Molecular Imaging: from cell to man

Size: px
Start display at page:

Download "Molecular Imaging: from cell to man"

Transcription

1 Molecular Imaging: from cell to man Torino, 24 November 2011 Advances in PET/SPECT technology for pre-clinical molecular imaging applications Alberto Del Guerra Dipartimento di Fisica "E.Fermi' Università di Pisa and INFN, Sezione di Pisa

2 CONTENTS MicroPET Technology MicroPET Tomographs MicroSPECT Technology MicroSPECT Tomographs MicroCT imaging Multimodality imaging

3 PET spatial resolution / 1 FWHM 1.2 d 2 2 b D 2 r 2 p 2 Crystal size Coding Non collinearity Positron Range Parallax error 1.2 : Degradation factor due to reconstruction d : Crystal pitch b : Coding error (range: 0-2 mm) D : Detector separation (i.e. gantry diameter) r : effective source size (including positron range) p : Parallax error * Derenzo & Moses, "Critical instrumentation issues for resolution <2mm, high sensitivity brain PET", in Quantification of Brain Function, Tracer Kinetics & Image Analysis in Brain PET, ed. Uemura et al, Elsevier, 1993, pp

4 From men to monkeys, to rats.. to mice Human PET human micropet *Images courtesy of Simon Cherry, UCLA mouse rat rat mouse infant monkey 4

5 Spatial resolution requirements 5

6 From the block detector to PSPMT s Block detector 1 st generation PSPMT Hamamatsu PS-PMT R mm Ø active area 16 x + 16 y anodes Small crystals can be used (down to d = 1mm) Used in the YAP-(S)PET (Univ of Ferrara Italy,1993) Large b Limitations on minimum d Flood field irradiation (511 kev) of a matrix of scintillator YAP:Ce, read by a Hamamatsu R2486 (resistive readout) 6

7 APD APD array Scintillator matrix (BGO/LSO) Coding problem: Individual coupling with APD s or light sharing with PSPMT s 2 nd generation PSPMT Hamamatsu PS-PMTR8520-C12 Active area 22 mm 22mm 6 x + 6 y anodes + High spatial resolution (b=0) + No Pile-up + No scattering in the crystals - Expensive - Many channels - Difficult tuning The detector module is composed by a matrix of 8 4 LSO crystals readout by a Hamamatsu S8550 (Pichler B., IEEE TNS 45 (1998) ) + Few channels to readout (resistive chain) + High gain and stability - Non negligible coding error - Pile-up increases with area - PS-PMT Hamamatsu R7600-C8 Matrix 8 8 square fibres 8 8 LSO matrix

8 Today: advanced light sharing photodetectors PET History PET Physics and technology light sharing technique: Hybrid position sensitive APD light sharing technique: Multi Anode flat panel PMT PSDs in PET: PMT PSDs in PET: solid state Advanced PET detectors: DOI and TOF Picture of a HPS-APD with four output connectors MA-PMT (8 8 ch s) Hamamatsu H8500 Active area 49 mm x 49 mm SiPMs for PET Conclusions Flood field image ( 241 Am, 60 kev ) obtained with a 4x4 and 8x8 CsI(Tl) scintillating matrices Flood field image (511 kev ) obtained with a 20x20 YAP:Ce scintillating matrix (resistive readout) 8

9 Spatial resolution of of commercial scanners Commercial scanners do not show large differences in spatial resolution YAP-(S)PET

10 Sensitivity requirements Requirements Imaging of low activity sources low uptake processes such as in gene research Possibility to study fast metabolic processes with characteristic time comparable with the scanning time Limitations Brain receptor saturation usually a maximum of 100 Ci can be injected to a mouse Limitation on the volume a maximum of 300 l can be injected to a mouse Solutions Utilization of radionuclides with a very high specific activity such as PET short half-life radioisotopes: 15 O (122 s), 13 N (10 min), 11 C (20 min), 18 F (110min) High geometry efficiency (large solid angle covered by detectors) High detection efficiency (e.g. for crystals: high/medium Z, high density) 10

11 Absolute sensitivity of commercial scanners Larger variations can be observed in the sensitivity figure of merit. YAP-(S)PET

