SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
|
|
|
- Alexina Marjorie Pope
- 9 years ago
- Views:
Transcription
1 WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory requirements as well as those of accreditation organizations such as the College of American Pathologists (CAP) and Joint Commission on Accreditation of HealthCare Organizations (JCAHO). LifeScan does not require linearity or AMR testing and the decision to perform this testing should be based on current accreditation and regulatory requirements and your institution s policies. Advantages of SureStep Pro Linearity Solutions SureStep Pro Linearity solutions are appropriate for SureStep Pro and SureStep Flexx meters in that they have test strip lot specific target values that minimize misinterpretation due to test sample matrix. Additionally, they require no sample preparation, have been manufactured so that the glucose is in a stable matrix, and contain five glucose levels that cover a wide concentration range. In addition to linearity testing, SureStep Pro linearity solutions may be used to verify the AMR. The linearity target values may not cover the entire AMR of the method (e.g., 0 to 500 mg/dl) and it is up to each institution to assess if the target values for the low (Level 1) and high (Level 5) linearity solutions meet your institution s needs. These solutions can also be used, depending on your needs, to assess precision, recovery, and use error (competency testing). Using SureStep Pro Linearity solutions to interpret and document the AMR Before using these solutions, assure that the operator performing the test is thoroughly familiar with the correct operation of the system(s) in use and refer to the package insert instructions for limitations and correct handling of these solutions. At least three levels must be used for verifying or validating the AMR (low, mid, and high). 1. Test each level in replicate (a minimum of 2) and calculate the mean value for each meter in use (initial verification). 2. For DataLink Control users, use the Linearity Report option in DataLink Control if desired. (See Appendix A) 3. For those that prefer to document the results manually, record the results on Form A (Form B at altitudes >2000 feet). Note: Use of this form is optional. 4. Compare the results to the acceptance range. If acceptable, document and sign the form. Retain records. If unacceptable, refer to the package insert for recommended actions to take or contact LifeScan customer support. criteria for LifeScan Linearity Solutions are for level 1 and for levels 2-5. The following example provided illustrates how these data may be documented. LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 1 of 6
2 Example: Acceptable Performance All values within acceptable range Target (Target ± ) Y Y Y Y Y Example: Unacceptable Performance Level 1 value is not within acceptable range Target (Target ± ) N Y Y Y Y For those institutions that require a more robust assessment for testing the linearity of a method, the Clinical Laboratory Standards Institute (CLSI previously referred to as the National Committee for Clinical Laboratory Standards or NCCLS) has published voluntary guidelines that provide a number of statistical approaches that may be used. IMPORTANT: While this represents what we believe is the proper interpretation of the requirements, each laboratory is responsible for requirements and checklists specific to its accrediting agency. Individual states may have additional requirements. Local authorities should be consulted as determined by the laboratory director to determine compliance with state/local regulations. This summary reflects information available as of April 2007 and may be superseded by newer regulations and requirements. LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 2 of 6
3 References 1 College of American Pathologists. Commission on Laboratory Accreditation, Point-of-Care- Testing Checklist. 10/ Sidebottom RA et al. Glucose determinations in plasma and serum: Potential error related to increased hematocrit. Clinical Chemistry 1982;28(1): SureStep Pro Linearity Test Kit Package Insert AW A, 9/02 4 LifeScan Technical Bulletin Effect of Altitude on Linearity Testing Using the SureStep Pro Linearity Test Kit, AW , LifeScan Technical Bulletin Linearity AW A, 6/97 6 NCCLS. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. NCCLS document EP6-A (ISBN ). NCCLS, 940 West Valley Road, Suite 1400, Wayne, PA USA, 200 LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 3 of 6
