Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report



Similar documents
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

How To Test For Contamination In Large Volume Water

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

Analysis of Phthalate Esters in Children's Toys Using GC-MS

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Analyzing Small Molecules by EI and GC-MS. July 2014

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Expectations for GC-MS Lab

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

LC-MS/MS for Chromatographers

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

New drugs of abuse? A case study on Phenazepam & Methoxetamine

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

SCION SQ Series GC-MS

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS No. SSI-GCMS Shilpi Chopra, Ph.D.

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Opiates in Urine by SAMHSA GC/MS

Overview. Purpose. Methods. Results

Pesticide Analysis by Mass Spectrometry

Oasis HLB Cartridges and 96-Well Plates

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

AMD Analysis & Technology AG

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

Chemical analysis service, Turner s Green Technology Group

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Czech J. Food Sci. Vol. 25, No. 5:

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Analysis of Liquid Samples on the Agilent GC-MS

Background Information

Supplementary Materials and Methods (Metabolomics analysis)

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Indoor Air Comfort Test Report

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Appendix 5 Overview of requirements in English

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Application of Structure-Based LC/MS Database Management for Forensic Analysis

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

LC-MS/MS, the new reference method for mycotoxin analysis

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Analytical Test Report

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

CORESTA RECOMMENDED METHOD N 72

GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs

SUCRALOSE. White to off-white, practically odourless crystalline powder

How To Use Gc-Ms

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

S olid Ph ase Extraction (SP E) Method:

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Waters Core Chromatography Training (2 Days)

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

LC Application Note.

Determination of Tetracycline Antibiotics in Milk Using a Simple Strong Cation-Exchange SPE Cleanup Procedure and LC-MS/MS Analysis

Purification of reaction mixtures using flash chromatography.

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Protocol: HPLC (amino acids)

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Transcription:

of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley Tel: 020 8943 7678 Prepared by: Checked by: Date:

of melamine-ware by GC-MS and LC-MS/MS Page 2 of 15 Contents 1. Executive Summary 3 2. Background..3 3. Objective...3 4. Materials and methods 3 4.1 Standards and reagents.. 3 4.2 Solvent standards preparation... 4 4.3 Samples and matrix calibration standards... 4 4.4 2,4- DNPH derivatisation.4 4.5 Instrumental analysis 5 5. Results..7 5.1 LC-MS approach.7 5.2 GC-MS approach 7 6. Conclusions.15 7. Recommendations for future work..15 8. References..15

of melamine-ware by GC-MS and LC-MS/MS Page 3 of 15 Method development for analysis of formaldehyde in food-simulant extracts of melamine-ware by GC-MS and LC-MS/MS 1 Executive Summary As part of Mass Spectrometry capability building, the Organic Analysis team were requested to assist in the development of a method for the analysis of formaldehyde in food-simulant extracts of melamine-ware as part of the Government Chemist programme. The analytical techniques investigated were LC-MS/MS (Q-TOF analyser) and GC-MS (single quadrupole analyser). 2 Background Melamine resin is a hard plastic material obtained from the polymerization of melamine and formaldehyde. Melamine resin is often used in kitchen utensils and food contact articles. Formaldehyde and melamine monomer have been found leaching from the resin after exposure to the food simulant (e.g. 3% acetic acid as the most aggressive simulant towards melamine plastics) [1,2]. Formaldehyde can exist naturally in food (e.g. vegetable, meat and seafood) and ingestion of a small amount is unlikely to cause acute adverse health effects. However, intake of large amounts may result in severe abdominal pain, coma or renal injury. One of the most efficient approaches for the determination of carbonyl compounds in aqueous samples is their derivatisation in acidic media with 2,4-dinitrophenylhydrazine (DNPH) to form the corresponding 2,4-dinitrophenylhydrazones. 2,4-dinitrophenylhydrazones can be analysed by reversed-phase HPLC and detected by UV Vis at wavelengths between 349 nm and 380 nm [3]. However, UV Vis detection lacks selectivity especially in complex matrices and doesn t often allow the quantification of carbonyl compounds with the required accuracy. Mass spectrometrybased methods can supply the needed selectivity and sensitivity for the detection and identification of carbonyl compounds and unknown sample constituents. 3 Objective Compliance of food contact melamine-ware is usually checked by evaluation of the Total Specific Migration Limit (SML (T)) of formaldehyde and hexamethylenetetramine (HMTA), expressed as formaldehyde, as required by the European Commission Directive 2002/72/EC [4]. The SML (T) for formaldehyde is 15 mg per Kg of food simulant, equivalent to 2.5 mg/dm 2. The aim of this study was to develop a mass spectrometry-based analytical method for the detection, confirmation and quantitation of formaldehyde at the SML (T) levels in melamine-ware aqueous extracts. Both LC-MS and GC-MS approaches have been investigated, highlighting some of the main issues involved in the sample preparation procedure for the two different methods. However, further development of the LC-MS method was not pursued due to the higher suitability of the GC-MS to this specific task. In fact, the GC-MS system allowed the combined Scan/SIM acquisition mode which could guarantee sensitive quantitation and ID confirmation within one run. For this reason, efforts to obtain a finalised method were focused on the GC-MS methodology. 4 Materials and methods 4.1 Standards and reagents Standards of formaldehyde-dinitrophenylhydrazone (FA-DNPH; Supelco 44-2597) and acetaldehyde- dinitrophenylhydrazone (AA-DNPH; Supelco 44-2434) were obtained from Sigma- Aldrich.

of melamine-ware by GC-MS and LC-MS/MS Page 4 of 15 The derivatising reagent 2,4-dinitrophenylhydrazine (2,4-DNPH) was also purchased from Sigma- Aldrich (Fluka 42215-250mL; phosphoric acid solution). Isotopically labeled formaldehyde 2,4- diphenylhydrazone-3,5,6-d3 (d3-fa-dnph; CDN isotopes D-7765/0.1; lot M28695; isotope purity 99.3 atom% D) as internal standard was supplied by CDN isotopes inc., Quebec, Canada via their UK distributor QMX Laboratories, Thaxted, Essex, UK. The chemical structures of the analytes, derivatising reagent and derivatives are shown in Figure 1 below. (a) (b) (c) (d) (e) Figure 1: Structures of formaldehyde (a), 2,4-DNPH derivatising reagent (b), natural (c) and labeled (d) formaldehyde 2,4- dinitrophenylhydrazones and acetaldehyde 2,4-dinitrophenylhydrazone 4.2 Solvent standards preparation Individual solvent standards of FA-DNPH, AA-DNPH and d3-fa-dnph were prepared in acetonitrile (LGC Promochem Optigrade SO-9184). FA-DNPH was prepared at concentration of 1.2 mg/ml, equivalent to 0.17 mg/ml of FA. The internal standard d3-fa-dnph was prepared at concentration 0.59 mg/ml, equivalent to d3-fa concentration of 0.08 mg/ml. AA-DNPH was also used as additional internal standard in the initial stage of the project when the labeled d3-fa- DNPH was unavailable. 4.3 Samples and matrix calibration standards Samples and matrix calibration standards were provided by FA30. Samples consisted of 3% acetic acid food simulant extract of melamine plates. The extraction was performed at 70 C for 2 hours. Matrix calibration standards were prepared in 3% acetic acid by spiking certain amounts of formaldehyde (FA) standards to achieve final concentrations in solution of 0, 1.5, 5.8, 10.2, 14.6 and 29.1 mg/ml. 4.4 2,4- DNPH derivatisation Prior to the instrumental analysis, samples and matrix standards were derivatised by with 2,4- DNPH [5]. Details of the finalized derivatisation procedure are reported below. Optimized Derivatisation Procedure (GC-MS analysis) 1. 0.5mL of sample were transferred into a 15mL falcon tube (BD falcon 352096). 2. 0.1mL of 2,4-DNPH derivatising reagent were added and the sample was vortex mixed for 30 seconds. 3. 0.175mL of aqueous NH 3 (32%, d= 0.88; VWR 153312K) were added to the mixture and the sample was vortex mixed for 20 seconds.

of melamine-ware by GC-MS and LC-MS/MS Page 5 of 15 4. 0.1mL of internal standard solution (0.59mg/mL d3-fa-dnph in acetonitrile) was added and the sample further vortex mixed for additional 20 seconds. 5. 1mL toluene (Acros Organics 332070025) was added to allow in situ liquid/liquid extraction and the mixture was then vortex mixed for 60 seconds. 6. An aliquot of the organic phase (toluene layer) was transferred to an autosampler vial and then injected into the GC system. The LC-MS/MS preliminary study also involved derivatisation of the samples and matrix standards prior to analysis in order to generate the corresponding DNPH derivatives. However, in this specific case, the sample preparation procedure did not require the in situ liquid/liquid extraction step with toluene to change injection solvent (no aqueous mixture can be injected on the GC-MS system). In its place, 1mL of acetonitrile was added after the derivatising reagent and the extract was injected on the LC-MS system directly. When internal standards are available, the derivatisation procedure shown above for GCMS can be employed with substitution of step 5 for the addition of 1mL of acetonitrile. 4.5 Instrumental analysis LC-MS/MS LC-(+)ESI-MS analysis were performed on an Agilent 1200 HPLC system coupled to an Agilent 6530 Q-TOF mass spectrometer. - HPLC separation conditions HPLC separation was performed using an Agilent C18 Zorbax, 4.6 mm x 50 mm, particle size 1.8 µm. An acetonitrile water binary gradient at a flow-rate of 1 ml/min, column temperature at 40 C using the gradient reported in Table 1. The injection volume was 5 µl. Time [min] % Mobile phase B [ACN] 0 40 1 40 5 80 6 80 6.5 40 8 40 Table 1: HPLC separation gradient - MS detection conditions MS tuning (source and acquisition conditions) for the analytes of interest was optimised by post column mobile phase infusion of solvent standard solutions. Details of the condition settings are reported in Table 2. Ion Source Jet stream (+ESI) Drying Gas N 2 flow [L/min] 8 Drying Gas N 2 T [ C ] 325 Sheath Gas N 2 flow [L/min] 11 Sheath Gas N 2 T [ C ] 350 Capillary voltage [V] 3500 Fragmentor[V] 175 Scan acquisition range [m/z] 100-1000 Scan acquisition rate [ spectra/sec] 1 Table 2: Ion source tuning conditions for 2,4-DNPH derivatives solvent standards

of melamine-ware by GC-MS and LC-MS/MS Page 6 of 15 GC-MS Gas chromatographic analysis was carried out on an Agilent 7890 gas chromatograph coupled to a MSD 5975C inert XL EI/CI with triple-axis detector. Instrumental conditions are reported in Table 3 below. Column: Inlet mode: Restek Rxi-HT 30m x 0.25mm ID x 0.25 µm df Split Inlet Temperature [ C] 260 Septum purge: Split flow: 3mL/min 100 ml/min Split ratio: 100:1 Injection mode: Automatic, Combi PAL G6500 CTC (CTC Analytics) Air volume: 1µL Pre-clean with i-octane: 2µL Pre-clean with sample: 2µL Injection volume: 1µL Filling speed: Filling strokes: 4 Injection speed: Pre-injection delay: Post-injection delay: Post-clean with acetone: 6 Carrier gas: Oven program: 2µL/s 50µL/s 0 ms 500 ms He, 1mL/min Transfer line: 300 C Solvent delay: 3min Source: 300 C 70 C held for 1min 35 C/min to 340 C 340 C held for 2 min Quad: 150 C Acquisition mode: Scan/SIM SIM Ions monitored [m/z]: 152, 155, 180, 183, 198, 210, 213 Dwell time: 15 ms Scan range [m/z]: 50 to 500 Run time [min]: 10.7 Table 3: GC-MS conditions for 2,4-DNPH derivatives analysis

of melamine-ware by GC-MS and LC-MS/MS Page 7 of 15 5 Results 5.1 LC-MS approach Initial work was carried out before the strength of the acid was known, this adversely affected the results from the derivatisation, leading to poor linearity and response of the analyte peaks. However a separation was achieved as is detailed in Figure 2. Figure 2: Extracted Ion Current Chromatogram and Extracted Spectra of FA-DNPH Due to the promise shown by the GC-MS approach, the investigation into the quantitative LC-MS approach was suspended. However, with the knowledge gained on the sample and its derivatisation from the GC method development it is probable that the method could be further developed rapidly. 5.2 GC-MS approach Derivatisation procedure optimisation During the optimisation process, some issues were encountered which needed to be overcome in order to generate a successful finalised method. The derivatising reagent 2,4-DNPH is supplied as 60% H 3 PO 4 solution which significantly decreases the ph of the reagent mixture. Initially, an additional step of ph neutralisation by NH 3 addition was added to the procedure in order to avoid degradation of the GC column stationary phase due to the high acidity of the extract. However, further experiments proved that the high acidity of the extract was also causing hydrogendeuterium exchange (HDX) on the d3-fa-dnph, producing natural FA-DNPH and consequently affecting the ratios between natural and labelled in the tested calibration curves. Figure 3 shows the comparison between results obtained with and without NH 3 neutralisation. Therefore, the neutralisation step was confirmed to be essential to guarantee quality data from the GC-MS analysis.

of melamine-ware by GC-MS and LC-MS/MS Page 8 of 15 Matrix calibration curve [no NH3 neutralization] 1st injection 2nd injection Linear (1st injection) Linear (2nd injection) Area ratio FA-DNPH/d3FA-DNPH 25.00 2 15.00 1 5.00 y = 0.1037x + 13.771 R² = 0.1076 y = 0.0577x + 14.496 R² = 0.0549 0 5 10 15 20 25 30 35 FA-DNPH concentration [ug/ml] Matrix calibration curve [plus NH3 neutralization] 1st injection 2nd injection Linear (1st injection) Linear (2nd injection) 2.50 Area ratio FA-DNPH/d3FA-DNPH 2.00 1.50 1.00 0.50 y = 0.0633x + 0.0526 R² = 0.9985 y = 0.0611x + 0.0713 R² = 0.9947 0 5 10 15 20 25 30 35 FA-DNPH concentration [ug/ml] Figure 3: NH 3 neutralisation effect on matrix calibration curves for FA-DNPH Analyte Scan/SIM characterisation The use of the combined Scan/SIM acquisition mode permits selective and sensitive quantitation (SIM mode) acquiring simultaneously full scan data which can be helpful for ID confirmation purposes. Scan/SIM chromatograms for a solvent standard solution 1 µg/ml and a matrix standard solution 1.5 µg/ml are reported in Figure 4 and Figure 5, respectively.

of melamine-ware by GC-MS and LC-MS/MS Page 9 of 15 Figure 4: Scan/SIM chromatogram for a FA-DNPH solvent standard solution 1 µg/ml (IS d3-fa-dnph 15 µg/ml)

of melamine-ware by GC-MS and LC-MS/MS Page 10 of 15 Figure 5: Scan/SIM chromatogram for a FA-DNPH matrix standard solution 1.5 µg/ml (IS d3-fa-dnph 15 µg/ml

