FT FIELD MEASUREMENT OF RESIDUAL CHLORINE

Size: px
Start display at page:

Download "FT 2000. FIELD MEASUREMENT OF RESIDUAL CHLORINE"

Transcription

1 FT FIELD MEASUREMENT OF RESIDUAL CHLORINE Use in conjunction with: FT 1000 General Field Testing and Measurement FQ 1000 Field Quality Control Requirements FS 1000 General Sampling Procedures FD 1000 Documentation Procedures Chlorinating water supplies and polluted waters at treatment facilities is a common practice to destroy or deactivate disease-producing microorganisms. This SOP describes the most commonly used field procedures for measuring the residual chlorine that remains after chlorination. FT DPD COLORIMETRIC METHOD 1. SCOPE AND APPLICABILITY: FDEP recommends this method if the testing levels that must be met are between 0.2 to 4.0 mg-cl/l. This method only measures total residual chlorine as opposed to the titrimetric and iodometric-electrode methods that can distinguish between free and combined forms of chlorine. The DPD-colorimetric is best suited for polluted waters (e.g., wastewater effluents) because it is the least affected by the presence of organic matter in the sample. 2. EQUIPMENT AND SUPPLIES 2.1. Field Instruments: Photometric equipment: Spectrophotometer, for use at a wavelength of 515 nm and a light path of 1 cm or longer; or Filter Photometer, equipped with a filter having maximum transmission in the wavelength range of 490 to 530 nm and providing a light path of 1 cm or longer The instrument must allow calibration using at least one standard and a blank Color Wheel Comparator (only approved for drinking water compliance) Sample Cells (cuvettes): Use sample cells or tubes of clear, colorless glass or plastic Keep cells scrupulously clean, both inside and out, and discard if scratched or etched Never handle them where the light beam strikes the sample Clean sample cells by thorough washing with laboratory soap (inside and out) followed by multiple rinses with distilled or de-ionized water, and let air-dry Mask minor imperfections or scratches in the cells with a very thin layer of silicone oil on the outside surfaces. Use silicone oil with the same refractive index of the glass; making sure the cell appear to be nearly dry with little or no visible signs of oil. Page 1 of 20 Revision Date: January 1, 2002

2 Because small differences between cells significantly impact measurement, use either matched pairs or the same cell for standardization and sample measurement Reagents and Standards: DPD reagent powder pillows (from commercial supplier) or Phosphate buffer solution and DPD indicator solution (prepared according to EPA or SM 4500-Cl-G) Standard Potassium Permanganate solutions ( Chlorine Standards ) prepared in the appropriate concentration range of interest according to method EPA or SM 4500-Cl-G If used, two sets of secondary standards from a commercial vendor Chlorine-demand-free water and appropriate glassware Recordkeeping and Documentation Supplies: Field notebook (w/ waterproof paper is recommended) Field/Lab record forms (e.g., forms FD , FD and FD ) Indelible pens 3. CALIBRATION AND USE 3.1. Interferences: Sample color and turbidity may interfere in all colorimetric procedures applied to natural and treated waters. Other possible interferences are from: bromine, chlorine dioxide, iodine, permanganate, hydrogen peroxide, and ozone. Bromine will always interfere when measuring residual chlorine in estuarine and marine waters Instrument Calibration (spectrophotometer or photometer): These procedures apply to photometric instruments; the color wheel comparator is factory calibrated See FT 1000, section 2.2 for specific quality control measures that must be observed Follow the manufacturer s operating instructions Initial Calibration: Use the chlorine standards (see above) for initial calibration of the photometers Use as many standards as allowed by the instrument. FDEP recommends 4 standard concentrations and a blank, however the instrument must have the ability to be calibrated with at least one standard and a blank If the sample concentration range is unknown, calibrate with standards in the 0.05 to 4.0 mg-cl/l range The correlation coefficient of the standard curve must be greater than or equal to If only one standard is used to calibrate, the standard must be greater than the highest expected sample concentration. Check the accuracy of the curve by measuring a standard like a sample, and comparing the result with the known standard value. This check standard must be near the expected concentration of the majority of the samples and the instrument reading must be within 10% of the standard value Continuing Calibration: Page 2 of 20 Revision Date: January 1, 2002

3 Use at least two standards at two different concentrations in the field for continuing calibration checks The standards may be secondary standards or chlorine standards (see above) If the stated manufacturer s accuracy of any secondary standard is greater than 10% of the true or stated value, the standard cannot be used for a calibration check At least one of the standards must be greater than the highest expected sample concentration. The second standard must be near the expected concentration of the majority of the samples Each check must be within 10% of the true or manufacturer s stated value Sample Analysis (spectrophotometer or photometer): Rinse the cuvette with a small amount of sample Discard the rinsing, and fill the cuvette with sample If samples are colored or turbid, compensate for these interferences by zeroing the meters with the sample before adding the chemicals for color development Add the reagents (DPD, buffer, etc.) to the sample according to the manufacturer s instructions Invert the container to mix Gently wipe the sides with a lint-free tissue and insert the cuvette into the instrument Follow the instrument s instructions for obtaining the sample value (either by direct read-out or by using a plot of the calibration curve) After recording the sample reading, remove the cuvette, discard the sample and rinse the cuvette 3 5 times with de-ionized or distilled water Repeat steps through until all samples have been tested Color Wheel Comparator: Follow the manufacturer s instructions for obtaining the sample value from a color wheel when used for drinking water sample analysis. 4. PREVENTIVE MAINTENANCE: Refer to FT 1000, section RECORDS: See FT 1000, section 4 for additional details Calibration: Record all calibration information (initial and continuing) in the field notebook or on a form specifically designed for this purpose (e.g. FD ). This information must be recorded: The calibration method (or SOP) and the type of standard(s) (including the concentrations) used Record each calibration check (initial, continuing or final) in the permanent field records (or calibration logs). At a minimum, these records must include: Date, time and location (e.g., monitoring well MW-X; laboratory; etc.) of each calibration check; Page 3 of 20 Revision Date: January 1, 2002