4 Appendix A - Interpretive Guidelines for AMR Verification using the DataLink Linearity Report When using the DataLink Linearity report for interpreting the Analytical Measurement Range (AMR), refer to the last column in the bottom table. Acceptable criteria for SureStep Pro Linearity Test solutions are for level 1 and for level 2 5 when compared to lot-specific reference (target) values. Example: Acceptable Performance Linearity Report Regression Statistics Summary Table N Slope Intercept Correlation Coeff. SE Glucose Glucose Reference Value Linearity Level Test Strip Lot Number Test Strip Exp. Date Reference Value N Mean SD %CV % Difference From Reference Values A002 12/ % +11.8% A002 12/ % -10.7% A002 12/ % 3.8% A002 12/ % 5.2% A002 12/ % 11.7% Level 1: mean within + 12 mg/dl of Ref. value +5.7 mg/dl (Note: Manually calculate mg/dl difference for Level 1) Levels 2-5: mean within + 20% of Ref. value LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 4 of 6
5 SureStep Pro Verification Record Form A Target (Target + ) Target (Target + ) Target (Target + ) Please make copies before use. LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 5 of 6
6 SureStep Pro Verification Record (For use at altitudes > 2000 feet above sea level ONLY) Form B Target Correction Factor New Target (Target + ) Target Correction Factor New Target (Target + ) Target Correction Factor New Target (Target + ) Please make copies before use. LifeScan, Inc. Milpitas, CA AW A 4/2007 Page 6 of 6
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD TABLE OF CONTENTS Introduction Applicability General Topics Proficiency Testing
Advanced Installation & Reference Guide. for OneTouch DataLink Data Management System and OneTouch DataLink Web
Advanced Installation & Reference Guide for OneTouch DataLink Data Management System and OneTouch DataLink Web Copyright 2010 LifeScan Inc. All rights reserved. No part of this publication may be reproduced,
Quality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD
SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision)
August 2012 Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision) This document provides guidance on how to verify comparability of quantitative
PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL
Quick Reference Guide
Quick Reference Guide Look inside for step-by-step instructions on how to: >> Transfer readings from your OneTouch Meter >> View and share any of the 11 available reports >> Add information to a patient
Quality Assurance for the Clinical Laboratory
EP Evaluator Quality Assurance for the Clinical Laboratory Simplified process, powerful results While we (clinical labs) are but a small component of the total healthcare system, our data control a large
Calibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK PROFICIENCY TESTING
FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK PROFICIENCY TESTING PRINCIPLE Laboratories may participate, by mandate or by choice, in external proficiency testing programs.
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM
For In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date
An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and
An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in
Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)
Topic: Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Click on the links below to be taken to a specific section of the FAQs. General
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25
system en Intact PTH 8K25 84-6434/R5 Intact PTH Read Highlighted Changes Revised November, 2008 Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.
Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care
Accuracy Test Omnitest Blood Glucose Monitoring System B. Braun Diabetes Care Accuracy Test Omnitest blood glucose monitoring system according to ISO 97 The accuracy study of Omnitest blood glucose monitoring
2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)-524-3270
SIEM EN S Abbreviated 510(k) Premnarket Notification IMMULITE 2000 PSA Calibration Verification Material 6. 510(k) Summary as Required by 21 CFR 807.92 K( - is-1 3 (v A. 5 10(k) Number: B. Purpose for
What Does Having a FDA Cleared Pregnancy Test Mean?
What Does Having a FDA Cleared Pregnancy Test Mean? Clinical Trials Transformation Initiative (CTTI) July 15-16, 2013 Denise N. Johnson-Lyles, Ph.D. Toxicology Branch Chief Division of Chemistry and Toxicology
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Biochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests
2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren
Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA
Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO
Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2
Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
General. What is a laboratory test? What is a waived test?
