American Association for Laboratory Accreditation
|
|
|
- Luke Ellis
- 10 years ago
- Views:
Transcription
1 Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number of colonies in a test sample. These examples apply for all quantitative microbiological methods, including plate count and MPN. To the extent possible they use conventional concepts for estimating uncertainty, but they are not the only possible methods for determining uncertainty of counting methods. These examples are based on 20 data points each, but larger datasets will produce more reliable estimates and smaller data sets may be used with caution. The coverage factor obtained from the Student t-tables should be used to estimate expanded uncertainty for smaller datasets. For each of these examples, the log base 10 result was used instead of the natural log for ease of calculation. Conversion to log base 10 is necessary because of the nature of the measurand (Colony Forming Units or CFU), which reproduces exponentially and the dilution scheme which is determined by serial dilution. Example 1: Reproducibility Replicates for Laboratory Control Samples This procedure illustrates the use of reproducibility replicates to estimate uncertainty for the same type of sample matrix analyzed. This technique captures various sources of uncertainty that can affect routine customer samples by having replicates be produced independently under as many different conditions as possible that normally occurs in a laboratory. This procedure presents the techniques recommended in ISO TS19036: Microbiology of foods and animal feeding stuffs Guidelines for the estimation of measurement uncertainty for quantitative determinations. The example data given in Table 1 are results from control samples, not test samples, which are analyzed through all of the steps of the test method and were set up on different days, in duplicate, by different analysts, using different equipment (e.g. balances, pipettors) and possibly using different batches of media/reagents. It is assumed that the matrix used to generate the data in Table 1 is representative of the samples analyzed by the laboratory and therefore this estimate may be used to estimate the uncertainty for any sample. Steps 7-9 below describe considerations for using the estimates for other samples of the same matrix type. This is useful for food or feed production laboratories where the same types of samples are analyzed for products to determine conformance to product specifications.
2 Page 2 of 12 Table 1: Replicates generated under reproducibility conditions Raw Data (actual CFU recovered) First Replicate Log 10 Value Raw Data (actual CFU recovered) Second Replicate Log 10 Value Difference between Replicates (Log 10 Value) Difference between Replicates Squared Step 1 Transform the raw data by taking the log base 10 value (log 10 ) of the data (column 2, 4). Step 2 Calculate the overall mean of 40 results in columns 2 and 4, this is Step 3 Calculate the difference between the transformed replicates (column 5). Step 4 Square the differences between the transformed replicates (column 6). Step 5 Add the differences together (column 6) and divide by 2n, where n = the total number of pairs of duplicates (for this example n = 20) to get Step 6 Take the square root of the result in step 5; this equals the pooled reproducibility standard deviation, which is Step 7 Convert this standard deviation into a relative standard deviation (RSD) by dividing by the mean from Step 2, which produces an RSD of.0499.
3 Page 3 of 12 Step 8 To provide a higher range of values that cover what is likely to be observed in other samples, apply the coverage factor (k=2 for 95% coverage) to the RSD to get the estimate of the expanded relative uncertainty, (Note this is a log 10 value). If using the student t-tables, with n=20 (19 degrees of freedom), the coverage factor k would be 2.09, which would provide an expanded relative uncertainty of Step 9 To calculate the uncertainty for any subsequent laboratory result, the result is first converted to the log base 10 value (log 10 ), multiplied by and then this expanded uncertainty is added and subtracted from the log value. Step 10 To estimate the MU of a sample, convert the log value for the sample measurement back to CFU for the reported result. This is accomplished by taking the anti-log of each of the endpoints of the interval (anti-log of x = 10 x ). NOTE: By convention the results are always rounded down for the value at the lower end of the range and always rounded up for the value at the upper end. In this way the minimal 95% coverage is preserved. For example, for a result of 150 CFU: 150 in log 10 = and the expanded uncertainty in log counts is times = Adding and subtracting from gives an interval from to ; transforming back to counts: =90.97, and = Therefore the uncertainty interval is 90 to 248 CFU. Example 2: Recovery Replicates for Laboratory Control Samples If it is safe to assume that recovery is reasonably constant for a particular organism in a given matrix, then the uncertainty in microbiological measurements can be estimated by looking at recovery over time. Recovery differences over time should reflect the various sources of uncertainty, including the following: Random error (common cause variability) Analyst differences Different equipment Different environmental conditions In this procedure (Table 2) the same amount of inoculum is plated both without matrix and with the matrix of interest. The difference between the counts from the plate without matrix and the plate with matrix samples is a measure of recovery of the organism, usually expressed as a percentage of the CFU count in the inoculated sample. The 20 replicates were of greatly different levels (as can be expected in some situations) but the recovery is reasonably constant. It is assumed that the matrix used for this study is representative of those samples where an estimate of uncertainty is needed.
