1 FlexiWeigh The balance is an integral part of the FlexiWeigh workstation. The role and limitations of the balance are important to understand in this environment. This technical note describes the different types of balances that have been integrated into the FlexiWeigh workstations and offers details of their specifications. Understanding the uncertainty of measurement associated with different types of weighing, both for static and dynamic systems, is important. The methods for calculating these factors are described in this Technical Note.
2 Table 1: Definition of Terms Readability: The smallest difference between two measured values that can be read on the display. Repeatability: The ability of a balance to provide the same result for repeated weighing of the same load under the same measurement conditions. Linearity: Cornerload: Accuracy: Precision: The capability of the balance to follow the linear relation between the load and the displayed value. Inaccuracy of the balance factor for objects that are not placed centrally on the balance. Closeness of the agreement between the determined result and the true value. Extent of mutual agreement between of several measured values, independent of the true value. Table 2: Comparison of Balance Specifications for Different Balance Types BALANCE: AG204 AG285 * SAG204/01 SAG 285/01 * Maximum Capacity 210 g 81 g 210 g 81 g Taring Range g 0-81 g g 0 81 g Readability 0.1 mg 0.01 mg 0.1 mg 0.01 mg Repeatability 0.1 mg 0.05 mg 0.1 mg 0.02 mg Linearity ± 0.2 mg ± 0.1 mg ± 0.2 mg ± 0.03 mg Typical 3σ value ** ± 0.25 mg ± mg ± 0.25 mg ± mg Stabilization time *** Sensitivity - temperature drift Sensitivity long term stability 3 s > 20 s 3 s > 20 s ± 1.5 ppm/ C ± 2 ppm/ C ± % ± % * Balances have an additional set of parameters for 0 g 210 g range, but these are not relevant for the FlexiWeigh application. ** 3σ value calculated using dynamic weighing method (see page 3). *** Stabilization time refers to the time taken for the balance to stabilise in an automated system and does not necessarily correspond with METTLER TOLEDO figures quoted for stand alone balance modules of this type Version 3 Dec-04 2
3 Static vs. Dynamic Weighing Remembering that a balance reading is not exact but it is ± another value holds significance. The uncertainty of this value will depend on a number of factors, including whether the measurement is static or dynamic. An example of a static measurement is: weigh a tube on a balance, remove from the balance, add powder to the tube, re-weigh the tube to determine the weight of the tube + powder. This is less accurate than a dynamic measurement. An example of a dynamic measurement is: weigh a tube, add powder whilst the tube is still on the balance, read the weight before removing the tube from the balance. Determination of the Uncertainty of Measurement (a) For a Dynamic Weighing The uncertainty value is known as 3σ (where σ is the statistical standard deviation, equal to the square root of the statistical variance: v= σ 2 ). The 3σ value is made up from four components, according to the following equation: 3σ = a + b + c + d where: a = readability b = repeatability c = linearity d = cornerload 2 For robotics, it can be assumed that d = 0, because vials or tubes are centred on the balance when they are weighed. In the case of dynamic weighing, i.e. when the balance is zeroed and the powder is added directly to a tube or vial on the balance, as is the case with FlexiWeigh, the calculation for the SAG204/01 balance, for example, is: a = 0.1mg b = 0.1 mg c = 0.2 mg d = 0.0 mg which gives 3σ = 0.25 mg Therefore, any value that is read on an SAG204/01 balance whose total weight is less than 81g is ± 0.25mg. This fact is not due to any robotics, but is general behaviour of all balances, which must be taken into account when a value is read on any balance. (b) For a Static Weighing The situation is different for a static weighing. In the case of static weighing, i.e. when a vessel is tared, taken off the balance, powder is added, then the vessel is re-weighed on the same balance in order to determine the amount of powder added, the standard deviation for a static weighing is calculated as follows: Version 3 Dec-04 3
4 From a statistical point of view, there are two totally mutually independent events E1 and E2. E1: The first weighing (tare of the vessel) T with a variance of V E2: The second weighing (gross weigh of the vessel), on the same balance, W with a variance of V Where V= σ 2 calculated according to the previous section. The purpose of this section is to calculate the variance V and deduce the 3σ associated with adding (or subtracting) several values relating to several mutually independent events. In this case, the quantity (Q) of powder added between events E1 and E2 is defined as Q=W-T. The error in the measurement of Q is: (1) ln(q)=ln(w-t) (2) Q Derivation: = Q W W T W T T (3) Q To calculate the error, the equation (2) becomes: = Q W T (4) Considering that Q=W-T, this is equivalent to: V =V+V=2V (5) Meanwhile V= σ 2 and V = σ 2 (6) Hence σ 2 =2 σ 2 σ = 2 x σ 3 σ = 2 x 3σ For