USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Size: px
Start display at page:

Download "USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY"

Transcription

1 USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory Medicine, Emory University School of Medicine

2 Learning Objectives Identify the seven performance characteristics that should be evaluated prior to reporting results from a new test method. Verify the claims of manufacturers regarding analytical performance by following CLSI guidelines. Demonstrate ongoing compliance with the method evaluation criteria contained in accreditation guidelines. Who is CLSI and what are these guidelines? Outline Method evaluation basic definitions and experiments Use of the CLSI Evaluation Protocols Use of StatisPro software in method evaluation

3 Who is CLSI? Clinical and Laboratory Standards Institute ANSI-accredited, global, nonprofit standards development organization CLSI has over 2,000 members organizations such as IVD manufacturers, hospital laboratories, reference laboratories, universities, professional associations, and government agencies We promote the development and use of voluntary consensus standards and guidelines within the health care community. Our documents help health care organizations meet their responsibilities with efficiency, effectiveness, and global acceptance.

4 We Make the Blue Books

5 Standards in the Clinical Laboratory Goal of standardization in the laboratory: The right laboratory test at the right time with the right result leads to quality diagnostics and improved patient care, and improved public health around the world. Standardized test Standardized procedure Standardized reporting Improved outcomes

6 How are CLSI Documents developed? CLSI documents are developed by volunteer experts from three distinct constituencies: professions, government, and industry. Under the supervision of a consensus committee, these volunteers work on: Document Development Committees or Standing subcommittees and Working groups

7 CLSI Consensus Process Government Balance Industry Professions

8 Why are CLSI Guidelines Important? The US Food and Drug Administration(FDA) recognizes over 100 CLSI documents. The College of American Pathologists (CAP) recognizes 80 CLSI documents. The Joint Commission recognizes over 144 CLSI documents. All Evaluation Protocol guidelines in this presentation are recognized by all three groups.

9 Why are Evaluation Protocols Important? Provide clear and thorough guidance. Evaluation protocols are guidelines for clinical laboratories and manufacturers to characterize the performance of their analytical systems. Ensure consistency with good laboratory practice. Good laboratory practice requires clinical laboratories to verify performance claims before reporting results used for decisions about patient care. Help you to comply with the law! Evaluation of performance characteristics is required by regulatory and accreditation bodies in the United States. See ( ).

10 CLSI and Evaluation Protocols CLSI has over 25 Evaluation Protocol Guidelines. These include: EP05 Evaluation of Precision EP06 Evaluation of Linearity EP09 Evaluation of Bias and Comparability Using Patient Samples EP10 Preliminary Evaluation (Bias, Carryover, Drift, Linearity) EP15 Verification of Precision and Trueness EP17 Limits of Detection and Limits of Quantitation C28 Defining, Establishing, and Verifying Reference Intervals

11 Performance Characteristics The seven performance characteristics that should be evaluated before reporting results of a new test method include: 1. Precision 2. Accuracy (measured bias) or comparability (measured differences) 3. Linearity over the measuring interval or analytical measurement range (AMR) 4. Limit of detection (LoD) and limit of quantitation (LoQ or analytical sensitivity) 5. Specificity or interference 6. Reagent or sample (analyte) carryover 7. Reference interval or decision value (interpretive information)

12 Precision & Accuracy

13 Apply a clinical perspective; set a target, an analytical goal, before you begin Perform experiments that gather representative data about a method s analytical performance Convert data into statistical estimates of errors Compare error estimates to specifications for medically allowable error for an objective assessment of the errors

14 Introducing a New Method Establish a Need Define the Quality Goal Method Selection Submit Specimen Method Evaluation Method Development Implementation Preventive Maintenance Routine Analysis Quality Control Practices Report Result

15 APPROACH IN METHOD EVALUATION: Evaluate imprecision and inaccuracy IMPRECISION Refers to Random Analytic Error (Lack of repeatability, reproducibility) INACCURACY Refers to Systematic Analytic Error (Lack of trueness) 1. Constant 2. Proportional TOTAL ERROR Combined error for a single result

