Biochemistry Validation Form
|
|
|
- Cornelius Curtis Phelps
- 9 years ago
- Views:
Transcription
1 Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August 2012) Evaluation performed by: Patricia Lockett, Senior Biomedical Scientist for Point of Care Testing & Caroline Addison, Principal Clinical Biochemist Date of Evaluation: August 2012 Introduction Background The point of care team was contacted by A&E with regard to increasing the point of care test repertoire in their department. The initial requirements were U&Es, LFTs, paracetamol & salicylate, Troponin T, BNP, INR, D-Dimer, CRP, hcg and amylase. The department agreed to look for suitable analysers. The Abaxis Piccolo Xpress chemistry analyser is a point of care device that delivers clinical chemistry profiles. Profiles are available in disks and include lipid, liver enzymes, kidney function, glucose and electrolytes (see picture 1). It was decided to evaluate this instrument for the A&E department in particular the Chemistry 13 reagent disk which offers albumin, ALP, ALT, amylase, AST, urea, creatinine, calcium, GGT, glucose, total bilirubin, total protein and urate. Sample type is heparinised whole blood, heparinised plasma or serum. Rationale for evaluation See background. Author: S Troup Page 1 of 26 Date Approved: 05/04/12
2 Picture 1: Piccolo Xpress chemistry analyser and reagent disk. Method Current New Method Various Chemistry 13 reagent disk Analyser Roche cobas c701 Abaxis Piccolo Xpress A limited number of reagent disks (50) was available for the evaluation. Patient comparison was performed by taking 10 random serum samples and analysing on both the Piccolo Xpress and laboratory analyser. A further 10 serum samples were then selected to ensure the measuring range had been evaluated (where possible). Comparison of lithium heparin whole blood versus lithium heparin plasma was also performed. Patient samples which had been sent to the laboratory in lithium heparin sample tubes (i.e. for ketone analysis) were analysed as whole blood on the Piccolo Xpress. These samples were then centrifuged to obtain the plasma, which was then analysed for all parameters on the Roche cobas c701. Due to a limited number of reagent disks this could only be performed for 8 patient samples. Precision was determined by analysing the 2 levels of Randox Chemistry internal quality control material which is used in the laboratory. Within day precision was assessed by measuring both levels of QC 5 times in one day. Between batch precision was assessed over a further 2 days, by measuring both levels of QC 3 times each. Review of kit insert Interferences/cross reactivities Samples containing EDTA, fluoride, oxalate or any anticoagulant containing ammonium ions will interfere with at least 1 chemistry in the chemistry 13 reagent disc Sodium heparin is not suitable. Samples with haematocrits in excess of 62-65% packed red cell volume (volume fraction of ) may give an inaccurate result Author: S Troup Page 2 of 26 Date Approved: 05/04/12
3 these samples may be reported as haemolysed and can be re-spun and the plasma analysed. Haemolysis, icterus and lipaemia can cause changes in reported concentrations of some analytes. The Piccolo Xpress suppresses any results that are affected by >10% interference from haemolysis, lipaemia or icterus. Table 1 shows substances with >10% interference and their effect on the chemistry 13 test(s). Refrigerating whole blood samples can cause significant changes in concentrations of AST, creatinine and glucose. Total bilirubin may be adversely affected by photodegradation. Samples not run immediately should be stored in the dark. Glucose concentrations decrease approximately 5-12mg/dl ( mmol/L) in 1 hour in uncentrifuged samples at room temperature. Linearity/dilutions Each analyte in the Chemistry 13 reagent disc is linear over the measuring range as shown in table 2. Samples above or below this range are reported as greater than (>) or less than (<) with an asterisk e.g. ALT >2000* U/L or ALT <5* U/L Any samples that exceed the assay range should be analysed by another approved method. DO NOT dilute the sample and re-run on the Piccolo Xpress analyser. Samples which are grossly beyond the measuring range will be printed as ~~~ instead of a result. Precision Results for within-run and total precision were determined by testing two levels of control material. Controls were run in duplicate twice each day for 20 days over a four week period. Precision data is shown in table 3. LOD/LOQ The lower limit of the reportable range for each analyte is shown in table 2. Sample type/handling Minimum required volume is 100µl of heparinised whole blood, heparinised plasma, serum or control material. The reagent disc chamber can contain up to 120µl of sample. Samples containing EDTA, fluoride, oxalate or any anticoagulant containing ammonium ions will interfere with at least 1 chemistry in the chemistry 13 reagent disc Sodium heparin is not suitable. Whole blood samples must be homogenous before transferring to a sample disc. Gently invert the collection tubes several times prior to inversion. DO NOT shake the tube as this may cause haemolysis. Author: S Troup Page 3 of 26 Date Approved: 05/04/12