example, the previous calculation showed that using the SAG204/01 balance, the 3σ = 0.25 for each balance reading where the quantity is less than 81g. In this example: Event E1: the tare of the tube: T Event E2: the weighing of the tube after adding powder: W Consequently, the 3σ for the quantity of powder added Q=W-T is: 3σ = 2 x 3σ 3σ = 2 x 0.25 = 0.35 This means that each quantity of powder Q measured using the static weighing method on an SAG204/01 balance is ± 0.35mg. Comparison of Dynamic vs Static In this worked example, using an SAG204/01 balance, the dynamic weighing method used by FlexiWeigh has a 3σ of 0.25 mg and the static has a 3σ of 0.35 mg. The dynamic weighing method is clearly more accurate, and care should be taken about validating dynamic weights using a static weighing method. For hints on the best way to approach this, see page Version 3 Dec-04 4
5 Validation of a Dynamic Weighing Consequently, if it is necessary to double check the weight reported by the FlexiWeigh, the user must be extremely cautious about how this is carried out. The calculations shown in this document should be taken into account before any testing is undertaken to validate the FlexiWeigh weight recorded. With an AG204, the dynamic weighing method used by FlexiWeigh has a 3σ of 0.25 mg and the static method to validate the weights measured by FlexiWeigh has a 3σ of 0.35 mg, as previously illustrated. This means that if the difference is less than = 0.60mg, the result is validated. Examples of similar calculations with different METTLER TOLEDO balance types are shown in the table below. Note: All the balances presented here are used at their best performances in terms 3σ. Table 3: Uncertainty Measurement (3σ) Calculations for Different Balance Types AG204 or SAG204/01 AG285 SAG285/01 Dynamic weighing method ± 0.25 mg ± mg ± mg Static weighing method ± 0.35 mg ± mg ± mg Maximum permitted difference between the same weight measured by two different methods ± 0.60 mg ± mg ± mg Confirming a Weighing When attempting to confirm a weighing on an external balance, the factors described above must be taken into consideration. Beware of comparing readings from the FlexiWeigh to another balance reading. Take into account the 3σ value of the other balance, as well as that from the FlexiWeigh. In addition, the balance itself should be considered a factor, as this will be subject to drift over time. The powder could also be a cause for some variation, as it may have hygroscopic tendencies, which will serve to alter the weight of the same sample over a period of time. Tips for attempting to confirm a FlexiWeigh powder dispensing measurement on another balance: Aim to use the same type of balance for both weight measurements Aim for the minimum delay between weight measurements to reduce potential error factors resulting from absorption, balance drift, temperature changes, etc Version 3 Dec-04 5
6 Additional Factors Affecting Weighing Other factors that should be taken into account with a balance reading are: Air flow Electrostaticity Humidity Temperature Pressure Bench stability The FlexiWeigh workstation design eliminates some of the factors listed above. For example, the factors associated with air flow are eliminated by having the enclosure; the electrostaticity is minimized by use of the in-built ionization system, and the issue of bench stability is easily addressed at an early stage. Effect of the Ionizer The ionizer, which is important for good powder dispensing, can have an additive (or subtractive) effect on the values shown on the balance when it is ON. The FlexiWeigh is able to detect these fluctuations and the powder dispensing algorithm is able to filter most of this effect. However, all the values recorded by FlexiWeigh are correct, as the initial weighing (Tare or Zero operation) and the final weighing are strictly measured under the same conditions. These conditions are: Ionizer OFF: no additive effect Nose Up Nose Out For example, this is the reason that a user can request a minimum dispense target of 2.0 mg. However, because the FlexiWeigh powder algorithm measures an additive effect when the ionizer is set to ON of, for example -0.3mg, the FlexiWeigh powder algorithm will decide to set a new minimum target of 1.7 mg. When the algorithm finishes its dispensing, it will turn off the ionizer, and the value on the balance will increase by +0.3mg, and the FlexiWeigh will record the final weight made without the ionizer effect. Calibration The balance should be calibrated regularly (e.g. once a day for internal calibration). The balance is capable of detecting automatically that it needs an internal calibration. At this point the AutoCal symbol flashes on the balance screen. The FlexiWeigh software will allow the balance to perform the Autocalibration, only if a process is not running at that time. If a process is running, then the balance will autocalibrate once the process has finished Version 3 Dec-04 6
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
Estimating Weighing Uncertainty From Balance Data Sheet Specifications
Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement
Estimates of Uncertainty of the Calibration of Balances
Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
Signal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
The Analytical Balance
Chemistry 119: Experiment 1 The Analytical Balance Operation of the Single-Pan Analytical Balance Receive instruction from your teaching assistant concerning the proper operation of the Sartorius BP 210S
Testing Your Laboratory Balance
Testing Your Laboratory Balance If you use a laboratory balance in your day-to-day work, you may have wondered whether it is accurate. Here are four tests that you can use to confirm your instrument's
Biostatistics: DESCRIPTIVE STATISTICS: 2, VARIABILITY
Biostatistics: DESCRIPTIVE STATISTICS: 2, VARIABILITY 1. Introduction Besides arriving at an appropriate expression of an average or consensus value for observations of a population, it is important to
Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS
Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal
MEASURES OF VARIATION
NORMAL DISTRIBTIONS MEASURES OF VARIATION In statistics, it is important to measure the spread of data. A simple way to measure spread is to find the range. But statisticians want to know if the data are
Determining Measurement Uncertainty for Dimensional Measurements
Determining Measurement Uncertainty for Dimensional Measurements The purpose of any measurement activity is to determine or quantify the size, location or amount of an object, substance or physical parameter
THE STATISTICAL TREATMENT OF EXPERIMENTAL DATA 1
THE STATISTICAL TREATMET OF EXPERIMETAL DATA Introduction The subject of statistical data analysis is regarded as crucial by most scientists, since error-free measurement is impossible in virtually all
User checks and maintenance of laboratory balances
Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National
Assessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm
Error Analysis and the Gaussian Distribution In experimental science theory lives or dies based on the results of experimental evidence and thus the analysis of this evidence is a critical part of the
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
4. Continuous Random Variables, the Pareto and Normal Distributions
4. Continuous Random Variables, the Pareto and Normal Distributions A continuous random variable X can take any value in a given range (e.g. height, weight, age). The distribution of a continuous random
Calculate Highest Common Factors(HCFs) & Least Common Multiples(LCMs) NA1
Calculate Highest Common Factors(HCFs) & Least Common Multiples(LCMs) NA1 What are the multiples of 5? The multiples are in the five times table What are the factors of 90? Each of these is a pair of factors.
High-Precision Platforms
High-Precision Platforms PBK9 and PFK9 Weighing Platforms High-precision weighing Reliable in harsh conditions Fast weighing processes Flexible for many applications High Accuracy For Harsh Environments
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Statistical Confidence Calculations
Statistical Confidence Calculations Statistical Methodology Omniture Test&Target utilizes standard statistics to calculate confidence, confidence intervals, and lift for each campaign. The student s T
The right edge of the box is the third quartile, Q 3, which is the median of the data values above the median. Maximum Median
CONDENSED LESSON 2.1 Box Plots In this lesson you will create and interpret box plots for sets of data use the interquartile range (IQR) to identify potential outliers and graph them on a modified box
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory * Paper prepared for the 4 th International Symposium on Fluid Flow
3.2. Solving quadratic equations. Introduction. Prerequisites. Learning Outcomes. Learning Style
Solving quadratic equations 3.2 Introduction A quadratic equation is one which can be written in the form ax 2 + bx + c = 0 where a, b and c are numbers and x is the unknown whose value(s) we wish to find.
Figure 1. A typical Laboratory Thermometer graduated in C.
SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES
Statistical estimation using confidence intervals
0894PP_ch06 15/3/02 11:02 am Page 135 6 Statistical estimation using confidence intervals In Chapter 2, the concept of the central nature and variability of data and the methods by which these two phenomena
Measurement and Calibration
Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding
Welcome to Physics 40!
Welcome to Physics 40! Physics for Scientists and Engineers Lab 1: Introduction to Measurement SI Quantities & Units In mechanics, three basic quantities are used Length, Mass, Time Will also use derived
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Experimental Uncertainties (Errors)
Experimental Uncertainties (Errors) Sources of Experimental Uncertainties (Experimental Errors): All measurements are subject to some uncertainty as a wide range of errors and inaccuracies can and do happen.
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION
1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.