16 RELIABLE DECISIONS ABOUT PERFORMANCE REQUIRE: 1. Standards for acceptable performance 2. Experimental protocols to estimate performance reliably 3. Criteria for comparing estimated performance with performance standards

17 PERFORMANCE STANDARD (PS) Specify: EA... Allowable error XC... Decision level Format: PS = EA at XC

18 ALLOWABLE ERROR (EA) The amount of error that can be tolerated without invalidating the medical usefulness of the result or causing the test to fail a proficiency testing event

19 DECISION LEVEL (XC) Any concentration of the analyte that is critical for medical interpretation whether for diagnosis, monitoring, or therapeutic decisions. Laboratory data are most carefully interpreted at these decision level concentrations.

20 DECISION LEVELS FOR GLUCOSE plasma glucose, mg/dl

21 DECISION LEVELS FOR GLUCOSE Performance standards for Glucose PS1 = mg/dl Medical Decision Hypoglycemia PS2 = 10% = mg/dl Impaired glucose control PS3 = mg/dl Poorly controlled diabetes

22 Sources of Allowable Errors 1. Proficiency testing requirements for acceptable performance 2. Literature guidelines a. based on physician surveys e.g.: Karon, Boyd & Klee, Glucose Meter Performance Criteria for Tight Glycemic Control Estimated by Simulation Modeling, Clin Chem, 2010; 56: b. based on intra-individual biological variation of analyte Ricos C et al., Scand J Clin Lab Invest,1999; 59: Fraser C, Biological Variation: From Principles to Practice, AACC, 2001 Internet at 3. Input from clinicians and/or professional judgment

23 Formulation of Criteria to Judge Analytic Errors General form: compare observed analytic error to the specification for allowable analytic error Performance is acceptable when: observed error < allowable error Performance is not acceptable when: observed error > allowable error

24 Performance Characteristics: Precision CLSI Guidelines for Precision EP15: a five-day procedure to verify that imprecision meets the claims of a measurement procedure (EP15 is most frequently used by clinical laboratories for method evaluation.) EP05: a 20-day procedure to establish the imprecision for a measurement procedure

25 Replication Experiment 1. Time period: within-run within-day day-to-day 2. Number of samples: minimum of Sample matrix: simulate patient sample 4. Analyte concentration: medical decision limit 5. Calculations: mean, standard deviation (SD), coefficient of variation (CV)

26 Performance Characteristics: Accuracy Accuracy [Trueness] (Measured as Bias) Bias: Estimate of a systematic measurement error; a quantitative measure of the average difference between results from a measurement procedure and results from an accepted reference measurement procedure. When a reference measurement procedure is not available for an analyte, a best-available comparative method may be used to measure bias. Frequently, clinical laboratories perform a comparison of patient sample results between a new and an existing measurement procedure. ( correlation studies )

27 Performance Characteristics: Accuracy CLSI Guidelines for Trueness (Measured as Bias) EP15: a method comparison to verify that a new method conforms to a manufacturer s claim for comparability to another procedure. (minimum of 20 patient samples) EP09: a method comparison to establish a claim for method comparability. (minimum of 40 patient samples)

28 Comparison of Methods Experiment CLSI EP9-A: User Comparison of Quantitative Clinical Laboratory Methods Using Patient Samples 1. Choice of comparative method: critical for the conclusions which can be made 2. Number of test samples: minimum N = 40 uniform distribution (EP9-A includes a table) a bin-box approach

29 Comparison of Methods Experiment Bin-box approach: 5 Number of samples 10

30 Comparison of Methods Experiment 3. Replicates: required for EP9-A: desirable, but not always practical 4. Time period: minimum of 5 days 5. Data analysis: review daily Check for maximum allowable differences between methods EP9-A includes a test for outliers within and between methods 6. EP9-A has a section on establishing manufacturer s claims

31 Three Approaches to Analyzing Comparison of Methods Data 1. correlation coefficient 2. t-test statistics 3. regression statistics