12 SIEMENS micropet Focus 220 / Inveon PET 18 F-Paclitaxel biodistribution in rat MicroPET Focus 220 is a PET only scanner using the fiber technique Rat heart 18F-FDG The SIEMENS Inveon is dockable with a CT scanner 12

13 YAP-(S)PET II small animal scanner Scanner configuration Configuration: Four rotating heads Scintillator: YAlO 3 :Ce (YAP:Ce) Crystal size: 27 x 27 (1.5 x 1.5 x 20 mm 3 each) Photodetector: Position Sensitive PMT Readout method: Resistive chain (4 channels) FoV size: 40.5 mm axial 40.5 mm Ø Collimators (SPECT): Lead (parallel holes) Head-to-head distance: cm Scanner installed at the Institute of clinical Physiology (IFC-CNR) within the framework of the Center of Excellence AmbiSEN of the University of Pisa, Italy

14 Heart and bone metabolism in mouse with 18 F-FDG and 18 F - Mouse with 18 F-FDG Mouse with 18 F - (post-mortem) Horizontal slices: Gray and color scale injection of 11 MBq of 18 F -, 120 min. uptake time Step-and-shoot acquisition 128 views/180 (Acquisition time 60 min) Transaxial sections Horizontal section Total body (MIP) 120 min. uptake time (Acquisition time 100 min) Voxel size 375 m 375 m 750 m 3D ML-EM reconstruction 14

15 Small animal SPECT instrumentation development The use of pinhole-collimators allows a large magnification obtaining a high spatial resolution on medium-small field of view. The implementation of multi-pinholes increases the sensitivity Tipically based on large area NaI gamma camera similar to clinical one. ================ Other solutions based on solid state detectors are available. In this case the detector has a high intrinsic spatial resolution (smaller magnification allowed) and are characterized by a high energy resolution (multi-isotope imaging allowed)

16 Small animal SPECT collimator geometry Parallel hole pinhole D h L d System spatial resolution System spatial resolution D d L R 2 int 2 D 2 1 L e d L Rint d 2 System sensitivity System sensitivity D D D e 3 L D h 4d sin 16

17 Geom. Resol. FOV size Small animal SPECT Effect of collimator-to-target distance Sensitivity High sensitivity of pinholes only at small d (small FOV) d d Parallel hole Pinhole d 17

18 MILabs U-SPECT II Bone scan: rat Bone scan: mouse Detector active area Crystal Number of detectors Number and size of the pinholes 510 x 381 mm NaI(Tl), 9.5 mm thick Continuous 3 stationary 75 / mm FOV 28 mm x 140 mm 60 mm x 240 mm Spatial resolution mm FWHM Based on a standard three heads NaI gamma camera (no rotation) equipped with multi-pinhole collimators. Sensitivity >1500 cps/mbq 18

19 Bioscan NanoSPECT Features Detector active area 230 x 215 mm Crystal NaI(Tl), 9.5 mm thick continuous Number of detectors 4 (1,2 or 4) Also available as HiSPECT: transforms a clinical SPECT camera into an Animal Imager Number and size of the pinholes 36 / 1.0 mm FOV 26 mm x 20 mm Spatial resolution Sensitivity 0.8 mm FWHM 1640 cps/mbq Multi-pinhole with elical scanning 19

20 Siemens Inveon SPECT Detector active area 150 x 150 mm Crystal Number of detectors 2 or 4 NaI(Tl), 10 mm thick Pixilated (2.2 mm pitch) Mouse bone scan Number and size of the pinholes FOV Spatial resolution Sensitivity 1 or more / mm variable variable variable 20

21 GE explore speczt Full-ring solid-state detector small animal SPECT Utilizes a cadmium zinc telluride (CZT) detector High-energy resolution to enable dual or triple radionuclide imaging. Stationary, full-ring, 10 detector design Interchangeable, rotating cylindrical collimators - Multi-slit: 80mm axial FOV, full 360-degree coverage - Multi-pinhole: high resolution, full 360-degree coverage Detector active area Crystal Number of detectors Number and size of the pinholes FOV 124 x 124 mm ev-czt pixilated 10 full ring stationary Multi pinhole Multi slit 80 mm axial with multi slit 21