of melamine-ware by GC-MS and LC-MS/MS Page 11 of 15 Full scan spectra of FA-DNPH and d3-fa-dnph are shown in Figure 6 and Figure 7 below. Figure 6: EI spectrum of FA-DNPH Figure 7: EI spectrum of d3-fa-dnph Internal standards evaluation Preliminary experiments assessed the relative performance of different internal standards for this analysis. Initially, AA-DNPH was evaluated for use as internal standard. Calibration solutions for the internal standard method evaluation were prepared in acetonitrile at different FA-DNPH concentrations (0, 1, 5, 10, 15, 25 and 30 µg/ml) and at constant internal standard concentration of 15 µg/ml. Matrix standards and samples were derivatised as described in details in the Materials and Methods Section 4.4. Calibration curves were obtained plotting on the y-axis the peak area ratios of FA-DNPH (m/z 180) to of AA-DNPH (m/z 180) and on the x-axis the FA- DNPH concentration. Figure 8 shows the solvent standard calibration curves obtained from 3 repeated injections of each standard solution.

of melamine-ware by GC-MS and LC-MS/MS Page 12 of 15 Injection 1 Injection 2 Injection 3 y = 0.2325x - 0.0992 R² = 0.9915 y = 0.2472x - 0.0528 R² = 0.9981 y = 0.2583x - 0.0543 R² = 0.9981 Figure 8: Solvent standard calibration curves using AA-DNPH as internal standard Injection 1 Injection 2 Injection 3 y = 0.1018x + 0.4 R² = 0.9865 y = 0.0629x + 0.763 R² = 0.7799 y = 0.0994x + 0.5889 R² = 0.9987 Figure 9: Matrix calibration curves using AA-DNPH as internal standard Calibration curves for 3 repeated injections of derivatised matrix standards are reported in Figure 9. Linearity and repeatability appeared satisfactory, with the exception of one outlier in the second injection dataset. However, background interference or contamination was noticed in the blank matrix standard (0 ug/ml FA-DNPH) as clearly revealed by intercept value above zero.

of melamine-ware by GC-MS and LC-MS/MS Page 13 of 15 Figure 10 shows the comparison between the 0 ug/ml FA-DNPH (m/z 210) matrix standard and 0 ug/ml FA-DNPH solvent standard using d3-fa-dnph (m/z 213) as internal standard. Clearly the matrix standard shows the presence of the natural FA-DNPH due to contamination whereas the same peak is absent in the solvent standard solution. Figure 10: Matrix standard contamination effect The isotopically labelled d3-fa-dnph was tested as internal standard as well. As previously mentioned in the Material and Methods sections, HDX issues were observed before the addition of the neutralisation step to the sample preparation methodology. The HDX caused the interconversion between natural and labelled consequently affecting the performance of d3-fa-ndph as internal standard. However, once the methodology had been optimised, d3-fa-dnph proved to be a good internal standard for the purposes of this study as shown in the solvent and matrix calibration curves in Figure 11. Both solvent and matrix calibration curves showed good linearity and repeatability of the results. RSD% ranged between 1 and 30 %.