32 Sensitivity of Statistical Parameters to Errors Parameter Random Constant Proportional LEAST SQUARES SLOPE Y-INTERCEPT STD. ERROR no no yes no yes no yes no no BIAS sd no yes yes no yes yes yes no no T-TEST CORRELATION COEFFICIENT r

33 Effect of range on the correlation coefficient H H Range 0 to to 110 Random Error 10 units 10 units Corr. Coef., r

34 Correlation coefficient, r Responds to random error. Value depends on the range of data. Does not estimate analytical bias or random error between methods. Merely presents the relationship of the range of the data to the scatter of the data between methods. Therefore, the correlation coefficient should NOT be used to judge acceptability of analytical methods in method comparison studies.

35 Linear regression statistics Subject to certain limitations: Data must be linear Outliers must be carefully examined Range of data must be wide: a. r > 0.99 (Waakers et al.) b. r > (CLSI EP9-A)

36 Recommendations for Method Comparison Summary Present graph of data Present slope, y-intercept, and Sy/x Present mean and standard deviation of X data Present correlation coefficient ONLY to show that least squares regression is applicable; if not, use Deming or Passing-Bablock regression statistics

37 Performance Characteristics: Linearity Linearity Measuring Interval or Analytical Measurement Range (AMR) A linearity study is used to establish or verify the measuring interval for a measurement method. Measuring Interval: the interval between lower and upper numerical values for which a method can produce quantitative results suitable for the intended clinical use. The measuring interval is verified by demonstrating a linear relationship between the measured and expected concentration relationships. CLSI Guideline for Linearity Measuring Interval EP06: procedures to verify or establish the linear measuring interval of a measurement procedure.

38 Performance Characteristics: LoD/LoQ Limit of Detection (LoD) & Limit of Quantitation (LoQ) (sometimes referred to as Analytical Sensitivity ) LoD: the lowest amount of analyte (measurand) in a sample that can be detected with a stated probability. LoQ: the lowest amount of analyte (measurand) in a sample that can be quantified with acceptable precision and bias under stated experimental conditions. Usually, laboratories review and accept the manufacturer s claims for LoD and LoQ. But these characteristics can be tested by laboratories using: CLSI Guideline for LoD and LoQ EP17: procedures for verifying or establishing the LoD and the LoQ

39 Performance Characteristics: Interference Interference: an artifactual increase or decrease in the apparent quantity of an analyte due to the presence of a substance that reacts nonspecifically with the measuring system. Most manufacturers evaluate a large number of substances known or suspected to be potential interferents. They report this information in the Instructions For Use (IFU). It is not practical for most clinical laboratories to repeat such an investigation and inspection of the manufacturer s information is frequently sufficient. But these characteristics can be tested by laboratories using: CLSI Guideline for Interference EP7: procedures for testing constant error due to interference

40 Interference Experiment: Factors 1. See CLSI EP7-A2 2. What to test: Literature review Always test hemolysis, lipemia, bilirubin Tube additives 3. Concentrations to test: Interferent: highest compatible with life Analyte: at medical decision levels 4. Volume of interferent <10% of sample 5. Replicates: Based on Effect / Stm (see EP7) 6. Validate technique with current method

41 Interference Experiment: N=? Number of Measurements / Replicates: at least several samples per interferent at least duplicates per sample EP7 lists a table of N as a function of bias/stm (EA,I/Stm ), with which one can determine how many replicates are necessary to reach 95% probability of observing a certain magnitude of error: EA,I/Stm No. Replicates EA,I/Stm No. Replicates

42 Performance Characteristics: Carryover Carryover: the discrete amount of reagent or analyte carried by the measuring system from one test into subsequent test(s), thereby erroneously affecting test results. Reagent carryover among different measurement procedures on multichannel automated analyzers is an evaluation that is usually conducted by measuring system manufacturers. But this characteristic can be tested by laboratories using: CLSI Guideline for Carryover EP10: includes an assessment of sample carryover along with other parameters. NOTE: EP10 is intended to determine if a device has unacceptable performance. It is recognized in the CAP Chemistry Checklist as an acceptable way to measure carryover.