22 CT Dipartimento di Fisica e IFC-CNR 22

23 PET/CT image fusion PET Co-registration CT Advantages Anatomical repere: CT provides high resolution morphological information map The map is scaled at 511 kev and blurred at the YAP-(S)PET spatial resolution Better quantification: Attenuation correction of PET data CT may provide the shape and size of the target for recovery coeff. correction Additional information w.r.t. PET Using CT as a stand alone modality for stem cell imaging 23

24 Imaging PET/CT nel ratto

25 Whole body PET/CT (topo)

26 Multimodality systems Bioscan NanoSPECT/CT The BIOSCAN NanoSPECT is available also in combination with a CT module 26

27 Multimodality systems SIEMENS Inveon Available as integrated PET/SPECT/CT or dockable PET + SPECT/CT (only 2 SPECT heads available when in combination with CT) 27

28 GE Triumph (Gammamedica) X-SPECT : CZT based SPECT Sub-System LabPET : APD based PET Sub-System X-O : Fast, Low Dose CT Sub-System Triple isotope SPECT + CT 18F-FDG PET + CT CT 28

29 Rationale for PET/MR PET High sensitivity (10-11 /10-12 mol/l) Good spatial resolution (4mm for clinical system) Functional info and quantitation MRI Good sensitivity (10-3 /10-5 mol/l) Excellent spatial resolution (1mm isotropic for clinical system) Better soft tissue contrast with respect to CT Anatomical info (but also functional) No radiation dose COMBINED PET/MR 29

30 Technical Challenges in PET/MRI Interference on PET (photomultiplier and electronics) Static magnetic field Electromagnetic interference from RF and gradients Interference on MR (homogeneity and gradients) Electromagnetic radiation from PET electronics Maintaining magnetic field homogeneity Eddy currents Susceptibility artifacts General Challenges Space Environmental factors (temperature, vibration ) Cost 30

31 Effect of magnetic field on positron range A) B) 0 Tesla Tesla (X-Y plane) 86 Y (E max = 3.15 MeV) A). Influence of the magnetic field on positron range, for 86Y (Emax=3.15 MeV) in water,illustrated by Monte Carlo simulations obtained at 0 Tesla (top) and 10 Tesla (bottom) field. The 3-D tracks are projected onto a plane perpendicular to the direction of the magnetic field Distance (mm) B). Simulated positron range reduction for I-124 (Emax=2.14 MeV) in a 0 Tesla (top) and 9 Tesla (bottom) magnetic field. 31

32 PET/MRI solutions Artistic cross-view of various potential designs of combined PET-MRI systems a) tandem: The two scanners are mounted together back-to-back allowing sequential (like PET/CT) rather than simultaneous acquisition, b) insert: The PET scanner is inserted between the RF-coil and gradient set of the MR system, c) full integration: the two systems are fully integrated within the same gantry 32

33 Philips tandem PET/MRI PET MRI PET MRI PET-MRI Photograph of the Philips whole-body Ingenuity TF PETMR system design installed at Geneva University Hospital. A turntable patient handling system facilitates patient motion between the Achieva X-series 3T MRI system on the right and the time-of-flight PET system on the left. 33 Whole-body MRI, PET and fused PET-MRI images are also shown.

34 The advent of Solid State Photodetectors CdZnTe Used succesfully in SPECT by GE and used in SPECT/MR prototype APD = Avalanche Photodiode Criticity: High performance Amplifier Variation with T particularly relevant SiPM (Silicon Photomultiplier) Geiger-Muller APD DSiPM = Digital SiPM 34

35 Prototype combined solution Combined small animal PET/MRI developed by the University of Tuebingen (Germany). The PET insert is fully integrated into a 7 Tesla MRI system (ClinScan, Bruker): (a) Drawing of PET/MRI combination, showing the PET insert placed inside the MRI scanner, matching the centers of both fields of view. (b) Photograph of the MRI compatible PET insert, consisting of ten detector modules. (c) Single PET detector module showing the LSO scintillator block, APD-array, and preamplifier built into a MRI compatible copper shielding. 35

