1 New Seismic Compliance Requirements for Hospitals Special Seismic Certification Preapproval Program Rebroadcast of June 17, 2010 Web Seminar
2 Welcome & Program Overview
3 Roger Richter Roger Richter is Senior Vice President for Professional Services, at California Hospital Association where he represents the members interests on energy, facilities, environmental and disaster planning issues for California hospitals and health systems. Mr. Richter serves as a liaison with state regulatory agencies, including the Office of Statewide t Health Planning and Development, Hospital Building Safety Board, California Buildings Standards Commission, State Fire Marshal, Cal/OSHA, Cal EPA, Emergency Medical Services Authority, California Department of Public Health, California Energy Commission and the California Public Utilities Commission. 3
4 Acknowledgments Special thanks to: Office of Statewide Health Planning and Development California Society for Healthcare Engineering American Institute of Architects, American Institute of Architects, California Council
5 Chris Tokas Chris Tokas MS, SE, is a supervising structural engineer for the Office of Statewide Health Planning and Development. Mr. Tokas has 25 years of experience in structural engineering and design, with special emphasis in the areas of seismic evaluation, analysis, and retrofit. He has designed and managed seismic upgrade projects for a wide variety of public, private and institutional clients and on a wide variety of structures. Mr. Tokas has been involved in all phases of seismic risk reduction programs, including development of criteria, evaluation of existing structural capacity, schematic design, and preparation p of construction documents. 5
6 Office of Statewide t Health Planning & Development Facilities Development Division The Building Department for California s Hospitals Update on the CBC 2010 Requirements Presented by Chris Tokas, MS, SE 6
7 Drivers of Change Earthquake Lessons Learned 7
8 Special Seismic Certification is Also Known As (AKA) Seismic Certification of Nonstructural Components (IBC 2009 Section and the CBC 2010 Section 1708A.4) Special Certification for Designated Seismic System (ASCE 7-05 Section ) 8
9 Codes and Standards Relationship 9
10 What is Designated Seismic System? Any Component or System with Importance Factor (I p )i is Greater than 1.0 is called Designated Seismic System (CBC 2010 Section 1702A) 10
11 Nonstructural Component Importance Factor Component Importance (I p ) is either 1.0 or 1.5 (ASCE 7-05 Section ) Component Importance (I p ) is 1.5 if: Component is required to function for life-safety Component contain hazardous materials Components is in essential facilities and its failure could impair the continued operation of the facility 11
12 Intent of Building Code Provide reasonable assurance of seismic performance that will: 1. Avoid serious injury and life loss, 2. Avoid loss of function in critical facilities, 3. Minimize structural and nonstructural repair costs where practical to do so. 12
13 Statutory Mandate for OSHPD HOSPITAL FACILITIES SEISMIC SAFETY ACT 1983 (HSSA 83) Hospital buildings that house patients who have less than the capacity of normally healthy persons to protect themselves must be reasonably capable of providing services to the public after a disaster 13
14 Seismic Performance of Nonstructural Components Damage to nonstructural components can result in: Significant economic losses Cost of building contents exceeds building cost Temporary partial/total loss of operation/functionality Injuries and Loss of Life Past Emphasis: Life Safety and Position Retention only Present Focus: Maintaining Functionality 14
15 Special Certification for Designated Seismic Systems Section of ASCE 7-05 Special Certification is required for Designated Seismic Systems assigned to Seismic i Design Categories C through F as follows: Active mechanical and electrical equipment that are required to function following the Design Earthquake Components containing hazardous contents 15
16 Seismic Certification of Nonstructural Components CBC 2010 Section 1708A.4: The registered design professional shall state the applicable seismic certification requirements for nonstructural components and designated seismic systems on the construction documents. CONT D 16
17 CBC 2010 Section 1708A.4 Seismic Certification of Nonstructural t Components The manufacturer of each designated seismic system components subject to the provisions of ASCE 7 Section shall test or analyze the component and its mounting system or anchorage and submit a certificate of compliance for review and acceptance by the registered design professional responsible for the design of the designated seismic system and for approval by the building official. CONT D 17
18 CBC 2010 Section 1708A.4 Certification shall be based on: An actual test on a shake table, Seismic Certification of Nonstructural Components By three-dimensional shock tests, By an analytical method using dynamic characteristics and forces, By the use of experience data (i.e., historical data demonstrating acceptable seismic performance), or By more rigorous analysis providing for equivalent safety. 18
19 Seismic Certification by Analysis 19
20 Seismic Certification by Analysis Seismic Certification of Tanks by Analysis Using ASME BPVC Tanks (without vibration isolators) designed by: 1. A registered design professional in accordance with ASME Boiler and Pressure Vessel Code, 2007 (BPVC 2007); and, 2. Satisfying the force and displacement requirements of Sections and of ASCE 7-05 having an importance factor, I = 1.5; and, 3. Reviewed and approved by OSHPD Shall be considered to satisfy the Special Seismic Certification requirements on the basis of ASCE 7-05 Section
21 Seismic Certification by Analysis Seismic Certification of Tanks by Analysis Using ASCE 7-05 Chapter 15 Tanks Classified as Nonbuilding Structures in ASCE 7-05 shall be considered to satisfy Special Seismic Certification requirements if all of the following conditions are satisfied: 1. Tank design is in accordance ASCE 7-05 Chapter 15 with R = 0 = C d = 1, and 2. Attachments and Supports are designed in accordance with ASCE 7-05 Chapter 15 with R, 0, and C d values provided therein, and 3. Tank design is reviewed and approved by OSHPD. 21
22 Seismic Certification by Analysis For a product line with multiple products analysis is too time consuming. Active and Energized components must be qualified by test or experience data. Analyzing a custom 40-units HVAC product line will be similar to designing 40-hospital buildings. Same HVAC product line can be qualified by just two shake table test. Shake Table test is not a destructive test, hence tested unit may be installed and used 22