General What is a laboratory test? What is a waived test? Laboratory tests are medical procedures that involve testing samples of blood, urine, or other tissues or substances from the body. Information
SAMPLE. Quality Management System: A Model for Laboratory Services; Approved Guideline Fourth Edition
June 2011 Quality Management System: A Model for Laboratory Services; Approved Guideline Fourth Edition This document provides a model for medical laboratories that will assist with implementation and
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
ADMINISTRATIVE MANUAL Policy and Procedure
ADMINISTRATIVE MANUAL Policy and Procedure TITLE: Point of Care Testing NUMBER: CH 30-111 (Laboratory Diagnostic Bedside Testing) Effective Date: January 2014 Page 1 of 6 Applies To: Holders of Administrative
Blood Glucose Monitoring: The Facts about Accuracy
Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM
Inform II BLOOD GLUCOSE MONITORING SYSTEM Quick Reference Guide for Healthcare Professionals 2 Table of Contents Important... 4 General Information... 6 ACCU-CHEK Inform II Meter...10 How to Perform Patient
SAMPLE. Training and Competence Assessment; Approved Guideline Third Edition
February 2009 Training and Competence Assessment; Approved Guideline Third Edition This document provides background information and recommended processes for the development of training and competence
CLSI TRAINING COURSES CATALOG
CLSI TRAINING COURSES CATALOG A Laboratory Development Resource CLINICAL AND LABORATORY STANDARDS INSTITUTE TRAINING COURSES CATALOG The Clinical and Laboratory Standards Institute (CLSI) is known for
OmniPod. Extension. for the CoPilot Health Management System. Version 1.1. For use with PDM model UST200
OmniPod Extension for the CoPilot Health Management System Version 1.1 For use with PDM model UST200 OmniPod Extension for CoPilot Quick Start Guide 1 Quick Start Guide Use the OmniPod Extension to download
College of American Pathologists Accreditation for Public Health Laboratories
College of American Pathologists Accreditation for Public Health Laboratories Romesh Gautom, PhD Director, Washington State Public Health Laboratory Garry McKee, PhD, MPH Director, Oklahoma State Public
SMF Awareness Seminar 2014
SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD
Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma
Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use
Quantifying measurement error from digital instruments
Quantifying measurement error from digital instruments W. BLAKE LAING AND SEAN BRYANT SOUTHERN ADVENTIST UNIVERSITY CHAT TANOOGA, TN What I m doing HELPING STUDENTS LEARN TO CONSTRUCT KNOWLEDGE First lab:
How To Test For Creatinine
Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified
Radioanalytical Data Validation Issues and Solutions
Radioanalytical Data Validation Issues and Solutions Presented by: Terry McKibbin and Kevin Harmon Validata Chemical Services, Inc. Purpose of Data Validation The Presentation is Based on this Goal for
Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff
Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff GUIDANCE guidance. Submit written comments to the Division of
XV.Quality Assurance in the Blood Bank
XV.Quality Assurance in the Blood Bank A. Overview 1. Goals 2. Terms B. Record Keeping a. Safe transfusion b. Careful adherence to SOPs by trained personnel c. Develop comprehensive guidelines to be in
US Diabetic Devices Industry Research Report: KenResearch
US Diabetic Devices Industry Research Report: KenResearch by KenResearch - Monday, February 24, 2014 https://www.kenresearch.com/blog/2014/02/us-diabetic-devices-industry-research-report-kenresearch/ The
Laboratory Computer Systems
LAP Audioconference Laboratory Computer Systems October 21, 2009 Bruce A. Beckwith, MD, FCAP 2009 College of American Pathologists. Materials are used with the permission of the faculty Learning Objectives
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
Sentinel I28 and Blackbelt Mass Flow Calibration and Verification
Service Bulletin: #001 Date: August 2012 Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Summary This Application Bulletin will give a short synopsis of the purpose of a calibration and
Blood Glucose Monitoring System USER GUIDE. Replaces Owner s Booklet
Blood Glucose Monitoring System USER GUIDE Replaces Owner s Booklet AW 06650201A Rev. date: 07/2009 Contents: Getting to know your system 1 Setting up your meter 4 Testing your blood glucose 8 Attaching
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff
Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff Document issued on: April 25, 2013 The draft of this document was issued on January 5, 2009. For questions regarding
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories () Edited by Richard Pang, PhD, FACB Senior Consultant Pro Q Consulting Services Hong Kong For Hong Kong
SAMPLE. Effects of Different Sample Types on Glucose Measurements
1st Edition POCT06 Effects of Different Sample Types on Glucose Measurements This report provides information to assist the clinical and point-ofcare staff in result and measurement procedure comparisons
2009 LAP Audioconference Series. Laboratory Computer Systems
2009 LAP Audioconference Series Objectives: After participating in this audioconference you will be able to: Describe the requirements for validating computer interfaces Understand the principles for validating
CLIA & Individualized Quality Control Plan (IQCP)
& Individualized Quality Control Plan (IQCP) Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services Centers for Medicare & Medicaid Services 1 Objectives Provide Background & History of
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
Who gets you to the next level?