36 A Whole body PET/MRI solution from SIEMENS B (A) Showing the basic components of the system where the PET detector ring is placed between the RF coil and the RF body coil. (B) Configuration of the detector block consisting of 8 8 LSO crystals readout by a matrix of 3 3 APDs. (Courtesy of Siemens Medical Solutions). 36

37 Silicon Photomultiplier (SiPM) as a the most promising solid state photodetector SiPM are p-n diodes operating in Geiger mode, which means that the bias voltage is above the diode breakdown voltage. In this way output is independent from input: the surface is divided into -cells (~1000/mm 2 ) The SiPM has all of the characteristics: speed, QE, granularity, flexibility, robustness for a successful implementation in small animal instrumentation. Signal N cell of hit cells LSO slab Light guide SiPM array + High gain + Low noise + Good proportionality if N photon < N cell Triple layer detector block SiPM are insensitive to magnetic fields compatible with MRI 37

38 PET/MRI 38

39 Thank you!

40

41 MRI based Attenuation Correction Whole-body MRI MR- map CT- map MRAC PET CTAC PET From left to right: - whole-body T1 weighted gradient echo MRI sequence co-registered to CT image of the same patient, - derived three-segment (soft tissue, lung and air) attenuation map (MRAC), - CT-based attenuation map (CTAC), - attenuation corrected PET images using MRAC -- attenuation corrected PET images using CTAC. 41

42 MRI based Attenuation Correction MR- map Modified MR- map CT- map LEFT: Attenuation correction maps derived from segmentation of T1 weighted MRI followed by assignment of known linear attenuation coefficients to the lung and soft tissue and addition of the scanner table template MIDDLE: same image shown on the left after non-rigid alignment to the CT attenuation map following removal of the PET-MR bed and addition of the CT scanner bed CT RIGHT: the CT-based attenuation map of the same patient. 42

43 Combined and simultaneous PET/SPECT with the YAP-(S)PET II One PET pair One SPECT pair Low energy shielding foil Thanks to the planar detectors and the YAP:Ce scintillator the YAP-(S)PET can perform SPECT imaging too on the same gantry by adding parallel hole collimators Removable collimators Cross contamination is reduced by: shielding the low energy single photons with a thin lead slab in front of the PET detectors With a dual window subtraction technique for selecting 99m Tc gamma s only in the SPECT data delguerra@df.unipi.it

44 Performance: Transaxial resolution Derenzo Phantom (SPECT) with 99m Tc 1.2 mm FBP (ramp filter) reconstruction was used on a mm 3 voxel space. Sinograms were build using kev energy window (37 cps/mbq). 1.5 mm thick slices delguerra@df.unipi.it

45 Dual tracer SPECT study (Myoview-Annexin) on rat heart 99m Tc-Annexin V (Apoptosis) 99m Tc-Myoview (Perfusion) Fusion 45 delguerra@df.unipi.it

46 Simultaneous PET/SPECT imaging with YAP-(S)PET SPECT ( 99m Tc) PET ( 18 F) 5:1 SPECT PET 10:1 30:1 Images of a section of an image quality phantom: the phantom is filled with 99m Tc while the two holes are filled with 18 F. ( 99m Tc/ 18 F activity ratio 30:1). 50:1 Images of two small cylinders simultaneously present in the FOV with different SPECT/PET isotopes activity concentration ratio Transaxial and coronal sections of a simultaneous PET/SPECT acquisition of two capillaries. The left one was filled with 18.5 MBq of 99m Tc, while the right with 370 kbq of 18 F ( 99m Tc/ 18 F activity ratio 50:1). delguerra@df.unipi.it

47 Molecular imaging technique for small animals A. PET Imaging on rats using 18 F-FDG showing glucose metabolism B. CT Imaging of a mouse abdomen after the injection of a contrast agent C. SPECT Imaging of a mouse abdomen after the injection of 99m Tc methylene diphosphonate showing the accumulation of the tracer in bones. D. Optical Imaging of a mouse showing the fluorescence of GFP from liver, abdomen, spinal chord and brain due to the presence of cancer cells. E. MRI image T2-weighted of a mouse brain. F. Bioluminescence optical imaging of a mouse superimposed to the picture of the animal. MULTIMODALITY