Accuracy and Pharmacutical
Accuracy and Pharmacutical Solution Weighing Team Contents Weighing Errors Weighing Technology Accurate scale with Pharmaceutical Why is accuracy / precision important? Knowing the error of your results
Thermochemistry: Calorimetry and Hess s Law
Thermochemistry: Calorimetry and Hess s Law Some chemical reactions are endothermic and proceed with absorption of heat while others are exothermic and proceed with an evolution of heat. The magnitude
EXPERIMENTAL ERROR AND DATA ANALYSIS
EXPERIMENTAL ERROR AND DATA ANALYSIS 1. INTRODUCTION: Laboratory experiments involve taking measurements of physical quantities. No measurement of any physical quantity is ever perfectly accurate, except
INTRODUCTION TO ERRORS AND ERROR ANALYSIS
INTRODUCTION TO ERRORS AND ERROR ANALYSIS To many students and to the public in general, an error is something they have done wrong. However, in science, the word error means the uncertainty which accompanies
Uncertainty of Force Measurements
European Association of National Metrology Institutes Uncertainty of Force Measurements EURAMET cg-4 Version.0 (03/011) Previously EA-10/04 Calibration Guide EURAMET cg-4 Version.0 (03/011) UNCERTAINTY
Guidelines on the Calibration of Static Torque Measuring Devices
European Association of National Metrology Institutes Guidelines on the Calibration of Static Torque Measuring Devices EURAMET cg-14 Version.0 (03/011) Previously EA-10/14 Calibration Guide EURAMET cg-14
CALCULATIONS & STATISTICS
CALCULATIONS & STATISTICS CALCULATION OF SCORES Conversion of 1-5 scale to 0-100 scores When you look at your report, you will notice that the scores are reported on a 0-100 scale, even though respondents
Part 2: Analysis of Relationship Between Two Variables
Part 2: Analysis of Relationship Between Two Variables Linear Regression Linear correlation Significance Tests Multiple regression Linear Regression Y = a X + b Dependent Variable Independent Variable
Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance
INL/CON-10-18643 PREPRINT Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance 2010 NCSL International Workshop and Symposium Michael Stears July 2010 This is
A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS
Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,
ROUTH S STABILITY CRITERION
ECE 680 Modern Automatic Control Routh s Stability Criterion June 13, 2007 1 ROUTH S STABILITY CRITERION Consider a closed-loop transfer function H(s) = b 0s m + b 1 s m 1 + + b m 1 s + b m a 0 s n + s
Zero: If P is a polynomial and if c is a number such that P (c) = 0 then c is a zero of P.
MATH 11011 FINDING REAL ZEROS KSU OF A POLYNOMIAL Definitions: Polynomial: is a function of the form P (x) = a n x n + a n 1 x n 1 + + a x + a 1 x + a 0. The numbers a n, a n 1,..., a 1, a 0 are called
Means, standard deviations and. and standard errors
CHAPTER 4 Means, standard deviations and standard errors 4.1 Introduction Change of units 4.2 Mean, median and mode Coefficient of variation 4.3 Measures of variation 4.4 Calculating the mean and standard
Descriptive Statistics
Y520 Robert S Michael Goal: Learn to calculate indicators and construct graphs that summarize and describe a large quantity of values. Using the textbook readings and other resources listed on the web
Instruction Manual Crane Scales PCE-HS Series
PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 [email protected] PCE Instruments UK Ltd. Units 12/13 Southpoint Business
MANUAL TR-110 TR-110
MANUAL TR-110 TR-110 CONTENTS 1. GENERAL INTRODUCTION 2 2. WORK PRINCIPLE 3 3. STANDARD DELIVERY 4 4. NAME OF EACH PART 5 5. OPERATION 6 5.1 Preparation before operation 6 5.2 Switch on, Switch off and
CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13
COMMON DESCRIPTIVE STATISTICS / 13 CHAPTER THREE COMMON DESCRIPTIVE STATISTICS The analysis of data begins with descriptive statistics such as the mean, median, mode, range, standard deviation, variance,
3.1. Solving linear equations. Introduction. Prerequisites. Learning Outcomes. Learning Style
Solving linear equations 3.1 Introduction Many problems in engineering reduce to the solution of an equation or a set of equations. An equation is a type of mathematical expression which contains one or
Northumberland Knowledge
Northumberland Knowledge Know Guide How to Analyse Data - November 2012 - This page has been left blank 2 About this guide The Know Guides are a suite of documents that provide useful information about
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
Rounding and Significant Figures
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Rounding and Significant Figures
Pipette Calibration. Accurate Pipettes Day by Day. Ensure Operational Reliability With Service XXL. www.mt.com For more information
Ensure Operational Reliability With Service XXL Qualify your balances and work safely from day one. METTLER TOLEDO offers customized solutions for working in regulated environments. EQ-Pac Conformity and
Measurement with Ratios
Grade 6 Mathematics, Quarter 2, Unit 2.1 Measurement with Ratios Overview Number of instructional days: 15 (1 day = 45 minutes) Content to be learned Use ratio reasoning to solve real-world and mathematical
School experiments. Weighing the Right Way. Proper Weighing with Laboratory Balances. P h y - Guide Book
Weighing the Right Way School experiments Experience Improved Weighing Flawless Operation Precise Results P h y - Guide Book Proper Weighing with Laboratory Balances Natural science laws experience live
THE LABORATORY NOTEBOOK
THE LABORATORY NOTEBOOK In scientific work keeping a permanent record of all raw data, observations, calculations, et cetera obtained during an experiment is important. Therefore, a student must become
ph Alkalinity of Water
ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.
Chapter 4. Evaluating Analytical Data. When using an analytical method we make three separate evaluations of experimental error.
Chapter 4 Evaluating Analytical Data Chapter Overview 4A Characterizing Measurements and Results 4B Characterizing Experimental Errors 4C Propagation of Uncertainty 4D The Distribution of Measurements
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
Characterizing Digital Cameras with the Photon Transfer Curve
Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently
Good Weighing Practice In Pharmaceutical Industry. Pharmacy Risk Assessment For Mass Measurement Processes - 1 -
Good Weighing Practice In Pharmaceutical Industry Pharmacy Risk Assessment For Mass Measurement Processes - 1 - Content Introduction...3 2. Measurement errors...4 2.1. Gross error...4 2.2. Systematic error...4
2 ESTIMATION. Objectives. 2.0 Introduction
2 ESTIMATION Chapter 2 Estimation Objectives After studying this chapter you should be able to calculate confidence intervals for the mean of a normal distribution with unknown variance; be able to calculate
Time series Forecasting using Holt-Winters Exponential Smoothing
Time series Forecasting using Holt-Winters Exponential Smoothing Prajakta S. Kalekar(04329008) Kanwal Rekhi School of Information Technology Under the guidance of Prof. Bernard December 6, 2004 Abstract
Week 3&4: Z tables and the Sampling Distribution of X
Week 3&4: Z tables and the Sampling Distribution of X 2 / 36 The Standard Normal Distribution, or Z Distribution, is the distribution of a random variable, Z N(0, 1 2 ). The distribution of any other normal
Training Material for Balances (1) Prepared by R&D Division 1. Marketing Department A&D Company, Limited
Training Material for Balances (1) Prepared by R&D Division 1 Marketing Department A&D Company, Limited 1 Contents Things you should know before using a balance --------------------------- P.3 1. Convection
Density Determinations and Various Methods to Measure
Density Determinations and Various Methods to Measure Volume GOAL AND OVERVIEW This lab provides an introduction to the concept and applications of density measurements. The densities of brass and aluminum
LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS
XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal
EdExcel Decision Mathematics 1
EdExcel Decision Mathematics 1 Linear Programming Section 1: Formulating and solving graphically Notes and Examples These notes contain subsections on: Formulating LP problems Solving LP problems Minimisation
COMMUNICATING using MEASUREMENTS In Engineering we use a great many measuring instruments.
COMMUNICATING using MEASUREMENTS In Engineering we use a great many measuring instruments. Scales Verniers Micrometers Gauges Comparators Thermometers Thermistors + Indicator Thermocouples + Indicator
MEASUREMENT. Historical records indicate that the first units of length were based on people s hands, feet and arms. The measurements were:
MEASUREMENT Introduction: People created systems of measurement to address practical problems such as finding the distance between two places, finding the length, width or height of a building, finding
Experiment 1: Measurement and Density
Experiment 1: Measurement and Density Chemistry 140 Learning Objectives Become familiar with laboratory equipment and glassware Begin to see the link between measurement and chemical knowledge Begin to
Confidence Intervals for the Difference Between Two Means
Chapter 47 Confidence Intervals for the Difference Between Two Means Introduction This procedure calculates the sample size necessary to achieve a specified distance from the difference in sample means
Conductivity Sensor Calibrations to Meet Water Industry Requirements
Conductivity Sensor Calibrations to Meet Water Industry Requirements Victor M. Braga Technical Service/Training Manager Mettler-Toledo Thornton Inc. 36 Middlesex Turnpike Bedford, Massachusetts 01730 Phone
Prentice Hall: Middle School Math, Course 1 2002 Correlated to: New York Mathematics Learning Standards (Intermediate)
New York Mathematics Learning Standards (Intermediate) Mathematical Reasoning Key Idea: Students use MATHEMATICAL REASONING to analyze mathematical situations, make conjectures, gather evidence, and construct
DESCRIPTIVE STATISTICS. The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses.