43 Performance Characteristics: Reference Intervals Reference Interval: interpretive information for laboratory test results that is frequently provided as the central 95% interval of results for a group of well-defined reference individuals. Laboratories can produce reference intervals in a variety of ways, including testing procedures found in CLSI Guideline for Reference Intervals or Decision Value C28: procedures for establishing a reference interval or verifying the suitability of a manufacturer-proposed reference interval

44 Reference Interval Determination Transference of established reference intervals to an individual laboratory or a new method may be accomplished in a variety of ways: 1. Subjective assessment by a responsible individual; the Medical Director (sometimes called by divine judgment ) 2. Donor testing a. Verify with ~ 20 donor samples b. Validate/Estimate using ~ 60 donor samples c. Establish using ~ 120 donor samples 3. Calculation use regression statistics from a comparison of methods study to calculate reference limits for the new method (Y) that correspond to the reference interval limits of the former method (X). Y=a+b X

45 CLSI Makes Life Easier with StatisPro In October 2010, CLSI released StatisPro software: Direct, faithful implementation of CLSI Evaluation Protocol Guidelines Study Advisor step-by-step help for each study Four steps to complete a study: Definition, Data Input, Analysis, and Signoff

46 StatisPro Pick a Study Type

47 StatisPro Study Design Study Goal Identifying Information Details of the Study Performance Claim to be Verified Description of Materials Used

48 StatisPro Data Entry Copying and Pasting from any spreadsheet application or Windows application with clipboard support is easy.

49 StatisPro Analysis 1 - Inspect group: Evaluate the data visually using various plots and tables. You can choose to show or hide excluded observations. 2 - Outliers group: Select an observation to exclude from the calculations. 3 - Study-specific group: Select commands that continue to evaluate the data and reach a study conclusion. 4 - Sign Off group: Add any comments, your name, and a signature line to the study report so it is ready for a handwritten signature when printed.

50 StatisPro Study Advisor

51 StatisPro Demonstration Demonstrate EP15 (method comparison) and EP06 (linearity).

52 User Experience with StatisPro StatisPro is useful when introducing new methods into your laboratory. StatisPro is useful when performing six-month linearity or calibration verification studies. By using StatisPro: You are demonstrating compliance with regulatory and accreditation bodies. You are ensuring that your laboratory delivers accurate results.

53 Thank You Questions?

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1 ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas

More information

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory

More information

Validation of measurement procedures

Validation of measurement procedures Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

To avoid potential inspection

To avoid potential inspection Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become

More information

SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision)

SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision) August 2012 Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision) This document provides guidance on how to verify comparability of quantitative

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail inab@inab.ie Web

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i

More information

SAMPLE. Effects of Different Sample Types on Glucose Measurements

SAMPLE. Effects of Different Sample Types on Glucose Measurements 1st Edition POCT06 Effects of Different Sample Types on Glucose Measurements This report provides information to assist the clinical and point-ofcare staff in result and measurement procedure comparisons

More information

SMF Awareness Seminar 2014

SMF Awareness Seminar 2014 SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified

More information

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Calibration Verification

Calibration Verification Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in

More information

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests 2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1 Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification

More information

Med TechNet Online Services East Amherst, NY November 1997

Med TechNet Online Services East Amherst, NY November 1997 Med TechNet Online Services East Amherst, NY November 1997 DECISION MAKING IN THE CLINICAL LABORATORY: A QUANTITATIVE AND STATISTICAL APPROACH FOR METHODS EVALUATION Associate Professor Department of Clinical

More information

QbD Approach to Assay Development and Method Validation

QbD Approach to Assay Development and Method Validation QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Quality Assurance for the Clinical Laboratory

Quality Assurance for the Clinical Laboratory EP Evaluator Quality Assurance for the Clinical Laboratory Simplified process, powerful results While we (clinical labs) are but a small component of the total healthcare system, our data control a large