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Positron Emission Tomography: Principles, Technology, and Recent Developments

Positron Emission Tomography: Principles, Technology, and Recent Developments Nuclear Physics A 752 (2005) 679c 687c Positron Emission Tomography: Principles, Technology, and Recent Developments Sibylle I. Ziegler Nuklearmedizinische Klinik, Klinikum rechts der Isar der Technischen

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity

More information

Experimental study of beam hardening artefacts in photon counting breast computed tomography

Experimental study of beam hardening artefacts in photon counting breast computed tomography Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE

Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using Parham Geramifar 1,3,4, Mohammad Reza Ay 2,3,4, Mojtaba Shamsaie Zafarghandi 1, George Loudos 5, Arman

More information

Truly digital PET imaging

Truly digital PET imaging Advanced Molecular Imaging Vereos PET/CT Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital

More information

PoS(PhotoDet 2012)068

PoS(PhotoDet 2012)068 Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento

More information

Family IMAGINGFORSCIENCE. the next generation four modality preclinical imaging platform SPECT/MRI, PET/MRI, SPECT/CT, PET/CT, SPECT/CT/PET

Family IMAGINGFORSCIENCE. the next generation four modality preclinical imaging platform SPECT/MRI, PET/MRI, SPECT/CT, PET/CT, SPECT/CT/PET Family the next generation four modality preclinical imaging platform SPECT/MRI, PET/MRI, SPECT/CT, PET/CT, SPECT/CT/PET IMAGINGFORSCIENCE nanoscan products are No.1 in the following: - achieve the breakthrough

More information

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier

More information

Truly digital PET imaging

Truly digital PET imaging Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital PET, resulting in significantly improved

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage

More information

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction

More information

Laue lens for Nuclear Medicine

Laue lens for Nuclear Medicine Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of

More information

Positron Range Modeling for Statistical PET Image Reconstruction

Positron Range Modeling for Statistical PET Image Reconstruction Positron Range Modeling for Statistical PET Image Reconstruction Bing Bai, Ananya Ruangma, Richard Laforest, Yuan-Chuan Tai and Richard M. Leahy Signal and Image Processing Institute, University of Southern

More information

Study of 3D position determination of the interaction point in monolithic scintillator blocks for PET

Study of 3D position determination of the interaction point in monolithic scintillator blocks for PET FACULTY OF SCIENCE Department of Physics Study of 3D position determination of the interaction point in monolithic scintillator blocks for PET Thesis submitted in the fulfillment of the requirements for

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

PET/CT-MRI First clinical experience

PET/CT-MRI First clinical experience 20 th April 2013, Barcelona, Sp PET/CT-MRI First clinical experience Philippe Appenzeller, MD Staff Radiologist and Nuclear Medicine Physician Department Medical Imaging, University Hospital Zurich PET/CT-MR

More information

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

PET, SPECT and CT for Small Animal Imaging

PET, SPECT and CT for Small Animal Imaging State-of of-the-art of PET, SPECT and CT for Small Animal Imaging Alberto Del Guerra Department of Physics and INFN, Pisa,Italy alberto.delguerra@df.unipi.it www.df.unipi.it/~fiig Outline of the talk A

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. 1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

Next Generation Small Animal PET Technology

Next Generation Small Animal PET Technology G4 PET/X-ray P R O D U C T N O T E Pre-clinical in vivo Imaging Key Benefits: High performance, ultra-sensitive PET technology designed for pre-clinical imaging 3D PET and X-ray whole body images Purpose

More information

An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout

An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout M. Anelli, W. Baldini, P. Ciambrone, M. Dallavalle, F. Fabbri, G. Lanfranchi, A. Montanari INFN-LNF, INFN-Ferrara,

More information

Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging

Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging Shaun S. Gleason, Ph.D. 1, James Goddard, Ph.D. 1, Michael Paulus, Ph.D. 1, Ryan Kerekes, B.S.