4 Whole blood venepuncture samples should be run within 60minutes of collection. If analysis cannot be performed in this time, the sample may be separated into plasma or serum and stored at 2-8 C. The reagent disc should be analysed within 10 minutes of transferring the sample to the reagent disc. Reagent storage/hazards Store reagent discs in their sealed pouches at 2-8 C. Do not expose to direct sunlight or temperatures above 32 C. Do not use beyond the expiration date. Reagent discs may be used directly from the refrigerator without warming. Do not allow discs to remain at room temperature for more than 48hours. Be careful not to touch the barcode ring located on top of the disc. A disc not used within 20 minutes of opening should be discarded. The reagent discs are plastic and may crack or chip if dropped. Never use a dropped disc as this may spray biohazardous material throughout the interior of the analyser. Level sensing? Not applicable Chemistry 13 discs contain all reagents Other In rare circumstances sample when applied to the disc may not flow smoothly into the sample chamber. Due to an uneven flow, an inadequate quantity of sample may be analysed and several results fall outside of the reference range. Re-run sample using new reagent disc. Author: S Troup Page 4 of 26 Date Approved: 05/04/12
5 Table 1: Substances with significant interference (>10%) in the Chemistry 13 reagent disc on the Piccolo Xpress analyser Author: S Troup Page 5 of 26 Date Approved: 05/04/12
6 Table 2: Manufacturer measuring range for Chemistry 13 reagent disc on Piccolo Xpress analyser Table 3: Manuafacturer Total precision data for the chemistry 13 reagent disc on the Piccolo Xpress analyser % CV (mean concentration) Analyte Level 1 Level 2 ALT 13.5% (21 U/L) 6.2% (52 U/L) Albumin 2.1% (56 g/l) 2.9% (37g/L) ALP 5.8% (39 U/L) 3.1% (281 U/L) Amylase 5.7% (46 U/L) 3.8% (300 U/L) AST 1.9% (49 U/L) 1.2% (147 U/L) Calcium 2.9% (8.6mg/dl/2.15mmol/L) 3.4% (11.8mg/dl/2.94mmol/L) Creatinine 13.1% (1.1mg/dl/97µmol/L) 5.2% (5.2mg/dl/460µmol/L) GGT 2.94% (25 U/L) 2.15% (106 U/L) Glucose 1.6% (66 mg/dl/3.66mmol/l) 1.4% (278 mg/dl/15.43mmol/l) Total bilirubin 9.3% (0.8 mg/dl/14µmol/l) 2.8% (5.2mg/dl/89µmol/L) Total protein 1.2% (68 g/l) 2.0% (47 g/l) Urea 2.1% (19mg/dl/6.8mmol/L) Urate 4.8% (3.8 mg/dl/0.226mmol/l) 1.8% (65mg/dl/23.2mmol/L) 3.9% (7.5mg/dl/0.446mmol/L) Author: S Troup Page 6 of 26 Date Approved: 05/04/12
7 Instrument download/calculations Details of information downloaded if applicable Details of results calculations if applicable Performed by: not applicable Checked by:not applicable Date: Date: Results/Data N.B. ensure data use is raw data from analyser, not from middleware or LIMS, as this may be rounded. Bias o Patient comparison Figures 1-13 show the Passing & Bablok regression analysis for all analytes included in the chemistry 13 reagent disc. Albumin shows a constant negative bias on the Piccolo Xpress compared to the laboratory assay (y = 0.9x -0.5). This bias was not considered to be significant. A proportional negative bias was seen for ALP. Again this bias was not considered to be significant. Good comparison was seen between the Piccolo Xpress and the laboratory for ALT (y = ) A positive bias was seen for amylase on the Piccolo Xpress (y = 1.36x ). This bias may be attributable to the differences in what each assay is measuring. The laboratory assay is specific for pancreatic amylase whereas the Piccolo Xpress measures total amylase. A small constant positive bias was seen between the Piccolo Xpress and the laboratory for AST (y = 1.06x ) however, this was not considered to be significant. Good comparison was seen between the Piccolo Xpress and the laboratory for urea (y = 1.03x 0.35). A mixed bias was seen for calcium when comparing the Piccolo Xpress and the laboratory assay. This assay would identify patients with calcium Author: S Troup Page 7 of 26 Date Approved: 05/04/12
8 concentrations at the extremes and therefore, is probably sufficient for a point of care device. It should be noted that a new calcium assay is currently under evaluation by the department. This new calcium assay is reported to be more stable than the current assay. A repeat comparison if this new assay is installed would be beneficial. A constant negative bias was seen for Creatinine (y = 0.92x -7.00). Again this bias was not considered to be significant. Good comparison was seen between the Piccolo Xpress and the laboratory for GGT (y = 1.01x ). Good comparison was seen between the Piccolo Xpress and the laboratory for Glucose (y = 0.98x ). A mixed bias was seen for total bilirubin (y = 0.88x ). This bias was not considered to be significant. A small proportional positive bias was seen for total protein, again this was not considered to be significant. Urate showed a small constant negative bias, but was not considered to be significant. Figure 1: Passing and Bablok regression for Albumin Author: S Troup Page 8 of 26 Date Approved: 05/04/12
9 Figure 2: Passing and Bablok regression for ALP Figure 3: Passing and Bablok regression for ALT Author: S Troup Page 9 of 26 Date Approved: 05/04/12
10 Figure 4: Passing and Bablok regression for Amylase Figure 5: Passing and Bablok regression for AST Author: S Troup Page 10 of 26 Date Approved: 05/04/12
