WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA. Vehicle Radiation Monitoring Systems for Medical Waste Disposal

Similar documents
Results from Studies on Radiation Alarms at Waste Management Facilities

RC4000: Vehicle Monitoring Radiation Detection Systems

Introduction to Geiger Counters

A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

Optimization of Digital Industrial Radiography (DIR) Techniques for Specific Applications: An IAEA Coordinated Research Project

Overview of Nuclear Detection Needs for Homeland Security. Abstract

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:


INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

SARA (Spectroscopic Ambient Radiation Detection) Spectroscopic Monitoring Systems for Online Environmental Radiation Monitoring Edition 2008

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

Atomic and Nuclear Physics Laboratory (Physics 4780)

WORLD CUSTOMS EXHIBITION AND FORUM, Budapest, th September 2003

Tutorial 4.6 Gamma Spectrum Analysis

National Radiation Air Monitoring Network National Environmental Monitoring Conference San Antonio -- August 6, 2013

GAMMA-RAY SPECTRA REFERENCES

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

RADIOACTIVE MATERIALS REGULATORY GUIDE

TEMPLE UNIVERSITY ENVIRONMENTAL HEALTH AND RADIATION SAFETY

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N

Panel on Economic Development Radiation monitoring after the Japanese Fukushima nuclear power plant incident

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

Radiation Monitoring System RMS

Cronos -1 Gamma Object / Tool Monitors

EDXRF of Used Automotive Catalytic Converters

Radiological mapping and characterization at the Barsebäck nuclear power plant

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Quality Assurance Testing of AtomLab Dose Calibrators

Nuclear accidents and radioactive contamination of foods 30 March 2011

Assessment of environmental radiation monitoring data in Hungary following the Fukushima accident

Radiation Detection and Measurement

identifinder R-Series Handheld Radiation Detectors

Alarm Response Guidelines for Radiation Portal Monitoring Systems

NORM MANAGEMENT GUIDELINES & PRINCIPLES

notes MERRTT IS-302 Radiological Survey Instruments and Dosimetry Devices

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks

ASTM C26 LONG RANGE PLAN Committee on the Nuclear Fuel Cycle

Homework #10 (749508)

READING DEVICE US-05

Treatment Centers for Radioactive Waste

CHAPTER 7 SCINTILLATION COUNTING

Thermo Scientific Matrix Mobile ARIS Mobile Radiation Detection System

Nuclear applications at the Hellenic Centre for Marine Research (HCMR): Radioactivity in the marine environment

Radiation Strip Thickness Measurement Systems

Gamma and X-Ray Detection

EPA Radionuclides Rule and the RadNet Program

Lessons learned from the radiological accident in Mayapuri, New Delhi, India

A Review of Emerging Gamma Detector Technologies for Airborne. Radiation Monitoring

RADIATION DETECTION & ANALYSIS SYSTEMS FOR WATER

Pedestrian and post-luggage monitors. Vehicle monitors. Mobile monitors. Integrated monitoring system

An Innovative, portable MCA based on Digital Signal Processing. R.D. Bingham, R.M. Keyser, T.R. Twomey

Environmental Radiation Monitoring in Taiwan

For convenience, we may consider an atom in two parts: the nucleus and the electrons.

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

EMERGENCY RESPONSE FOR THE AREA SURROUNDING THE CATTENOM NUCLEAR POWER PLANT

(including ASBU RADIATION SUMMARY HISTORY WORLD INTERNATIONAL CIVIL AVIATION ORGANIZATION MET/14-IP/ /20. 4/6/14 English only

Thyroid Screening for Radioiodine

Laboratory Technologies, Inc. Multi-Wiper. Time Saving Multiple Detector Wipe Counter

EMERGENCY PROCEDURES FOR RADIOACTIVE MATERIAL. 2. Prevent the spread of contamination by covering the spill with absorbent paper.

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software

Measurement solutions for nuclear safety and security. software customization services. Nuclear MeasureMeNts BusiNess unit of areva

PNNL Radioxenon Detector. January MP Foxe JI McIntyre

Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient

WM2012 Conference, February 26-March 1, 2012, Phoenix, Arizona, USA

Hazard Classification of the Remote Handled Low Level Waste Disposal Facility

ORTEC. High Resolution Detection Systems for Interdiction of Nuclear Material Trafficking. Abstract. Introduction

Captus 3000 Thyroid Uptake And Well System

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

MCQ - ENERGY and CLIMATE

Prepared for the Department of Energy Office of Transportation and Emergency Management. 02B p65

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

notes Hazard Recognition

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

REEVALUATING ELEMENTAL ABUNDANCE ON THE MOON. Kenneth E. Dobbs Department of Political Science University of Hawai i at Mānoa Honolulu, HI 96822

Post Fukushima Off Site Radiological Monitoring Enhancements. January, 2015 Fuji Electric Corp. of America