More information

A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories

A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories () Edited by Richard Pang, PhD, FACB Senior Consultant Pro Q Consulting Services Hong Kong For Hong Kong

More information

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by: How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated

More information

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

SAMPLE. Gas Chromatography/Mass Spectrometry Confirmation of Drugs; Approved Guideline Second Edition

SAMPLE. Gas Chromatography/Mass Spectrometry Confirmation of Drugs; Approved Guideline Second Edition March 2010 Gas Chromatography/Mass Spectrometry Confirmation of Drugs; Approved Guideline Second Edition This document provides guidance on establishing uniform practices necessary to produce quality data

More information

Fairfield Public Schools

Fairfield Public Schools Mathematics Fairfield Public Schools AP Statistics AP Statistics BOE Approved 04/08/2014 1 AP STATISTICS Critical Areas of Focus AP Statistics is a rigorous course that offers advanced students an opportunity

More information

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Prepared by the Advisory Committee on Laboratory Medicine College of Physicians & Surgeons of Alberta Serving the Public by guiding

More information

Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)

Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

Guidance for Industry and FDA Staff

Guidance for Industry and FDA Staff Guidance for Industry and FDA Staff Review Criteria for Assessment of C-Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hscrp) and Cardiac C-Reactive Protein (ccrp) Assays Document issued

More information

MEDICAL LABORATORY TECHNICIAN COMPETENCY PROFILE

MEDICAL LABORATORY TECHNICIAN COMPETENCY PROFILE Description of Work: Positions in this banded class support or perform laboratory tests that are used in the diagnosis and treatment of patients and animals. Duties performed include: receiving or procuring

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi Instrument/Method Validation QUESTION AND ANSWER PRESENTATION Instrument Validation What is instrument validation?

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

1.0 What Are the Purpose and Applicability of Performance Specification 11?

1.0 What Are the Purpose and Applicability of Performance Specification 11? While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r),

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r), Chapter 0 Key Ideas Correlation, Correlation Coefficient (r), Section 0-: Overview We have already explored the basics of describing single variable data sets. However, when two quantitative variables

More information

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1) CORRELATION AND REGRESSION / 47 CHAPTER EIGHT CORRELATION AND REGRESSION Correlation and regression are statistical methods that are commonly used in the medical literature to compare two or more variables.

More information

AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK

AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK ENVIRONMENT PROTECTION AUTHORITY June 1999 AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK, June 1999 David Robinson

More information

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD

More information

USE OF REFERENCE MATERIALS IN THE LABORATORY

USE OF REFERENCE MATERIALS IN THE LABORATORY USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from

More information

6. MEASURING EFFECTS OVERVIEW CHOOSE APPROPRIATE METRICS

6. MEASURING EFFECTS OVERVIEW CHOOSE APPROPRIATE METRICS 45 6. MEASURING EFFECTS OVERVIEW In Section 4, we provided an overview of how to select metrics for monitoring implementation progress. This section provides additional detail on metric selection and offers

More information

Six Sigma Metric Analysis for Analytical Testing Processes

Six Sigma Metric Analysis for Analytical Testing Processes Six Sigma Metric Analysis for Analytical Testing Processes Sten Westgard, MS, Westgard QC Introduction Laboratories seek objective assessment and comparison of analytical methods and instrumentation performance.

More information

1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96

1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96 1 Final Review 2 Review 2.1 CI 1-propZint Scenario 1 A TV manufacturer claims in its warranty brochure that in the past not more than 10 percent of its TV sets needed any repair during the first two years

More information

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2 Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

More information

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations General Information Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations Our Background Foundation Laboratory is an advanced, accredited and certified clinical diagnostic

More information

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract

James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract Clinical Chemistry / ANALYTIC QUALITY TODAY The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA

More information

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few

More information

Great Lakes National Program Office and Office of Water Quality Management Training Modules