11 Figure 6: Passing and Bablok regression for Urea Figure 7: Passing and Bablok regression for Calcium Author: S Troup Page 11 of 26 Date Approved: 05/04/12
12 Figure 8: Passing and Bablok regression for Creatinine Figure 9: Passing and Bablok regression for GGT Author: S Troup Page 12 of 26 Date Approved: 05/04/12
13 Figure 10: Passing and Bablok regression for Glucose Figure 11: Passing and Bablok regression for Total bilirubin Author: S Troup Page 13 of 26 Date Approved: 05/04/12
14 Figure 12: Passing and Bablok regression for Total protein Figure 13: Passing and Bablok regression for Urate Author: S Troup Page 14 of 26 Date Approved: 05/04/12
15 o Patient comparison lithium heparin whole blood vs Lithium heparin plasma Patient samples which had been sent to the laboratory in lithium heparin sample tubes (i.e. for ketone analysis) were analysed as whole blood on the Piccolo Xpress. These samples were then centrifuged to obtain the plasma, which was then analysed for all parameters on the Roche cobas c701. Due to a limited number of reagent disks this could only be performed for 8 patient samples. Further comparison may be appropriate. Figures show the Passing & Bablok regression analysis for lithium heparin whole blood versus lithium heparin plasma for all analytes included in the chemistry 13 reagent disk. Good comparison was seen between lithium heparin whole blood and plasma samples for albumin (y = 0.97x ). Good comparison was seen between lithium heparin whole blood and plasma samples for ALP (y = 0.97x ). Good comparison was seen between lithium heparin whole blood and plasma samples for ALT (y = 0.97x ). A positive bias was seen for amylase on whole blood lithium heparin versus lithium heparin plasma. Again, this bias may be attributable to the differences in what each assay is measuring. The laboratory assay is specific for pancreatic amylase whereas the Piccolo Xpress measures total amylase. A small constant positive bias was seen between whole blood lithium heparin versus lithium heparin plasma for AST, however, this was not considered to be significant. Good comparison was seen between lithium heparin whole blood and plasma samples for urea (y = 1.03x 0.35). A mixed bias was seen for calcium on whole blood lithium heparin versus lithium heparin plasma. Further comparison of samples would be beneficial, especially if the more stable laboratory calcium assay is introduced. A negative bias was seen for Creatinine (y = 0.92x -7.00). Again this bias was not considered to be significant. Good comparison was seen between lithium heparin whole blood and plasma samples for GGT (y = 1.10x ). Author: S Troup Page 15 of 26 Date Approved: 05/04/12
16 Good comparison was seen between lithium heparin whole blood and plasma samples for Glucose (y = 0.98x ). A positive bias was seen for total bilirubin. This bias was not considered to be significant. A positive bias was seen for total protein, again this was not considered to be significant. Good comparison was seen between lithium heparin whole blood and plasma samples for Urate. Figure 14: Passing and Bablok regression for Albumin, lithium heparin whole blood vs lithium heparin Author: S Troup Page 16 of 26 Date Approved: 05/04/12
17 Figure 15: Passing and Bablok regression for ALP, lithium heparin whole blood vs lithium heparin Figure 16: Passing and Bablok regression for ALT, lithium heparin whole blood vs lithium heparin Author: S Troup Page 17 of 26 Date Approved: 05/04/12
18 Figure 17: Passing and Bablok regression for Amylase, lithium heparin whole blood vs lithium heparin Figure 18: Passing and Bablok regression for AST, lithium heparin whole blood vs lithium heparin Author: S Troup Page 18 of 26 Date Approved: 05/04/12
19 Figure 19: Passing and Bablok regression for Urea, lithium heparin whole blood vs lithium heparin Figure 20: Passing and Bablok regression for Calcium, lithium heparin whole blood vs lithium heparin Author: S Troup Page 19 of 26 Date Approved: 05/04/12
20 Figure 21: Passing and Bablok regression for Creatinine, lithium heparin whole blood vs lithium heparin Figure 22: Passing and Bablok regression for GGT, lithium heparin whole blood vs lithium heparin Author: S Troup Page 20 of 26 Date Approved: 05/04/12
21 Figure 23: Passing and Bablok regression for Glucose, lithium heparin whole blood vs lithium heparin Figure 24: Passing and Bablok regression for Total Bilirubin, lithium heparin whole blood vs lithium heparin Author: S Troup Page 21 of 26 Date Approved: 05/04/12
22 Figure 25: Passing and Bablok regression for Total protein, lithium heparin whole blood vs lithium heparin Figure 26: Passing and Bablok regression for Urate, lithium heparin whole blood vs lithium heparin Author: S Troup Page 22 of 26 Date Approved: 05/04/12
23 o EQA comparison Due to the number of reagent discs available, EQA performance was not assessed. o Dilution check Samples with results outside of the measuring range should not be diluted. Imprecision o Within run see between run o Between run Precision was found to be acceptable for all analytes (<5 %CV) except creatinine and ALT - see table 1 for data. The level 1 QC for creatinine had a mean concentration of 42.8µmol/L and gave a %CV of This is higher than is desirable even for a point of care instrument and has been fed back to the manufacturer. The level 1 QC for ALT had a mean concentration of 26.8 U/L and gave a %CV of This is probably acceptable precision for a point of care device. o Measurement uncertainty Measurement uncertainty is estimated as being the intermediate imprecision expressed as 1 SD. The intermediate imprecision is the imprecision of the analytical method over a period of time, taking into account changes in reagent lots and calibration lots. Cannot be determined at this time and to be determined at a later date. Other (if applicable, e.g. non-ce marked method, or change of use of CE marked method) Not applicable. o Limit of detection o Limit of quantitation o Linear range o Recovery o Interference o Analyte stability Author: S Troup Page 23 of 26 Date Approved: 05/04/12
24 Table 1: Precision data for all analytes in the chemistry 13 panel. Level 1 Level 2 Within Day Between Day Within Day Between Day Mean %CV %CV Mean %CV %CV Albumin ALP ALT Amylase AST Urea Calcium Creatinine GGT Glucose Total bilirubin Total Protein Urate Author: S Troup Page 24 of 26 Date Approved: 05/04/12
25 Cost implications Instrumentation & IT connectivity The capital cost of each instrument does not include interfacing of the Piccolo Xpress to the LIMS. Quality Control Randox Chemistry QC material which is currently used by the laboratory is suitable for use on the Piccolo Xpress. The cost of the QC material is QC costs are included in the laboratory cost per test. Turn around times This assay is intended to be performed in the A&E department. It is therefore expected for the turnaround time to improve. This does rely on suitably trained staff being available to perform the analysis as well as machine availability. The chemistry 12 panel takes approx. 12 minutes. Conclusion The Piccolo Xpress chemistry analyser and chemistry 13 reagent disk have been evaluated by the Point of Care department. The evaluation has shown that the assays are suitable, in terms of performance for use in the A&E department. Further work to be carried out include 1) Analysis of EQA material 2) Repeat comparison of calcium on the Piccolo versus the laboratory assay, particularly if a more stable calcium assay is introduced into the laboratory. 3) Repeat precision for creatinine and ALT. Funding to cover the costs of purchase, reagents, QC and maintenance needs to be identified. Author: S Troup Page 25 of 26 Date Approved: 05/04/12
26 Confirmation Method Evaluation Date of Introduction: Date Signature Staff Name (print): Departmental BMS 4 Name (print): Head of Department Name (print) Author: S Troup Page 26 of 26 Date Approved: 05/04/12
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD
PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL
For In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB
HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.
CTSC Clinical Laboratory. Clinical Laboratory Services CTSC. Clinical & Translational Science Center
Clinical Laboratory The Clinical Laboratory develops and carries out research-related sample analyses to UNM HSC investigators, researchers throughout the United States and world, as well as corporate
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
FULL LABORATORY to HOT LABORATORY DR. DAVID OLEESKY
FULL LABORATORY to HOT LABORATORY Remodelling the biochemistry service at Macclesfield District General Hospital within a hub and spoke managed network --------------------------------------------------------------------------------------------
Certificate of Analysis NFKK Reference Serum X Components in Human Serum, Lot Number: NFKK2002a
Scandinavian Society of Clinical Chemistry Certificate of Analysis NFKK Reference Serum X Components in Human Serum, Lot Number: NFKK2002a Version 7 Description of the material The NFKK Reference Serum
Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO
Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban
Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA
KAMIYA BIOMEDICAL COMPANY Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA For the quantitative determination of mouse CKMB in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-57681
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:
Liver Function Tests. Dr Stephen Butler Paediatric Advance Trainee TDHB
Liver Function Tests Dr Stephen Butler Paediatric Advance Trainee TDHB Introduction Case presentation What is the liver? Overview of tests used to measure liver function RJ 10 month old European girl
cobas c 111 analyzer Small box. Big performance
cobas c 111 analyzer Small box. Big performance His journey back to health begins in your laboratory cobas solutions making a difference when it matters most Every patient prognosis depends on several
Room Temp: 8 Hr. Room Temp: 8Hr. Room Temp: ASAP. Refrigerated (2-8C): 3D. Room Temp: ASAP. Refrigerated (2-8C): 3D
Diabetes Chem Panel (Glu, Trig, Chol, HDL, LDL, Chol:HDL Ratio, AST, ALT, BUN, Creat, GFR, Na, K, Cl, CO2, Ca) ANMC Lab Test Requirements Updated 4/7/15 Key: Room Temp (20-25C), Refrigerated (2-8C), (-15
Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.