More information

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM

More information

Proton tracking for medical imaging and dosimetry

Proton tracking for medical imaging and dosimetry Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

The idea of merging data sets to

The idea of merging data sets to CE Techniques, Benefits, and Challenges of PET-MR Positron emission tomography (PET) scans can now be acquired in unison with magnetic resonance (MR) scans as a single resource. This hybrid PET-MR solution

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

ISTITUTO NAZIONALE DI FISICA NUCLEARE

ISTITUTO NAZIONALE DI FISICA NUCLEARE ISTITUTO NAZIONALE DI FISICA NUCLEARE Sezione di Catania INFN/TC-11/02 February 28, 2011 PRELIMINARY TESTS OF A SCINTILLATOR-BASED MINI-STATION FOR EXTENSIVE AIR SHOWERS MEASUREMENTS S.Aiola 1, P. La Rocca

More information

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100) Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know

More information

ASNC IMAGING GUIDELINES FOR NUCLEAR CARDIOLOGY PROCEDURES. Single photon-emission computed tomography

ASNC IMAGING GUIDELINES FOR NUCLEAR CARDIOLOGY PROCEDURES. Single photon-emission computed tomography ASNC IMAGING GUIDELINES FOR NUCLEAR CARDIOLOGY PROCEDURES Thomas A. Holly, MD, a Brian G. Abbott, MD, b Mouaz Al-Mallah, MD, c Dennis A. Calnon, MD, d Mylan C. Cohen, MD, MPH, e Frank P. DiFilippo, PhD,

More information

BME 501 - Introduction to BME. Bioelectrical Engineering Part: Medical Imaging

BME 501 - Introduction to BME. Bioelectrical Engineering Part: Medical Imaging BME 501 - Introduction to BME Bioelectrical Engineering Part: Medical Imaging Reference Textbook: Principles of Medical Imaging, by Shung, Smith and Tsui Lecturer: Murat EYÜBOĞLU, Ph.D. Dept. of Electrical

More information

Monte-Carlo simulation based analysis of performance parameters in the MiniPET scanner, developed for preclinical studies. by Sándor Attila Kis

Monte-Carlo simulation based analysis of performance parameters in the MiniPET scanner, developed for preclinical studies. by Sándor Attila Kis Thesis for the Degree of doctor of philosophy (Ph.D.) Monte-Carlo simulation based analysis of performance parameters in the MiniPET scanner, developed for preclinical studies by Sándor Attila Kis Supervisor:

More information

Information about the T9 beam line and experimental facilities

Information about the T9 beam line and experimental facilities Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions

More information

Cardius. XPO Series. Cardius 3 XPO

Cardius. XPO Series. Cardius 3 XPO Cardius XPO Series Cardius 3 XPO Three imaging systems, all of which provide outstanding image quality, superior efficiency, and increased patient comfort. Cardius 2 XPO Includes the standard features

More information

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution

More information

Performance Characteristics of PET Scanners

Performance Characteristics of PET Scanners 6 Performance Characteristics of PET Scanners A major goal of the PET studies is to obtain a good quality and detailed image of an object by the PET scanner, and so it depends on how well the scanner performs

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831

More information

Three Channel Optical Incremental Encoder Modules Technical Data

Three Channel Optical Incremental Encoder Modules Technical Data Three Channel Optical Incremental Encoder Modules Technical Data HEDS-9040 HEDS-9140 Features Two Channel Quadrature Output with Index Pulse Resolution Up to 2000 CPR Counts Per Revolution Low Cost Easy

More information

MRI for Paediatric Surgeons

MRI for Paediatric Surgeons MRI for Paediatric Surgeons Starship David Perry Paediatric Radiologist Starship Children s Hospital CHILDREN S HEALTH What determines the brightness of a pixel in MRI? i.e. What determines the strength

More information

SAIL Small Animal Imaging Lab

SAIL Small Animal Imaging Lab Prof. T. Helbich/PD. P. Brader SAIL Small Animal Imaging Lab Small Animal Imaging - Warum? Identifikation und Entwicklung von neuen Diagnostika (Kontrastmitteln/Tracern),Therapeutika und deren Kombination

More information

Radiation Strip Thickness Measurement Systems

Radiation Strip Thickness Measurement Systems Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,