APPENDIX B SUPPLEMENTAL INSPECTION PROGRAM A. OBJECTIVES AND PHILOSOPHY OF THE SUPPLEMENTAL INSPECTION PROGRAM

USER MANUAL OPERATION AND USE OF CAR WITH. Diego G3 / NEVO SEQUENTIAL GAS INJECTION SYSTEM

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

UK RESPONSE PLAN AND RADIOACTIVE INCIDENT MONITORING NETWORK (RIMNET) 3: BRIEFING NOTE

Results of Airborne Monitoring by the Ministry of Education, Culture, Sports, Science and Technology and the U.S. Department of Energy

Fukushima Daiichi Nuclear Power Station Unit 1-3 Evaluation method of the present amount of radioactive material released from the Reactor Building

RADIATION MONITORING NETWORK OF THE CZECH REPUBLIC

Examination of National and International Impacts of Adoption of ICRP Recommendations

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY

Ramon 2.2 Radon-Monitor

GRID AND PRISM SPECTROMETERS

Clean, Crisp Water Bottled Water Versus Tap Water

FREQUENTLY ASKED QUESTIONS: National Response to I-131 from Japan Nuclear Power Plant March 27, Environmental Protection Agency (EPA)

Radiological Protection Principles concerning the Natural Radioactivity of Building Materials

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

The following assumptions are to be considered when reviewing this model procedure:

Radiation Monitoring System DLMon

Activities at temporary job site utilizing U.S. NRC Radioactive Material License # Amendment (,

IBC Nuclear Energy Liability Exclusion. Explained

Drayton Digistat +2RF/+3RF

X Ray Flourescence (XRF)

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software

Evaluation of 55-gallon Lead-Lined Calibration for the Gamma Modality of the High Efficiency Neutron Counter

Transcription:

Vehicle Radiation Monitoring Systems for Medical Waste Disposal - 12102 Vladislav S. Kondrashov and Steve A. Steranka RadComm Systems Corp., 2931 Portland Dr., Oakville, ON L6H 5S4, CANADA ABSTRACT Hospitals often declare their waste as being non-radioactive ; however this material often has excessive levels of radiation caused either by an accident or lack of control. To ensure the best possible protection against the accidental receipt of radioactive materials and as a safety precaution for their employees, waste-handling companies have installed large-scale radiation portal monitors at their weigh scales or entry gates of the incinerator plant, waste transfer station, and/or landfill. Large-volume plastic scintillator-based systems can be used to monitor radiation levels at entry points to companies handling medical waste. INTRODUCTION In accordance with Canada s regulation [1], hospitals can store low-level radioactive waste with short half-life isotopes (e.g., Tc-99m, I-131) until they have decayed and then dispose of it as clean waste [2]. In terms of document [1], non-radioactive means that the material produces equivalent dose less than 0.05 msv/a. The hospital waste is then transported, handled, and either incinerated or placed in a landfill by government-owned or privately held companies. Despite declared as non-radioactive, the hospital waste sometimes has levels of radiation higher than would be expected due to the lack of control or an accident. To ensure the best possible protection against the accidental receipt of radioactive materials and as a safety precaution for their employees, waste-handling companies only accept waste that is less than double the background level. There are two reasons that have driven this level of acceptance: first, the problem of accumulated radiological waste at the same facility; second, materials passing into the United States from Canada have been subjected to severe fines after setting off the security-based Border radiation portal monitors. For these reasons companies have installed large radiation portal monitors at their weigh scales or entry gates of the Incinerator plant, waste transfer station, and/or landfill. For large radiation portal monitors the most practical solution is the use of high grade plastic scintillation detectors. The reasons for the widespread popularity of plastic scintillators are as follows: relatively good sensitivity, easy to fabricate and most important its low cost compared to other scintillation materials. The popular plastic scintillator polyvinyl toluene (PVT) has an order of magnitude lower cost in comparison with a higher efficiency inorganic scintillator such as sodium iodide (NaI) A plastic scintillator has poor energy resolution and cannot be used for nuclide identification by gamma-spectra. Regardless of this drawback, good radiation sensitivity (20-30% of NaI) makes them effective tools as radiation screening devices. 1

Despite the naturally degradation processes which is associated with plastic scintillators, the life-span of PVT radiation detectors is considered to be 10-12 years, making this material a long-term solution for waste-management companies. This presentation describes these systems and addresses the problem of radiation dose assessment with a plastic scintillator. METHOD Radiation detection technique The plastic scintillator (PVT based) RC4000 systems are typically configured with two detector panels, a dedicated PC with a network connection allowing remote monitoring, data retrieval and maintenance/service functions. The internal software and hardware designs are extremely flexible, allowing remote software updates and electronic hardware adjustments when necessary. Supervisors can monitor the system operation in real-time to ensure normal system operation is always maintained. if a network connection is provided, the system has the capability of emailing alarms and system malfunctions. Detailed logging of critical historic information is a useful feature of these systems. A case in point is Japan, where historical data had been retrieved before and during the earthquake, during the tsunami and the months that followed. The data provided critical information about the radiation in the atmosphere in specific areas. In addition, course energy-analysis information was determined to assist with improved alarm threshold analyses. The principal block-diagram of a typical portal radiation-monitoring system offered for waste-handling companies is shown in Figure 1. 2