K-ASSAY Mouse IgM ELISA For the quantitative determination of IgM in mouse biological samples Cat. No. KT-407 For Research Use Only. 1 Rev. 072309 K-ASSAY PRODUCT INFORMATION Mouse IgM ELISA Cat. No. KT-407
Human Free Testosterone(F-TESTO) ELISA Kit
Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be
IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.
IgM ELISA For the quantitative determination of IgM in human serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calibrator
ab83369 Alkaline Phosphatase Assay kit (Colorimetric)
ab83369 Alkaline Phosphatase Assay kit (Colorimetric) Instructions for use: For the rapid, sensitive and accurate measurement of Alkaline Phosphatase in various samples. This product is for research use
Aarhus University Hospital Validation Tempus600 Transport system for blood tube samples. Pediatric blood samples
Aarhus University Hospital Validation Tempus6 Transport system for blood tube samples Pediatric blood samples Department of Clinical Biochemistry, SKS January 213 Aarhus Universitetshospital Inflammationscenteret
How To Test For Creatinine
Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed
Pointe Scientific, Inc. Instrument Application
Test: ACID PHOS. (NP) Catalog # : A7503 TEST NAME: ACID PHOS. (NP) TEST CODE: [AP (NP)] UNITS: [U/L] PRECISION (DECIMAL): [0] Kinetic MATH MODEL: Fixed Factor FACTOR: 1318.00 [
SAMPLE PROCEDURE 1029-3, 09/11
SAMPLE PROCEDURE This Sample Procedure is not intended as a substitute for your facility s Procedure Manual or reagent labeling, but rather as a model for your use in customizing for your laboratory s
Unleash Your Potential
Unleash Your Potential Efficiency. Power. Flexibility. Small In Size Big In Performance. The respons 910 vet A Leap Forward For Your Laboratory. System solution consisting of the respons 910 vet analyzer,
Intact PTH. system. Intact PTH 84-6434/R5. Key to symbols used 8K25
system en Intact PTH 8K25 84-6434/R5 Intact PTH Read Highlighted Changes Revised November, 2008 Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA
UNIT: Total and Direct Bilirubin
UNIT: Total and Direct Bilirubin 13bili.wpd Task Determination of total and direct bilirubin. Objectives Upon completion of this exercise, the student will be able to: 1. Explain formation, excretion,
TOTAL PROTEIN FIBRINOGEN
UNIT: Proteins 16tproteins.wpd Task Determination of Total Protein, Albumin and Globulins Objectives Upon completion of this exercise, the student will be able to: 1. Explain the ratio of albumin and globulin
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Malondialdehyde (MDA) ELISA
Malondialdehyde (MDA) ELISA For the quantitative determination of MDA in serum, plasma and other biological fluids Cat. No. KT-21493 For Research Use Only. Not for use in diagnostic procedures. Page 1
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2
Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: [email protected] ELISA Kit for Human Prostaglandin E1(PG-E1) Instruction manual Cat. No.: E0904Hu
Procedure Manual for the i-stat System
Procedure Manual for the i-stat System This Procedure Manual is intended to be a template for the Procedure Manual required by CLIA and laboratory accreditation bodies. This Procedure Manual should be
Discover Your Potential
Discover Your Potential Efficiency. Power. Flexibility. Small In Size Big In Performance. respons 910 A Leap Forward For Your Laboratory. System solution consisting of respons 910 analyzer, intuitive software,
Discover Your Potential
Discover Your Potential Efficiency. Power. Flexibility. Small In Size Big In Performance. respons 910 A Leap Forward For Your Laboratory. Superior system solution including respons system reagents, analyzer,
Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07
Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma
Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use
Pointe Scientific, Inc. Instrument Application