More information

Powder diffraction and synchrotron radiation

Powder diffraction and synchrotron radiation Powder diffraction and synchrotron radiation Gilberto Artioli Dip. Geoscienze UNIPD CIRCe Center for Cement Materials single xl diffraction powder diffraction Ideal powder Powder averaging Textured sample

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

Assessing Radiation Dose: How to Do It Right

Assessing Radiation Dose: How to Do It Right Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine

More information

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images

multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images InterView FUSION InterView FUSION is a new visualization and evaluation software developed by Mediso built on state of

More information

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES)

DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES) Features: DIRECTIONAL FIBER OPTIC POWER MONITORS (TAPS/PHOTODIODES) PATENT NUMBERS: CANADA 2,494,133, USA 7095931, 7295731 AND CHINA 1672073 Telcordia GR-468 qualified Available in versions for any wavelength

More information

CHAPTER 7 SCINTILLATION COUNTING

CHAPTER 7 SCINTILLATION COUNTING CHAPTER 7 SCINTILLATION COUNTING Radiation of various types is widely utilized for non-destructive inspection and testing such as in medical diagnosis, industrial inspection, material analysis and other

More information

TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS

TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS P. Moskal, T. Bednarski, P. Białas, M. Ciszewska, E. Czerwiński, A. Heczko, M. Kajetanowicz, Ł. Kapłon, A. Kochanowski, G. Konopka-Cupiał, G. Korcyl,

More information

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Medical and Industrial Tomography - W.B.Gilboy

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Medical and Industrial Tomography - W.B.Gilboy MEDICAL AND INDUSTRIAL TOMOGRAPHY Department of Physics, University of Surrey, Guildford, Surrey, U.K. Keywords: Radiography, transmission tomography, emission tomography, microtomography, SPECT (single

More information

MEDICAL IMAGING 2nd Part Computed Tomography

MEDICAL IMAGING 2nd Part Computed Tomography MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission MAVEN Science Community Workshop December 2, 2012 Particles and Fields Package Solar Energetic Particle Instrument (SEP) Davin Larson and the SEP

More information

Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters

Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters C-A/AP/#122 November 23 Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters J. Beebe-Wang, P. Vaska, F.A. Dilmanian, S.G. Peggs and D.J. Schlyer "This paper

More information

Breast MRI Quality Control

Breast MRI Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

Avalanche Photodiodes: A User's Guide

Avalanche Photodiodes: A User's Guide !"#$%& Abstract Avalanche Photodiodes: A User's Guide Avalanche photodiode detectors have and will continue to be used in many diverse applications such as laser range finders and photon correlation studies.

More information

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays J. E. Wilhjelm Ørsted TU Technical University of enmark, Bldg. 348, K-2800 Kongens Lyngby, enmark. X-ray and Computed Tomography Contents History and characterization of X-rays Conventional (projection)

More information

Exploitation of geometric occlusion and covariance spectroscopy in a gamma sensor array

Exploitation of geometric occlusion and covariance spectroscopy in a gamma sensor array DOE/NV/25946-1846 SPIE ' # Optics+Photonics Exploitation of geometric occlusion and covariance spectroscopy in a gamma sensor array Sanjoy Mukhopadhyay, Richard Maurer, Ronald Wolff, Stephen Mitchell,

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

Improvements in quality and quantification of 3D PET images

Improvements in quality and quantification of 3D PET images Improvements in quality and quantification of 3D PET images The resolution of PET depends on some effects impossible to be readily corrected The implementation of 3D spatially-variant PSF (radially asymmetric)

More information

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Tomohiro Sahara, Hiroshi Tsutsui, Shigehiro Hashimoto, Shuichi Mochizuki, Kenichi Yamasaki, Hideo Kondo, Teruo Miyazaki Department

More information

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999 Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray

More information

Optimization of image quality through online motion correction

Optimization of image quality through online motion correction Advanced imaging technologies a revolution in diagnosis and treatment Optimization of image quality through online motion correction Oline Vinter Olesen December 10, 2010 Medical Solutions, Siemens Institute

More information

Seminar 7. Medical application of radioisotopes - radiotherapy

Seminar 7. Medical application of radioisotopes - radiotherapy Seminar 7 Medical application of radioisotopes - radiotherapy Radioisotopes in medical diagnosis. Gamma camera. Radionuclide imaging (PET, SPECT). Radiotherapy. Sources of radiation. Treatment planning.