4 Page 4 of 12 Table 2: Recovery replicates CFU Inoculated Log 10 Value CFU recovered in Spike Log 10 Value % Recovery of Log Values 30, , , , , , , , , , , , , , , , , , , , , , , , , , , , , Step 1 Transform the CFU values (columns 1, 3) to log base 10 value (log 10 ) (columns 2, 4). Step 2 Calculate the % recovery of the log 10 values by dividing column 4 by column 2 and multiplying by 100 (column 5). Step 3 Calculate the mean and standard deviation (SD) of the % recovery of the log 10 values. The mean recovery is 97.0% and the SD of % recovery is 3.6%. The SD is an estimate of the combined standard uncertainty which can be used as a relative uncertainty. Step 4 For reporting purposes, apply the coverage factor to the SD to obtain the expanded uncertainty. For 95% coverage we use a coverage factor of k=2. The expanded uncertainty in this example is thus 7.2%. If using the student t-tables, with n=20 (19 degrees of freedom), the coverage factor k would be 2.09, which would provide an expanded uncertainty of 7.5%. Step 5 Because the recovery is expressed as a percentage, when calculating the expanded uncertainty for a sample, this percentage needs to be multiplied by the log 10 value in order to estimate the uncertainty in log units. For example, for a result of 150 CFU: 150 in log 10 = and the expanded uncertainty in log counts is times 0.072= Adding and subtracting from gives an interval from
5 Page 5 of to ; transforming back to counts: =104.6, and = Therefore the uncertainty interval is 104 to 216 CFU. Note regarding % Recovery: In this example, recovery is close to 100% and therefore there is no indication of significant bias. If average recovery were much lower (e.g., 80%) then there would be an indication of consistent bias that should be investigated. However, it is not current conventional practice to correct for bias in quantitative microbiology results. Discussion of Examples 1 and 2: The intervals produced in examples 1 and 2 can appear to be similar, but they are quite different and the calculations for estimating uncertainty are very different; in the first example (reproducibility replicates) the sources of uncertainty are reflected in the differences between replicates, so the difference between each pair of log counts is converted into a variance (squared difference) and these variances are pooled and then converted into an estimate of the combined standard uncertainty. In the second example (recovery replicates), the main observation is the recovery on each trial and the sources of uncertainty are reflected in the difference between different sets of replicates. A single SD is calculated as the estimate of combined standard uncertainty. Two other types of replicates are sometimes used by some laboratories to estimate uncertainty, but these are not recommended because they do not capture all the sources of uncertainty. Therefore these types of replicates could significantly underestimate the combined uncertainty: Plate replicates: dilutions on one matrix are carried out by one analyst, and duplicate dilution plates of the single control sample are made. True replicates: the original sample is split and diluted in two separate and independent series of dilutions by one analyst, and only one set of plates is prepared from each sample. The sources of measurement uncertainty that are adequately captured in the estimate by the various types of replicates are listed in the table below: Table 3: Sources of uncertainty reflected in different types of replicates MU Source Type of Replicate Reproducibility Recovery True Plate Random error X X X X Counting error X X X X Dilutions X X X Environment X X Equipment X X Analyst X X
6 Page 6 of 12 Example 3: Use of Laboratory Control Samples with Same Target Values This approach is defined in P103b, Annex: Policy on Estimating Measurement Uncertainty for Life Sciences Testing Laboratories for Category III methods. Laboratory control sample (LCS) results may be used to estimate Measurement Uncertainty (MU), provided the samples are an appropriate matrix and concentration. 1. When the LCS has been through all method steps, then the laboratory can use the pooled standard deviation (SDP) from the LCS intermediate precision data as an estimate of combined standard uncertainty. A relative SD or Coefficient of Variation (CV) may also be used. 2. When the LCS have not been run through all method steps, then the laboratory should incorporate any appropriate additional components or considerations in the uncertainty calculations, for example, those uncertainty components from sub-sampling, aliquoting or sample preparation. The additional components should be combined with SDP using the root sum square (RSS) method. 