Pointe Scientific, nc. nstrument Application Analyzer: Chemwell Test: Albumin Catalog # : A7502 Options: Assignment Order: 8-way in groups of 4 Optimized order execution Temperature Control to 37C Dispense
User Guide. Quality Check Messages and Codes (continued) (01)00054749001286. Message Cause Action
Quality Check Messages and Codes (continued) Message Cause Action i-stat Analyzer Error, Use Electronic Simulator Analyzer Error, See Manual No display Cartridge Locked does not disappear after test cycle
Blood Glucose Monitoring: The Facts about Accuracy
Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined
Bovine Vitamin B12 (VB12) ELISA Kit
Bovine Vitamin B12 (VB12) ELISA Kit Catalog Number. CSB-E14089B For the quantitative determination of endogenic bovine vitamin B12 (VB12) concentrations in serum, plasma, tissue homogenates. This package
Southern Derbyshire Shared Care Pathology Guidelines. Hyperkalaemia
Southern Derbyshire Shared Care Pathology Guidelines Hyperkalaemia Purpose of Guideline Dealing with adult patients with Hyperkalaemia in the community Definition Serum potassium normal range is 3.5 5.3
510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2
Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Vitamin B12. ADVIA Centaur System
Vitamin B12 Principle of the Test ADVIA Centaur System The ADVIA Centaur VB12 assay is a competitive immunoassay using direct chemiluminescent technology in which vitamin B 12 from the patient sample competes
Chapter 14 Urinalysis, Body Fluids and Other Specimens. Objectives:
EXERCISE 15: CHEMICAL EXAMINATION OF URINE Textbook: Skill: Chapter 14 Urinalysis, Body Fluids and Other Specimens 15 points Objectives: 1. Name 10 routine chemical tests performed on urine and list a
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Catalog Number. CSB-E14405r For the quantitative determination of rat creatine kinase MM isoenzyme (CK-MM) concentrations in serum, plasma and tissue
STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST
Before performing testing, all operators MUST read and become familiar with Universal Precautions for Prevention of Transmission of Human Immunodeficiency Virus, Hepatitis B Virus, and other Blood-borne
URINE DIPSTICK Multistix 10SG
The Johns Hopkins Hospital Point-of-Care Testing Program URINE DIPSTICK Multistix 10SG Operator Competency Only operators who have completed a defined training program and can demonstrate competence will
LIAISON XL HCV Ab Accurate diagnosis of the early stage of HCV infection
Hepatitis and Retrovirus LIAISON XL HCV Ab Accurate diagnosis of the early stage of HCV infection FOR OUTSIDE THE US AND CANADA ONLY LIAISON XL HCV Ab is Your solution LIAISON XL murex HCV Ab main features
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
MEDICAL NUTRITION THERAPY (MNT) CLINICAL NUTRITION THERAPY Service Time CPT Code
MEDICAL NUTRITION THERAPY (MNT) CLINICAL NUTRITION THERAPY Service Time CPT Code Initial Assessment And Intervention This Code Can Be Used Only Once A Year For First Appointment Medical Nutrition Therapy
Ammonia (plasma, blood)
Ammonia (plasma, blood) 1 Name and description of analyte 1.1 Name of analyte Ammonia 1.2 Alternative names None 1.3 NLMC code 1.4 Description of analyte Ammonia has the formula NH 3. At physiological
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Catalog Number. CSB-E14405r For the quantitative determination of rat creatine kinase MM isoenzyme (CK-MM) concentrations in serum, plasma, tissue homogenates.
Please note: Contact Coppe Laboratories at 262-574-0701 if archival plasma samples need to be tested.
Collecting a Coppe Laboratories Sample The Coppe Laboratories Sample Kit contains: Test Request Form (TRF) Heparin tube Absorbent sheet Biohazard bag Foil pouch Label for blood tube Box Seal Instructions
Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit
Canine creatine kinase MB isoenzyme (CK-MB)ELISA Kit Catalog No. CSB-E15852c (96T) This immunoassay kit allows for the in vitro quantitative determination of canine CK-MB concentrations in serum and plasma.
Reference Range: 0.5-1.6 mmol/l (arterial) 0.7-2.1 mmol/l (venous) CPT Code: 83605
LACTIC ACID Fasting, arterial specimen preferred. Please note whether arterial or venous. 0.5 ml heparinized plasma. Green top or PST must be drawn, placed on ice, and spun within 15 minutes. Immediately
Carcinoembryonic Antigen (CEA)
Principle of the Test Carcinoembryonic Antigen (CEA) ADVIA Centaur System The ADVIA Centaur CEA assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts
Contents. POCT Program Massachusetts General Hospital - Pathology Service 55 Fruit Street, Boston, MA 02114
Title: DCA Vantage HgbA1C Procedure Cross References: DCA Vantage HgbA1C Training and Competency Assessment Record DCA Vantage Operator Training Checklist DCA Vantage Maintenance Log DCA Vantage Optical
Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory in India: A Pilot Study
Ind J Clin Biochem (Apr-June 2011) 26(2):131 135 DOI 10.1007/s12291-010-0083-1 ORIGINAL ARTICLE Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory
LIAISON XL HBsAg Quant
Hepatitis and Retrovirus LIAISON XL HBsAg Quant Reliable detection of HBsAg mutants and genotypes for accurate differential diagnosis of the stage of infection FOR OUTSIDE THE US AND CANADA ONLY LIAISON
TECHNICAL BULLETIN. Glucagon EIA Kit for serum, plasma, culture supernatant, and cell lysates. Catalog Number RAB0202 Storage Temperature 20 C
Glucagon EIA Kit for serum, plasma, culture supernatant, and cell lysates Catalog Number RAB0202 Storage Temperature 20 C TECHNICAL BULLETIN Product Description The Glucagon Enzyme Immunoassay (EIA) Kit
BIOCHEMISTRY DEPARTMENT
ALBUMIN 33 46 g/l ALK PHOSPHATASE Up to 11 months Male 133 426 U/L Female 196 553 U/L 11+ months 2 years Male 133 287 U/L Female 196 510 U/L 2+ 8 years Male 133 287 U/L Female 196 426 U/L 8+ 11 years Male
Hepatitis and Retrovirus. LIAISON XL Accurate detection of HIV infection. HIV Ab/Ag FOR OUTSIDE THE US AND CANADA ONLY
Hepatitis and Retrovirus LIAISON XL Accurate detection of HIV infection HIV Ab/Ag FOR OUTSIDE THE US AND CANADA ONLY LIAISON XL HIV Ab/Ag is Your solution LIAISON XL murex HIV Ab/Ag main features The LIAISON
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
TECHNICAL BULLETIN. Angiotensin II EIA Kit for serum, plasma, culture supernatant, and cell lysates. Catalog Number RAB0010 Storage Temperature 20 C
Angiotensin II EIA Kit for serum, plasma, culture supernatant, and cell lysates Catalog Number RAB0010 Storage Temperature 20 C TECHNICAL BULLETIN Product Description The Angiotensin II Enzyme Immunoassay
Phosphate (serum, plasma, urine)
Phosphate (serum, plasma, urine) 1 Name and description of analyte 1.1 Name of analyte Phosphate 1.2 Alternative names Phosphorus, inorganic phosphorus (Pi), PO 4. 1.3 NMLC number 1.4 Description of analyte
Cold Agglutination Titer detecting Cold Reacting Antibodies
Objectives: Cold Agglutination Titer detecting Cold Reacting Antibodies 1. Perform a serial dilution to determine the amount of cold reacting antibody present in a patient specimen with the results obtained
Albumin (serum, plasma)
Albumin (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Albumin (plasma or serum) 1.2 Alternative names None (note that albumen is a protein found in avian eggs) 1.3 NLMC code 1.4
Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit
Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Catalog No. MBS702810 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of mouse KLH(IgM)antibody concentrations
Human Parathyroid Hormone (PTH) ELISA Kit
AssayMax TM Human Parathyroid Hormone (PTH) ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St. Charles, MO 63301 T (636) 447-9175 F (636) 395-7419 www.assaypro.com For any questions regarding troubleshooting
Title: Fingerstick Glucose by Accu-Chek Inform
Title: Fingerstick Glucose by Accu-Chek Inform Target Audience: This module is available to aid in assessing competency for all clinical staff who perform fingerstick glucose testing. Contents Instructions...
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods
COBAS INTEGRA 400 plus
COBAS INTEGRA 400 plus COBAS INTEGRA 400 plus Advantages Real security of results with the COBAS INTEGRA clot detection system cobas c packs make reagent preparation a task of the past Workflow consolidation
Individualized Quality Control Plan
NOVEMBER 2014 CLIA Individualized Quality Control Plan WHAT IS AN IQCP? OVERVIEW Discovering, Deciphering and Designing an Individualized Quality Control Plan (IQCP) IQCP provides a framework for customizing
INMA LABORATORY MANUAL
INMA LABORATORY MANUAL NUTRITION BIOMARKERS Cohorte-VALENCIA This manual describes the collection, separation, storage and transport of blood samples for the INMA study. Careful collection of samples is
SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory
BY UV SPECTROPHOTOMETRY
1 of 9 1. PURPOSE This Standard Operating Procedure describes the method for measuring the concentration of viral particles using UV/Vis spectrophotometry at 260 and 280 nm. 2. SCOPE This procedure is
cobas c 311 analyzer
cobas c 311 analyzer cobas brand Roche Diagnostics introduces the cobas brand as the umbrella for products used to complete or expand the screening, diagnostic and monitoring applications of the professional
OneStep Fecal Occult Blood RapiDip InstaTest. Cat # 13020-1
CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: [email protected] Web site: www.rapidtest.com See external label
Anti-Zona Pellucida Antibody Latex Agglutination Test
Instructions for use Anti-Zona Pellucida Antibody Latex Agglutination Test Cat. No.: BS-20-10 Size: 50 Determinations Storage: 2 C 8 C (36 F 46 F) Screening test for the determination of anti-zona pellucida
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Portex Arterial Blood Sampling Kits
Portex Arterial Blood Sampling Kits Safe Application. Simple Activation. Consistent Results. Portex Arterial Blood Sampling Kits > Compatible with Multi-parameter Analyzers > Balanced Heparin Provides
TECHNICAL BULLETIN. Cholecystokinin (CCK) EIA Kit for serum, plasma, culture supernatant, and cell lysates
Cholecystokinin (CCK) EIA Kit for serum, plasma, culture supernatant, and cell lysates Catalog Number RAB0039 Storage Temperature 20 C TECHNICAL BULLETIN Product Description The Cholecystokinin (CCK) Enzyme
for patient monitoring
Thermo Scientific Indiko Reagents for Immunosuppressant Drug Testing Complete solutions for patient monitoring The cost-effective Thermo Scientific Indiko benchtop analyzer and bar-coded reagent is designed
Mouse krebs von den lungen 6 (KL-6) ELISA
KAMIYA BIOMEDICAL COMPANY Mouse krebs von den lungen 6 (KL-6) ELISA For the quantitative determination of mouse KL-6 in serum, plasma, cell culture supernatants, body fluid and tissue homogenate Cat. No.
Adams Memorial Hospital Decatur, Indiana EXPLANATION OF LABORATORY TESTS
Adams Memorial Hospital Decatur, Indiana EXPLANATION OF LABORATORY TESTS Your health is important to us! The test descriptions listed below are for educational purposes only. Laboratory test interpretation
LIVER FUNCTION TESTS
MODULE Liver Function Tests 17 LIVER FUNCTION TESTS 17.1 INTRODUCTION Liver function tests are a group of tests done to assess the functional capacity of the liver as well as any cellular damage to the
Rat Fibronectin ELISA Kit
AssayMax TM Rat Fibronectin ELISA Kit Assaypro LLC 3400 Harry S Truman Blvd St. Charles, MO 63301 T (636) 447-9175 F (636) 395-7419 www.assaypro.com For any questions regarding troubleshooting or performing
Quality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
METHODS OF VITAMIN ANALYSIS
METHODS OF VITAMIN ANALYSIS Specimen requirements; Fasting plasma or serum Lithium Heparin is the anticoagulant of choice for vitamins such as; thiamine, riboflavin, retinol, tocopherols & cholecalciferol
Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit
Mouse glycated hemoglobin A1c(GHbA1c) ELISA Kit Catalog Number. CSB-E08141m For the quantitative determination of mouse glycated hemoglobin A1c(GHbA1c) concentrations in lysate for RBC. This package insert
SKUP Scandinavian evaluation of laboratory equipment for primary health care
SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Aviva Meter and test strips designed for glucose self-measurement Manufactured by Roche Diagnostics GmbH Report from
Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA
Rat Creatine Kinase MB Isoenzyme (CKMB) ELISA For the quantitative determination of rat CKMB in serum, plasma, tissue homogenates and other biological fluids. Cat. No. KT-12247 For Research Use Only. Not
Calcium (serum, plasma, blood)
Calcium (serum, plasma, blood) 1 Name and description of analyte 1.1 Name of analyte Calcium (total in serum, plasma, ionised in blood (see 2.1 (2)). 1.2 Alternative names None 1.3 NMLC code 1.4 Description
Lab 2. Spectrophotometric Measurement of Glucose
Lab 2 Spectrophotometric Measurement of Glucose Objectives 1. Learn how to use a spectrophotometer. 2. Produce a glucose standard curve. 3. Perform a glucose assay. Safety Precautions Glucose Color Reagent
And Now, Presenting...
Presentation and Handling of Clinical Laboratory Data From Test Tube to Table Randall K. Carlson, Wilmington, DE and Nate Freimark, Lakewood, NJ Omnicare Clinical Research. Inc. INTRODUCTION In human clinical
Routine Investigations for Liver Disease a guide
fighting childhood liver disease Routine Investigations for Liver Disease a guide Medical Information Series Welcome This leaflet has been written specifically for: Parents/carers of a child with a liver
Cancer Antigen CA 125
Principle of the Test Cancer Antigen CA 125 ADVIA Centaur System The ADVIA Centaur CA 125 II assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses two monoclonal