4 Individual performing the check; Results of each check, including the concentration/type of standard, expected reading, and the actual reading; Whether the check met or failed acceptance criteria; Readings associated with a failed check; and Corrective actions associated with failed check (such as recalibration, removal from use, etc.) Optionally, enter the meter name, model number, and identification number (if applicable) Samples: Record the sample results on the appropriate field parameter sheets (e.g., Form FD ) or in a field notebook. For each sample record the following information: Project Name Date and time the measurement was collected Location of the sample (description, monitor well number, etc.) Analyte Name Reporting Units Sample Value Initials of the person taking the measurements 5.3. Reagent and Standards Documentation: Maintain documentation on calibration standards (e.g., permanganate, gel standards) and other reagents. FT At a minimum, note the date of receipt, expiration dates (on the bottle label), and date of first use (on the standard container) Follow expiration dates: If any standard or chemical is used after the expiration date, there must be documentation showing that the reagent is providing an acceptable response. SPECIFIC REQUIREMENTS FOR WASTEWATER FACILITIES USING SECONDARY STANDARDS NOTE: FT 2011 ONLY APPLIES TO DOMESTIC WATEWATER FACILITIES WITH A RESIDUAL CHLORINE REQUIREMENT EQUAL TO OR GREATER THAN 0.5 mg-cl/l. EPA and SM 4500-Cl G require the use of chlorine or potassium permanganate solutions to calibrate the spectrophotometer. These must be prepared on a daily basis, and require skill in preparing the volumetric dilutions. Secondary standards, such as gel or dyes, are usually marketed in three dilutions, plus a blank. These can be used without additional preparation and are much easier to use and handle. The problem is that the low level standard does not have sufficient accuracy to be used if low-level chlorine measurements are required. Secondary standards may be used if a plant must meet basic level disinfection requirements and must maintain a residual chlorine level of at least 0.5 mg/l. Use secondary standards for daily calibration checks only if the procedures discussed below are Page 4 of 20 Revision Date: January 1, 2002

5 followed. Facilities having permit requirements with chlorine residual standards of less than 0.5 mg/l cannot use secondary standards, but must use a titration method if the permit limit is less than 0.2 mg/l or standardize the instrument using either chlorine or permanganate standards as discussed in FT These procedures are divided into two sections: Section 1 deals with the activities that are associated with the storage, and use of standards in the field. Section 2 deals with the procedures, tests and calibrations that must be performed by: A laboratory with certification for residual chlorine in water from the Florida Department of Health Environmental Laboratory Certification Program (DoH ELCP) or A certified operator holding an A or B wastewater operator's license in the State of Florida. 1. FIELD PROCEDURES 1.1. Calibration Verification: Verify all new standard sets before first use (see section below) On each day of use check the instrument calibration: Zero the meter using the blank provided by the manufacturer Read the mid-range standard as if it was a sample and record the reading. The standard must be within 10% of the stated manufacturer s value Read the high and low standards as if they were samples and record the readings. The values must be within the tolerances allowed by the manufacturer If the standards do not meet these criteria, the meter is not usable. Either recalibrate the meter or return to the manufacturer. In addition, re-verify the standards to ensure accuracy. DO NOT ATTEMPT to adjust the calibration Perform this calibration check at intervals of no greater than 24 hours. Perform an additional check at the end of the day, after the last sample measurement has been taken Unacceptable calibration checks during, or at the end of the day mean that the standards and/or the meter are no longer functioning properly. Do not use the standards and the meter until the standard concentrations have been re-verified and the working condition of the meter has been checked (see above). All measurements taken between the last acceptable calibration check and the failed calibration check may not be accurate, and if reported, the letter J must follow each value At least once a week, check a blank, prepared with chlorine-free water and the DPD reagent after the meter has been checked as described above. This blank must read no more than 0.05 mg/l. If the blank is greater than 0.05 mg/l, discard the DPD reagent because it is no longer producing the expected result Required documentation for the daily calibration verification: Lot number of standards. Page 5 of 20 Revision Date: January 1, 2002

6 Acceptance range for each of the standard concentrations. Example: if the mid standard is 0.7 mg/l, then calculate the acceptance range by taking 10% of 0.7 (0.07) and adding and subtracting the value from the original concentration. The acceptance range for a 0.7 mid-range standard would be 0.63 to 0.77 mg/l Values for the low, medium and high standards Indication of whether the standards meet the acceptance criteria Value for the water blank (see above) Care of Secondary Standards: Protect standards from direct sunlight, heat and scratches Store the standards in the upright position. The most convenient way is to keep the standards in the original storage box provided by the manufacturer During transport to and from the field, keep the standards in an air-conditioned vehicle away from the sun (not on the dashboard or in the trunk) If the vehicle is not air conditioned, the standards may be placed in a watertight plastic bag and stored on ice or on a cold blue ice pack in a cooler When not in use, store the standards in an upright position in air conditioning or in ambient air not to exceed 90 o F Avoid getting fingerprints on the glass vial by handling the standards only by the plastic cap Before each day s use, inspect the secondary standard vials and clean the outside of the standard vials with a dilute solution of soapy water, followed by a tap water rinse. Wipe dry with lint-free paper or cloth Discard the standard set when: You notice scratches in the glass vial; You notice that a secondary standard has melted, and/or has melted and solidified in a different position from the original standards; and The standards have exceeded the manufacturer s expiration date Equipment and Standard Maintenance: Instrument: At least once a quarter, perform the following maintenance tasks on the spectrophotometer/colorimeter. These tasks must be performed by a laboratory with DoH ELCP certification for residual chlorine or by a certified operator holding an A or B wastewater operator's license in the State of Florida. Request the following: Check the instrument to ensure proper working condition including performing any maintenance tasks necessary to ensure proper operation Perform an initial calibration using primary (chlorine or permanganate) standards Issue a statement certifying that the instrument is in proper working condition, and has been calibrated using primary standards. The statement must bear the signature of the laboratory director (or certified A or B operator), the date that the instrument was calibrated, and the certification number of the laboratory (if performed by a laboratory). Page 6 of 20 Revision Date: January 1, 2002

