Wide-field, high-resolution Fourier ptychographic microscopy
|
|
|
- Claire York
- 10 years ago
- Views:
Transcription
1 Wide-field, high-resolution Fourier ptychographic microscopy Guoan Zheng*, Roarke Horstmeyer, and Changhuei Yang Electrical Engineering, California Institute of Technology, Pasadena, CA 91125, USA *Correspondence should be addressed to: Keywords: Ptychography; high-throughput imaging; digital wavefront correction; digital pathology Manuscript information: 11 text pages, 4 figures Supporting materials: 2 text pages, 8 figures, 1 video Abstract: In this article, we report an imaging method, termed Fourier ptychographic microscopy (FPM), which iteratively stitches together a number of variably illuminated, low-resolution intensity images in Fourier space to produce a wide-field, high-resolution complex sample image. By adopting a wavefront correction strategy, the FPM method can also correct for aberrations and digitally extend a microscope's depth-of-focus beyond the physical limitations of its optics. As a demonstration, we built a microscope prototype with a resolution of 0.78 μm, a field-of-view of approximately 120 mm 2, and a resolution-invariant depth-of-focus of 0.3 mm (characterized at 632 nm). Gigapixel color images of histology slides verify FPM's successful operation. The reported imaging procedure transforms the general challenge of high-throughput, highresolution microscopy from one that is coupled to the physical limitations of the system's optics to one that is solvable through computation. The throughput of an imaging platform is fundamentally limited by its optical system s spacebandwidth product (SBP) 1, defined as the number of degrees of freedom it can extract from an optical signal. The SBP of a conventional microscope platform is typically in megapixels, regardless of its employed magnification factor or numerical aperture (NA). As a reference point, a standard 20 microscope objective (MPLN 20, 0.4 NA, Olympus) has a resolution of 0.8 µm and a 1.1-mm diameter field-of-view (FOV), corresponding to an SBP of approximately 7 megapixels. Increasing the SBP of a microscope is fundamentally confounded by the scaledependent geometric aberrations of its optical elements 1, thus resulting in a compromise between achievable image resolution and FOV. A large SBP in microscopy, however, is highly desired in biomedical applications, such as digital pathology, hematology, immunohistochemistry, and neuroanatomy. The strong need in biomedicine and neuroscience to digitally image large numbers of histology slides for analysis has prompted the commercial development of sophisticated mechanical scanning microscope systems and lensless microscopy setups. Artificially increasing an imaging system s SBP by 1
2 mechanical means is suboptimal, as it requires precise control over actuation, optical alignment, and motion tracking. Furthermore, a mechanical solution simply accepts the intrinsic resolution limit and SBP of a conventional microscope s optics, neglecting the computationally addressable problem of resolution enhancement. Likewise, lensless microscopy methods, such as digital in-line holography 2,3 and contact-imaging microscopy 4-6, offer unique imaging capabilities but also exhibit certain drawbacks. Whereas optically compact, digital in-line holography works well for sparse samples, contact-imaging microscopy requires a sample to be in close proximity to the sensor. Here, we present a computational imaging method that is capable of providing a scalable SBP for most existing microscopes without involving mechanical scanning or phase measurements. While the method does require the acquisition of a plurality of images, it does so non-mechanically and accomplishes its SBP improvement using the plural data to overcome the physical limitations of the system s optics. The imaging method we introduce and demonstrate, termed Fourier ptychographic microscopy (FPM), shares its roots with interferometric synthetic aperture microscopy 7-20, ptychography 21-27, phase retrieval 28-31, light-field imaging 32-35, structured illumination 36, and adaptive optics 37. It works by iteratively stitching together a number of low-resolution images in Fourier space to recover an accurate high-resolution, high-sbp output image. Unlike systems designed to increase the SBP of a single acquisition 38, combining time-sequential measurements for the same goal allows our setup to maintain a simple and compact form factor. The main design strategy of FPM is similar to that of interferometric synthetic aperture microscopy 7-20 : expanding the SBP in Fourier space through multi-image fusion. However, because no measured phase information is needed for FPM, our setup eliminates the design challenges associated with interferometric detection schemes. Furthermore, the image recovery procedure of FPM follows a strategy common with ptychography (i.e., scanning diffraction microscopy) : iteratively solving for a sample estimate that is consistent with many intensity measurements. Unlike ptychography, however, FPM s object support constraints are imposed in the Fourier domain, offering several unique advantages and opportunities. By adding a simple light emitting diode (LED) matrix illumination module and applying FPM s reconstruction algorithm, we transform a conventional optical microscope into a highresolution (0.78 µm, 0.5 NA), wide-fov (~120 mm 2 ) microscope with a final SBP of approximately 1 gigapixel. Our joint optical-digital solution further allows us to exploit adaptive optics-based wavefront correction strategies to compensate for aberrations and expand depthof-focus beyond conventional optical limits. Specifically, we use our FPM procedure to extend a conventional microscope s 80-µm depth-of-focus to approximately 0.3 mm, creating a platform with a large tolerance to microscope slide placement errors. We will briefly outline FPM s operation and experimental setup, discuss how to apply FPM s digital wavefront correction technique, and demonstrate successful gigapixel imaging of a pathology slide. Results Principle of FPM The data collection procedure of FPM is straightforward. We place a 2D sample at the focal plane of a low-na microscope objective and collect a sequence of N images, with the sample successively illuminated by plane waves at N different angles. As a key distinction from other 2
3 synthetic aperture techniques, we only acquire intensity images of the sample no interferometric measurements are needed. The use of a low-na objective lens allows a large FOV to be captured at the expense of a low spatial resolution. In this section, we assume the sample is located at the in-focus position of the microscope objective. Later, we will demonstrate that computational refocusing of a mispositioned sample is also possible. Based on N collected low-resolution intensity images, we computationally reconstruct a high-resolution image of the sample following the recovery procedure shown in Figure 1. Before explaining the procedure, we first note that our recovery process alternates between the spatial (x y) and Fourier (k x k y ) domains, where k represents wavenumber. Second, we assume that illuminating a thin sample by an oblique plane wave with a wave vector (k x, k y ) is equivalent to shifting the center of the sample s spectrum by (k x, k y ) in the Fourier domain. Third, we assume our objective lens s filtering function (i.e., coherent optical transfer function) in Fourier space is a circular pupil with a radius of NA*k 0, where k 0 = 2π/λ is the wavenumber in a vacuum. FPM generates a high-resolution image I h from a set of N low-resolution measurements, I lm k i x, k i y (indexed by their illumination wave vector k i x, k i y, with i = 1, 2 N) as follows. Subscripts h, l, and m denote high-resolution, low-resolution, and measurement, respectively. First, the FPM method starts by making an initial guess of the high-resolution object function in the spatial domain, I h e iφ h. A good starting point is to select φ h =0 and I h as any up-sampled low-resolution image (an initial guess with constant value also works). The Fourier transform of the initial guess creates a broad spectrum in the Fourier domain (Fig. 1, left). Second, we select a small subregion of this spectrum, equivalent to a low-pass filter, and apply Fourier transformation to generate a new low-resolution target image I l e iφ l. The applied low-pass filter shape is a circular pupil, given by the coherent transfer function of the objective lens. The position of the low-pass filter is selected to correspond to a particular angle of illumination. For example, the subregion enclosed by the red circle in Figure 1 corresponds to an image collected under normally incident illumination (k 1 x = 0, k 1 y = 0). Third, following phase retrieval concepts developed by Fienup 28-31, we replace the target image s amplitude component I l with the square root of the low-resolution measurement obtained under illumination angle i, I lm, to form an updated, low-resolution target image I lm e iφ l. We then apply Fourier transformation to this updated target I lm e iφ l and replace its corresponding subregion of the high-resolution Fourier space. In other words, for i = 1, we update the area enclosed by the red circle in Figure 1 with image I lm k 1 x, k 1 y, where k 1 x = 0, k 1 y = 0. Fourth, for the i th -shifted subregion, we repeat steps 2 and 3 (select a small, circular region of k-space and update it with measured image data). Other examples are represented by the green and blue circles in Figure 1. Each shifted subregion corresponds to a unique, lowresolution intensity measurement I lm k i x, k i y, and each subregion must overlap with neighboring subregions to assure convergence. This data redundancy requirement is also present in ptychography 22,39. This iterative update continues for all N images, at which point the entire high-resolution image in Fourier space has been modified with data from all low-resolution intensity measurements. 3
4 Lastly, steps 2 4 are repeated until a self-consistent solution is achieved (for the simulation and experimental data shown in this paper, we only repeat these steps once). At the end of this iterative recovery process, the converged solution in Fourier space is transformed to the spatial domain to recover a high-resolution field I h e iφ h, offering an accurate image of the targeted 2D sample (Fig. 1, right) with a dramatically increased SBP (high-resolution with wide- FOV). A discussion of the computational cost of the above recovery procedure can be found in Supplementary Note 1. We also performed a set of numerical simulations to validate the proposed FPM method in Supplementary Note 2. Drawing connections and distinctions between this iterative process and two related modalities, light-field imaging and ptychography 21-27, helps clarify FPM s principle of operation. A light-field microscope uses a microlens array at its image plane to project M subimages onto its sensor. By extracting different pixels from each subimage, different perspective views can be synthesized, each corresponding to a small aperture of the objective s pupil plane 32,33. Similar to a light-field microscope, the FPM setup also captures multiple perspective images of a sample, corresponding to different, small apertures in the Fourier domain. However, three key differences allow the FPM to create a high-resolution output. First, FPM records larger scattering angles than does a standard light-field microscope. Second, lightfield microscopes sacrifice spatial resolution to acquire all perspective images in one single snapshot 32,33, whereas FPM acquires each perspective over time. Third, light-field microscopes use the interplay between spatial and angular resolvability to achieve refocusing through a 3D sample, whereas FPM applies this interplay to achieve a different goal: different angular perspectives are synthesized to increase a 2D object s spatial resolution. We also note that an FPM dataset of a 3D object can be processed in a similar way as a light-field microscope to achieve 3D sample refocusing and rendering 40. Ptychography is a lensless imaging method originally proposed for transmission electron microscopy and brought to fruition by Faulkner and Rodenburg with the introduction of transverse translation diversity 22,39. The basic idea of ptychography is to illuminate a sample with a focused beam and repeatedly record its far-field diffraction pattern as a function of sample position. Iterative retrieval methods are then applied to invert the diffraction process and recover the sample s amplitude and phase from this set of measurements. It is clear that FPM and ptychography both iteratively seek a complex field solution that is consistent with many intensity measurements. With ptychography, the object support for phase retrieval is provided by the confined illumination probe in the spatial domain; therefore, the sample (or the probe) must be mechanically scanned through the desired FOV. With FPM, however, the object support is provided by the confined NA in the Fourier domain (a circular pupil). In this regard, FPM appears as the Fourier counterpart of ptychography, justifying the proposed name. By imposing object support in the Fourier domain, FPM naturally offers a large, fixed FOV, a higher signal-to-noise ratio (with focusing elements), and no mechanical scanning as compared to conventional ptychography. Furthermore, as discussed in below, FPM can also digitally correct for aberrations common to simple low-na focusing elements. Experimental setup and characterization To experimentally validate the FPM method, we used an Olympus BX 41 microscope, a 2 apochromatic objective lens (Plan APO, 0.08 NA, Olympus), and an interline CCD camera 4