Who gets you to the next level? Ideal OEM-partner for Global Corporations and Start-ups, founded in 1961, is part of. LRE Medical is a leading developer and manufacturer of analytical instrumentation for
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical
Air Quality Performance Test Guidelines. Kansas Department of Health and Environment Bureau of Air and Radiation
Air Quality Performance Test Guidelines Kansas Department of Health and Environment Bureau of Air and Radiation December 22, 2004 TABLE OF CONTENTS 1. INTRODUCTION 2. TEST PLANNING 2.1 EPA Test Methods
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE A. 510(k) Number: K041502 B. Purpose for Submission: To change the name of a previously cleared analyzer, and
Center for Clinical Standards and Quality/Survey & Certification Group
DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Abstract... i. Committee Membership... iii. Foreword... vii. 1 Scope... 1. 2 Introduction... 1. 3 Standard Precautions... 1. 4 Definitions...
Vol. 27 No. 26 Replaces H3-A5 Vol. 23 No. 32 Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition This document provides procedures for the collection
PRINCIPLE REF 475136. ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Rheumatoid Factor REF 475136 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
Experiment #1, Analyze Data using Excel, Calculator and Graphs.
Physics 182 - Fall 2014 - Experiment #1 1 Experiment #1, Analyze Data using Excel, Calculator and Graphs. 1 Purpose (5 Points, Including Title. Points apply to your lab report.) Before we start measuring
This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System.
CREATINE KINASE 7D63-20 and 7D63-30 30-3938/R4 CREATINE KINASE This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System. NOTE: Changes
College of American Pathologists Laboratory Accreditation Program. Document Control Management
College of American Pathologists Laboratory Accreditation Program Document Control Management William Castellani, MD, FCAP Inter-Regional Commissioner Laboratory Accreditation Program September 19, 2007
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
Blood Glucose Monitoring System USER GUIDE
Blood Glucose Monitoring System USER GUIDE AW 06629001B Rev. date: 07/2009 Contents: Getting to know your system 1 Setting the time and date, and coding your meter 4 Testing your blood glucose 8 Reviewing
APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN
APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN This outline is recommended for use by UST consultants/contractors in preparing a generic Quality Assurance Project Plan (QAPP) for use
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED Linda C. Bruno, M.A., MT(ASCP) Director, Microbiology and Molecular Labs ACL Laboratories, Rosemont, IL June 1, 2015 Disclosures - No disclosures 2 IQCP
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:
ETV Joint Verification Statement
THE ENVIRONMENTAL TECHNOLOGY VERIFICATION PROGRAM U.S. Environmental Protection Agency ETV Joint Verification Statement TECHNOLOGY TYPE: ON-BOARD EMISSIONS MONITOR APPLICATION: MEASURING VEHICLE EXHAUST
Content Sheet 16-1: Introduction to Documents & Records
Content Sheet 16-1: Introduction to Documents & Records Role in quality management system The management of documents and records is one of the 12 essential elements of the quality system. The management
Unique Standards and Documentation Required for Accredited CLT/MLT Programs
Unique Standards and Documentation Required for Accredited CLT/MLT Programs Section III (CLT/MLT) Page 1 UNIQUE STANDARDS AND THE REQUIRED DOCUMENTATION Objectives Clinical Laboratory Technician/Medical
General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations
General Information Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations Our Background Foundation Laboratory is an advanced, accredited and certified clinical diagnostic
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate When is a CLIA Certificate Required? NOTE: Congress passed the Clinical Laboratory Improvement Amendments (CLIA) in 1988
Accu-Chek Inform II system. Professional glucose testing for the wireless age
Accu-Chek Inform II system Professional glucose testing for the wireless age The Accu-Chek Inform II system Real time control The Accu-Chek Inform II System overview Quality Advanced chemistry with multiple
PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are
1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96
1 Final Review 2 Review 2.1 CI 1-propZint Scenario 1 A TV manufacturer claims in its warranty brochure that in the past not more than 10 percent of its TV sets needed any repair during the first two years
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
QA Transactions and Reports
QA Transactions and Reports Monitoring Quality Assurance QA Transactions & Reports National Ambient Air Monitoring Conference August 2014 1 Background The QA requirements for ambient air quality monitoring