7 Standards: Each of the standards in a set must be clearly identified as belonging to that set, and may not be mixed with standards from other sets On a monthly basis, check one of the two sets of standard standards against primary standards. These checks must be performed by a certified laboratory or a certified operator holding an A or B wastewater operator's license in the State of Florida. These tasks must be performed: Check the secondary standards to ensure that: The standards have not expired; The standards have not been mishandled (e.g., stored so that the gel has melted into a different form); and The glass cuvettes are free from fingerprints or scratches. If the standards have exceeded the expiration date, gel standards appear to be deformed, the standards appear to be discolored, or the cuvettes are damaged, discard the standard set Check the accuracy of the secondary standards by: Testing each of the secondary standards with a properly calibrated spectrophotometer as if they were samples and record the value. The standard at the high end and low end must meet manufacturer specifications. The mid level standard (~ 0.78 mg/l) must be within 10% of the stated value to be acceptable. If any of the standards do not meet the stated criteria, the user must be notified that a new set of standards must be purchased. Standards that do not meet the calibration acceptance criteria must be discarded Issue a statement signed by the laboratory director (or certified A or B operator) certifying that: The standard cuvettes and standards are in acceptable condition; and Each individual standard meets the acceptance criteria in above. The certificate must also report the manufacturer s value and the value obtained by the laboratory (or A or B operator), the date that the test was performed and if performed by a laboratory, the environmental certification number Retain all certificates (both instrument and standard calibration) from the laboratory (or operator) in a permanent file for a period of five years after they have been issued Records: Maintain the following records: Daily Calibration Checks: Date and time of calibration check The true value and the instrument value of each standard Lot number of calibration standards. Page 7 of 20 Revision Date: January 1, 2002

8 Acceptance range for each standard (may be listed when the standards are newly purchased) Secondary Standard Records: Date standards were originally received Expiration date of standards (per manufacturer) Original certificate of tolerance provided by the manufacturer Dates standards were sent to the laboratory (or Class A or B operator) for certification The laboratory name and DoH certification number used for certification or the name of the Class A or B operator The laboratory (or operator) results and certification statement for each recertification Instrument Records: Date instrument was originally received Dates instrument was sent to the laboratory (or Class A or B Operator) for certification The laboratory name and DoH certification number used for certification or the name of the Class A or B operator The laboratory (or operator) results and certification statement for each certification Preventative maintenance and repairs performed on the instrument Retain all records for a period of at least five (5) years from the date that the information was generated. 2. LABORATORY OR CLASS A AND B OPERATOR RESPONSIBILITIES: 2.1. When the spectrophotometer is received: Check the instrument to ensure proper working condition; Perform an initial calibration using primary (chlorine or permanganate) standards and make adjustments necessary to ensure that the instrument is properly calibrated If the initial calibration is acceptable (see FT 2010, section 3.2.3), issue a statement certifying that the instrument is in proper working condition, and has been calibrated using primary standards. The statement must bear the signature of the laboratory director (or A or B operator), the date that the instrument was calibrated, and the certification number of the laboratory (if performed by a laboratory) On a monthly basis, one of the two sets of standards must be submitted to check against primary standards: Check the standards to ensure that: The standards have not expired; The standards have not been mishandled (e.g., stored such that the gel has melted into a different form); and Page 8 of 20 Revision Date: January 1, 2002

9 The glass vials are free from fingerprints or scratches If the standards have exceeded the expiration date, gel standards appear to be deformed, the color is not correct, or the cuvettes are damaged, notify the user that a new set of standards must be purchased Check the accuracy of the secondary standards by testing each of the standards as if they were samples with a spectrophotometer calibrated according to the analytical method (i.e. permanganate or chlorine standards). The standard at the high and low end must meet manufacturer specifications. The mid-level standard (~ 0.78 mg/l) must be within 10% of the stated value to be acceptable If any of the standards do not meet the criteria in 2.3 above, notify the user that a new set of standards must be purchased Issue a certification signed by the laboratory director (or Certified A or B operator) that certifies that: The standards (vials or cuvettes and the gel or dye) are in acceptable condition; and Each individual standard meets the acceptance criteria in 2.3 above The certificate should also report the manufacturer s value and the value obtained by the laboratory, the date that the test was performed and the environmental certification number of the laboratory In addition to the analytical records associated with the secondary standards, laboratories (or A or B operators) must retain records on the initial calibration and maintenance that are performed on the meter and the clients to whom the instrument and standard calibration services are provided. 3. RECORDS 3.1. Record all calibrations (initial and continuing) in the field notebook or in a form specifically designed for this purpose Record field measurements in the appropriate field/lab parameter sheets (e.g., forms FD , FD and FD ) or in the field notebook. FT TITRIMETRIC METHODS Residual chlorine can be measured using several different titrimetric methods. These methods take more time, require some skill in making chemical measurements, require several different reagents and must be done in the field. Unless the expected concentration levels are below the detectability of colorimeters, FDEP recommends using the DPD colorimetric method (FT 2010). If a titrimetric method must be used, select one of the following methods: EPA Methods 330.1, 330.2, 330.3, or 330.4, or SM 4500-Cl B, SM 4500-Cl C, SM 4500-Cl D, or SM 4500-Cl F. The following general discussions deal with using the titrimetric equipment only. 1. The volumetric glassware that is used to transfer and to titrate must be thoroughly clean Use class A pipets for transferring samples and reagents; use a class A buret for titration The volumetric pipets and the buret must be free flowing. This means that there cannot be any droplets remaining on the sides of the glassware when the liquid is dispensed. Page 9 of 20 Revision Date: January 1, 2002