3. When a method has a known consistent bias that is inherent to the method (e.g. low recovery on difficult analytes) the bias must not be added to the uncertainty calculations. The bias shall, however, be clearly stated and recorded along with the uncertainty estimate. If a bias adjustment is made prior to reporting a result (e.g., adjusting for recovery on a sample that is spiked with a known amount of substance), then an additional source of uncertainty is introduced and must be included in the uncertainty estimate. However, if LCS data routinely include adjustments for recovery, then the error from the adjustment is already included in SDP and does not need to be added again. It is recommended that 20 or more individual LCS data points be obtained to estimate SDP. The estimate of combined uncertainty is then expanded using the formula: Measurement Uncertainty for a Defined Matrix (LCS) = k x SDP, where k (the coverage factor) equals 2 (for 95% confidence). If fewer than 20 LCS results are available, the coverage factor should be the appropriate t statistic for 95% confidence for the associated number of degrees of freedom (10=2.23, 20=2.09, 30=2.04, 40=2.02, 60=2.000, 120=1.98 & =1.96).
7 Page 7 of 12 Table 4: Laboratory Control Data with Same Target Values (e.g., 100 CFU) Raw Data (actual CFU Log 10 Value recovered) Estimating Uncertainty using the Standard Deviation (SD): Step 1 Transform the CFU values (column 1) to log base 10 value (log 10 ) (column 2). Step 2 Calculate the standard deviation of the log 10 values. The SD of this data set is Step 3 For reporting purposes, apply the coverage factor to the SD to obtain the expanded uncertainty. For 95% coverage we use a coverage factor of k=2. The expanded uncertainty in this example is thus If using the student t-tables, with n=20 (19 degrees of freedom), the coverage factor k would be 2.09, which would provide an expanded uncertainty of Step 4 To calculate the uncertainty for any subsequent laboratory result using SD x k, the result is first converted to log base 10 value (log 10 ), and the expanded uncertainty of is added and subtracted from the log value. Step 5 To estimate the MU of a sample result, calculate the uncertainty interval for the logarithmic value, by adding and subtracting the expanded uncertainty from the sample log (base 10) value; then convert the log value for the sample measurement back to CFU for the reported result. This is accomplished by taking the anti-log of each of the endpoints of the interval (anti-log of x = 10 x ). NOTE: By convention the results are always rounded down for the value at the lower end of the range and always rounded up for the value at the upper end. In this way the minimal 95% coverage is preserved.
8 Page 8 of 12 For example, estimating the uncertainty using SD x k for a result of 150 CFU: 150 in log 10 = Adding and subtracting from gives an interval from to ; transforming back to counts: =32.10, and = Therefore the uncertainty interval is 32 to 701 CFU. Estimating Uncertainty using the Relative Standard Deviation (RSD): Step 1 Transform the CFU values (column 1) to log base 10 value (log 10 ) (column 2). Step 2 Calculate the mean and standard deviation (SD) of the log 10 values. The SD of this data set is The mean of this dataset (log 10 ) is , or 77 CFU. This could also be expressed as a percentage SD (or relative SD), in log units: / = , or 17.8%; an expanded relative uncertainty is 2*0.1775=0.3550, or 35.5%. Step 3 For reporting purposes, apply the coverage factor to the SD to obtain the expanded uncertainty. For 95% coverage we use a coverage factor of k=2. The expanded uncertainty in this example is thus If using the student t-tables, with n=20 (19 degrees of freedom), the coverage factor k would be 2.09, which would provide an expanded uncertainty of Step 4 To estimate the uncertainty for any subsequent laboratory result using RSD x k, the result is first converted to the log base 10 value (log 10 ), multiplied by (35.5% converted to decimal) and then this expanded uncertainty is added and subtracted from the log value. NOTE: By convention the results are always rounded down for the value at the lower end of the range and always rounded up for the value at the upper end. In this way the minimal 95% coverage is preserved. For example, if estimating the uncertainty using RSD x k for a result of 150 CFU: 150 in log 10 = and the expanded uncertainty in log counts is times = Adding and subtracting from gives