More information

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - nzarrin@qiau.ac.ir

More information

Health Policy Advisory Committee on Technology

Health Policy Advisory Committee on Technology Health Policy Advisory Committee on Technology Technology Brief PET MRI integrated hybrid scanners February 2012 State of Queensland (Queensland Health) 2012 This work is licensed under a Creative Commons

More information

PET and PET/CT in Clinical Trials

PET and PET/CT in Clinical Trials PET and PET/CT in Clinical Trials Nathan C. Hall, M.D., Ph.D. The Ohio State University Medical Center CALGB Imaging Core Lab CALGB Imaging Committee Outline Introduction to PET Positron Emitter Physics

More information

Electron-Muon Ranger (EMR)

Electron-Muon Ranger (EMR) Electron-Muon Ranger (EMR) Status of EMR Project R.Asfandiyarov 1, P.Béné 1, R.Bloch 1, A.Blondel 1, D.Bolognini 3 F.Cadoux 1, S.Débieux 1, J-S.Graulich 1, C.Husi 1, D.Lietti 3, F.Masciocchi 1, L.Nicola

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

Concepts for High-Resolution Low-Dose CT of the Breast

Concepts for High-Resolution Low-Dose CT of the Breast RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi

More information

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami

More information

Medical Imaging, Acquisition and Processing

Medical Imaging, Acquisition and Processing Medical Imaging, Acquisition and Processing J. Kybic Center for Machine Perception, CTU Prague Medical Imaging, Acquisition and Processing p.1/59 Overview Introduction and motivation Modalities (acquisition

More information

Scientific Data Visualization Foundation

Scientific Data Visualization Foundation Scientific Data Visualization Foundation Data Sources Scientific Visualization Pipelines Data Acquisition Methods GeoVisualization 1 Scientific Data Sources Common data sources: Scanning devices Computation

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang C1 Medical Imaging Modalities & Characteristics 4005-759 Linwei Wang Major Types of Medical Imaging Modalities X-ray Imaging Computed Tomography (CT) Magnetic Resonance Imaging (MRI) Nuclear Imaging Positron

More information

QAV-PET: A Free Software for Quantitative Analysis and Visualization of PET Images

QAV-PET: A Free Software for Quantitative Analysis and Visualization of PET Images QAV-PET: A Free Software for Quantitative Analysis and Visualization of PET Images Brent Foster, Ulas Bagci, and Daniel J. Mollura 1 Getting Started 1.1 What is QAV-PET used for? Quantitative Analysis

More information

MI Software. Innovation with Integrity. High Performance Image Analysis and Publication Tools. Preclinical Imaging

MI Software. Innovation with Integrity. High Performance Image Analysis and Publication Tools. Preclinical Imaging MI Software High Performance Image Analysis and Publication Tools Innovation with Integrity Preclinical Imaging Molecular Imaging Software Molecular Imaging (MI) Software provides high performance image

More information

Imaging Science. Prof. Dr. rer. nat. Bernd Pichler

Imaging Science. Prof. Dr. rer. nat. Bernd Pichler Imaging Science Prof. Dr. rer. nat. Bernd Pichler Department Radiologie Diagnostic and lnterventional Radiology Prof. Nikolaou Diagnostic and lnterventional Neuroradiology Prof. Ernemann Nuclear Medicine

More information

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Cone Beam Reconstruction Jiang Hsieh, Ph.D. Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Advanced MRI methods in diagnostics of spinal cord pathology

Advanced MRI methods in diagnostics of spinal cord pathology Advanced MRI methods in diagnostics of spinal cord pathology Stanisław Kwieciński Department of Magnetic Resonance MR IMAGING LAB MRI /MRS IN BIOMEDICAL RESEARCH ON HUMANS AND ANIMAL MODELS IN VIVO Equipment:

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry

More information