10 1.3. Check the glassware before use by filling with de-ionized or distilled water and allowing to drain. Do not use the glassware if droplets appear Clean by using hot sudsy water, and a buret brush or soaking in sudsy water Rinse with tap water to remove the soap Rinse with de-ionized water and check for water droplets Additional cleaning with chromic acid, or alcoholic potassium hydroxide may be needed to remove stubborn oil or films. 2. Rinse all glassware with a small amount of the solution to be dispensed/titrated, and discard the rinse Carefully fill (or pipet) with the solution Read the volume from the bottom of the meniscus [figure ] After filling the buret, record the initial volume. 3. When performing the titration, add the titrant slowly Swirl the flask gently while adding the titrant As the solution comes closer the endpoint, the color change will last for a longer period of time. When this occurs, add the titrant dropwise and mix thoroughly between drops The reaction is complete when the color change is permanent. Take the reading on the buret at the completion of the titration and record the reading. 4. Titrate at least two aliquots of each sample. 5. Titrate at least one blank. 6. Record the preparation of all reagents in the field notebook or in a form specifically designed for this purpose. 7. For each sample and blank, record the volume of each sample, and the beginning and ending readings on the buret. FT SELECTIVE ION ELECTRODE METHODS 1. INTRODUCTION AND SCOPE: Adapted from the Orion Research Instruction Manual, Residual chlorine Electrode Model 97-70, This method may be used if a method detection limit study verifies that the method can achieve the desired permit limits. This is not recommended for on-site use. 2. EQUIPMENT AND SUPPLIES 2.1. Instrument ph-millivolt meter capable of an expanded 0.1 mv reading Platinum-iodide combination electrode (recommended) or A platinum and an iodide ion selective electrodes Magnetic Stirrer 2.2. Reagents and Standards Page 10 of 20 Revision Date: January 1, 2002

11 ph 4.0 Buffer Acetic acid Potassium Iodide Solution: Dissolve 42 g potassium iodide and 0.2 g sodium carbonate in 500 ml chlorine demand-free deionized/distilled water. Potassium iodate (Primary Standard N): Dissolve g potassium iodate (KIO 3 ) in chlorine demand-free deionized/distilled water. Dilute to 1000 ml Recordkeeping and Documentation Supplies: Field notebook (w/ waterproof paper is recommended) Field/Lab record forms (e.g., forms FD , FD and FD ) Indelible pens 3. CALIBRATION AND USE 3.1. Standard Preparation: prepare standards by volumetrically pipetting 0.2, 1.0 and 5.0 ml pf the potassium iodate into 100 ml volumetric flasks Add 1 ml of the buffer and 1 ml potassium iodide solution to each Swirl and let stand for 2 minutes Bring to 100 ml, stopper and invert several times to mix Prepare a blank by adding all reagents except the potassium iodate to a 100 ml volumetric and bring the volume to 100 ml Initial Calibration: Follow Manufacturer s instructions on the use of the instrument. Use the three standards and the blank to calibrate. Plot the curve on semilogarithmic graph paper or enter values into a direct-reading ion meter Continuing Calibration: Use at least one of the standards to check the stability of the standard curve at the end of the analytical day Follow the quality control requirements for calibration (see FT 1000, section 2.2) Sample Analysis: Adjust the ph of the sample with acetic acid to 4 5 ph units Pipet 1 ml buffer and 1mL potassium iodide into a 100 ml volumetric flask. Swirl to mix and let stand for at least 2 minutes Bring to the volume to 100 ml with the ph-adjusted sample, stopper and invert several times to mix Follow manufacturer s instructions for the use of the instrument Calculations: Read the mg chlorine/l (R) from the calibration curve: mg chlorine/l = R X 100/mL sample used 4. PREVENTIVE MAINTENANCE: Refer to FT 1000, section RECORDS: See FT 1000, section 4 for additional details. Page 11 of 20 Revision Date: January 1, 2002

12 5.1. Calibration: Record all calibration information (initial and continuing) in the field notebook or on a form specifically designed for this purpose (e.g. FD ). This information must be recorded: The calibration method (or SOP) and the type of standard(s) (including the concentrations) used Record each calibration check (initial, continuing or final) in the permanent field records (or calibration logs). At a minimum, these records must include: Date, time and location (e.g., monitoring well MW-X; laboratory; etc.) of each calibration check; Individual performing the check; Results of each check, including the concentration/type of standard, expected reading, and the actual reading; Whether the check met or failed acceptance criteria; Readings associated with a failed check; and Corrective actions associated with failed check (such as recalibration, removal from use, etc.) Optionally, enter the meter name, model number, and identification number (if applicable) Samples: Record the sample results on the appropriate field parameter sheets (e.g., Form FD ) or in a field notebook. For each sample record the following information: Project Name Date and time the measurement was collected Location of the sample (description, monitor well number, etc.) Analyte Name Reporting Units Sample Value Initials of the person taking the measurements 5.3. Reagent and Standards Documentation: Maintain documentation on calibration standards (e.g., [potassium iodate) and other reagents. FT At a minimum, note the date of receipt, expiration dates (on the bottle label), and date of first use (on the standard container) Follow expiration dates: If any standard or chemical is used after the expiration date, there must be documentation showing that the reagent is providing an acceptable response. USING THE METHOD OF STANDARD ADDITIONS TO VERIFY THE ABSENCE OF ANY RESIDUAL CHLORINE 1. INTRODUCTION AND SCOPE: In some cases, wastewater treatment plants may need to determine the complete absence of residual chlorine after dechlorination. Some metered Page 12 of 20 Revision Date: January 1, 2002

13 instruments may not have sufficient sensitivity to measure below 0.2-mg/L chlorine. This procedure was developed by Region 4 of the Environmental Protection Agency to be used to verify, but not report the absence of residual chlorine. 2. METHOD: Use any residual chlorine method. If a colorimetric method (FT 2010) is used, the spectrophotometer or colorimeter must be able to distinguish between 0.02 and 0.05 mg/l chlorine. To document the sensitivity, perform a method detection limit study to verify that the method detection limit is at least 0.02 mg/l. 3. PROCEDURE: 3.1. If used, calibrate the meter following FT Otherwise, use a titrimetric method (FT 2020) Perform all tests within 15 minutes of collecting the samples Collect a grab sample at a point just before dechlorination or the effluent of the chlorine contact basin. This is sample A Collect a second grab sample after dechlorination. This should correspond to the final effluent monitoring point (sample B) Sample A and Sample B must differ only by the dechlorination treatment. There cannot be any waste streams that enter the system between the two collection points Measure and pour equal volumes of Sample A and Sample B into a clean container that can be sealed. Use a container that leaves little or no headspace when filled. This is sample C Cap and gently invert the container (sample C) several times to mix the contents. Do not shake vigorously Measure the total residual chlorine in samples A and C Calculate the concentration in sample B: Residual Chlorine in B (mg/l) = [C*2] A 3.8. The absence of residual chlorine is verified if the residual chlorine in B is less than or equal to ½ A*1.10. Refer to figure FT for illustrative examples If the value results in a negative number and a sulfite salt (e.g., sodium bisulfite, sodium metabisulfite) was used to dechlorinate the effluent, less dechlorination agent can be added. FT REFERENCES 1. U.S. Environmental Protection Agency, EPA 330.1, Methods for the Chemical Analysis of Water and Waste, EPA-600/ , Revised U.S. Environmental Protection Agency, EPA 330.2, Methods for the Chemical Analysis of Water and Waste, EPA-600/ , Revised U.S. Environmental Protection Agency, EPA 330.3, Methods for the Chemical Analysis of Water and Waste, EPA-600/ , Revised U.S. Environmental Protection Agency, EPA 330.4, Methods for the Chemical Analysis of Water and Waste, EPA-600/ , Revised Page 13 of 20 Revision Date: January 1, 2002

14 5. U.S. Environmental Protection Agency, EPA 330.5, Methods for the Chemical Analysis of Water and Waste, EPA-600/ , Revised American Public Health Association, American Water Works Association, Water Pollution Control Federation, SM 4500-Cl B, Standard Methods for the Examination of Water and Wastewater, 20 th Edition, American Public Health Association, American Water Works Association, Water Pollution Control Federation, SM 4500-Cl C, Standard Methods for the Examination of Water and Wastewater, 20 th Edition, American Public Health Association, American Water Works Association, Water Pollution Control Federation, SM 4500-Cl D, Standard Methods for the Examination of Water and Wastewater, 20 th Edition, American Public Health Association, American Water Works Association, Water Pollution Control Federation, SM 4500-Cl F Standard Methods for the Examination of Water and Wastewater, 20 th Edition, Florida Department of Environmental Protection Environmental Assessment Section, The Use of Secondary Gel Standards for Residual Chlorine Calibration, DEP-QAS-99-02, September 22, U.S. Environmental Protection Agency, Region 4, Monitoring Chlorine and Sulfur Dioxide at Wastewater Treatment Plants, undated Paper. 12. Thermo-Orion, Orion Research Instruction Manual, Residual Chlorine Electrode Model 97-70, Page 14 of 20 Revision Date: January 1, 2002

15 Appendix FT 2000 Tables, Figures and Forms Figure FT Scenario for Total Chlorine Analysis Form FD Maintenance and Calibration Log for Residual Cl Spectrophotometer Form FD Field Calibration Check Log for Residual Cl Spectrophotometer Form FD Residual Cl Secondary Standard Calibration Log Page 15 of 20 Revision Date: January 1, 2002

16 Figure FT SCENARIO FOR TOTAL CHLORINE ANALYSIS (TRC) SO 2 DECHLORINATION FINAL DISCHARGE CHLORINE CONTACT BASIN * TRC = 0.5 MG/L * TRC =0.0 MG/L * Sample A Sample B A = Grab sample collected just prior to dechlorination. Sample C = TRC = 0.25 MG/L B = Grab sample collected downstream of dechlorination at final discharge point. C = Equal volumes of A & B mixed without aeration and the expectant TRC concentration mg/l = A + B = = If 1/2 A = C, then the TRC at Sample B = <0.01 or the MDL value determined on the instrument whichever was greater. If 1/2 A > C, then excess SO 2 was added. The TRC concentration would be <0.25 mg/l at Sample C and the TRC for Sample B = 0.0 mg/l and would be reported as a < value. If C > 1/2 A, then Sample B would have TRC present in final effluent. To calculate the Sample B value: [C * 2] - A = TRC concentration. Example: If the TRC concentration of C = 0.27 and A = 0.5, then 0.27 * 2 = 0.54, and the TRC at Sample B = [ ] = 0.04 mg/l. Page 16 of 20 Revision Date: January 1, 2002

17 Page of Maintenance and Calibration Log for Residual Chlorine Spectrophotometer FACILITY NAME: INSTRUMENT MAKER: INSTRUMENT MODEL #: DATE RECEIVED (YY/MM/DD) BY FDEP PERMIT #: DATE PURCHASED: SERIAL NUMBER: CALIBRATED BY (LABORATORY / OPERATOR NAME) DOH CERT # ACCEPTABLE CONDITION? YES NO NOTE: IN ADDITION TO THIS LOG, THE FOLLOWING DOCUMENTS MUST ALSO BE RETAINED BY THE FACILITY: THE LABORATORY CERTIFICATION STATEMENTS, AND THE MANUFACTURER S MANUAL AND WARRANTIES. Page 17 of 20 Form FD (June 1, 2001) Revision Date: January 1, 2002

18 ENTER ALL MAINTENANCE AND REPAIR ACTIVITIES ON THIS PAGE Page of DATE NAME/SIGNATURE DESCRIPTION OF ACTIVITY Page 18 of 20 Form FD (June 1, 2001) Revision Date: January 1, 2002

19 Field Calibration Check Log for Residual Cl Spectrophotometer Page of CALIBRATION CHECKS MUST BE PERFORMED AT THE BEGINNING OF EACH WORKING DAY AND AT FOUR- HOUR INTERVALS THEREAFTER. A FINAL CALIBRATION CHECK MUST BE PERFORMED AFTER THE LAST MEASUREMENT HAS BEEN MADE IN THE DAY OF USE. DATE YY/MM/DD TIME HR:MIN OPERATOR SIGNATURE LOT # OF STANDARDS TRUE VALUE (MG/L) METER VALUE (MG/L) ACCEPTANCE RANGE ACCEPTABLE? YES NO Page 19 of 20 Form FD (June 1, 2001) Revision Date: January 1, 2002

20 Residual Cl Secondary Standard Calibration Log Page of A DOH-CERTIFIED LABORATORY (OR A CERTIFIED A OR B OPERATOR) MUST, ON A MONTHLY BASIS, CHECK SECONDARY STANDARDS. COMPLETE ONE PAGE OF THIS LOG FOR EACH SET OF STANDARDS. RETAIN THE MANUFACTURER S CERTIFICATIONS AND REPORTS FROM THE LABORATORY (OR A OR B OPERATOR) ON THE RESULTS OF THE STANDARD CHECKS. VENDOR: LOT #: PURCHASE DATE: EXPIRATION DATE: DATE CHECKED YY/MM/DD CHECKED BY (LABORATORY NAME OR A OR B OPERATOR) DOH ACCEPTABLE? CERTIFICATION NUMBER YES NO INITIALS Page 20 of 20 Form FD (June 1, 2001) Revision Date: January 1, 2002

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506 Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking

More information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,

More information

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater

More information

Color, True and Apparent

Color, True and Apparent Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method

More information

Acid Base Titrations

Acid Base Titrations Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually

More information

A Volumetric Analysis (Redox Titration) of Hypochlorite in Bleach

A Volumetric Analysis (Redox Titration) of Hypochlorite in Bleach CHEM 311L Quantitative Analysis Laboratory Revision 2.3 A Volumetric Analysis (Redox Titration) of Hypochlorite in Bleach In this laboratory exercise, we will determine the concentration of the active

More information

Phosphorus, Reactive (Orthophosphate)

Phosphorus, Reactive (Orthophosphate) Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from

More information

DR/4000 PROCEDURE. CHLORINE, Free

DR/4000 PROCEDURE. CHLORINE, Free DR/4000 PROCEDURE CHLORINE, Free Method 10069 DPD method* UHR (0.1 10.0 mg/l as Cl 2 ) Scope and Application: For testing higher levels of free chlorine (hypochlorous acid and hypochlorite ion) in drinking

More information

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required

More information

Quaternary Ammonium Compounds DOC316.53.01128

Quaternary Ammonium Compounds DOC316.53.01128 Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower

More information

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin An Introduction to Standards and Quality Control for the Laboratory Barbara Martin Hach Company, 2002. All rights reserved. Lit. No. 2426 eac/rb 8/02 1ed In memory of Clifford C. Hach (1919-1990) inventor,

More information

Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution

Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Volatile Acids DOC316.53.01144 Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Scope and application: For digestor sludges. 1 Adapted from The Analyst, 87, 949 (1962).

More information

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2

More information

COMMON LABORATORY APPARATUS

COMMON LABORATORY APPARATUS COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen

More information

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows

FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for

More information

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1 Experiment 13H FV 1/25/2011(2-run) THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:

More information

Determination of Citric Acid in Powdered Drink Mixes

Determination of Citric Acid in Powdered Drink Mixes Determination of Citric Acid in Powdered Drink Mixes Citric acid and its salts (sodium citrate and potassium citrate) are found in many foods, drinks, pharmaceuticals, shampoos, and cosmetics. The tartness

More information

Measurement by Ion Selective Electrodes

Measurement by Ion Selective Electrodes Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable

More information

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination

More information

Private Water Supplies Sampling Manual. A Field Guide

Private Water Supplies Sampling Manual. A Field Guide Private Water Supplies Sampling Manual A Field Guide Foreword This sampling manual details standard procedures for the collection, storage and transportation of samples at private water supplies in accordance

More information

Determination of the Mass Percentage of Copper in a Penny. Introduction

Determination of the Mass Percentage of Copper in a Penny. Introduction Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny

More information

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856

Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows

More information

Chem 131A: Absorbance of Riboflavin

Chem 131A: Absorbance of Riboflavin Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry EXPERIMENT #5 THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 I. PURPOSE OF THE EXPERIMENT In this experiment

More information

Phosphates (ortho- and total)

Phosphates (ortho- and total) INTRODUCTION Phosphates (ortho- and total) Phosphorus is an essential nutrient for all aquatic plants and algae. Only a very small amount is needed, however, so an excess of phosphorus can easily occur.

More information

CHEM 161: Beer s Law and Analysis of a Sports Drink

CHEM 161: Beer s Law and Analysis of a Sports Drink CHEM 161: Beer s Law and Analysis of a Sports Drink Introduction Although sunlight appears white, it contains a spectrum of colors. A rainbow actually shows this range of colors in visible light: violet,

More information

Calibration of Volumetric Glassware

Calibration of Volumetric Glassware CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;

More information

Analyzing the Acid in Vinegar

Analyzing the Acid in Vinegar Analyzing the Acid in Vinegar Purpose: This experiment will analyze the percentage of acetic acid in store bought vinegar using titration. Introduction: Vinegar can be found in almost any home. It can

More information

Calcium Analysis by EDTA Titration

Calcium Analysis by EDTA Titration Calcium Analysis by EDTA Titration ne of the factors that establish the quality of a water supply is its degree of hardness. The hardness of water is defined in terms of its content of calcium and magnesium

More information

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq) Weak Acid Titration v120413 You are encouraged to carefully read the following sections in Tro (2 nd ed.) to prepare for this experiment: Sec 4.8, pp 158-159 (Acid/Base Titrations), Sec 16.4, pp 729-43

More information

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND #3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric

More information

Vitamin C Content of Fruit Juice

Vitamin C Content of Fruit Juice 1 Vitamin C Content of Fruit Juice Introduction Vitamin C Vitamins are organic compounds that have important biological functions. For instance, in humans they enable a variety of enzymes in the body to

More information

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container. TITRATION: STANDARDIZATION OF A BASE AND ANALYSIS OF STOMACH ANTACID TABLETS 2009, 1996, 1973 by David A. Katz. All rights reserved. Reproduction permitted for education use provided original copyright