Fig 1. This diagram shows a typical layout for a vehicle radiation-detection monitor with two detectors. Dose rate assessment Waste-management companies require displayed radiation levels and/or alarm assessment primarily in dose rates, unlike the systems for steel and metal scrap plants where total count (CPS) information is considered the normal units of measured radiation levels. To achieve this goal RadComm Systems has designed and implemented a special approach for plastic scintillators (PVT). This approach incorporates real-time analysis algorithms providing energy compensation and calibration for assessment of Dose Rates. The major Dose Rate correction factor originates from the spectral response from the plastic scintillator Compton Edges. Unlike an inorganic scintillator or semiconductor detectors, the plastic has no Photo Peaks and experimental spectral response can be seen in Figure 2 (for Cs- 137, 662 kev). Compton edge of 662 kev Count rate, cps Energy, kev Fig 2. Actual experimental response of plastic scintillator (PVT) for Cs-137 radioisotope The Compton edges are the primary source of information produced by plastic scintillators. In order to properly evaluate Dose Rates, the correction factors of incident photons must be introduced. The Compton edge shifts the incident spectral response to lower energy regions. As the vehicle passes through the radiation detection scanning zone and if a photon produced by a radioactive source in the load of waste reaches the plastic scintillator, as an approximation 3

the direct hit by the incident photon on the scintillator s electron would be assumed for further consideration. In this situation the backscatter photon has well-known predictable energy E b_sc, kev: _ = 1 + 2 511 (. 1) where the E 0 is incident photon energy in kev. Therefore, the Compton electron energy becomes the maximal: or, _ = _ (. 2) _ = 1 1 + 511 2 (. 3) The Compton electron deposits this energy into the plastic scintillator and generates the scintillator response. However, as it follows from (Eq. 3) this energy is lower in comparison with the incident photon by the factor outlined below: Therefore, to correct the estimated dose rate based on PVT Compton energy response, one should multiply the evaluated dose rate on incident photons by the following correction factor: = 1 + (. 4) Where the Dose corr is the dose correction for incident photons and Dose scint is directly obtained from plastic scintillator. For example, the correction factor for Cs-137 producing 662 kev is 1.39. RESULTS AND DICUSSION To satisfy waste-management companies with their requirement for alarm thresholds and dose rate displayed measurements,i.e., to alarm if radiation from load of waste materials exceeds twice the ambient background and to provide the readings in dose rate units, two principal objectives were required: A) Control the maximum allowable radiation response from a passing vehicle 4

B) Permanent dose rate assessment To achieve the first objective, a reliable and continuous real-time background tracking algorithm was developed. This algorithm incorporates an especially designed locking radiation response technique to evaluate maximum amplitude of radiation signals for alarm threshold settings. The second objective was achieved as described earlier in the Dose Rate Assessment section for plastic scintillators. The typical response of the PVT-based system where the measured radiation signal has exceeded the allowable limit is shown on Figure 3. Fig 3. The background updating and alarm response of the PVT-based systems (time radiation profile) Figure 3 shows the Real-Time vehicle radiation profile (blue line) of the systems installed at landfills. The red line represents the accurate updating alarm threshold, which locks when the measured radiation Dose Rate from the detected radioactive source inside the vehicle exceeds twice background. The result causes the system to set an alarm condition. 5

Fig 4. View of the standard User Interface with radiation time profile (green lines) displayed in Equivalent Dose Rate units (nsv/hr). Nevertheless, the plastic scintillator based radiation detection systems cannot properly differentiate between medical and non-medical isotopes. To address this issue of isotopic identification and for detailed investigation of waste which has generated an alarm condition on the PVT-based portal radiation detection system, RadComm offers the handheld Gamma-Ray Spectrometer with nuclide ID and neutron detection capability. CONCLUSION The recent and intensive field tests together with the thousands of accumulated hours of actual real-life vehicle scanning have proven that the plastic scintillation based system is an appropriate radiation control instrument for waste management companies. The Real-Time background compensation algorithm is flexible with automatic adjustable coefficients that will 6

response to rapidly changing environmental and weather conditions maintaining the preset alarm threshold levels. The Dose Rate correction algorithms further enhance the system s ability to meet the stringent requirements of the waste industries need for Dose Rate measurements. REFERENCES 1. Radiation Safety Requisites for Exemption of Certain Radioactive Materials from further licensing upon transferral for disposal. Ottawa: Atomic Energy Control Board, 1989. Regulatory document R-85 2. Assessing the Long Term Safety of Radioactive Waste Management. Canadian Nuclear Safety Commission, 2006. Regulatory Guide G-320 7