23 ASCE/SEI 7-05 Referenced Seismic Qualification Testing Protocol 23
24 AC-156 Seismic Qualification by Shake Table Testing of Nonstructural Components Companion Document to 2009 IBC/ASCE 7-05 Acceptance Criteria published by ICC Evaluation Services First published in 2000, latest version 2007 Provides testing protocol and test spectra definition iti Test Spectra is tied directly to F p force equation Acceptance Criteria tied to I p factor. For I p = 1.5, Equipment to remain functional following the test 24
25 Translating Building Code Requirement to Test Protocol 25
26 Seismic Simulation Test Results and Acceptance Criteria Example 26
27 Seismic Simulation Test Results and Acceptance Criteria Acceptance Criteria Account for post event pass/fail criteria Post event functionality for critical facilities Photo by: Philip Caldwell 27
28 Seismic Simulation Test Results and Acceptance Criteria Account for post event pass/fail criteria Ip= 1.5 Post test functionality verified Photo by: Philip Caldwell 28
29 Seismic Certification by Experience Requirements for Qualification by Experience Data (for OSHPD) Characterization of experience motions as measured by instruments t Verification of post-earthquake Functionality and Structural Integrity Equipment must be of equal or better construction Project seismic demand must be less than Certification basis Requires maintenance of data base for continuous evaluation of the performance of installed equipment experiencing new earthquake 29
30 HOSPITAL FACILITIES SEISMIC SAFETY ACT 1983 (HSSA 83) Hospitals housing patients less able to protect themselves; Legislative intent It is the intent of the Legislature that hospital buildings that house patients who have less than the capacity of normally healthy persons to protect themselves, and that must be reasonably capable of providing services to the public after a disaster. In order to accomplish this purpose, the office shall propose proper building standards for earthquake resistance based upon current knowledge 30
31 CBC 2010, Section 1708A.4 List of Systems Requiring Special Seismic Certification: 1. Emergency and standby power systems including generators, turbines, fuel tanks, and automatic transfer switches. 2. Elevator equipment (excluding elevator cabs). 31
32 CBC 2010, Section 1708A.4 List of Equipment and Components Requiring Special Seismic Certification: 3. Components with hazardous contents (excluding pipes, ducts, and underground tanks). 4. Smoke control fans. 5. Exhaust fans New addition in the CBC 2010 to Ensure That Pressurization in Building will not Interfere with Opening or Closing of Exit Doors. 6. Switchgear. 7. Motor control centers. 8. X-Ray machines in fluoroscopy rooms. 9. CT (Computerized Tomography) Scanners New addition in the CBC 2010 to Ensure that Stroke/Head Injury Patients and Trauma Center/Emergency Rooms will have Adequate Access to Imaging (Virtually all Manufacturer s of CT scanners will have scanners with special seismic certification by July, 2010) 32
33 CBC 2010, Section 1708A.4 List of Equipment and Components Requiring Special Seismic Certification: 10.Air conditioning units. 11. Air handling units. 12.Chillers. 13.Cooling Towers (excluding Cooling Towers designed as nonbuilding structures). 14. Transformers. 15.Electrical substations. 16.UPS (Inverters) and associated batteries. 17.Distribution panels including electrical panelboards, 18.Control panels including fire alarm, fire suppression, preaction, and auxiliary or remote power supplies. 33
34 CBC 2010, Section 1708A.4 Components Exempt from Special Seismic Certification: 1. Equipment and components installed in nonconforming buildings unless the equipment or component provides a service/system or utility to conforming buildings, or building is designated as SPC 3 or higher. 1. Equipment and components weighing not more than 20lbs. supported directly on structures (and not mounted on other equipment or components) with supports and attachments in accordance with a SCE 7 Chapter 13 as modified by Section 1615A. 34
35 CBC 2010, Section 1708A.4 Special Seismic Certification for OSHPD 2 and 3 Buildings: Special Seismic Certification is required for Designated Seismic System (components with importance factor equal to 1.5). Typical components that require special seismic certification in OSHPD 2 & 3 buildings include: Emergency generators Automatic transfer switches Smoke control fans Medical gas systems Fire alarm panels 35
36 CBC 2010 Section 1708A.4/ASCE 7-05 Section : 2 Special Seismic i Certification Special Seismic Certification for Designated Seismic System, ASCE/SEI 7-05 Section and CBC 2010 Section 1708A.4 Analysis Testing Experience Data ASCE/SEI7-05 Sections 13.3, 13.4, 13.5, & 13.6 ASCE/SEI 7-05 Section ASCE/SEI 7-05 Section Project Specific Submittal to OSHPD OSHPD Special Seismic Certification Preapproval (OSP) 36
37 OSHPD Special Seismic Certification Preapproval (OSP) OSP requirements of a product line with similar structural configuration: Testing of at least two units (typically the smallest & largest units), and Verification of similarities il iti for interpolated t units (CBC 2010 Section 1708A.4). 37
38 Special Seismic Certification By a Single Test One dynamic test is Acceptable when: A single product is certified NOT a product line with more than one product with variations; and, Manufacturing process is ISO 9001 certified 38
39 OSP Requirements (Continued) OSHPD strongly recommends: The Owner and his California licensed structural engineer meet and review the proposed test plan with OSHPD prior to any testing to ensure that there will be no issues with the OSP approval when submitted. 39
40 Statutory OSHPD Review Requirements Hospitals housing patients less able to protect themselves; Legislative Intent (a) It is the intent of the Legislature that hospital buildings that house patients who have less than the capacity of normally healthy persons to protect themselves, and that must be reasonably capable of providing services to the public after a disaster, shall be designed and constructed t to resist, insofar as practical, the forces generated by earthquakes, gravity, and winds. In order to accomplish this purpose, the office shall propose proper building standards for earthquake resistance based upon current knowledge, and provide an independent d review of the design and construction of hospital buildings. 40
41 Regulatory OSHPD Approval Requirements CBC A.3 Enforcement Agency Approval. In addition to requirements of California Code of Regulations (C.C.R.) Title 24, Parts 1 & 2, any aspect of project design, construction, quality assurance, or quality control programs for which this code requires approval by the design professional, are also subject to approval by the enforcement agency. 41
42 Is Third Party Review Subject to OSHPD Review and Approval? Yes Required by Statutet t Required by the CBC 2010 Section 1601A.3 42
43 No Can the Analysis be Bookended Similar to Testing? Analysis is only good for the unit analyzed limited variations and/or grouping may be acceptable i.e. analysis of the smallest and largest buildings (under OSHPD jurisdiction) can t be used for permitting all OSHPD buildings (of a given structural t configuration). However, testing of a few connections to verify an analysis methodology can justify all connections of similar configurations 43
44 What R p Can be Used for Special Seismic Certification? R p = 1.0 & I p = 1.5 for Special Seismic Certification by Analysis (Attachments design can use R p in ASCE 7), Since Functionality can t be Verified by Analysis R p/i p = 1.0 for Special Seismic Certification by Shake Table Test. Both Structural Integrity and Functionality shall be maintained after shake table test. 44
45 Why is R p based on the AC-156 Requirements? ASCE 7-05 does not provide R p values for special seismic certification of active electrical and mechanical equipment, since it does not provide analysis as an option for special certification of active mechanical and electrical equipment. ICC-ES AC 156 requires R p/i p = 1 for Shake Table Test. Require Post-Test Structural Integrity for Ip = 1 Require Post-Test Structural Integrity and Functionality for Ip = 1.5 Critical equipment and components should remain essentially elastic for post-earthquake functionality 45
46 Can the Active Mechanical & Electrical Equipment get an OSP by analysis alone? No Active and energized components must be certified by shake table testing or experience data. ONLY connecting elements, attachments, and supports can be justified by supporting analysis for active and energized components 46
47 Does the Design need to comply with the CBC 2010 for OSP? Yes All components shall be designed in accordance with the CBC requirements Both demand and capacity shall be based on the CBC and referenced material standards. Stability and local affects shall be part of the design as required by the code. The CBC does not permit 1/3 rd allowable stress increase for most materials, including steel and concrete. 47
48 OSHPD Special Seismic Certification Preapproval (OSP) at OSHPD Website OSHPD Special Seismic Certification Preapproval (OSP) Requirements and Application form is available at: List of Equipment and Components with OSHPD Special Seismic Certification Preapproval (OSP) is available at: Approval/special_seismic_cert_pre-approval.html 48
49 Progress in OSHPD Special Seismic Certification Preapproval (OSP) More than seventy Special Seismic Certification Pre-Approvals to date (OSP Program started in July, 09). 10,000 Equipment Currently processing on average more than three (3) new OSP s per week Most Categories have Multiple Pre-approvals 49
50 Options When No Equipment with Special Seismic Certification is Available Differ the Special Seismic Certification and Require the Manufacturer to Provide the Certification Certification must be complete prior to installation. Emergency Waiver OSHPD may waive the requirements for emergency replacements 50
51 Yes Does HVAC Unit for CT Scanner Room Require Special Seismic Certification for the CBC 2007? The CBC 2007 Section 1708A.5 require special seismic certification of all equipment with importance factor (I p ) of 1.5. The CBC 2010 will require CT Scanners to have special seismic certification (Every Manufacturer of CT scanners are expected to have qualified scanners by July, 2010) If CT Scanners are out of service, that room can be converted to another service following earthquake provided all support equipment are functional. All rooms may become essential after an earthquake. Loss of equipment such as acid filled UPS batteries may create a hazardous materials situation, which may affect other functions in the hospital. 51
52 Closing Comments Thank you for viewing the web seminar. For questions regarding seismic issues, contact Roger Richter at (916) or Other questions, please contact Liz Mekjavich at (916) or
Facilities Development Division. The Building Department for California s Hospitals
OSHPD Special Seismic Certification Preapproval (OSP) Preapproval of Special Seismic Certification by OSHPD is a voluntary program for Equipment Manufacturers! Facilities Development Division The Building
More informationAT&T Internet Data Center Site Specification Chicago Area (Lisle, IL)
AT&T Internet Data Center Site Specification Chicago Area (Lisle, IL) Site Biography Market Area: Chicago, IL Location: Lisle, IL 60532 Address: 4513 Western Avenue CLLI,POP, NPA/NXX Code: LSLEILAA, EQLAILAA
More informationKey Considerations for Selecting Server Cabinets for Seismic Areas
Key Considerations for Selecting Server Cabinets for Seismic Areas Protect Up to 1500 of Critical Network Equipment without Sacrificing Flexibility Planning a data center in a seismically active area requires
More informationThe Fundamentals of Seismic Compliance and Heating, Ventilation and Air-Conditioning Systems (HVAC)
Carrier Engineering Newsletter Volume 1, Issue 2 The Fundamentals of Seismic Compliance and Heating, Ventilation and Air-Conditioning Systems (HVAC) Earthquakes are a real problem and not just in California
More informationSUBJECT CAN: 2-108. Temporary Systems, Utilities and Equipment. Effective: 4/09/2013. CODE SECTION 2010 California Building Code (CBC)
Facilities Development Division Office of Statewide Health Planning and Development 400 R Street, Suite 200 Sacramento, CA 95811 (916) 440-8300 700 N. Alameda Street, Suite 2-500 Los Angeles, CA 90012
More informationAT&T Internet Data Center Site Specification Washington-DC Area (Ashburn, VA)
AT&T Internet Data Center Site Specification Washington-DC Area (Ashburn, VA) Site Biography Market Area: Washington, DC Area Location: Ashburn, VA 20147-6011 Address: All deliveries should be sent to
More informationHow To Visit An Internet Data Center
AT&T Internet Data Center Site Specification Atlanta Area (Lithia Springs, GA) Site Biography Market Area: Atlanta and surrounding areas (Georgia, Alabama, Tennessee, Florida, North and South Carolina
More informationAT&T Internet Data Center Site Specification - Phoenix Area (Mesa, AZ)
AT&T Internet Data Center Site Specification - Phoenix Area (Mesa, AZ) Site Biography Market Area: Phoenix, Mesa, Tempe, Chandler, Scottsdale, AZ Location: Mesa, AZ 85201 Address: 1301 University Blvd.