More information

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy Determining the Free Chlorine Content of Swimming Pool Water Computer 33 Physicians in the nineteenth century used chlorine water as a disinfectant. Upon the discovery that certain diseases were transmitted

More information

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE

Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure

More information

Experiment 4 (Future - Lab needs an unknown)

Experiment 4 (Future - Lab needs an unknown) Experiment 4 (Future - Lab needs an unknown) USING A ph TITRATION TO DETERMINE THE ACID CONTENT OF SOFT DRINKS 2 lab periods Reading: Chapter 9, 185-197; Chapter 10, pg 212-218; Chapter 14 pg 317-323,

More information

(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid

(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid The Determination of Hypochlorite in Bleach Reading assignment: Chang, Chemistry 10 th edition, pages 156-159. We will study an example of a redox titration in order to determine the concentration of sodium

More information

Preparation of frequently used solutions

Preparation of frequently used solutions Preparation of frequently used solutions Content 1. Diluting Concentrated Acids (Last Login: 08/08/2009) 2. Indicators (Last Login: 27/07/2009) 3. Standard Buffer Solutions (Last Login: 27/07/2009) 4.

More information

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

USEPA Electrode Method Method 8156

USEPA Electrode Method Method 8156 ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method

More information

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 OBJECTIVE The objective of this experiment will be the standardization of sodium hydroxide using potassium hydrogen phthalate by the titration

More information

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date: PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA

More information

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate

Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate Oxygen Scavengers DOC316.53.01105 Iron Reduction Method Method 8140 3 to 450 µg/l DEHA; 5 to 600 µg/l carbohydrazide; 9 to 1000 µg/l hydroquinone; 13 to 1500 µg/l iso-ascorbic acid (ISA); 15 to 1000 µg/l

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

Experiment 7: Titration of an Antacid

Experiment 7: Titration of an Antacid 1 Experiment 7: Titration of an Antacid Objective: In this experiment, you will standardize a solution of base using the analytical technique known as titration. Using this standardized solution, you will

More information

Acid Dissociation Constants and the Titration of a Weak Acid

Acid Dissociation Constants and the Titration of a Weak Acid Acid Dissociation Constants and the Titration of a Weak Acid One of the most important applications of equilibria is the chemistry of acids and bases. The Brønsted-Lowry acid-base theory defines an acid

More information

Juice Titration. Background. Acid/Base Titration

Juice Titration. Background. Acid/Base Titration Juice Titration Background Acids in Juice Juice contains both citric and ascorbic acids. Citric acid is used as a natural preservative and provides a sour taste. Ascorbic acid is a water-soluble vitamin

More information

ph: Measurement and Uses

ph: Measurement and Uses ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic

More information

Lab #11: Determination of a Chemical Equilibrium Constant

Lab #11: Determination of a Chemical Equilibrium Constant Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a

More information

Experiment 17: Potentiometric Titration

Experiment 17: Potentiometric Titration 1 Experiment 17: Potentiometric Titration Objective: In this experiment, you will use a ph meter to follow the course of acid-base titrations. From the resulting titration curves, you will determine the

More information

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Salicylate Method * 1. Enter the stored program number for ammonia nitrogen (NH 3 -N). 2. Press: 64 ENTER The display

More information

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.

15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration. S HIFT INTO NEUTRAL 15. Acid-Base Titration Shift into Neutral Student Instruction Sheet Challenge Discover the concentration of an unknown acid solution using acid-base titration. Equipment and Materials

More information

Calibration of Glassware

Calibration of Glassware Calibration of Glassware Introduction Glassware is commonly calibrated using a liquid of known, specific density, and an analytical balance. The procedure is to determine the mass of liquid the glassware

More information

Factors Affecting Enzyme Activity

Factors Affecting Enzyme Activity INTRODUCTION Factors Affecting Enzyme Activity The chemical reactions occurring in living things are controlled by enzymes. An enzyme is a protein in the cell which lowers the activation energy of a catalyzed

More information

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy 2-1 Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy Introduction: The kinetics of a decomposition reaction involving hydroxide ion and crystal violet, an organic dye used

More information

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration EXPERIMENT 7 Identifying a Substance by Acid-Base Titration SAFETY WARNING In this experiment you will be working with NaOH pellets and using 0.25 M NaOH as a titrant. Sodium hydroxide is extremely basic,

More information

OXIDATION-REDUCTION TITRATIONS-Permanganometry

OXIDATION-REDUCTION TITRATIONS-Permanganometry Experiment No. Date OXIDATION-REDUCTION TITRATIONS-Permanganometry INTRODUCTION Potassium permanganate, KMnO 4, is probably the most widely used of all volumetric oxidizing agents. It is a powerful oxidant

More information

Analysis of Vitamin C Using Iodine. Introduction

Analysis of Vitamin C Using Iodine. Introduction Analysis of Vitamin C Using Iodine Introduction Vitamin C (ascorbic acid) is oxidized to dehydroascorbic acid using a mild oxidizing agent such as iodine. The oxidation is a two- electron process, following

More information

Odor (Threshold Odor, Consistent Series)

Odor (Threshold Odor, Consistent Series) METHOD #: 140.1 (Issued 1971) TITLE: ANALYTE: INSTRUMENTATION: Odor (Threshold Odor, Consistent Series) Odor N/A STORET No. 60 C: 00086 Room Temp: 00085 1.0 Scope and Application 1.1 This method is applicable

More information

Reaction Stoichiometry and the Formation of a Metal Ion Complex

Reaction Stoichiometry and the Formation of a Metal Ion Complex Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry

More information

Phenolphthalein-NaOH Kinetics

Phenolphthalein-NaOH Kinetics Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.

More information

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration

More information

Solubility Product Constants

Solubility Product Constants Solubility Product Constants PURPOSE To measure the solubility product constant (K sp ) of copper (II) iodate, Cu(IO 3 ) 2. GOALS 1 To measure the molar solubility of a sparingly soluble salt in water.