an interval from to ; transforming back to counts: =25.33, and = Therefore the uncertainty interval is 25 to 889 CFU. Note: When using relative standard deviation of data sets, the uncertainty value needs to be multiplied by the test result it is being applied to, as the value is being compared and relative to the data produced. When using standard deviation of data sets, the uncertainty value does not need to be multiplied by the test result it is being applied to. The expectation is that the uncertainty is not relative to the concentration of the sample, and the uncertainty would be the same across all concentrations for that method. Discussion of Example 3: Though the procedure used for estimating MU in Examples 1-3 are quite different, there are some similarities between the methods. Whether the replicates are derived from reproducibility replicates, as recovery replicates or as control data replicates, they all include the same sources of uncertainty listed in Table 3. Since the process in Example 3 only compares data over time, but not against another source as do the first two examples, it is expected that the MU would be larger using the formulas described in Example 3. Since all three procedures are valid, and no one process is favored over the other, it is up to
9 Page 9 of 12 the laboratory to determine whether the MU estimate obtained from any of these processes is reasonable and whether it meets the needs of its customers. Example 4 Use of Method Validation Data This example is a way by which the estimate of reproducibility from an interlaboratory method validation study can be used as an estimate of measurement uncertainty. If a laboratory can demonstrate competence with the method and if the uncertainty estimate is suitable for the laboratory s use, then no additional data need to be generated (other than to estimate repeatability and bias, which should be done anyway). The procedure comes from ISO 21748: Guidance for the use of repeatability, reproducibility and trueness estimates in measurement uncertainty estimation. The ISO document requires that the interlaboratory study to validate the method was conducted by competent laboratories using appropriate materials, and in accord with ISO : Accuracy (trueness and precision) of measurement methods and results Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method. This standard sets basic minimal numbers for laboratories, materials, and replicates, and has specific requirements for statistical analysis. Laboratories using measurement methods that are validated according to AOAC Harmonized Collaborative Laboratory Validation (HCLV), may also use this procedure for estimating uncertainty. The justification for this approach is that testing conditions in a single laboratory vary less over time than do conditions in different laboratories at a single point in time. Therefore, the estimate of reproducibility from a validation study provides a realistic over-estimate of uncertainty. If this estimate is acceptable for a laboratory then the reproducibility estimate from the study will suffice. If it is not suitable, then laboratories can generate their own estimates as they gain experience with the method. The procedure requires that laboratories demonstrate that their competence with the method is consistent with the laboratories in the validation study. This is accomplished with checks for bias and repeatability, as described below. Finally, this procedure assumes that the validation study included all parts of the measurement procedure, including (for example) sample preparation. If this last assumption is not satisfied then separate components can be added to the reproducibility, as discussed below. Suggested Protocol for Estimating Measurement Uncertainty using Statistics from a Validation Study 1. Assure there was an appropriate validation study design and data analysis (including outlier removal, statistical calculations, analysis for concentration effects, etc.), and that the estimates of Repeatability (S r ) and Reproducibility (S R ) are suitable for use in the laboratory. 2. Take the estimate of Reproducibility as a provisional estimate of measurement uncertainty (u ): u = S R. 3. Use the estimates of Reproducibility and Repeatability to calculate the between laboratory SD (S L ), as follows: S L = (S R 2 S r 2 ). 4. Estimate the laboratory bias (B L ) from repeated measurements of reference materials, comparison with a reference laboratory, or from proficiency testing: B L = (Laboratory Mean Reference Value). This should be calculated with log-transformed counts.
10 Page 10 of Estimate laboratory repeatability (S i ) from an internal study (see discussion above about replicates), which could have been done in the past. The study should be based on at least 10 replicates. Note 1: if the laboratory estimate of bias (B L ) is less than S L then step 6 may be skipped. If repeatability S i is smaller than S r, then step 7 may be skipped (but consider step 7c). 6. Calculate the acceptable criterion for the laboratory bias, as follows: Bias Limit = 2xS L If B L < Bias Limit then laboratory bias is acceptable for use of this procedure. If B L is larger than the limit then the procedure cannot be used. Check the reason for the bias and correct it if possible. 7. Calculate the acceptance criterion for repeatability: Precision Limit = 1.5xS r a. If S i < Precision Limit then repeatability is acceptable for this procedure b. If S i > Precision Limit then the procedure may still be used, but the provisional uncertainty estimate must be expanded as follows: u = (S L 2 + S i 2 ). c. If S i is much less than S r than the laboratory may wish to lower the provisional estimate of uncertainty using the same calculation in step 7.b, above. Note 2: ISO TS actually requires a statistical F test here, based on the number of results in the validation study and the number of results used by the laboratory to estimate repeatability. This requires statistical complexity that is beyond this simplified description. For the minimum number of results in an acceptable study, and at least 10 true replicates in the repeatability study, the Precision Limit in this step is the tightest criterion that would be calculated. 8. Add in any components of variability (S a1, S a2, etc.) that were not included in the validation experiment, such as subsampling or sample preparation. There might be more than one additional component. Add the additional component(s) to the provisional estimate (u ) to produce the final estimate of combined standard uncertainty (u). u = (u 2 + S a1 2 + S a2 2 ) 9. Calculate expanded uncertainty (U) with 95% coverage and k = 2, as follows: U = 2xu Note 3: If the uncertainty estimate is a percentage, the actual uncertainty will need to be calculated for every sample, as appropriate for its level. Note 4: Endpoints of the uncertainty interval are calculated with log 10 values and transformed back to CFU. Example: Using AOAC method : Aerobic Plate Count data This method was validated by a study that used 8 laboratories, 6 foods with different levels of contamination, 2 samples per food, and 2 replicates per sample. The data analysis was consistent with ISO , and the validation study included all steps in the testing process, except the step to choose
11 Page 11 of 12 an exact sub-sample size (measured samples were provided). Following are the reported estimates of Repeatability and Reproducibility for three of the foods, given as percentages: Food Reproducibility S R Repeatability S r Shrimp 11.1 % 9.8 % Vegetables 9.2 % 6.3 % Flour 5.8 % 5.3 % Calculations described above in Steps 2, 3, 6 and 7 are used to generate u, S L, Bias Limit and Precision Limit Food Provisional Uncertainty (u ) Between Lab SD (S L ) Bias Limit (2S L ) Precision Limit (1.5S r ) Shrimp 11.1 % 5.2 % 10.4 % 14.7 % Vegetables 9.2 % 6.7 % 13.4 % 9.4 % Flour 5.8 % 2.4 % 4.7% 7.9 % To estimate bias (Step 4), assume that the laboratory does a comparison study with a reference laboratory and results for vegetables and shrimp are always within 10% (B L < 10.0 %). The comparison with a flour sample shows results 5% apart (B L < 5.0%), so bias is judged to be acceptable. To estimate repeatability (Step 5) the laboratory generates estimates with a series of 10 replicates, and repeatability for all foods is 5% or less (S i < 5.0 %). It is decided, therefore, that repeatability is acceptable, but lower estimates of provisional uncertainty can be calculated, as described in Step 7c (using the formula in Step 7b): Food Initial Provisional Uncertainty (u ) Laboratory SD (S L ) Repeatability (S i ) Final Provisional Uncertainty (u ) Shrimp 11.1 % 5.2 % 5.0% 7.2 % Vegetables 9.2 % 6.7 % 5.0% 8.4 % Flour 5.8 % 2.4 % 5.0 % 5.6 % In consideration of additional components (Step 8), we assume that sample preparation (sub-sampling, weighing) has been estimated (or is suspected) to add an additional 3.0% to the uncertainty (based on expert opinion). This component is then added as described in Step 8, and the Final Uncertainty is expanded as described in Step 9. Food Final provisional uncertainty (u ) Adjusted final uncertainty (u) Expanded Uncertainty (U) Shrimp 7.2 % 7.8% 15.6% Vegetables 8.4 % 8.9% 17.8% Flour 5.6 % 6.4% 12.8% Calculate uncertainty intervals for samples at 150 CFU as an example of how this is accomplished. The steps for this are described in procedures 1 and 2 above.
12 Page 12 of 12 Food Expanded Uncertainty (U) Log 10 of 150 CFU Uncertainty in log 10 Uncertainty interval in log 10 Final uncertainty interval in CFU Shrimp 15.6% to to 328 Vegetables 17.8% to to 366 Flour 12.8% to to 285 General considerations for estimates of uncertainty Sampling is considered to be a large source of uncertainty in microbiological results. However it is not considered by any of the procedures in this document. This is because for most measurement procedures, sampling is a separate process, and error due to sampling is usually not considered to be part of the uncertainty in the measurement. Some procedures include sampling as a part or measurement uncertainty, but most procedures do not. Sampling error is best studied separately, for purposes of control of that important source. Estimates of uncertainty should always be confirmed if possible and the primary means is through experienced judgment:. The calculated uncertainty intervals should agree with expert experience for typical samples. Proficiency testing can also provide useful confirmation of uncertainty estimates. Any result graded as unacceptable should be investigated to see if it is also outside calculated uncertainty interval, and then investigate appropriately. Contributors: A2LA wishes to acknowledge the efforts of Dawn Mettler and Dan Tholen in the development and editorial input of this document by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written permission of A2LA. Document Revision History Date Description July 17, 2012 Changed calculation step 2 of procedure 2 on page 3 September 3, 2014 Added use of student t-table in in step 8 of example 1 and step 4 p of example 2, added new example 3 on use of lab control sample data, some text editing for clarity.
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample
General and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Principles of Microbiological Testing: Methodological Concepts, Classes and Considerations
Principles of Microbiological Testing: Methodological Concepts, Classes and Considerations Russell S. Flowers Silliker Group Corp. Relating Microbiological Testing and Microbiological Criteria to Public
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014
Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
Appendix H: Probability of Detection (POD) as a Statistical Model for the Validation of Qualitative Methods
Appendix H: Probability of Detection (POD) as a Statistical Model for the Validation of Qualitative Methods European validation organizations utilize ISO standard 16140:003(E), Protocol for the Validation
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Determining Measurement Uncertainty for Dimensional Measurements
Determining Measurement Uncertainty for Dimensional Measurements The purpose of any measurement activity is to determine or quantify the size, location or amount of an object, substance or physical parameter
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
Assessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Testing Surface Disinfectants
Testing Surface Disinfectants This series of knowledge sharing articles is a project of the Standardized Biofilm Methods Laboratory in the CBE KSA-SM-06 Enumerating viable cells by pooling counts for several
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Standard Deviation Estimator
CSS.com Chapter 905 Standard Deviation Estimator Introduction Even though it is not of primary interest, an estimate of the standard deviation (SD) is needed when calculating the power or sample size of
Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs
Building Credibility: Quality Assurance & Quality Control for Volunteer Monitoring Programs Elizabeth Herron URI Watershed Watch/ National Facilitation of CSREES Volunteer Monitoring Ingrid Harrald Cook
Biostatistics: DESCRIPTIVE STATISTICS: 2, VARIABILITY
Biostatistics: DESCRIPTIVE STATISTICS: 2, VARIABILITY 1. Introduction Besides arriving at an appropriate expression of an average or consensus value for observations of a population, it is important to
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program
Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
Design of Experiments for Analytical Method Development and Validation
Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Calculation example mean, median, midrange, mode, variance, and standard deviation for raw and grouped data
Calculation example mean, median, midrange, mode, variance, and standard deviation for raw and grouped data Raw data: 7, 8, 6, 3, 5, 5, 1, 6, 4, 10 Sorted data: 1, 3, 4, 5, 5, 6, 6, 7, 8, 10 Number of
Estimating Uncertainties in Testing