of melamine-ware by GC-MS and LC-MS/MS Page 14 of 15 Peak area ratio FA-DNPH/d3-FA-DNPH 3.00 2.50 2.00 1.50 1.00 0.50 Matrix calibration curve [m/z210/] 1st injection 2nd injection Linear (1st injection) Linear (2nd injection) y = 0.075x + 0.2081 R² = 0.9985 y = 0.0709x + 0.3166 R² = 0.9509 0 5 10 15 20 25 30 35 FA-DNPH concentration [µg/ml] Peak area ratio FA-DNPH/d3-FA-DNPH 3.50 3.00 2.50 2.00 1.50 1.00 0.50 Solvent calibration curve [m/z 210/213] y = 0.0906x - 0.0189 R² = 0.9997 0 5 10 15 20 25 30 35 40 FA-DNPH concentration [µg/ml] Peak area ratio FA-DNPH/d3-FA-DNPH 2.50 2.00 1.50 1.00 0.50 Matrix calibration curve [m/z 152/155] 1st injection 2nd injection Linear (1st injection) Linear (2nd injection) y = 0.0632x + 0.3135 R² = 0.9982 y = 0.0621x + 0.4146 R² = 0.9658 0 5 10 15 20 25 30 35 FA-DNPH concentration [µg/ml] Peak area ratio FA-DNPH/d3-FA-DNPH 3.00 2.50 2.00 1.50 1.00 0.50 Solvent calibration curve [m/z 152/155] y = 0.0768x + 0.1281 R² = 0.9997 0 5 10 15 20 25 30 35 40 FA-DNPH concentration [µg/ml] Peak area ratio FA-DNPH/d3-FA-DNPH 0.70 0.60 0.50 0.40 0.30 0.20 0.10 Matrix calibration curve [m/z 180/183] 1st injection 2nd injection Linear (1st injection) Linear (2nd injection) y = 0.016x + 0.0774 R² = 0.9993 y = 0.0172x + 0.0893 R² = 0.9975 0 5 10 15 20 25 30 35 FA-DNPH concentration [µg/ml] Peak area ratio FA-DNPH/d3-FA-DNPH 2.50 2.00 1.50 1.00 0.50 Solvent calibration curve [m/z 180/183] y = 0.0651x + 0.0681 R² = 0.9988 0 5 10 15 20 25 30 35 40 FA-DNPH concentration [µg/ml] Figure 11: Solvent and matrix standard calibration curves for natural and d3-labeled FA-DNPH

of melamine-ware by GC-MS and LC-MS/MS Page 15 of 15 6 Conclusions Two mass spectrometry-based analytical methods were developed for the determination of formaldehyde in food-simulant extracts of melamine-ware. A derivatisation procedure was optimised for the detection of formaldehyde as 2,4-dinitrophenylhydrazone. An LC-MS approach was preliminarily evaluated but not further developed due to the superiority of GC-MS methodology in this case. Two different internal standards were tested for this study, both of them showing good results in terms of linearity and repeatability in solvent and matrix calibration curves. A fully quantitative procedure by GC-MS was developed using the labeled d3- formaldehyde-2,4-dinitrophenylhydrazone as internal standard. The method developed is applicable to food-simulants, however there is scope for application to other aqueous extractions or formulations, such as cosmetics, with further investigation into matrix effects during the analysis stage. 7 Recommendations for further work Investigation into the source of background contamination and methods for its reduction or elimination would be beneficial since the 0 µg/ml matrix calibration solutions showed significant levels of the analyte FA-DNPH. Optional clean-up procedures might be investigated to reduce the contamination of the GC-MS system and deliver higher robustness throughout the analysis. 8 References [1] K. H. Lund & J. H. Petersen, Migration of formaldehyde and melamine monomers from kitchen- and tableware made of melamine plastic, Food Additives and Contaminants, 2006, 23 (9): 948-955 [2] E. L Bradley, V Boughtflower, T. L Smith, D. R Speck & L Castle, Survey of the migration of melamine and formaldehyde from melamine food contact articles available on the UK market, Food Additives and Contaminants, 2005, 22 (6): 597-606 [3] W. Pötter, U. Karst, Identification of Chemical Interferences in Aldehyde and Ketone Determination Using Dual-Wavelength Detection, Analytical Chemistry, 1996, 68:3354-3358 [4] European Commission Directive 2002/72/EC of 6th August 2002 relating to plastic materials and articles intended to come into contact with foodstuffs, Article 7. Official Journal of the European Communities (2002) L220 [5] G. Zurek, U. Karst, 2,4-Dinitro-3,5,6-trideuterophenylhydrazones for the quantitation of aldehydes and ketones in air samples by liquid chromatography mass spectrometry, Journal of Chromatography A, 2000, 869: 251 259