More information

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure

More information

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy EXPERIMENT 4 Determination of Sodium by Flame Atomic-Emission Spectroscopy USE ONLY DEIONIZED WATER (NOT DISTILLED WATER!) THROUGHOUT THE ENTIRE EXPERIMENT Distilled water actually has too much sodium

More information

EXPERIMENT 10: TITRATION AND STANDARDIZATION

EXPERIMENT 10: TITRATION AND STANDARDIZATION EXPERIMENT 10: TITRATION AND STANDARDIZATION PURPOSE To determine the molarity of a NaOH solution by titrating it with a standard HCl solution. To determine the molarity of acetic acid in vinegar using

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

18 Conductometric Titration

18 Conductometric Titration Lab Activity 18 CONDUCTOMETRIC TITRATION LAB ACTIVITY 18 Conductometric Titration Background Titration is the a method of determining the concentration of an unknown solution (the analyte) by reacting

More information

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration GOAL AND OVERVIEW Antacids are bases that react stoichiometrically with acid. The number of moles of acid that

More information

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For NANYANG TECHNOLOGICAL UNIVERSITY School of Civil and Environmental Engineering CV2701: Laboratory 2A Laboratory Manual For Experiment Lab 2A-6(ENV) Wastewater Quality Analysis Location : Environment Laboratory,

More information

Determination of calcium by Standardized EDTA Solution

Determination of calcium by Standardized EDTA Solution Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic

More information

Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared!

Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared! Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared! Introduction Vitamin C (also known as ascorbic acid, HC6H7O6) is a necessary ingredient

More information

GA/7 Potentiometric Titration

GA/7 Potentiometric Titration Rev. 7/99 7-1 INTRODUCTION GA/7 Potentiometric Titration The potentiometric titration is a useful means of characterizing an acid. The ph of a solution is measured as a function of the amount of titrant

More information

Colorimetric Determination of Iron in Vitamin Tablets

Colorimetric Determination of Iron in Vitamin Tablets Colorimetric Determination of Iron in Vitamin Tablets Big Picture Conceptual Approach Vitamin Tablet How much Fe? ph = 3.5 Vitamin Tablet How much Fe? Too difficult to eyeball so will have the colorimeter

More information

NITRIC OXIDE and NITROGEN DIOXIDE 6014

NITRIC OXIDE and NITROGEN DIOXIDE 6014 NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm

More information

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3

More information

9. Analysis of an Acid-Base Titration Curve: The Gran Plot

9. Analysis of an Acid-Base Titration Curve: The Gran Plot 9. Analysis of an Acid-Base Titration Curve: The Gran Plot In this experiment, you will titrate a sample of pure potassium hydrogen phthalate (Table 10-4) with standard NaOH. A Gran plot will be used to

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

Keeping Your Swimming Pool and Spa Healthy Environmental Health Guide

Keeping Your Swimming Pool and Spa Healthy Environmental Health Guide Keeping Your Swimming Pool and Spa Healthy Environmental Health Guide A swimming pool or spa can be an asset to your home, and provide the opportunity to take part in healthy relaxing activities. However,

More information

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of total nitrogen in food and crude protein calculation (Kjeldahl method) Responsible person: Assoc.Prof. Ing.Kateřina Riddellová,

More information

BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use.

BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. USP proposes the revision and development of a suite of plastic packaging system standards in the current issue of PF. General test chapter

More information

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS

5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS 5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2

More information

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l.

1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l. Luminescence Measurement of Dissolved Oxygen in Water and Wastewater and for Use in the Determination of BOD 5 and cbod 5 Revision 1.2 October 2011 1. Scope and Application 1.1 This method is for the measurement

More information

THE ACTIVITY OF LACTASE

THE ACTIVITY OF LACTASE THE ACTIVITY OF LACTASE Lab VIS-8 From Juniata College Science in Motion Enzymes are protein molecules which act to catalyze the chemical reactions in living things. These chemical reactions make up the

More information

Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material

Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division

More information

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Parameter and code: Phosphorus, total-in-bottom-material, dry weight, I-6600-88 (mg/kg as P): 00668 1. Application This method is used

More information

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets

Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets Chemistry 119: Experiment 7 Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets Vitamin C is another name for ascorbic acid (C 6 H 8 O 6, see below ), a weak acid that can be determined by titration

More information

Lab #10 How much Acetic Acid (%) is in Vinegar?

Lab #10 How much Acetic Acid (%) is in Vinegar? Lab #10 How much Acetic Acid (%) is in Vinegar? SAMPLE CALCULATIONS NEED TO BE DONE BEFORE LAB MEETS!!!! Purpose: You will determine the amount of acetic acid in white vinegar (sold in grocery stores)

More information

Experiment 6 Titration II Acid Dissociation Constant

Experiment 6 Titration II Acid Dissociation Constant 6-1 Experiment 6 Titration II Acid Dissociation Constant Introduction: An acid/base titration can be monitored with an indicator or with a ph meter. In either case, the goal is to determine the equivalence

More information

FAJANS DETERMINATION OF CHLORIDE

FAJANS DETERMINATION OF CHLORIDE EXPERIMENT 3 FAJANS DETERMINATION OF CHLORIDE Silver chloride is very insoluble in water. Addition of AgNO 3 to a solution containing chloride ions results in formation of a finely divided white precipitate

More information

SenTix 950 SenTix 980

SenTix 950 SenTix 980 SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the

More information

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES

4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES 4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES Sl. No. Contents Preamble 4.1 Aim 4.2 Introduction 4.2.1 Environmental Significance 4.3 Principle 4.4 Materials Required 4.4.1 Apparatus Required 4.4.2 Chemicals

More information

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL 1 of 8 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.

More information

The Influence of Carbon Dioxide on Algae Growth

The Influence of Carbon Dioxide on Algae Growth The Influence of Carbon Dioxide on Algae Growth The first objective of this experiment is to show that increased atmospheric concentrations of carbon dioxide, CO 2, can stimulate algae growth. The second

More information