More informationEarthquakes and Data Centers
7x24 Exchange Fall Symposium September 11, 2013 Hilton Bellevue Andrew W. Taylor, Ph.D., S.E., FACI Earthquake Hazards 2 September 11, 2013 1 Cascadia Earthquake Sources Figure Credit: Craig Weaver, Pacific
More informationAT&T Internet Data Center Site Specification San Diego Koll (San Diego, CA)
AT&T Internet Data Center Site Specification San Diego Koll (San Diego, CA) Site Biography Market Area: San Diego, CA Location: San Diego, CA 92121 Address: 5732 Pacific Center Blvd CLLI,POP, NPA/NXX Code:
More informationELECTRICAL EMERGENCY SYSTEM PLAN REVIEW OVERVIEW OF CODE REQUIREMENTS
ELECTRICAL EMERGENCY SYSTEM PLAN REVIEW What to Look for during Plan Review OVERVIEW OF CODE REQUIREMENTS First determine the type of emergency power system; is the system designed for Emergency or Legally
More informationAT&T Internet Data Center Site Specification Dallas Area (Allen, Tx)
AT&T Internet Data Center Site Specification Dallas Area (Allen, Tx) Site Biography Market Area: Dallas and Fort Worth Areas of Texas Location: Allen, Tx 75013 Address: 900 Ventura Drive CLLI, POP, NPA/NXX
More informationThe State Fire Marshal's Office and the VDH/Office of Licensure and Certification. Generator Testing in Nursing Facilities
Generator Testing in Nursing Facilities Introduction Recent emergencies and natural disasters have emphasized the need to ensure that nursing facilities have an adequate EPSS in place should the loss of
More informationSeismic Assessment of Engineering Systems in Hospitals A Challenge for Operational Continuity
Seismic Assessment of Engineering Systems in Hospitals A Challenge for Operational Continuity S.R. Uma GNS Science, Lower Hutt. G.J. Beattie Building Research Association of New Zealand, Porirua. 2010
More informationRESIDENTIAL PHOTOVOLTAIC SUBMITTAL REQUIREMENTS
DEVELOPMENT AND RESOURCE MANAGEMENT DEPARTMENT BUILDING AND SAFETY SERVICES 2600 Fresno Street, Third Floor Fresno, California 93721-3604 (559) 621-8082 FAX (559) 498-4357 RESIDENTIAL PHOTOVOLTAIC SUBMITTAL
More informationPrepared For San Francisco Community College District 33 Gough Street San Francisco, California 94103. Prepared By
Project Structural Conditions Survey and Seismic Vulnerability Assessment For SFCC Civic Center Campus 750 Eddy Street San Francisco, California 94109 Prepared For San Francisco Community College District
More informationFacilities Maintenance Standards rev 9-1-2015
Facilities Maintenance Standards rev 9-1-2015 Purpose The purpose of these standards is to outline the minimum requirements for maintaining state owned facilities and infrastructures in a manner that will
More informationGuidelines for doing Mechanical & Electrical work At the University of New Mexico
Guidelines for doing Mechanical & Electrical work At the University of New Mexico {Revised on 04/14/2015 @ LHC} Table of Contents Cover letter...2 Contractor Information: Performing Work at the University
More informationINSPECTION AND TESTING OF EMERGENCY GENERATORS
INSPECTION AND TESTING OF EMERGENCY GENERATORS Certification requirements NFPA 101(00), Sec. 7.9.2.3 requires that emergency generators be installed, tested and maintained in accordance with NFPA 110,
More informationFire Alarm System Plans Submittal Guidelines For New and Existing Systems
Fire Alarm System Plans Submittal Guidelines For New and Existing Systems SCOPE The Temecula Fire Prevention Bureau (TFPB) has established the following requirements for the submittal of all fire alarms,
More informationNational Archives and Records Administration (NARA) Facility Standards for Federal Records Storage Facilities
National Archives and Records Administration (NARA) Facility Standards for Federal Records Storage Facilities A quick overview of what Federal Records Officers need to know about: What the Records Storage
More informationRhode Island State Fire Marshal Fire Protection Systems Narrative Reports. What is a Narrative Report?
RHODE ISLAND DIVISION OF STATE FIRE MARSHAL 1951 Smith Street, North Providence, RI 02911 Telephone: (401) 383-7717 Fax: (401) 383-8879 John E. Chartier, EFO, CFI State Fire Marshal Rhode Island State
More informationRecommended Product Specifications Fuel Day Tank System Tramont UTRS Fuel Day Tank
Recommended Product Specifications Fuel Day Tank System Tramont UTRS Fuel Day Tank This specification describes requirements for a Diesel Fuel Day Tank System consisting of one or more fuel tanks, an Electronic
More informationNeuStar Ultra Services Physical Security Overview
NeuStar Ultra Services Physical Security Overview Mission-critical network operations must have the highest standards of quality, security and reliability. NeuStar Ultra Services currently locates its
More informationAT&T Internet Data Center Site Specification New York III (Secaucus, NJ)
AT&T Internet Data Center Site Specification New York III (Secaucus, NJ) Site Biography Market Area: New York City, New Jersey, Upstate NY and Lower Connecticut Location: Secaucus, NJ 07094 Address: 15
More informationThe ACI 562 Repair Code
The ACI 562 Repair Code Code Requirements for Evaluation, Repair and Rehabilitation of Concrete Buildings by Keith Kesner 1 Chair ACI 562 Lawrence Kahn 2 Former Chair ACI 562 1. Associate WDP & Associates,
More informationASME QRO Certification Program
ASME QRO Certification Program ASME QRO Certification For Operators of Resource Recovery Facilities Applicant Information Table of Contents Table of Contents Page QRO Certification Overview...1 Types of
More informationSEISMIC EVALUATION OF HOSPITAL PIPING SYSTEMS
13 th World Conference on Earthquake Engineering Vancouver, B.C., Canada August 1-6, 2004 Paper No. 1081 SEISMIC EVALUATION OF HOSPITAL PIPING SYSTEMS Elliott GOODWIN 1, Emmanuel MARAGAKIS 2, Ahmad ITANI
More informationWinnipeg Fire Department Fire Prevention Branch
Winnipeg Fire Department Fire Prevention Branch Manitoba Fire Code Life-Safety Equipment Maintenance Requirements The Fire Prevention Branch of the Winnipeg Fire Department has prepared this document of
More informationGeneral Fire Code Requirements Within Commercial Tenant Spaces
FIRE DEPARTMENT in cooperation with Scope: General Fire Code Requirements Within Commercial Tenant Spaces This handout is designed to assist owners, architects, contractors and others in understanding
More informationFLORIDA ATLANTIC UNIVERSITY FIRE ALARM SYSTEM INSTALLATION MANUAL
FLORIDA ATLANTIC UNIVERSITY FIRE ALARM SYSTEM INSTALLATION MANUAL January, 2003 Environmental Health and Safety Florida Atlantic University 777 Glades Rd. Boca Raton, FL 33431 Phone: 561-297-3129 Fax:
More informationSECTION 8 Industrial Facilities Field Investigation
SECTION 8 Industrial Facilities Field Investigation Types of Data to Be Collected and Recorded Warehouses, manufacturing facilities, energy-producing facilities, and factories should be inspected for performance
More informationAppendix B. 2010 Florida Building Code Building Section 423.25 Public Shelter Design Criteria
Appendix B 2010 Florida Building Code Building Section 423.25 Public Shelter Design Criteria 423.25 Public shelter design criteria. 423.25.1 New facilities. New educational facilities for school boards
More informationABSTRACT DCMCCT PROJECT OVERVIEW
Title: Case Study of the Austin Energy Dell Regional Children s Medical Center of Central Texas Packaged Hybrid CHP Energy Plant in Austin, Texas Authors: Mr. Ed Mardiat, Burns & McDonnell Mr. Jim Teigen,
More information26 3213.13 Diesel Engine Driven Generators Page 1 of 6
Last Update: December 8, 2014 A. Description of System Consultant s Handbook Page 1 of 6 1. Provide a diesel engine driven electric generating unit, factory assembled, tested and certified to operate at
More informationSafety, Health and Accessibility Design Guidelines For Architects, Engineers and Design Firms
Safety, Health and Accessibility Design Guidelines For Architects, Engineers and Design Firms U.S. Department of the Interior Bureau of Indian Affairs Division of Safety and Risk Management Introduction
More informationSECTION 13850 (28 31 00) FIRE DETECTION AND ALARM SYSTEM
SECTION 13850 (28 31 00) FIRE DETECTION AND ALARM SYSTEM ENGINEERING SPECIFICATION INTELLIGENT REPORTING FIRE DETECTION SYSTEM PART 1 GENERAL 1.1 DESCRIPTION A. This specification includes the furnishing,
More informationEXPRESS PERMITS (No Plan Check Required)
EXPRESS PERMITS (No Plan Check Required) Following are the lists of different types of construction for which an express permit can be issued. Express permits can be issued where work does not require
More informationEnergy-efficient data center achieves 2N reliability; MTU Onsite Energy generator sets provide critical backup
Press Release Contact: Gary Mason Tognum America Inc. Phone: +1 248 560 8480 E-mail: gary.mason@tognum.com Agency contact: Robert E. Sheldon, APR Creative Communications Consultants, Inc. Phone: +1 210
More informationISO IEC 27002 2005 (17799 2005) INFORMATION SECURITY AUDIT TOOL
9.1 USE SECURITY AREAS TO PROTECT FACILITIES 1 GOAL Do you use physical methods to prevent unauthorized access to your organization s information and premises? 2 GOAL Do you use physical methods to prevent
More informationDELAWARE STATE FIRE MARSHAL FIRE ALARM SYSTEM CERTIFICATE OF INSTALLATION
Phone System Owner Phone System Engineered by: of Record Drawings: of System Manuals: of Test Reports: System Installer 1. GENERAL INFORMATION Phone Delaware License No: Certificate Holder No. System Supplier
More information2012 Michigan Rehabilitation Code for Existing Buildings ISBN: 978-1-60983-208-7
2012 Michigan Rehabilitation Code for Existing Buildings First Printing: June 2014 ISBN: 978-1-60983-208-7 COPYRIGHT 2014 by INTERNATIONAL CODE COUNCIL, INC. ALL RIGHTS RESERVED. This 2012 Michigan Rehabilitation
More informationAFTER THE BIG ONE, WILL CRITICAL COUNTY SERVICES SURVIVE?