Great Lakes National Program Office and Office of Water Quality Management Training Modules GLOSSARY Assessment - the evaluation process used to measure the performance or effectiveness of a system and its elements. As used here, assessment is an all-inclusive term used to denote any of the following:

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Topic: Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Click on the links below to be taken to a specific section of the FAQs. General

More information

Basic Lessons in Laboratory Quality Control

Basic Lessons in Laboratory Quality Control Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and

More information

UCLA PATHOLOGY & LABORATORY MEDICINE QUALITY MANAGEMENT PLAN

UCLA PATHOLOGY & LABORATORY MEDICINE QUALITY MANAGEMENT PLAN Page 1 of 11 UCLA PATHOLOGY & LABORATORY MEDICINE QUALITY MANAGEMENT PLAN INTRODUCTION A quality management system can be described as a set of key quality elements that must be in place for an organization

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

CALIBRATION PRINCIPLES

CALIBRATION PRINCIPLES 1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they

More information

International Consortium for Harmonization of Clinical Laboratory Results. Operating Procedures

International Consortium for Harmonization of Clinical Laboratory Results. Operating Procedures International Consortium for Harmonization of Clinical Laboratory Results Operating Procedures Approved: February 11, 2014 Background Results from clinical laboratory measurement procedures should be comparable

More information

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,

More information

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014

G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014 Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

ALACC Frequently Asked Questions (FAQs)

ALACC Frequently Asked Questions (FAQs) Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC

More information

Overview of ISO 17511 for 'Biologicals'

Overview of ISO 17511 for 'Biologicals' JCTLM Symposium on Reference Measurement Systems for Biologicals International Bureau of Weights and Measures Pavillon de Breteuil, Sèvres, FR 2004-12-15 Overview of ISO 17511 for 'Biologicals' R. Dybkaer

More information

Quantifying measurement error from digital instruments

Quantifying measurement error from digital instruments Quantifying measurement error from digital instruments W. BLAKE LAING AND SEAN BRYANT SOUTHERN ADVENTIST UNIVERSITY CHAT TANOOGA, TN What I m doing HELPING STUDENTS LEARN TO CONSTRUCT KNOWLEDGE First lab:

More information

Experiment #1, Analyze Data using Excel, Calculator and Graphs.

Experiment #1, Analyze Data using Excel, Calculator and Graphs. Physics 182 - Fall 2014 - Experiment #1 1 Experiment #1, Analyze Data using Excel, Calculator and Graphs. 1 Purpose (5 Points, Including Title. Points apply to your lab report.) Before we start measuring

More information

Managing POCT: Trying to Control Testing in an Out-of-Control Environment. William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11

Managing POCT: Trying to Control Testing in an Out-of-Control Environment. William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11 Managing POCT: Trying to Control Testing in an Out-of-Control Environment William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11 Learning Objectives Participants should be able to: Discuss

More information

Six Sigma Quality New Metrics vs Old Myths

Six Sigma Quality New Metrics vs Old Myths Six Sigma Quality New Metrics vs Old Myths James O. Westgard, PhD Professor, Dept Path & Lab Medicine University of Wisconsin Medical School President, Westgard QC, Inc. Madison, WI http://www.westgard.com/sixsigprimer.htm

More information

Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD

Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

The Facts on Equine Drug Testing

The Facts on Equine Drug Testing The Facts on Equine Drug Testing CONTACT Travis Mays, MS Interim Section Head Analytical Chemistry, Drug Testing Laboratory Toxicology Laboratory 979.845.3414 tmays@ LIMITATIONS IN DRUG TESTING While advancements

More information

Implementing New USP Chapters for Analytical Method Validation

Implementing New USP Chapters for Analytical Method Validation Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: Ludwig_Huber@labcompliance.com Today s Agenda Handling Method Changes vs. Adjustments

More information

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 Paul Bachner, MD, FCAP Professor & Chair Department of Pathology & Laboratory Medicine University of Kentucky

More information

Contents. Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions...

Contents. Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... Vol. 28 No. 20 Replaces H47-A Vol. 16 No. 3 One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline Second Edition This document provides guidelines

More information