Estimating Uncertainties in Testing An Intermediate Guide to Estimating and Reporting Uncertainty of Measurement in Testing Keith Birch British Measurement and Testing Association At least two numbers
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
NCSS Statistical Software Principal Components Regression. In ordinary least squares, the regression coefficients are estimated using the formula ( )
Chapter 340 Principal Components Regression Introduction is a technique for analyzing multiple regression data that suffer from multicollinearity. When multicollinearity occurs, least squares estimates
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
NABL NATIONAL ACCREDITATION
NABL 162 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDELINES FOR PROFICIENCY TESTING PROGRAM for TESTING AND CALIBRATION LABORATORIES ISSUE NO : 04 AMENDMENT NO : 00 ISSUE
Assay Development and Method Validation Essentials
Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
ESTIMATION OF THE EFFECTIVE DEGREES OF FREEDOM IN T-TYPE TESTS FOR COMPLEX DATA
m ESTIMATION OF THE EFFECTIVE DEGREES OF FREEDOM IN T-TYPE TESTS FOR COMPLEX DATA Jiahe Qian, Educational Testing Service Rosedale Road, MS 02-T, Princeton, NJ 08541 Key Words" Complex sampling, NAEP data,
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Northumberland Knowledge
Northumberland Knowledge Know Guide How to Analyse Data - November 2012 - This page has been left blank 2 About this guide The Know Guides are a suite of documents that provide useful information about
Estimating Weighing Uncertainty From Balance Data Sheet Specifications
Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement
USP Certificate PREDNISONE TABLETS. USP Catalog No.: 1559505 USP Lot No.: R031Y0 (10 mg nominal prednisone content per tablet)
Certificate PREDNISONE TABLETS USP Catalog No.: 1559505 USP Lot No.: R031Y0 (10 mg nominal prednisone content per tablet) FOR DISSOLUTION PERFORMANCE VERIFICATION TEST (PVT) Period of validity: This certificate
Figure 1. A typical Laboratory Thermometer graduated in C.
SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES
Quantitative Inventory Uncertainty
Quantitative Inventory Uncertainty It is a requirement in the Product Standard and a recommendation in the Value Chain (Scope 3) Standard that companies perform and report qualitative uncertainty. This
Validating Methods using Waters Empower TM 2 Method. Validation. Manager
Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information
Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods
Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness
Quantifying Uncertainty in Analytical Measurement
EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison
IAEA-TECDOC-1350. Development and use of reference materials and quality control materials
IAEA-TECDOC-1350 Development and use of reference materials and quality control materials April 2003 The originating Section of this publication in the IAEA was: Industrial Applications and Chemistry Section
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
PROCEDURES FOR HANDLING OOS RESULTS
SOP # A-195--0100 STANDARD OPERATING PROCEDURES Page: 1 of 6. PROCEDURES FOR HANDLING OOS RESULTS 1. PURPOSE The purpose of this Standard Operation Procedure is to establish a procedure for the routine
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT
SPECIFICATIONS AND CONTROL TESTS ON THE FINISHED PRODUCT Guideline Title Specifications and Control Tests on the Finished Product Legislative basis Directive 75/318/EEC as amended Date of first adoption
Statistics I for QBIC. Contents and Objectives. Chapters 1 7. Revised: August 2013
Statistics I for QBIC Text Book: Biostatistics, 10 th edition, by Daniel & Cross Contents and Objectives Chapters 1 7 Revised: August 2013 Chapter 1: Nature of Statistics (sections 1.1-1.6) Objectives
NCSS Statistical Software
Chapter 06 Introduction This procedure provides several reports for the comparison of two distributions, including confidence intervals for the difference in means, two-sample t-tests, the z-test, the
Simple Random Sampling
Source: Frerichs, R.R. Rapid Surveys (unpublished), 2008. NOT FOR COMMERCIAL DISTRIBUTION 3 Simple Random Sampling 3.1 INTRODUCTION Everyone mentions simple random sampling, but few use this method for
CALCULATIONS & STATISTICS
CALCULATIONS & STATISTICS CALCULATION OF SCORES Conversion of 1-5 scale to 0-100 scores When you look at your report, you will notice that the scores are reported on a 0-100 scale, even though respondents
Magruder Statistics & Data Analysis
Magruder Statistics & Data Analysis Caution: There will be Equations! Based Closely On: Program Model The International Harmonized Protocol for the Proficiency Testing of Analytical Laboratories, 2006
ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:
How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated
Simple Regression Theory II 2010 Samuel L. Baker
SIMPLE REGRESSION THEORY II 1 Simple Regression Theory II 2010 Samuel L. Baker Assessing how good the regression equation is likely to be Assignment 1A gets into drawing inferences about how close the
Solving Exponential Equations
Solving Exponential Equations Deciding How to Solve Exponential Equations When asked to solve an exponential equation such as x + 6 = or x = 18, the first thing we need to do is to decide which way is
ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits
ANALYTICAL DETECTION LIMIT GUIDANCE & Laboratory Guide for Determining Method Detection Limits Wisconsin Department of Natural Resources Laboratory Certification Program April 1996 PUBL-TS-056-96 introduction
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
How to Conduct the Method Validation with a New IPT (In-Process Testing) Method
How to Conduct the Method Validation with a New IPT (In-Process Testing) Method Weifeng Frank Zhang QA Engineer BMS Ernest Lee Senior Manager, Facilities & Engineering Medarex, a fully owned subsidiary
DATA QUALITY OBJECTIVES
The Wastewater Treatment Plant Operators Guide to Biosolids Sampling Plans CHAPTER 5 DATA QUALITY OBJECTIVES Goals of the Sampling Plan ESSENTIAL ELEMENTS OF A SAMPLING PLAN Description of the Facility
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 1 TABLE OF CONTENTS FOREWORD... 1 1.0 SCOPE... 1 1.5 Safety Requirements... 1 4.0 MANAGEMENT REQUIREMENTS... 1 4.1
R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES
Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007
Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140
Penalized regression: Introduction
Penalized regression: Introduction Patrick Breheny August 30 Patrick Breheny BST 764: Applied Statistical Modeling 1/19 Maximum likelihood Much of 20th-century statistics dealt with maximum likelihood
DATA QUALITY ASSESSMENT AND DATA USABILITY EVALUATION TECHNICAL GUIDANCE
New Jersey Department of Environmental Protection Site Remediation Program DATA QUALITY ASSESSMENT AND DATA USABILITY EVALUATION TECHNICAL GUIDANCE Version 1.0 April 2014 Preamble The results of analyses
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting Method Number: ELAP method code 7037 SOP Number: Revision Number: SOP Date: Personnel / Data
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
Missing Data: Part 1 What to Do? Carol B. Thompson Johns Hopkins Biostatistics Center SON Brown Bag 3/20/13
Missing Data: Part 1 What to Do? Carol B. Thompson Johns Hopkins Biostatistics Center SON Brown Bag 3/20/13 Overview Missingness and impact on statistical analysis Missing data assumptions/mechanisms Conventional
Quantifying Bacterial Concentration using a Calibrated Growth Curve
BTEC 4200 Lab 2. Quantifying Bacterial Concentration using a Calibrated Growth Curve Background and References Bacterial concentration can be measured by several methods, all of which you have studied
The replication of empirical research is a critical
RESEARCH TECHNICAL COMMENT PSYCHOLOGY Comment on Estimating the reproducibility of psychological science Daniel T. Gilbert, 1 * Gary King, 1 Stephen Pettigrew, 1 Timothy D. Wilson 2 A paper from the Open
Annex 6 BEST PRACTICE EXAMPLES FOCUSING ON SAMPLE SIZE AND RELIABILITY CALCULATIONS AND SAMPLING FOR VALIDATION/VERIFICATION. (Version 01.
Page 1 BEST PRACTICE EXAMPLES FOCUSING ON SAMPLE SIZE AND RELIABILITY CALCULATIONS AND SAMPLING FOR VALIDATION/VERIFICATION (Version 01.1) I. Introduction 1. The clean development mechanism (CDM) Executive
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
Evaluation & Validation: Credibility: Evaluating what has been learned
Evaluation & Validation: Credibility: Evaluating what has been learned How predictive is a learned model? How can we evaluate a model Test the model Statistical tests Considerations in evaluating a Model
Chapter 4 and 5 solutions
Chapter 4 and 5 solutions 4.4. Three different washing solutions are being compared to study their effectiveness in retarding bacteria growth in five gallon milk containers. The analysis is done in a laboratory,
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
Measurement Information Model
mcgarry02.qxd 9/7/01 1:27 PM Page 13 2 Information Model This chapter describes one of the fundamental measurement concepts of Practical Software, the Information Model. The Information Model provides
Data Analysis Tools. Tools for Summarizing Data
Data Analysis Tools This section of the notes is meant to introduce you to many of the tools that are provided by Excel under the Tools/Data Analysis menu item. If your computer does not have that tool
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