AFTER THE BIG ONE, WILL CRITICAL COUNTY SERVICES SURVIVE? SUMMARY There is a 67% probability that Southern California will be hit by a magnitude 6.7 or larger earthquake in the next 30 years according
More informationMaintaining Fire Protection Systems
Maintaining Fire Protection Systems Fire continues to be a major threat to your business and your personal safety. Each year several thousand people are injured or killed by fires and billions of dollars
More informationALBERTA FIRE CODE SAFETY EQUIPMENT MAINTENANCE REQUIREMENTS
ALBERTA FIRE CODE SAFETY EQUIPMENT MAINTENANCE REQUIREMENTS In this document are the fire safety equipment maintenance requirements that are found in Division B of the Alberta Fire Code (AFC 2006) for
More informationEXETER TOWNSHIP DRY OR WET CHEMICAL FIRE SUPPRESSION SYSTEMS
DRY OR WET CHEMICAL FIRE SUPPRESSION SYSTEMS POLICY STATEMENT This policy applies to NEW Dry or Wet chemical fire suppression systems. The Authority Having Jurisdiction will not waive major criteria in
More informationFire Alarm Systems Certification
Fire Alarm Systems Certification The skills and knowledge listed under each task are suggestive of those involved in that task, but are not intended to constitute an exhaustive listing. 2.1 Submittal Preparation
More informationFire Pump Plan Review March 2010
Fire Pump Plan Review March 2010 Date of Review: / / Permit Number: Business/Building Name: Address of Project: Designer Name: Designer s Phone: Contractor: Contractor s Phone: Occupancy Classification:
More informationHOW TO EVALUATE BUILDINGS AND DETERMINE RETROFIT COSTS
David Bell PJHM Architects, Inc. Young Nam Daniel Traub Thornton Tomasetti how to evaluate buildings and determine retrofit costs HOW TO EVALUATE BUILDINGS AND DETERMINE RETROFIT COSTS Presented by: David
More informationSINGLE-FAMILY EARTHQUAKE (SEISMIC) RETROFITS
CITY OF BEAVERTON Community Development Department Building Division 12725 SW Millikan Way / PO Box 4755, Beaverton, OR 97076 Phone: (503) 526-2493 Fax: (503) 526-2550 General Information (503) 526-2222
More informationSECTION 16721 FIRE ALARM AND DETECTION SYSTEM
1. GENERAL 1.1. Related Documents: 1.2. Scope: 1.1.1. The general provisions of the Contract, including General and Supplementary Conditions and General Requirements apply to the work specified in this
More informationCAN/ULC S1001-11 Integrated Systems Testing of Fire Protection and Life Safety Systems and Fire Protection Commissioning
CAN/ULC S1001-11 Integrated Systems Testing of Fire Protection and Life Safety Systems and Fire Protection Commissioning Presentation To: Canadian Fire Alarm Association (CFAA) Alberta Technical Seminar
More informationAPEC CAPABILITIES PROJECT MANAGEMENT / SUPPORT
PROJECT MANAGEMENT / SUPPORT Capital Cost Savings Construction Monitoring and Assistance In-Plant Production Assistance In-Plant Maintenance Assistance Procurement Assistance Project Cost Estimating o
More informationLocation. Central Business District. Ataturk Airport IFC. Sabiha Gökçen Airport
Welcome Zenium Istanbul One provides 12,000 square metres of high specification technical space across three purpose-built data centers. It is Turkey s first wholesale data center offering; the data halls
More information527 CMR: BOARD OF FIRE PREVENTION REGULATIONS
527 CMR 31.00: CARBON MONOXIDE ALARMS Section 31.01: Purpose and Scope 31.02: Definitions 31.03: General Installation Provisions 31.04: Specific Installation Provisions 31.05: Carbon Monoxide Protection:
More informationCyberFortress Data Center at Innovation Park Infrastructure and Topology Guide. Innovation Park Charlotte. NC 28262. DRAFT V0.
Innovation Park Charlotte. NC 28262 DRAFT V0.1 Page 2 CyberFortress Data Center at Innovation Park BECO South purchased Innovation Park in March 2010 and immediately launched a dramatic top-to-bottom $30M
More informationinimi ing damage and repair osts to manufa tured omes during an eart ua e
inimi ing damage and repair osts to manufa tured omes during an eart ua e Prepared for T e Department of Housing and Urban Development Offi e of Poli y Development and eesear by t e National Conferen e
More informationG. Michele Calvi IUSS Pavia
CONVEGNO SISMA ED ELEMENTI NON STRUTTURALI Approcci, Stati Limite e Verifiche Prestazionali Bologna 24 ottobre 2014 PROGETTO DI ELEMENTI NON STRUTTURALI SOGGETTI AD AZIONI SISMICHE G. Michele Calvi IUSS
More informationPLAN REVIEW GUIDE FOR FIRE ALARM
PLAN REVIEW GUIDE FOR FIRE ALARM PROJECT NAME: PERMIT # PROJECT ADDRESS: CONTACT PERSON: PHONE Fire alarm system installation information shall be provided on the appropriate architectural and electrical
More informationCITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS
CITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS PROPERTY ADDRESS (BRT Address Required) TESTING CONTRACTOR (Name and Address) License No. ANNUAL
More informationTELECOMMUNICATION SYSTEM HAZARD MITIGATION STRATEGIC PLANNING
TELECOMMUNICATION SYSTEM HAZARD MITIGATION STRATEGIC PLANNING A.K. Tang 1 1 President, L&T Engineering and Project Management Consultant, Mississauga. Canada Email: alexktang@mac.com ABSTRACT: Telecommunication
More informationNEW JERSEY CENTER OF EXCELLENCE
Overview NEW JERSEY Page 1 The Plant (CUP) provides electricity via a cogeneration system, chilled water for environmental cooling, steam for heating, and compressed air primarily for HVAC control Serves
More informationArc Terminator Active Arc-Resistant Switchgear
Arc Terminator Active Arc-Resistant Switchgear Increasing safety and productivity by extinguishing internal arcing faults within the switchgear. The Square D Arc Terminator from Schneider Electric offers
More informationPrime Power Services, Inc. Georgia, Carolinas 800 741 0128. NFPA 110 Standards for Emergency & Stand by Generators June 1, 2010
Prime Power Services, Inc. Georgia, Carolinas 800 741 0128 NFPA 110 Standards for Emergency & Stand by Generators June 1, 2010 NFPA Disclaimer The views and opinions expressed in this seminar are purely
More informationUCLA Policy 811: Environmental Health and Safety
UCLA Policy 811: Environmental Health and Safety Issuing Officer: Assistant Vice Chancellor, General Services Responsible Dept: Environment, Health & Safety Effective Date: September 5, 2007 Supersedes:
More informationMOBILE FIRE - RESCUE DEPARTMENT FIRE CODE ADMINISTRATION
MOBILE FIRE - RESCUE DEPARTMENT FIRE CODE ADMINISTRATION Fire Pump Plan Review 2009 International Fire Code and NFPA 20 Date of Review / / BLD201 - Project Address: Project Name: Contractor s Business
More informationNATIONAL FIRE ALARM AND SIGNALING CODE
GAP.11.1.1.0 A Publication of Global Asset Protection Services LLC NATIONAL FIRE ALARM AND SIGNALING CODE INTRODUCTION National Fire Protection Association (NFPA) documents describe a level of fire protection
More informationUniversity Design Standard. Fire Alarm Systems
University Design Standard Fire Alarm Systems 1.0 Overview This document provides Virginia Tech s standards for the design and installation of fire alarm systems and equipment in new construction, retrofit
More informationButch Browning State Fire Marshal, Bobby Jindal Governor - Licensing Requirements
LOUISIANA DEPARTMENT OF PUBLIC SAFETY AND CORRECTIONS OFFICE OF STATE FIRE MARSHAL, CODE ENFORCEMENT AND BUILDING SAFETY H. Butch Browning State Fire Marshal Bobby Jindal Governor Licensing Law and Rules
More informationCanadian Regulatory Compliance for Stationary Battery Systems North American Update
Canadian Regulatory Compliance for Stationary Battery Systems North American Update ComplianceKnowledgeCenter.com What is Compliance? In general: compliance means conforming to a rule, such as: a specification,
More informationDESIGN CRITERIA. Determination of earthquake demand. Building importance level. Component categories SEISMICALLY RESILIENT NON-STRUCTURAL ELEMENTS
FACT SHEET 2 NZS 4219:2009 Seismic performance of engineering systems in buildings requires the arrangement and layout of engineering systems within a building to allow for earthquake actions. NZS 4219:2009
More informationNational Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook Trainee:... Training Provider:... Year:... Trainee Contact no:... Competency Based Skill Training Logbook Introduction
More informationNIST Disaster Resilience Framework
NIST Disaster Resilience Framework Communication and Information Systems Presenter: David Mizzen Applied Research Associates Chapter Goals Provide guidance to: 1. Understand potential vulnerabilities/damage
More informationBuilding Code of New York State
Welcome Christopher R. Paolini, P.E., Sr. Vice President CME Associates, Inc. 6035 Corporate Drive East Syracuse, New York 13057 (315) 701-0522 (315) 701-0526 (Fax) cpaolini@cmeassociates.com 1 Building
More informationWelcome UST Owners & Operators!!
Welcome UST Owners & Operators!! Underground Storage Tank Owner & Operator Workshop Presented By: Kern County Environmental Health CUPA Kern County Environmental Health Phone: (661) 862-8740 Email: hazmatprogram@co.kern.ca.us
More informationUL Fire Door Inspection Service. UL Fire Alarm Certification Program. A valuable but misunderstood resource for jurisdictions
2013 Issue 2 2 A Strong Foundation 3 Introducing UL Fire Door Inspection Service 6 Enhanced UL Marks Coming Your Way Soon UL Fire Alarm Certification Program A valuable but misunderstood resource for jurisdictions
More informationENGINEERING-BASED EARTHQUAKE RISK MANAGEMENT
ENGINEERING-BASED EARTHQUAKE RISK MANAGEMENT MRP Engineering Newsletter February 2012 The world recently experienced several major earthquakes, which caused severe local impacts and major worldwide repercussions.
More informationSample Company SECTION 019113 GENERAL COMMISSIONING REQUIREMENTS PART 1 GENERAL 1.1 SUMMARY AND CONTRACT RESPONSIBILITIES
SECTION 019113 GENERAL COMMISSIONING REQUIREMENTS PART 1 GENERAL 1.1 SUMMARY AND CONTRACT RESPONSIBILITIES A. The Owner will contract separately and directly for the majority of the commissioning work
More informationBUILDING OCCUPANCY RESUMPTION PROGRAM
BUILDING OCCUPANCY RESUMPTION PROGRAM Revised JANUARY 2011 CITY AND COUNTY OF SAN FRANCISCO DEPARTMENT OF BUILDING INSPECTION EMERGENCY OPERATIONS PLAN For any questions regarding this program, please
More informationENVIRONMENTAL HEALTH & SAFETY 6 East Dr., Amherst MA 01002-5000
AMHERST COLLEGE ENVIRONMENTAL HEALTH & SAFETY 6 East Dr., Amherst MA 01002-5000 Michael Zlogar April 11, 2001 Assistant Fire Chief Amherst Fire Department 68 North Pleasant St. Amherst, MA 01004 FIRE ALARM
More informationPublished in the Official State Gazette (BOE) number 166 of July 10th 2009 [1]
Nuclear Safety Council Instruction number IS-22, of July 1st 2009, on safety requirements for the management of ageing and long-term operation of nuclear power plants Published in the Official State Gazette
More informationCx for Enhanced Healthcare Project Delivery. Presentation Format. Project features and timeline
Cx for Enhanced Healthcare Project Delivery Rick Staley Sr. Construction Manager PH&S Joe Webb Construction Manager PH&S Richard Beam Director, Energy Management Services PH&S Todd McGuire, PE - Sr. Project
More informationBuilding Permit Application
Building Permit Application Administrative Use Only PARENT PERMIT NUMBER (Required if part of another project) PERMITTEE: PERMITTEE ADDRESS: E-MAIL ADDRESS: ARCHITECT/ENGINEER: PHONE: PHONE: ARCHITECT/ENGINEER
More informationDirector of Operations. Lattice Communications & Industrial Energy Applications
Director of Operations Lattice Communications & Industrial Energy Applications The Case for Remote Monitoring How confident are you that your generator will be available when you need it? How can you be
More informationSECTION 26 05 49 VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS
SECTION 26 05 49 VIBRATION AND SEISMIC CONTROLS FOR ELECTRICAL SYSTEMS PART 1 - GENERAL 1.1 SUMMARY A. This Section includes seismic restraints and other earthquake-damage-reduction measures for electrical
More informationISSUE BRIEF. Understanding the Hospital Planning, Design, and Construction Process. Introduction. Background. Phase 1: Project Definition and Planning
Understanding the Hospital Planning, Design, and Construction Process Introduction California s acute care hospitals are an essential health care resource that serve a wide range of patients during times
More informationFIRE ALARM AND DETECTION SYSTEMS SECTION 16721
PART 1 - GENERAL 1.01 WORK INCLUDED FIRE ALARM AND DETECTION SYSTEMS SECTION 16721 A. Provide a complete fully addressable, power limited, fire detection and evacuation system. The system shall be connected
More informationState of Iowa Electrician Licensing Program Brian Young Interim Executive Secretary Of The Electrical Examining Board Division Of The State Fire Marshal s Office Iowa Code 103 Electrician Licensing & Inspections
More informationPOSITION STATEMENT THE ENGINEER AND THE ENGINEERING TECHNICIAN DESIGNING FIRE PROTECTION SYSTEMS. July 28, 2008
SOCIETY OF FIRE PROTECTION ENGINEERS (SFPE) NATIONAL SOCIETY OF PROFESSIONAL ENGINEERS (NSPE) NATIONAL INSTITUTE FOR CERTIFICATION IN ENGINEERING TECHNOLOGIES (NICET) POSITION STATEMENT THE ENGINEER AND
More information2012 Ontario Building Code Changes Affecting Fire Alarm Systems
2012 Ontario Building Code Changes Affecting Fire Alarm Systems CFAA Annual Technical Seminar May 31, 2013 Ken Baird VP Systems www.lrifire.com LRI 2010 Developments in Life Safety Introduction Code Development
More information2015 C&I HVAC RETROFIT APPLICATION FOR PRESCRIPTIVE REBATES
2015 C&I HVAC RETROFIT APPLICATION FOR PRESCRIPTIVE REBATES WELCOME TO THE POWER MOVES C&I HVAC RETROFIT PROGRAM. We look forward to working with you. If you have questions after reading this application,
More informationHERNANDO COUNTY FIRE ALARM GUIDELINES
HERNANDO COUNTY FIRE ALARM GUIDELINES To be used for all commercial building fire alarm system plan review. The procedures set forth in this document are the minimum requirements necessary to ensure a
More informationThis is a Master s degree level apprenticeship which includes academic learning combined workplace learning and training.
Power Engineer Degree Apprenticeship The UK s energy sector is integral to the prosperity and stability of the UK economy as a whole, delivering an absolutely fundamental service for all its citizens.
More informationELECTRICAL ENGINEERING DESIGN CRITERIA APPENDIX F
ELECTRICAL ENGINEERING DESIGN CRITERIA APPENDIX F TABLE OF CONTENTS Appendix F - Electrical Engineering Design Criteria F.1 Introduction...F-1 F.2 Codes and Standards...F-1 F.3 Switchyard and Transformers...F-1
More informationBT300 HVAC Drives Conventional Bypass (C-Bypass) Options
BT300 HVAC Drives Conventional Bypass (C-Bypass) Options Description The BT300 Conventional Bypass is a companion package for the family of BT300 HVAC Drives. For information on the family of BT300 HVAC
More informationSECTION THREE CONSTRUCTION CODES
SECTION THREE CONSTRUCTION CODES CHAPTER 36: Construction Codes in Michigan Purpose and Applicability of Regulations Manufacturing facilities and other businesses must be designed according to code in
More informationEnergy Design Resources Commissioning Plan Outline Template
IMPORTANT NOTICE: This sample document is provided for instructional purposes only. CCC is not rendering advice concerning any commission project or practices. This document is neither approved nor intended
More informationAPPLICATION PACKAGE FOR LISTING AS A WATER AND/OR WASTEWATER INFRASTRUCTURE PROVIDER FOR DESIGNERS, CONSTRUCTORS
APPLICATION PACKAGE FOR LISTING AS A WATER AND/OR WASTEWATER INFRASTRUCTURE PROVIDER FOR DESIGNERS, CONSTRUCTORS (Major & minor works) TO THE MANAGING NEW DEVELOPMENTS PROCESS For information regarding
More information