Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi
|
|
|
- Matthew Dean
- 9 years ago
- Views:
Transcription
1 Instrument/Method Validation Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi
2 Instrument/Method Validation QUESTION AND ANSWER PRESENTATION
3 Instrument Validation What is instrument validation?
4 Instrument Validation Definition: It is a series of processes through which you test your system to verify or validate the performance specifications published by the manufacturer of the instrument.
5 Instrument Validation Why is it important to validate the instrument?
6 Why Validate your instrument/method? It is required by the CLIA regulation (CLIA 42 CFR part ). International sites participate in DAIDS/NIH funded studies, we must be in compliance with the U.S. regulations. You want to validate the manufacturer s claims for their method performance characteristics, under your current environmental conditions, e.g. temp, humidity, water, electricity, operator skills etc.
7 Why validate your instrument? You want to ensure that the amount of error present in the system won t affect the interpretation of the test result and compromise patient care. Ensure that effects of shipment and storage did not affect your instrument performance.
8 Instrument Validation When do you validate your instrument?
9 When to validate instrument Initial installment of the instrument. When a new method or test is implemented. When a major part is changed e.g. photometer of a chemistry instrument. When an instrument is borrowed. When you MOVE your instrument from one location to another. Annually (if time allows).
10 Three validation principles IQ, OQ, PQ
11 IQ- Installation Qualification Installation qualification verifies that the equipment and its subsystems have been installed in accordance to manufacturer s specifications. Usually performed by the company s field representative. Examples: Verification that all components parts are functional Verification that local supply voltages conform to instrument Verify that the ambient conditions exist for optimal instrument performance. Verify correct versions of software were issued with the instrument. Setting up of instrument parameters (date, time, language, test protoc.
12 OQ- Operation Qualification This validation provides evidence that the instrument operates as expected and confirms that installation was successful. Done by the field representative and lab technicians. Training of instrument operation. Testing of controls, calibrators and a few patient samples.
13 Instrument Validation - PQ What experiments are performed in Performance Qualification (Method Validation)?
14 Instrument Validation - Experiments Precision (Replication) Linearity (to verify reportable range) Accuracy (correlation or comparison) Reference Intervals (Normal values) Analytical Sensitivity Analytical Specificity
15 Instrument Validation What are CLIA guidelines for validation?
16 CLIA Regulations CLIA regulations are based on the complexity of the test method. 3 categories: 1. Waived tests does not require validation. 2. Non-waived unmodified tests moderate and high complexity tests needs validation. 3. Non-waived, modified (or in-house developed) tests validation is required. Information about classification of specific tests is available at :
17 Instrument Validation - Linearity Definition of Linearity?
18 Instrument Validation- Linearity Linearity or AMR(analytical measurement range) is an assessment of the lowest and highest levels at which an analyte can be accurately measured without any type of dilution or concentration. It is important to validate the manufacturer s claims for reportable range of their system/method.
19 Instrument Validation- Precision Definition of Precision?
20 Instrument Validation- Precision Definition Precision is the degree of reproducibility among several independent measurements of the same sample for the same analyte.
21 Instrument Validation- Precision Precision experiment is performed to estimate the imprecision or random error of the analytical method. Precision is measured in terms of coefficient of variation (CV) and standard deviation (SD). The smaller the CV and SD, the better the precision.
22 Instrument Validation - Accuracy Define Accuracy?
23 Instrument Validation - Accuracy Definition: A measurement of the exactness of an analytical method, or the closeness of agreement between the measured value and the true value.
24 Instrument Validation - Accuracy Performed to estimate inaccuracy or systematic error of the new method. Experiment is performed by analyzing forty or more patient samples by the new method (test method) and a comparative method, then estimate the systematic errors on the basis of the differences observed between the two methods.
25 Accuracy/Correlation Experiment Westgard has a set of statistical tools on the internet to help calculate the statistics:
26 Reference Range Intervals or NRR Definition of a Reference Range Intervals?
27 Reference Range Intervals or NRR Definition: Reference ranges are a measured set of values determined to occur in a healthy nondiseased population.
28 Reference Range Intervals or NRR Why should you validate the manufacturer s reference intervals or establish your own NRR?
29 Reference Range Intervals or NRR Differences in demographics of your population due to environmental or genetic factors. Differences in test methods and instruments.
30 HPTN 035 and MTN 015 protocol required tests What validation exercises are required for: 1. ALP, ALT, AST, T.Bil and Creatinine? 2. CBC/DIFF? 3. CD4? 4. APTT, PT, INR? 5. RPR/TPHA? 6. HIV ELISA? 7. Western Blot?
31 HPTN 035 and MTN 015 protocol required tests 8. RNA PCR? 9. Probe-Tec CT/GC? 10. HIV Rapid Tests? 11. HSV-2?
32 Validation Summary Need to write up a validation summary and file together with results data in a binder. Report should be signed by performing technician and lab Manager/Director or designee. Submit validation data and summary to Network Lab for approval.
33 Reasons why we don t validate our instruments/methods? Discussion
34 NEW METHOD BC CX-5 PRO
35 NEW METHOD BC CX-5 PRO
36 REFERENCES Method Validation by Westgard HPTN Lab Manual by Network Lab
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Laboratory Director Responsibilities
Clinical Laboratory Improvement Amendments (CLIA) Laboratory Director Responsibilities What Are My Responsibilities As A Laboratory Director NOTE: Congress passed the Clinical Laboratory Improvement Amendments
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
OPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Installation and Operational Qualification Protocol (Reference: SOP )
Project Name Equipment Process Line/Location Project Number Serial Number Model Number Protocol number WRITTEN BY: REVIEWED BY: Position APPROVAL TO EXECUTE: Position: PROTOCOL COMPLETION APPROVAL: Position:
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories () Edited by Richard Pang, PhD, FACB Senior Consultant Pro Q Consulting Services Hong Kong For Hong Kong
Test Reliability Indicates More than Just Consistency
Assessment Brief 015.03 Test Indicates More than Just Consistency by Dr. Timothy Vansickle April 015 Introduction is the extent to which an experiment, test, or measuring procedure yields the same results
2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests
2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification
White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
Basic Lessons in Laboratory Quality Control
Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and
Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren
Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency
Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3
Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3 Prepared by Date Adopted Supersedes Procedure # Estelle Piwowar-Manning Quality Assessment vers 2 Sarah Dawson Paul Richardson
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING
NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials
Assay Development and Method Validation Essentials
Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding
Calibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
Measurement Systems Correlation MSC for Suppliers
Measurement Systems Correlation MSC for Suppliers Copyright 2003-2007 Raytheon Company. All rights reserved. R6σ is a Raytheon trademark registered in the United States and Europe. Raytheon Six Sigma is
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice
EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 Working Party on Control
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory in India: A Pilot Study
Ind J Clin Biochem (Apr-June 2011) 26(2):131 135 DOI 10.1007/s12291-010-0083-1 ORIGINAL ARTICLE Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
QbD Approach to Assay Development and Method Validation
QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Six Sigma Quality New Metrics vs Old Myths
Six Sigma Quality New Metrics vs Old Myths James O. Westgard, PhD Professor, Dept Path & Lab Medicine University of Wisconsin Medical School President, Westgard QC, Inc. Madison, WI http://www.westgard.com/sixsigprimer.htm
Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs
Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs Elizabeth Herron URI Watershed Watch/ National Facilitation of CSREES Volunteer Monitoring Ingrid Harrald Cook
Flow Cytometry. CLIA Compliance Manual*
Flow Cytometry CLIA Compliance Manual* Revised for the International Society for Advancement of Cytometry (ISAC) and Clinical Cytometry Society (CCS) By Michael Keeney ART, FIMLS Teri Oldaker B.A., CLS(NCA)QCYM
1 For more information T: +1.847.664.1499 / E: [email protected] / W: ul.com/dap
DATA ACCEPTANCE PROGRAM UL Client Test Data Program Investigations of products by UL typically involve the testing of product samples. While UL maintains extensive facilities for these purposes, customers
Agilent Cary 60 UV-Vis
Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current
CHAPTER 7 QUALITY ASSESSMENT
CHAPTER 7 QUALITY ASSESSMENT Chapter 7 QUALITY ASSESSMENT 7.1 OVERVIEW OF CBB-SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL PROGRAM An extensive, site-specific Quality Assurance/Quality Control (QA/QC) program
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical
Design of Experiments for Analytical Method Development and Validation
Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD TABLE OF CONTENTS Introduction Applicability General Topics Proficiency Testing
Process Validation for Medical Devices
Process Validation for Medical Devices Dan O Leary CBA, CQE, CRE, CSSBB, CIRM, LLC 603-209-0600 [email protected] Copyright 2010 by, LLC Process Validation for Medical Devices 1 Instructor Introduction
A.1 Sensor Calibration Considerations
Wind Speed and Direction Appendix A. Quality Assurance A major pre-requisite for establishing quality control standards for wind measurements is a strong quality assurance program. Remember the mantra
Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
Laboratory and Technician Certification. by Nabil Kamel, M.A.Sc., P.Eng. CCIL Program Manager
Laboratory and Technician Certification by Nabil Kamel, M.A.Sc., P.Eng. CCIL Program Manager Laboratory and Technician Certification Table of contents What is lab certification? Why lab certification?
Quality Assurance for the Clinical Laboratory
EP Evaluator Quality Assurance for the Clinical Laboratory Simplified process, powerful results While we (clinical labs) are but a small component of the total healthcare system, our data control a large
Rashika Maharaj. Strategies for maintaining high standards for Good Clinical Laboratory Practices
Strategies for maintaining high standards for Good Clinical Laboratory Practices Rashika Maharaj MRC HPRU Laboratory Manager MTN Regional Meeting 10 September 2008 What is GCLP? Is your lab GCLP compliant?
Contents. Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions...
Vol. 28 No. 20 Replaces H47-A Vol. 16 No. 3 One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline Second Edition This document provides guidelines
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency? http://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/clia_compbrochure_508.pdf
Assessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
Inspections: Central Laboratory of the Regions
Page 1 Inspections: Central Laboratory of the Regions Protocol for Open Laboratory Review of a CDOT Materials Testing Laboratory Protocol for the Inspection of Region Materials Laboratories by the Central
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract
Clinical Chemistry / ANALYTIC QUALITY TODAY The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA
INTEROPERABILITY UNIT
INTEROPERABILITY UNIT MODULES FOR THE PROCEDURES FOR ASSESSMENT OF CONFORMITY, SUITABILITY FOR USE AND EC VERIFICATION TO BE USED IN THE TECHNICAL SPECIFICATIONS FOR INTEROPERABILITY Reference: Version
Quality Assurance Procedures for Construction and Materials Sampling and Testing 14
Page 1 Quality Assurance Procedures for Construction and Materials Sampling and Testing 14 1. PURPOSE 1.1 To prescribe policies, procedures, and guidelines to assure the quality of materials on all Colorado
DATA QUALITY OBJECTIVES
The Wastewater Treatment Plant Operators Guide to Biosolids Sampling Plans CHAPTER 5 DATA QUALITY OBJECTIVES Goals of the Sampling Plan ESSENTIAL ELEMENTS OF A SAMPLING PLAN Description of the Facility
Quality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
Process Validation: Practical Aspects of the New FDA Guidance
Process Validation: Practical Aspects of the New FDA Guidance ISPE Boston Chapter Meeting April 18, 2013 Rusty Morrison Commissioning Agents, Inc. Objectives / Summary What is Process Validation? Regulatory
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
4 QUALITY ASSURANCE 4.1 INTRODUCTION
4 QUALITY ASSURANCE 4.1 INTRODUCTION BNL has an established Quality Assurance/Quality Control (QA/QC) Program to ensure that the accuracy, precision, and reliability of environmental monitoring data are
No. 706. Page 1 of 5. Issue Date 4/21/2014
Subject: ROUTINE MONITORING VISITS Standard Operating Procedure Prepared By: Beaumont Research Coordinating Center, Research Institute PURPOSE Prior Issue Date 8/15/2011 No. 706 Issue Date 4/21/2014 Page
Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma
Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use
VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR LABORATORY TESTING
Test Procedure for VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR Texas Department of Transportation TxDOT Designation: Tex-901-K Effective Date: August 1999 1. SCOPE 1.1 All weighing devices shall
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
ORACLE CONSULTING GROUP
ORACLE CONSULTING GROUP 9 Golder Ranch Rd., Ste. 1 Tucson, Arizona 9 Web Site: E-mail: 20-2-0 20-2-0 (FAX) CONSULTING MEMORANDUM QUALITY SYSTEM INSPECTION TECHNIQUE
The Regulatory Binder/Trial Master File: Essential Records for the Conduct of a Clinical Trial
The Regulatory Binder/Trial Master File: Essential Records for the Conduct of a Clinical Trial Clinical Research Operations & Regulatory Support (CRORS) Ann Glasse, RN, BSN, MBA Director-CRORS Objectives
Qualification Study CHO 360-HCP ELISA (Type A to D)
Short Report Qualification Study CHO 360-HCP ELISA (Type A to D) Presented by: http://www.biogenes.de Version 01 Issue date: 27.11.2013 Version 01 Page 1 of 8 Table of Content TABLE OF CONTENT... 2 1 INTRODUCTION...
Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program
Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly
Considerations When Validating Your Analyst Software Per GAMP 5
WHITE PAPER Analyst Software Validation Service Considerations When Validating Your Analyst Software Per GAMP 5 Blair C. James, Stacy D. Nelson Introduction The purpose of this white paper is to assist
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Document Control for Control Freaks. Kristina Martin, MLS(ASCP) CM,MS
Document Control for Control Freaks Kristina Martin, MLS(ASCP) CM,MS Objectives 1. Define document control 2. List components of document control 3. Identify regulatory and standards associated with document
ASVCP Guidelines: Quality Assurance for Point of Care Testing in Veterinary Medicine
ASVCP Guidelines: Quality Assurance for Point of Care Testing in Veterinary Medicine Version 1.0 (May 2013) Developed by the American Society for Veterinary Clinical Pathology (ASVCP) Quality Assurance
Introduction to Cloud Computing What is SaaS? Conventional vs. SaaS Methodologies Validation Requirements Change Management Q&A
Best Practices for Validation of a Software as a Service (SaaS) Customer Relationship Management (CRM) Solution Presented By: Gregg Mauriello Validation Manager Elise Miner Associate Validation Manager
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
IMPAACT INTERNATIONAL LABORATORY QUALITY ASSURANCE PROGRAM
SOP Number: LAB-6008-02 Title: IMPAACT International Laboratory Quality Assurance Program Version: 2.00 Effective Date: March 4. 2010 Approved By: Brooks Jackson, M.D. Last Review Date: February 24, 2010
What Does Having a FDA Cleared Pregnancy Test Mean?
What Does Having a FDA Cleared Pregnancy Test Mean? Clinical Trials Transformation Initiative (CTTI) July 15-16, 2013 Denise N. Johnson-Lyles, Ph.D. Toxicology Branch Chief Division of Chemistry and Toxicology
Biochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
SAGE SCIENCE INC. Pippin Prep IQ/OQ. 460008 Rev. A Originated: March 5, 2015. Sheet 1 of 11. Prepared by: Alex Vira SUMMARY
SAGE SCIENCE INC Pippin Prep IQ/OQ 460008 Rev. A Originated: March 5, 2015 Sheet 1 of 11 Prepared by: Alex Vira SUMMARY This document provides instructions and records for Installation Qualification (IQ)
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
Installation Qualification/Operational Qualification Protocols and Instructions. Experion Automated Electrophoresis Station and Software
Installation Qualification/Operational Qualification Protocols and Instructions Experion Automated Electrophoresis Station and Software Validation Kit (Catalog #700-7051) Security Edition Software (Catalog
Guido Baum Phone: +49(0)221-806 2592 Mail: [email protected] www.qal1.de
Determination of CO 2 and N 2 O emissions with CEM Systems within the scope of greenhouse gas trading CEM 2014 Istanbul Thursday, 15 th May 2014 TÜV Rheinland Energie und Umwelt GmbH Guido Baum Phone:
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter
Computerised Systems. Seeing the Wood from the Trees
Computerised Systems Seeing the Wood from the Trees Scope WHAT IS A COMPUTERISED SYSTEM? WHY DO WE NEED VALIDATED SYSTEMS? WHAT NEEDS VALIDATING? HOW DO WE PERFORM CSV? WHO DOES WHAT? IT S VALIDATED -
ISO 9001:2008 QUALITY MANUAL. Revision B
ISO 9001:2008 QUALITY MANUAL Revision B Because we want you to achieve the highest levels of performance, we connect care Because with leading we want science you to achieve continuously the highest improve,
Epidemiology of HCV and HIV/HCV Infection in Guangzhou, China. Title. Charles Wang, M.D.
Epidemiology of HCV and HIV/HCV Infection in Guangzhou, China Title Charles Wang, M.D. Background Hepatitis C infects 170 million worldwide & 3 million Chinese HIV infects 33.3 million worldwide & 740,000
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
Basic research methods. Basic research methods. Question: BRM.2. Question: BRM.1
BRM.1 The proportion of individuals with a particular disease who die from that condition is called... BRM.2 This study design examines factors that may contribute to a condition by comparing subjects
Qualification, calibration and maintenance of stability chambers. Dr. Mathias Kramer
Qualification, calibration and maintenance of stability chambers Dr. Mathias Kramer Agenda 1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability