1-(2-Pyridylazo)-2-Napthol (PAN) Method 1 Method to 2.00 mg/l Co Powder Pillows

Similar documents
USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Quaternary Ammonium Compounds DOC

Phosphorus, Reactive (Orthophosphate)

Color, True and Apparent

Color, True and Apparent

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate

DR/4000 PROCEDURE. CHLORINE, Free

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

USEPA Electrode Method Method 8156

Solids, Volatile Dissolved and Fixed Dissolved

Total Aerobic Bacteria, Yeasts and Molds

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

Calcium Analysis by EDTA Titration

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

Preparation of frequently used solutions

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Experiment 12- Classification of Matter Experiment

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Solutions and Dilutions

Total Water Hardness

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Determination of the Mass Percentage of Copper in a Penny. Introduction

Chapter 16: Tests for ions and gases

OXIDATION-REDUCTION TITRATIONS-Permanganometry

Complexometric Titrations

Determination of Citric Acid in Powdered Drink Mixes

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Pocket Colorimeter II

Additional Lecture: TITRATION BASICS

Analysis of Vitamin C Using Iodine. Introduction

Complexometric Titration Analysis of Ca 2+ and Mg 2+ in seawater

(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid

Chapter 8 - Chemical Equations and Reactions

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE

Mixtures and Pure Substances

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS

Colorimetric Determination of Iron in Vitamin Tablets

SODIUM CARBOXYMETHYL CELLULOSE

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

Experiment 8 - Double Displacement Reactions

Solubility Product Constants

FT FIELD MEASUREMENT OF RESIDUAL CHLORINE

Standard Methods for the Examination of Water and Wastewater

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Factors Affecting Enzyme Activity

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Determination of calcium by Standardized EDTA Solution

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Reaction Stoichiometry and the Formation of a Metal Ion Complex

Experiment 16-Acids, Bases and ph

Colorimetric Determination of Iron in Vitamin Tablets

Where the exp subscripts refer to the experimental temperature and pressure acquired in the laboratory.

Ion Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp , 699

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

The Determination of an Equilibrium Constant

Phenolphthalein-NaOH Kinetics

The Molar Mass of a Gas

A Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado

Carolina s Solution Preparation Manual

Thermo Scientific Orion AQUAfast Powder and Tablet Reagent Chemistry User Guide

MOISTURE (Karl Fischer, Buffered)

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

1. Qualitative Analysis of Chromium, Iron, and Copper

Syllabus OC18 Use litmus or a universal indicator to test a variety of solutions, and classify these as acidic, basic or neutral

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

EXPERIMENT 8: Activity Series (Single Displacement Reactions)

COMMON LABORATORY APPARATUS

PROTEINS (LOWRY) PROTOCOL

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Neutralizing an Acid and a Base

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Odor (Threshold Odor, Consistent Series)

EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet. by Professor David Cash. September, 2008

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

The volume of a penny will be calculated from its mass and density.

THE ACTIVITY OF LACTASE

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

ENCM Environmental Impact Assessment and Environmental Monitoring. Determination of nitrate and phosphate levels in well water/surface water

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

Experiment 7: Titration of an Antacid

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media

Notes Chapter 9 Limiting Reagent Sample Problems Page 1

Transcription:

Cobalt DOC316.53.01036 1-(2-Pyridylazo)-2-Napthol (PAN) Method 1 Method 8078 0.01 to 2.00 mg/l Co Powder Pillows Scope and application: For water and wastewater. Digestion is required to determine total recoverable cobalt. If EDTA is in the sample, use the vigorous digestion. 1 Adapted from Watanabe, H., Talanta, 21 295 (1974). Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows sample cell and orientation requirements for specific instruments. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR 3900 The fill line is to the right. 2495402 The fill line is toward the user. Before starting To make sure that all forms of the metal are measured, digest the sample with heat and acid. Use the mild or vigorous digestion. Refer to the Water Analysis Guide for more information. The recommended temperature for samples and reagents is 15 25 C (59 77 F). For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option. This method can measure the nickel concentration on the same sample with Program Number 340. Make sure to use a reagent blank for the nickel procedure. The Pour-Thru Cell can only be used (for applicable instruments) with reagents for 25-mL samples. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Cobalt/Nickel Reagent Set, PAN 1 Water, deionized 10 ml 1

Items to collect (continued) Description Quantity Stopper, Neoprene, solid, size #1 2 Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.) 2 Refer to Consumables and replacement items on page 5 for order information. Sample collection and storage Test procedure Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1) hydrochloric acid and rinsed with deionized water. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated nitric acid (approximately 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at room temperature for a maximum of 6 months. Before analysis, adjust the ph to 3 8 with 5.0 N sodium hydroxide standard solution. Do not exceed ph 8 to prevent precipitation of the metal. Correct the test result for the dilution caused by the volume additions. Start 1. Start program 110 Cobalt. For information about sample cells, adapters or light shields, refer to Instrument-specific information on page 1. 2. Prepare the blank: Fill a sample cell with 10 ml of deionized water. 3. Prepare the sample: Fill a second sample cell with 10 ml of sample. 4. Add the contents of one Phthalate-Phosphate Reagent Powder Pillow to each cell. Note: Although the program name can be different between instruments, the program number does not change. 2 Cobalt, PAN Method (2.00 mg/l)

5. Close the sample cells. Immediately shake to dissolve the reagent. If the sample contains iron, make sure that all the powder is dissolved before the PAN Indicator Solution is added. 6. Add 0.5 ml of 0.3% PAN Indicator Solution to each cell. 7. Put the stopper on the sample cells. Invert several times to mix. 8. Start the instrument timer. A 3-minute reaction time starts. During color development the sample solution color can change from green to dark red, based on the chemical composition of the sample. The blank will be yellow. 9. When the timer expires, add the contents of one EDTA Reagent Powder Pillow to each cell. 10. Close the sample cells. Shake to dissolve the reagent powder. 11. Clean the blank sample cell. 12. Insert the blank into the cell holder. Zero Read 13. Push ZERO. The display shows 0.00 mg/l Co. 14. Clean the prepared sample cell. 15. Insert the prepared sample into the cell holder. 16. Push READ. Results show in mg/l Co. Interferences Interfering substance Interference level Al 3+ 32 mg/l Ca 2+ 1000 mg/l as (CaCO 3 ) Cd 2+ Cl 8000 mg/l Cobalt, PAN Method (2.00 mg/l) 3

Interfering substance Chelating agents (e.g., EDTA) Cr 3+ Cr 6+ Cu 2+ F Fe 3+ Fe 2+ K + Mg 2+ Mn 2+ Mo 6+ Na + Pb 2+ Zn 2+ Highly buffered samples or extreme sample ph Interference level Interfere at all levels. Use either the Digesdahl or vigorous digestion to remove this interference. 40 mg/l 15 mg/l 10 mg/l. If the sample contains iron, make sure that all the powder is dissolved before the PAN Indicator is added. Interferes directly and must not be present. 500 mg/l 400 mg/l 25 mg/l 60 mg/l 5000 mg/l 30 mg/l Can prevent the correct ph adjustment of the sample by the reagents. Sample pre-treatment may be necessary. Accuracy check Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: 1000-mg/L Cobalt Standard Solution 1-L volumetric flask, Class A 100-mL volumetric flask, Class A 10-mL volumetric pipet, Class A and pipet filler safety bulb Deionized water 1. Prepare a 10.00-mg/L cobalt stock solution as follows: a. Use a pipet to add 10.00 ml of a 1000-mg/L cobalt standard solution into a 1-L volumetric flask. b. Dilute to the mark with deionized water. Mix well. Prepare the stock solution each day. 2. Prepare a 1.00-mg/L cobalt standard solution as follows: a. Use a pipet to add 10.00 ml of the 10.00-mg/L cobalt stock solution into a 100- ml volumetric flask. b. Dilute to the mark with deionized water. Mix well. Prepare the standard solution each day. 3. Use the test procedure to measure the concentration of the prepared standard solution. 4. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are slight variations in the reagents or instruments. 4 Cobalt, PAN Method (2.00 mg/l)

Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% confidence interval) Sensitivity Concentration change per 0.010 Abs change 110 1.00 mg/l Co 0.99 1.01 mg/l Co 0.01 mg/l Co Summary of Method After the sample is buffered and pyrophosphate is added to mask any Fe 3+, the cobalt reacts with 1-(2-Pyridylazo)-2-Naphthol indicator. The indicator forms complexes with most metals. After color development, EDTA is added to destroy all metal-pan complexes except nickel and cobalt which can be determined with the same sample preparation. The measurement wavelength is 620 nm. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Cobalt/Nickel Reagent Set, PAN, 10-mL, includes: 100/pkg 2651600 EDTA Reagent Powder Pillow 2 100/pkg 700599 Phthalate-Phosphate Reagent Powder Pillow, 10-mL 2 100/pkg 2615199 PAN Indicator Solution, 0.3% 1 ml 100 ml MDB 2150232 Water, deionized varies 4 L 27256 Required apparatus Description Quantity/test Unit Item no. Stoppers for 18-mm tubes and AccuVac Ampuls 2 6/pkg 173106 Recommended standards Description Unit Item no. Cobalt Standard Solution, 1000-mg/L Co 100 ml 2150342 Optional reagents and apparatus Description Unit Item no. Flask, volumetric, Class A, 100-mL glass each 1457442 Flask, volumetric, Class A, 1000-mL glass each 1457453 Pipet, volumetric, Class A, 10-mL each 1451538 Pipet filler, safety bulb each 1465100 Nitric Acid, concentrated 500 ml 15249 Sodium Hydroxide Standard Solution, 5.0 N 100 ml MDB 245032 Cobalt, PAN Method (2.00 mg/l) 5

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 2007, 2010, 2012, 2014. All rights reserved. 03/2014, Edition 8