5 (Kodak KAI-29050, 5.5-µm pixel size) as our experimental setup. We then introduced a programmable color LED matrix placed approximately 8 cm under the sample stage as a variable illumination source (Fig. 2a and 2b, also refer to Methods). Resolution improvement provided by the FPM method is demonstrated with a USAF resolution target imaging experiment in Figure 2c and 2d (also refer to Supplementary Video 1). Figure 2c1 shows a full-fov raw intensity image acquired by the FPM platform. Figure 2c2 shows a magnified view of the raw data, with a pixel size of 2.75 µm at the object plane (CCD pixel size divided by the magnification factor). The corresponding high-resolution FPM reconstruction is shown in Figure 2d for comparison, with a maximum synthetic NA is 0.5 set by the maximum angle between the optical axis and an LED. In our FPM reconstruction, the feature of group 9, element 3 on the USAF target (0.78 µm line width) is clearly resolved. This verifies our prototype platform s expected synthetic NA of 0.5, following the Rayleigh criterion (refer to Fig. S2 for FPM reconstructions with different synthetic NAs and Fig. S5 for reconstructed image line traces). In Figure S3, we further determine the depth-of-focus of the proposed platform to be approximately 100 µm without any computational correction applied, which is approximately equal to the 80 µm depth-of-focus associated with the 2 objective used in the experiment, but is approximately 25 fold longer than that of a conventional microscope objective with a similar 0.5 NA. Digital wavefront correction While the FPM method does not require phase information as input, its operation implicitly accommodates phase during iterative reconstruction. As we will demonstrate, the depth-offocus of our FPM prototype can be significantly extended beyond that of the employed objective lens using a numerical strategy to compensate for aberrations in the pupil function 19,41. This digital correction process is inspired by similar wavefront correction concepts in adaptive optics 37. The basic idea is to digitally introduce a phase map to our coherent optical transfer function to compensate for aberrations at the pupil plane during the iterative image recovery process. The FPM algorithm incorporates this compensation into two additional multiplication steps (steps 2 and 5 in Fig. 3a and 3b). Specifically, step 2 models the connection between the actual sample profile and the captured intensity data (with included aberrations) through multiplication with a pupil function e i φ(k x,k y ), whereas step 5 inverts such a connection to achieve an aberration-free reconstructed image. Sample defocus is essentially equivalent to introducing a second-order Zernike mode, or a quadratic phase factor, to the pupil plane (i.e., a defocus aberration 42 ): e i φ(k x,k y ) = e i (2π/λ)2 k 2 x k2 y z 0, k 2 x + k 2 y < (NA 2π/λ) 2, (1) where k x and k y are the wave numbers at the pupil plane, z 0 is the defocus distance, and NA is the objective s numerical aperture. Simulations of the proposed digital wavefront correction strategy are provided in Supplementary Note 2, whereas Figure 3 experimentally demonstrates FPM s ability to fully resolve an object given a set of intensity images defocused by 150 µm. The significance of wavefront correction is made clear by comparing reconstruction results without (Fig. 3d) or with (Fig. 3e) digital addition of a defocused pupil. We note that, in Figure 3e, the defocus distance is known a priori. If the defocus distance is unknown, we can digitally adjust the z parameter to 5
6 different values, reconstruct the corresponding FPM images, and pick the sharpest image through observation or by a computer algorithm. This approach can be extended for tilted samples as well. Here, we digitally adjust the z parameter to achieve acuity for each vignette of the whole image and combine the in-focus vignette to form a focused image of the tilted sample. From Figure 3, we conclude that our FPM prototype can achieve a resolution-invariant depth-offocus of approximately 0.3 mm with digital wavefront correction (Fig. S5). In contrast, the natural depth-of-focus of the employed 2 objective lens (0.08 NA) is approximately 80 µm. The improvement is even more remarkable if compared to an objective lens with a resolutionmatching 0.5 NA, where the FPM prototype s 0.3 mm depth-of-focus offers an approximate factor of improvement of 75. Finally, we note that alternate digital multiplicative phase factors can be included in steps 2 and 5 to correct for a variety of aberrations, as long as they correctly model the employed optics. Following this strategy, we also correct for astigmatism aberrations of our prototype s objective lens (refer to Methods and Supplementary Note 4 for details on pupil function measurement). In Figure S6, we establish that the effective FOV of our prototype is approximately 1.25 cm in diameter (~120 mm 2 ). Gigapixel color imaging of histology slides Color FPM images can be created by simply combining results from red, green, and blue LED illumination into each corresponding color channel. We demonstrate color FPM with our prototype by acquiring a wide-fov color image of a pathology slide (human adenocarcinoma of breast section, Carolina), as shown in Figure 4. Vignette high-resolution views are provided in Figure 4b 4d with a reconstructed pixel size of µm. The imaging FOV is approximately 120 mm 2, the same as that from a 2 objective (Plan APO, 0.08 NA, Olympus), whereas the maximum achieved NA is 0.5, similar to that of a typical 20 objective (MPLN, 0.4 NA, Olympus). The conventional microscope images taken with 20 and 2 lenses are shown for comparison in Figure 4c2 and 4c3. In Figure S7, we include a detailed comparison between raw data, FPM reconstruction, and a conventional microscope image for two samples: a pathology slide and a blood smear. The demonstrated SBP of our FPM prototype is approximately 0.9 gigapixels (120 mm 2 FOV divided by µm 2 Nyquist pixel area, characterized at 632 nm wavelength; refer to Supplementary Note 3). Such a SBP is orders of magnitude larger than that of its constituent 2 objective (16 megapixels) and that of a typical 20 objective (7 megapixels). From another perspective, our FPM prototype can be considered a microscope that combines the FOV advantage of a 2 objective with the resolution advantage of a 20 objective. Discussion We have demonstrated a simple and cost-effective microscopy imaging method, termed Fourier ptychographic microscopy (FPM). This computation-based method is capable of providing a scalable SBP for most conventional microscopes without requiring mechanical scanning. Fundamentally, it transforms the general challenge of high-throughput microscopy from one that is coupled to the physical limitations of optics to one that is solvable through computation. FPM can be applied to most standard digital microscopes through retrofitting with a readily available 6
7 LED matrix. Our discussion has focused on generating a high-sbp intensity image; the capability of FPM-enabled phase imaging will be detailed in future work. Our FPM prototype has not been optimized for performance speed. At present, our imaging speed is limited by the low illumination intensities provided by the LEDs at the array s edges to the sample. This issue can be addressed by either angling the LEDs inwards or using higher power LEDs. The processing speed can be significantly improved by employing a GPU, as the algorithm process is highly parallelizable. The FPM method requires an overlap of the spectrum in Fourier domains encompassed by each raw image (~60% spectrum overlap in our implementation). This redundancy is necessary as it promotes fast image convergence. It would be worth exploring the exact relationship between data redundancy and convergence in the future. In our current FPM method, we assume that samples are effectively two-dimensional. We believe there are FPM variants that can be developed to handle extended samples. We would also like to reiterate that the current FPM method is not a fluorescence technique, as fluorescent emission profiles would remain unchanged under angle-varied illuminations. However, we believe that it is possible to use patterned illumination with FPM variants to increase the SBP of a fluorescence image. Finally, we believe that higher order aberration corrections and more accurate back aperture characterizations would be worth implementing in future systems to improve the image quality of the FPM reconstruction. FPM s ability to significantly increase the SBP of a conventional microscope is highly useful for addressing the wide-fov imaging needs that dominate digital pathology and neuroscience. Furthermore, FPM s digital wavefront correction procedure lends extra flexibility to many biomedicine experiments, by largely eliminating the need to maintain a precise working distance between the sample slide and the microscope objective. However, we believe that FPM is potentially even more broadly transformative. Conventionally, the quality of an imaging system is largely defined by the physical limitations of its optical elements. For example, spatial resolution is generally characterized by the sharpness of the optical system s point-spread function. The proposed FPM method reduces the optical system to a filtering transfer function of the complex field employed in an iterative recovery process, through which the characteristics of this complex optical transfer function are rendered nominally irrelevant. As long as the low-pass pupil function is accurately characterized, this link between the actual sample profile and captured data may iteratively improve image resolution. It is this underlying robustness that allows our FPM prototype to render high-resolution images with a low-na objective which is conventionally incapable of optically providing such a narrow point-spread function and long depth-of-focus. More broadly speaking, FPM can be potentially applied to systems with severe but known aberrations to render high-quality images. Our demonstration of digital wavefront correction provides a viable strategy in this respect. We believe that the development of a general aberration correction procedure using our iterative complex field recovery strategy would be very interesting and useful. Additionally, it can potentially significantly improve X-ray and THz imaging setups that are generally limited by poor and aberrative focusing elements. Methods Experimental setup 7
8 The measured distance between the sample stage and the LED array was about 8 cm, and the measured working distance of the objective lens was about 6 mm. The LED matrix contains 32*32 surface mounted, full-color LEDs (SMD 3528), and the lateral distance between two adjacent LEDs is 4 mm. The central wavelengths of the full-color LED are 632 nm (red), 532 nm (green), and 472 nm (blue), each offering an approximately spatially coherent quasimonochromatic source with an approximate 15 nm bandwidth. We used an Atmel ATMEGA-328 microcontroller to provide the logical control for the LED matrix. To achieve maximum brightness, the matrix was driven statically rather than in normal scanning mode, eliminating the duty cycle and boosting currents through the LEDs at a maximum level. The measured light intensities were 0.7, 1.0, and 0.4 W/m 2 for the red, green, and blue colors, respectively. Measured intensities of different individual LEDs were also used to normalize each corresponding intensity image. Image acquisition and reconstruction In all figures shown, variable pixel gain was removed by flat-field correction, and hot pixels were identified and removed by interpolation. The sampling requirement of raw images is λ/(2 NA obj ), where NA obj denotes the NA of the employed objective lens (refer to Supplementary Note 3). To reconstruct a high-resolution image with a maximum synthetic NA of 0.5, we use 137 LEDs for illumination (each LED corresponds to a circle in Fig. S2c2). Due to low light intensities of the LEDs, a long exposure time is required by our prototype, limiting the speed of image acquisition. For the central 49 (7 by 7) LEDs, we acquired three images with three different exposure times (0.005 s, 0.1 s, and 0.5 s), and combined them to obtain a 14-bit high-dynamic range (HDR) image for FPM reconstruction. For LEDs outside this central area, we acquired two images with two different exposure times (0.1 s and 0.8 s) to create an 11-bit HDR image. The HDR combination process is used to suppress the saturation error caused by the overexposed pixels 43. The total acquisition time for the current prototype is about 3 minutes. With a brighter LED matrix, the maximum throughput will ultimately be determined by the sensor s data transfer rate. For example, using a commercially available 53 fps full-frame camera (VC-25MX, Vieworks), an acquisition time of several seconds can be achieved for a gigapixel image. During the reconstruction process, we divided each full FOV raw image ( pixels) into smaller image segments ( pixels each). Each set of image segments was then independently processed by the FPM recovery procedure to create a high-resolution image segment ( pixels). Finally, all high-resolution image segments were combined into one full FOV, high-resolution image (Fig. S8). The benefits of dividing the raw image into smaller segments include the following: 1) Each segment of the raw image can be processed independently, a requirement for parallel computing. 2) Memory requirements for computation are reduced. 3) The light from each LED can be accurately treated as a plane wave for each image segment of the raw image. The incident wave vector for each segment can be expressed as (k i x, k i y ) = 2π (x c x i ), (y c y i ) λ (x c x i ) 2 +(x c x i ) 2 +h 2 (y c y i ) 2 +(y c y i ) 2 +h 2, where (x c, y c ) is the central position of each small segment of the raw image, (x i, y i ) is the position of the i th LED, and h is the distance between the LED matrix and the sample. 8
9 4) Each small portion can be assigned a specific aberration-correcting pupil function, a common strategy used in wide field imaging 44 (see Supplementary Note 4 for pupil function measurement details). Using a personal computer with an Intel i7 CPU (no GPU), the processing time for each high-resolution image segment (converting raw pixels to pixels) is about 2 seconds in Matlab. The total processing time for creating a final full FOV image is about 10 minutes. For color imaging via FPM, we acquired the red, green, and blue channels using their corresponding color LEDs, processing each channel independently. Thus, the total acquisition and processing time for a color image must be multiplied by a factor of 3. References: 1 Lohmann, A. W., Dorsch, R. G., Mendlovic, D., Zalevsky, Z. & Ferreira, C. Space-bandwidth product of optical signals and systems. J. Opt. Soc. Am. A 13, (1996). 2 Denis, L., Lorenz, D., Thiébaut, E., Fournier, C. & Trede, D. Inline hologram reconstruction with sparsity constraints. Optics letters 34, (2009). 3 Xu, W., Jericho, M., Meinertzhagen, I. & Kreuzer, H. Digital in-line holography for biological applications. Proceedings of the National Academy of Sciences 98, (2001). 4 Heng, X. et al. Optofluidic microscopy a method for implementing a high resolution optical microscope on a chip. Lab on a Chip 6, (2006). 5 Zheng, G., Lee, S. A., Antebi, Y., Elowitz, M. B. & Yang, C. The epetri dish, an on-chip cell imaging platform based on subpixel perspective sweeping microscopy (SPSM). Proceedings of the National Academy of Sciences 108, (2011). 6 Zheng, G., Lee, S. A., Yang, S. & Yang, C. Sub-pixel resolving optofluidic microscope for on-chip cell imaging. Lab on a Chip 10, (2010). 7 Turpin, T., Gesell, L., Lapides, J. & Price, C. Theory of the synthetic aperture microscope. Proc. SPIE 2566, (1995). 8 Di, J. et al. High resolution digital holographic microscopy with a wide field of view based on a synthetic aperture technique and use of linear CCD scanning. Appl. Opt. 47, (2008). 9 Hillman, T. R., Gutzler, T., Alexandrov, S. A. & Sampson, D. D. High-resolution, wide-field object reconstruction with synthetic aperture Fourier holographic optical microscopy. Opt. Express 17, (2009). 10 Granero, L., Micó, V., Zalevsky, Z. & García, J. Synthetic aperture superresolved microscopy in digital lensless Fourier holography by time and angular multiplexing of the object information. Appl. Opt. 49, (2010). 11 Kim, M. et al. High-speed synthetic aperture microscopy for live cell imaging. Opt. Lett. 36, (2011). 12 Schwarz, C. J., Kuznetsova, Y. & Brueck, S. Imaging interferometric microscopy. Optics letters 28, (2003). 13 Feng, P., Wen, X. & Lu, R. Long-working-distance synthetic aperture Fresnel off-axis digital holography. Optics Express 17, (2009). 14 Mico, V., Zalevsky, Z., García-Martínez, P. & García, J. Synthetic aperture superresolution with multiple off-axis holograms. JOSA A 23, (2006). 15 Yuan, C., Zhai, H. & Liu, H. Angular multiplexing in pulsed digital holography for aperture synthesis. Optics letters 33, (2008). 9
10 16 Mico, V., Zalevsky, Z. & García, J. Synthetic aperture microscopy using off-axis illumination and polarization coding. Optics communications 276, (2007). 17 Alexandrov, S. & Sampson, D. Spatial information transmission beyond a system's diffraction limit using optical spectral encoding of the spatial frequency. Journal of Optics A: Pure and Applied Optics 10, (2008). 18 Tippie, A. E., Kumar, A. & Fienup, J. R. High-resolution synthetic-aperture digital holography with digital phase and pupil correction. Optics express 19, (2011). 19 Gutzler, T., Hillman, T. R., Alexandrov, S. A. & Sampson, D. D. Coherent aperture-synthesis, widefield, high-resolution holographic microscopy of biological tissue. Opt. Lett. 35, (2010). 20 Alexandrov, S. A., Hillman, T. R., Gutzler, T. & Sampson, D. D. Synthetic aperture Fourier holographic optical microscopy. Physical review letters 97, (2006). 21 Rodenburg, J. M. & Bates, R. H. T. The Theory of Super-Resolution Electron Microscopy Via Wigner-Distribution Deconvolution. Philosophical Transactions of the Royal Society of London. Series A: Physical and Engineering Sciences 339, (1992). 22 Faulkner, H. M. L. & Rodenburg, J. M. Movable Aperture Lensless Transmission Microscopy: A Novel Phase Retrieval Algorithm. Physical Review Letters 93, (2004). 23 Rodenburg, J. M. et al. Hard-X-Ray Lensless Imaging of Extended Objects. Physical Review Letters 98, (2007). 24 Thibault, P. et al. High-Resolution Scanning X-ray Diffraction Microscopy. Science 321, (2008). 25 Dierolf, M. et al. Ptychographic coherent diffractive imaging of weakly scattering specimens. New Journal of Physics 12, (2010). 26 Maiden, A. M., Rodenburg, J. M. & Humphry, M. J. Optical ptychography: a practical implementation with useful resolution. Optics letters 35, (2010). 27 Humphry, M., Kraus, B., Hurst, A., Maiden, A. & Rodenburg, J. Ptychographic electron microscopy using high-angle dark-field scattering for sub-nanometre resolution imaging. Nature communications 3, 730 (2012). 28 Fienup, J. R. Phase retrieval algorithms: a comparison. Appl. Opt. 21, (1982). 29 Fienup, J. R. Reconstruction of a complex-valued object from the modulus of its Fourier transform using a support constraint. JOSA A 4, (1987). 30 Fienup, J. R. Reconstruction of an object from the modulus of its Fourier transform. Optics letters 3, (1978). 31 Fienup, J. R. Lensless coherent imaging by phase retrieval with an illumination pattern constraint. Opt. Express 14, (2006). 32 Levoy, M., Ng, R., Adams, A., Footer, M. & Horowitz, M. Light field microscopy. ACM Transactions on Graphics 25, (2006). 33 Levoy, M., Zhang, Z. & McDowall, I. Recording and controlling the 4D light field in a microscope using microlens arrays. Journal of Microscopy 235, (2009). 34 Arimoto, H. & Javidi, B. Integral three-dimensional imaging with digital reconstruction. Optics letters 26, (2001). 35 Hong, S.-H., Jang, J.-S. & Javidi, B. Three-dimensional volumetric object reconstruction using computational integral imaging. Optics Express 12, (2004). 36 Gustafsson, M. G. Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy. Journal of microscopy 198, (2000). 37 Tyson, R. Principles of adaptive optics. (CRC Press, 2010). 38 Brady, D. et al. Multiscale gigapixel photography. Nature 486, (2012). 10
11 39 Guizar-Sicairos, M. & Fienup, J. R. Phase retrieval with transverse translation diversity: a nonlinear optimization approach. Optics express 16, (2008). 40 Zheng, G., Kolner, C. & Yang, C. Microscopy refocusing and dark-field imaging by using a simple LED array. Opt. Lett. 36, (2011). 41 Colomb, T. et al. Automatic procedure for aberration compensation in digital holographic microscopy and applications to specimen shape compensation. Applied optics 45, (2006). 42 Liang, R. Optical design for biomedical imaging. (SPIE Press, 2010). 43 Reinhard, E. et al. High dynamic range imaging: acquisition, display, and image-based lighting. (Morgan Kaufmann, 2010). 44 Gunturk, B. K. & Li, X. Image Restoration: Fundamentals and Advances. Vol. 7 (CRC Press, 2012). Acknowledgements We are grateful for the constructive discussions and generous help from Mr. Xiaoze Ou, Ms. Ying Min Wang and Mr. Chris Kolner from Caltech. We acknowledge funding from NIH 1DP2OD Author contributions: G.Z. conceived the initial idea, designed and implemented the experiment. G.Z., R.H., and C.Y. developed and refined the concept, and wrote the paper. 11
12 Figure 1: Fig. 1 FPM s iterative recovery procedure. Steps 1 5 illustrate FPM s algorithm, following principles from phase retrieval. N low-resolution intensity images captured under variable illumination are used to recover one high-resolution intensity image and one high-resolution phase map. 12
13 Figure 2: Fig. 2 FPM prototype setup. (a) Diagram of the setup. A programmable LED matrix is placed beneath the sample. The i th LED illuminates the sample with wave-vector k x i. (b) The LED matrix and microscope used in experiment, where (Inset) each LED can provide red, green, and blue narrow-band illumination. (c1) A full-fov raw image of a USAF resolution target. (c2) A magnified view of the raw image, exhibiting a pixel size of 2.75 µm. (d) Our FPM reconstruction of the same region, where we achieve a reconstructed pixel size of µm (refer to the discussion of FPM s sampling requirement in Supplementary Note 3). In this reconstruction, the corresponding maximum synthetic NA of the reconstructed image is 0.5, set by the maximum angle between the optical axis and an LED. The entire recovery process is demonstrated in Supplementary Video 1. 13
14 Figure 3: Fig. 3 Extending depth-of-focus with digital wavefront correction. (a) The principle of FPM s digital wavefront correction technique. A digital pupil function is introduced in steps 2 and 5 to model the connection between the actual sample profile and the captured intensity data, which may exhibit aberrations caused by defocus. (b) Diagram of FPM s iterative recovery algorithm with the addition of digital wavefront correction. (c) One raw low-resolution image of the USAF target placed at z 0 = -150 µm. High-resolution FPM reconstructions without (d) and with (e) steps 2 and 5 added to the iterative recovery procedure. 14
15 Figure 4: Fig. 4 Gigapixel color imaging via FPM. (a) A wide-fov color image of a pathology slide, with a SBP of approximately 0.9 gigapixels. (b, c1, d, and e): Vignette high-resolution views of the image in (a). Images taken by a conventional microscope with a 20 (c2) and a 2 (c3) objective lens, for comparison. A color image sensor (DFK 61BUC02, Image Source Inc.) is used for capturing (c2 and c3). 15
Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998)
Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998) Geoff Andersen and R. J. Knize Laser and Optics Research Center
Color holographic 3D display unit with aperture field division
Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University
Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.
Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining
WHITE PAPER. Are More Pixels Better? www.basler-ipcam.com. Resolution Does it Really Matter?
WHITE PAPER www.basler-ipcam.com Are More Pixels Better? The most frequently asked question when buying a new digital security camera is, What resolution does the camera provide? The resolution is indeed
Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.
Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
Advances in scmos Camera Technology Benefit Bio Research
Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical
Near-field scanning optical microscopy (SNOM)
Adviser: dr. Maja Remškar Institut Jožef Stefan January 2010 1 2 3 4 5 6 Fluorescence Raman and surface enhanced Raman 7 Conventional optical microscopy-limited resolution Two broad classes of techniques
Optical laser beam scanner lens relay system
1. Introduction Optical laser beam scanner lens relay system Laser beam scanning is used most often by far in confocal microscopes. There are many ways by which a laser beam can be scanned across the back
Diffraction of a Circular Aperture
Diffraction of a Circular Aperture Diffraction can be understood by considering the wave nature of light. Huygen's principle, illustrated in the image below, states that each point on a propagating wavefront
Physics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
Fig.1. The DAWN spacecraft
Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics
RAY TRACING UNIFIED FIELD TRACING
RAY TRACING Start to investigate the performance of your optical system using 3D ray distributions, dot diagrams of ray positions and directions, and optical path length. GEOMETRIC FIELD TRACING Switch
Realization of a UV fisheye hyperspectral camera
Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical
Endoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
Spherical Beam Volume Holograms Recorded in Reflection Geometry for Diffuse Source Spectroscopy
Spherical Beam Volume Holograms Recorded in Reflection Geometry for Diffuse Source Spectroscopy Sundeep Jolly A Proposal Presented to the Academic Faculty in Partial Fulfillment of the Requirements for
Understanding astigmatism Spring 2003
MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest
Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD
2008 AGI-Information Management Consultants May be used for personal purporses only or by libraries associated to dandelon.com network. Optical Metrology Third Edition Kjell J. Gasvik Spectra Vision AS,
Rodenstock Photo Optics
Rogonar Rogonar-S Rodagon Apo-Rodagon N Rodagon-WA Apo-Rodagon-D Accessories: Modular-Focus Lenses for Enlarging, CCD Photos and Video To reproduce analog photographs as pictures on paper requires two
Arrayoptik für ultraflache Datensichtbrillen
Arrayoptik für ultraflache Datensichtbrillen Peter Schreiber, Fraunhofer Institut für Angewandte Optik und Feinmechanik IOF, Jena 1. Introduction 2. Conventional near-to-eye (N2E) projection optics - first
Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications
Screw thread image generated by WLI Steep PSS angles WLI color imaging Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications 3D optical microscopy is a mainstay metrology
Lenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012
Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 You are sitting at your microscope working at high magnification trying to sort out the three-dimensional compartmentalization
Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images
FORMATEX 2007 A. Méndez-Vilas and J. Díaz (Eds.) Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images Jörg Piper Clinic Meduna, Department for Internal
AxioCam HR The Camera that Challenges your Microscope
Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,
PHYS 39a Lab 3: Microscope Optics
PHYS 39a Lab 3: Microscope Optics Trevor Kafka December 15, 2014 Abstract In this lab task, we sought to use critical illumination and Köhler illumination techniques to view the image of a 1000 lines-per-inch
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus
Grazing incidence wavefront sensing and verification of X-ray optics performance
Grazing incidence wavefront sensing and verification of X-ray optics performance Timo T. Saha, Scott Rohrbach, and William W. Zhang, NASA Goddard Space Flight Center, Greenbelt, Md 20771 Evaluation of
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Preface Light Microscopy X-ray Diffraction Methods
Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective
Resolution for Color photography
Resolution for Color photography Paul M. Hubel a and Markus Bautsch b a Foveon, Inc., 282 San Tomas Expressway, Santa Clara, CA, USA 955; b Stiftung Warentest, Luetzowplatz -3, D-785 Berlin-Tiergarten,
How To Fix Out Of Focus And Blur Images With A Dynamic Template Matching Algorithm
IJSTE - International Journal of Science Technology & Engineering Volume 1 Issue 10 April 2015 ISSN (online): 2349-784X Image Estimation Algorithm for Out of Focus and Blur Images to Retrieve the Barcode
We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement
MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors Rudolph J. Guttosch Foveon, Inc. Santa Clara, CA Abstract The reproduction
AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy
Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
High-resolution Imaging System for Omnidirectional Illuminant Estimation
High-resolution Imaging System for Omnidirectional Illuminant Estimation Shoji Tominaga*, Tsuyoshi Fukuda**, and Akira Kimachi** *Graduate School of Advanced Integration Science, Chiba University, Chiba
Inspection and Illumination Systems for Visual Quality Assurance. Large selection Attractive prices Tailored solutions. optometron.
Inspection and Illumination Systems for Visual Quality Assurance Large selection Attractive prices Tailored solutions optometron.de en Optometron fulfils a key role Visual quality assurance is essential
Lecture 12: Cameras and Geometry. CAP 5415 Fall 2010
Lecture 12: Cameras and Geometry CAP 5415 Fall 2010 The midterm What does the response of a derivative filter tell me about whether there is an edge or not? Things aren't working Did you look at the filters?
Imaging techniques with refractive beam shaping optics
Imaging techniques with refractive beam shaping optics Alexander Laskin, Vadim Laskin AdlOptica GmbH, Rudower Chaussee 29, 12489 Berlin, Germany ABSTRACT Applying of the refractive beam shapers in real
Performance Verification of Super-Resolution Image Reconstruction
Performance Verification of Super-Resolution Image Reconstruction Masaki Sugie Department of Information Science, Kogakuin University Tokyo, Japan Email: [email protected] Seiichi Gohshi Department
Light-Field Displays: Technology and Representation of 3D visual information. Péter Tamás Kovács Holografika
Light-Field Displays: Technology and Representation of 3D visual information Péter Tamás Kovács Holografika JPEG PLENO Workshop Warsaw, Poland 23 June, 2015 Holografika Hungarian company, active in the
White Paper. "See" what is important
Bear this in mind when selecting a book scanner "See" what is important Books, magazines and historical documents come in hugely different colors, shapes and sizes; for libraries, archives and museums,
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
Introduction to Optics
Second Edition Introduction to Optics FRANK L. PEDROTTI, S.J. Marquette University Milwaukee, Wisconsin Vatican Radio, Rome LENO S. PEDROTTI Center for Occupational Research and Development Waco, Texas
Image Compression through DCT and Huffman Coding Technique
International Journal of Current Engineering and Technology E-ISSN 2277 4106, P-ISSN 2347 5161 2015 INPRESSCO, All Rights Reserved Available at http://inpressco.com/category/ijcet Research Article Rahul
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
WAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
Synthetic Aperture Radar: Principles and Applications of AI in Automatic Target Recognition
Synthetic Aperture Radar: Principles and Applications of AI in Automatic Target Recognition Paulo Marques 1 Instituto Superior de Engenharia de Lisboa / Instituto de Telecomunicações R. Conselheiro Emídio
Introduction to acoustic imaging
Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3
Measuring large areas by white light interferometry at the nanopositioning and nanomeasuring machine (NPMM)
Image Processing, Image Analysis and Computer Vision Measuring large areas by white light interferometry at the nanopositioning and nanomeasuring machine (NPMM) Authors: Daniel Kapusi 1 Torsten Machleidt
Optical Design using Fresnel Lenses
Optical Design using Fresnel Lenses Basic principles and some practical examples Arthur Davis and Frank Kühnlenz Reflexite Optical Solutions Business Abstract The fresnel lens can be used in a wide variety
Acoustic GHz-Microscopy: Potential, Challenges and Applications
Acoustic GHz-Microscopy: Potential, Challenges and Applications A Joint Development of PVA TePLa Analytical Systems GmbH and Fraunhofer IWM-Halle Dr. Sebastian Brand (Ph.D.) Fraunhofer CAM Fraunhofer Institute
3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY
3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY Prepared by Duanjie Li, PhD & Andrea Novitsky 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.
MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize
Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision!
Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision! 1 In this presentation We are going to go over lens basics Explain figures of merit of lenses Show how
Understanding Line Scan Camera Applications
Understanding Line Scan Camera Applications Discover the benefits of line scan cameras, including perfect, high resolution images, and the ability to image large objects. A line scan camera has a single
http://dx.doi.org/10.1117/12.906346
Stephanie Fullerton ; Keith Bennett ; Eiji Toda and Teruo Takahashi "Camera simulation engine enables efficient system optimization for super-resolution imaging", Proc. SPIE 8228, Single Molecule Spectroscopy
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
Resolution Enhancement of Photogrammetric Digital Images
DICTA2002: Digital Image Computing Techniques and Applications, 21--22 January 2002, Melbourne, Australia 1 Resolution Enhancement of Photogrammetric Digital Images John G. FRYER and Gabriele SCARMANA
Biomedical & X-ray Physics Kjell Carlsson. Light Microscopy. Compendium compiled for course SK2500, Physics of Biomedical Microscopy.
Biomedical & X-ray Physics Kjell Carlsson Light Microscopy Compendium compiled for course SK2500, Physics of Biomedical Microscopy by Kjell Carlsson Applied Physics Dept., KTH, Stockholm, 2007 No part
High Quality Image Deblurring Panchromatic Pixels
High Quality Image Deblurring Panchromatic Pixels ACM Transaction on Graphics vol. 31, No. 5, 2012 Sen Wang, Tingbo Hou, John Border, Hong Qin, and Rodney Miller Presented by Bong-Seok Choi School of Electrical
Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis
Cam CANIMPEX CPX-SOLUTIONS 3D Digital Microscope Camera FAST PRECISE AFFORDABLE 3D CAMERA FOR MICROSCOPY Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis
LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes
LBS-300 Beam Sampler for C-mount Cameras P/N SP90183, SP90184, SP90185 and SP90186 YAG Focal Spot Analysis Adapter P/N SP90187, SP90188, SP90189, SP90190, SP90191 User Notes Ophir-Spiricon Inc. 60 West
Laser expander design of highly efficient Blu-ray disc pickup head
Laser expander design of highly efficient Blu-ray disc pickup head Wen-Shing Sun, 1,* Kun-Di Liu, 1 Jui-Wen Pan, 1 Chuen-Lin Tien, 2 and Min-Sheng Hsieh 1 1 Department of Optics and Photonics, National
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,
Fiber optic communication
Fiber optic communication Fiber optic communication Outline Introduction Properties of single- and multi-mode fiber Optical fiber manufacture Optical network concepts Robert R. McLeod, University of Colorado
Optical Design Tools for Backlight Displays
Optical Design Tools for Backlight Displays Introduction Backlights are used for compact, portable, electronic devices with flat panel Liquid Crystal Displays (LCDs) that require illumination from behind.
Experimental and modeling studies of imaging with curvilinear electronic eye cameras
Experimental and modeling studies of imaging with curvilinear electronic eye cameras Viktor Malyarchuk, 1 Inhwa Jung, 1 John A. Rogers, 1,* Gunchul Shin, 2 and Jeong Sook Ha 2 1 Department of Materials
Digital Holography using a Laser Pointer and Consumer Digital Camera Report date: June 22 nd, 2004
Digital Holography using a Laser Pointer and Consumer Digital Camera Report date: June 22 nd, 2004 by David Garber, M.E. undergraduate, Johns Hopkins University, [email protected] Available online at http://pegasus.me.jhu.edu/~lefd/shc/lpholo/lphindex.htm
Application Report: Running µshape TM on a VF-20 Interferometer
: Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20
Example #1: Controller for Frequency Modulated Spectroscopy
Progress Report Examples The following examples are drawn from past student reports, and illustrate how the general guidelines can be applied to a variety of design projects. The technical details have
Study Guide for Exam on Light
Name: Class: Date: Study Guide for Exam on Light Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Which portion of the electromagnetic spectrum is used
Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications
Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications Saulius Marcinkevičius Optics, ICT, KTH 1 Outline Optical near field. Principle of scanning near field optical microscope
High-resolution imaging of the human retina with a Fourier deconvolution technique
D. Catlin and C. Dainty Vol. 19, No. 8/August 2002/J. Opt. Soc. Am. A 1515 High-resolution imaging of the human retina with a Fourier deconvolution technique David Catlin and Christopher Dainty The Blackett
INFITEC - A NEW STEREOSCOPIC VISUALISATION TOOL BY WAVELENGTH MULTIPLEX IMAGING
INFITEC - A NEW STEREOSCOPIC VISUALISATION TOOL BY WAVELENGTH MULTIPLEX IMAGING Helmut Jorke, Markus Fritz INFITEC GmbH, Lise-Meitner-Straße 9, 89081 Ulm [email protected] Phone +49 731 550299 56 Fax _
Encoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
CREOL, College of Optics & Photonics, University of Central Florida
OSE6650 - Optical Properties of Nanostructured Materials Optical Properties of Nanostructured Materials Fall 2013 Class 3 slide 1 Challenge: excite and detect the near field Thus far: Nanostructured materials
Modern Classical Optics
Modern Classical Optics GEOFFREY BROOKER Department of Physics University of Oxford OXPORD UNIVERSITY PRESS Contents 1 Electromagnetism and basic optics 1 1.1 Introduction 1 1.2 The Maxwell equations 1
Optical Communications
Optical Communications Telecommunication Engineering School of Engineering University of Rome La Sapienza Rome, Italy 2005-2006 Lecture #2, May 2 2006 The Optical Communication System BLOCK DIAGRAM OF
5 Factors Affecting the Signal-to-Noise Ratio
5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed
Video Camera Image Quality in Physical Electronic Security Systems
Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems In the second decade of the 21st century, annual revenue for the global
Today. next two weeks
Today Temporal and spatial coherence Spatially incoherent imaging The incoherent PSF The Optical Transfer Function (OTF) and Modulation Transfer Function (MTF) MTF and contrast comparison of spatially
Bio 321 Lightmicroscopy Electronmicrosopy Image Processing
Bio 321 Lightmicroscopy Electronmicrosopy Image Processing Urs Ziegler Center for Microscopy and Image Analysis Light microscopy (Confocal Laser Scanning Microscopy) Light microscopy (Confocal Laser Scanning
Theremino System Theremino Spectrometer Technology
Theremino System Theremino Spectrometer Technology theremino System - Theremino Spectrometer Technology - August 15, 2014 - Page 1 Operation principles By placing a digital camera with a diffraction grating
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property
Holographic data storage at 2+ Tbit/in 2
Holographic data storage at + Tbit/in Mark R. Ayres *, Ken Anderson, Fred Askham, Brad Sissom, Adam C. Urness Akonia Holographics, LLC, Miller Dr., Longmont, CO, USA, 85 ABSTRACT The onslaught of big data
Theory and Methods of Lightfield Photography SIGGRAPH 2009
Theory and Methods of Lightfield Photography SIGGRAPH 2009 Todor Georgiev Adobe Systems [email protected] Andrew Lumsdaine Indiana University [email protected] 1 Web Page http://www.tgeorgiev.net/asia2009/
High Quality Image Magnification using Cross-Scale Self-Similarity
High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg
The Olympus stereology system. The Computer Assisted Stereological Toolbox
The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,
Rodenstock Photo Optics
Apo-Sironar-S Apo-Macro-Sironar Apo-Grandagon Grandagon-N Accessories: Center filters Accessories: Focus-Mount Lenses for Analog Professional Photography Even in the age of digital photography, the professional
Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
Integrated sensors for robotic laser welding
Proceedings of the Third International WLT-Conference on Lasers in Manufacturing 2005,Munich, June 2005 Integrated sensors for robotic laser welding D. Iakovou *, R.G.K.M Aarts, J. Meijer University of
HSI BASED COLOUR IMAGE EQUALIZATION USING ITERATIVE n th ROOT AND n th POWER
HSI BASED COLOUR IMAGE EQUALIZATION USING ITERATIVE n th ROOT AND n th POWER Gholamreza Anbarjafari icv Group, IMS Lab, Institute of Technology, University of Tartu, Tartu 50411, Estonia [email protected]
SUPER RESOLUTION FROM MULTIPLE LOW RESOLUTION IMAGES
SUPER RESOLUTION FROM MULTIPLE LOW RESOLUTION IMAGES ABSTRACT Florin Manaila 1 Costin-Anton Boiangiu 2 Ion Bucur 3 Although the technology of optical instruments is constantly advancing, the capture of
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The