DESCRIPTIVE STATISTICS The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses. DESCRIPTIVE VS. INFERENTIAL STATISTICS Descriptive To organize,
Chapter 1 Lecture Notes: Science and Measurements
Educational Goals Chapter 1 Lecture Notes: Science and Measurements 1. Explain, compare, and contrast the terms scientific method, hypothesis, and experiment. 2. Compare and contrast scientific theory
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Introduction Many metals react with acids to form hydrogen gas. In this experiment, you will use the reactions
5.1 Radical Notation and Rational Exponents
Section 5.1 Radical Notation and Rational Exponents 1 5.1 Radical Notation and Rational Exponents We now review how exponents can be used to describe not only powers (such as 5 2 and 2 3 ), but also roots
Lean Six Sigma Analyze Phase Introduction. TECH 50800 QUALITY and PRODUCTIVITY in INDUSTRY and TECHNOLOGY
TECH 50800 QUALITY and PRODUCTIVITY in INDUSTRY and TECHNOLOGY Before we begin: Turn on the sound on your computer. There is audio to accompany this presentation. Audio will accompany most of the online
CALIBRATION PRINCIPLES
1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they
Gage Studies for Continuous Data
1 Gage Studies for Continuous Data Objectives Determine the adequacy of measurement systems. Calculate statistics to assess the linearity and bias of a measurement system. 1-1 Contents Contents Examples
Temperature Scales. The metric system that we are now using includes a unit that is specific for the representation of measured temperatures.
Temperature Scales INTRODUCTION The metric system that we are now using includes a unit that is specific for the representation of measured temperatures. The unit of temperature in the metric system is
1.3 Algebraic Expressions
1.3 Algebraic Expressions A polynomial is an expression of the form: a n x n + a n 1 x n 1 +... + a 2 x 2 + a 1 x + a 0 The numbers a 1, a 2,..., a n are called coefficients. Each of the separate parts,
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
MBA 611 STATISTICS AND QUANTITATIVE METHODS
MBA 611 STATISTICS AND QUANTITATIVE METHODS Part I. Review of Basic Statistics (Chapters 1-11) A. Introduction (Chapter 1) Uncertainty: Decisions are often based on incomplete information from uncertain
Standard Deviation Estimator
CSS.com Chapter 905 Standard Deviation Estimator Introduction Even though it is not of primary interest, an estimate of the standard deviation (SD) is needed when calculating the power or sample size of
High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors
High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last
When your applications demand more than a standard balance, the OHAUS Voyager Pro is up to the task.
Voyager Pro Analytical and Precision Balances When your applications demand more than a standard balance, the OHAUS Voyager Pro is up to the task. Voyager Pro s Advanced Applications simplify even the
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
AN_6521_035 APRIL 2009
71M6521 Energy Meter IC A Maxim Integrated Products Brand APPLICATION NOTE AN_6521_035 APRIL 2009 This document describes how to use software to compensate the real time clock (RTC) in Teridian meter chips.
WHERE DOES THE 10% CONDITION COME FROM?
1 WHERE DOES THE 10% CONDITION COME FROM? The text has mentioned The 10% Condition (at least) twice so far: p. 407 Bernoulli trials must be independent. If that assumption is violated, it is still okay
Assay Development and Method Validation Essentials
Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Training Manual. Pre-Employment Math. Version 1.1
Training Manual Pre-Employment Math Version 1.1 Created April 2012 1 Table of Contents Item # Training Topic Page # 1. Operations with Whole Numbers... 3 2. Operations with Decimal Numbers... 4 3. Operations
NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments
NITP 6.1 to 6.4 National Instrument Test Procedures for Non-automatic Weighing Instruments Commonwealth of Australia 2012 NSC V 1 First edition 2000 Second edition October 2002 Third edition February 2004
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND