective image reconstruction methods results in improved uniformity of myocardium wall in cardiac SPECT and specifically CDR compensation provides better recovery of lesion contrast in variety of clinical applications [8, 16]. The effect of CDR compensation on reconstructed images has been studied from qualitative [7], quantitative [8, 9, 10], and clinical task-based [11, 13, 15] prospective. It has been found that CDR corrected images demonstrate not only improved resolution and signal-to-noise ratio, but also lower noise variability in reconstructed images (Fig. 5). It has also been shown that the noise texture with CDR compensation changes, exhibiting wider correlations than non-corrected images, resulting in improved resolution without corresponding increase in noise. CDR is quite sensitive to incorrect modeling of the CDR function or the use of incorrect acquisition parameters. It is important that these parameters are accurately passed from the camera to the iterative reconstruction process. Figure 6. Reconstructed clinical Tc-99m myocardium SPECT images. Top line - attenuation correction only applied, bottom line attenuation and CDR compensation. Numbers mark number of updates (product of iterations number and subsets number) applied. From: E.C. Frey and B.M.W. Tsui, Collimator-detector response compensation in SPECT Mathematical and human observer studies demonstrated improvement in cold and hot lesion detectability with application of CDR compensation [15, 16]. Clinical Application CDR has been implemented on the Xeleris workstation for use in the Bone SPECT and Whole Body SPECT protocols. The application of CDR in shorter acquisition times has been shown to demonstrate equivalent or improved image quality over standard, full time acquisitions (16). Image improvement of half time CDR reconstructed SPECT over full time standard OSEM iterative reconstruction is demonstrated in (fig. 9). Figure 5. Hot sphere phantom images. Left: schematic presentation of insert locations; middle: slice of conventional reconstruction, right: slice of corrective reconstruction with collimator-detector response compensation. No post filtering was applied. Significant improvement of the resolution properties requires more iterations of the reconstruction method with compensation than are usually recommended for the same
5 Figure 8. Changes in Area under ROC curve (AUC) across acquisition/processing combinations for selected 10mm 2:1 contrast lesions results from Channelized Hotelling Observer study. CDR OSEM Figure 7. Examples of reconstructed image slices of Tc-99m bone SPECT containing 10 mm 6:1 contrast lesions; simulated data. Figures 7 and 8 show results from an investigation of image properties where application of CDR compensation gave rise to studies of potential usefulness of this compensation in shorter than usual clinical acquisition time to ensure nondegraded diagnostic quality [16, 17]. Figure 9. Demonstrated improvement in WB SPECT half time acquisition over standard OSEM iterative reconstruction process
6 Conclusion Detailed description of the effect of the collimator detector response on properties and diagnostic quality of SPECT images is a subject of extensive research. So far, significant evidence has been accumulated that diagnostic quality of NM images is substantially improved by incorporation of collimator-detector system modeling into iterative image reconstruction. References 1. Metz, C.E., The geometric transfer function component for scintillation camera collimators with straight parallel holes. Phys. Med. Biol., (6): p Tsui, B.M.W. and G.T. Gullberg, The Geometric Transfer- Function for Cone and Fan Beam Collimators. Physics in Medicine and Biology, (1): p Tsui, B.M.W., et al., Implementation of simultaneous attenuation and detector response correction in SPECT. IEEE Transactions on Nuclear Science, (1): p Tsui, B.M.W., et al., The importance and implementation of accurate three-dimensional compensation methods for quantitative SPECT. Phys Med Biol, (3): p Tsui, B.M.W., et al., Characteristics of reconstructed point response in three-dimensional spatially variant detector response compensation in SPECT, in Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, P. Grangeat and J.-L. Amans, Editors. 1996, Kluwer Academic Publishers. p Wilson, D.W. and H.H. Barrett, The effects of incorrect modeling on noise and resolution properties of ML-EM images. Nuclear Science, IEEE Transactions on, (3): p Tsui BMW, Zhao XD*, Frey EC and McCartney WH. Quantitative SPECT: Basics and Clinical Considerations. Seminar in Nuclear Medicine, Vol. XXIV, No 1 (January), pp 38-65, Pretorius, P.H., et al., Reducing the influence of the partial volume effect on SPECT activity quantitation with 3D modelling of spatial resolution in iterative reconstruction. Physics in Medicine and Biology, (2): p Kohli, V., et al., Compensation for distance-dependent resolution in cardiac-perfusion SPECT: impact on uniformity of wall counts and wall thickness. Nuclear Science, IEEE Transactions on, (3): p Pretorius, P.H., et al., Comparison of detection accuracy of perfusion defects in SPECT for different reconstruction strategies using polar-map quantitation. Ieee Transactions on Nuclear Science, (5): p Narayanan, M.V., et al., Human-observer receiveroperating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for Tc-99m myocardial perfusion imaging. Journal of Nuclear Medicine, (11): p Sankaran, S., et al., Optimum compensation method and filter cutoff frequency in myocardial SPECT: A human observer study. Journal of Nuclear Medicine, (3): p Frey, E.C., K.L. Gilland, and B.M.W. Tsui, Application of taskbased measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Transactions on Medical Imaging, (9): p Gifford, H.C., et al., LROC analysis of detector-response compensation in SPECT. IEEE Transactions on Medical Imaging, (5): p He, X, et al., A mathematical observer study for the evaluation and optimization of compensation methods for Myocardial SPECT using a phantom population that realistically models patient variability. IEEE Transactions on Nuclear Science, (1): Volokh, L., et al., Efficacy of corrective reconstruction with collimator detector response compensation for short Tc- 99m bone SPECT acquisition in a bone lesion detection task./ Abstract presented at SNM Meeting, 2005/ 17. Keidar, Z. et al., Half-time bone SPECT acquisition - Assessment of a new Collimator Detector Response (CDR) reconstruction algorithm / Abstract submitted to RSNA Meeting, 2005 / 18. Frey, E.C., Tsui, B.M.W., Collimator-Detector Response Compensation in SPECT. in: Quantitative Analysis in Nuclear Medicine Imaging. Springer 2005 GE Healthcare Waukesha, WI U.S.A General Electric Company All rights reserved. GE Healthcare, a division of General Electric Company. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE Representative for the most current information.
Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
More informationCone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
More informationA Three-Dimensional Correlation Method for Registration of Medical Images in Radiology
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami
More informationPurchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage
More informationComputed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
More informationImprovements in quality and quantification of 3D PET images
Improvements in quality and quantification of 3D PET images The resolution of PET depends on some effects impossible to be readily corrected The implementation of 3D spatially-variant PSF (radially asymmetric)
More informationASNC IMAGING GUIDELINES FOR NUCLEAR CARDIOLOGY PROCEDURES. Single photon-emission computed tomography
ASNC IMAGING GUIDELINES FOR NUCLEAR CARDIOLOGY PROCEDURES Thomas A. Holly, MD, a Brian G. Abbott, MD, b Mouaz Al-Mallah, MD, c Dennis A. Calnon, MD, d Mylan C. Cohen, MD, MPH, e Frank P. DiFilippo, PhD,
More informationHow To Improve Your Ct Image Quality
Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
More informationPET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationLaue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
More informationExperimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationCommunication Optimization and Auto Load Balancing in Parallel OSEM Algorithm for Fully 3-D SPECT Reconstruction
2005 IEEE Nuclear Science Symposium Conference Record M11-330 Communication Optimization and Auto Load Balancing in Parallel OSEM Algorithm for Fully 3-D SPECT Reconstruction Ma Tianyu, Zhou Rong, Jin
More informationCT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
More informationPhysics in Nuclear Medicine
~.~ ~ Physics in Nuclear Medicine Second Edition James A. Sorenson, Ph.D. Director, Medical Physics Professor of Radiology Depat1meht of Radiology University of Utah Medical Center Salt Lake City, Utah
More informationPerformance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE
Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using Parham Geramifar 1,3,4, Mohammad Reza Ay 2,3,4, Mojtaba Shamsaie Zafarghandi 1, George Loudos 5, Arman
More informationFor many years, physicians have been faced with the. SPECT/CT Physical Principles and Attenuation Correction*
SPECT/CT Physical Principles and Attenuation Correction* James A. Patton 1 and Timothy G. Turkington 2 1 Vanderbilt University Medical Center, Nashville, Tennessee; and 2 Duke University Medical Center,
More informationCardius. XPO Series. Cardius 3 XPO
Cardius XPO Series Cardius 3 XPO Three imaging systems, all of which provide outstanding image quality, superior efficiency, and increased patient comfort. Cardius 2 XPO Includes the standard features
More informationFeasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
More informationLecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
More informationMODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada
More informationTruly digital PET imaging
Advanced Molecular Imaging Vereos PET/CT Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital
More informationImage Analysis for Volumetric Industrial Inspection and Interaction
Image Analysis for Volumetric Industrial Inspection and Interaction Rasmus Larsen Mads F. Hansen, Hildur Olafsdottir, Thomas H. Mosbech, Lasse F. Laursen DTU Informatics (CT) reconstruction from few projections
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationExperimental and modeling studies of imaging with curvilinear electronic eye cameras
Experimental and modeling studies of imaging with curvilinear electronic eye cameras Viktor Malyarchuk, 1 Inhwa Jung, 1 John A. Rogers, 1,* Gunchul Shin, 2 and Jeong Sook Ha 2 1 Department of Materials
More informationPart A: Structure and Organization
Part A: Structure and Organization Radioactive Materials License NRC (or State) radioactive materials license present Personnel Medical Director Meets Medical Director responsibility requirements Meets
More informationPositron Range Modeling for Statistical PET Image Reconstruction
Positron Range Modeling for Statistical PET Image Reconstruction Bing Bai, Ananya Ruangma, Richard Laforest, Yuan-Chuan Tai and Richard M. Leahy Signal and Image Processing Institute, University of Southern
More informationHigh-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard
More informationAN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
More informationINTRODUCTION TO RENDERING TECHNIQUES
INTRODUCTION TO RENDERING TECHNIQUES 22 Mar. 212 Yanir Kleiman What is 3D Graphics? Why 3D? Draw one frame at a time Model only once X 24 frames per second Color / texture only once 15, frames for a feature
More informationTruly digital PET imaging
Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital PET, resulting in significantly improved
More informationCALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
More informationA PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - nzarrin@qiau.ac.ir
More informationMeasuring Size and Distance in Digital Mammography
On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval
More informationOptimization of image quality through online motion correction
Advanced imaging technologies a revolution in diagnosis and treatment Optimization of image quality through online motion correction Oline Vinter Olesen December 10, 2010 Medical Solutions, Siemens Institute
More informationQuantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
More informationDynamic Neutron Imaging
Wir schaffen Wissen heute für morgen Pierre Boillat Neutron Imaging & Activation Group (NIAG) and Electrochemistry Laboratory (LEC), Paul Scherrer Institute, Switzerland Dynamic Neutron Imaging PSI, 8.
More informationMDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
More informationCT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More informationHow To Analyze Ball Blur On A Ball Image
Single Image 3D Reconstruction of Ball Motion and Spin From Motion Blur An Experiment in Motion from Blur Giacomo Boracchi, Vincenzo Caglioti, Alessandro Giusti Objective From a single image, reconstruct:
More informationSuper-Resolution Reconstruction in MRI: Better Images Faster?
Super-Resolution Reconstruction in MRI: Better Images Faster? Esben Plenge 1,, Dirk H. J. Poot 1, Monique Bernsen 2, Gyula Kotek 2, Gavin Houston 2, Piotr Wielopolski 2, Louise van der Weerd 3, Wiro J.
More informationCS 534: Computer Vision 3D Model-based recognition
CS 534: Computer Vision 3D Model-based recognition Ahmed Elgammal Dept of Computer Science CS 534 3D Model-based Vision - 1 High Level Vision Object Recognition: What it means? Two main recognition tasks:!
More informationNuclear PACS: Problems, Solutions
Nuclear PACS: Problems, Solutions by Jerold W. Wallis, MD Introduction Reprinted with permission: Wallis JW. Nuclear PACS: Problems, Solutions. Decisions in Imaging Economics. 2004;6;21-25. There is increasing
More informationOptical Design Tools for Backlight Displays
Optical Design Tools for Backlight Displays Introduction Backlights are used for compact, portable, electronic devices with flat panel Liquid Crystal Displays (LCDs) that require illumination from behind.
More informationNuclear Medicine Accreditation Program Requirements
Nuclear Medicine Accreditation Program Requirements OVERVIEW... 2 MEDICARE IMPROVEMENT F PATIENTS AND PROVIDERS ACT OF 2008 (MIPPA)... 2 MANDATY ACCREDITATION TIME REQUIREMENTS... 3 WITHDRAWN, ADDED, REPLACEMENT
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationHigh Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful
More informationTwo years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions Laurent Desbat TIMC-IMAG, UMR 5525, Grenoble University, France Laurent.Desbat@imag.fr Rolf
More informationManufacturing Process and Cost Estimation through Process Detection by Applying Image Processing Technique
Manufacturing Process and Cost Estimation through Process Detection by Applying Image Processing Technique Chalakorn Chitsaart, Suchada Rianmora, Noppawat Vongpiyasatit Abstract In order to reduce the
More informationFundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
More informationCT for the production floor: Recent developments in Hard- and Software for industrial CT Systems
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies
More informationConcepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
More informationRemoving chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches.
Plastic Threads Technical University of Gabrovo Yordanka Atanasova Threads in plastic products can be produced in three ways: a) by direct moulding with thread punch or die; b) by placing a threaded metal
More informationLow-dose CT for Pulmonary Embolism
Low-dose CT for Pulmonary Embolism Gautham Gautham P. P. Reddy, Reddy, MD, MD, MPH MPH University University of of Washington Washington Introduction Introduction CT CT accounts accounts for for > 50%
More informationTEXTURE AND BUMP MAPPING
Department of Applied Mathematics and Computational Sciences University of Cantabria UC-CAGD Group COMPUTER-AIDED GEOMETRIC DESIGN AND COMPUTER GRAPHICS: TEXTURE AND BUMP MAPPING Andrés Iglesias e-mail:
More information1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003. Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors
1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST 2003 Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors Carolyn E. Lehner, Student Member, IEEE, Zhong He, Senior
More informationEndoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
More informationDICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
More informationThe promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
More informationAdvanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
More informationGPR Polarization Simulation with 3D HO FDTD
Progress In Electromagnetics Research Symposium Proceedings, Xi an, China, March 6, 00 999 GPR Polarization Simulation with 3D HO FDTD Jing Li, Zhao-Fa Zeng,, Ling Huang, and Fengshan Liu College of Geoexploration
More informationAn introduction to OBJECTIVE ASSESSMENT OF IMAGE QUALITY. Harrison H. Barrett University of Arizona Tucson, AZ
An introduction to OBJECTIVE ASSESSMENT OF IMAGE QUALITY Harrison H. Barrett University of Arizona Tucson, AZ Outline! Approaches to image quality! Why not fidelity?! Basic premises of the task-based approach!
More informationReflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.
T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting
More informationImage Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
More informationModelling, Extraction and Description of Intrinsic Cues of High Resolution Satellite Images: Independent Component Analysis based approaches
Modelling, Extraction and Description of Intrinsic Cues of High Resolution Satellite Images: Independent Component Analysis based approaches PhD Thesis by Payam Birjandi Director: Prof. Mihai Datcu Problematic
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationThe Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
More informationMODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT
MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT C. Reinhart, C. Poliwoda, T. Guenther, W. Roemer, S. Maass, C. Gosch all Volume Graphics GmbH, Heidelberg, Germany Abstract:
More informationProjection Center Calibration for a Co-located Projector Camera System
Projection Center Calibration for a Co-located Camera System Toshiyuki Amano Department of Computer and Communication Science Faculty of Systems Engineering, Wakayama University Sakaedani 930, Wakayama,
More informationPhysics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
More informationChapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
More informationFig.1. The DAWN spacecraft
Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics
More informationCT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
More informationClassification of Fingerprints. Sarat C. Dass Department of Statistics & Probability
Classification of Fingerprints Sarat C. Dass Department of Statistics & Probability Fingerprint Classification Fingerprint classification is a coarse level partitioning of a fingerprint database into smaller
More informationLow-resolution Character Recognition by Video-based Super-resolution
2009 10th International Conference on Document Analysis and Recognition Low-resolution Character Recognition by Video-based Super-resolution Ataru Ohkura 1, Daisuke Deguchi 1, Tomokazu Takahashi 2, Ichiro
More informationSpiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray
More informationDISCLAIMER. This document was prepared as an account of work sponsored by an agency of the United States
DISCLAIMER This document was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor the University of California nor any of their
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationPHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Medical and Industrial Tomography - W.B.Gilboy
MEDICAL AND INDUSTRIAL TOMOGRAPHY Department of Physics, University of Surrey, Guildford, Surrey, U.K. Keywords: Radiography, transmission tomography, emission tomography, microtomography, SPECT (single
More informationInformation about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
More informationComputerized Tomography, X-rays, and the Radon Transform
1 Computerized Tomography, X-rays, and the Radon Transform 1.1 Introduction The purpose of this chapter is to give an informal introduction to the subject of tomography. There are very few mathematical
More informationTemporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board
More informationStudy of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus
More informationDETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm
URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-199:0 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 DETECTION OF SUBSURFACE DAMAGE
More informationAlgebra 1 2008. Academic Content Standards Grade Eight and Grade Nine Ohio. Grade Eight. Number, Number Sense and Operations Standard
Academic Content Standards Grade Eight and Grade Nine Ohio Algebra 1 2008 Grade Eight STANDARDS Number, Number Sense and Operations Standard Number and Number Systems 1. Use scientific notation to express
More informationDevelopment of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
More informationPHOTOGRAMMETRIC TECHNIQUES FOR MEASUREMENTS IN WOODWORKING INDUSTRY
PHOTOGRAMMETRIC TECHNIQUES FOR MEASUREMENTS IN WOODWORKING INDUSTRY V. Knyaz a, *, Yu. Visilter, S. Zheltov a State Research Institute for Aviation System (GosNIIAS), 7, Victorenko str., Moscow, Russia
More informationVariance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
More informationCHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
More informationNeutron Detection Setups proposed for
Neutron Detection Setups proposed for DESPEC D. Cano-Ott on behalf of the WG members CIEMAT, IFIC, LNL, FYL, UPC, UU, UW Motivation GOAL: to measure neutron emission probabilities and energies for neutron
More informationALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE
ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE Gevaldo Lisboa de Almeida, Carlos Feu Alvim Agência Brasileiro-Argentina de
More informationwaves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object
PHYS1000 Optics 1 Optics Light and its interaction with lenses and mirrors. We assume that we can ignore the wave properties of light. waves rays We represent the light as rays, and ignore diffraction.
More informationIntroduction to Medical Imaging. Lecture 11: Cone-Beam CT Theory. Introduction. Available cone-beam reconstruction methods: Our discussion:
Introduction Introduction to Medical Imaging Lecture 11: Cone-Beam CT Theory Klaus Mueller Available cone-beam reconstruction methods: exact approximate algebraic Our discussion: exact (now) approximate
More informationQuality control of CT systems by automated monitoring of key performance indicators: a two-year study
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,
More informationGE Healthcare. Revolution EVO. More than just high tech. Higher purpose.
GE Healthcare Revolution EVO More than just high tech. Higher purpose. Revolution EVO. Designed with purpose. Today s healthcare environment is about creating new solutions to pressing needs. It s about
More informationHigh-definition 3D Image Processing Technology for Ultrasound Diagnostic Scanners
Hitachi Review Vol. 63 (204), No. 9 584 Featured Articles High-definition 3D Image Processing Technology for Ultrasound Diagnostic Scanners Realistic 3D Fetal Imaging Masahiro Ogino Takuma Shibahara, Ph.D.
More informationResolution Enhancement of Photogrammetric Digital Images
DICTA2002: Digital Image Computing Techniques and Applications, 21--22 January 2002, Melbourne, Australia 1 Resolution Enhancement of Photogrammetric Digital Images John G. FRYER and Gabriele SCARMANA
More information